15U21180-S1V3 SAR_App B System Check Plots

FCC ID: C9O-QOLL2

RF Exposure Info

Download: PDF
FCCID_2819034

Test Laboratory: UL Verification Services Inc., SAR Lab 1                    Date/Time: 8/10/2015 8:08:32 PM

20150810 SystemPerformanceCheck-D835V2 SN 4d142

Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 1.016 S/m; εr = 52.738; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1352; Calibrated: 11/7/2014
- Probe: EX3DV4 - SN7356; ConvF(9.97, 9.97, 9.97); Calibrated: 4/22/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI-A v5.0; Type: QDOVA002AA; Serial: TP 1194

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 0.946 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 30.825 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 1.28 W/kg
SAR(1 g) = 0.864 W/kg; SAR(10 g) = 0.571 W/kg
Maximum value of SAR (measured) = 1.05 W/kg




                      0 dB = 1.05 W/kg = 0.21 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc., SAR Lab 1   Date/Time: 8/10/2015 8:27:46 PM

20150810 SystemPerformanceCheck-D835V2 SN 4d142
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 0.950 W/kg




                                                                                              Plot No. 2


Test Laboratory: UL Verification Services Inc., SAR Lab 2                    Date/Time: 8/11/2015 6:36:03 PM

20150811_SystemPerformanceCheck-D1750V2 SN 1050

Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.476 S/m; εr = 52.02; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Probe: EX3DV4 - SN3990; ConvF(8.08, 8.08, 8.08); Calibrated: 3/18/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI A v5.0; Type: QDOVA002AA; Serial: S/n:1198

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 4.89 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.524 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 6.95 W/kg
SAR(1 g) = 3.89 W/kg; SAR(10 g) = 2.07 W/kg
Maximum value of SAR (measured) = 5.23 W/kg




                      0 dB = 5.23 W/kg = 7.19 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc., SAR Lab 2   Date/Time: 8/11/2015 6:57:57 PM

20150811_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.87 W/kg




                                                                                              Plot No. 4


Test Laboratory: UL Verification Services Inc., SAR Lab 2                    Date/Time: 8/28/2015 2:04:15 PM

20150828 SystemPerformanceCheck-D1750V2 SN 1053

Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.502 S/m; εr = 51.801; ρ = 1000 kg/m
                                                                                3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1259; Calibrated: 1/14/2015
- Probe: EX3DV4 - SN3990; ConvF(8.08, 8.08, 8.08); Calibrated: 3/18/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI A v5.0; Type: QDOVA002AA; Serial: S/n:1198

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 4.95 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.162 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 6.97 W/kg
SAR(1 g) = 3.9 W/kg; SAR(10 g) = 2.08 W/kg
Maximum value of SAR (measured) = 5.23 W/kg




                      0 dB = 5.23 W/kg = 7.19 dBW/kg




                                                                                                               Plot No. 5


Test Laboratory: UL Verification Services Inc., SAR Lab 2   Date/Time: 8/28/2015 2:26:17 PM

20150828 SystemPerformanceCheck-D1750V2 SN 1053
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.92 W/kg




                                                                                              Plot No. 6


Test Laboratory: UL Verification Services Inc., SAR Lab 3                    Date/Time: 8/11/2015 11:28:57 AM

20150811_SystemPerformanceCheck-D1900V2 SN 5d163

Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.52 S/m; εr = 52.018; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1434; Calibrated: 4/16/2015
- Probe: EX3DV4 - SN3749; ConvF(7.09, 7.09, 7.09); Calibrated: 1/26/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI-B v5.0; Type: QDOVA001BB; Serial: 1215

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.60 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 61.18 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 7.75 W/kg
SAR(1 g) = 4.19 W/kg; SAR(10 g) = 2.18 W/kg
Maximum value of SAR (measured) = 5.65 W/kg




                      0 dB = 5.65 W/kg = 7.52 dBW/kg




                                                                                                               Plot No. 7


Test Laboratory: UL Verification Services Inc., SAR Lab 3   Date/Time: 8/11/2015 11:46:21 AM

20150811_SystemPerformanceCheck-D1900V2 SN 5d163
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.66 W/kg




                                                                                           Plot No. 8


Test Laboratory: UL Verification Services Inc., SAR Lab 5                    Date/Time: 9/1/2015 12:04:13 PM

20150901_SystemPerformanceCheck-D835V2 SN 4d142

Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 1.008 S/m; εr = 53.639; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 4/16/2015
- Probe: EX3DV4 - SN3773; ConvF(8.26, 8.26, 8.26); Calibrated: 4/22/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 (A); Type: QDOVA001BB; Serial: 1213

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.08 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 33.34 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 1.40 W/kg
SAR(1 g) = 0.959 W/kg; SAR(10 g) = 0.632 W/kg
Maximum value of SAR (measured) = 1.16 W/kg




                      0 dB = 1.16 W/kg = 0.64 dBW/kg




                                                                                                               Plot No. 9


Test Laboratory: UL Verification Services Inc., SAR Lab 5   Date/Time: 9/1/2015 12:20:25 PM

20150901_SystemPerformanceCheck-D835V2 SN 4d142
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.08 W/kg




                                                                                          Plot No. 10


Test Laboratory: UL Verification Services Inc. SAR Lab E                     Date/Time: 11/16/2015 9:56:50 AM

20151116_SystemPerformanceCheck-D835V2 SN 4d117

Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 0.982 S/m; εr = 53.589; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1439; Calibrated: 7/30/2015
- Probe: EX3DV4 - SN3772; ConvF(8.46, 8.46, 8.46); Calibrated: 2/23/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 B; Type: QDOVA002AA; Serial: 1180

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.13 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 35.915 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 1.38 W/kg
SAR(1 g) = 0.933 W/kg; SAR(10 g) = 0.615 W/kg




                      0 dB = 1.13 W/kg = 0.53 dBW/kg




                                                                                                               Plot No. 11


Test Laboratory: UL Verification Services Inc. SAR Lab E    Date/Time: 11/16/2015 10:16:23 AM

20151116_SystemPerformanceCheck-D835V2 SN 4d117
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.12 W/kg




                                                                                         Plot No. 12


Test Laboratory: UL Verification Services Inc. SAR Lab F                    Date/Time: 11/2/2015 10:18:30 PM

20151102_SystemPerformanceCheck-D835V2 SN 4d002

Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 1.013 S/m; εr = 53.183; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/18/2015
- Probe: EX3DV4 - SN3929; ConvF(8.57, 8.57, 8.57); Calibrated: 4/22/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 (B); Type: QDOVA001BB; Serial: 1118

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.20 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 35.452 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 1.42 W/kg
SAR(1 g) = 0.984 W/kg; SAR(10 g) = 0.651 W/kg
Maximum value of SAR (measured) = 1.19 W/kg




                      0 dB = 1.19 W/kg = 0.76 dBW/kg




                                                                                                               Plot No. 13


Test Laboratory: UL Verification Services Inc. SAR Lab F   Date/Time: 11/2/2015 10:38:40 PM

20151102_SystemPerformanceCheck-D835V2 SN 4d002
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.18 W/kg




                                                                                         Plot No. 14


Test Laboratory: UL Verification Services Inc. SAR Lab G                    Date/Time: 11/2/2015 9:41:49 PM

20151102_SystemPerformanceCheck-D1900V2 SN 5d043

Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.57 S/m; εr = 51.377; ρ = 1000 kg/m
                                                                               3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/12/2015
- Probe: EX3DV4 - SN3991; ConvF(7.73, 7.73, 7.73); Calibrated: 5/19/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: QDOVA002AA; Serial: 1258

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.23 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 56.716 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 7.59 W/kg
SAR(1 g) = 4.11 W/kg; SAR(10 g) = 2.12 W/kg
Maximum value of SAR (measured) = 5.59 W/kg




                      0 dB = 5.59 W/kg = 7.47 dBW/kg




                                                                                                               Plot No. 15


Test Laboratory: UL Verification Services Inc. SAR Lab G    Date/Time: 11/2/2015 10:00:14 PM

20151102_SystemPerformanceCheck-D1900V2 SN 5d043
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.25 W/kg




                                                                                          Plot No. 16


Test Laboratory: UL Verification Services Inc. SAR Lab G                    Date/Time: 11/3/2015 4:13:24 PM

20151103_SystemPerformanceCheck-D1750V2 SN 1053

Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.487 S/m; εr = 51.148; ρ = 1000 kg/m
                                                                                3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/12/2015
- Probe: EX3DV4 - SN3991; ConvF(8.01, 8.01, 8.01); Calibrated: 5/19/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: QDOVA002AA; Serial: 1258

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 4.79 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.142 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 6.83 W/kg
SAR(1 g) = 3.83 W/kg; SAR(10 g) = 2.03 W/kg
Maximum value of SAR (measured) = 5.13 W/kg




                      0 dB = 5.13 W/kg = 7.10 dBW/kg




                                                                                                               Plot No. 17


Test Laboratory: UL Verification Services Inc. SAR Lab G    Date/Time: 11/3/2015 4:31:56 PM

20151103_SystemPerformanceCheck-D1750V2 SN 1053
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.78 W/kg




                                                                                          Plot No. 18


Test Laboratory: UL Verification Services Inc. SAR Lab G                    Date/Time: 11/16/2015 10:48:25 AM

20151116_SystemPerformanceCheck-D1900V2 SN 5d163

Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.551 S/m; εr = 51.897; ρ = 1000 kg/m
                                                                                3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/12/2015
- Probe: EX3DV4 - SN3991; ConvF(7.73, 7.73, 7.73); Calibrated: 5/19/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: QDOVA002AA; Serial: 1258

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.57 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.638 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 7.65 W/kg
SAR(1 g) = 4.1 W/kg; SAR(10 g) = 2.1 W/kg
Maximum value of SAR (measured) = 5.58 W/kg




                      0 dB = 5.58 W/kg = 7.47 dBW/kg




                                                                                                               Plot No. 19


Test Laboratory: UL Verification Services Inc. SAR Lab G    Date/Time: 11/16/2015 11:06:48 AM

20151116_SystemPerformanceCheck-D1900V2 SN 5d163
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.55 W/kg




                                                                                          Plot No. 20


Test Laboratory: UL Verification Services Inc. SAR Lab G                    Date/Time: 11/18/2015 6:41:45 PM

20151118_SystemPerformanceCheck-D1750V2 SN 1053

Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.469 S/m; εr = 51.479; ρ = 1000 kg/m
                                                                                3

DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/12/2015
- Probe: EX3DV4 - SN3991; ConvF(8.01, 8.01, 8.01); Calibrated: 5/19/2015;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: QDOVA002AA; Serial: 1258

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.26 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.505 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 7.19 W/kg
SAR(1 g) = 3.99 W/kg; SAR(10 g) = 2.12 W/kg
Maximum value of SAR (measured) = 5.34 W/kg




                      0 dB = 5.34 W/kg = 7.28 dBW/kg




                                                                                                               Plot No. 21


Test Laboratory: UL Verification Services Inc. SAR Lab G    Date/Time: 11/18/2015 7:00:09 PM

20151118_SystemPerformanceCheck-D1750V2 SN 1053
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.27 W/kg




                                                                                          Plot No. 22