RF_Annex D Plots of Maximum SAR Test

FCC ID: 2ARHS-W562

RF Exposure Info

Download: PDF
FCCID_4114581

                                                                                               REPORT No.:SZ18100025S01



          Annex D Plots of Maximum SAR Test Results




NOTE: This document is issued by MORLAB, the test report shall not be reproduced except in full without prior written permission of the
company. The test results apply only to the particular sample(s) tested and to the specific tests carried out which is available on request for
validation and information confirmed at our website.

                         SHENZHEN MORLAB COMMUNICATIONS TECHNOLOGY Co., Ltd.                   Tel: 86-755-36698555   Fax: 86-755-36698525
 MORLAB                  FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,
                         Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.cn   E-mail: service@morlab.cn


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.18

WLAN2.4GHz_802.11b 1Mbps_Front Side_10mm_Ch11



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2462 MHz;Duty Cycle: 1:1
Medium: HSL_2450_181018 Medium parameters used: f = 2462 MHz; σ = 1.878 S/m; εr = 40.042; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(4.61, 4.61, 4.61); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch11/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0976 W/kg

Ch11/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 0.35 V/m; Power Drift = 0.10 dB
Peak SAR (extrapolated) = 0.209 W/kg
SAR(1 g) = 0.106 W/kg; SAR(10 g) = 0.052 W/kg
Maximum value of SAR (measured) = 0.112 W/kg




               0 dB = 0.112 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 1_802.11n-HT20 6.5Mbps_Front Side_10mm_Ch36



Communication System: UID 0, WLAN 5GHz (0); Frequency: 5180 MHz;Duty Cycle: 1:1
Medium: HSL_5200_181115 Medium parameters used: f = 5180 MHz; σ = 4.597 S/m; εr = 37.051; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(5.28, 5.28, 5.28); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch36/Area Scan (61x61x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.0443 W/kg

Ch36/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 0.451 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 0.340 W/kg
SAR(1 g) = 0.073 W/kg; SAR(10 g) = 0.022 W/kg
Maximum value of SAR (measured) = 0.0616 W/kg




               0 dB = 0.0616 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 2_802.11a 6Mbps_Front Side_10mm_Ch52



Communication System: UID 0, WLAN 5GHz (0); Frequency: 5260 MHz;Duty Cycle: 1:1
Medium: HSL_5200_181115 Medium parameters used: f = 5260 MHz; σ = 5.191 S/m; εr = 47.427; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(5.28, 5.28, 5.28); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch52/Area Scan (61x61x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.0149 W/kg

Ch52/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 0.456 V/m; Power Drift = -0.09 dB
Peak SAR (extrapolated) = 0.323 W/kg
SAR(1 g) = 0.069 W/kg; SAR(10 g) = 0.019 W/kg
Maximum value of SAR (measured) = 0.0652 W/kg




               0 dB = 0.0652 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 4_802.11a 6Mbps_Front Side_10mm_Ch149



Communication System: UID 0, WLAN 5GHz (0); Frequency: 5745 MHz;Duty Cycle: 1:1
Medium: HSL_5200_181115 Medium parameters used: f = 5745 MHz; σ = 5.451 S/m; εr = 35.785; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(4.60, 4.60, 4.60); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch149/Area Scan (61x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.0211 W/kg

Ch149/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 0.145 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 0.188 W/kg
SAR(1 g) = 0.036 W/kg; SAR(10 g) = 0.00952 W/kg
Maximum value of SAR (measured) = 0.0324 W/kg




               0 dB = 0.0324 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.10.18

WLAN2.4GHz_802.11b 1Mbps_Back Side_0mm_Ch11



Communication System: UID 0, WLAN 2.4GHz 802.11b (0); Frequency: 2462 MHz;Duty Cycle: 1:1
Medium: MSL_2450_181018 Medium parameters used: f = 2462 MHz; σ = 2.058 S/m; εr = 50.542; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3154; ConvF(4.28, 4.28, 4.28); Calibrated: 2017.10.30;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch11/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 1.44 W/kg

Ch11/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.873 V/m; Power Drift = 0.29 dB
Peak SAR (extrapolated) = 3.13 W/kg
SAR(1 g) = 1.33 W/kg; SAR(10 g) = 0.531 W/kg
Maximum value of SAR (measured) = 1.59 W/kg




               0 dB = 1.59 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 1_802.11n-HT20 6.5Mbps_Back Side_0mm_Ch36



Communication System: UID 0, WLAN 5GHz (0); Frequency: 5180 MHz;Duty Cycle: 1:1
Medium: MSL_5200_181115 Medium parameters used: f = 5180 MHz; σ = 5.357 S/m; εr = 48.461; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(4.73, 4.73, 4.73); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch36/Area Scan (61x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.755 W/kg

Ch36/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.560 V/m; Power Drift = 2.58 dB
Peak SAR (extrapolated) = 1.80 W/kg
SAR(1 g) = 0.613 W/kg; SAR(10 g) = 0.221 W/kg
Maximum value of SAR (measured) = 0.703 W/kg




               0 dB = 0.703 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 2_802.11a 6Mbps_Back Side_0mm_Ch52

Communication System: UID 0, WLAN 5GHz (0); Frequency: 5260 MHz;Duty Cycle: 1:1
Medium: MSL_5200_181115 Medium parameters used: f = 5260 MHz; σ = 5.501 S/m; εr = 47.727; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(4.73, 4.73, 4.73); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 2; Type: QD000P40CC; Serial: TP:1464
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch52/Area Scan (61x61x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.12 W/kg

Ch52/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.487 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 2.21 W/kg
SAR(1 g) = 0.737 W/kg; SAR(10 g) = 0.236 W/kg
Maximum value of SAR (measured) = 0.911 W/kg




               0 dB = 0.911 W/kg


Test Laboratory: SAR Lab. of Shenzhen Morlab Communications Technology Co., Ltd.   Date: 2018.11.15

WLAN 5GHz Band 4_802.11a 6Mbps_Back Side_0mm_Ch149



Communication System: UID 0, WLAN 5GHz (0); Frequency: 5745 MHz;Duty Cycle: 1:1
Medium: MSL_5200_181115 Medium parameters used: f = 5745 MHz; σ = 6.151 S/m; εr = 46.888; ρ
= 1000 kg/m3
Ambient Temperature:23.1 ℃; Liquid Temperature:21.1 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN3823; ConvF(3.98, 3.98, 3.98); Calibrated: 2018.11.12;
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn1516; Calibrated: 2018.07.14
- Phantom: SAM 1; Type: QD000P40CC; Serial: TP:1471
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Ch149/Area Scan (61x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.563 W/kg

Ch149/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 1.076 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 1.26 W/kg
SAR(1 g) = 0.402 W/kg; SAR(10 g) = 0.142 W/kg
Maximum value of SAR (measured) = 0.426 W/kg




               0 dB = 0.563 W/kg