FCC ID: A3LSMA205W

RF Exposure Info

Download: PDF
FCCID_4274485

Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                         Date: 2019-04-25

20190425_SystemPerformanceCheck-D2450V2 SN 939

Frequency: 2450 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 2450 MHz; σ = 2.022 S/m; εr = 52.995; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1447; Calibrated: 2019-03-21
- Probe: EX3DV4 - SN7376; ConvF(7.5, 7.5, 7.5); Calibrated: 2018-09-26;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI V6.0 (20deg); Type: QD OVA 003 AA; Serial: 2013

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.79 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm
Reference Value = 62.06 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 10.7 W/kg
SAR(1 g) = 5.32 W/kg; SAR(10 g) = 2.51 W/kg
Maximum value of SAR (measured) = 7.56 W/kg




                      0 dB = 7.56 W/kg = 8.79 dBW/kg




                                                                                                            Plot No. 1


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-25

20190425_SystemPerformanceCheck-D2450V2 SN 939

Frequency: 2450 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 7.95 W/kg




                                                                                  Plot No. 2


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                         Date: 2019-05-02

20190502_SystemPerformanceCheck-D2600V2 SN 1097

Frequency: 2600 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 2600 MHz; σ = 2.241 S/m; εr = 51.212; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1447; Calibrated: 2019-03-21
- Probe: EX3DV4 - SN7376; ConvF(7.49, 7.49, 7.49); Calibrated: 2018-09-26;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: ELI V6.0 (20deg); Type: QD OVA 003 AA; Serial: 2013

Body/Pin=100 mW/Area Scan (8x8x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 5.66 W/kg

Body/Pin=100 mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 61.02 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 12.1 W/kg
SAR(1 g) = 5.68 W/kg; SAR(10 g) = 2.55 W/kg
Maximum value of SAR (measured) = 8.20 W/kg




                      0 dB = 8.20 W/kg = 9.14 dBW/kg




                                                                                                            Plot No. 3


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-05-02

20190502_SystemPerformanceCheck-D2600V2 SN 1097

Frequency: 2600 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 8.24 W/kg




                                                                                  Plot No. 4


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-05-06

20190506_SystemPerformanceCheck-D5GHzV2 SN 1184

Frequency: 5600 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 5600 MHz; σ = 4.902 S/m; εr = 35.815; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1447; Calibrated: 2019-03-21
- Probe: EX3DV4 - SN7376; ConvF(4.6, 4.6, 4.6); Calibrated: 2018-09-26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: SAM (20deg probe tilt) with CRP v5.0_Back; Type: QD000P40CD; Serial: TP:1882

Head/5.6 GHz, Pin=100mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 21.0 W/kg

Head/5.6 GHz, Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 72.67 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 36.9 W/kg
SAR(1 g) = 7.84 W/kg; SAR(10 g) = 2.2 W/kg
Maximum value of SAR (measured) = 19.1 W/kg




                      0 dB = 19.1 W/kg = 12.81 dBW/kg




                                                                                                             Plot No. 5


Test Laboratory: UL Korea, Ltd. Suwon Laboratory              Date: 2019-05-06

20190506_SystemPerformanceCheck-D5GHzV2 SN 1184

Frequency: 5600 MHz; Duty Cycle: 1:1

Head/5.6 GHz, Pin=100mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 13.9 W/kg




                                                                                      Plot No. 6


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-04-18

20190418_SystemPerformanceCheck-D835V2 SN 4d174

Frequency: 835 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 835 MHz; σ = 0.988 S/m; εr = 54.087; ρ = 1000 kg/m 3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1494; Calibrated: 2018-07-23
- Probe: EX3DV4 - SN7330; ConvF(10.45, 10.45, 10.45); Calibrated: 2019-01-31;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1166

Body/Pin=100 mW/Area Scan (7x13x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.17 W/kg

Body/Pin=100 mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 34.71 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 1.43 W/kg
SAR(1 g) = 0.973 W/kg; SAR(10 g) = 0.641 W/kg
Maximum value of SAR (measured) = 1.18 W/kg




                      0 dB = 1.18 W/kg = 0.72 dBW/kg




                                                                                                             Plot No. 7


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-18

20190418_SystemPerformanceCheck-D835V2 SN 4d174

Frequency: 835 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.15 W/kg




                                                                                  Plot No. 8


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-04-18

20190418_SystemPerformanceCheck-D750V3 SN 1122

Frequency: 750 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 750 MHz; σ = 0.962 S/m; εr = 56.782; ρ = 1000 kg/m 3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1468; Calibrated: 2018-08-22
- Probe: EX3DV4 - SN7314; ConvF(10.05, 10.05, 10.05); Calibrated: 2018-08-30;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1166

Body/Pin=100 mW/Area Scan (7x14x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.01 W/kg

Body/Pin=100 mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 32.72 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 1.21 W/kg
SAR(1 g) = 0.836 W/kg; SAR(10 g) = 0.560 W/kg




                      0 dB = 1.01 W/kg = 0.04 dBW/kg




                                                                                                             Plot No. 9


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-18

20190418_SystemPerformanceCheck-D750V3 SN 1122

Frequency: 750 MHz; Duty Cycle: 1:1

Body/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.00 W/kg




                                                                                  Plot No. 10


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-04-19

20190419_SystemPerformanceCheck-D835V2 SN 4d174

Frequency: 835 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 835 MHz; σ = 0.916 S/m; εr = 43.047; ρ = 1000 kg/m 3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1468; Calibrated: 2018-08-22
- Probe: EX3DV4 - SN7314; ConvF(9.47, 9.47, 9.47); Calibrated: 2018-08-30;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: SAM Phantom CRP v5.0(Right); Type: QD000P40CD; Serial: TP:1855

Head/Pin=100 mW/Area Scan (7x13x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 1.21 W/kg

Head/Pin=100 mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 36.80 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 1.48 W/kg
SAR(1 g) = 0.996 W/kg; SAR(10 g) = 0.654 W/kg




                      0 dB = 1.21 W/kg = 0.83 dBW/kg




                                                                                                         Plot No. 11


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-19

20190419_SystemPerformanceCheck-D835V2 SN 4d174

Frequency: 835 MHz; Duty Cycle: 1:1

Head/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.26 W/kg




                                                                                  Plot No. 12


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-04-18

20190418_SystemPerformanceCheck-D1750V2 SN 1125

Frequency: 1750 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 1750 MHz; σ = 1.369 S/m; εr = 38.512; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1259; Calibrated: 2018-07-26
- Probe: EX3DV4 - SN3991; ConvF(8.81, 8.81, 8.81); Calibrated: 2018-05-24;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Twin-SAM V5.0 (30deg probe tilt)_20190404; Type: QD 000 P40 CD; Serial: 1829

Head/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 4.62 W/kg

Head/Pin=100 mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.60 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 6.39 W/kg
SAR(1 g) = 3.51 W/kg; SAR(10 g) = 1.85 W/kg
Maximum value of SAR (measured) = 4.68 W/kg




                      0 dB = 4.68 W/kg = 6.70 dBW/kg




                                                                                                         Plot No. 13


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-18

20190418_SystemPerformanceCheck-D1750V2 SN 1125

Frequency: 1750 MHz; Duty Cycle: 1:1

Head/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 4.60 W/kg




                                                                                  Plot No. 14


Test Laboratory: UL Korea, Ltd. Suwon Laboratory                                          Date: 2019-04-18

20190418_SystemPerformanceCheck-D1900V2 SN 5d190

Frequency: 1900 MHz; Duty Cycle: 1:1;  Room Ambient Temperature: 23.0C; Liquid Temperature: 22.0C
Medium parameters used: f = 1900 MHz; σ = 1.463 S/m; εr = 38.301; ρ = 1000 kg/m3
DASY5 Configuration:
- Area Scan Setting: Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Averaged Fast SAR: Polynomial fit
- Electronics: DAE4 Sn1259; Calibrated: 2018-07-26
- Probe: EX3DV4 - SN3991; ConvF(8.43, 8.43, 8.43); Calibrated: 2018-05-24;
- Sensor-Surface: 2.5mm (Mechanical Surface Detection)
- Phantom: Twin-SAM V5.0 (30deg probe tilt)_20190404; Type: QD 000 P40 CD; Serial: 1829

Head/Pin=100 mW/Area Scan (7x7x1): Measurement grid: dx=15mm, dy=15mm
Maximum value of SAR (measured) = 5.51 W/kg

Head/Pin=100 mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 62.12 V/m; Power Drift = -0.00 dB
Peak SAR (extrapolated) = 7.75 W/kg
SAR(1 g) = 4.13 W/kg; SAR(10 g) = 2.13 W/kg
Maximum value of SAR (measured) = 5.65 W/kg




                      0 dB = 5.65 W/kg = 7.52 dBW/kg




                                                                                                         Plot No. 15


Test Laboratory: UL Korea, Ltd. Suwon Laboratory               Date: 2019-04-18

20190418_SystemPerformanceCheck-D1900V2 SN 5d190

Frequency: 1900 MHz; Duty Cycle: 1:1

Head/Pin=100 mW/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 5.50 W/kg




                                                                                  Plot No. 16



Document Created: 2019-05-14 22:02:19
Document Modified: 2019-05-14 22:02:19