Dipole cal certs

FCC ID: C3K1793

RF Exposure Info

Download: PDF
FCCID_3611212

Calibration Laboratory of                                          NUz                                      S        Schweizerischer Kalibrierdienst
Schmid & Partner                                                  m‘\
                                                                   wA                                                Service suisse d‘étalonnage
  Engineering AG                                                                                            C        Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                         4/fi\“\                                 S        gwiss Calibration Service
                                                                      thilabs®



Accredited by the Swiss Accreditation Service (SAS)                                                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client      Microsoft                                                                             Certificate No: D2450V2—917_Apr17

CALIBRATION CERTIFICATE
 Object                                  D2450V2 — SN: 917


 Calibration procedure(s}                QA CAL—O05.v9
                                         Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                       April 21, 2017


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


  All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


  Calibration Equipment used (M&TE critical for callbration)


  Primary Standards                      | ID #                    Cal Date (Certificate No.)                             Scheduled Calibration
  Power meter NRP                         SN:     104778           04—Apr—17     (No.   217—02521/02522)                  Apr—18
  Power sensor NRP—Z91                    SN:     103244           04—Apr—17     (No.   217—02521)                        Apr—18
  Power sensor NRP—Z91                    SN:     103245           04—Apr—17     (No.   217—02522)                        Apr—18
  Reference 20 dB Attenuator              SN:     5058 (20k)       07—Apr—17     (No.   217—02528)                        Apr—18
  Type—N mismatch combination             SN: 5047.2 / 06327       07—Apr—17 (No. 217—02529)                              Apr—18
  Reference Probe EX3DV4                  SN: 7349                 31—Dec—16 (No. EX3—7349_Dec16)                         Dec—17
  DAE4                                    SN: 601                  28—Mar—17 (No. DAE4—601_Mar17)                         Mar—18


  Secondary Standards                     ID #                     Check Date (in house)                                  Scheduled Check
  Power meter EPM—442A                    SN: GB37480704           07—Oct—15 (in house check Oct—16)                      In house check: Oct—18
  Power sensor HP 8481A                   SN: US37202783           07—Oct—15 (in house check Oct—16)                      in house check: Oct—18
  Power sensor HP 8481A                   SN: MY41092317           07—Oct—15 (in house check Oct—16)                      In house check: Oct—18
  RF generator R&S SMT—06                 SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
  Network Analyzer HP 8753E               SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                          Name                                     Function                                Signature
  Calibrated by:                          Michael Weber                           Laboratory Technician                   I&r



  Approved by:                            KatJa Pokovic                           Technical Manager                                    4_’


                                                                                                                          Issued: April 21, 2017
 L This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


 maainccal aiol maamavin mase        k     }~


Calibration
     .      Laboratory of                                   3\‘Uz
                                                               \GZ/z       g   Schweizerischer Kalibrierdienst
Schmid & Partner                                           %               c   Service suisse d‘étalonnage
  Engineering AG                                           T//:\\[             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                */I//\\‘\\s‘   S   swiss Calibration Service
                                                               alutabs

Accredited by the Swiss Accreditation Service (SAS)                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", June 2013
     b) 1EC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
        devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
         February 2005
     c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
         communication devices used in close proximity to the human body (frequency range of 30
        MHz to 6 GHz)", March 2010
     d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
     e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
      SAR measured: SAR measured at the stated antenna input power.
      SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     *    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
   The reported uncertainty of measurement is stated as the standard uncertainty of measurement
   multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
   probability of approximately 95%.


Measurement Conditions
  DASY system contfiguration, as far as not given on page 1.
    DASY Version                                               DASY5                              V52.10.0
    Extrapolation                                   Advanced Extrapolation
    Phantom                                          Modular Flat Phantom

    Distance Dipole Center — TSL                               10 mm                            with Spacer
    Zoom Scan Resolution                               dx, dy, dz =5 mm
    Frequency                                          2450 MHz +1 MHz



Head TSL parameters
  The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

    Nominal Head TSL parameters                                   22.0 °C           39.2               1.80 mho/m
    Measured Head TSL parameters                               (22.0 + 0.2) °C   37.7 16 %          1.87 mho/m + 6 %
    Head TSL temperature change during test                       <0.5 °C            ~—                      ———


SAR result with Head TSL

    SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
    SAR measured                                         250 mW input power                     13.3 W/kg
    SAR for nominal Head TSL parameters                      normalized to 1W         51.8 Wikg + 17.0 % (k=2)

    SAR averaged over 10 cm* (10 g) of Head TSL                  condition
     SAR measured                                        250 mW input power                  6.16 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.2 Wikg £ 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                   22.0 °C           52.7              1.95 mho/m
     Measured Body TSL parameters                              (22.0 £ 0.2) °C    52.5 +6 %         2.03 mho/m x 6 %

     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     12.9 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W         50.6 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Body TSL                 condition
     SAR measured                                           250 mW input power               6.02 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         23.8 Wikg + 16.5 % (k=2)


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           57.3 Q + 2.7 jQ
      Return Loss                                                                       — 22.8 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           52.1 Q + 4.0 jQ
      Return Loss                                                                      —27.2 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.157 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

       Manufactured by                                                                   SPEAG
       Manufactured on                                                             December 19, 2012


DASY5 Validation Report for Head TSL

                                                                                  Date: 21.04.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 917

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.87 $/m; &, = 37.7; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(7.72, 7.72, 7.72); Calibrated: 31.12.2016;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   *   DASY52 52.10.0(1442); SEMCAD X 14.6.10(7413)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 112.4 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 27.2 W/kg
SAR(I g) = 13.3 W/kg; SAR(10 g) = 6.16 W/kg
Maximum value of SAR (measured) = 21.3 W/kg


       dB
        0

       —5.00

       ~10.00

       —15.00

       —20.00

       —25.00
                 0 dB = 21.3 W/kg = 13.28 dBW/kg


Impedance Measurement Plot for Head TSL




                                                               24 Apr 2017 11:50156
       (CHZ) S11    1 U FS           4:87.324 e   26543 a   172.43 pH      2 450.000 000 MHz




       Del

       Ca




       fvg
       16


       HILd


       CH2




       Ca




       Aw
       16°

       H1d

                   START 2 250,000 000 MHz                           STOP 2 650.000 000 MHz


DASY5 Validation Report for Body TSL

                                                                                  Date: 21.04.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 917

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.03 S/m; s = 52.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(7.79, 7.79, 7.79); Calibrated: 31.12.2016;

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 Sn601; Calibrated: 28.03.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

       DASY52 52.10.0(1442); SEMCAD X 14.6.10(7413)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 103.3 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 25.3 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) = 6.02 W/kg
Maximum value of SAR (measured) = 20.0 W/kg


       dB
        0

       —5.00

       ~10.00

       —15.00

       —20.00

       —25.00
                 0 dB = 20.0 W/kg = 13.01 dBW/kg


Impedance Measurement Plot for Body TSL




                                                                21 Apr 2017     11:S0:02
       CBZ) si1    1 U FS$          1: 52.105 a   3.9609 e   257.91 pH         2 450.000 @00 MHz
                                             w    “-—T‘\-\




       Del

       Ca




       five
       16


       H1d




                  START 2 250.000 000 MHz                                STOP 2 650.000 008 MHz


Calibration Laboratory of                                            ly                                    S       Schweizerischer Kalibrierdienst




                                                                 7
Schmid & Partner                                                     y




                                                                              p
                                                                                                           Cc      Service suisse d‘étalonnage




                                                             $
  Engineering AG                                                 Lz e                                              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      Hi(y                                     S Swiss Calibration Service
                                                                      ues

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      Microsoft                                                                          Certificate No: D5GHzV2—1159_Apr17

|CALIBRATION CERTIFICATE                                                                                                                             l
 Object                                D5SGHzV2 — SN: 1159


 Calibration procedure(s)             QA CAL—22.v2
                                       Calibration procedure for dipole validation kits between 3—6 GHz



 Calibration date:                     April 25, 2017


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1)
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                     | ID #                     Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN:     104778            04—Apr—17   (No.   217—02521/02522)                   Apr—18
 Power sensor NRP—Z91                   SN:     103244            04—Apr—17   (No.   217—02521)                         Apr—18
 Power sensor NRP—291                   SN:     103245            04—Apr—17   (No.   217—02522)                         Apr—18
 Reference 20 dB Attenuator             SN:     5058 (20k)        07—Apr—17   (No.   217—02528)                         Apr—18
 Type—N mismaich combination            SN:     5047.2 / 06327    07—Apr—17   (No.   217—02529)                         Apr—18
 Reference Probe EX3DV4                 SN: 3503                  31—Dec—16 (No. EX3—3503_Dec16)                        Dec—17
 DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18

 Secondary Standards                JID #                         Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN:     GB37480704        07—Oct—15 (in house check Oct—16)                     in   house check:   Oct—18
 Power sensor HP 8481A                  SN:     US37292783        07—Oct—15 (in house check Oct—16)                     In   house check:   Oct—18
 Power sensor HP 8481A                  SN:     MY41092317        07—Oct—15 (in house check Oct—16)                     In   house check:   Oct—18
 RF generator R&S SMT—06                SN:     100972            15—Jun—15 (in house check Oct—16)                     in   house check:   Oct—18
 Network Analyzer HP 8753E              SN:     US37390585        18—Oct—01 (in house check Oct—16)                     In   house check:   Oct—17


                                        Name                                   Function                                  Signature
 Calibrated by:                         Michael Weber                          Laboratory Technician                     ”M_


 Approved by:                           Katia Pokovic                          Technical Manager                   m



                                                                                                                        Issued: April 25, 2017
[Thns calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D5GHzV2—1159_Apr17                                    Page 1 of 13


Calibration
      ‘     Laboratory of                                   willy‘s/,,’
                                                            e                 S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           ;E%&               c   Service suisse d‘étalonnage
  Engineerlng AG                                           T //:\\ s              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                7'/,,,fi\‘y}       S   swiss Calibration Service
                                                               rhiba®s

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates
Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y.z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
   * SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D5GHzV2—1159_Apr17                             Page 2 of 13


Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
     DASY Version                                              DASY5                              V52.10.0
     Extrapolation                                   Advanced Extrapolation

     Phantom                                       Modular Flat Phantom V5.0
     Distance Dipole Center— TSL                               10 mm                             with Spacer
     Zoom Scan Resolution                        dx, dy = 4.0 mm, dz =1.4 mm         Graded Ratio = 1.4 (Z direction)
                                                       5250 MHz + 1 MHz
     Frequency                                         5600 MHz + 1 MHz
                                                       5750 MHz + 1 MHz



Head TSL parameters at 5250 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.9               4.71 mho/m
     Measured Head TSL parameters                              (22.0 £ 0.2) °C   34.6 16 %          4.52 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C


SAR result with Head TSL at 5250 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     7.81 Wikg

     SAR for nominal Head TSL parameters                     normalized to 1W        774 Wikg £ 19.9 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition

     SAR measured                                        100 mW input power                  2.24 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         22.1 Wikg #19.5 % (k=2)




Certificate No: D5GHzV2—1159_Apr17                    Page 3 of 13


Head TSL parameters at 5600 MHz
   The followingparameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.5               5.07 mho/m

     Measured Head TSL parameters                              (22.0 + 0.2) °C   34.2 16 %          4.86 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C                                   ~——


SAR result with Head TSL at 5600 MHz

     SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     8.10 Wikg

     SAR for nominal Head TSL parameters                     normalized to 1W        80.2 W / kg + 19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                     2.32 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        22.9 Wikg + 19.5 % (k=2)


Head TSL parameters at 5750 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.4               5.22 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   33.9 +6 %          5.06 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL at 5750 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
     SAR measured                                        100 mW input power                     7.84 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         77.6 Wikg 2 19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Head TSL                 condition

     SAR measured                                        100 mW input power                  2.23 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         22.0 Wikg £ 19.5 % (k=2)




Certificate No: D5GHzV2—1159_Apr17                    Page 4 of 13


Body TSL parameters at 5250 MHz
   The followingparameters and calculations were applied.
                                                              Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                 22.0 °C           48.9               5.36 mho/m

     Measured Body TSL parameters                             (22.0 # 0.2) °C   47.7 £6 %          5.50 mho/m + 6 %
     Body TSL temperature change during test                     <0.5 °C

SAR result with Body TSL at 5250 MHz

     SAR averaged over 1 cm‘ (1 g) of Body TSL                 Condition

     SAR measured                                       100 mW input power                     7.68 W/kg
     SAR for nominal Body TSL parameters                    normalized to 1W        76.5 Wikg + 19.9 % (k=2)


     SAR averaged over 10 cm* (10 g) of Body TSL                condition
     SAR measured                                       100 mW input power                 2.14 Wikg
     SAR for nominal Body TSL parameters                    normalized to 1W        21.3 Wikg £ 19.5 % (k=2)

Body TSL parameters at 5600 MHz
   The followingparameters and calculations were applied.
                                                              Temperature       Permittivity          Conductivity

     Nominal Body TSL parameters                                 22.0 °C            48.5              5.77 mho/m
     Measured Body TSL parameters                             (22.0 + 0.2) °C   47.0 £ 6 %         5.97 mho/m + 6 %
     Body TSL temperature change during test                     <0.5 °C

SAR result with Body TSL at 5600 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
     SAR measured                                       100 mW input power                     8.02 W/kg
     SAR for nominal Body TSL parameters                    normalized to 1W         79.8 Wikg £ 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                condition
     SAR measured                                       100 mW input power                  2.25 Wikg
     SAR for nominal Body TSL parameters                    normalized to 1W         22.3 Wikg £ 19.5 % (k=2)




Certificate No: D5GHzV2—1159_Apri7                   Page 5 of 13


Body TSL parameters at 5750 MHz
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C          48.3               5.94 mho/m
     Measured Body TSL parameters                             (22.0 + 0.2) °C   46.8 +6 %          6.18 mho/m   6%

     Body TSL temperature change during test                     <0.5 °C


SAR result with Body TSL at 5750 MHz

     SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
     SAR measured                                        100 mW input power                    7.73 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W        76.9 W/kg £ 19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Body TSL                condition

     SAR measured                                        100 mW input power                 215 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        21.4 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1159_Apr17                    Page 6 of 13


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL at 5250 MHz

      Impedance, transformed to feed point                                           47.7 Q — 6.5 jQ

      Return Loss                                                                       — 23.0 dB


Antenna Parameters with Head TSL at 5600 MHz

      Impedance, transformed to feed point                                           52.2 Q — 1.3 jQ
      Return Loss                                                                       — 32.0 dB


Antenna Parameters with Head TSL at 5750 MHz

      impedance, transformed to feed point                                           53.0 Q — 2.9 jQ

      Return Loss                                                                       —27.9 dB


Antenna Parameters with Body TSL at 5250 MHz

      Impedance, transformed to feed point                                           47.7 Q — 5.3 jQ
      Return Loss                                                                       — 24.6 dB


Antenna Parameters with Body TSL at 5600 MHz

      Impedance, transformed to feed point                                           53.7 Q + 1.6 jQ
      Return Loss                                                                      —28.1 dB


Antenna Parameters with Body TSL at 5750 MHz

      Impedance, transformed to feed point                                           53.3 Q — 0.9 jQ
      Return Loss                                                                      — 29.6 dB



General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.201 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole tength is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.

Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 June 06, 2013



Certificate No: D5GHzV2—1159_Apr17                       Page 7 of 13


DASY5 Validation Report for Head TSL

                                                                                       Date: 25.04.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole D5GHzV2; Type: DSGHzV2; Serial: D5GHzV2 — SN: 1159

Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
Medium parameters used: f = 5250 MHz; 0 = 4.52 $/m:; & = 34.6; p = 1000 kg/m‘,
Medium parameters used: f = 5600 MHz; 0 = 4.86 S/m; & = 34.2; p = 1000 kg/m? ,
Medium parameters used: f = 5750 MHz; 0 = 5.06 S/m; & = 33.9; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN3503; ConvF(5.58, 5.58, 5.58); Calibrated: 31.12.2016, ConvF(5.09, 5.09,
        5.09); Calibrated: 31.12.2016, ConvF(5.02, 5.02, 5.02); Calibrated: 31.12.2016;

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *«   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   «_   DASY52 52.10.0(1442); SEMCAD X 14.6.10(7413)


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5250 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0; Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 71.48 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 28.8 W/kg
SAR(I1 g) =7.81 W/kg; SAR(10 g) = 2.24 W/kg
Maximum value of SAR (measured) = 17.6 W/ikg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1 .4mm
Reference Value = 71.77 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 31.7 W/kg
SAR(1 g) = 8.1 W/kg; SAR(10 g) = 2.32 W/kg
Maximum value of SAR (measured) = 18.7 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 70.36 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 31.5 W/kg
SAR(I g) = 7.84 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 18.5 W/kg




Certificate No: D5GHzV2—1159_Apr17             Page 8 of 13


       —10.00

       ~15.00

       —20.00

       —25.00
                  0 dB = 18.5 W/kg = 12.67 dBW/kg




Certificate No: D5GHzV2—1159_Apr17            Page 9 of 13


Impedance Measurement Plot for Head TSL




                                                               25 fApr 2017     12:29133
         (CH® sii     i uU fs         1147,686 n   —6.4961 0 4.6667 pF         5 250.000 000 MHz


         *
                                                                                                   CH1 Markers
         Del
                                                                                                    2: 52219 0
                                                                                                      ~1,2793 4
         for                                                                                       5.60000 GHz
                                                                                                    3: $2,973 a
                                                                                                      —2.8906 a
                                                                                                   5.75000 GHz

         Ave
         16

         H1d

         CH2



                                                                                                   CH2 Markers

                                                                                                   2:—32.019 dB
         Cor                                                                                        5.60000 GHz
                                                                                                   3:—27.999 dB
                                                                                                    5.75000 GHz

         fiw
         tCs


         H1d

                    START 5 000.000 0G0 MHz                              STOP 6 000.000 800 MHz




Certificate No: D5GHzV2—1159_Apr17                 Page 10 of 13


DASY5 Validation Report for Body TSL

                                                                                          Date: 18.04.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole DSGHzV2; Type: D5GHzV2; Serial: DSGHzV2 — SN: 1159

Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
Medium parameters used: f = 5250 MHz; 0 = 5.5 $/m; & =47.7; p = 1000 kg/m‘ ,
Medium parameters used: f = 5600 MHz; 0 = 5.97 S$/m; & = 47; p = 1000 kg/m‘ ,
Medium parameters used: f = 5750 MHz; 0 = 6.18 $/m; & = 46.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   «_   Probe: EX3DV4 — SN3503; ConvF(5.14, 5.14, 5.14); Calibrated: 31.12.2016, ConvF(4.57, 4.57,
        4.57), Calibrated: 31.12.2016, ConvF(4.51, 4.51, 4.51); Calibrated: 31.12.2016;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

        DASY52 52.10.0(1442); SEMCAD X 14.6.10(7413)


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5250MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 66.10 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 29.6 W/kg
SAR(1 g) =7.68 W/kg; SAR(10 g) = 2.14 W/kg
Maximum value of SAR (measured) = 18.1 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 66.45 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 33.7 W/kg
SAR(1 g) = 8.02 W/kg; SAR(10 g) = 2.25 W/kg
Maximum value of SAR (measured) = 19.6 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 64.64 V/m; Power Drift = —0.9 dB
Peak SAR (extrapolated) =34.1 W/kg
SAR(I g) =7.73 W/kg; SAR(10 g) = 2.15 W/kg
Maximum value of SAR (measured) = 20.1 W/kg




Certificate No: D5GHzV2—1159_Apr17               Page 11 of 13


       —5.00

       —10.00

       ~15.00

       —20.00

       —25.00
                  0 dB = 20.1 W/kg = 13.03 dBW/kg




Certificate No: D5GHzV2—1159_Apr17           Page 12 of 13


Impedance Measurement Plot for Body TSL




                                                              19 Apr 2017     15:28111
         CHZ sn      i uors          4: 47.709 a   52559 a 5.7679 pF         5 250.000 @00 MHz




                                                                                                 CH1 Markers
         Del
                                                                                                  2: 53.727 a
                                                                                                     1.6113 o
         Cor                                                                                     5.60000 GHz
                                                                                                  3: 53.299 a
                                                                                                  ~893.92 mo
                                                                                                 5,75000 GHz


         fivg
         16


         H1g


         CH2



                                                                                                 CH2 Markers

         Cor                                                                                     2:—28.148 d8
                                                                                                 5.60000 GHz
                                                                                                 3:—29.628 dB
                                                                                                 5.75000 GHz

         Av
         16°

         H1d

                   START 5 000,000 000 MHz                             STOP 6 000.000 090 MHz




Certificate No: D5GHzV2—1159_Apr17                 Page 13 of 13



Document Created: 2017-09-18 10:54:46
Document Modified: 2017-09-18 10:54:46

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC