Test Report

FCC ID: 2ATHKES005

Test Report

Download: PDF
FCCID_4362439

                                           FCC PART 15.231

                                              TEST REPORT
                                                             For

                                 ETROG SYSTEMS LTD.
                                        124 Hertzel street, Jerusalem, Israel


                                         FCC ID: 2ATHKES005


            Report Type:                                      Product Type:
            Original Report                                   Medical Alert Sy stem



                   Report Number:           RSZ190606002-00B


                        Report Date:        2019-07-16

                                            Simon Wang
                       Reviewed By:         RF Engineer

                        Prepared By:        Bay Area Compliance Laboratories Corp. (Shenzhen)
                                            6/F., West Wing, Third Phase of Wanli Industrial
                                            Building, Shihua Road, Futian Free Trade Zone,
                                            Shenzhen, Guangdong, China
                                            Tel: +86-755-33320018
                                            Fax: +86-755-33320008
                                            www.baclcorp.com.cn




Note: This report must not be used by the customer to claim product certification, approval, or endorsement by A2LA* or any
agency of the Federal Government. * This report may contain data that are not covered by the A2LA accreditation and are marked
with an asterisk “*”.
      The information marked # is provided by the applicant, the laboratory is not responsible for its authenticity.


Bay Area Compliance Laboratories Corp. (Shenzhen)                                                                          Report No.: RSZ190606002-00B




                                                         TABLE OF CONTENTS


GENERAL INFORMATION .......................................................................................................................................3
 PRODUCT DESCRIPTION FOR EQUIPMENT UNDER TEST (EUT) .....................................................................................3
 OBJECTIVE ...................................................................................................................................................................3
 RELATED SUBMITTAL(S)/GRANT(S) .............................................................................................................................3
 TEST METHODOLOGY ..................................................................................................................................................3
 MEASUREMENT UNCERTAINTY ....................................................................................................................................4
SYSTEM TEST CONFIGURATION ..........................................................................................................................5
  JUSTIFICATION .............................................................................................................................................................5
  SPECIAL ACCESSORIES .................................................................................................................................................5
  EQUIPMENT MODIFICATIONS .......................................................................................................................................5
  SUPPORT EQUIPMENT LIST AND DETAILS ....................................................................................................................5
  EXTERNAL I/O CABLE..................................................................................................................................................5
  BLOCK DIAGRAM OF TEST SETUP ................................................................................................................................6
SUMMARY OF TEST RESULTS ...............................................................................................................................7
TEST EQUIPMENT LIST AND DETAILS ...............................................................................................................8
FCC §15.203 - ANTENNA REQUIREMENT.............................................................................................................9
  APPLICABLE STANDARD ..............................................................................................................................................9
  ANTENNA CONNECTOR CONSTRUCTION ......................................................................................................................9
FCC §15.207 (A) – AC LINE CONDUCTED EMISSIONS ....................................................................................10
  APPLICABLE STANDARD ............................................................................................................................................10
  EUT SETUP ................................................................................................................................................................10
  EMI TEST RECEIVER SETUP .......................................................................................................................................10
  TEST PROCEDURE ......................................................................................................................................................10
  CORRECTED FACTOR & MARGIN CALCULATION .......................................................................................................11
  TEST RESULTS SUMMARY ..........................................................................................................................................11
  TEST DATA ................................................................................................................................................................11
FCC §15.205, §15.209, §15.231 (B) - RADIATED EMISSIONS .............................................................................14
  APPLICABLE STANDARD ............................................................................................................................................14
  EUT SETUP ................................................................................................................................................................14
  EMI TEST RECEIVER SETUP .......................................................................................................................................15
  TEST PROCEDURE ......................................................................................................................................................15
  CORRECTED AMPLITUDE & MARGIN CALCULATION .................................................................................................16
  TEST RESULTS SUMMARY ..........................................................................................................................................16
  TEST DATA ................................................................................................................................................................16
FCC §15.231(A) (1) - DEACTIVATION TESTING.................................................................................................20
  APPLICABLE STANDARD ............................................................................................................................................20
  TEST PROCEDURE ......................................................................................................................................................20
  TEST DATA ................................................................................................................................................................20
FCC §15.231(C) – 20 DB EMISSION BANDWIDTH TESTING ...........................................................................22
  APPLICABLE STANDARD ............................................................................................................................................22
  TEST PROCEDURE ......................................................................................................................................................22
  TEST DATA ................................................................................................................................................................22


FCC Part 15.231                                                                                                                                         Page 2 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                 Report No.: RSZ190606002-00B

GENERAL INFORMATION
Product Description for Equipment under Test (EUT)

             Product                  Medical Alert System
           Tested Model               ES005
         Multiple Model#              ES005-BP, ES005-BC, ES005-BW, ES005-BM
         Frequency Range              434.21MHz
          Field Strength              80 dBuV/m@3m
       Modulation Technique           ASK
       Antenna Specification          2 dBi
          Voltage Range               DC 3.7V from battery or DC 5V from adapter
            Date of Test              2019-06-18 ~ 2019-07-16
       Sample serial number           190606002
          Received date               2019-06-06
        Sample/EUT Status             Good condition
                                      Model: JF005WR-0500100UH
         Adapter information          Input: AC 100-240V, 50/60Hz, 0.18A
                                      Output: DC 5V, 1A

Manufacturer information
Company Name: Shenzhen Joy Technology Co. Ltd
Company Address: 5/F,3rd Building, Shunchengji Park, Huayun Road of Dalang, Longhua, Shenzhen, China

Notes: This series products model: ES005-BP, ES005-BC, ES005-BW, ES005-BM and ES005 are identical
schematics, they have different model name. Model ES005 was selected for fully testing, the detailed information can
be referred to the declaration.


Objective
This test report is prepared on behalf of ETROG SYSTEMS LTD.. All the test measurements were
performed according to the measurement procedure described in ANSI C63.10 - 2013.

The tests were performed in order to determine compliance with FCC Part 15, Subpart C, section 15.203,
15.205,15.207, 15.209, 15.35(c) and 15.231 rules.


Related Submittal(s)/Grant(s)
Not Related Submittal(s)/Grant(s).


Test Methodology
All measurements contained in this report were conducted with ANSI C63.10 - 2013, American National
Standard of Procedures for Compliance Testing of Unlicensed Wireless Devices.

All emissions measurement was performed at Bay Area Compliance Laboratories Corp. (Shenzhen). The
radiated testing was performed at an antenna-to-EUT distance of 3 meters.




FCC Part 15.231                                                                                       Page 3 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                 Report No.: RSZ190606002-00B


Measurement Uncertainty

                       Parameter                                    Uncertainty
             Occupied Channel Bandwidth                                 ±5%
          RF Output Power with Power meter                            ±0.5dB
           RF conducted test with spectrum                            ±1.5dB
        AC Power Lines Conducted Emissions                            ±1.95dB
           Radiated         Below 1GHz                                ±4.75dB
          Emissions         Above 1GHz                                ±4.88dB
                   Temperature                                         ±3℃
                       Humidity                                         ±6%
                    Supply voltages                                    ±0.4%

Note: Otherwise required by the applicant or Product Regulations, Decision Rule in this report did not consider the
uncertainty.


Test Facility
The test site used by Bay Area Compliance Laboratories Corp. (Shenzhen) to collect test data is located on
the 6/F., West Wing, Third Phase of Wanli Industrial Building, Shihua Road, Futian Free Trade Zone,
Shenzhen, Guangdong, China.
The test site has been approved by the FCC under the KDB 974614 D01 and is listed in the FCC Public
Access Link (PAL) database, FCC Registration No.: 342867, the FCC Designation No.: CN1221.
The test site has been registered with ISED Canada under ISED Canada Registration Number 3062B.




FCC Part 15.231                                                                                       Page 4 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                             Report No.: RSZ190606002-00B

SYSTEM TEST CONFIGURATION
Justification
The system was configured for testing by manufacturer.
Operating frequency:434.21 MHz.


Special Accessories
No special accessories was used


Equipment Modifications
No modification was made to the EUT.


Support Equipment List and Details


        Manufacturer                  Description                Model                Serial Number

           Yeastar                       PBX                 MY PBX_SOHO                 1103020
           Kinhao                     Telephone                  N/A                     14589612
           Herene                     Telephone                  HD-70                     N/A


External I/O Cable

                  Cable Description                 Length (m)         From / Port            To

         Un-shielding detachable RJ11 cable            1.5                 EUT                PBX
         Un-shielding detachable RJ11 cable            2.0               Telephone            PBX
         Un-shielding detachable RJ11 cable            2.0                 EUT             Telephone
        Un-shielding Un-detachable AC cable            1.2                 PBX              LISN 2
        Un-shielding Un-detachable DC cable            1.0                 EUT              Adapter




FCC Part 15.231                                                                               Page 5 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                              Report No.: RSZ190606002-00B

Block Diagram of Test Setup



       LISN 2                                                                                                 LISN 1




                    Telephone          PBX            Telephone          EUT           Adapter
                                                                                        Adapter
                                10cm         10cm                 10cm          10cm
                                                                                       Socket




                                                                                                  1.0 meter
                  Non-Conductive Table
                  80/150 cm above Ground Plane

                                                    1.5 meters




FCC Part 15.231                                                                                   Page 6 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                            Report No.: RSZ190606002-00B


SUMMARY OF TEST RESULTS

                      FCC Rules                        Description of Test              Result

                        §15.203                       Antenna Requirement             Compliance

                      §15.207 (a)                     Conducted Emissions             Compliance

              §15.205, §15.209, §15.231(b)             Radiated Emissions             Compliance

                      §15.231 (c)                   20dB Emission Bandwidth           Compliance

                     §15.231 (a) (1)                      Deactivation                Compliance




FCC Part 15.231                                                                              Page 7 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                   Report No.: RSZ190606002-00B


TEST EQUIPMENT LIST AND DETAILS

                                                                         Serial        Calibration   Calibration
      Manufacturer              Description             Model
                                                                        Number            Date        Due Date
                                       AC Line Conducted Emission Test
    Rohde & Schwarz          EMI Test Receiver         ESCS30            100176         2018-07-11   2019-07-11
                                                                    3560.6650.12-
    Rohde & Schwarz                LISN                ENV216                           2019-01-25   2020-01-25
                                                                     101613-Yb
    Rohde & Schwarz          Transient Limiter         ESH3Z2           DE25985         2019-03-02   2020-03-01
                            Conducted Emission                      UF A210B-1-
        Unknown                                         78652                           2018-11-12   2019-11-12
                                  Cable                             0720-504504
    Rohde & Schwarz          CE Test software          EMC 32           V 09.10.0         NCR           NCR
                                               Radiated Emission Test
       A.H. System             Horn Antenna          SAS-200/571           135          2018-09-01   2021-08-31
    Rohde & Schwarz           Signal Analyzer           FSEM         845987/005         2018-06-23   2019-06-23
    Rohde & Schwarz           Signal Analyzer           FSEM         845987/005         2019-06-23   2020-06-23
      COM-POWER                Pre-amplifier           PA-122            181919         2018-11-12   2019-11-12
   Sonoma Instrument             Amplifier              310N             186238         2018-11-12   2019-11-12
      Sunol Sciences        Broadband Antenna            JB1            A040904-1       2017-12-22   2020-12-21
                                                                   1316.3003K03
    Rohde & Schwarz          EMI Test Receiver           ESR                            2018-07-11   2019-07-11
                                                                    -101746-zn
                                                                   1316.3003K03
    Rohde & Schwarz          EMI Test Receiver           ESR                            2019-07-11   2020-07-11
                                                                    -101746-zn

    Rohde & Schwarz          Spectrum Analyzer          FSU26            200120         2019-03-02   2020-03-01
   UTiFLEX MICRO-                                     UFA147A-       MFR64639
                                 RF Cable                                               2018-11-12   2019-11-12
        C0AX                                         2362-100100     231029-003
      Ducommun
                                 RF Cable              104PEA           218124002       2018-11-12   2019-11-12
     Technologies
      Ducommun
                                 RF Cable              RG-214              1            2018-11-19   2019-05-21
     Technologies
      Ducommun
                                 RF Cable              RG-214              1            2019-05-21   2019-11-19
     Technologies
      Ducommun
                                 RF Cable              RG-214              2            2018-11-12   2019-11-12
     Technologies

* Statement of Traceability: Bay Area Compliance Laboratories Corp. (Shenzhen) attests that all calibrations have
 been performed in accordance to requirements that traceable to National Primary Standards and International
 System of Units (SI).




FCC Part 15.231                                                                                      Page 8 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                           Report No.: RSZ190606002-00B

FCC §15.203 - ANTENNA REQUIREMENT
Applicable Standard
An intentional radiator shall be designed to ensure that no antenna other than that furnished by the
responsible party shall be used with the device. The use of a permanently attached antenna or of an
antenna that uses a unique coupling to the intentional radiator shall be considered sufficient to comply with
the provisions of this section.


Antenna Connector Construction
The EUT has one internal antenna arrangement, which used a unique coupling to this product. And the
antenna is 2dBi; fulfill the requirement of this section. Please refer to EUT photos.
Result: Compliant.




FCC Part 15.231                                                                                Page 9 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                           Report No.: RSZ190606002-00B

FCC §15.207 (a) – AC LINE CONDUCTED EMISSIONS
Applicable Standard
FCC§15.207


EUT Setup




The setup of EUT is according with per ANSI C63.4-2014 measurement procedure. The specification used
was with the FCC Part 15.207 limits.

The adapter was connected to a 120 VAC/60 Hz power source.


EMI Test Receiver Setup
The EMI test receiver was set to investigate the spectrum from 150 kHz to 30 MHz.

During the conducted emission test, the EMI test receiver was set with the following configurations:

                              Frequency Range                      IF B/W
                              150 kHz – 30 MHz                      9 kHz



Test Procedure
During the conducted emission test, the adapter was connected to the outlet of the LISN.

Maximizing procedure was performed on the six (6) highest emissions of the EUT.

All final data was recorded in the Quasi-peak and average detection mode.

FCC Part 15.231                                                                              Page 10 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                               Report No.: RSZ190606002-00B

Corrected Factor & Margin Calculation
The Corrected factor is calculated by adding LISN VDF (Voltage Division Factor), Cable Loss and
Transient Limiter Attenuation. The basic equation is as follows:

      Correction Factor = LISN VDF + Cable Loss + Transient Limiter Attenuation

The “Margin” column of the following data tables indicates the degree of compliance with the applicable
limit. For example, a margin of 7 dB means the emission is 7 dB below the limit. The equation for margin
calculation is as follows:

      Margin = Limit – Corrected Amplitude

Test Results Summary
According to the recorded data in following table, the EUT complied with the FCC Part 15.207.
Refer to CISPR16-4-2:2011 and CISPR 16-4-1:2009, the measured level complies with the limit if

                                            Lm + U(Lm) ≤ Llim + Ucispr

In BACL, U(Lm) is less than Ucispr, if Lm is less than Llim, it implies that the EUT complies with the limit.


Test Data
     Environmental Conditions

             Temperature:                25 ℃
        Relative Humidity:               50 %
           ATM Pressure:               101.0 kPa

     The testing was performed by Haiguo Li on 2019-06-26.

     EUT operation mode: Transmitting &Talking




FCC Part 15.231                                                                                    Page 11 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                Report No.: RSZ190606002-00B


     AC 120V/60 Hz, Line



                         90


                         80


                         70

                                                                                                   QP
                         60
          Level in dBμ




                                                                                                     AV
                         50

                         40


                         30

                         20


                         10

                         0
                         150k   300 400 500   800 1M        2M     3M 4M 5M 6     8 10M      20M     30M
                                                        Frequency in Hz


                                Corrected     Correction
         Frequency                                                Limit     Margin          Detector
                                Amplitude      Factor
           (MHz)                                                 (dBμV)      (dB)         (PK/Ave./QP)
                                 (dBμV)         (dB)
           0.213500               40.4           19.8             63.1          22.7          QP
           0.305350               39.8           19.7             60.1          20.3          QP
           0.340870               38.4           19.9             59.2          20.8          QP
           0.348690               36.8           19.9             59.0          22.2          QP
           0.996730               33.2           19.9             56.0          22.8          QP
           1.010730               33.3           19.9             56.0          22.7          QP
           0.213500               34.9           19.8             53.1          18.2          Ave.
           0.305350               33.3           19.7             50.1          16.8          Ave.
           0.340870               32.6           19.9             49.2          16.6          Ave.
           0.348690               30.4           19.9             49.0          18.6          Ave.
           0.996730               26.1           19.9             46.0          19.9          Ave.
           1.010730               25.3           19.9             46.0          20.7          Ave.




FCC Part 15.231                                                                                    Page 12 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                Report No.: RSZ190606002-00B


     AC 120V/60 Hz, Neutral



                         90


                         80


                         70

                                                                                                   QP
                         60
          Level in dBμ




                                                                                                     AV
                         50

                         40


                         30

                         20


                         10

                         0
                         150k   300 400 500   800 1M        2M     3M 4M 5M 6     8 10M      20M     30M
                                                        Frequency in Hz



                                Corrected     Correction
         Frequency                                                Limit     Margin          Detector
                                Amplitude      Factor
           (MHz)                                                 (dBμV)      (dB)         (PK/Ave./QP)
                                 (dBμV)         (dB)
           0.217500               42.0           19.8             62.9          20.9          QP
           0.309290               38.8           19.7             60.0          21.2          QP
           0.340870               38.8           19.8             59.2          20.4          QP
           0.372330               34.8           19.8             58.4          23.6          QP
           0.400030               35.2           19.8             57.9          22.7          QP
           0.959390               32.6           19.8             56.0          23.4          QP
           0.217500               37.3           19.8             52.9          15.6          Ave.
           0.309290               32.9           19.7             50.0          17.1          Ave.
           0.340870               32.8           19.8             49.2          16.4          Ave.
           0.372330               28.2           19.8             48.4          20.2          Ave.
           0.400030               27.9           19.8             47.9          20.0          Ave.
           0.959390               25.6           19.8             46.0          20.4          Ave.

     Note:
     1) Correction Factor =LISN VDF (Voltage Division Factor) + Cable Loss + Transient Limiter Attenuation
     2) Corrected Amplitude = Reading + Correction Factor
     3) Margin = Limit – Corrected Amplitude
FCC Part 15.231                                                                                    Page 13 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                             Report No.: RSZ190606002-00B

FCC §15.205, §15.209, §15.231 (b) - RADIATED EMISSIONS
Applicable Standard
FCC §15.205, §15.209, §15.231 (b)
According to FCC §15.231(b), the field strength of emissions from intentional radiators operated under
this section shall not exceed the following:

           Fundamental frequency             Field Strength of             Field Strength of spurious
                                          Fundamental (Microvolts            emissions ((Microvolts
                 (MHz)                            /meter)                            /meter)
                  40.66-40.70                        2250                              225
                     70-130                          1250                              125
                    130-174                     1250 to 3750**                    125 to 375**
                    174-260                          3750                              375
                    260-470                    3750 to 12500**                   375 to 1250**
                  Above 470                         12500                             1250

       *Linear interpolations.

The above field strength limits are specified at a distance of 3-meters the tighter limits apply at the band
edges.

EUT Setup
Below 1 GHz:




FCC Part 15.231                                                                                  Page 14 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                              Report No.: RSZ190606002-00B

Above 1 GHz:




The radiated emission tests were performed in the 3 meters test site, using the setup accordance with the
ANSI C63.10 - 2013. The specification used was the FCC 15 § 15.209, 15.205 and 15.231.


EMI Test Receiver Setup
The system was investigated from 30 MHz to 5 GHz.

During the radiated emission test, the test receiver was set with the following configurations:

               Frequency Range            RBW          Video B/W        IF B/W       Measurement

              30MHz – 1000 MHz           100 kHz        300 kHz        120 kHz            QP
                  Above 1 GHz             1 MHz          3 MHz             /              PK

Test Procedure

Maximizing procedure was performed on the highest emissions to ensure that the EUT complied with all
installation combinations.
All final data was recorded in the Quasi-peak detection mode from 30MHz to 1GHz, Peak and average
detection mode above 1 GHz.




FCC Part 15.231                                                                                   Page 15 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                              Report No.: RSZ190606002-00B

Corrected Amplitude & Margin Calculation
The Corrected Amplitude is calculated by adding the Antenna Loss and Cable Loss, and subtracting the
Amplifier Gain from the Meter Reading. The basic equation is as follows:
        Corrected Amplitude = Meter Reading + Antenna Loss + Cable Loss - Amplifier Gain
The “Margin” column of the following data tables indicates the degree of compliance with the applicable
limit. For example, a margin of 5.8 dB means the emission is 5.8 dB below the limit. The equation for
margin calculation is as follows:
        Margin = Limit –Corrected Amplitude


Test Results Summary
According to the data in the following table, the EUT complied with the FCC §15.205, §15.209, §15.231
(b)
Refer to CISPR16-4-2:2011 and CISPR 16-4-1:2009, the measured level complies with the limit if

                                          Lm + U(Lm ) ≤ Llim + Ucispr

In BACL, U(Lm ) is less than Ucispr, if Lm is less than Llim, it implies that the EUT complies with the limit.


Test Data
 Environmental Conditions

            Temperature:              23~25 ℃
       Relative Humidity:             49~50 %
          ATM Pressure:            100.0~101.0 kPa

     The testing was performed by Alan He from 2019-06-18 to 2019-07-16.

     Test mode: Transmitting&talking




FCC Part 15.231                                                                                   Page 16 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                               Report No.: RSZ190606002-00B




                            80
                                                                     Fundamental

                            70



                            60



                            50                                      Electric Field Strength QP-3m
          Level in dB¦ÌV/




                            40



                            30



                            20



                            10



                             0
                             30M         50   60   80   100M               200        300      400    500    800     1G
                                                                 Frequency in Hz




                            Corrected                   Antenna          Turntable Correction
Frequency                                  Detector             Antenna                         Limit                     Margin
                            Amplitude                    Height           Position  Factor
  (MHz)                                  (PK/QP/Ave.)           Polarity                      (dBμV/m)                     (dB)
                            (dBμV/m)                      (cm)            (degree)  (dB/m)
30.975375                        17.70        QP         389.0         H           179.0             -8.2    22.30        40.00
153.599750                       13.13        QP         365.0         H           119.0             -14.3   13.13        43.50
497.874750                       12.92        QP         302.0         V           284.0             -7.2    12.92        46.00
626.145000                       18.23        QP         275.0         V            0.0              -2.6    18.23        46.00
832.569875                       25.42        QP         278.0         V           223.0             5.2     25.42        46.00
868.016750                       42.61        QP         105.0         V            44.0              6.0    3.39         46.00




FCC Part 15.231                                                                                                    Page 17 of 23


   Bay Area Compliance Laboratories Corp. (Shenzhen)                            Report No.: RSZ190606002-00B



Frequency
                     Receiver             Rx Antenna                            FCC Part 15.231(b)
  (MHz)                                              Corrected Corrected
                                Turntable
                                                      Factor Amplitude
            Reading  Detector    Degree Height Polar                       Limit   Margin
                                                      (dB/m) (dBμV/m)                          Comment
            (dBμV) (PK/QP/Ave.)           (cm) (H/V)                     (dBμV/m)    (dB)

 434.21      89.86          PK       359       1       H      -8.9      80.00       80.8       0.80    Fundamental
 434.21      79.35          PK       165      1.1      V      -8.9      70.45       80.8      10.35    Fundamental
 1302.63     50.44          PK       349      1.6      H     -4.31      46.13       54         7.87    Harmonic
 1302.63     46.93          PK        98      2.0      V     -4.31      42.62       54        11.38    Harmonic
 1736.84     53.86          PK       285      2.2      H     -2.01      51.85       60.8       8.95    Harmonic
 1736.84     53.00          PK       151      2.4      V     -2.01      50.99       60.8       9.81    Harmonic


   Note:
   Peak detector data complies with the average limit, average results are not necessary to provide.
   Corrected Amplitude = Corrected Factor + Reading
   Corrected Factor = Antenna factor (Rx) + cable loss – amplifier factor
   Margin = Limit - Corr. Amplitude




   FCC Part 15.231                                                                               Page 18 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)              Report No.: RSZ190606002-00B


                                                Horizontal




             Date: 16.JUL.2019   20:16:47




                                                    Vertical




             Date: 16.JUL.2019   20:10:48




FCC Part 15.231                                                                Page 19 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                         Report No.: RSZ190606002-00B


FCC §15.231(a) (1) - DEACTIVATION TESTING
Applicable Standard
Per FCC §15.231(a) (1), A manually operated transmitter shall employ a switch that will automatically
deactivate the transmitter within not more than 5 seconds of being released.

Test Procedure
1.   Remove the antenna from the EUT and then connect a low loss RF cable from the antenna port to the
     spectrum analyzer.
2.   Set center frequency of spectrum analyzer=operating frequency.
3.   Set the spectrum analyzer as RBW=100kHz/ VBW=300kHz/ Span=0Hz.
4.   Repeat above procedures until all frequency measured was complete.


Test Data
 Environmental Conditions

            Temperature:               25 ℃
       Relative Humidity:              50 %
          ATM Pressure:              101.0 kPa

     The testing was performed by Andy Yu on 2019-06-26.

Test mode: Transmitting




FCC Part 15.231                                                                             Page 20 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                                Report No.: RSZ190606002-00B


Test Result: Compliant. This product will cease transmission within 5 seconds after activation. Please
refer to following plots.


                                                                    RBW 100 kHz    Delta 2 [T1 ]
                                                                   * VBW 300 kHz               0.31 dB
                      Ref   99 dBµV              * Att   10 dB      SWT 10 s               5.000000 s

                                                                                   Marker 1 [T1 ]
                                                                                             29.20 dBµV
                      90                                                                   1.020641 s      A
                                                                                   Delta 1 [T1 ]
               1 PK                                                                           -0.10 dB
               MAXH   80                                                                 929.487179 ms


                      70



                      60



                      50


                                                                                                          3DB
                      40


                              1        1                                 2
                      30



                      20



                      10


                      0

                      Center 434.21 MHz                          1 s/




              Date: 26.JUN.2019       16:04:23




FCC Part 15.231                                                                                                  Page 21 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                       Report No.: RSZ190606002-00B

FCC §15.231(c) – 20 dB EMISSION BANDWIDTH TESTING
Applicable Standard
Per 15.231(c), The bandwidth of the emission shall be no wider than 0.25% of the center frequency for
devices operating above 70 MHz and below 900 MHz. Bandwidth is determined at the points 20 dB down
from the modulated carrier.


Test Procedure
With the EUT’s antenna attached, the waveforzm was received by the test antenna which was connected to
the spectrum analyzer, plot the 20 dB bandwidth.


Test Data
 Environmental Conditions

            Temperature:               25 ℃
       Relative Humidity:              50 %
          ATM Pressure:              101.0 kPa

     The testing was performed by Andy Yu on 2019-06-26.

    Test Mode: Transmitting

    Please refer to following table and plots.




FCC Part 15.231                                                                          Page 22 of 23


Bay Area Compliance Laboratories Corp. (Shenzhen)                                                      Report No.: RSZ190606002-00B


         Channel Frequency                  20 dB Emission Bandwidth                         <Limit
                                                                                                                      Result
              (MHz )                                 (kHz)                                   (kHz)
                  434.21                                        93.59                        1085.525                  Pass


                                                   20 dB Emission Bandwidth
                                                                         * RBW 3 kHz    Delta 1 [T1 ]
                                                                         * VBW 10 kHz               0.46 dB
                      Ref   96 dBµV              * Att       10 dB        SWT 45 ms         93.589743590 kHz

                                                                                        Marker 1 [T1 ]
                               D1 92.66 dBµV
                      90                                                                          72.37 dBµV
                                                                                           434.158717949 MHz     A

               1 PK   80
               MAXH
                                                         1                     1
                                  D2 72.66 dBµV
                      70



                      60



                      50



                      40                                                                                        3DB



                      30



                      20



                      10



                      0

                      Center   434.21 MHz                            40 kHz/                   Span   400 kHz




              Date: 26.JUN.2019       15:41:41




                                            ***** END OF REPORT *****




FCC Part 15.231                                                                                                           Page 23 of 23



Document Created: 2019-07-18 23:55:47
Document Modified: 2019-07-18 23:55:47

© 2024 FCC.report