HAC RF test report 2 of 7

FCC ID: BCG-E3234A

RF Exposure Info

Download: PDF
FCCID_3976358

Test Laboratory: UL Verification Services Inc. SAR Lab C                           Date: 7/9/2018

HAC-RF Emission
Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: EF3DV3 - SN4041; ConvF(1, 1, 1); Calibrated: 3/16/2018;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1472; Calibrated: 3/8/2018
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field measurement 835MHz/835 MHz/Hearing Aid Compatibility Test at 15mm
distance (41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 111.9 V/m; Power Drift = -0.00 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 108.8 V/m
Near-field category: M4 (AWF 0 dB)

                                         PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          88.14 V/m        90.35 V/m   89.55 V/m
                                          Grid 4 M4        Grid 5 M4   Grid 6 M4
                                          58.45 V/m        59.83 V/m   59.47 V/m
                                          Grid 7 M4        Grid 8 M4   Grid 9 M4
                                          106.1 V/m        108.8 V/m   107.6 V/m




                      0 dB = 108.8 V/m = 40.73 dBV/m



                                                                                                    Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab C                      Date/Time: 7/9/2018 1:43:50 AM

HAC-RF Emission
Frequency: 1880 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.0C
Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1472; Calibrated: 3/8/2018
- Probe: EF3DV3 - SN4041; ConvF(1, 1, 1); Calibrated: 3/16/2018;
- Sensor-Surface: (Fix Surface)
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB; Serial: 1155

Dipole E-Field Measurement 1880MHz/1880 MHz/Hearing Aid Compatibility Test at
15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 114.6 V/m; Power Drift = -0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 80.50 V/m
Near-field category: M3 (AWF 0 dB)

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          76.84 V/m        78.84 V/m   78.47 V/m
                                          Grid 4 M4        Grid 5 M3   Grid 6 M3
                                          62.94 V/m        63.84 V/m   63.68 V/m
                                          Grid 7 M3        Grid 8 M3   Grid 9 M3
                                          78.59 V/m        80.50 V/m   79.88 V/m




                      0 dB = 80.50 V/m = 38.12 dBV/m



                                                                                                               Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab C                           Date: 7/9/2018

HAC-RF Emission
Communication System: UID 0, CW (0); Frequency: 2600 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: EF3DV3 - SN4041; ConvF(1, 1, 1); Calibrated: 3/16/2018;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1472; Calibrated: 3/8/2018
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field Measurement 2600MHz/2600 MHz/Hearing Aid Compatibility Test at
15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 58.10 V/m; Power Drift = -0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 81.28 V/m
Near-field category: M3 (AWF 0 dB)

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          79.39 V/m        81.28 V/m   80.99 V/m
                                          Grid 4 M3        Grid 5 M3   Grid 6 M3
                                          75.16 V/m        76.40 V/m   75.90 V/m
                                          Grid 7 M3        Grid 8 M3   Grid 9 M3
                                          77.57 V/m        79.10 V/m   78.45 V/m




                      0 dB = 81.28 V/m = 38.20 dBV/m



                                                                                                    Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab C                           Date: 7/11/2018

HAC-RF Emission
Communication System: UID 0, CW (0); Frequency: 2450 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: EF3DV3 - SN4041; ConvF(1, 1, 1); Calibrated: 3/16/2018;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1472; Calibrated: 3/8/2018
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field Measurement 2450MHz/2450 MHz/Hearing Aid Compatibility Test at
15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 83.64 V/m; Power Drift = -0.04 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.11 V/m
Near-field category: M3 (AWF 0 dB)

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          88.12 V/m        91.07 V/m   90.41 V/m
                                          Grid 4 M3        Grid 5 M3   Grid 6 M3
                                          78.47 V/m        80.00 V/m   79.76 V/m
                                          Grid 7 M3        Grid 8 M3   Grid 9 M3
                                          89.30 V/m        92.11 V/m   91.06 V/m




                      0 dB = 92.11 V/m = 39.29 dBV/m




                                                                                                     Plot No. 4


Test Laboratory: UL Verification Services Inc. SAR Lab C                           Date: 7/11/2018

HAC-RF Emission
Communication System: UID 0, CW (0); Frequency: 5500 MHz; Duty Cycle: 1:1
Phantom section: RF Section
DASY5 Configuration:
- Probe: EF3DV3 - SN4041 (5-6 GHz); ConvF(1, 1, 1); Calibrated: 3/16/2018;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1472; Calibrated: 3/8/2018
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BB
- Measurement SW: DASY52, Version 52.8 (7);SEMCAD X Version 14.6.10 (7164)

Dipole E-Field Measurement 5.5GHz/5.5GHz/Hearing Aid Compatibility Test at 15mm
distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 21.97 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 98.38 V/m
Near-field category: M3 (AWF 0 dB)

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          78.27 V/m        79.52 V/m   78.18 V/m
                                          Grid 4 M3        Grid 5 M3   Grid 6 M3
                                          97.07 V/m        98.38 V/m   95.66 V/m
                                          Grid 7 M3        Grid 8 M3   Grid 9 M3
                                          79.68 V/m        81.31 V/m   79.45 V/m




                      0 dB = 98.38 V/m = 39.86 dBV/m




                                                                                                     Plot No. 5