RF Exposure Appendix A

FCC ID: EJE-EM7455D1

RF Exposure Info

Download: PDF
FCCID_4217094

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/17


System Check_Body_750MHz

DUT: D750V3—1012

Communication System: CW ; Frequency: 750 MHz;Duty Cycle: 1: 1
Medium: MSL_750_190117 Medium parameters used: £= 750 MHz; 0 = 0.963 S/m; &, = 54.264; p = 1000
kg/m>
Ambient Temperature : 23.4 °C ; Liquid Temperature : 22.4 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(10.17, 10.17, 10.17) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.05 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.87 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.43 W/kg
SAR(1 g) = 2.32 W/kg; SAR(10 g) = 1.55 W/kg
Maximum value of SAR (measured) = 3.06 W/kg

        dB
        0


        —2.00


        —4.00


        —6.00


        —8.00
                       22

        —10.00        C
                    0 dB = 3.06 W/kg = 4.86 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/23


System Check_Body_750MHz

DUT: D750V3—1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1: 1
Medium: MSL_750_190123 Medium parameters used: £= 750 MHz; 0 = 0.969 S/m; s, = 55.292; p = 1000
kg/m>
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(10.17, 10.17, 10.17) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.07 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.87 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.46 W/kg
SAR(1 g) = 2.33 W/kg; SAR(10 g) = 1.56 W/kg
Maximum value of SAR (measured) = 3.08 W/kg

        dB
        0


        —2.22


        —4.43


        —6.65


        —B.86
                       22

        —11.08        C
                    0 dB = 3.08 W/kg = 4.89 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/17


System Check_Body_835MHz

DUT: D835V2—499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1: 1
Medium: MSL_850_190117 Medium parameters used: f= 835 MHz; 0 = 1.013 S/m; s, = 56.678; p = 1000
kg/m>
Ambient Temperature : 23.4 °C ; Liquid Temperature : 22.4 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(9.92, 9.92, 9.92) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.56 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 61.50 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 4.01 W/kg
SAR(1 g) =2.67 W/kg; SAR(10 g) = 1.76 W/kg
Maximum value of SAR (measured) = 3.55 W/kg

        dB
        0


        —2.07


        —4.15


        —6.22


        —8.30
                       22
                      K*                             &
        —10.37
                    0 dB = 3.55 W/kg = 5.50 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/17


System Check_Body_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_190117 Medium parameters used: £= 1750 MHz; 0 = 1.527 S/m; g, = 54.161; p =
1000 kg/m>
Ambient Temperature : 23.4 °C ; Liquid Temperature : 22.4 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(8.4, 8.4, 8.4) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v4.0; Type: QDOVAOOIBB; Serial: 1026
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.7 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 101.2 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) =9.89 W/kg; SAR(10 g) = 5.37 W/kg
Maximum value of SAR (measured) = 14.8 W/kg

        dB
        0


        —3.14


        —6.28


        —9.43


        12.57
                       22

        —15.71        6.
                    0 dB = 14.8 W/kg = 11.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/17


System Check_Body_1900MHz

DUT: D1900V2—5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_190117 Medium parameters used: £= 1900 MHz; 0 = 1.559 S$/m; g, = 51.91; p = 1000
kg/m>
Ambient Temperature : 23.4 °C ; Liquid Temperature : 22.4 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(8, 8, 8) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v4.0; Type: QDOVAOOIBB; Serial: 1026
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 17.0 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 109.3 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 19.8 W/kg
SAR(1 g) =10.7 W/kg; SAR(10 g) = 5.89 W/kg
Maximum value of SAR (measured) = 16.9 W/kg

        dB
        0


        —3.38


        —6.76


        —10.14


        —13.52
                       22

        —16.90        .
                    0 dB = 16.9 W/kg = 12.28 dBW/zg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/1/17


System Check_Body_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600_190117 Medium parameters used: £ = 2600 MHz; 0 = 2.234 S/m; &, = 51.728; p =
1000 kg/m>
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3931; ConvF(7.43, 7.43, 7.43) ; Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2018/11/16
— Phantom: ELI v4.0; Type: QDOVAOOIBB; Serial: 1026
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 22.1 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 99.37 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 29.9 W/kg
SAR(1 g) = 14.2 W/kg; SAR(10 g) = 6.34 W/kg
Maximum value of SAR (measured) = 21.9 W/kg

        dB
        0


        —4.80


        —9.59


        —14.39


        —19.18
                       22

        —23.98        6.
                    0 dB = 21.9 W/kg = 13.40 dBW/kg



Document Created: 2019-03-28 20:23:45
Document Modified: 2019-03-28 20:23:45