FZ-L1AC_RFExp_Appendix A

FCC ID: ACJFZL1A

RF Exposure Info

Download: PDF
FCCID_4086036

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/10/30


System Check_Body_2450MHz

DUT: D2450V2—736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450_181030 Medium parameters used: £ = 2450 MHz; 0 = 2.024 S/m; s = 52.188; p =
1000 kg/m>
Ambient Temperature : 23.8 °C ; Liquid Temperature : 22.8 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3925; ConvF(7.63, 7.63, 7.63) ; Calibrated: 2018/5/31
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1326; Calibrated: 2018/9/18
— Phantom: ELI v4.0; Type: QDOVAOOIBB; Serial: 1026
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 22.2 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.3 V/m; Power Drift = 0.10 dB
Peak SAR (extrapolated) = 27.5 W/kg
SAR(I g) = 13.6 W/kg; SAR(10 g) = 6.32 W/keg
Maximum value of SAR (measured) = 22.5 W/kg

        dB
         0


        —4.29


        —8.58


        —12.86


        —17.15


        —21.44
                    0 dB = 22.5 W/kg = 13.52 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/10/30


System Check_Body_5250MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_181030 Medium parameters used: £ = 5250 MHz; 0 = 5.457 S/m; s = 49.208; p = 1000
kg/m3
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3925; ConvF(4.44, 4.44, 4.44) ; Calibrated: 2018/5/31
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1326; Calibrated: 2018/9/18
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1):; Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.2 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 70.17 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 31.2 W/kg
SAR(1 g) =8.15 W/kg; SAR(10 g) = 2.22 W/kg
Maximum value of SAR (measured) = 19.4 W/kg

        dB
         0


        —10.00


        —20.00


        —30.00


        —40.00


        —50.00
                    0 dB = 19.4 W/kg = 12.88 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/1


System Check_Body_5250MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_181101 Medium parameters used: £= 5250 MHz; 0 = 5.443 S/m; g, = 49.078; p = 1000
kg/m3
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3642; ConvF(3.87, 3.87, 3.87) ; Calibrated: 2018/4/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2017/11/16
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1):; Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 18.1 W/kg

Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 62.93 V/m; Power Drift =0.13 dB
Peak SAR (extrapolated) = 29.4 W/kg
SAR(I g) =7.77 W/kg; SAR(10 g) =2.2 W/kg
Maximum value of SAR (measured) = 18.6 W/kg

        dB
         0


        —9.02


        —18.04


        —27.07


        —36.09


        —45.11
                    0 dB = 18.6 W/kg = 12.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/10/30


System Check_Body_5600MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: MSL_5G_181030 Medium parameters used: £ = 5600 MHz; 0 = 5.943 S/m; s = 48.579; p = 1000
kg/m3
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3925; ConvF(4.08, 4.08, 4.08) ; Calibrated: 2018/5/31
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1326; Calibrated: 2018/9/18
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1):; Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 19.9 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.14 V/m; Power Drift =0.12 dB
Peak SAR (extrapolated) = 36.1 W/kg
SAR(1 g) =8.03 W/kg; SAR(10 g) = 2.24 W/kg
Maximum value of SAR (measured) = 21.1 W/kg

        dB
         0


        —10.00


        —20.00


        —30.00


        —40.00


        —50.00
                    0 dB =21.1 W/kg = 13.24 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/1


System Check_Body_5600MHz

DUT: DSGHzV2—1006

Communication System: CW ; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: MSL_5SG_181101 Medium parameters used: £= 5600 MHz; 0 = 5.927 S/m; g, = 48.449; p = 1000
kg/m3
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3642; ConvF(3.38, 3.38, 3.38) ; Calibrated: 2018/4/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1399; Calibrated: 2017/11/16
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1):; Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.6 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 62.54 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 35.1 W/kg
SAR(1 g) =8.31 W/kg; SAR(10 g) = 2.32 W/keg
Maximum value of SAR (measured) = 21.5 W/kg

        dB
         0


        —8.22


        —16.44


        —24.67


        —32.89


        —41.11
                    0 dB =21.5 W/kg = 13.32 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/10/30


System Check_Body_5750MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: MSL_5G_181030 Medium parameters used: £ = 5750 MHz; 0 = 6.162 S/m; s = 48.312; p = 1000
kg/m3
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3925; ConvF(4.17, 4.17, 4.17) ; Calibrated: 2018/5/31
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 Sn1326; Calibrated: 2018/9/18
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1):; Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.0 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 68.79 V/m; Power Drift = —0.15 dB
Peak SAR (extrapolated) = 37.6 W/kg
SAR(1 g) = 8.25 W/kg; SAR(10 g) = 2.17 W/kg
Maximum value of SAR (measured) = 21.6 W/kg

        dB
         0


        —7.06


        —14.12


        —21.17


        —28.23


        —35.29
                    0 dB =21.6 W/kg = 13.34 dBW/kg