11735596-S1V1 SAR_App B System Check Plots

FCC ID: C3K1807

RF Exposure Info

Download: PDF
FCCID_3592313

Test Laboratory: UL Verification Services Inc., SAR Lab 2                    Date/Time: 8/17/2017 6:03:45 PM

20170817_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.029 S/m; εr = 51.352; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/8/2017
- Probe: EX3DV4 - SN3991; ConvF(7.95, 7.95, 7.95); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI B v4.0; Type: QDOVA002AA; Serial: 1196

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 7.26 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 52.051 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 11.3 W/kg
SAR(1 g) = 5.46 W/kg; SAR(10 g) = 2.52 W/kg
Maximum value of SAR (measured) = 7.81 W/kg




                      0 dB = 7.81 W/kg = 8.93 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc., SAR Lab 2   Date/Time: 8/17/2017 6:32:11 PM

20170817_SystemPerformanceCheck-D2450V2 SN 899
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.43 W/kg




                                                                                              Plot No. 2


Test Laboratory: UL Verification Services Inc., SAR Lab 2                    Date/Time: 9/11/2017 10:09:14 PM

20170911_SystemPerformanceCheck-D2300V2 SN 1002
Frequency: 2300 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2300 MHz; σ = 1.834 S/m; εr = 51.851; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/8/2017
- Probe: EX3DV4 - SN3991; ConvF(8.05, 8.05, 8.05); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI B v4.0; Type: QDOVA002AA; Serial: 1196

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.80 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 50.265 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 8.87 W/kg
SAR(1 g) = 4.43 W/kg; SAR(10 g) = 2.11 W/kg
Maximum value of SAR (measured) = 6.23 W/kg




                      0 dB = 6.23 W/kg = 7.94 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc., SAR Lab 2   Date/Time: 9/11/2017 10:37:29 PM

20170911_SystemPerformanceCheck-D2300V2 SN 1002
Frequency: 2300 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.62 W/kg




                                                                                           Plot No. 4


Test Laboratory: UL Verification Services Inc., SAR Lab 2                    Date/Time: 9/22/2017 2:12:16 PM

20170922_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5600 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5600 MHz; σ = 5.713 S/m; εr = 49.128; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1433; Calibrated: 3/8/2017
- Probe: EX3DV4 - SN3991; ConvF(4.07, 4.07, 4.07); Calibrated: 5/30/2017;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI B v4.0; Type: QDOVA002AA; Serial: 1196

Body/5.6 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 17.9 W/kg

Body/5.6 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 41.015 V/m; Power Drift = -0.19 dB
Peak SAR (extrapolated) = 36.0 W/kg
SAR(1 g) = 8.27 W/kg; SAR(10 g) = 2.27 W/kg
Maximum value of SAR (measured) = 20.3 W/kg




                      0 dB = 20.3 W/kg = 13.07 dBW/kg




                                                                                                               Plot No. 5


Test Laboratory: UL Verification Services Inc., SAR Lab 2   Date/Time: 9/22/2017 2:38:07 PM

20170922_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5600 MHz; Duty Cycle: 1:1

Body/5.6 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 0.688 W/kg




                                                                                              Plot No. 6


Test Laboratory: UL Verification Services Inc. SAR Lab 3                     Date/Time: 9/5/2017 10:49:10 AM

20170905_SystemPerformanceCheck-D835V2 SN 4d117
Frequency: 835 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 835 MHz; σ = 1.018 S/m; εr = 55.407; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(10.28, 10.28, 10.28); Calibrated: 5/30/2017;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI B v5.0; Type: QD OVA 002 AA; Serial: 1196

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.41 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 35.638 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 1.61 W/kg
SAR(1 g) = 1.07 W/kg; SAR(10 g) = 0.704 W/kg
Maximum value of SAR (measured) = 1.43 W/kg




                      0 dB = 1.43 W/kg = 1.55 dBW/kg




                                                                                                               Plot No. 7


Test Laboratory: UL Verification Services Inc. SAR Lab 3    Date/Time: 9/5/2017 11:07:43 AM

20170905_SystemPerformanceCheck-D835V2 SN 4d117
Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.27 W/kg




                                                                                              Plot No. 8


Test Laboratory: UL Verification Services Inc. SAR Lab 3                     Date/Time: 9/7/2017 10:43:07 AM

20170907_SystemPerformanceCheck-D1900V2 SN 5d140
Frequency: 1900 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1900 MHz; σ = 1.574 S/m; εr = 50.958; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(8.23, 8.23, 8.23); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI A v5.0; Type: QD OVA 002 AA; Serial: 1194

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.25 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.138 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 7.25 W/kg
SAR(1 g) = 3.91 W/kg; SAR(10 g) = 2.01 W/kg
Maximum value of SAR (measured) = 5.31 W/kg




                      0 dB = 5.31 W/kg = 7.25 dBW/kg




                                                                                                               Plot No. 9


Test Laboratory: UL Verification Services Inc. SAR Lab 3    Date/Time: 9/7/2017 11:02:13 AM

20170907_SystemPerformanceCheck-D1900V2 SN 5d140
Frequency: 1900 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.26 W/kg




                                                                                          Plot No. 10


Test Laboratory: UL Verification Services Inc. SAR Lab 3                     Date/Time: 9/7/2017 7:55:17 PM

20170907_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 1750 MHz; σ = 1.54 S/m; εr = 52.048; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(8.48, 8.48, 8.48); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI A v5.0; Type: QD OVA 002 AA; Serial: 1194

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 4.72 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 55.981 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 6.39 W/kg
SAR(1 g) = 3.59 W/kg; SAR(10 g) = 1.92 W/kg
Maximum value of SAR (measured) = 4.80 W/kg




                      0 dB = 4.80 W/kg = 6.81 dBW/kg




                                                                                                               Plot No. 11


Test Laboratory: UL Verification Services Inc. SAR Lab 3    Date/Time: 9/7/2017 8:14:21 PM

20170907_SystemPerformanceCheck-D1750V2 SN 1050
Frequency: 1750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.73 W/kg




                                                                                             Plot No. 12


Test Laboratory: UL Verification Services Inc. SAR Lab 3                    Date/Time: 9/8/2017 12:06:33 AM

20170908_SystemPerformanceCheck-D750V3 SN 1024
Frequency: 750 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 750 MHz; σ = 1.025 S/m; εr = 55.264; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(10.66, 10.66, 10.66); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm
(Mechanical Surface Detection)
- Phantom: ELI B v5.0; Type: QD OVA 002 AA; Serial: 1196

Body/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.09 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 32.808 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 1.34 W/kg
SAR(1 g) = 0.909 W/kg; SAR(10 g) = 0.606 W/kg




                      0 dB = 1.09 W/kg = 0.37 dBW/kg




                                                                                                           Plot No. 13


Test Laboratory: UL Verification Services Inc. SAR Lab 3    Date/Time: 9/8/2017 12:23:02 AM

20170907_SystemPerformanceCheck-D750V3 SN 1024
Frequency: 750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.09 W/kg




                                                                                          Plot No. 14


Test Laboratory: UL Verification Services Inc. SAR Lab 3                     Date/Time: 9/15/2017 6:54:57 PM

20170915_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2450 MHz; σ = 2.002 S/m; εr = 51.694; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(7.89, 7.89, 7.89); Calibrated: 5/30/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI A v5.0; Type: QD OVA 002 AA; Serial: 1194

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.58 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.217 V/m; Power Drift = 0.10 dB
Peak SAR (extrapolated) = 11.2 W/kg
SAR(1 g) = 5.34 W/kg; SAR(10 g) = 2.46 W/kg
Maximum value of SAR (measured) = 7.62 W/kg




                      0 dB = 7.62 W/kg = 8.82 dBW/kg




                                                                                                               Plot No. 15


Test Laboratory: UL Verification Services Inc. SAR Lab 3    Date/Time: 9/15/2017 7:15:32 PM

20170915_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.84 W/kg




                                                                                          Plot No. 16


Test Laboratory: UL Verification Services Inc. SAR Lab 3                    Date/Time: 9/18/2017 6:45:37 PM

20170918_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5200 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5200 MHz; σ = 5.42 S/m; εr = 48.008; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1239; Calibrated: 7/27/2017
- Probe: EX3DV4 - SN3902; ConvF(4.89, 4.89, 4.89); Calibrated: 5/30/2017;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 1.4mm
(Mechanical Surface Detection)
- Phantom: ELI B v5.0; Type: QD OVA 002 AA; Serial: 1196

Body/5.2 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 18.6 W/kg

Body/5.2 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 50.888 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 33.5 W/kg
SAR(1 g) = 7.67 W/kg; SAR(10 g) = 2.15 W/kg




                      0 dB = 18.6 W/kg = 12.70 dBW/kg




                                                                                                           Plot No. 17


Test Laboratory: UL Verification Services Inc. SAR Lab 3   Date/Time: 9/18/2017 7:06:04 PM

20170918_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5200 MHz; Duty Cycle: 1:1

Body/5.2 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 12.5 W/kg




                                                                                         Plot No. 18


Test Laboratory: UL Verification Services Inc. SAR Lab 4                     Date/Time: 8/21/2017 1:16:34 PM

20170821_SystemPerformanceCheck-D5GHzV2 SN 1138
Frequency: 5800 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5800 MHz; σ = 5.897 S/m; εr = 47.257; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1258; Calibrated: 5/12/2017
- Probe: EX3DV4 - SN3990; ConvF(4.42, 4.42, 4.42); Calibrated: 3/15/2017;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 B; Type: QDOVA002AA; Serial: 1248

Body/5.8 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 21.4 W/kg

Body/5.8 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 52.264 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 30.5 W/kg
SAR(1 g) = 7.88 W/kg; SAR(10 g) = 2.28 W/kg
Maximum value of SAR (measured) = 18.0 W/kg




                      0 dB = 18.0 W/kg = 12.55 dBW/kg




                                                                                                               Plot No. 19


Test Laboratory: UL Verification Services Inc. SAR Lab 4   Date/Time: 8/21/2017 2:29:23 PM

20170821_SystemPerformanceCheck-D5GHzV2 SN 1138
Frequency: 5800 MHz; Duty Cycle: 1:1

Body/5.8 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 13.8 W/kg




                                                                                         Plot No. 20


Test Laboratory: UL Verification Services Inc. SAR Lab 4                     Date/Time: 8/22/2017 3:54:38 PM

20170822_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5200 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 5200 MHz; σ = 5.361 S/m; εr = 47.453; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1258; Calibrated: 5/12/2017
- Probe: EX3DV4 - SN3990; ConvF(4.84, 4.84, 4.84); Calibrated: 3/15/2017;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: SM 000 T01 DA; Serial: TP:1247

Body/5.2 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 19.5 W/kg

Body/5.2 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 54.372 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 31.7 W/kg
SAR(1 g) = 7.69 W/kg; SAR(10 g) = 2.17 W/kg
Maximum value of SAR (measured) = 18.6 W/kg




                      0 dB = 18.6 W/kg = 12.70 dBW/kg




                                                                                                               Plot No. 21


Test Laboratory: UL Verification Services Inc. SAR Lab 4   Date/Time: 8/22/2017 4:25:38 PM

20170822_SystemPerformanceCheck-D5GHzV2 SN 1003
Frequency: 5200 MHz; Duty Cycle: 1:1

Body/5.2 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 14.0 W/kg




                                                                                         Plot No. 22


Test Laboratory: UL Verification Services Inc. SAR Lab 4                    Date/Time: 9/1/2017 12:06:00 AM

20170831_SystemPerformanceCheck-D2600V2 SN 1036
Frequency: 2600 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used: f = 2600 MHz; σ = 2.151 S/m; εr = 50.321; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1258; Calibrated: 5/12/2017
- Probe: EX3DV4 - SN3990; ConvF(7.75, 7.75, 7.75); Calibrated: 3/15/2017;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm
(Mechanical Surface Detection)
- Phantom: ELI v5.0 A; Type: SM 000 T01 DA; Serial: TP:1247

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.59 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.898 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 11.6 W/kg
SAR(1 g) = 5.25 W/kg; SAR(10 g) = 2.27 W/kg
Maximum value of SAR (measured) = 7.65 W/kg




                      0 dB = 7.65 W/kg = 8.84 dBW/kg




                                                                                                           Plot No. 23


Test Laboratory: UL Verification Services Inc. SAR Lab 4    Date/Time: 9/1/2017 12:26:54 AM

20170831_SystemPerformanceCheck-D2600V2 SN 1036
Frequency: 2600 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.17 W/kg




                                                                                          Plot No. 24