SAR Appendix 2

FCC ID: AFJ405400

RF Exposure Info

Download: PDF
FCCID_4232522

 Schmid & Partner Engineering AG                               s    £          e        a       g

 Zeughausstrasse 43, 8004 Zurich, Switzerland
 Phone +41 44 245 9700, Fax +41 44 245 9779




                                                                                               874
 info@speag.com, http://www.speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4

The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the E—stop. To prevent Estop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environment if not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configuration file.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:            ®
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                11.12.2009


 Calibration Laboratory of                                                                                         Schweizerischer Kalibrierdienst
 Schmid & Partner                                                                                                  Service suisse d‘étalonnage
   Engineering AG                                                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


 Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

 Client       UlItratech Labs                                                                    Certificate No: DAE4—874_Aug18

[CALIBRATION CERTIFICATE
  Object                                DAE4 — SD 000 DO4 BJ — SN: 874

  Calibration procedure(s)              QA CAL—O6.v29
                                        Calibration procedure for the data acquisition electronics (DAE)



 Calibration date:                      August 14, 2018




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (Sl).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                       ID #                      Cal Date (Certificate No.)                          Scheduled Calibration
 Keithley Multimeter Type 2001           SN: 0810278               31—Aug—17 (No:21092)                                Aug—18

 Secondary Standards                     ID #                      Check Date (in house)                               Scheduled Check
 Auto DAE Calibration Unit               SE UWS 053 AA 1001        O04—Jan—18 (in house check)                         In house check: Jan—19
 Calibrator Box V2.1                     SE UMS 006 AA 1002         04—Jan—18 (in house check)                         In house check: Jan—19




                                        Name                                    Function                               Signature
 Calibrated by:                         Adrian Gehring                         Laboratory Technician                         5:/,%


 Approved by:                           Sven Kihn                              ‘Deputy Manager                           £   /     (fl s uL7
                                                                                                                         iJ          WW

                                                                                                                       Issued: August 14, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—874_Aug18                                        Page 1 of 5


Calibration Laboratory of                                     oS               6      Schweizerischer Kalibrierdienst
Schmid & Partner                                             il;:\\—\—‘//!
                                                                       IEE//;:        Service suisse d‘étalonnage
  Engi neering AG                                            z.//—\\6                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                  7 ///_\\\ y             Swiss Calibration Service
                                                                * uhudluls N



 Accredited by the Swiss Accreditation Service (SAS)                               Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                           data acquisition electronics
Connector angle               information used in DASY system to align probe sensor X to the robot
                              coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
          comparison with a calibrated instrument traceable to national standards. The figure given
          corresponds to the full scale range of the voltmeter in the respective range.
     e    Connector angle: The angle of the connector is assessed measuring the angle
          mechanically by a tool inserted. Uncertainty is not required.
     e    The following parameters as documented in the Appendix contain technical information as a
          result from the performance test and require no uncertainty.
               DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
               Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
                Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
               AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
               Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
               Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
               Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
               Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
               Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—874_Aug18                                  Page 2 of 5


 DC Voltage Measurement
     A/D — Converter Resolution nominal!
         High Range:          1LSB =     6.1uV ,       full range =  —100...+300 mV
         Low Range:           1LSB =      6inV ,       full range =  —1....... +3mV
     DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


       Calibration Factors            KX                           Y                    Z
       High Range            404.303 + 0.02% (k=2)    404.787 + 0.02% (k=2)   404.363 + 0.02% (k=2)
       Low Range             3.97851 + 1.50% (k=2)    4.01725 + 1.50% (k=2)   4.01560 + 1.50% (k=2)



Connector Angle

      Connector Angle to be used in DASY system                                   116.5 °+1 °




Certificate No: DAE4—874_Aug18                       Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)
 1. DC Voltage Linearity
      High Range                               Reading (uV)          Difference (uV)               Error (%)
      Channel X       + Input                   199997.28                   —0.43                  —0.00
      Channel X       + Input                    20002.82                   0.40                   0.00
      Channel X       — Input                   —20000.19                   0.51                   —0.00
      Channel Y       + Input                   199998.66                   0.96                   0.00
      Channel Y       + Input                   20001.51                    —0.89                  —0.00
      Channel Y       — Input                   —20000.87                   —0.01                  0.00
      Channel Z       + Input                   199998.67                   0.43                   0.00
      Channel Z       + Input                   20000.77                    —1.59                  —0.01
      Channel Z       — Input                   —20002.13                   —1.25                  0.01

      Low Range                                Reading (uV)         Difference (uV)               Error (%)
      Channel X       + Input                    2003.07                    1.10                   0.05
      Channel X       + Input                    202.62                     0.41                   0.20
      Channel X       — Input                    —197.10                    0.63                   —0.32
      Channel Y       + Input                    2002.52                    0.63                   0.03
      Channel Y       + Input                    201.77                     —0.22                 —0.11
      Channel Y       — Input                    —198.38                    —0.50                  0.25
      Channel Z       + Input                    2002.31                    0.66                   0.03
      Channel Z       + Input                    200.94                    —0.89                  —0.44

      Channel Z       — Input                    —199.11                   —1.06                   0.54



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                       High Range                           Low Range
                        Input Voltage (mV)            Average Reading (uV)                Average Reading (uV)

     Channel X                          200                      —9.11                         —10.42
                                       — 200                     1215                           10.40
     Channel Y                          200                      3.91                           3.71
                                       — 200                     —4.67                          —5.18
     Channel Z                          200                      4.63                           4.93
                                       — 200                     —7.79                          —7.45



3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)        Channel X (uV)          Channel Y (uV)       Channel Z (uV)

     Channel X                   200                       —                 —0.72                 —3.54
    Channel Y                    200                   5.82                     —                  —0.50
    Channel Z                    200                   8.99                   3.40                      —


Certificate No: DAE4—874_Aug18                             Page 4 of 5


4. AD—Converter Values with inputs shorted
     DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                     High Range (LSB)                     Low Range (LSB)

     Channel X                                               15806                             16071

     Channel Y                                               15909                             15063

     Channel Z                                               16127                             16042



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MCQ

                            Average (uV)        min. Offset (uV)         max. Offset (uv)       *"* '():\‘,’;am"
     Channel X                   —0.33                119                        0.58                 0.38
     Channel Y                   0.59                 —0.37                      3.77                 0.54
     Channel Z                   0.54                 —0.39                      2.08                 0.53

6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                         Measuring (MOhm)
     Channel X                                             200                                 200
     Channel Y                                             200                                 200
     Channel Z                                             200                                 200



8. Low Battery Alarm Voltage (Typical values for information)
     Typical values                               Alarm Level (VDC)
     Supply (+ Veoe)                                                              +7.9

     Supply (— Vec)                                                               —7.6



9. Power Consumption (Typical values for information)
     Typical values                         Switched off (mA¥)       Stand by (mA)          Transmitting (mA)
     Supply (+ Vec)                                +0.01                    +6                     +14
     Supply (— Vec)                               —0.01                     —8                       —9




Certificate No: DAE4—874_Aug18                             Page 5 of 5


 Calibration Laboratory of                                             XR
                                                                           P1 /m
                                                                               N\     Z
                                                                                                                              e             Sb   staeas
                                                                                                                       Schweizerischer Kalibrierdienst
                                                                                                                   S




                                                              §
                                                                    l/’/




                                                                                                  i“‘\\\
 Schmid & Partner                                                                                                  C   Service suisse d‘étalonnage
   Engineering AG                                                                                                      Servizio svizzero di taratura
                                                                                                                   S




                                                                                          I,/",
                                                                    *
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                 7M      .N                               Swiss Calibration Service
                                                                           *Atul u\“\\\


 Accredited by the Swiss Accreditation Service (SAS)                                                               Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

 Client      Ultratech Labs                                                                                Certificate No: EX3—3673_Aug18

CALIBRATION CERTIFICATE                                                                                                                                   |
 Object                                EX3DV4— SN:3673

 Calibration procedure(s)              QA CAL—01.v9, QA CAL—12.v9, QA CAL—14.v4, QA CAL—23.v5,
                                       QA CAL—25.v6
                                       Calibration procedure for dosimetric E—field probes

 Calibration date:                    August 23, 2018

 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (Sl).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70% .


 Calibration Equipment used (M&TE critical for calibration)



 Primary Standards                   ID                                      Cal Date (Certificate No.)                  Scheduled Calibration
 Power meter NRP                     SN: 104778                              04—Apr—18 (No. 217—02672/02673)             Apr—19
 Power sensor NRP—Z91                SN: 103244                              04—Apr—18 (No. 217—02672)                   Apr—19
 Power sensor NRP—Z91                SN: 103245                              04—Apr—18 (No. 217—02673)                  Apr—19
 Reference 20 dB Attenuator          SN: S5277 (20x%)                        04—Apr—18 (No. 217—02682)                  Apr—19
 Reference Probe ES3DV2              SN: 3013                                30—Dec—17 (No. ES3—3013_Dec17)              Dec—18
 DAE4                                SN: 660                                 21—Dec—17 (No. DAE4—660_Dec17)              Dec—18

 Secondary Standards                 ID                                      Check Date (in house)                      Scheduled Check
 Power meter E4419B                  SN: GB41293874                          06—Apr—16 (in house check Jun—18)          In house check: Jun—20
 Power sensor E4412A                 SN: MY41498087                          06—Apr—16 (in house check Jun—18)          In house check: Jun—20
 Power sensor E4412A                 SN: 000110210                           06—Apr—16 (in house check Jun—18)          In house check: Jun—20
 RF generator HP 8648C              .SN: US3642U01700                        04—Aug—99 (in house check Jun—18)          In house check: Jun—20
 Network Analyzer E8358A             SN: US41080477                          31—Mar—14 (in house check Oct—17)          In house check: Oct—18

                                       Name                                         Function                              Signature
Calibrated by:                         Jeton Kastrati                               Laboratory Technician                          V


Approved by:                           Katia Pokovic                                Technical Manager
                                                                                                                         yeestec
                                                                                                                          Issued: August 24, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: EX3—3673_Aug18                                        Page 1 of 11


 Callbration Laboratory of                                      \x‘\‘@/’:&,                S   Schweizerischer Kalibrierdienst
 Schmid & Partner                                             ;E"\\\‘-—'—'//El
                                                                          IE/?             C   Service suisse d‘étalonnage
    Engineering AG                                            T m cla                          Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                  ";/,,//7——/:'\\\\;\\3“      S   Swiss Calibration Service
                                                                    nhatis


 Accredited by the Swiss Accreditation Service (SAS)                                        Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

 Glossary:
TSL                           tissue simulating liquid
NORMx,y,z                     sensitivity in free space
ConvF                         sensitivity in TSL / NORMx,y,z
DCP                           diode compression point
CF                            crest factor (1/duty_cycle) of the RF signal
A, B, C, D                    modulation dependent linearization parameters
Polarization o                @ rotation around probe axis
Polarization 8                8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                              ie., 8 = 0 is normal to probe axis
Connector Angle               information used in DASY system to align probe sensor X to the robot coordinate system

Calibration is Performed According to the Following Standards:
     a)   IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
          Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
          Techniques", June 2013
     b)   IEC 62209—1, ", "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from hand—
          held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)", July 2016
    c)    EC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
          used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March 2010
    d)    KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
          NORMx,y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; f > 1800 MHz: R22 waveguide).
          NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E®—field
          uncertainty inside TSL (see below ConvF).
          NORM(P)x,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
          implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
          in the stated uncertainty of Convr.
          DCPx,y,z:;: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
          signal (no uncertainty required). DCP does not depend on frequency nor media.
          PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
          characteristics
          Ax,y,z; Bx,y,z; Cx,y,z; Dx,y,z; VRx,y,z: A, B, C, D are numerical linearization parameters assessed based on
          the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
          media. VR is the maximum calibration range expressed in RMS voltage across the diode.
          ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
          Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
          measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
          boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
          used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
          to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
          ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
          MHz.
          Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
          exposed by a patch antenna.
          Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
          (on probe axis). No tolerance required.
          Connector Angle: The angle is assessed using the information gained by determining the NORMx ({no
          uncertainty required).


Certificate No: EX3—3673_Aug18                                  Page 2 of 11


 EX3DV4 — SN:3673                                                               August 23, 2018




                       Probe EX3DV4

                                          SN:3673

                                  Manufactured:             September 9, 2008
                                  Calibrated:               August 23, 2018



                                 Calibrated for DASY/EASY Systems
                                    (Note: non—compatible with DASY2 system!)




Certificate No: EX3—3673_Aug18                    Page 3 of 11


EX3DV4— SN:3673                                                                                                                August 23, 2018




 DASY/EASY — Parameters of Probe: EX3DV4 — SN:3673
Basic Calibration Parameters
                                                 Sensor X                        Sensor Y                     Sensor Z             Unc (k=2)
  Norm (uV/(V/m))*                                 0.53                            0.47                         0.46               +10.1 %
  DCP (mV)"                                        98.7                           102.5                         95.3

Modulation Calibration Parameters
  UID           Communication System Name                                   A             B          C           D         VR          Unc®
                                                                           dB        dBVuV                      dB         mV          (k=2)
 0              Cw                                                xX       0.0          0.0         1.0        0.00       1784        +3.8 %
                                                                  Y        0.0          0.0         1.0                   186.9
                                                                  Z        0.0          0.0         1.0                   193.8


 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


* The uncertainties of Norm X,Y,Z do not affect the E*—field uncertainty inside TSL (see Pages 5 and 6).
® Numerical linearization parameter: uncertainty not required.
& Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: EX3—3673_Aug18                                    Page 4 of 11


 EX3DV4— SN:3673                                                                                                               August 23, 2018




 DASY/EASY — Parameters of Probe: EX3DV4 — SN:3673


Calibration Parameter Determined in Head Tissue Simulating Media
                         Relative         Conductivity                                                                 Depth °          Unc
      f (MHz)®©       Permittivity "         (S/m)"             ConvF X     ConvFY         ConvFZ        Aipha®         (mm)           (k=2)

         150               52.3                0.76              10.71        10.71          10.71         0.00         1.00         + 13.3 %

         450               43.5                0.87              10.14        10.14          10.14         0.13         1.20         + 13.3 %

         600               42.7                0.88              9.28          9.28          9.28          0.10         1.20         + 13.3 %

         835               41.5                0.90              9.39          9.39          9.39          0.53         0.80        + 12.0 %

         900               41.5                0.97              9.08          9.08          9.08          0.52         0.80        + 12.0 %

        1640               40.2                1.31              8.04          8.04          8.04          0.37         0.86        + 12.0 %

        1810               40.0                1.40              7.87          7.87          7.87          0.31         0.87        + 12.0 %

        2450               39.2                1.80              7.23         7.23           7.23          0.35         0.89        + 12.0 %

        5200               36.0                4.66              4.86         4.86           4.86          0.40         1.80        + 13.1 %

        5300              35.9                 4.76              4.68         4.68           4.68          0.40         1.80        +13.1 %

        5500              35.6                 4.96              4.94         4.94           4.94          0.40         1.80        + 13.1 %

        5600              35.5                 5.07              4.78         4.78           4.78          0.40         1.80        +13.1 %

        5800              35.3                 5.27              4.38         4.38           4.38          0.40         1.80        + 13.1 %


© Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and o) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3673_Aug18                                     Page 5 of 11


 EX3DV4— SN:3673                                                                                                                August 23, 2018




 DASY/EASY — Parameters of Probe: EX3DV4 — SN:3673


 Calibration Parameter Determined in Body Tissue Simulating Media
                         Relative         Conductivity                                                                 Depth °          Unc
     _f(MHz)°         |_Permittivity"         (S/m)"          ConvF X        ConvF Y       ConvFZ         Alpha®        _(mm)          (k=2)
         150               61.9                0.80             10.55         10.55          10.55         0.00         1.00         + 13.3 %
         450               56.7                0.94             10.31         10.31          10.31         0.08         1.20         +13.3 %
         600               56.1                0.95             9.89           9.89          9.89          0.10         1.20         +13.3 %
         835               55.2                0.97             9.15           9.15          9.15          0.46         0.86         +12.0 %
         900               55.0                1.05             9.31           9.31          9.31          0.47         0.85         +12.0 %
        1640               53.7                1.42             8.05           8.05          8.05          0.39         0.86         +12.0 %
        1810               53.3                1.52             7.52           7.52          7.52          0.40         0.86        + 12.0 %
        2450               52.7                1.95             7.18           7.18          7.18          0.37         0.88        +12.0 %
        5200              49.0                 5.30             4.33          4.33           4.33          0.50         1.90        +131 %
        5300              48.9                 5.42             4.16          4.16           4.16          0.50         1.90        +131 %
        5500              48.6                 5.65             4.30          4.30           4.30          0.50         1.90        +131 %
        5600              48.5                 5.77             4.13          4.13           4.13          0.50         1.90        +131 %
        5800              48.2                 6.00             3.89          3.89           3.89          0.50         1.90        +131 %


C Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and 0) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3673_Aug18                                     Page 6 of 11


 EX3DV4— SN:3673                                                                                                                                                                                                 August 23, 2018


                                                                         Frequency Response of E—Field.
                                                                                   (TEM—Cell:ifi110 EXX, Waveguide: R22)




                                             hn
           Frequency response {normalized)




                                             1A PfrrrminsntressssressoespororrrinnstrmmsrronsaefersmmnmrmenmmesirtsferrnrnnnrnsrserereceeesJ
                                             1Q]mlSmtenetum._frmirrrrRiefrrmereercoomeemunipecee
                                                                                        oi       nm m enrerfrmesineig oocBo++
                                             ) O ] essm ne en en nncb nn snn nenennsnne en en onchoon en en enonsesns eeenend en neeee e en se nee en en enchons en eenenee ns enn snn edecneee enenen snn e en dene en




                                                                                         rem                                                                              a
                                                                     Uncertainty of Frequency Response of E—field: *# 6.3% (k=2)




Certificate No: EX3—3673_Aug18                                                                                       Page 7 of 11


 EX3DVA4— SN:3673                                                                                                                                                                                                                                                                                                       August 23, 2018




                                                                                   Receiving Pattern (¢), 8 = 0°

                                              f=600 MHz,TEM                                                                                                                                                                     f=1800 MHz,R22
                                                                     e                                                                                                                                                                                        90_

                           135                                   yA 0 OA                                                           45                                                                    135                                              "*"**"**                     _                            45
                            &                           "*;                                       s 5o                                  9                                                                                                se¥ *                                        *a & ~*
                                                        %                      «        *                        *        *                                                                          J                      *               #                         A /.                         &
                                     ®                                         h                                                   *                                                                 j              s                                                 #                                             *
                                                        a                                                        *                       \                                                       &                                          &                     5       ‘-’                      4                    %


                                .!            x~~*               *        *                                  ‘_'                               ’    4             1                          *                 ';                    w *#                     |                            a®,
                                *+                               w        *L                  *                      &                      f                                       ‘                          4 ;                                        *a iL                            &                            i
               180—             Ck 3                                      :Q 6              %                                          ‘i\‘.            IO                       1 80                          + ..\-                                     :       .‘              *                                 #         it
                                                             .            * .                   D&           040     osvVaos                            .                           $                     $o                    u.                            ** "o2,                           oc         oe           0s
                            (                   *z>~                               y **%ec®                          e                              T                                   to               d                           |          ge   4*                           ***ig.%..                4e
                      ro    s                   4                         *J                                 f                     f                                                                     *+                                1
                                &                                                                                              ®                                                                                                                                  #       .                                     &   ‘
                                                    l                 #        &    .                        l           x ‘                                                                                                               1                                                       *
                                              e«oly _«                                  »..       k 4e                                                                                                                                   e b.zs                                                .       o



                           205 _                                 es lc uze *                                              L os5s                                                                         235 _                                       *n                   le@**                             —o345

                                                                     ~~3%                                                                                                                                                                                 ~3%0
                e                                        e                                               e                                          e                              e                                                            &                                          e                                 e
                Tot                                     X                                                Y                                          2Z                           Tot                                                            X                                          Y¥                                Z




             0.5——
       o                                                                                                                                                                                                                                                                                                                i
       vou                                                                                                                                                                                                                                                                                                              i
       5     0.047
                                                                                                                                                                                                                                                                                                                        i
       i                                                                                                                                                                                                                                                                                                                i
       Lid                                                                                                                                                                                                                                                                                                              i
             —0.5—                                                                                                                                                                                                                                                                                                      f
                     [                    ]     L             uU                                     p           O                                  $        Lfi          I        £_KLJI                                J                _3..3._.}                            |       Lo                                ]    1|
                                         —150                                      —100                                                     —50                          0                                          50                                                100                                           150
                                                                                                                                                                      Roll ["]                           8
                                                *#                                                                                                 [#                                                    _*®
                                              100 MHz                                                                              600 MHz                                                           1800 MHz                                                                                  2500 MHz



                                                                     Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificate No: EX3—3673_Aug18                                                                                                                               Page 8 of 11


 EX3DV4—SN 3673                                                                                                        August 23, 2018




                                                           Dynamic Range f(SARneaq)
                  Input Signal [uV]
                                                                    (TEM cell , feyai= 1900 MHz)




                                                                                        C
                                                                                       O
                                                                                       —A




                                                                                 SAR [mW/icm3]

                                                             not compensated                       compensated
                                      Error [dB]




                                                   103        102        10—1          109       101      102    103
                                                                                SAR [mW/cm3]

                                                             not compensated                     compensated


                                                         Uncertainty of Li nearlity Assessment: + 0.6% (k=2)




Certificate No EX3—3673_Aug18                                                   Page 9 of 11


 EX3DV4— SN:3673                                                                                                                                                                                                            August 23, 2018


                                                           Conversion Factor Assessment
                           f= 835 MHz,WGLS RYQ ({H_convF)                                                                                    f= 1810 MHz,WGLS R22 (H_convF)


             3.5
                       |                                                                                                       »|      i



             30—it*                                                                                                            25|13


             25 [                                                                                                              20 ‘        '.'
        3.                                                                                                                 B          Ob     *
        @           0                                                                                                      &          |
        3 20—                                                                                                              3
        >    1                                                                                                             é 15{
        ®o.0                                       &                                                                       B    +\
          150
                                                                                                                               10|
             10                                                   ho 8.                                                            1


                                                                                                                               50                                                   Tv~«
             0.5—                                                                                          +.                                                                                        ty
                                                                                                                                  |                                                                               ns.
                                                                                                                                  |                                                                                               *~         hove
             0.0                                   t        bropondommininmtnerednd                   dolemiad.|               Lup Ad...srntomopomivsidhndormmmemecdeni       Sriol oph m Amals wb        Aroolleoningnnndmeoob   sinttaidehomdonite
                   0       5      10                   1     20            2w            30      35             43               0               5          0             1          20              25           30              35             40
                                                           z {mm]                       e                                                                 x)                     z {mmj                         (%)
                                   #                                                    : 0                                                               | &                                                      @
                               analytical                                             measured                                                         analytical                                           measured



                                                       Deviation from Isotropy in Liquid
                                                                                          Error (6, 9), f = 900 MHz
                                       Deviation




                                           —1.0            —0.8         —0.86            —0.4     —0.2               0.0       02                0.4         0.6              0.8           1.0
                                                   Uncertainty of Spherical Isotropy Assessment: + 2.6% (k=2)




Certificate No: EX3—3673_Aug18                                                                          Page 10 of 11


 EX3DV4— SN:3673                                            August 23, 2018




 DASY/EASY — Parameters of Probe: EX3DV4 — SN:3673

Other Probe Parameters
  Sensor Arrangement                                           Triangular
 Connector Angle (°)                                                 16.9
 Mechanical Surface Detection Mode                               enabled
 Optical Surface Detection Mode                                 disabled
 Probe Overall Length                                            337 mm
 Probe Body Diameter                                              10 mm
 Tip Length                                                        9 mm
 Tip Diameter                                                    2.5 mm
 Probe Tip to Sensor X Calibration Point                           1 mm
 Probe Tip to Sensor Y Calibration Point                           1 mm
 Probe Tip to Sensor Z Calibration Point                           1 mm
 Recommended Measurement Distance from Surface                   1.4 mm




Certificate No: EX3—3673_Aug18              Page 11 of 11


 Calibration Laboratory of                                                                                    S    Schweizerischer Kalibrierdienst
 Schmid & Partner                                                                                             Cc   Service suisse d‘étalonnage
   Engineering AG                                                                                                  Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                 S    swiss Calibration Service


 Accredited by the Swiss Accreditation Service (SAS)                                                          Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

 Client       Ultratech Labs                                                                       Certificate No: CLA150—4006_Aug16

 CALIBRATION CERTIFICATE
  Object                                CLA150 — SN: 4006


  Calibration procedure(s)              QA CAL—15.v8
                                        Calibration procedure for system validation sources below 700 MHz



  Calibration date:                     August 24, 2016


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                       ID #                      Cal Date (Certificate No.)                 s       Scheduled Calibration
 Power meter NRP                         SN: 104778                06—Apr—16 (No. 217—02288/02289)                    Apr—17
 Power sensor NRP—Z91                    SN:    108244             06—Apr—16   (No.   217—02288)                      Apr—17
 Power sensor NRP—Z91                    SN:    103245             06—Apr—16   (No.   217—02289)                      Apr—17
 Reference 30 dB Attenuator              SN:    5129 (3Ob)         05—Apr—16   (No.   217—02294)                      Apr—17
 Type—N mismatch combination             SN:    5047.2 / 06327     05—Apr—16   (No.   217—02295)                      Apr—17
 Reference Probe EX3DV4                  SN: 3877                  31—Dec—15 (No. EX3—3877_Dect5)                     Dec—16
 DAE4                                    SN: 654                   12—Aug—16 (No. DAE4—654_Aug16)                     Aug~17

 Secondary Standards                     ID #                      Check Date (in house)                              Scheduled Check
 Power meter E44198                     SN:     GB41203874         06—Apr—16 (No. 217—02285/02284)                    In house check: Jun—18
 Power sensor E4412A                    SN:     MY41498087         06—Apr—16 (No. 217—02285)                          In house check: Jun—18
 Power sensor E4412A                    SN:     000110210          06—Apr—16 (No. 217—02284                           In house check: Jun—18
 RF generator HP 8648C                  SN:     US3642U01700       04—Aug—99 (in house check Jun—16)                  In house check: Jun—18
 Network Analyzer HP 8753E              SN: US37390585             18—Oct—01 (in house check Oct—15)                  In house check: Oct—16


                                        Name                                    Function                              Signature
 Calibrated by:                         Jeton Kastrati                          Laboratory Technician "~~*



 Approved by:                           Katia Pokovic                           Technical Manager



                                                                                                                     Issued: August 26, 2016
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: CLA150—4006_Aug16                                     Page 1 of 8


 Calibration
       .     Laboratory of                                    swl />,
                                                                \G\/          S   Schweizerischer Kalibrierdienst
 Schmid & Partner                                            ifi//mfi           c   Service suisse d‘étalonnage
    Englneering AG                                           proy                 Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                44/,///fi\\\\\}   S   Swiss Calibration Service
                                                                  talnts®



 Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

 Glossary:
 TSL                         tissue simulating liquid
 ConvF                       sensitivity in TSL / NORM x.y,z
 N/A                         not applicable or not measured

 Calibration is Performed According to the Following Standards:
     a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
        Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
        Communications Devices: Measurement Techniques", June 2013
     b) 1EC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
        devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
        February 2005
     c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
        communication devices used in close proximity to the human body (frequency range of 30
        MHz to 6 GHz)", March 2010
     d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The source is mounted in a touch configuration below the
     center marking of the flat phantom.
   * Return Loss: This parameter is measured with the source positioned under the liquid filled
     phantom (as described in the measurement condition clause). The Return Loss ensures low
     reflected power. No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
  * SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
  «_ SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.

The reported uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.




Certificate No: CLA150—4006_Aug16                              Page 2 of 8


 Measurement Conditions
     DASY system configuration, as far as not given on page 1.
       DASY Version                                               DASY5                               V52.8.8
       Extrapolation                                  Advanced Extrapolation
       Phantom                                             EL4 Flat Phantom               Shell thickness: 2 + 0.2 mm
       EUT Positioning                                        Touch Position
      Zoom Scan Resolution                         dx, dy = 4.0 mm, dz =1.4 mm          Graded Ratio = 1.4 (Z direction)
       Frequency                                           150 MHz + 1 MHz



Head TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                    22.0 °C           52.3               0.76 mho/m

      Measured Head TSL parameters                               (22.0 + 0.2) °C    50.2 26 %        | 0.76 mho/im +6 %
      Head TSL temperature change during test                       <0.5 °C


SAR result with Head TSL

      SAR averaged over 1 cm* {1 g) of Head TSL                    Condition
      SAR measured                                             1 W input power                    3.85 W/kg
      SAR for nominal Head TSL parameters                     normalized to 1W          3.82 Wikg + 18.4 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                  condition
      SAR measured                                             1 W input power                 2.57 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W          2.55 Wikg + 18.0 % (k=2)

Body TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity          Conductivity

      Nominal Body TSL parameters                                   22.0 °C           61.9               0.80 mho/m
      Measured Body TSL parameters                               (22.0 + 0.2) °C   61.1 +6 %          0.84 mho/m x 6 %
      Body TSL temperature change during test                       <0.5 °C


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
      SAR measured                                             1 W input power                    4.01 W/kg

    'L SAR for nominal Body TSL parameters             |      normalized to 1W         3.85 W/kg * 18.4 %(k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                            1 W input power                 2.67 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W         2.57 Wikg * 18.0 %(k=2)



Certificate No: CLA150—4006_Aug16                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

       Impedance, transformed to feed point                  43.6 Q — 5.8 jQ

      Retum Loss                                               —20.7 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                  48.3 Q — 6.3 jQ
      Return Loss                                              —23.6 dB



Additional EUT Data

      Manufactured by                                           SPEAG
      Manufactured on                                       August 23, 2013




Certificate No: CLA150—4006_Aug16             Page 4 of 8


 DASYS5 Validation Report for Head TSL

                                                                                      Date: 24.08.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: CLA150; Type: CLA150; Serial: CLA150 — SN: 4006

Communication System: UID 0 — CW; Frequency: 150 MHz
Medium parameters used: £ = 150 MHz; 0 = 0.76 $/m; &, = 50.2; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN3877; ConvF(12.02, 12.02, 12.02); Calibrated: 31.12.2015;

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n654; Calibrated: 12.08.2016

        Phantom: ELI v4.0; Type: QDOVAOO1BB; Serial: TP:1003

        DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


CLA Calibration for HSL—LF Tissue/CLA150, touch configuration, Pin=1W/Area Scan
(81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 5.34 W/kg

CLA Calibration for HSL—LF Tissue/CLA150, touch configuration, Pin=1W/Zoom Scan,
dist=1.4mm (8x10x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 84.06 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) =7.10 W/kg
SAR(I g) = 3.85 W/kg; SAR(10 g) = 2.57 W/kg
Maximum value of SAR (measured) = 5.36 W/kg


       dB
        0

        —2.96

        —5.93

        —8.89

        —11.86

        —14.82
                   0 dB = 5.34 W/kg = 7.28 dBW/kg



Certificate No: CLA150—4006_Aug16              Page 5 of 8


 Impedance Measurement Plot for Head TSL




                                                                                 24 fug 2016 11:27:18
           CHI Sii       1 U FS           1143.555 0      —5.9105 0   182.60 pF               150.000 800 MHz


           #


                                                                                      4

                                          |                  1               f            *
           Cor                                                   —

                                          i—          i    ze gwl




                                                 %%        N mu/ 4
          Hld                                         in>
           CH2   Si1 _   LO6 3,5 QSI_F_REF‘42’_[51_3____.~__ 1='Z§;<?2_CLB_._15.3_:_.9!3._975@_715.2,_[

                          +       fi——NT           t »-—»—T|_~__1_:___                     cwe..s 4—-——}»——]
                                                                         !            [           |


          Cor




          Aiv
          16°
                                                                                              |
          H1d                                                                         s _.T,

                                                                                                      1
                                                                                   ~ STOP 200.000 080 MHz




Certificate No: CLA150—4006_Aug16                          Page 6 of 8


 DASY5 Validation Report for Body TSL

                                                                                      Date: 24.08.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: €CLA150; Type: CLA150; Serial: CLA150 — SN: 4006

Communication System: UID 0 — CW; Frequency: 150 MHz
Medium parameters used: f = 150 MHz; 0 = 0.84 S/m; & = 61.1; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    *   Probe: EX3DV4 — SN3877; ConvF(11.44, 11.44, 11.44); Calibrated: 31.12.2015;

    &   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e   Electronics: DAF4 $n654; Calibrated: 12.08.2016

    &   Phantom: ELI v4.0; Type: QDOVAOOIBB; Serial: TP:1003

    &   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


CLA Calibration for MSL—LF Tissue/CLA150, touch configuration, Pin=1W/Area Scan
(81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 5.62 W/kg

CLA Calibration for MSL—LF Tissue/CLA150, touch configuration, Pin=1 W/Zoom Scan,
dist=1.4mm (8x10x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = §1.98 V/m; Power Drift =—0.06 dB
Peak SAR (extrapolated) =7.47 W/kg
SAR(I g) = 4.01 W/kg; SAR(10 g) = 2.67 W/kg
Maximum value of SAR (measured) = 5.59 W/kg




        —2.93

        —5.85

        —8.78

        11.70

        —14.63
                   0 dB = 5.62 W/kg = 7.50 dBW/kg



Certificate No: CLA150—4006_Aug16              Page 7 of 8


 Impedance Measurement Plot for Body TSL




                                                                             24 fAug 2016        11145:35
           (CHI)   Si1     i U FS          1148.273 6     ~6.3425 e   168.05 pF                 150.000 000 MHz


           #



                                                                 .                    &
           tor                                                   w                     t
                                                                                      ai



           Av                                                                         I/
           159                                                                    4



           H1d




                   *                                                                           y"*7"
           CH2     $41     LOG                                                            dB    159.000 000 MH

                       .._{fsss=                  s                                            _4      sife=s
                            j                                                ob                                   :
                   ———+                                                                —f——!—
                   |                                                                                                  x
          Cor
                   ...i                                                                   c cafecsp_

                   fo————{———+—             .         4   ——I4           f..._L                 +.    _j

          s [—|—]—f—i—f—[   }       f.._.%._f
                                                                                      —f—<4
                                                                         ———|—_—_——y__—+.                        _|
          Hla               4    .——.—T—.%                                       aallws._.. 5k L_~ j
                            1                         1      |           |
                         START 160.000 d6n MHz                                        STOP 200.000 on MHz




Certificate No: CLA150—4006_Aug16                          Page 8 of 8



Document Created: 2018-10-15 15:49:18
Document Modified: 2018-10-15 15:49:18

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC