SAR Report Appendix 1

FCC ID: CNFSPJB1

RF Exposure Info

Download: PDF
FCCID_4328824

Appendix C – Dipole Calibration Certificates
2.4GHz Dipole


Calibration Laboratory of                                   Kw
Schmid & Partner
                                                                //           G   Schweizerischer Kalibrierdienst
                                                                             g   Service suisse d‘étatonnage
  Engineering AG                                                                 Servizio svizzero di taratura
Zeughausstrasse 48, 8004 Zurich, Switzerland                                 S   Swiss Calibration Service



Aceredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certficates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear(frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Condiitions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurementis stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—1005_Nov17                             Page 2 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          54.10+1.6 0
      Retum Loss                                                                       +275 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          51.2 0+ 3.5 jn
      Retum Loss                                                                       —28.7 dB




General Antenna Parameters and Design

    | Electrical Detay (anedirection)                       |                           1.157 ns                            ]
After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard seririgid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions* paragraph. The SAR data are not affected by this change. The overall dipols length is stil
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   SPEAG

      Manufactured on                                                             December 29, 2016




Certiicate No: D2450V2—1005_Nov17                        Page 4 of 8


UASY5 Validation Heport tor Head TSL

                                                                                   Date: 07.11.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:1005

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.86 S/m; s = 37.8; p = 1000 kg/mJ
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   +    Probe: EX3DV4 — SN7349; ConvF(8.12, 8.12, 8.12); Calibrated: 31.05.2017;

   +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   +    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   «_   DASY32 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 114.6 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) =27.1 W/kg
SAR(I g) = 13.4 W/kg; SAR(10 g) = 6.19 W/kg
Maximum value of SAR (measured) = 22.0 W/kg




        —4.00


        —8.00


        —12.00


        —16.00


        —20.00
                  0 dB =22.0 W/kg = 13.42 dB W/kg




Certificate No: D2450V2—1005_Novi7              Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                     7 Nev 2017 14:00:12
                sit     a uo rs         1: 54.096 o   1.5625 n   101.50 pH      2 450,000 000 MHz
                                                      eT
         §                                             es

         Del                                                         —
                                                                                  A
         Cs                             [                 1                   T



                                        |                 »              A~       *q
         Av                                                           L           4
         16°                                                         *z

         Hid                                                     s

          cn2




         Ca




         Ay
         16°

         H1d

                      START 2 250.000 000 M4z                                     SToP 2 650.000 000 MHz




Certificate No: D2450V2—1005_Nov17                      Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                Date: 07.11.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:1005

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: £= 2450 MHz; 0 = 2.02 S/m; s = 51.5; p = 1000 kg/m“
Phantom section: Flat Section
Measurement Standard: DASY3 (IEEE/IEC/ANSI C63.19—2011)

DASY32 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(8.1, 8.1, 8.1); Calibrated: 31.05.2017;

   *«   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   «+   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.8 V/m; Power Drift =—0.05 dB
Peak SAR (extrapolated) = 25.6 W/kg
SAR(I g) = 12.8 W/kg:; SAR(10 g) = 6 W/kg
Maximum value of SAR (measured)= 20.7 W/kg




        —4.00


        —8.00


        —12.00


        16.00


        —20.00
                  0 dB = 20.7 W/kg = 13.16 dBW/kg




Certificate No: D2450V2—1005_Nov17              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                7 Nev 2017 13:57:57
          CB3 si      i U Fs         11 51.209 a   25098    227.99 pH      2 450,000 000 MHz




          Del

          Ca                          [        4




          B
          159


          HLd

          cH2




          by
          16°

          Hid

                   STaRT 2 250.000 000 hiz                           SToP 2 650.000 000 MHz




Certificate No: D2450V2—1005_Nov17                  Page 8 of 8


Calibration Laboratory of                                           lt
                                                                    L
                                                                    w
                                                                                                            _      Schweizerischer Kalibrierdienst




                                                                 P
Schmid & Partner




                                                               J              w
                                                                                                            g Service suisse d‘étalonnage
  Engineering AG                                                     s                                             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                        ¥,
                                                                      Tltals®
                                                                             s                              S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Ctient       BACL                                                                               Certificate No: D5GHzV2—1001_Sep17

CALIBRATION CERTIFICATE
 Object                                D5GHzV2 — SN: 1001

 Calibration procedure(s)             QA CAL—22.v2
                                      Calibration procedure for dipole validation kits between 3—6 GHz



 Calibration date:                     September 18, 2017


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facilty: environmenttemperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      1D #                      Cal Date (Cortificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                Od—Apr—17 (No. 217—02521/02522)                       Apr—18
 Power sensor NRP—291                   SN: 108244                O4—Apr—17 (No. 217—02521)                             Apr—18
 Power sensor NRP—291                   SN: 103245                Od—Apr—17 (No. 217—02522)                             Apr—18
 Reference 20 dB Attenuator             SN: 5058 (204)            07—Apr—17 (No. 217—02528)                             Apr—18
 Type—N mismatch combination            SN: 5047.2 / o6g27        07—Apr—17 (No. 217—02529)                             Apr—18
 Reference Probe EXGDV4                 SN: 3503                  31—Dec—16 (No. EX3—3503_Dect6)                        Dec—17
 DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar18


 Secondary Standards                    tb #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM~442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     in house check: Oct—18
 Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                     in house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     in house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     in house check: Oct—18
 Network Analyzar HP 8763E              SN: US37300585            18—0ct—01 (in housecheck Oct—16)                      in house check: Oct—17


                                       Name                                    Function                                 Signature
 Calibrated by:                        Jeton Kastrati                          Laboratory Technician
                                                                                                           o*/

                                                                                                                   NKA
 Approved by:                           Katja Pokovie                          Technical Manager



                                                                                                                       Issued: September 19, 2017
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D5GHzV2—1001_Sep17                                   Page 1 of 13


Calibration Laboratory of                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                               S
                                                                                   Service   suissed‘étalonnage
  Engineering AG                                             i ns              C   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 "/,,/r_\\\ y      S   Swiss Calibration Service
                                                               Zalitabet®

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM xy,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
   * SAR measured: SAR measured at the stated antenna input power.
   * SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
   «_ SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.
  The reported uncertainty of measurementis stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D5GHzV2—1001_Sep17                              Page 2 of 13




Head TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C          35.5               5.07 mhoim
     Measured Head TSL parameters                              (22.0 £0.2) °C   36.2 2 6 %         4.95 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C                                  ——~—


SAR result with Head TSL at 5600 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                    8.13 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W       81.5 W / kg x 19.9 % (k=2)

     SAR averaged over 10 om‘ (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                    2.32 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        23.3 W/kg £19.5 % (k=2)


Head TSL parameters at 5750 MHz
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.4              5.22 mhoim
     Measured Head TSL parameters                              (22.0 £0.2) °C   36.0 £ 6 %         5.12 mho/m # 6 %
     Head TSL temperature change during test                      <0.5°C            —~                     ——

SAR result with Head TSL at 5750 MHz

     SAR averaged over 1 em‘ (1 g) of Head TSL                   Condition
     SAR measured                                        100 mW input power                    7.98 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        80.0 W/kg #19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                    2.26 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        22.7 Wikg #19.5 % (k=2)




Certificate No: D5GHzV2—1001_Sep17                    Page 4 of 13


Body TSL parameters at 5250 MHz
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C          48.9               5.36 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C   47.0 26 %          5.49 mho/m £6 %
     Body TSL temperature change during test                      <0.5 °C           ~—                     «>

SAR result with Body TSL at 5250 MHz

     SAR averaged aver 1 om‘ {1 g) of Body TSL                  Condition
     SAR measured                                        100 mW input power                    7.63 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W       75.7 Wikg # 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                condition
     SAR measured                                        100 mW input power                    214 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W       21.2 Wikg # 19.5 % (k=2)

Body TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C          48.5               5.77 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C   46.4 x6 %          5.96 mho/m # 6 %
     Body TSL temperature change during test                      <0.5 °C           ——                     ————

SAR result with Body TSL at 5600 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
     SAR measured                                        100 mW input power                    7.96 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        79.0 Wikg 2 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                 2.24 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        22.2 Wikg a 19.5 % (k=2)




Certificate No: DSGHzV2—1001_Sep17                    Page 5 of 13


Body TSL parameters at 5750 MHz
  The following parameters and calculations were applied.
                                                              Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                 22.0 °C           48.3               5.94 mho/m
     Measured Body TSL parameters                             (22.0 0.2) °C     46.1 £6 %          6.17 mho/m x 6 %
     Body TSL temperature change during test                     <0.5 °C            ——                     —~—

SAR result with Body TSL at 5750 MHz

     SAR averaged over 1 em‘ (1 g) of Body TSL                 Condition
     SAR measured                                       100 mW input power                     7.64 Wikg
     SAR for nominal Body TSL parameters                    rnormalized to 1W        75.8 Wikg #19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Body TSL                condition

     SAR measured                                       100 mW input power                  213 Wikg
     SAR for nominal Body TSL parameters                    normalized to 1W         21.1 Wikg £19.5 % (k=2)




Certificate No: D5GHzV2—1001_Sep17                   Page 6 of 13


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL at 5250 MHz

      Impedance, transformed to feed point                                           §1.4 0 — 6.7 |Q
      Retum Loss                                                                        —23.4 dB

Antenna Parameters with Head TSL at 5600 MHz

      Impedance, transformed to feed point                                           53.9 0 —1.7 j|Q
      Return Loss                                                                       —27.7 dB

Antenna Parameters with Head TSL at 5750 MHz

      Impedance, transformed to feed point                                           55.9 0 +3.1 in
      Return Loss                                                                       ~24.1 dB


Antenna Parameters with Body TSL at 5250 MHz

      Impedance, transformed to feed point                                           51.1 0 —5.4 jQ
      Return Loss                                                                       —25.3 dB

Antenna Parameters with Body TSL at 5600 MHz

      Impedance, transformed to feed point                                           55.7 0 — 1.9 jQ
      Retum Loss                                                                       —24.9 dB

Antenna Parameters with Body TSL at 5750 MHz

      Impedance, transformed to feed point                                           56.7 0 + 2.3 jQ
      Retum Loss                                                                        —23.5 dB

General Antenna Parameters and Design

      Electrical Detay (one direction)                                                   1.202 ns                        J

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxia! cable. The center conductor of the feeding line is direcily connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On someof the dipoles, small end caps
are added to the dipole arms in orderto improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections nearthe
feedpoint may be damaged.

Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 April 02, 2003


Certificate No: D5SGHzV2—1001_Sep17                      Page 7 of 13


DASY5 Validation Report for Head TSL

                                                                                      Date: 15.09.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 5GHz; Type: DSGHzV2; Serial; DSGHzV2 — SN: 1001
Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
Medium parameters used: f = 5250 MHz; 0 = 4.59 S/m; s; = 36.7; p = 1000 kg/m*
Medium parameters used: £= 5600 MHz; 0 = 4.95 $/m; s; = 36.2; p = 1000 kg/m*
Medium parameters used: f= 5750 MHz; 0 = 5.12 S/m; ; = 36.0; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2011)
DASY52 Configuration:

   *   Probe: EX3DV4 — SN3503; ConvF(5.58, 5.58, 5.58); Calibrated: 31.12.2016, ConvF(5.09, 5.09,
       5.09), Calibrated: 31.12.2016, ConvF(5.02, 5.02, 5.02); Calibrated: 31.12.2016;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAEA4 $n601; Calibrated: 28.03.2017

   &   Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50AA; Serial: 1001

   *   DASY52 52.10.0(1446);, SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5250 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.13 V/m; Power Drift =—0.07 dB
Peak SAR (extrapolated) = 28.5 W/kg
SAR(1 g) =7.85 W/kg; SAR(10 g) = 2.25 W/kg
Maximum value of SAR (measured) = 18.0 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurementgrid: dx=4mm, dy=4mm, dz=1 4mm
Reference Value = 67.01 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 31.5 W/kg
SAR(L g) =8.13 W/kg; SAR(10 g) = 2.32 W/kg
Maximum value of SAR (measured) = 19.2 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurementgrid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 65.65 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 31.7 W/kg
SAR(I g) = 7.98 W/kg; SAR(10 g) = 2.26 W/kg
Maximum value of SAR (measured) = 19.1 W/kg




Certificate No: D5GHzV2—1001_Sep17             Page 8 of 13


        5.00


       —10.00


       —15.00


       —20.00


       25.00
                  0 dB = 19.1 W/kg= 12.81 dBW/kg




Certificate No: DSGHzV2—1001_Sep17          Page 9 of 13


Impedance Measurement Plot for Head TSL




                                                               15 Sep 2oi7? idrdo:42
                it     i U Fs          11519790   —669920 4.9252pF         5 250.000 000 Kz
                                                   T_

                                                                                              CHL Markers
                                                                                               2 $3,928 a
                                                                                                ~4,5835 a
                                                                                              5.60000 GHz
                                                                                               3 55.991 a
                                                                                                 2.0762 a
                                                                                              5.75000 GHz

         hy
         16 ®


         HLd

          CH2



                                                                                              CH2 Markers

         Cor                                                                                  %—27.718 08
                                                                                              $.60000 GHz
                                                                                              a:—24.050 dB
                                                                                              $.75000 GHz

         Avi
         16°

         H1d

                     STaRT 5 000.000 000 Mz                          SToP 6 000,000 000 MHz




Certificate No: DSGHzV2—1001_Sep17                 Page 10 of 13


DASYS5 Validation Report for Body TSL

                                                                                          Date: 18.09.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 5GHz; Type: D5SGHzV2; Serial: DSGHzV2 — SN: 1001
Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
Medium parameters used: f = 5250 MHz; 0 = 5.49 S/m; s= 47.0; p = 1000 kg/m*
Medium parameters used: £= 5600 MHz; 0 = 5.96 $/m; £;= 46.4; p = 1000 kg/m*
Medium parameters used: f= 5750 MHz; 0 = 6.17 S/m; &,= 46.1; p = 1000 kg/m’
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)
DASY52 Configuration:

   «_   Probe: EX3DV4 — SN3503; ConvF(5.14, 5.14, 5.14); Calibrated: 31.12.2016, ConvF(4.57, 4.57,
        4.57); Calibrated: 31.12.2016, ConvF(4.51, 4.51, 4.51); Calibrated: 31.12.2016;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   «_   Electronics: DAEA4 Sn601; Calibrated: 28.03.2017

   +    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   &    DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5250 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 59.82 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 29.2 W/kg
SAR(L g) = 7.63 W/kg; SAR(1O0 g) = 2.14 W/kg
Maximum value of SAR (measured) = 17.3 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 59.65 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 33.1 W/kg
SAR(I g) = 7.96 W/kg; SAR(10 g) = 2.24 W/kg
Maximum value of SAR (measured) = 18.7 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 58.36 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 33.1 W/kg
SAR(1 g) = 7.64 W/kg; SAR(10 g) =2.13 W/kg
Maximum value of SAR (measured) = 18.3 W/kg




Certificate No: DSGHzV2—1001_Sep17               Page 23 of 23


       —10.00


       ~20.00


       ~30.00


       ~40.00


       —50.00
                  0 dB = 17.3 W/kg = 12.38 dBW/kg




Certificate No: D5GHzV2—1001_Sep17           Page 12 of 18


                                           E1 10 51 abeq               Lides~Looi—sAzH9sa :on steomeq




                ZHM 800 600°000 5 4015                          zi 008 000°000 5 14w15

                                                                                            PTH



                                                                                            3
zh9 ©0oSz‘S
ap ereec—t
Tho eo009‘s
ab Ob6 be—12                                                                                w09
sAe2tH THO




                                                                                            THo

                                                                                            PTH


                                                                                             ar
                                                                                            Bau



 9 conge‘s                                                                                  09
 €5907—
                                                                                            t°o
suexen tho


                                               es
               zuL ce cocrest 5        adessors torke      vetmte it         si n t      ts EBJ
                     evsorizt     atoz dos st




                                                    791 Apog 101 101q juowainseapy eouepaduj



Document Created: 2019-06-22 06:20:02
Document Modified: 2019-06-22 06:20:02

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC