SAR APPENDIX C3

FCC ID: EJE-WL0029

RF Exposure Info

Download: PDF
FCCID_1793730

DASY5 Validation Report for Body

                                                                        Date/Time: 09.12.2010 13:23:23

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:724

Communication System: CW; Frequency: 2450 MHz; Duty Cycle: 1: 1
Medium: MSL U12 BB
Medium parameters used: £ = 2450 MHz; 0 = 1.92 mho/m; &= 52.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASYS5 Configuration:

        Probe: ES3DV3 — SN3205; ConvF(4.31, 4.31, 4.31); Calibrated: 30.04.2010

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 10.06.2010

        Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

        Measurement SW: DASY52, V52.2 Build 0, Version 52.2.0 (163)

        Postprocessing SW: SEMCAD X, V14.2 Build 2, Version 14.2.2 (1685)


Pin=250 mW /d=10mm, dist=3.0mm (ES—Probe)/Zoom Scan (7x7x7) /Cube 0: Measurement
grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 97.3 V/m; Power Drift = —0.059 dB
Peak SAR (extrapolated) = 27.3 W/kg
SAR(1 g) = 13 mW/g; SAR(10 g) = 6.04 mW/g
Maximum value of SAR (measured) = 17.2 mW/g




                         48

                                    0 dB = 17.2mW/g

Certificate No: D2450V2—724_Dec10               Page 5 of 6


Impedance Measurement Plot for Body TSL




                                                                9 Dec 2010    10:54:44
     CHI sii     4 uo ors           3: 48.301 a   4.3281 o 28116 pH          2 450.000 000 MHz



     *
     De1


     Ca
                                                     y



     fivg
     16

         +

     CH2




     Ca




     Av
     16°


         1

                TART 2 250.000   0B MHz




Certificate No: D2450V2—724_Dec10                        Page 6 of 6


                                                                           28 June 2012

  CRA sii       i.een us       41 49.400 a   S.4144 0. 354.70 pH          2450.000 000 nz
        Ap                                                                                            HoLD



  Del

                                                                                                SINGLE
  Cor



  PRa                                                                                        NUNBER of
                                                                                                srours
  Av
  159

                                                                                            conTINuous
  Hid


  CH2

                                                                                              TRIGGER:
                                                                                              Trie orr

  Ser                                                                                         ext tere
                                                                                              O SHEEP


  PRm                                                                                               I
  m                                                                                           EXT TRIG
   v
  16°                                                                                         on Poxlrn-

  Hid                                                                                       naMUAL.    TRe
                                                                                              on PorhT
             sTaRt 2 250.000 000 MHz                               stor 2 sse.a0e 00e hiz
p




NATA


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                              Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client             EMC Technologies                                                              Certificate No: DAE3—442_Dec11


[CALIBRATION CERTIFICATE
 Object                                DAE3 — SD 000 DO3 AE — SN: 442


 Calibration procedure(s)              QA CAL—06.v23
                                       Calibration procedure for the data acquisition electronics (DAE)



 Calibration date:                     December 5, 2011




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               28—Sep—11 (No:11450)                                 Sep—12


 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Calibrator Box V1.1                    SE UMS 006 AB 1004        08—Jun—11 (in house check)                           In house check: Jun—12




                                        Name                                   Function                                 Signature   _
 Calibrated by:                         Andrea Guntli                          Technician                                     j’ %
                                                                                                                    _Aetlll
 Approved by:                           Fin Bomholt                             R&D Director
                                                                                                                        L a¥ IQQ (QMAU

                                                                                                                        Issued: December 5, 2011
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE3—442_Dec11                                        Page 1 of 5




Auresutce No: sn


Calibration Laboratory of                                    ‘\\\\‘&'@>       S   Schweizerischer Kalibrierdienst
                                                             S mz
Schmid & Partner                                            i'E:\_/!53:       c   Service suisse d‘étalonnage
  Engineering AG                                            yzz— o.               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 /'/,l///_\\\ w   S   Swiss Calibration Service
                                                                thitabs®"

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 108
The Swiss Accreditation Service Is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates



Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
     comparison with a calibrated instrument traceable to national standards. The figure given
     corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    *    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         e    DC Voltage Measurement Linearity: Veritication of the Linearity at +10% and —10% of
              the nominal calibration voltage. Influence of offset voltage is included in this
              measurement.
         e    Common mode sensitivity: Influence of a positive or negative common mode voltage on
              the differential measurement.
         e    Channel separation: Influence of a voltage on the neighbor channels not subject to an
              input voltage.
         e    AD Converter Values with inputs shorted: Values on the internal AD converter
              corresponding to zero input voltage
        e     Input Offset Measurement. Output voltage and statistical results over a large number of
              zero voltage measurements.
        *     Input Offset Current: Typical value for information; Maximum channel input offset
              current, not considering the input resistance.
        e     Input resistance: Typical value for information: DAE input resistance at the connector,
              during internal auto—zeroing and during measurement.
        e     Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
              alarm signal is generated.
        +     Power consumption: Typical value for information. Supply currents in various operating
              modes.




Certificate No: DAE3—442_Dec11                                  Page 2 of 5


DC Voltage Measurement
      A/D — Converter Resolution nominal
          High Range:          1LSB =     6.AuV ,       full range =  ~100...+300 mV
          Low Range:           1LSB=       61nV ,       full range =  Bossiesé +3mV
      DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

         Calibration Factors              X                           Y                   Z
         High Range              404.367 + 0.1%(k=2) 405.009 + 0.1% (k=2)       405.229 + 0.1% (k=2)
         Low Range               3.98363 £ 0.7% (k=2)     3.98114 £0.7% (k=2)   3.98948 + 0.7% (k=2)

Connector Angle

         Connector Angle to be used in DASY system                                     §7.0°%1"




Certificate No: DAE3—442_Dec11                          Page 3 of 5




Acrestie No se


Appendix

1. DC Voltage Linearity
      High Range                               Reading (uV)           Difference (1V)               Error (%)
     Channel X        + Input                    200002.2                    —0.05                  0.00
     Channel X        + Input                    20000.16                    0.66                    0.00
     Channel X        — Input                   —~19997.14                   2.86                   —0.01
     Channel Y        + Input                    200008.3                    215                    —0.00
     Channel Y        + Input                    19996.72                    —2.68                  —0.01
     Channel Y        — Input                   ~19998.92                    0.08                   —0.00

     Channel Z        + Input                    200008.5                    —0.80                  —0.00
     Channel Z        + Input                    20000.01                    —0.09                  —0.00
     Channel Z        — Input                   —19998.00                     1.90                  —0.01

     Low Range                                 Reading (1V)           Difference (uV)               Error (%)
     Channel X        + Input                     1999.8                     —0.20                  —0.01
     Channel X        + Input                     200.22                     0.22                    0.11
     Channel X        — Input                    ~198.99                      1.01                  —0.50
     Channel Y        + Input                     2000.6                     0.94                    0.05
     Channel Y        + Input                     199.59                     —0.51                  —0.26
     Channel Y        — Input                    —200.74                     —0.84                   0.42
     Channel Z        + Input                     2000.0                     —0.14                  —0.01
     Channel Z        + Input                     198.71                     ~1.29                  —0.64
     Channel Z        — Input                    ~200.84                     —0.94                   0.47



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                       High Range                             Low Range
                        Input Voltage (mV)            Average Reading (uV)                  Average Reading (1V)

     Channel X                          200                       —8.70                          —10.53
                                       — 200                      11.41                           10.05
     Channel Y                          200                        0.01                           —0.31
                                       —200                       ~1.37                           ~1.76
     Channel Z                          200                       5.64                            5.59
                                       ~200                        3.08                           5.29


3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)         Channel X (uV)           Channel Y (uV)       Channel Z (1V)

    Channel X                    200                        —                   1.76                 ~1.72
    Channel Y                    200                   1.76                          —                1.74
    Channel Z                    200                   2.90                     —0.48                     —


Certificate No: DAE3—442_Dec11                               Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                    High Range (LSB)                        Low Range (LSB)

     Channel X                                              15778                                16839
     Channel Y                                              19772                                16308
     Channel Z                                              15590                                16770



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                            Average (uV)        min. Offset (uV)         max. Offset (uv)        °** ?:\)’;am“
     Channel X                   —0.87                —2.04                       0.18                   0.54
     Channel Y                   —1.01                —2.34                   —0.08                     0.42
     Channel Z                   —1.28                —3.05                       1.11                  0.70



6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA


7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                           Measuring (MOhm)
     Channel X                                             200                                   200
     Channel Y                                             200                                   200
     Channel Z                                             200                                   200



8. Low Battery Alarm Voltage (Typical values for information)
     Typical values                               Alarm Level (VDC)
     Supply (+ Vee)                                                                +7.9
     Supply (— Vee)                                                                 —7.6



9. Power Consumption (Typical values for information)
     Typical values                         Switched off (mA)        Stand by (mA)            Transmitting (mA)
     Supply (+ Vee)                                +0.01                    +6                         +14
     Supply (— Vee)                               —0.01                      —8                        —9




Certificate No: DAE3—442_Dec11                             Page 5 of 5