SAR test report Appendix 1

FCC ID: CNFSPCC1

RF Exposure Info

Download: PDF
FCCID_4312869

                  Appendix A: DAE & Probe & Dipole Calibration Certificate


1.1. DAE4 Calibration Certificate




                                           1 of 36


           ;??Ti in Collaboration with
           .                        s_p_ e a q
           iess                     CALIBRATION LAEORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com        Hitp:/www.chinattl.en

Glossary:
DAE                             data acquisition electronics
Connector angle                  information used in DASY system to align probe sensor X
                                to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e DC Voltage Measurement: Calibration Factor assessed for use in DASY
   system by comparison with a calibrated instrument traceable to national
   standards. The figure given corresponds to the full scale range of the
  voltmeter in the respective range.

e    Connector angle: The angle of the connector is assessed measuring the
     angle mechanically by a tool inserted. Uncertainty is not required.

e    The report provide only calibration results for DAE, it does not contain other
     performance test results.




Certificate No: Z18—60157                               Page 2 of 3


           =A®                       in Collaboration with

          =‘/"[‘], a
          ie CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com        Hittp:/www.chinattl.en



 DC Voltage Measurement
     AID — Converter Resolution nominal
         High Range:       1LSB =       6.1uV ,              full range =    —100... +300 mV
         Low Range:        1LSB =        61nV ,              full range =    AAsiclll +3mV
     DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors                     X                              Y                     Z
      High Range                   404.484 + 0.15% (k=2)           404.709 + 0.15% (k=2)   404.531 + 0.15% (k=2)

      Low Range                    3.97937 + 0.7% (k=2)            4.00986 £0.7% (k=2)     4.02022 + 0.7% (k=2)



Connector Angle


      Connector Angle to be used in DASY system                                                 192°%1"




Certificate No: Z18—60157                               Page 3 of 3


                                                                                       wl m                                  L

                 &7
                 camn 7J, _
                          s
                 _h

                                     e 4  a
                                         In Collaboration with

                                              PSrRACNAS :




                                                                                 7+



                                                                                               \®
                                                                                  $ \“&J//"’/ 2l
                                                                                                                        ;:;:{

                          CALIBRATION LABORAT                      ORY            zy                   v CALIBRATION




                                                                                                Th
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China        "/,,,m\‘.\“                       CNAS LO570
             Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
             E—mail: cttl@chinattl.com        Hitp://www.chinattl.cn
       Client            GTI                                                Certificate No:         2Z18—60160

CALIBRATION CERTIFICATE

Object                                     EX3DV4 — SN:3974

Calibration Procedure(s)                   FF—211—004—01

                                           Calibration Procedures for Dosimetric E—field Probes

Calibration date:                          June 21, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘c and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)
Primary Standards                      ID #          Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 Power Meter   NRP2               101919               27—Jun—17 (CTTL, No.J17X05857)                       Jun—18
 Power sensor NRP—Z91             101547               27—Jun—17 (CTTL, No.J17X05857)                       Jun—18
 Power sensor NRP—Z91             101548               27—Jun—17 (CTTL, No.J17X05857)                       Jun—18
 Reference10dBAttenuator          18N50OW—10dB         09—Feb—18(CTTL, No.J18X01133)                        Feb—20
 Reference20dBAttenuator          18NSOW—20dB           09—Feb—18(CTTL, No.J18X01132)                       Feb—20
 Reference Probe EX3DV4           SN 3846              25—Jan—18(SPEAG,No.EX3—3846_Jan18)                   Jan—19
 DAE4                             SN 777               15—Dec—17(SPEAG, No.DAE4—777_Dec17)                  Dec —18

 Secondary Standards             ID #                  Cal Date(Calibrated by, Certificate No.)          Scheduled Calibration
 SignalGeneratorMG3700A          6201052605            27—Jun—17 (CTTL, No.J17X05858)                     Jun—18
 Network Analyzer E5071C         MY46110673            14—Jan—18 (CTTL, No.J18X00561)                     Jan —19
                               Name                          Function                                      Signature
Calibrated by:                  Yu Zongying                 SAR Test Engineer                              /;},,44?

Reviewed by:                    Lin Hao                     SAR Test Engineer                             'Tfif,%)

Approved by:                    Qi Dianyuan                 SAR Project Leader                        woi
                                                                                       Issued: June 23, 2018
This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.



    Certificate No: Z18—60160                               Page 1 of 11


                                    In Collaboration with

             [/"["/, a
          r\
          =‘
          ligpzzem>" CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Hitp://www.chinattl. en


Glossary:
TSL                   tissue simulating liquid
NORMx,y,z             sensitivity in free space
ConvF                 sensitivity in TSL / NORMx,y,z
DCP                diode compression point
CF                 crest factor (1/duty_cycle) of the RF signal
A,B,.C,D           modulation dependent linearization parameters
Polarization ®     ® rotation around probe axis
Polarization 0     0 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                   0=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
   Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
   July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e    NORMx,y,z: Assessed for E—field polarization 6=0 (f900MHz in TEM—cell; > 1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
    E" field uncertainty inside TSL (see below ConvF).
e   NORM(Mx,y,z = NORMx,y,z* frequency_response (see Frequency Response Chart). This
    linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
    frequency response is included in the stated uncertainty of ConvF.
e   DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
e   PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
    characteristics.
e   Axyz; Bx,yz; Cx,yz;VRx,y,z:A,B,C are numericallinearization parameters assessed based on the
    data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
    media. VR is the maximum calibration range expressed in RMS voltage across the diode.
e   ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
    Transfer Standard for fs800MHz) and inside waveguide using analyticalfield distributions based on
    power measurements for f >800MHz. The same setups are used for assessment of the parameters
    applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
    These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
    The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
    that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
    allows extending the validity from+50MHz to£100MHz.
e   Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
    phantom exposed by a patch antenna.
e   Sensor Offset: The sensoroffset corresponds to the offset of virtual measurement center from the
    probe tip (on probe axis). No tolerance required.
e   Connector Angle: The angle is assessed using the information gained by determining the NORMsx
    (no uncertainty required).

Certificate No: Z18—60160                              Page 2 of 11


           m                       In Collaboration with

          ==® ‘/"/"/
          chee                     S      _p      e        a
                                                           a g ____
                                   CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hitp://www.chinattl. en




                       Probe EX3DV4


                                               SN: 3974


                                        Calibrated: June 21, 2018

                              Calibrated for DASY/EASY Systems
                                       (Note: non—compatible with DASY2 system!)




Certificate No: Z18—60160                             Page 3 of 11


           r                        In Collaboration with


           »
                 TTL a              CALIBRATION LABORATORY

         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com      Hitp://wwrw.chinattl. en



      DASY/!EASY — Parameters of Probe: EX3DV4 — SN: 3974

Basic Calibration Parameters
                                     Sensor X                Sensor Y              Sensor Z     Unc (k=2)
 Norm(uV/(V/im))A                    0.48                    0.46                  0.52         £10.0%
 DCP(mV)®                            103.7                   104.3                 102.4


Modulation Calibration Parameters
  uiD         Communication                        A            B            6         D      VR       Unc®
              System Name                          dB           dB/uV                  dB     mV       (k=2)
  0           Cw                          X        0.0          0.0          1.0       0.00   163.9    +2.2%
                                          Y        0.0          0.0          1.0              162.1
                                          2        0.0          0.0          1.0              172.5



      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement multiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probatbility of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6).
® Numerical linearization parameter: uncertainty not required.
& Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z18—60160                                Page 4 of 11


             r                         in Collaboration with

             =‘/"[‘/, a
              v                        CALIBRATION LABORATORY
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com        Hitp://www.chinattl.en


       DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3974

Calibration Parameter Determined in Head Tissue Simulating Media
                        c                      m   04                                                      6

  fMHz]® Pef:"i:i'\‘l’;yF c°"'::7l:‘)":y ConvF X ConvF Y ConvF Z Alpha® D(;p':;                                ::(';c;)
    750            41.9                 0.89             1042            1042   1042     040        0.80       £121%
    835            41.5                 0.90             1012            1012   1012     014        145        £121%
    900            41.5                 0.97             10.00           1000   10.00    017        1.38       £121%
   1750            40.1                 1.37             8.78            8.78   878      024        1.07       £121%
   1900            40.0                 1.40             8.49            8.49   849      0.24       1.08       £121%
   2450            39.2                 1.80             8.01            8.01   801      034        1.04       £121%
   2600            39.0                 1.96             779             7.79   779      0.39       1.00       £121%
   5200            36.0                 4.66             5.72            5.72   5.72     040        170        £13.3%
   5300            35.9                 4.76             5.48            5.48   548      0.40       1.60       £13.3%
   5500            35.6                 4.96             5.04            5.04   504      040        1.80       £13.3%
   5600            35.5                 5.07             4.85            4.85   485      040        1.70       +13.3%
   5800            35.3                 5.27             5.01            5.01   501      045        1.65       £13.3%

6 Frequency validity above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
*50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
"At frequency below 3 GHz, the validity of tissue parameters (¢ and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (€ and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
GAIpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




  Certificate No: Z18—60160                               Page 5 of 11


             !\                        In Collaboration with

             =‘/"/"/, s p_ q
                           e a
                                       CALIBRATION LABORATORY
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com        Hitp://www.chinattl. en



       DASY/!EASY — Parameters of Probe: EX3DV4 — SN: 3974

Calibration Parameter Determined in Body Tissue Simulating Media
                        Tz                    x                                                          6
 f [MHz]® Pe':;'i‘::i"‘l';yF °°"'::Z:‘)"Fty ConvF X ConvF Y ConvF Z Alpha® D(:°;")                           :‘:(':;)
    750         55.5              0.96       1062    i1062   i1062   040     0.80                            +121%
    835         55.2              0.97       1033    1033    1033    018     1.38                            £121%
    900         55.0              1.05       i0227   1027    1027    020     125                             £121%
   1750         53.4              1.49        8.42    8.42    842    028     1.00                            £121%
   1900         53.3              1.52        8.09    8.09    809    022     117                             £121%
   2450         52.7              1.95        8.00    8.00    Bo0    042     1.03                            £121%
   2600         525               216         7.52    7.52    752    052     0.88                            £121%
   5200         49.0              5.30        5.24    5.24    5.24   0.50    1.55                            +13.3%
   5300         48.9              5.42        515     515     515    050     145                             +13.3%
   5500         48.6              5.65        4.57    4.57   457     050     1.65                            £13.3%
   5600         485               5.77        4.46    4.46   446     0.50    1.55                            £13.3%
   5800         48.2              6.00        4.58    4.58   458     0.50    1.60                            £13.3%

© Frequency validity above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is £ 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
"At frequency below 3 GHz, the validity of tissue parameters (e and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




  Certificate No: Z18—60160                               Page 6 of 11


                                                            In Collaboration with

                               aa [TL a
                                     v                      CALIBRATION LABORATORY

               Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
               Tel: +86—10—62304633—2512                         Fax: +86—10—62304633—2504
               E—mail: cttl@chinattl.com                         Hitp://www.chinattl.on



                                                      Frequency Response of E—Field
                                                      (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normalized)




                                         0.5           +    T              T               T        T            T
                                               0           500           1000            1500     2000         2500

                                                                 Cs—                    f [MHz]          «g>
                                                                 TEM                                     R22

                                                   Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z18—60160                                                      Page 7 of 11


             !\                             In Collaboration with
             =/"7‘J, a
             imz CALIERATION LABEORATORY
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com      Hitp://www.chinattl.en



                                   Receiving Pattern (®), 0=0°

                 f=600 MHz, TEM                                                      f=1800 MHz, R22




     1.0



    0.5


m                                           ~#—uuhg—.                                               ;         —lx—;m ie
E   0.0               i ‘.’J..,.AI»I    *          ts   .\.‘“‘\".              ;      hss *     a       199            #   a?“\‘h.»             i
o              hn t                                                    ® l\»‘gr—;"lf"                                           j     mewseng
L
    —0.5



     1.0              T                 T          *       T                   T            I                     T             T
                  +150                 +100               —50                  0           50                    100           150
                                                                         Rollf]
                              [—*~1100MkHz___— _600MHz                             _ _+ 1800MHz ___—            __2500MHz]
                              — Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z18—60160                                       Page 8 of 11


                                       In Collaboration with

           =‘[/"["/ a
           imz CALIERATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com            Hitp://www.chinattl. en


                                     Dynamic Range f(SARneaq)
                                         (TEM cell, f = 900 MHz)
                       10 4




              2C
              £
              J
              L
              w
              a
              [=M
              £




                       10                                i—i—HHARH——i—I—H                           i        ——I—i
                            10°                                 10°                         10°              10°         10°
                                                                      SAR[mWcm*]
                                                         CMO/not compensated      —@— compensated




                  1

          <                     I  #  }                           Pj
          J 0               .—H—QA—. Ge—'—.—.-t:--:—-O s —040=»: « ,..."
                                                                       Mess

          w                                                                  m
               A                                                                 (94j
                                                                                        m

               24
                                       i
                                     mevey           —        em        o   —mmemg
                                                                                    0_——i—               y       —   n
                      10°               10°                     10                ; 10                     10°       10°
                                                 _                   SAR[mW/icm‘]
                                  |___—=—notcoi
                                      —=—not compensated.                                         ® compensated _|
                                    Uncertainty of Linearity Assessrge nt: £0.9% (k=2)
Certificate No: Z18—60160                                    Page 9 of 11


                                             In Collaboration with
          é TTL a
                                             CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hitp://www.chinattl.on



                            Conversion Factor Assessment
       f=750 MHz, WGLS R9(H_convF)                                                           f=1750 MHz, WGLS R22(H_convF)
              350 (—                                                                              20.00                                 1                              1
                  +
              t# §                                                                                2s 0o 4
              as \                                                                                            \4
                                                                                                                                                                  s        ~
                 |—4                                                                                                |
                  ft                                                                          : #4 1|A
                                                                                                     | ————
            %'—‘M E—             j                        =~=                                 %    X
            3     |                                                                           S 1500                M                                             3        =
            §1.84 |——¥i———                                      se                            €
            8|                                                                                &
                                                                                                  10.00             +                       ——i—                  —        —
              tou }              _
                  t               \
                  t
              om {                                          '                      |               500                         i‘


              boa L._._A_____S                                                     )               a 00                                              Femsenissmmmtine.
                  0             20           40            60         80          100                     0         10        20    30              40     5          s0   id
                                                   zmm}                                                                                     zimm]
                                      —#—measured         —— snalytica                                                    —&—measured               — analjfica




                       Deviation from Isotropy in Liquid
                                     1.0

                                     0.8

                                     06

                                     0.4

                                      0.2
                       7 Axis




                                      0.0
                                     —0.2
                                      —0.4
                                      —0.6
                                      —0.8
                                       1f

                                                                                                                         30
                                                          150                                                      8           $
                                                                                                          1000           [¥
                                                                200                                                            y
                                                       *4*@           250
                                                                            300
                                                                                       aso        _°
                                                  10    080     .060 .040 .020          0     020      oa0         oco   O80       10
                                 Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z18—60160                                             Page 10 of 11


                    ’I'rrV""‘"‘- mm en m n n on n n un n n                  n n   n




         fil ;Collaboratjogmth a

          y                         CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hitp://www.chinattl.on


   DASY/!EASY — Parameters of Probe: EX3DV4 — SN: 3974


     Other Probe Parameters
    Sensor Arrangement                                                                Triangular

    Connector Angle (°)                                                                   133.9

    Mechanical Surface Detection Mode                                                   enabled

    Optical Surface Detection Mode                                                      disable

    Probe Overall Length                                                                337mm

    Probe Body Diameter                                                                   10mm

    Tip Length                                                                             9mm

    Tip Diameter                                                                         2.5mm

    Probe Tip to Sensor X Calibration Point                                                1mm

    Probe Tip to Sensor Y Calibration Point                                                1mm

    Probe Tip to Sensor Z Calibration Point                                                1mm

    Recommended Measurement Distance from Surface                                        1.4mm




Certificate No: Z18—60160                              Page 11 of 11


                      TTLf e
                           (btokbcaicout «l73 P PRA
                                         s\U»   {&}!
                                         CALIBRATION

             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   -",'//:\\\¢                      CALIBRATION
             Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504               *Arludubs®®                    CNAS LO570
             E—mail: ettl@chinattl.com            :!Iwww.chinattl.en

      Client               GTI                                              Certificate No:      218—60198
                                                                                 Aoaioarticane
                                                                                       o iess                                ®:
                                                                                       o.


Object                                     D2450V2 — SN: 928

Calibration Procedure(s)                   FEz11—003.01

                                           Calibration Procedures for dipole validation kits

Calibration date:                          October 12, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22:3)C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #               Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Power Meter        NRP2          101919            27—Jun—17 (CTTL, No.J16X04777)                          Jun—18
 Power sensor       NRP—Z91       101547            27—Jun—17 (CTTL, No.J16X04777)                          Jun—18
 Reference Probe ES3DV3           SN 3149            15—Apr—17(CTTL—SPEAG,No.J16—97035)                      Apr—18
 DAE4                             SN 777            22—Aug—17(CTTL—SPEAG,No.216—97138)                       Aug—18

 Secondary Standards              ID #              Cal Date(Calibrated by, Certificate No.)         Scheduled Calibration
 Signal Generator E4438C          MY49071430        01—Feb—17 (CTTL, No.J16X00893)                         Jan—18
 Network Analyzer E5071C          MY46110673          26—Jan—17 (CTTL, No.J16X00894)                       Jan—18



                                   Name                         Function                                  Signature
patiorated by:                   Zhao Jing                  SAR Test Engineer                           K¥
Reviewed by:                     Qi Dianyuan                SAR Project Leader                       73?,67}
                                                                                                          C3

Approved by:                     Liu Wei                   Deputy Director of SEM Department          . Jyn|i%>       .
                                                                                       issued: October 14, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z18—60198                               Page 1 of 8


         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


Glossary:
TSL                        tissue simulating liquid
ConvF                      sensitivity in TSL / NORNMx,y,z
N/A                        not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
    exposure to radio frequency fields from hand—held and body—mounted wireless
    communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
    6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e  Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
e     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
*     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
      SAR measured: SAR measured at the stated antenna input power.
      SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
e     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.

Certificate No: Z18—60198                               Page 2 of 8


                TTL CALIBRATION
                    ;_n;v:c_a_  LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                        DASY52                                     52.110.1.1476

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                         10 mm                                   with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                      2450 MHz & 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature              Permittivity           Conductivity

     Nominal Head TSL parameters                             22.0 °C                  39.2                1.80 mho/m

     Measured Head TSL parameters                        (22.0 £ 0.2) °C            39.0 £ 6 %          1.78 mho/m + 6 %

     Head TSL temperature change during test                 <1.0 °C                   ~——                      «——
SAR result with Head TSL
     SAR averaged over 1 _Cm" _(1 g) of Head TSL                       Condition
     SAR measured                                                  250 mW input power                 12.9 mW /g

     SAR for nominal Head TSL parameters                            normalized to 1W         51.9 mW /g £ 20.8 % (k=2)
     SAR averaged over 10 cm* (10 g) of Head TSL                       Condition
     SAR measured                                                  250 mW input power                 6.08 mW /g
     SAR for nominal Head TSL parameters                            normalized to 1W         24.4 mW ig + 20.4 % (k=2)
Body TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature              Permittivity           Conductivity

     Nominal Body TSL parameters                             22.0 °C                  52.7                1.95 mho/m

     Measured Body TSL parameters                        (22.0 £ 0.2) °C            52.9 46 %           1.92 mho/m £ 6 %

     Body TSL temperature change during test                 <1.0 °C                   ——                       ————
SAR result with Body TSL
     SAR averaged over 1_cm" _(1 g) of Body TSL                        Condition
     SAR measured                                                  250 mW input power                 12.6 mW /g

     SAR for nominal Body TSL parameters                            normalized to 1W         50.8 mW /g * 20.8 % (k=2)

     SAR averaged over 10 em‘ (10 g) of Body TSL                       Condition

     SAR measured                                                  250 mW input power                 5.96 mW /g

     SAR for nominal Body TSL parameters                            normalized to 1W         23.9 mW /g * 20.4 % (k=2)


Certificate No: Z18—60198                            Page 3 of 8


                TTL ;_D;Q:L.n_
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


 Appendix

 Antenna Parameters with Head TSL

     Impedance, transformed to feed point                                   52.50+ 4.52j0

     Return Loss                                                             —25.90B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   51.10+ 5.39)0

     Return Loss                                                             —25.30B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.263 ns



 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some
 ofthe dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z18—60198                            Page 4 of 8



Document Created: 2019-06-06 15:35:06
Document Modified: 2019-06-06 15:35:06

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC