12458150-S2V1_App C_SAR Highest Test Plots

FCC ID: BCG-A2032

RF Exposure Info

Download: PDF
FCCID_4204551

Test Laboratory: UL Verification Services Inc. SAR Lab 4                     Date/Time: 3/15/2019 3:36:22 PM

Bluetooth Head
Frequency: 2441 MHz; Duty Cycle: 1:1.29033;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used (interpolated): f = 2441 MHz; σ = 1.797 S/m; ε r = 37.812; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1547; Calibrated: 5/3/2018
- Probe: EX3DV4 - SN7356; ConvF(8.11, 8.11, 8.11) @ 2441 MHz; Calibrated: 4/24/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 AA; Serial: 1956

REP /Touch_ Bluetooth _ch 39/Area Scan (7x7x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.119 W/kg

REP /Touch_Bluetooth_ch 39/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm
Reference Value = 8.131 V/m; Power Drift = 0.14 dB
Peak SAR (extrapolated) = 0.367 W/kg
SAR(1 g) = 0.095 W/kg; SAR(10 g) = 0.045 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.134 W/kg




                      0 dB = 0.134 W/kg = -8.73 dBW/kg




                                                                                                                Plot No. 1


Test Laboratory: UL Verification Services Inc. SAR Lab 2                    Date/Time: 3/15/2019 5:01:15 PM

Bluetooth Body
Frequency: 2441 MHz; Duty Cycle: 1:1.29033;      Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C
Medium parameters used (interpolated): f = 2441 MHz; σ = 2.021 S/m; ε r = 51.685; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn1545; Calibrated: 4/13/2018
- Probe: EX3DV4 - SN3686; ConvF(7.02, 7.02, 7.02) @ 2441 MHz; Calibrated: 8/28/2018
- Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm
(Mechanical Surface Detection)
- Phantom: ELI V8.0 (20deg probe tilt) Back; Type: QD OVA 004 AA; Serial: 2082

Flat RHS Bottom/Bluetooth_ch 39/Area Scan (6x7x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.464 W/kg

Flat RHS Bottom/Bluetooth_ch 39/Zoom Scan (7x9x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 15.70 V/m; Power Drift = -0.19 dB
Peak SAR (extrapolated) = 2.88 W/kg
SAR(1 g) = 0.581 W/kg; SAR(10 g) = 0.169 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.10 W/kg




                      0 dB = 1.10 W/kg = 0.41 dBW/kg




                                                                                                              Plot No. 2



Document Created: 2019-03-15 23:23:22
Document Modified: 2019-03-15 23:23:22