Test Report

FCC ID: F6O-INBODY-ON2

Test Report

Download: PDF
FCCID_4260965

                                                                Page 1 of 29                         Report No.      : OT-193-RWD-010




        ELECTROMAGNETIC EMISSION
           COMPLIANCE REPORT
 FOR LOW-POWER, NON-LICENSED TRANSMITTER

      Test Report No.              : OT-193-RWD-010

      AGR No.                      : A188A-224

      Applicant                    : InBody Co., Ltd.

      Address                      : InBody Bldg., 54, Nonhyeon-ro 2-gil, Gangnam-gu, Seoul, 135-960, Korea

      Manufacturer                 : InBody Co., Ltd.

      Address                      : 15, Heugam-gil, Ipjang-myeon, Seobuk-gu, Cheonan-si, Chungcheongnam-do, 31025,
                                     KOREA

      Type of Equipment            : Body Composition Analyzer

      FCC ID.                      : F6O-INBODY-ON2

      Model Name                   : InBodyON2

      Multiple Model Name : N/A

      Serial number                : N/A

      Total page of Report : 29 pages (including this page)

      Date of Incoming             : February 01, 2019

      Date of issue                : March 07, 2019



      SUMMARY
      The equipment complies with the regulation; FCC PART 15 SUBPART C Section 15.247
      This test report only contains the result of a single test of the sample supplied for the examination.
      It is not a generally valid assessment of the features of the respective products of the mass-production.




        Reviewed by:                                                            Approved by:
                             Ki-Hong, Nam / Asst, Chief Engineer                                 Keun-Young, Choi / Vice President
                             ONETECH Corp.                                                       ONETECH Corp.



It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


onETECH
        Ns                                                                  Page 2 of 29                                      Report No.           _: OT—193—RWD—010


                                                                            CONTENTS

                                                                                                                                                                        PAGE

1. VERIFICATION OF COMPLIANCE                                                                                                                                                   5

2. TEST SUMMARY                                                                                                                                                                 6

  2.1 TEST ITEMS AND RESULTS as2.00.00.00.00000000 00e en en nnennernne en en en en en en en en en en en en nnennennennnnennennennnnennennennnnnnennnnennennennen 6

  2.2 ADDITIONS, DEVIATIONS, EXCLUSIONS FROM STANDARDS.............00.00n0000000000 en en eneeneeneeneeneeneeneeneeneeneencencencencencencencencen 6

  2.3 RELATED SUBMITTAL(S) / (GRANT(S) .. 22.22.00.000.00.00000000000000en0en0ennnnnne en en en en en en en en en en en en en en en en en en en eneenennennennenne 6

  2.4 PURPOSE OF THE TEST ..........00.00.0000000e00 en ene en nnennenne en en nnenne en en nne en en en nne en nnennennnnennnnennennennnnennennennnnnnennnnennennennen 6

  2.5 TEST METHODOLOGY.22.20.00.00000000000 00 n0en0ennnnnnr en en en ennr en en en en en en en en en en en en en en en en en en nnennnne en en nnennennennennennennennen 6

  2.6 TEST FACILATY .22 00.00e00 00 en en en nnr en en en en en en en en nne en nnennenne en en en en en nnenne en en nnennnnennnne en nnennnnenne en nnennennennennennennennen 6

3. GENERAL INFORMATION                                                                                                                                                          7

  3.1 PRODUCT DESCRIPTION.:.22.22.00.00.0000000000000en0en0ennennennne n en en en en en en en en en en en en en en en nne en en nnenne en nne en en nnenne en en nnennenneene 7

  3.2 ALTERNATIVE TYPE(S)/MODEL(S); ALSO COVERED BY THIS TEST REPORT..................0000000000000e0en en enmenmenmenmnmnmenme 7

4. EUT MODIFICATIONS                                                                                                                                                            7

5. SYSTEM TEST CONFIGURATION                                                                                                                                                    8

  5.1 JUSTIFICATION .. 22. 00. 0000000000 en en n nne en en en en en nnene n nne en nne en en nne en en nne en nne en nnnne en nnennennnnennennennennnnennennennennennennenneene 8

  5.2 PERIPHERAL EQUIPMENT a2.22.22.00.00.00.0000000000000000en0en0 en ennnnnnnennnnnnnnnnnnennennennennennennenneananeaneaneananeaneaneaneaneanenneanennennenne 8

  5.3 MODE OF OPERATION DURING THE TEST .........00.00.0000 0000000e00en0eneenennenneene en en en en en en en en en en en en nne en en en anennennennennennenne 8

  5.4 CONFIGURATION OF TEST SYSTEM ... 22. 00.00.00.0000000000000 e en nneene en en en en enrenr en en en en en en en en en en en en en en en en ane en en nnennennenne 8

6. PRELIMINARY TEST                                                                                                                                                             9

  6.1 AC POWER LINE CONDUCTED EMISSIONS TESTS.....2..2.00.00.00.0000000000000en0ennennenrenrenrenrenrenrenrenrenrenrenrenrenreneeneenennennennennenne 9
  6.2 GENERAL RADIATED EMISSIONS TESTS ... 2202000000000 0000000en0 en nneneenr en en enrenrenrenrenrenreneenrene en en en en en en en en en nnennennenne 9

7. MINIMUM 6 DB BANDWIDTH                                                                                                                                                     10

  7.1 OPERATING ENVIRONMENT 12. 20.00.00.00.00000000000000en0en0 en en en en en en en en en en en nne en nnennennennennnnnnennnnennnnennnnennennnnnnennennennennen 10

  T.2 TEST SETAUP 22000000 en ennnnnnennenne en en en en en en en en nnennen n nne en nne en nn en nnennennennennennnnennennnnnnennennennnnennnnennennnnnnennennennennen 10

  7.3 TEST EQUIPMENT USED ....22.00.00.00000 000000000000 00000000 en0 en en n en n n nnnnnnennennennennennennennennenneannnenneanennnneannnennennennennenne 10

  7.A TEST DATA .. .00 .00 0n n0 en en nnnnr en en en en en en en en en en nnennene n nnennenne en nnnnennennennennennennnnennennnnnnennennennnnennnnennennnnnnennennennennen 11

8. MAXIMUM PEAK OUTPUT POWER                                                                                                                                                  13

  8.1 OPERATING ENVIRONMENT ... 22. 00.00.00.0000000000000en0en0 en en rneene en en en en en en en en en en en en nne en nnennennennennnnennnnennennnnnnennennennennen 13

  8.2 TEST SETAUP 2220000000 e0 en en en en en en en en en en en en nnenernvnne en en en en en nnennenne en nne en nnennennnnennennnnennnnennnnennennnnnnennennennennen 13
  8.3 TEST EQUIPMENT USED ....22.22.00.00.00000000 000000000 000000 en0 en annnennnnennnnennnnnnnnennennennennennennennennenneannnenneaneannneannnennennennennenne 13

  8.4 TEST DATA....

9. 100 KHZ BANDWIDTH OUTSIDE THE FREQUENCY BAND                                                                                                                               16

  9.1 OPERATING ENVIRONMENT ... 20. 00.00.00.0000000000000en0 en en en rneenr en en en en en en en en en en en en en en nne en en nne en en nnennnnennnnnnnnennennennen 16

It should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                                                    EMC—003 (Rev.2)
ONETECH Corp.; 43—14, Jinsaegol—gil, Chowol—cup, Gwangju—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799—9599)


onETECH
         Ns                                                                   Page 3 of 29                                        Report No.            _: OT—193—RWD—010

  9.2 TEST SET—UP FOR CONDUCTED MEASUREMENT .....02.0.00.00.00.0000000000000en0 en en en nnennennennennennnnnnnnnnnnnnnnnnnnnnnnennenne 16

  9.3 TEST SET—UP FOR RADIATED MEASUREMENT::..2.00.00.00.00.0000000000000000en0 en en en en nnnnennnnennennennnnnnnnnnnnnnnnnnnnnnnnennenne 16

  9.4 TEST EQUIPMENT USED ... 22. 2000000000000 0000000000000 en0 en annnnnnnennnnennnnnnnnennennennennennennennennenneannnenneaneannneannnennennennennenne 16

  9.5 TEST DATA FOR CONDUCTED EMISSION .. ... 22. 00.00.00.0000 0000000 en en en nnnnennennennenne en en en en en nnnnnnnnnnnnnnnnnnnnnnnnennenne 17

  9.6 TEST DATA FOR RADIATED EMISSION                                                                                                                                            ..22

      9.6.1 Radiated EmisSiOn Which f@ll in the ReStriCLA BOMA.....................scccccviccierrerrensensenrensenrenrenrenrenrennennenrennenrencenne 22

      9.6.2 SPUTiOUS & HQPMORiG RAdiQb@d EIISSLON ...........0. 0s ons enseneennernernennennennennencencencencencencencencencencencencencencencencencencenccnne 23

10. PEAK POWER SPECTRAL DENSITY                                                                                                                                                    24

   10.1 OPERATING ENVIRONMENT .. 22. 00.00.00.00.00000000000000en0en0eneenenne en en en en en en en en en en en en ane en nnenneane en en en annneanennennennennennennenne 24

   10.2 TEST SETAUP o. 22. 0000000 en en en nnnne en en en en en en en en en nnenennenne en en en en en en nne en nne en nnennnnennennennennnne en nnennnnenne en en en nnennennenne 24

   10.3 TEST EQUIPMENT USED ... 22.22.00.00.00000000 000000000000 en en en nnnnnnnnennennennennennnnennennennenneannneaneaneaneaneaneaneane en en aneaneanennennenne 24

   10.4 TEST DATA...

11. RADIATED EMISSION TEST                                                                                                                                                         27

   11.A OPERATING ENVIRONMENT .. 22. 00.00.00.0000000000000000en0en0enennenne en en en en en en en en ane en en en anennennenneane en en en annneanennennennennennennenne 27

   1L.2 TEST SETAUP o. 20. 000000 en enennnnenne en nnr en en en en en en en nnenennenne en en en en nnennenne en nne en nnennnnennennennennnne en nnennnnennennenne en nnennennenne 27

   11.3 TEST EQUIPMENT USED ... 22. 02.00.00.00000000 000000000000 en en en nnnnnnnnennennennnnennnnennennennenneannneaneaneaneaneaneaneane en en aneaneanennennenne 27

   11.4 TEST DATA FOR TRANSMITTING MODE ..2.22.00.00.00.000.0000000000000en0en0rnerne en en en enrenrenrenrenrenrenrene en en en en en en en en en en nnennenne 28

      11.4.1 TeSt d@t@t fOr 30 MHZ ~ 1 GHZ .......c.0sueensenvevenmenrrnrrnrrnrrarrerrenrenrenranrnnrnnennennennennennennennennennennennvnnennennennennennencenne 28

      11.4.2 Test data for Below 30 MHz ....                                                                                                                                      .29

      11.4.3 TeSt d@tt fOF QbOVG 1 GHZ ...........csocsonsensenmensensevensensensrnnrnnrnnrnnrnnrnnennennennennennennennennennennennennennennennvnnennennennennennencene 29




It should not be reproduced exceptin full, withoutthe written approval of ONETECH Corp.                                                                          EMC—003 (Rev.2)
ONETECH Corp.; 43—14, Jinsaegol—gil, Chowol—cup, Gwangju—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799—9599)


                                                                Page 4 of 29                          Report No.     : OT-193-RWD-010




                                                             Revision History
 Rev. No.          Issue Report No.                     Issued Date                           Revisions                 Section Affected
     0            OT-193-RWD-010                      March 07, 2019                        Initial Release                     All




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 5 of 29                         Report No.      : OT-193-RWD-010


1. VERIFICATION OF COMPLIANCE
Applicant           : InBody Co., Ltd.
Address             : InBody Bldg., 54, Nonhyeon-ro 2-gil, Gangnam-gu, Seoul, 135-960, Korea
Contact Person : Dong-Hyun Woo / Quality Approval Team / Employee
Telephone No. : +82-2-2182-1836
FCC ID              : F6O-INBODY-ON2
Model Name          : InBodyON2


Brand Name          :
Serial Number : N/A
Date                : March 07, 2019

 EQUIPMENT CLASS                                            DTS – DIGITAL TRNSMISSION SYSTEM
 E.U.T. DESCRIPTION                                        Body Composition Analyzer
 THIS REPORT CONCERNS                                       Original Grant
 MEASUREMENT PROCEDURES                                     ANSI C63.10: 2013
 TYPE OF EQUIPMENT TESTED                                   Pre-Production
 KIND OF EQUIPMENT
                                                            Certification
 AUTHORIZATION REQUESTED
 EQUIPMENT WILL BE OPERATED
                                                            FCC PART 15 SUBPART C Section 15.247
 UNDER FCC RULES PART(S)
 Modifications on the Equipment to            Achieve
                                                            None
 Compliance
 Final Test was Conducted On                                3 m, Semi Anechoic Chamber

-. The above equipment was tested by ONETECH Corp. for compliance with the requirement set forth in the FCC Rules
   and Regulations. This said equipment in the configuration described in this report, shows the maximum emission levels
   emanating from equipment are within the compliance requirements.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 6 of 29                         Report No.      : OT-193-RWD-010


2. TEST SUMMARY

2.1 Test items and results
      SECTION                                                 TEST ITEMS                                               RESULTS

  15.247 (a) (2)           Minimum 6 dB Bandwidth                                                                 Met the Limit / PASS
  15.247 (b) (3)           Maximum Peak Conducted Output Power                                                    Met the Limit / PASS
  15.247 (d)               100 kHz Bandwidth Outside the Frequency Band                                           Met the Limit / PASS
  15.247 (d)               Radiated Emission which fall in the Restricted Band                                    Met the Limit / PASS
  15.247 (e)               Peak Power Spectral Density                                                            Met the Limit / PASS
  15.209                   Radiated Emission Limits                                                               Met the Limit / PASS
  15.207                   Conducted Limits                                                                           N/A (See Note)
  15.203                   Antenna Requirement                                                                   Met requirement / PASS
Note: This test is not performed because the EUT is operated by DC battery.

2.2 Additions, deviations, exclusions from standards
No additions, deviations or exclusions have been made from standard.

2.3 Related Submittal(s) / Grant(s)
Original submittal only

2.4 Purpose of the test
To determine whether the equipment under test fulfills the requirements of the regulation stated in FCC PART 15
SUBPART C Section 15.247.

2.5 Test Methodology
Both conducted and radiated testing was performed according to the procedures in ANSI C63.10: 2013. Radiated testing
was performed at a distance of 3 m from EUT to the antenna.

2.6 Test Facility
The Onetech Corp. has been designated to perform equipment testing in compliance with ISO/IEC 17025.
The Electromagnetic compatibility measurement facilities are located at 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si,
Gyeonggi-do, 12735, Korea
-. Site Filing:
VCCI (Voluntary Control Council for Interference) – Registration No. R-4112/ C-14617/ G-10666 / T-1842
IC (Industry Canada) – Registration No. Site# 3736A-3
-. Site Accreditation:
KOLAS (Korea Laboratory Accreditation Scheme) - Accreditation NO. KT085
FCC (Federal Communications Commission) - Accreditation No. KR0013
RRA (Radio Research Agency) – Designation No. KR0013


It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 7 of 29                         Report No.      : OT-193-RWD-010


3. GENERAL INFORMATION

3.1 Product Description
The InBody Co., Ltd., Model InBodyON2 (referred to as the EUT in this report) is a Body Composition Analyzer. The
product specification described herein was obtained from product data sheet or user’s manual.

  Device Type                             Body Composition Analyzer
  Temperature Range                       10 °C ~ 40 °C
  Operating Frequency                     2 402 MHz ~ 2 480 MHz
  RF Output Power                         -11.72 dBm
  Number of Channel                       40 Channel
  Modulation Type                         GFSK (Bluetooth LE)
  Antenna Type                            Chip Antenna
  Antenna Gain                            1.99 dBi
  List of each Osc. or crystal
                                          16 MHz
  Freq.(Freq. >= 1 MHz)

3.2 Alternative type(s)/model(s); also covered by this test report.
-. None


4. EUT MODIFICATIONS
-. None




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 8 of 29                         Report No.      : OT-193-RWD-010


5. SYSTEM TEST CONFIGURATION

5.1 Justification
This device was configured for testing in a typical way as a normal customer is supposed to be used. During the test, the
following components were installed inside of the EUT.
      DEVICE TYPE                         MANUFACTURER                             MODEL/PART NUMBER                         FCC ID

         Main Board                          InBody Co., Ltd.                        INBODY_H20N_E2417                          N/A
          Sub Board                          InBody Co., Ltd.                          INBODY_H20 B2213                         N/A

5.2 Peripheral equipment
Defined as equipment needed for correct operation of the EUT, but not considered as tested: None

5.3 Mode of operation during the test
For the testing, software used to control the EUT for staying in continuous transmitting is programmed.
For final testing, the EUT was set at 2 402 MHz, 2 440 MHz, and 2 480 MHz to get a maximum emission levels from the
EUT. The EUT was moved throughout the XY, XZ, and YZ planes and the worst case is “XZ” axis, but the worst data was
recorded in this report.

5.4 Configuration of Test System
Line Conducted Test:                     It is not need to test this requirement, because the EUT shall be operated by DC battery.


Radiated Emission Test:                  Preliminary radiated emissions test were conducted using the procedure in ANSI C63.10:
                                         2013 to determine the worse operating conditions. Final radiated emission tests were
                                         conducted at 3 meter Semi Anechoic Chamber.
                                         The turntable was rotated through 360 degrees and the EUT was tested by positioned
                                         three orthogonal planes to obtain the highest reading on the field strength meter. Once
                                         maximum reading was determined, the search antenna was raised and lowered in both
                                         vertical and horizontal polarization.


5.5 Antenna Requirement
For intentional device, according to section 15.203, an intentional radiator shall be designed to ensure that no antenna other
than that furnished by the responsible party shall be used with the device.
Antenna Construction:
The antenna of the EUT is a Chip Antenna on the main board in the EUT, so no consideration of replacement by the user.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 9 of 29                         Report No.      : OT-193-RWD-010


6. PRELIMINARY TEST

6.1 AC Power line Conducted Emissions Tests
During Preliminary Tests, the following operating mode was investigated

                        Operation Mode                                    The Worse operating condition (Please check one only)

                    It is not need to test this requirement, because the power of the EUT is supplied by battery.

6.2 General Radiated Emissions Tests
During Preliminary Tests, the following operating modes were investigated

                        Operation Mode                                    The Worse operating condition (Please check one only)

                       Transmitting Mode                                                               X




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 10 of 29                           Report No.   : OT-193-RWD-010


7. MINIMUM 6 dB BANDWIDTH

7.1 Operating environment
Temperature                          :       24.3 °C
Relative humidity                    :       43.9 % R.H.

7.2 Test set-up
The antenna output of the EUT was connected to the spectrum analyzer. The resolution bandwidth is set to 100 kHz, and
peak detection was used. The 6 dB bandwidth is defined as the total spectrum over which the power is higher than the peak
power minus 6 dB.




                             EUT                                                            Spectrum analyzer



7.3 Test equipment used
       Model Number               Manufacturer                 Description                        Serial Number      Last Cal.
-     FSV40                      Rohde & Schwarz              Signal Analyzer                    101009             Mar. 14, 2018 (1Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 11 of 29                        Report No.      : OT-193-RWD-010


7.4 Test data
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Test Result                  : Pass

 CHANNEL            FREQUENCY(MHz)                  MEASURED VALUE (kHz)                    LIMIT (kHz)               MARGIN (kHz)
      Low                   2 402.00                             689.30                        500.00                       189.30
    Middle                  2 440.00                             694.30                        500.00                       194.30
     High                   2 480.00                             694.30                        500.00                       194.30
Remark. Margin = Measured Value - Limit




                                                                                              Tested by: Tae-Ho, Kim / Manager




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 12 of 29                        Report No.      : OT-193-RWD-010




                                                                 Middle Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


onETECH
       Ns                                                      Page 13 of 29                         Report No.      _: OT—193—RWD—010

8. MAXIMUM PEAK OUTPUT POWER

8.1 Operating environment
Temperature                          :      24.3 °C

Relative humidity                    :      43.9 % RH.

8.2 Test set—up
The antenna output of the EUT was connected to the spectrum analyzer.
The resolution bandwidth is set to 2 DTS Bandwidth, the video bandwidth is set to 3 times the resolution bandwidth.




                               EUT                                                         Spectrum Analyzer




8.3 Test equipment used
      Model Number                Manufacturer                Description                       Serial Number        Last Cal.

m—    FSV40                       Rohde & Schwarz              Signal Analyzer                  101009               Mar. 14, 2018 (1Y)

All test equipment used is calibrated on a regularbasis.




It should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                    EMC—003 (Rev.2)
ONETECH Corp.; 43—14, Jinsaegol—gil, Chowol—cup, Gwangju—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799—9599)


                                                                Page 14 of 29                        Report No.      : OT-193-RWD-010


8.4 Test data
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Test Result                  : Pass

                          FREQUENCY                   MEASURED VALUE                        LIMIT                       MARGIN
  CHANNEL
                               (MHz)                           (dBm)                        (dBm)                          (dB)
     LOW                      2 402.00                         -14.15                       30.00                          44.15
   MIDDLE                     2 440.00                         -12.92                       30.00                          42.92
     HIGH                     2 480.00                         -11.72                       30.00                          41.72
Remark. Margin = Limit – Measured Value (=Receiver Reading + Cable Loss)




                                                                                              Tested by: Tae-Ho, Kim / Manager




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 15 of 29                        Report No.      : OT-193-RWD-010




                                                                Middle Channel




                                                                  High Channel



It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 16 of 29                            Report No.   : OT-193-RWD-010


9. 100 kHz BANDWIDTH OUTSIDE THE FREQUENCY BAND

9.1 Operating environment
Temperature                          :       24.3 °C
Relative humidity                    :       43.9 % R.H.


9.2 Test set-up for conducted measurement
The antenna output of the EUT was connected to the spectrum analyzer. The resolution bandwidth is set to 100 kHz, the
video bandwidth is set to 3 times the resolution bandwidth and peak detection was used.



                             EUT                                                            Spectrum analyzer



9.3 Test set-up for radiated measurement
The radiated emissions measurements were performed on the 3 m semi anechoic chamber. The EUT was placed on
turntable approximately 1.5 m above the ground plane.

The frequency spectrum from 30 MHz to 26.5 GHz was scanned and maximum emission levels at each frequency recorded.
The system was rotated 360°, and the antenna was varied in the height between 1.0 m and 4.0 m in order to determine the
maximum emission levels. This procedure was performed for horizontal and vertical polarization of the receiving antenna.


9.4 Test equipment used
       Model Number              Manufacturer               Description                            Serial Number      Last Cal.
-     FSV40                     Rohde & Schwarz            Signal Analyzer                        101009             Mar. 14, 2018 (1Y)
-     ESU                       Rohde & Schwarz            EMI Test Receiver                      100261             Mar. 29, 2018 (1Y)
-     310N                      Sonoma Instrument          Pre-Amplifier                          312544             Mar. 28, 2018 (1Y)
-     BBV9718                   Schwarzbeck                Amplifier                              310                Mar. 30, 2018 (1Y)
      SCU40A                    Rohde & Schwarz            Signal Conditioning unit               100436             Mar. 15, 2018 (1Y)
-     DT3000-3t                 Innco System               Turn Table                             DT3000/093         N/A
-     MA-4000XPET               Innco System               Antenna Master                         MA4000/509         N/A
-     VULB9163                  Schwarzbeck                TRILOG Broadband Antenna               9163-419           Aug. 09, 2018 (2Y)
-     BBHA9120D                 Schwarzbeck                Horn Antenna                           BBHA9120D295 Aug. 16, 2017 (2Y)
-     BBHA9170                  Schwarzbeck                Horn Antenna                           BBHA9170179        Jul. 28, 2017 (2Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 17 of 29                        Report No.      : OT-193-RWD-010


9.5 Test data for conducted emission




                                                                  Low Channel




                                                                Middle Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 18 of 29                        Report No.      : OT-193-RWD-010




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 19 of 29                        Report No.      : OT-193-RWD-010




                                                                  Low Channel




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 20 of 29                        Report No.      : OT-193-RWD-010




                                                                Middle Channel




                                                                Middle Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 21 of 29                        Report No.      : OT-193-RWD-010




                                                                  High Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 22 of 29                        Report No.      : OT-193-RWD-010


9.6 Test data for radiated emission

9.6.1 Radiated Emission which fall in the Restricted Band
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Resolution bandwidth         : 1 MHz and Peak Detector for Peak Mode
                                  1 MHz and RMS Detector for Average Mode
-. Video bandwidth              : 3 MHz for Peak and Average Mode
-. Measurement distance         :3m
-. Result                       : PASSED

Frequency        Reading         Detector        Ant. Pol.         Ant.         Cable       Amp        Total         Limits         Margin
   (GHz)         (dBμV)           Mode             (H/V)          Factor         Loss       Gain    (dBμV/m)       (dBμV/m)          (dB)

                                                          Test Data for Low Channel

  2 361.19         50.85           Peak               H            26.94         9.20       34.76      52.23          74.00         21.77
  2 360.87         35.41         Average              H            26.91         9.17       34.72      36.77          54.00         17.23
  2 360.71         46.44           Peak               V            26.91         9.17       34.72      47.80          74.00         26.20
  2 349.20         33.20         Average              V            26.91         9.17       34.72      34.56          54.00         19.44

                                                          Test Data for High Channel

  2 483.51         53.54           Peak               H            27.47         9.49       35.51      54.99          74.00         19.01
  2 483.51         37.15         Average              H            27.47         9.49       35.51      38.60          54.00         15.40
  2 483.56         49.70           Peak               V            27.48         9.49       35.52      51.15          74.00         22.85
  2 483.51         34.61         Average              V            27.47         9.49       35.51      36.06          54.00         17.94

                                                   Tabulated test data for Restricted Band
Remark: “H”: Horizontal, “V”: Vertical
             Margin (dB) = Limits (dBμV/m) - Total Level (dBμV/m)
             Total Level = Reading + Antenna Factor + Cable Loss – Pre-Amplifier Gain




                                                                                               Tested by: Tae-Ho, Kim / Manager




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 23 of 29                        Report No.      : OT-193-RWD-010


9.6.2 Spurious & Harmonic Radiated Emission
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Resolution bandwidth         : 1 MHz and Peak Detector for Peak Mode for the emissions fall in restricted band,
                                  1 MHz and RMS Detector for Average Mode for the emissions fall in restricted band
                                 100 kHz for Peak Mode for the emissions outside restricted band
-. Video bandwidth              : 3 MHz for Peak and Average Mode
-. Frequency range              : 1 GHz ~ 26.5 GHz
-. Measurement distance         :3m
-. Result                       : PASSED

 Frequency         Reading        Detector Ant. Pol.            Ant.       Cable       Amp       Total           Limits             Margin
    (GHz)           (dBμV)         Mode          (H/V)        Factor        Loss       Gain    (dBμV/m)        (dBμV/m)              (dB)
                                                         Test Data for Low Channel

                     44.84          Peak            H                                            52.25            74.00             21.75
                     35.17        Average           H                                            42.58            54.00             11.42
  4 804.00                                                     30.84       12.31       35.74
                     45.03          Peak            V                                            52.44            74.00             21.56
                     34.86        Average           V                                            42.27            54.00             11.73
                                                        Test Data for Middle Channel

                     44.50          Peak            H                                            51.14            74.00             22.86
                     34.77        Average           H                                            41.41            54.00             12.59
  4 880.00                                                     30.01       12.43       35.80
                     44.92          Peak            V                                            51.56            74.00             22.44
                     35.03        Average           V                                            41.67            54.00             12.33
                                                         Test Data for High Channel
                     44.79          Peak            H                                            52.79            74.00             21.21
                     35.03        Average           H                                            43.03            54.00             10.97
  4 960.00                                                    31.15       12.81       35.96
                     45.18          Peak            V                                            53.18            74.00             20.82
                     35.56        Average           V                                            43.56            54.00             10.44
                                                   Tabulated test data for Restricted Band
Remark: “H”: Horizontal, “V”: Vertical
            Margin (dB) = Limits (dBμV/m) - Total Level (dBμV/m)
            Total Level = Reading + Antenna Factor + Cable Loss – Pre-Amplifier Gain




                                                                                               Tested by: Tae-Ho, Kim / Manager

It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


onETECH
       Ns                                                      Page 24 of 29                           Report No.    _: OT—193—RWD—010

10. PEAK POWER SPECTRAL DENSITY

10.1 Operating environment
Temperature                          :      24.3 °C

Relative humidity                    :      43.9 % RH.

10.2 Test set—up
The antenna output of the EUT was connected to the spectrum analyzer.
The resolution bandwidth is set to 3 kHz S RBW <100 kHz, the video bandwidth is set to 3 times the resolution

bandwidth.




                             EUT                                                           Spectrum analyzer



10.3 Test equipment used
      Model Number                Manufacturer                Description                        Serial Number        Last Cal.

m—    FSV40                       Rohde & Schwarz              Signal Analyzer                   101009               Mar. 14, 2018 (1Y)

All test equipment used is calibrated on a regularbasis.




It should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                    EMC—003 (Rev.2)
ONETECH Corp.; 43—14, Jinsaegol—gil, Chowol—cup, Gwangju—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799—9599)


                                                                Page 25 of 29                        Report No.      : OT-193-RWD-010


10.4 Test data
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Test Result                  : Pass
-. Operating Condition          : Continuous transmitting mode
  CHANNEL             FREQUENCY(MHz)                   MEASURED VALUE (dBm)                      LIMIT (dBm)              MARGIN (dB)
      Low                     2 402.00                               -25.28                            8.00                    33.28
    Middle                    2 440.00                               -24.15                            8.00                    32.15
      High                    2 480.00                               -22.70                            8.00                    30.70
Remark. Margin = Limit – Measured value




                                                                                              Tested by: Tae-Ho, Kim / Manager




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 26 of 29                        Report No.      : OT-193-RWD-010




                                                                Middle Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 27 of 29                        Report No.      : OT-193-RWD-010


11. RADIATED EMISSION TEST

11.1 Operating environment
Temperature                          :       24.3 °C
Relative humidity                    :       43.9 % R.H.

11.2 Test set-up
The radiated emissions measurements were on the 3 m semi anechoic chamber. The EUT and other support equipment were
placed on a non-conductive turntable above the ground plane. The interconnecting cables from outside test site were
inserted into ferrite clamps at the point where the cables reach the turntable.
The frequency spectrum from 30 MHz to 26.5 GHz was scanned and emission levels maximized at each frequency recorded.
The system was rotated 360°, and the antenna was varied in height between 1.0 m and 4.0 m in order to determine the
maximum emission levels. This procedure was performed for both horizontal and vertical polarization of the receiving
antenna.


11.3 Test equipment used
       Model Number              Manufacturer               Description                        Serial Number         Last Cal.
-     FSV40                     Rohde & Schwarz            Signal Analyzer                    101009                Mar. 14, 2018 (1Y)
-     ESU                       Rohde & Schwarz            EMI Test Receiver                  100261                Mar. 29, 2018 (1Y)
-     310N                      Sonoma Instrument          Pre-Amplifier                      312544                Mar. 28, 2018 (1Y)
-     BBV9718                   Schwarzbeck                Amplifier                          310                   Mar. 30, 2018 (1Y)
-     DT3000-3t                 Innco System               Turn Table                         DT3000/093            N/A
-     MA-4000XPET               Innco System               Antenna Master                     MA4000/509            N/A
-     VULB9163                  Schwarzbeck                TRILOG Broadband Antenna           9163-419              Aug. 09, 2018 (2Y)
-     BBHA9120D                 Schwarzbeck                Horn Antenna                       BBHA9120D295 Aug. 16, 2017 (2Y)
-     BBHA9170                  Schwarzbeck                Horn Antenna                       BBHA9170179           Jul. 28, 2017 (2Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 28 of 29                        Report No.      : OT-193-RWD-010


11.4 Test data for Transmitting Mode

11.4.1 Test data for 30 MHz ~ 1 GHz
Humidity Level                    : 43.9 % R.H.                                                                      Temperature: 23.9 °C
Limits apply to                   : FCC CFR 47, PART 15, SUBPART C, SECTION 15.247
Result                            : PASSED

EUT                               : Body Composition Analyzer                               Date: February 07, 2019 ~ February 12, 2019
Detector                          : CISPR Quasi-Peak (6 dB Bandwidth: 120 kHz)




                                                                                                Tested by: Tae-Ho, Kim / Manager

It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 29 of 29                        Report No.      : OT-193-RWD-010


11.4.2 Test data for Below 30 MHz
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Resolution bandwidth         : 200 Hz (from 9 kHz to 0.15 MHz), 9 kHz (from 0.15 MHz to 30 MHz)
-. Frequency range              : 9 kHz ~ 30 MHz
-. Measurement distance         :3m
-.Operating mode                : Transmitting mode

Frequency Reading Ant. Pol. Ant.                            Angle     Ant. Factor       Cable      Emission     Limits              Margin
  (MHz)   (dBμV)   (H/V) Height (m)                          (°)       (dB/m)           Loss    Level(dBμV/m) (dBμV/m)               (dB)



                                            It was not observed any emissions from the EUT.




11.4.3 Test data for above 1 GHz
-. Test Date                    : February 07, 2019 ~ February 12, 2019
-. Resolution bandwidth         : 1 MHz for Peak and Average Mode
-. Video bandwidth              : 1 MHz for Peak Mode, 10 Hz for Average Mode
-. Frequency range              : 1 GHz ~ 26.5 GHz
-. Measurement distance         :3m
-.Operating mode                : Transmitting mode

Frequency Reading Ant. Pol. Ant.                            Angle     Ant. Factor       Cable      Emission     Limits              Margin
  (MHz)   (dBμV)   (H/V) Height (m)                          (°)       (dB/m)           Loss    Level(dBμV/m) (dBμV/m)               (dB)



                                            It was not observed any emissions from the EUT.




                                                                                                Tested by: Tae-Ho, Kim / Manager




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)



Document Created: 2019-03-07 14:10:17
Document Modified: 2019-03-07 14:10:17