12802195_App B_SAR System Check Plots

FCC ID: C3K1876

RF Exposure Info

Download: PDF
FCCID_4447890

Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 7/16/2019 7:25:35 PM

2019-07-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5250 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5250 MHz; σ = 5.343 S/m; εr = 48.993; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(4.33, 4.33, 4.33) @ 5250 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/5.25 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 19.7 W/kg

Body/5.25 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 53.64 V/m; Power Drift = -0.16 dB
Peak SAR (extrapolated) = 32.6 W/kg
SAR(1 g) = 8.13 W/kg; SAR(10 g) = 2.28 W/kg
Maximum value of SAR (measured) = 19.8 W/kg




                      0 dB = 19.8 W/kg = 12.97 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 7/16/2019 7:45:53 PM

2019-07-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5250 MHz; Duty Cycle: 1:1

Body/5.25 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 13.5 W/kg




                                                                                             Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 7/22/2019 8:09:30 PM

2019-07-22_SystemPerformanceCheck-D2450V2 SN 748
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.022 S/m; εr = 51.961; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2450 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.40 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.39 V/m; Power Drift = -0.00 dB
Peak SAR (extrapolated) = 11.2 W/kg
SAR(1 g) = 5.44 W/kg; SAR(10 g) = 2.51 W/kg
Maximum value of SAR (measured) = 7.76 W/kg




                      0 dB = 7.76 W/kg = 8.90 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 7/22/2019 8:24:19 PM

2019-07-22_SystemPerformanceCheck-D2450V2 SN 748
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.75 W/kg




                                                                                             Plot No. 4


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/2/2019 2:06:22 PM

20190802_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.433 S/m; εr = 53.616; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.92, 7.92, 7.92) @ 1750 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.09 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 59.76 V/m; Power Drift = -0.12 dB
Peak SAR (extrapolated) = 6.61 W/kg
SAR(1 g) = 3.67 W/kg; SAR(10 g) = 1.95 W/kg
Maximum value of SAR (measured) = 4.92 W/kg




                      0 dB = 4.92 W/kg = 6.92 dBW/kg




                                                                                                               Plot No. 5


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/2/2019 2:20:16 PM

20190802_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.91 W/kg




                                                                                            Plot No. 6


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/2/2019 2:30:42 PM

2019-08-02_SystemPerformanceCheck-D1900V2 SN 5d140
Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.549 S/m; εr = 53.381; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.6, 7.6, 7.6) @ 1900 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.93 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.54 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 7.98 W/kg
SAR(1 g) = 4.37 W/kg; SAR(10 g) = 2.27 W/kg
Maximum value of SAR (measured) = 5.92 W/kg




                      0 dB = 5.92 W/kg = 7.72 dBW/kg




                                                                                                               Plot No. 7


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/2/2019 2:44:37 PM

2019-08-02_SystemPerformanceCheck-D1900V2 SN 5d140
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.91 W/kg




                                                                                            Plot No. 8


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/2/2019 3:53:15 PM

2019-08-02_SystemPerformanceCheck-D2600V2 SN 1036
Frequency: 2600 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2600 MHz; σ = 2.183 S/m; εr = 52.432; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.29, 7.29, 7.29) @ 2600 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.52 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.94 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 11.1 W/kg
SAR(1 g) = 5.16 W/kg; SAR(10 g) = 2.3 W/kg
Maximum value of SAR (measured) = 7.47 W/kg




                      0 dB = 7.47 W/kg = 8.73 dBW/kg




                                                                                                               Plot No. 9


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/2/2019 4:08:28 PM

2019-08-02_SystemPerformanceCheck-D2600V2 SN 1036
Frequency: 2600 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.47 W/kg




                                                                                            Plot No. 10


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/6/2019 8:48:52 AM

2019-08-06_SystemPerformanceCheck-D1900V2 SN 5d043
Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.57 S/m; εr = 55.202; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.6, 7.6, 7.6) @ 1900 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.93 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.39 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 8.06 W/kg
SAR(1 g) = 4.41 W/kg; SAR(10 g) = 2.29 W/kg
Maximum value of SAR (measured) = 5.98 W/kg




                      0 dB = 5.98 W/kg = 7.77 dBW/kg




                                                                                                               Plot No. 11


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/6/2019 9:02:44 AM

2019-08-06_SystemPerformanceCheck-D1900V2 SN 5d043
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.95 W/kg




                                                                                            Plot No. 12


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/7/2019 5:46:18 PM

20190807_SystemPerformanceCheck-D1750V2 SN 1053
Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.459 S/m; εr = 51.931; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.92, 7.92, 7.92) @ 1750 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.24 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.59 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 7.06 W/kg
SAR(1 g) = 3.89 W/kg; SAR(10 g) = 2.05 W/kg
Maximum value of SAR (measured) = 5.25 W/kg




                      0 dB = 5.25 W/kg = 7.20 dBW/kg




                                                                                                               Plot No. 13


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/7/2019 6:00:36 PM

20190807_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.25 W/kg




                                                                                            Plot No. 14


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/16/2019 3:46:25 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5750 MHz; σ = 5.926 S/m; εr = 47.422; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(4.16, 4.16, 4.16) @ 5750 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/5.75 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 21.1 W/kg

Body/5.75 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 52.61 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 36.9 W/kg
SAR(1 g) = 8.03 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 19.9 W/kg




                      0 dB = 19.9 W/kg = 12.99 dBW/kg




                                                                                                               Plot No. 15


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/16/2019 4:03:03 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5750 MHz; Duty Cycle: 1:1

Body/5.75 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 14.3 W/kg




                                                                                         Plot No. 16


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/16/2019 7:17:29 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5600 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5600 MHz; σ = 5.718 S/m; εr = 47.692; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(3.9, 3.9, 3.9) @ 5600 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/5.6 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 18.6 W/kg

Body/5.6 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 50.52 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 32.1 W/kg
SAR(1 g) = 7.4 W/kg; SAR(10 g) = 2.06 W/kg
Maximum value of SAR (measured) = 17.8 W/kg




                      0 dB = 17.8 W/kg = 12.50 dBW/kg




                                                                                                               Plot No. 17


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/16/2019 7:33:57 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5600 MHz; Duty Cycle: 1:1

Body/5.6 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 13.0 W/kg




                                                                                         Plot No. 18


Test Laboratory: UL Verification Services Inc. SAR Lab 1                    Date/Time: 8/21/2019 10:11:45 PM

2019-08-21_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.014 S/m; εr = 50.741; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1544; Calibrated: 3/19/2019
- Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2450 MHz; Calibrated: 9/18/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm
(Mechanical Surface Detection)
- Phantom: ELI Front; Type: QDOVA002AA; Serial: TP:2085

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.08 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 63.20 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 11.4 W/kg
SAR(1 g) = 5.45 W/kg; SAR(10 g) = 2.51 W/kg
Maximum value of SAR (measured) = 7.77 W/kg




                      0 dB = 7.77 W/kg = 8.90 dBW/kg




                                                                                                           Plot No. 19


Test Laboratory: UL Verification Services Inc. SAR Lab 1   Date/Time: 8/21/2019 10:24:16 PM

2019-08-21_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 8.00 W/kg




                                                                                         Plot No. 20


Test Laboratory: UL Verification Services Inc. SAR Lab 2                    Date/Time: 8/5/2019 10:16:34 AM

20190805_SystemPerformanceCheck-D750V3 SN 1024
Frequency: 750 MHz; Duty Cycle: 1:1;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 750 MHz; σ = 0.994 S/m; εr = 53.048; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/15/2019
- Probe: EX3DV4 - SN7483; ConvF(10, 10, 10) @ 750 MHz; Calibrated: 11/14/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm
(Mechanical Surface Detection)
- Phantom: ELI V8.0 (20deg probe tilt) Front; Type: QD OVA 004 AA; Serial: 2081

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.12 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 34.04 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 1.40 W/kg
SAR(1 g) = 0.935 W/kg; SAR(10 g) = 0.616 W/kg
Maximum value of SAR (measured) = 1.14 W/kg




                      0 dB = 1.14 W/kg = 0.57 dBW/kg




                                                                                                           Plot No. 21


Test Laboratory: UL Verification Services Inc. SAR Lab 2   Date/Time: 8/5/2019 10:28:30 AM

20190805_SystemPerformanceCheck-D750V3 SN 1024
Frequency: 750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.12 W/kg




                                                                                         Plot No. 22


Test Laboratory: UL Verification Services Inc. SAR Lab 2                    Date/Time: 8/6/2019 8:36:03 AM

20190806_SystemPerformanceCheck-D835V2 SN 4d002
Frequency: 835 MHz; Duty Cycle: 1:1;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 1.017 S/m; εr = 54.034; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/15/2019
- Probe: EX3DV4 - SN7483; ConvF(9.67, 9.67, 9.67) @ 835 MHz; Calibrated: 11/14/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI V8.0 (20deg probe tilt) Front; Type: QD OVA 004 AA; Serial: 2081

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.23 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 34.56 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 1.54 W/kg
SAR(1 g) = 1.02 W/kg; SAR(10 g) = 0.664 W/kg
Maximum value of SAR (measured) = 1.25 W/kg




                      0 dB = 1.25 W/kg = 0.97 dBW/kg




                                                                                                               Plot No. 23


Test Laboratory: UL Verification Services Inc. SAR Lab 2   Date/Time: 8/6/2019 8:49:59 AM

20190806_SystemPerformanceCheck-D835V2 SN 4d002
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.20 W/kg




                                                                                            Plot No. 24


Test Laboratory: UL Verification Services Inc. SAR Lab 2                    Date/Time: 8/16/2019 10:40:03 AM

20190816_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.036 S/m; εr = 52.292; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/15/2019
- Probe: EX3DV4 - SN7483; ConvF(7.97, 7.97, 7.97) @ 2450 MHz; Calibrated: 11/14/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI V8.0 (20deg probe tilt) Front; Type: QD OVA 004 AA; Serial: 2081

Body/Pin=100 mW 2 2/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.10 W/kg

Body/Pin=100 mW 2 2/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 59.79 V/m; Power Drift = 0.19 dB
Peak SAR (extrapolated) = 11.2 W/kg
SAR(1 g) = 5.48 W/kg; SAR(10 g) = 2.51 W/kg
Maximum value of SAR (measured) = 7.79 W/kg




                      0 dB = 7.79 W/kg = 8.92 dBW/kg




                                                                                                               Plot No. 25


Test Laboratory: UL Verification Services Inc. SAR Lab 2    Date/Time: 8/16/2019 10:55:08 AM

20190816_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW 2 2/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.52 W/kg




                                                                                          Plot No. 26


Test Laboratory: UL Verification Services Inc. SAR Lab 2                    Date/Time: 8/16/2019 1:43:32 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5250 MHz; Duty Cycle: 1:1;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5250 MHz; σ = 5.358 S/m; εr = 47.205; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1359; Calibrated: 2/15/2019
- Probe: EX3DV4 - SN7483; ConvF(4.92, 4.92, 4.92) @ 5250 MHz; Calibrated: 11/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI V8.0 (20deg probe tilt) Front; Type: QD OVA 004 AA; Serial: 2081

Body/5.25 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 17.1 W/kg

Body/5.25 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 52.35 V/m; Power Drift = -0.14 dB
Peak SAR (extrapolated) = 30.8 W/kg
SAR(1 g) = 7.97 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 18.5 W/kg




                      0 dB = 18.5 W/kg = 12.67 dBW/kg




                                                                                                               Plot No. 27


Test Laboratory: UL Verification Services Inc. SAR Lab 2   Date/Time: 8/16/2019 2:00:12 PM

2019-08-16_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5250 MHz; Duty Cycle: 1:1

Body/5.25 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 12.3 W/kg




                                                                                         Plot No. 28