SAR test report

FCC ID: C3KMN720

RF Exposure Info

Download: PDF
FCCID_340807

                                         SAR Test Report
                                           No. SAR_499_2003_FCC_2450


                                                                       for the
                                                      Microsoft Corporation

The Microsoft Broadband Networking Wireless Notebook Adapter (Mod
                        el number MN-720)

                                                         FCC ID: C3KMN720




                                            Date of Report:                       06/18/2003
                                            Date of issue:                        06/18/2003
                                             Report Copy No.:                     01




                                                                                                                         FCC listed # 101450
                                                          Bluetooth Qualification                                        IC recognized # 3925
                                                               Test Facility
TTI-P-G 081/94-A0                                                (BQTF)
Accredited according to ISO/IEC          17025


                                                                 CETECOM Inc.
                                            411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.
                 Phone: + 1 (408) 586 6200 Š Fax: + 1 (408) 586 6299 Š E-mail: info@cetecomusa.com Š http://www.cetecom.com
                                    CETECOM Inc. is a Delaware Corporation with Corporation number: 2113686
                                     Board of Directors: Dr. Harald Ansorge, Dr. Klaus Matkey, Hans Peter May


V.2.16M-2002-03-12 f:\emc\projects\broadcom\499_2003_mn_720_cardbus\report\sar_499_2003_fcc_2450.doc     © Copyright by CETECOM


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 2 of 28


Contents


1 ASSESSMENT ...................................................................................................................4


2 ADMINISTRATIVE DATA...................................................................................................5

     2.1       Identification of the Testing Laboratory Issuing the SAR Assessment Report ...............5
     2.2       Identification of the Client...........................................................................................5
     2.3       Identification of the Manufacturer...............................................................................5


3 EQUIPMENT UNDER INVESTIGATION (EUI)...................................................................6

     3.1       Identification of the Equipment under Investigation ...................................................6
     3.2       Front View of the Equipment under Investigation ......................................................7


4 SUBJECT OF INVESTIGATION ........................................................................................8

     4.1       The IEEE Standard C95.1 and the FCC Exposure Criteria .......................................8
     4.2       Distinction Between Exposed Population, Duration of Exposure and Frequencies...8
     4.3       Distinction between Maximum Permissible Exposure and SAR Limits......................9
     4.4       SAR Limit ...................................................................................................................9


5 THE FCC MEASUREMENT PROCEDURE .....................................................................10

     5.1       General Requirements.............................................................................................10
     5.2       Device Operating Next to a Person’s Ear ................................................................10
     5.3       Test Positions ..........................................................................................................11
     5.4       Test positions of device relative to head..................................................................11
     5.5       Test to be Performed ...............................................................................................14
     5.6       Body-worn and Other Configurations.......................................................................14
     5.7       Procedure for assessing the peak spatial-average SAR .........................................15
     5.8       Determination of the largest peak spatial-average SAR ..........................................17


6 THE MEASUREMENT SYSTEM ......................................................................................18

     6.1       Robot system specification ......................................................................................18
     6.2       Probe and amplifier specification .............................................................................19
     6.3       Phantoms.................................................................................................................19
     6.4       SAR measurement procedure .................................................................................20
Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 3 of 28


     6.5       SARA2 Interpolation and Extrapolation schemes ....................................................20
     6.6       Interpolation of 2D area scan...................................................................................21
     6.7       Extrapolation of 3D scan..........................................................................................21
     6.8       Interpolation of 3D scan and volume averaging.......................................................21


7 UNCERTAINTY ASSESSMENT.......................................................................................23

     7.1       Measurement Uncertainty Budget ...........................................................................24


8 TEST RESULTS SUMMARY............................................................................................25

     8.1       Conducted Output Power.........................................................................................25
     8.2       Test Positions and Configurations ...........................................................................25
     8.3       2450 MHz band Results for MN-720 in Dell Inspiron 4100......................................26
     8.4       2450 MHz band Results for MN-720 in Dell Latitude X300 .....................................26
     8.5       2450 MHz band Results for MN-720 in Toshiba Protégé 2000 ...............................26
     8.6       Validation Check Results .........................................................................................27


9 REFERENCES..................................................................................................................28




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 4 of 28




1 Assessment
The Microsoft Broadband Networking Wireless Notebook Adapter (Model number MN-
720) is in compliance with the exposure criteria specified in Federal Communications
Commission (FCC) Guidelines [FCC 2001] for uncontrolled exposure.




                          06/18/2003                                                                       06/18/2003
                          Pete Krebill                                                                   Lothar Schmidt
                          Project Leader                                                                  Test Lab Manager




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 5 of 28


2 Administrative Data

2.1      Identification of the Testing Laboratory Issuing the SAR Assessment Report

 Company Name:                                                              CETECOM Inc.
 Department:                                                                SAR
 Address:                                                                   411 Dixon Landing Road
                                                                            Milpitas, CA 95035
                                                                            U.S.A.
 Telephone:                                                                 +1 (408) 586 6200
 Fax:                                                                       +1 (408) 586 6299
 Responsible Test Lab Manager:                                              Lothar Schmidt


2.2      Identification of the Client

Applicant’s Name:                         Microsoft Corporation
Address:                                  One Microsoft Way
                                          Redmond, WA 98052-6399, USA
Contact Person:                           Stephen Stegner
Phone No.                                 +1 425 706 2697
Fax:                                      N/A
e-mail:                                   sstegner@microsoft.com



2.3      Identification of the Manufacturer

Manufacturer’s Name:                      Accton Technology Corporation
Address:                                  1, Creation 3rd Rd., Science-Based Industrial Park, Hsinchu, 300,
                                          Taiwan, R.O.C




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 6 of 28


3 Equipment under Investigation (EUI)

3.1      Identification of the Equipment under Investigation

 Product Type                                       The Microsoft Broadband Networking Wireless Notebook A
                                                    dapter (Model number MN-720)
 Model No:                                          MN-720
 FCC-ID:                                            C3KMN720
 Frequency Range:                                   2400-2483.5MHz
 Type(s) of Modulation:                             As per draft 802.11g specification (Direct sequence, OFDM)

 Number of Channels:                                11
 Antenna Type:                                      integral
 Output Power:                                      15 dBm (31.6mW) (average conducted packet power)




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 7 of 28




3.2      Front View of the Equipment under Investigation




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                  CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 8 of 28

4 Subject of Investigation
The Microsoft Broadband Networking Wireless Notebook Adapter (Model number MN-720)
from Microsoft Corporation operating in the 2400-2483.5MHz frequency ranges. The objective
of the measurements done by Cetecom Inc. was the dosimetric assessment of one device. The
tests were performed with the device configured in three different laptop computers. The tests
were performed in configurations for devices operated next to a person’s body. The examinations
were carried out with the dosimetric assessment system SARA2 described below.


4.1 The IEEE Standard C95.1 and the FCC Exposure Criteria
In the USA the recent FCC exposure criteria [FCC 2001] are based upon the IEEE Standard
C95.1 [IEEE 1999]. The IEEE standard C95.1 sets limits for human exposure to radio frequency
electromagnetic fields in the frequency range 3 kHz to 300 GHz.


4.2      Distinction Between Exposed Population, Duration of Exposure and
         Frequencies

The American Standard [IEEE 1999] distinguishes between controlled and uncontrolled
environment. Controlled environments are locations where there is exposure that may be incurred
by persons who are aware of the potential for exposure as a concomitant of employment or by
other cognizant persons. Uncontrolled environments are locations where there is the exposure of
individuals who have no knowledge or control of their exposure. The exposures may occur in
living quarters or workplaces. For exposure in controlled environments higher field strengths are
admissible. In addition the duration of exposure is considered. Due to the influence of frequency
on important parameters, as the penetration depth of the electromagnetic fields into the human
body and the absorption capability of different tissues, the
limits in general vary with frequency.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 9 of 28



4.3      Distinction between Maximum Permissible Exposure and SAR Limits

The biological relevant parameter describing the effects of electromagnetic fields in the
frequency range of interest is the specific absorption rate SAR (dimension: power/mass). It is a
measure of the power absorbed per unit mass. The SAR may be spatially averaged over the total
mass of an exposed body or its parts. The SAR is calculated from the r.m.s. electric field strength
E inside the human body, the conductivity σ and the mass density ρ of the biological tissue:

                  E2             ∂T
SAR =         σ   — =       c—
                    ρ            ∂t      t →0+


The specific absorption rate describes the initial rate of temperature rise ∂T / ∂t as a function of
the specific heat capacity c of the tissue. A limitation of the specific absorption rate prevents an
excessive heating of the human body by electromagnetic energy.

As it is sometimes difficult to determine the SAR directly by measurement (e.g. whole body
averaged SAR), the standard specifies more readily measurable maximum permissible exposures
in terms of external electric E and magnetic field strength H and power density S, derived from
the SAR limits. The limits for E, H and S have been fixed so that even under worst case
conditions, the limits for the specific absorption rate SAR are not exceeded.

For the relevant frequency range the maximum permissible exposure may be exceeded if the
exposure can be shown by appropriate techniques to produce SAR values below the
corresponding limits.


4.4      SAR Limit

In this report the comparison between the American exposure limits and the measured data is
made using the spatial peak SAR; the power level of the device under test guarantees that the
whole body averaged SAR is not exceeded.

Having in mind a worst case consideration, the SAR limit is valid for uncontrolled environment
and mobile respectively portable transmitters. According to Table 1 the SAR values have to be
averaged over a mass of 1 g (SAR1g) with the shape of a cube.

               Standard                                                           Status                 SAR limit (W/kg )
               IEEE C95.1                                                        In force                       1.6

Table 1: Relevant spatial peak SAR limit averaged over a mass of 1 g




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 10 of 28

5 The FCC Measurement Procedure
The Federal Communications Commission (FCC) has published a report and order on the 1st of
August 1996 [FCC 1996], which requires routine dosimetric assessment of mobile telecom-
communications devices, either by laboratory measurement techniques or by computational
modeling, prior to equipment authorization or use. In 2001 the Commission’s Office of
Engineering and Technology has released Edition 01-01 of Supplement C to OET Bulletin 65.
This revised edition, which replaces Edition 97-01, provides additional guidance and information
for evaluating compliance of mobile and portable devices with FCC limits for human exposure
to radiofrequency emissions [FCC 2001].


5.1      General Requirements

The test shall be performed in a laboratory with an environment which avoids influence on SAR
measurements by ambient EM sources and any reflection from the environment itself. The
ambient temperature shall be in the range of 20°C to 26°C and 30-70% humidity.


5.2      Device Operating Next to a Person’s Ear

Phantom Requirements

The phantom is a simplified representation of the human anatomy and comprised of material with
electrical properties similar to the corresponding tissues. The physical characteristics of the
phantom model shall resemble the head and the neck of a user since the shape is a dominant
parameter for exposure.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 11 of 28



5.3      Test Positions



5.4      Test positions of device relative to head

FCC’s OET Bulletin supplement C requires two test positions for the handset against the head
phantom, the “cheek” position and the “tilted” position. These two test positions are defined
below. The handset should be tested in both positions on the left and right sides of the SAM
phantom.
                              vertical
                                                                                                          vertical
                             center line
                                                                                                         center line
                      wt/2 wt/2                                                                    wt/2 wt/2


                       .                  horizontal
                                             line
                                                                              horizontal
                                                                                 line
                                                                                                         .
                                         A                                             A                        acoustic
   acoustic                                                                                                      output
    output                               B                                bottom of
                                              bottom of                   handset                                 B
                                              handset


                     wb/2 wb/2                                                                     wb/2 wb/2




Figure 1a – Handset vertical and                                           Figure 1b – Handset vertical and
horizontal reference lines – fixed case                                    horizontal reference lines – “clam-shell”


Definition of the “cheek” position
The “cheek” position is defined as follows:
      a) Ready the handset for talk operation.
      b) Define two imaginary lines on the handset: the vertical centerline and the horizontal line.
         The vertical centerline passes through two points on the front side of the handset: the
         midpoint of the width wt of the handset at the level of the acoustic output (point A on
         Figures 1a and 1b), and the midpoint of the width wb of the bottom of the handset (point
         B). The horizontal line is perpendicular to the vertical centerline and passes through the
         center of the acoustic output (see Figure 1a). The two lines intersect at point A. Note that
         for many handsets, point A coincides with the center of the acoustic output. However, the
         acoustic output may be located elsewhere on the horizontal line. Also note that the
         vertical centerline is not necessarily parallel to the front face of the handset (see Figure
         1b), especially for clamshell handsets, handsets with flip pieces, and other irregularly-
         shaped handsets.
Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                             CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 12 of 28

     c) Position the handset close to the surface of the phantom such that point A is on the
        (virtual) extension of the line passing through points RE and LE on the phantom (see
        Figure 2), such that the plane defined by the vertical center line and the horizontal line of
        the handset is approximately parallel to the sagittal plane of the phantom.
     d) Translate the handset towards the phantom along the line passing through RE and LE
        until the handset touches the pinna.
     e) While maintaining the handset in this plane, rotate it around the LE-RE line until the
        vertical centerline is in the plane normal to MB-NF including the line MB (called the
        reference plane).
     f) Rotate the handset around the vertical centerline until the handset (horizontal line) is
        symmetrical with respect to the line NF.
     g) While maintaining the vertical centerline in the reference plane, keeping point A on the
        line passing through RE and LE and maintaining the handset contact with the pinna, rotate
        the handset about the line NF until any point on the handset is in contact with a phantom
        point below the pinna (cheek). See Figure 2. The physical angles of rotation should be
        noted.




                                                                                 Figure 2
– Phone position 1, “cheek” or “touch” position. The reference points for the right ear
(RE), left ear (LE) and mouth (M), which define the reference plane for handset
positioning, are indicated.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 13 of 28




                                                                                     Figure 3
– Phone position 2, “tilted” position. The reference points for the right ear (RE), left ear
(LE) and mouth (M), which define the reference plane for handset positioning, are
indicated.


Definition of the “tilted” position
The “tilted” position is defined as follows:
     a) Repeat steps (a) – (g) of cheek position section above to place the device in the “cheek
        position.”
     b) While maintaining the orientation of the handset move the handset away from the pinna
        along the line passing through RE and LE in order to enable a rotation of the handset by
        15 degrees.
     c) Rotate the handset around the horizontal line by 15 degrees.
     d) While maintaining the orientation of the handset, move the handset towards the phantom
        on a line passing through RE and LE until any part of the handset touches the ear. The
        tilted position is obtained when the contact is on the pinna. If the contact is at any
        location other than the pinna, e.g., the antenna with the back of the phantom head, the
        angle of the handset should be reduced. In this case, the tilted position is obtained if any
        part of the handset is in contact with the pinna as well as a second part of the handset is
        contact with the phantom, e.g., the antenna with the back of the head.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 14 of 28



5.5      Test to be Performed

The SAR test shall be performed with both phone positions described above, on the left and right
side of the phantom. The device shall be measured for all modes operating when the device is
next to the ear, even if the different modes operate in the same frequency band.

For devices with retractable antenna the SAR test shall be performed with the antenna fully
extended and fully retracted. Other factors that may affect the exposure shall also be tested. For
example, optional antennas or optional battery packs which may significantly change the volume,
lengths, flip open/closed, etc. of the device, or any other accessories which might have the
potential to considerably increase the peak spatial-average SAR value.

The SAR test shall be performed at the high, middle and low frequency channels of each
operating mode. If the SAR measured at the middle channel for each test configuration is at least
2.0 dB lower than the SAR limit, testing at the high and low channels is optional.


5.6      Body-worn and Other Configurations

Phantom Requirements
For body-worn and other configurations a flat phantom shall be used which is comprised of
material with electrical properties similar to the corresponding tissues.

Test Position
The body-worn configurations shall be tested with the supplied accessories (belt-clips, holsters,
etc.) attached to the device in normal use configuration. Devices with a headset output shall be
tested with a connected headset.

Test to be Performed
For purpose of determining test requirements, accessories may be divided into two categories:
those that do not contain metallic components and those that do. For multiple accessories that do
not contain metallic components, the device may be tested only with that accessory which
provides the closest spacing to the body. For multiple accessories that contain metallic
components, the device must be tested with each accessory that contains a unique metallic
component. If multiple accessories share an identical metallic component, only the accessory that
provides the closest spacing to the body must be tested. If the manufacturer provides none body-
worn accessories a separation distance of 1.5 cm between the back of the device and the flat
phantom is recommended. Other separation distances may be used, but they shall not exceed 2.5
cm. In these cases, the device may use body-worn accessories that provide a separation distance
greater than that tested for the device provided however that the accessory contains no metallic
components.

For devices with retractable antenna the SAR test shall be performed with the antenna fully
extended and fully retracted. Other factors that may affect the exposure shall also be tested. For
example, optional antennas or optional battery packs which may significantly change the volume,
Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 15 of 28

lengths, flip open/closed, etc. of the device, or any other accessories which might have the
potential to considerably increase the peak spatial-average SAR value.


5.7      Procedure for assessing the peak spatial-average SAR

      Step 1: Power reference measurement:
      Prior to the SAR test, a local SAR measurement should be taken at a user-selected spatial
      reference point to monitor power variations during testing. For example, this power reference
      point can be spaced 10 mm or less in the normal direction from the liquid-shell interface and
      within ± 10 mm transverse to the normal line at the ear reference point.
      Step 2: Area scan
      The measurement procedures for evaluating SAR associated with wireless handsets typically
      start with a coarse measurement grid in order to determine the approximate location of the
      local peak SAR values. This is referred to as the "area scan" procedure. The SAR
      distribution is scanned along the inside surface of typically half of the head of the phantom
      but at least larger than the areas projected (normal to the phantom’s surface) by the handset
      and antenna. An example grid is given in Figure 4. The distance between the measured
      points and phantom surface should be less than 8 mm, and should remain constant (variation
      less than ± 1 mm) during the entire scan in order to determine the locations of the local peak
      SAR with sufficient precision. The distance between the measurement points should enable
      the detection of the location of local maximum with an accuracy of better than half the linear
      dimension of the tissue cube after interpolation. The resolution can also be tested using the
      functions in Annex E (see E.5.2). The approximate locations of the peak SARs should be
      determined from area scan. Since a given amplitude local peak with steep gradients may
      produce lower spatial-average SAR than slightly lower amplitude peaks with less steep
      gradients, it is necessary to evaluate the other peaks as well. However, since the spatial
      gradients of local SAR peaks are a function of wavelength inside the tissue simulating liquid
      and incident magnetic field strength, it is not necessary to evaluate peaks that are less than –
      2dB of the local maximum. Two-dimensional spline algorithms [Press, et al, 1996],
      [Brishoual, 2001] are typically used to determine the peaks and gradients within the scanned
      area. If the peak is closer than one-half of the linear dimension of the 1 g or 10 g tissue cube
      to the scan border, the measurement area should be enlarged if possible, e.g., by tilting the
      probe or the phantom (see Figure 5).




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 16 of 28




Figure 4 – Example of an area scan including the position of the handset. The scanned area
(white dots) should be larger than the area projected by the handset and antenna.

Step 3: Zoom scan
In order to assess the peak spatial SAR values averaged over a 1 g and 10 g cube, fine resolution
volume scans, called "zoom scans", are performed at the peak SAR locations determined during
the “area scan.” The zoom scan volume should have at least 1.5 times the linear dimension of
either a 1 g or a 10 g tissue cube for whichever peak spatial-average SAR is being evaluated. The
peak local SAR locations that were determined in the area scan (interpolated value) should be on
the centerline of the zoom scans. The centerline is the line that is normal to the surface and in the
center of the volume scan. If this is not possible, the zoom scan can be shifted but not by more
than half the dimension of the 1 g or a 10 g tissue cube.
The maximum spatial-average SAR is determined by a numerical analysis of the SAR values
obtained in the volume of the zoom scan, whereby interpolation (between measured points) and
extrapolation (between surface and closest measured points) routines should be applied. A 3-D-
spline algorithm [Press, et al, 1996], [Kreyszig, 1983], [Brishoual, 2001] can be used for
interpolation and a trapezoidal algorithm for the integration (averaging). Scan resolutions of
larger than 2 mm can be used provided the uncertainty is evaluated according to E (see E.5).
In some areas of the phantom, such as the jaw and upper head region, the angle of the probe with
respect to the line normal to the surface might become large, e.g., at angles larger than ± 30º (see
Figure 5), which may increase the boundary effect to an unacceptable level. In these cases, a
change in the orientation of the probe and/or the phantom is recommended during the zoom scan
so that the angle between the probe housing tube and the line normal to the surface is
significantly reduced (<30º).




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 17 of 28

      Step 4: Power reference measurement
      The local SAR should be measured at exactly the same location as in Step 1. The absolute
      value of the measurement drift (the difference between the SAR measured in Step 4 and Step
      1) should be recorded in the uncertainty budget. It is recommended that the drift be kept
      within ± 5%. If this is not possible, even with repeat testing, additional information may be
      used to demonstrate the power stability during the test. Power reference measurements can be
      taken after each zoom scan, if more than one zoom scan is needed. However, the drift should
      always be referred to the initial state with fully charged battery.


5.8      Determination of the largest peak spatial-average SAR

      In order to determine the largest value of the peak spatial-average SAR of a handset, all
      device positions, configurations and operational modes should be tested for each frequency
      band according to steps 1 to 3 below.
      Step 1: The tests of 6.4 should be conducted at the channel that is closest to the center of the
      transmit frequency band (fc) for:
      a) all device positions (cheek and tilt, for both left and right sides of the SAM phantom,
      b) all configurations for each device position in (a), e.g. antenna extended and retracted, and
      c) all operational modes for each device position in (a) and configuration in (b) in each
         frequency band, e.g. analog and digital.
      If more than three frequencies need to be tested, (i.e., Nc > 3), then all frequencies,
      configurations and modes must be tested for all of the above positions.
      Step 2: For the condition providing highest spatial peak SAR determined in Step 1 conduct
      all tests of 6.4 at all other test frequencies, e.g. lowest and highest frequencies. In addition,
      for all other conditions (device position, configuration and operational mode) where the
      spatial peak SAR value determined in Step 1 is within 3dB of the applicable SAR limit, it is
      recommended that all other test frequencies should be tested as well1.
      Step 3: Examine all data to determine the largest value of the peak spatial-average SAR
      found in Steps 1 to 2.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 18 of 28

6 The Measurement System

6.1      Robot system specification

The SAR measurement system being used is the IndexSAR SARA2 system, which consists of a
Mitsubishi RV-E2 6-axis robot arm and controller, IndexSAR probe and amplifier and SAM
phantom Head Shape. The robot is used to articulate the probe to programmed positions inside
the phantom head to obtain the SAR readings from the DUT.

The system is controlled remotely from a PC, which contains the software to control the robot
and data acquisition equipment. The software also displays the data obtained from test scans.




                           Figure 5: Schematic diagram of the SAR measurement system

The position and digitised shape of the phantom heads are made available to the software for
accurate positioning of the probe and reduction of set-up time.
The SAM phantom heads are individually digitised using a Mitutoyo CMM machine to a
precision of 0.001mm. The data is then converted into a shape format for the software,
providing an accurate description of the phantom shell.

In operation, the system first does an area (2D) scan at a fixed depth within the liquid from the
inside wall of the phantom. When the maximum SAR point has been found, the system will then
carry out a 3D scan centred at that point to determine volume averaged SAR level.

Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 19 of 28

6.2      Probe and amplifier specification


IXP-050 Indexsar isotropic immersible SAR probe
The probes are constructed using three orthogonal dipole sensors arranged on an interlocking,
triangular prism core. The probes have built-in shielding against static charges and are contained
within a PEEK cylindrical enclosure material at the tip. Probe calibration is described in the
probe’s calibration certificate (see appendix C.). The system uses diode compression potential
(DCP) to determine SAR values for different types of modulation. Crest factor is not used for
determining SAR values. The DCP for different types of modulation is determined during the
probe calibration procedure. For a more detailed explanation see IndexSAR Immesible SAR Probe
Calibration Report included in Appendix C of this report.

IXP-010 Amplifier
The amplifier unit has a multi-pole connector to connect to the probe and a multiplexer selects
between the 3-channel single-ended inputs. A 16-bit AtoD converter with programmable gain is
used along with an on-board micro-controller with non-volatile firmware. Battery life is around
150 hours and data are transferred to the PC via 3m of duplex optical fibre and a self-powered
RS232 to optical converter.


6.3      Phantoms

The Specific Anthropomorphic Mannequin (SAM) Upright Phantom is fabricated using moulds
generated from the CAD files as specified by CENELEC EN50361. It is mounted via a rotation
base to a supporting table, which also holds the robotic positioner. The phantom and robot
alignment is assured by both mechanical and laser registration systems. The box phantom used
for body testing and for validation is manufactured from Perspex. The material is 2 mm in
thickness on the test surfaces and 4 mm in thickness on the other surfaces. Its dimensions are:
X=21 cm., Y=20.5 cm., Z=16 cm.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 20 of 28



6.4      SAR measurement procedure




                      Figure 6: Principal components of the SAR measurement test bench

The major components of the test bench are shown in the picture above. A test set and dipole
antenna control the handset via an air link and a low-mass phone holder can position the phone at
either ear. Graduated scales are provided to set the phone in the 15 degree position. The upright
phantom head holds approx. 7 litres of simulant liquid. The phantom is filled and emptied
through a 45mm diameter penetration hole in the top of the head.

After an area scan has been done at a fixed distance of 8mm from the surface of the phantom on
the source side, a 3D scan is set up around the location of the maximum spot SAR. First, a point
within the scan area is visited by the probe and a SAR reading taken at the start of testing. At the
end of testing, the probe is returned to the same point and a second reading is taken. Comparison
between these start and end readings enables the power drift during measurement to be assessed.


6.5      SARA2 Interpolation and Extrapolation schemes

(see support document IXS-0202)
SARA2 software contains support for both 2D cubic B-spline interpolation as well as 3D cubic
B-spline interpolation. In addition, for extrapolation purposes, a general n-th order polynomial
fitting routine is implemented following a singular value decomposition algorithm presented in
[4]. A 4th order polynomial fit is used by default for data extrapolation, but a linear-logarithmic
fitting function can be selected as an option. The polynomial fitting procedures have been tested
by comparing the fitting coefficients generated by the SARA2 procedures with those obtained
using the polynomial fit functions of Microsoft Excel when applied to the same test input data.


Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 21 of 28

6.6      Interpolation of 2D area scan

The 2D cubic B-spline interpolation is used after the initial area scan at fixed distance from the
phantom shell wall. The initial scan data are collected with approx. 10mm spatial resolution and
spline interpolation is used to find the location of the local maximum to within a 1mm resolution
for positioning the subsequent 3D scanning.


6.7      Extrapolation of 3D scan

For the 3D scan, data are collected on a spatially regular 3D grid having (by default) 6.4 mm
steps in the lateral dimensions and 3.5 mm steps in the depth direction (away from the source).
SARA2 enables full control over the selection of alternative step sizes in all directions.

The digitised shape of the head is available to the SARA2 software, which decides which points
in the 3D array are sufficiently well within the shell wall to be ‘visited’ by the SAR probe. After
the data collection, the data are extrapolated in the depth direction to assign values to points in
the 3D array closer to the shell wall. A notional extrapolation value is also assigned to the first
point outside the shell wall so that subsequent interpolation schemes will be applicable right up to
the shell wall boundary.



6.8      Interpolation of 3D scan and volume averaging

The procedure used for defining the shape of the volumes used for SAR averaging in the SARA2
software follow the method of adapting the surface of the ‘cube’ to conform with the curved
inner surface of the phantom (see Appendix D in FCC Supplement C edition 01-01 to OET
Bulletin 65 edition 97-01). This is called, here, the conformal scheme.

For each row of data in the depth direction, the data are extrapolated and interpolated to less than
1mm spacing and average values are calculated from the phantom surface for the row of data
over distances corresponding to the requisite depth for 10g and 1g cubes. This results in two 2D
arrays of data, which are then cubic B-spline interpolated to sub mm lateral resolution. A search
routine then moves an averaging square around through the 2D array and records the maximum
value of the corresponding 1g and 10g volume averages. For the definition of the surface in this
procedure, the digitised position of the headshell surface is used for measurement in head-shaped
phantoms. For measurements in rectangular, box phantoms, the distance between the phantom
wall and the closest set of gridded data points is entered into the software. For measurements in
box-shaped phantoms, this distance is under the control of the user. The effective distance must
be greater than 2.5mm as this is the tip-sensor distance and to avoid interface proximity effects, it
should be at least 5mm. A value of 6 or 8mm is recommended. This distance is called dbe.

For automated measurements inside the head, the distance cannot be less than 2.5mm, which is
the radius of the probe tip and to avoid interface proximity effects, a minimum clearance distance
of x mm is retained. The actual value of dbe will vary from point to point depending upon how
the spatially-regular 3D grid points fit within the shell. The greatest separation is when a grid
Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 22 of 28

point is just not visited due to the probe tip dimensions. In this case the distance could be as large
as the step-size plus the minimum clearance distance (i.e with x=5 and a step size of 3.5, dbe will
be between 3.5 and 8.5mm).

The default step size (dstep) used is 3.5mm, but this is under user-control. The compromise is
with time of scan, so it is not practical to make it much smaller or scan times become long and
power-drop influences become larger.

The robot positioning system specification for the repeatability of the positioning (dss) is +/-
0.04mm.

The phantom shell is made by an industrial moulding process from the CAD files of the SAM
shape, with both internal and external moulds. For the upright phantoms, the external shape is
subsequently digitised on a Mitutoyo CMM machine (Euro an ultrasonic sensor indicate that the
shell thickness (dph) away from the ear is 2.0 +/- 0.1mm. The ultrasonic measurements were
calibrated using additional mechanical measurements on available cut surfaces of the phantom
shells. See support document IXS-020x.

For the upright phantom, the alignment is based upon registration of the rotation axis of the
phantom on its 253mm diameter baseplate bearing and the position of the probe axis when
commanded to go to the axial position. A laser alignment tool is provided (procedure detailed
elsewhere). This enables the registration of the phantom tip (dmis) to be assured to within
approx. 0.2mm. This alignment is done with reference to the actual probe tip after installation and
probe alignment. The rotational positioning of the phantom is variable – offering advantages for
special studies, but locating pins ensure accurate repositioning at the principal positions (LH and
RH ears).




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 23 of 28

7 Uncertainty Assessment
A measurement uncertainty assessment has been undertaken following guidance given in draft
IEEE 1528. Indexsar Ltd has supplied a generic uncertainty analysis for the SARA2 system in the
form of a spreadsheet and the supporting assessments are documented in an Indexsar document
IXS-2028. Additionally, uncertainties resulting from the probe positioning system and the upright
phantom geometry are discussed in additional documents.

Some of the uncertainty contributions are site-specific and, for these, Cetecom, Inc. have assessed
the uncertainty contributions arising from local environmental and procedural factors.

The resultant uncertainty budget, following the assessment template given in draft IEEE 1528 is
shown below:




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                                 Page 24 of 28


7.1        Measurement Uncertainty Budget

                                                                                                                                        Standard
                                                                                Probability     Divisor      Divisor                                  Stand      Stand
      Error Sources              Description          Uncertainty                                                        ci     ci^2   Uncertainty
                                                                                Distribution   (descrip)     (value)                                 Uncert^2   Uncert^2
                                                                                                                                           (%)


                                               (dB)                 (%)

Measurement equipment

         Calibration               7.2.1.1                          10            Normal        1 or k          2        1       1        5.00        25.00      25.00
          Isotropy                 7.2.1.2     0.5      12.20    5.30           Rectangular       √3          1.73      1.00     1        3.06         9.36       9.36
        Linearity                  7.2.1.3     0.04     0.93     2.92           Rectangular       √2          1.73      1.00     1        1.69         2.84       2.84
     Probe Stability                                             2.50           Rectangular       √3          1.73      1.00     1        1.44         2.08       2.08
     Detection limits              7.2.1.4                         0            Rectangular       √3          1.73      1.00     1        0.00         0.00       0.00
    Boundary effects               7.2.1.5                        1.7           Rectangular       √3          1.73      1.00     1        0.98         0.96       0.96
   Measurement device              7.2.1.6                         0              Normal        1 or k        1.96      1.00     1        0.00         0.00       0.00
     Response time                 7.2.1.7                         0              Normal          1           1.00      1.00     1        0.00         0.00       0.00
          Noise                    7.2.1.8                         0              Normal          1           1.00      1.00     1        0.00         0.00       0.00
    Intergration time              7.2.1.9                        0.4             Normal          1           1.00      1.00     1        0.40         0.16       0.16



 Mechanical Constraints

     Scanning system               7.2.2.1                       0.57           Rectangular       √3          1.73      1.00     1        0.33         0.11       0.11
       Phantom shell               7.2.2.2                       1.43           Rectangular       √3          1.73      1.00     1        0.83         0.68       0.68
 Matching between probe
                                   7.2.2.3                       2.86           Rectangular       √3          1.73      1.00     1        1.65         2.73       2.73
      and phantom
  Positioning of the phone         7.2.2.4                          10            Normal          1             1       1.00     1       10.00        100.00     100.00


   Physical parameters

     Liquid conductivity
                                   7.2.3.2                          5           Rectangular       √3          1.73      0.50    0.25      1.44         2.08       0.52
   (Deviation from target)

     Liquid conductivity
                                   7.2.3.2                          5           Rectangular       √3          1.73      0.50    0.25      1.44         2.08       0.52
    (measurement error)

Liquid permittivity (Deviation
                                   7.2.3.3                          5           Rectangular       √3          1.73      0.50    0.25      1.44         2.08       0.52
        from target)

     Liquid permittivity
                                   7.2.3.3                          5           Rectangular       √3          1.73      0.50    0.25      1.44         2.08       0.52
    (measurement error)

Drifts in output power of the
phone, probe, temperature          7.2.3.4                          5           Rectangular       √3          1.73      1.00     1        2.89         8.33       8.33
         and humidity
    Perturbation by the
                                   7.2.3.5                          3           Rectangular       √3          1.73      1.00     1        1.73         3.00       3.00
       environment


     Post processing

   SAR interpolation and
                                   7.2.4.1                          8           Rectangular       √3          1.73      1.00     1        4.62        21.33      21.33
      extrapolation
 Maximum SAR evaluation            7.2.4.2                          2.4         Rectangular       √3          1.73      1.00     1        1.39         1.92       1.92


       Other factors
                                                                    0             Normal        1 or k        1.96      1.00     1        0.00         0.00       0.00
                                                                    0             Normal        1 or k        1.96      1.00     1        0.00         0.00       0.00
                                                                    0           Rectangular       √3          1.73      1.00     1        0.00         0.00       0.00
                                                                    0             Normal          1           1.00      1.00     1        0.00         0.00       0.00


                                                                          m
   Combined standard
      uncertainty                                               uc = √ ∑ ci2 . ui2                                      13.67

                                                                          i=1

  Expanded uncertainty
  (confidence interval of
          95%)                                                  Normal              k=           1.96        ue=k* uc   26.79    %



Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                                 CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 25 of 28

8 Test results summary

The highest SAR measured is shown highlited below.


8.1      Conducted Output Power

Prior to testing the conducted RF output power was measured. The results are shown below.

           2450 MHz band:
           Power levels were set to >15 dBm average power in the packet for all channels prior to
           each SAR test.


8.2      Test Positions and Configurations

The EUT was configured in three different laptop PCs. In each laptop PC tests were performed
with the bottom of the PC touching the phantom (lap position) and with the edge of the EUT
5 mm from the phantom (bystander position). The EUT was installed in the PCMCIA slot nearest
the bottom of the PC. The distance from the bottom of the PCMCIA slot to the bottom of the PC
is shown in the far right column below. This distance includes the feet on the bottom of the PC.

                                                                                                                  Distance slot
                                   Manufacturer                       Model                              FCC ID   to PC bottom
Laptop PC#1:                       Dell                               Inspiron 4100                      DOC      6 mm
Laptop PC#2:                       Dell                               Latitude X300                      DOC      10 mm
Laptop PC#3:                       Toshiba                            Protégé 2000                       DOC      2 mm




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 26 of 28



8.3      2450 MHz band Results for MN-720 in Dell Inspiron 4100

      PC                 Position                Device Channel               Max. 1g          Area scan   Positioning
                                                    Number /                   SAR               (See       photo (See
                                                Frequency (MHz)               (W/kg)           Appendix    Appendix B)
                                                                                                  A)
       #1          Bystander 5 mm                      6 / 2437                 0.056           Plot 1     Photo 3 & 4
       #1                   Lap                        6 / 2437                 0.248             Plot 2     Photo 5
       #1                   Lap                        1 / 2412                 0.333             Plot 3     Photo 5
       #1                   Lap                       11 / 2462                 0.149             Plot 4     Photo 5



8.4      2450 MHz band Results for MN-720 in Dell Latitude X300

      PC                 Position                Device Channel               Max. 1g          Area scan   Positioning
                                                    Number /                   SAR               (See       photo (See
                                                Frequency (MHz)               (W/kg)           Appendix    Appendix B)
                                                                                                  A)
       #2          Bystander 5 mm                      6 / 2437                 0.049           Plot 5     Photo 6 & 7
       #2                   Lap                        6 / 2437                 0.146             Plot 6     Photo 8
       #2                   Lap                        1 / 2412                 0.159             Plot 7     Photo 8
       #2                   Lap                       11 / 2462                 0.136             Plot 8     Photo 8



8.5      2450 MHz band Results for MN-720 in Toshiba Protégé 2000

      PC                 Position                Device Channel               Max. 1g          Area scan   Positioning
                                                    Number /                   SAR               (See       photo (See
                                                Frequency (MHz)               (W/kg)           Appendix    Appendix B)
                                                                                                  A)
       #3          Bystander 5 mm                      6 / 2437                 0.076           Plot 9     Photo 9 & 10
       #3                   Lap                        6 / 2437                 0.585            Plot 10     Photo 11
       #3                   Lap                        1 / 2412                 0.693            Plot 11     Photo 11
       #3                   Lap                       11 / 2462                 0.591            Plot 12     Photo 11




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 27 of 28

[W/kg]]) Temperature


8.6      Validation Check Results

Prior to formal testing at each frequency a system verification was performed in accordance with
IEEE 1528. The balanced dipole source was placed at the specified distance in horizontal
orientation. All of the testing described in this report was performed within 24 hours of the
system verification. The following results were obtained:

                                                                                                                        Difference
                                                                                   Max                       1 Watt
                                                                                                                         reference
                                               CW input                Max     measured1g                   reference
                                                                                                                        SAR value
                       Frequency               at dipole            measured1g    SAR                      SAR value
      Date                                                                                                                   to
                         (MHz)                    feed                 SAR     normalized                  from IEEE
                                                                                                                        normalized
                                                (Watts)              (W/kg)     to 1 Watt                     1528
                                                                                                                           SAR
                                                                                 (W/kg)                      (W/kg)
06/05/2003                  2450                    0.5                 26.365                  52.73        52.4        +0.63%

[W/k
g] (




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.


SAR Test Report No. SAR_499_2003_FCC_2450
Date of Report: 06/18/2003                                                                 Page 28 of 28

9 References
 [FCC 2001] Federal Communications Commission: Evaluating Compliance with FCC
Guidelines for Human Exposure to Radiofrequency Electromagnetic Fields,
Supplement C (Edition 01-01) to OET Bulletin 65 (Edition 97-01), FCC, 2001.
[IEEE 1999] IEEE Std C95.1-1999: IEEE Standard for Safety Levels with Respect to
Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to
300 GHz, Inst. of Electrical and Electronics Engineers, Inc., 1999.
[IEEE 200x] IEEE Std 1528-200x: DRAFT Recommended Practice for Determining the
Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Body
Due to Wireless Communications Devices: Experimental Techniques. Draft
6.2, Inst. of Electrical and Electronics Engineers, Inc., 2000.
[NIST 1994] NIST: Guidelines for Evaluating and Expressing the Uncertainty of NIST
Measurement Results, Technical Note 1297 (TN1297), United States
Department of Commerce Technology Administration, National Institute of
Standards and Technology, 1994.




Error!   Reference This report shall not be reproduced except in full without the written approval of:
source not found.
                                            CETECOM Inc. Š SAR Š 411 Dixon Landing Road Š Milpitas, CA 95035 Š U.S.A.



Document Created: 2003-06-18 16:35:20
Document Modified: 2003-06-18 16:35:20

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC