12458150-S2V1_App F_SAR Dipole Certificates

FCC ID: BCG-A2032

RF Exposure Info

Download: PDF
FCCID_4204554

Calibration
      s     Laboratory of                                          i ty/7
                                                                  NR\                                       S      Schweizerischer Kalibrierdienst
                                                                  S              es
Schmid & Partner                                                 i’lfififlé                                    c      Service suisse d‘étalonnage
   Englneenng AG                                                 > ¢_—\\ §3                                        Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      ‘f/,,/f\\\\\\@                            S      Swiss Calibration Service
                                                                      Aulabs®



Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       UL CCS USA                                                                           Certificate No: D2450V2—706_May18

CALIBRATION CERTIFICATE
 Object                               D2450V2 — SN:706


 Calibration procedure(s)             QA CAL—O5.v10
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    May 18, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN:    104778             04—Apr—18     (No.   217—02672/02673)                 Apr—19
 Power sensor NRP—Z91                   SN:    103244             04—Apr—18     (No.   217—02672)                       Apr—19
 Power sensor NRP—Z91                   SN:    103245             04—Apr—18     (No.   217—02673)                       Apr—19
 Reference 20 dB Attenuator             SN:    5058 (20k)         04—Apr—18     (No.   217—02682)                       Apr—19
 Type—N mismatch combination            SN:    5047.2 / 06327     04—Apr—18     (No.   217—02683)                       Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX8—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15     (in   house check Oct—16)               In   house check:   Oct—18
 Power sensor HP 8481A                  SN:    US37292783         07—Oct—15     (in   house check Oct—16)               In   house check:   Oct—18
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15     (in   house check Oct—16)               In   house check:   Oct—18
 RF generator R&S SMT—O6                SN:    100972             15—Jun—15     (in   house check Oct—16)               In   house check:   Oct—18
 Network Analyzer HP 8753E              SN:    US37390585         18—Oct—01     (in   house check Oct—17)               In   house check:   Oct—18


                                        Name                                      Function                               Signature
 Calibrated by:                         Manu Seitz                               Laboratory Technician                                a



 Approved by:                           Katja Pokovic                            Technical Manager              /f@



                                                                                                                        Issued: May 22, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D2450V2—706_May18                                     Page 1 of 9


Calibration
      ( l
            Laboratory of                                     uyY
                                                             oS\
                                                                 AN
                                                                                    S
                                                                                              .   &          Het    ‘
                                                                                        Schweizerischer Kalibrierdienst
                                                             &                &
Schmid & Partner                                            ifii                     c   Service suisse d‘étalonnage
   Engineering AG                                           y                           Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                t/fi\\\v‘                S   Swiss Calibration Service
                                                                  Alnl uks®



Accredited by the Swiss Accreditation Service (SAS)                                 Accreditatior: No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multitateral Agreement for the recognition of calikration certificates

Glossary:
TSL                          tissue simulating liquid
Convr                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
         communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   «& Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    »    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    s    SAR measured: SAR measured at the stated antenna input power.
    s    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—706_May18                                     Page 2 of 9


Measurement Conditions
   DASY system configuration, as far as not given onpage 1.
     DASY Version                                              DASY5                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2450 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   38.2 + 6 %         1.85 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     13.4 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         52.6 W/ikg + 17.0 % (k=2)


     SAR averaged over 10 cm‘ (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.22 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.6 Wikg + 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C            52.7              1.95 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C    52.3 +6 %         1.99 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C             es                       ————


SAR result with Body TSL

     SAR averaged over 1 cm‘® (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                     12.8 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         50.6 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                  5.96 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         23.7 Wikg + 16.5 % (k=2)



Certificate No: D2450V2—706_May18                     Page 3 of 9


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.1 Q + 6.8 |Q

      Return Loss                                                                       — 23.1 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           47.1 0 + 6.6 jQ
      Return Loss                                                                      — 22.6 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.143 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 May 28, 2002




Certificate No: D2450V2—706_May18                         Page 4 of 9


DASY5 Validation Report for Head TSL

                                                                                   Date: 18.05.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT : Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:706

Communication System: UID 0 — CW ; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; o = 1.85 S$/m; & = 38.2; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 Sn601; Calibrated: 26.10.2017

       Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

       DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)
   ®




Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 116.9 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 26.7 W/kg
SAR(1 g) = 13.4 W/kg; SAR(10 g) = 6.22 W/kg
Maximum value of SAR (measured) = 22.0 W/kg


       dB
        0

       —5.00

       —10.00

       —15.00

       —20.00

       —25.00
                  0 dB = 22.0 W/kg = 13.42 dBW/kg




Certificate No: D2450V2—706_May18              Page 5 of 9


Impedance Measurement Plot for Head TSL




                                                                     18 May 2018    @§8:55:31
          CH1) s11     1 U FS           1: 52117 &   68047 a   442.04 pH           2 4590.000 900 MHz




          De 1

          Cor




          A
          169


          H1 d


          CH2Z




          Ay
          ast

          H1d

                     START 2 250.000 06BB MHz                              STOP 2 650.0990 dod MHz




Certificate No: D2450V2—706_May18                      Page 6 of 9


DASY5 Validation Report for Body TSL

                                                                                     Date: 18.05.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:706

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.99 S/m; & = 52.3; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   s     Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

   e     Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e     Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   e     Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   »_—   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.2 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 25.3 W/kg
SAR(1 g) = 12.8 W/kg; SAR(10 g) = 5.96 W/kg
Maximum value of SAR (measured) = 20.5 W/kg


         dB
          0

         —5.00

         —10.00

         —15.00

         —20.00

         —25.00
                  0 dB = 20.5 W/kg = 13.12 dBW/kg




Certificate No: D2450V2—706_May18               Page 7 of 9


Impedance Measurement Plot for Body TSL




                                                                                      18 May 2018    88:52155
         CHI) s4i     $ UUFS          1: 47.086 a           6.6133 m             429.61 pH          2 450.900 @B0 MHz

                                               @7      -—”-'-
                                                                    ¢        k‘-\\\
                                                                                   ~
                                               +                    +        _         }~
         La                               B#       T            e       TT
                                                                        \        * Jd :\\

         Del

         Cor




         Ava
         16


         H1d

          CH2




         Cor




         Av
         16 2



         H1d

                    START 2 259.000 0G0 MHz                                                  STOP 2 650.000 008 MHz




Certificate No: D2450V2—706_May18                               Page 8 of 9


Appendix (Additional assessments outside the scope of SCS 0108)

Evaluation Condition

     Phantom                                       SAM Head Phantom       For usage with cSAR3SDV2—R/L


SAR result with SAM Head (Top)

     SAR averaged over 1 cm‘ (1 g) of Head TSL            Condition

     SAR (average measured)                          250 mW input power          13.9 W/kg
     SAR for nominal Head TSL parameters              normalized to 1W    55.9 Wikg + 17.5 % (k=2)

     SAR averaged over 10 cm® (10 g) of Head TSL           condition

     SAR measured                                    250 mW input power          6.49 Wikg
     SAR for nominal Head TSL parameters              normalized to 1W    26.1 W/kg + 16.9 % (k=2)

SAR result with SAM Head (Mouth)

     SAR averaged over 1 cm® (1 g) of Head TSL             Condition

     SAR (average measured)                          250 mW input power          14.2 Wikg
     SAR for nominal Head TSL parameters               normalized to 1W   57.0 Wikg * 17.5 % (k=2)

     SAR averaged over 10 cm® (10 g) of Head TSL           condition

     SAR (average measured)                          250 mW input power          6.82 W/kg
     SAR for nominal Head TSL parameters               normalized to 1W   27.4 Wikg * 16.9 % (k=2)

SAR result with SAM Head (Neck)

     SAR averaged over 1 cm* (1 g) of Head TSL             Condition

     SAR (average measured)                          250 mW input power          13.3 W/kg
     SAR for nominal Head TSL parameters               normalized to 1W   53.7 Wikg * 17.5 % (k=2)

     SAR averaged over 10 cm® (10 g) of Head TSL           condition

     SAR (average measured)                          250 mW input power          6.22 Wikg
     SAR for nominal Head TSL parameters               normalized to 1W   25.0 Wikg + 16.9 % (k=2)

SAR result with SAM Head (Ear)

     SAR averaged over 1 cm® (1 g) of Head TSL             Condition

     SAR (average measured)                          250 mW input power          8.56 W/kg
     SAR for nominal Head TSL parameters               normalized to 1W   34.4 Wikg + 17.5 % (k=2)

     SAR averaged over 10 cm® (10 g) of Head TSL           condition

     SAR (average measured)                          250 mW input power          4.32 Wikg
     SAR for nominal Head TSL parameters               normalized to 1W   17.4 Wikg + 16.9 % (k=2)


Certificate No: D2450V2—706_May18                  Page 9 of 9