SAR report 2 of 4

FCC ID: AK8145899511

RF Exposure Info

Download: PDF
FCCID_3835780

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/12/19


System Check_Body_750MHz
DUT: D750V3-1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: MSL_750_171219 Medium parameters used: f = 750 MHz; σ = 0.976 S/m; εr = 54.334; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(10.56, 10.56, 10.56); Calibrated: 2017/5/24;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: ELI v5.0_Right; Type: QDOVA002AA; Serial: TP:1029
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.09 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 59.88 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 3.16 W/kg
SAR(1 g) = 2.19 W/kg; SAR(10 g) = 1.49 W/kg
Maximum value of SAR (measured) = 2.84 W/kg




                    0 dB = 2.84 W/kg = 4.53 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/12/19


System Check_Body_835MHz
DUT: D835V2-499

Communication System: CW ; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_850_171219 Medium parameters used: f = 835 MHz; σ = 0.975 S/m; εr = 56.301; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(10.29, 10.29, 10.29); Calibrated: 2017/5/24;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: ELI v5.0_Right; Type: QDOVA002AA; Serial: TP:1029
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.13 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.74 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 3.37 W/kg
SAR(1 g) = 2.33 W/kg; SAR(10 g) = 1.56 W/kg
Maximum value of SAR (measured) = 3.01 W/kg




                    0 dB = 3.01 W/kg = 4.79 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/12/19


System Check_Body_1750MHz
DUT: D1750V2-1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_171219 Medium parameters used: f = 1750 MHz; σ = 1.444 S/m; εr = 55.14; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(8.51, 8.51, 8.51); Calibrated: 2017/5/24;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: 1131
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.4 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 103.3 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 16.8 W/kg
SAR(1 g) = 9.63 W/kg; SAR(10 g) = 5.2 W/kg
Maximum value of SAR (measured) = 14.2 W/kg




                    0 dB = 14.2 W/kg = 11.52 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2017/12/19


System Check_Body_1900MHz
DUT: D1900V2-5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_171219 Medium parameters used: f = 1900 MHz; σ = 1.513 S/m; εr = 55.751; ρ =
1000 kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3925; ConvF(8.25, 8.25, 8.25); Calibrated: 2017/5/24;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: 1131
- Measurement SW: DASY52, Version 52.8 (8);SEMCAD X Version 14.6.10 (7373)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 16.0 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 106.0 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 17.8 W/kg
SAR(1 g) = 9.9 W/kg; SAR(10 g) = 5.17 W/kg
Maximum value of SAR (measured) = 15.0 W/kg




                    0 dB = 15.0 W/kg = 11.76 dBW/kg