Calibration certificate probe 3885

FCC ID: A4R-X1

RF Exposure Info

Download: PDF
FCCID_1910693

Calibration Laboratory of                                           wl/7
                                                                         v5jH7


                                                                                                            S      Schweizerischer Kalibrierdienst




                                                             §                    a8
Schmid & Partner                                                   >




                                                                                     T
                                                                    ctmmrertedl
                                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                  Eeez                                             Servizio svizzero di taratura
                                                                                                            S




                                                                                     tu}
Zeughausstrasse 43, 8004 Zurich, Switzerland




                                                                                 ’//,/
                                                                   TA    *                                         Swiss Calibration Service
                                                                     ”’"/nl uy

Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client




Object



Calibration procedure(s)




Calibration date:


This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)



 Primary Standards                   ID                                  Cal Date (Certificate No.)                   Scheduled Calibration
 Power meter E4419B                  GB41293874                          29—Mar—12 (No. 217—01508)                    Apr—13
 Power sensor E4412A                 MY41498087                          29—Mar—12 (No. 217—01508)                    Apr—13
 Reference 3 dB Attenuator           SN: $5054 (3c)                      27—Mar—12 (No. 217—01531)                    Apr—13
 Reference 20 dB Attenuator          SN:   S5086 (20b)                   27—Mar—12 (No. 217—01529)                    Apr—13
 Reference 30 dB Attenuator          SN:   $5129 (30b)                   27—Mar—12 (No. 217—01532)                    Apr—13
 Reference Probe ES3DV2              SN:   3013                          29—Dec—11 (No. ES3—3013_Dec11)               Dec—12
 DAE4                                SN:   660                           20—Jun—12 (No. DAE4—660_Jun12)               Jun—13


 Secondary Standards                 ID                                  Check Date (in house)                        Scheduled Check
 RF generator HP 8648C               US3642U01700                        4—Aug—99 (in house check Apr—11)             In house check: Apr—13
 Network Analyzer HP 8753E           US37390585                          18—0ct—01 {in house check Oct—11)            in house check: Oct—12

                                          Name                                     Function                             Signature
 Calibrated by:




 Approved by:




                                                                                                                        Issued: October 9, 2012
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: EX3—3885_Oct12                                          Page 1 of 11


                                                                      s Euy
 Calibration Laboratory of                                      cMOUIZ                          Schweizerischer Kalibrierdienst




                                                             JE
                                                               T




                                                                                  a®
 Schmid & Partner                                                   \\__/
                                                                                                Service suisse d‘étalonnage
   Engineering AG                                                   zn                          Servizio svizzero di taratura




                                                              €\
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                                   Swiss Calibration Service




                                                                                  77
                                                               8x
                                                                4./              US
                                                                    ,'/"lu 1 u\“\\\



  Accredited by the Swiss Accreditation Service (SAS)                                        Accreditation No.: SCS 108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multilateral Agreement for the recognition of calibration certificates

 Glossary:
——Tok:———                     tissue simulating diquid
 NORMx,y,z                     sensitivity in free space
 ConvF                         sensitivity in TSL / NORMx,y,z
 DCP                           diode compression point
 CF                            crest factor (1/duty_cycle) of the RF signal
 A, B, C                       modulation dependent linearization parameters
 Polarization q                @ rotation around probe axis
 Polarization 8                8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                               i.e., 8 = 0 is normal to probe axis

 Calibration is Performed According to the Following Standards:
      a)   IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
           Absorption Rate (SAR) in the Human Head from Wireiless Communications Devices: Measurement
           Techniques", December 2003
      b)   IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held devices used in close
           proximity to the ear (frequency range of 300 MHz to 3 GHz)", February 2005

 Methods Applied and Interpretation of Parameters:
           NORMx,y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; f > 1800 MHz: R22 waveguide).
           NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E—field
           uncertainty inside TSL (see below ConvF).
           NORM(Px,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
           implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
           in the stated uncertainty of ConvF.
           DCPx,y,z: DGP are numerical linearization parameters assessed based on the data of power sweep with CW
           signal (no uncertainty required). DCP does not depend on frequency nor media.
           PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
           characteristics
           Ax,y,z; Bx,y,z; Cx.y,z, VRx,y,z: A, B, C are numerical linearization parameters assessed based on the data of
           power sweep for specific modulation signal. The parameters do not depend on frequency nor media. VR is the
           maximum calibration range expressed in RMS voltage across the diode.
           ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
           Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
           measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
           boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
           used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
           to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
           ConvF is used in DASY version 4.4 and higher which allows extending the validity from * 50 MHz to £ 100
           MHz.
           Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
           exposed by a patch antenna.
            Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
            {on probe axis). No tolerance required.




  Certificate No: EX3—3885_Oct12                                      Page 2 of 11


EX3DV4 — SN:3885                                                                October 9, 2012




                     Probe EX3DV4

                                         SN:3885

                                  Manufactured:            April 30, 2012
                                  Calibrated:              October 9, 2012




                                 Calibrated for DASY/EASY Systems
                                    (Note: non—compatible with DASY2 system!)




Certificate No: EX3—3885_Oct12                     Page 3 of 11


EX3DV4— SN:3885                                                                                                              October 9, 2012




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3885

Basic Calibration Parameters
                                                Sensor X                         Sensor Y                   Sensor Z             Unc (k=2)
 Norm (uV/(V/im)")*                               0.42                             0.41                       0.28               £10.1 %
 DCP (mV)"                                        105.3                            98.0                       103.4



Modulation Calibration Parameters
 UID           Communication System Name                         PAR                 A            B           C           VR         Unc®
                                                                                     dB          dB           dB          mV         (k=2)
 0             Cw                                                0.00       x       0.00         0.00        100        1455        +3.0%
                                                                            ¥y      0.00         0.00        1.00       142.6
                                                                            z       0.00         0.00        1.00       151.1



 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


* The uncertainties of NormX,Y,Z do not affect the E*—field uncertainty inside TSL (see Pages 5 and 6).
® Numerical linearization parameter: uncertainty not required.
 Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: EX3—3885_Oct12                                      Page 4 of 11


EX3DV4— SN:3885                                                                                                                October 9, 2012




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3885


Calibration Parameter Determined in Head Tissue Simulating Media
                        Relative         Conductivity                                                                 Depth           Unct.
     f (MHz) ©       Permittivity"           ($/m)"          ConvF X        ConvF Y       ConvFZ         Alpha         {mm)           (Kk=2)

        450            T435 ° CCC             0.87              (9.85 ~C|~~9.85             9.85         7(012 T       CMMO CC £#13M4% CCC

        750               41.9                0.89              9.68          9.68           9.68         0.80         0.50         + 12.0 %

        835               41.5                0.90              9.35          9.35           9.35         0.80         0.50         + 12.0 %

        900               41.5                0.97              9.16          9.16           9.16         0.51         0.67         + 12.0 %

       1450               40.5                1.20              8.19          8.19           8.19         0.52         0.75         +12.0 %

       1640               40.3                1.29              8.19          8.19           8.19         0.65         0.62         +# 12.0 %

       1750               40.1                1.37              7.98          7.98           7.98         0.67         0.60         £12.0 %

       1900               40.0                1.40              7.66          7.66           7.66         0.80         0.58         + 12.0 %

       1950               40.0                1.40              7.44          7.44           7.44         0.45         0.73         + 12.0 %

       2000               40.0                1.40              7.62          7.62           7.62         0.74         0.60         + 12.0 %

       2300               39.5                1.67              7.34          7.34           7.34          0.61        0.65         £#12.0 %

       2450               39.2                1.80              6.97          6.97           6.97          0.41        0.81         £12.0 %

       2600               39.0                 1.96             6.80          6.80           6.80          0.43        0.80         * 12.0 %

       3500               37.9                2.91              6.81          6.81           6.81          0.76        0.63         £+13.1 %

       3700               37.7                3.12              6.36          6.36           6.36          0.45        0.83         *#13.1 %

       4950               36.3                4.40              5.18          5.18           5.18          0.35         1.80        * 13.1 %

       5200               36.0                4.66              4.99          4.99           4.99          0.35         1.80        + 13.1 %

       5300               35.9                4.76              4.91          4.91           4.91          0.30         1.80        +13.1 %

       5500               35.6                4.96              4.85          4.85           4.85          0.40         1.80        +131 %

       5600               35.5                 5.07              4.67         4.67           4.67          0.40         1.80        £13.1 %

       5800               35.3                 5.27              4.34          4.34          4.34          0.55         1.80         £13.1 %


© Frequency validity of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The uncertainty is the RSS
of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band.
" At frequencies below 3 GHz, the validity of tissue parameters (s and a) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.




Certificate No: EX3—3885_Oct12                                     Page 5 of 11


EX3DV4— SN:3885                                                                                                               October 9, 2012




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3885


Calibration Parameter Determined in Body Tissue Simulating Media
                       Relative          Conductivity                                                                Depth           Unct.
     f (MHz)°        Permittivity®           (S/im)"        ConvFX         ConvF Y¥      ConvFZ         Alipha       {mm)            (k=2)
        450             ~s67 <—|—~~0:94                     <40.82 —| o:e2 |—o:sz —| o:os /                          110           £134 %~

        750              55.5                 0.96              9.50         9.50           9.50         0.43         0.80         + 12.0 %

        835              55.2                 0.97              9.36         9.36           9.36         0.80         0.59         + 12.0 %

        900              55.0                 1.05              9.19         9.19           9.19         0.55         0.71         + 12.0 %

       1450               54.0                1.30              8.67         8.67           8.67         0.31         12111        +12.0 %

       1640               53.8                1.40              8.78         8.78           8.78         0.48         0.84         +12.0 %

       1750               53.4                1.49              8.03         8.03           8.03         0.29         1.04         £12.0 %

       1900               53.3                1.52              7.67         7.67           7.67         0.38         0.93         + 12.0 %

       1950               53.3                1.52              7.73         7.73           7.73         0.46         0.78         + 12.0 %

       2000               53.3                1.52              7.60         7.60           7.60         0.22         1113         +# 12.0 %

       2300               52.9                1.81              715          7.15           7.15          0.41        0.78         + 12.0 %

       2450               52.7                1.95              6.80          6.80          6.80          0.80        0.50         + 12.0 %

       2600               52.5                216               6.67          6.67          6.67          0.80        0.55         # 12.0 %

       3500               51.3                3.31              6.49          6.49          6.49          1.00        0.44         *# 13.1 %

       3700               51.0                3.55              6.63          6.63          6.63          1.00        0.50         £13.1 %

       4950               49.4                5.01              4.62)         4.62          4.62          0.40         1.90        £13.1 %

       5200               49.0                5.30              4.50          4.50          4.50          0.40         1.90        *# 13.1 %

       5300               48.9                5.42              4.28          4.28          4.28          0.50         1.90        £13.1 %

       5500               48.6                5.65              4.10          4.10          4.10          0.40         1.90        £13.1 %

       5600               48.5                5.77              4.26          4.26           4.26         0.30         1.90        £13.1 %

       5800               48.2                6.00              4.14          4.14           4.14         0.50         1.90         * 13.1 %



© Frequency validity of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The uncertainty is the RSS
of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band.
" At frequencies below 3 GHz, the validity of tissue parameters (s and a) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.




Certificate No: EX3—3885_Oct12                                     Page 6 of 11


EX3DV4— SN:3885                                                                                                                                                                             October 9, 2012


                                                                   Frequency Response of E—Field
                                                                            (TEM—Cell:ifi110 EXX, Waveguide: R22)




                                           14— ....................................................................................................................................................

                                           13—
         Frequency response {normalized)




                                           1.2—
                                                  %                                             f
                                           1117 mmprtr d
                                                  C s—___                      i
                                                  |—      mETooghoocclllllnlnlhlllll,
                                           1.0—

                                           0.9—7 ..................................................................................................................................................

                                           0.8—

                                           0.7—

                                           0.6—

                                           0.5                                                                         j                       i
                                                                                                                    1500                   2000
                                                                                                                    f [MHz]
                                                                                   (a)
                                                                                   TEM


                                                                 Uncertainty of Frequency Response of E—field: £ 6.3% (k=2)




Certificate No: EX3—3885_Oct12                                                                                Page 7 of 11


EX3DV4— SN:3885                                                                                                                             October 9, 2012




                                         Receiving Pattern (¢), 8 = 0°

                         f=600 MHz,TEM                                                                £=1800 MHz.R22
                                       90
                                     snmss.
                               l.—        =




            "Tot                x              ~y              7                   "Tot                 x~               ~y                       7




      x                                                                     |
      3.                                                                    i
                                                                           »:

                                                                            :
       kn

       0                                                   i
      ud                                                   |                f                     f                f                         f
       s                                                   1                1                     1                1



            hoefmmmmmmmmmmmmnoeesferremmmnisisieeformminiinineessoe
               [ 1__:     {—_L_L_L_1       l    ob   c     1   Lo           1   Lo        d       |    Lob   ol    i    L     o_   L    4         L
                        —150             —100            —50               0                  50                  100                       150
                               ca)                                      Roll [*]            e
                                £                         [#]                                 &                                    ce
                          100 MHz                        600 MHz                          1800 MHz                              2500 MHz


                                       Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificate No: EX3—3885_Oct12                                     Page 8 of 11


EX3DV4— SN:3885                                                                                                      October 9, 2012




                                                    Dynamic Range f(SARpneaq)
                                                               (TEM cell , f= 900 MHz)

                  Input Signal [uV]




                                      102




                                            103        102          107        10          10           10     10°
                                                                          SAR [mWicm3]
                                                              {¢]
                                                      not compensated                       compensated




                                      a
                                      C,
                                      8
                                      ui


                                             103        102         101          100      101      102       103
                                                                          SAR [mW/icm3]
                                                              *
                                                       not cofnpensated                   compensated


                                                   Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certificate No: EX3—3885_Oct12                                            Page 9 of 11


EX3DV4— SN:3885                                                                                                                                                                          October 9, 2012


                                                                   Conversion Factor Assessment
                                  f= 835 MHz,WGLS RQ (H_convF)                                                                       f= 1900 MHz,WGLS R22 (H_convF)
            4.07.
                   F                                                                                                    ao—||         :;
            £.57

            C'i.O"?le"'t;E§                                                                                             257 : :


            25—
                   Ps                   ss
                                                                                                                              Je
                                                                                                                             Lo      y
       5%          >              "*\                                                                               % * :            %“L
       % 204                           k                                                                            2|                 E'"\\
                   f—                       ~                                                                       & is                     \o
       §           F                        \,,\                                                                    3         L                x
            15                    he    ~~‘m%-\                                                                               T                   &

                                                               ~                                                        10—
            1.0                                                    >\‘\"v\\
                                                                        —                                                        .
                                                                                                                         s                                          w
            0.5                                                                _                                                                                        hk Tsl
                                                                                       ~_m“““*m«..._»%                                                                            es_ol
            0.0               1                                —                   —       1       .    :                c~!|          “",!":!1                         /x]-"l       ‘S(tlfi:w“"

                  o                    10                  20           30         40              5o        eo              0        5            10        15      20      25     30       35   40
                                                L2                    2 {mm}              Wl                                                                       z {mm}

                                            analylical                                  measured                                               analytical                         measured



                                                                   Deviation from Isotropy in Liquid
                                                                                            Error (6, 8), f = 900 MHz
                                                   Deviation




                                                           —1.0       —0.8     —0.6            —0.4     —0.2      0.0        0.2           0.4          06        0.3       1.0
                                                                   Uncertainty of Spherical Isotropy Assessment: £ 2.6% (k=2)




Certificate No: EX3—3885_Oct12                                                                              Page 10 of 11


EX3DV4— SN:3885                                              October 9, 2012




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3885

Other Probe Parameters
 Sensor Arrangement                                             Triangular
 Connector Angle (°)                                                 115.8
 Mechanical Surface Detection Mode                                enabled
 Optical Surface Detection Mode                                   disabled
 Probe Overall Length                                             337 mm
 Probe Body Diameter                                               10 mm
 Tip Length                                                          9 mm
 Tip Diameter                                                      2.5 mm
 Probe Tip to Sensor X Calibration Point                             1 mm
 Probe Tip to Sensor Y Calibration Point                             1 mm
 Probe Tip to Sensor Z Calibration Point                             1 mm
 Recommended Measurement Distance from Surface                       2 mm




Certificate No: EX3—3885_Oct12               Page 11 of 11