12681939_App B_SAR System Check Plots

FCC ID: BCG-A2084

RF Exposure Info

Download: PDF
FCCID_4475942

Test Laboratory: UL Verification Services Inc. SAR Lab H                    Date/Time: 9/12/2019 11:16:59 AM

20190912_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.0C
Medium parameters used: f = 2450 MHz; σ = 1.787 S/m; εr = 37.957; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1545; Calibrated: 4/12/2019
- Probe: EX3DV4 - SN7356; ConvF(7.98, 7.98, 7.98) @ 2450 MHz; Calibrated: 4/17/2019
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD; Serial: 1602

Head/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.16 W/kg

Head/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 63.55 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 10.7 W/kg
SAR(1 g) = 5.25 W/kg; SAR(10 g) = 2.44 W/kg
Maximum value of SAR (measured) = 7.46 W/kg




                      0 dB = 7.46 W/kg = 8.73 dBW/kg




                                                                                                               Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab H    Date/Time: 9/12/2019 11:32:12 AM

20190912_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1

Head/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.23 W/kg




                                                                                           Plot No. 2


Test Laboratory: UL Verification Services Inc. SAR Lab F                    Date/Time: 9/12/2019 10:09:35 AM

20190912_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.0C
Medium parameters used: f = 2450 MHz; σ = 2.046 S/m; εr = 51.001; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1548; Calibrated: 4/16/2019
- Probe: EX3DV4 - SN7482; ConvF(7.27, 7.27, 7.27); Calibrated: 7/18/2019;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI V5.0 Slot 6/7; Type: QD OVA 002 AA; Serial: 1247

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 6.22 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.699 V/m; Power Drift = -0.10 dB
Peak SAR (extrapolated) = 11.2 W/kg
SAR(1 g) = 5.38 W/kg; SAR(10 g) = 2.46 W/kg
Maximum value of SAR (measured) = 7.70 W/kg




                      0 dB = 7.70 W/kg = 8.86 dBW/kg




                                                                                                               Plot No. 3


Test Laboratory: UL Verification Services Inc. SAR Lab F   Date/Time: 9/12/2019 10:31:11 AM

20190912_SystemPerformanceCheck-D2450V2 SN 706
Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.50 W/kg




                                                                                          Plot No. 4



Document Created: 2019-09-25 18:57:53
Document Modified: 2019-09-25 18:57:53