test plot SAR

FCC ID: 2ARHHDOTR3000

Test Report

Download: PDF
FCCID_4402801

Report No.: DRRFCC1908-0066      FCC ID: 2ARHHDOTR3000           Date of issue: Aug. 09, 2019



SAR Plots

    Verification Plots


    SAR Test Plots




                              Copyright Ⓒ 2019, DT&C Co., Ltd.


                                      DT&C Co., Ltd.
             DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 - SN:1049
            Communication System: UID 0, CW (0); Frequency: 750 MHz;Duty Cycle: 1:1
           Medium parameters used: f = 750 MHz; σ = 0.962 S/m; εr = 56.591; ρ = 1000 kg/m3
                                    Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.43, 10.43, 10.43); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-05; Ambient Temp: 21.5; Tissue Temp: 21.7

                         750 MHz System Body Verification (250mW)

                Area Scan (6x15x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                      Power Drift = -0.12 dB
                               Peak SAR (extrapolated) = 3.52 W/kg
                           SAR(1 g) = 2.19 W/kg; SAR(10 g) = 1.42 W/kg




               0 dB = 2.86 W/kg


                                      DT&C Co., Ltd.
            DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN:4d159
              Communication System: UID 0, CW; Frequency: 835 MHz;Duty Cycle: 1:1
           Medium parameters used: f = 835 MHz; σ = 1.008 S/m; εr = 56.598; ρ = 1000 kg/m3
                                    Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.27, 10.27, 10.27); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-01; Ambient Temp: 21.2; Tissue Temp: 21.5

                         835 MHz System Body Verification (250mW)

                Area Scan (6x11x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                       Power Drift = 0.08 dB
                               Peak SAR (extrapolated) = 3.58 W/kg
                           SAR(1 g) = 2.34 W/kg; SAR(10 g) = 1.48 W/kg




               0 dB = 3.01 W/kg


                                    DT&C Co., Ltd.
          DUT: Dipole 900 MHz; Type: D900V2; Serial: D900V2 - SN:1d175
            Communication System: UID 0, CW; Frequency: 900 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 900 MHz; σ = 1.066 S/m; εr = 56.947; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.2, 10.2, 10.2); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-22; Ambient Temp: 21.1; Tissue Temp: 21.5

                       900 MHz System Body Verification (250mW)

               Area Scan (5x12x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                    Power Drift = -0.01 dB
                             Peak SAR (extrapolated) = 3.76 W/kg
                         SAR(1 g) = 2.85 W/kg; SAR(10 g) = 1.72 W/kg




              0 dB = 2.79 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 1800 MHz; Type: D1800V2; Serial: D1800V2 - SN:2d047
          Communication System: UID 0, CW (0); Frequency: 1800 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 1800 MHz; σ = 1.563 S/m; εr = 52.17; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.62, 8.62, 8.62); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-03; Ambient Temp: 21.3; Tissue Temp: 21.5

                      1800 MHz System Body Verification (100mW)

               Area Scan (7x10x1): Measurement grid: dx=10mm, dy=10mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.00 dB
                             Peak SAR (extrapolated) = 8.95 W/kg
                         SAR(1 g) = 4.11 W/kg; SAR(10 g) = 1.83 W/kg




              0 dB = 6.43 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN:5d176
            Communication System: UID 0, CW; Frequency: 1900 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 1900 MHz; σ = 1.537 S/m; εr = 52.039; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.21, 8.21, 8.21); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-02; Ambient Temp: 21.4; Tissue Temp: 21.6

                      1900 MHz System Body Verification (100mW)

               Area Scan (7x10x1): Measurement grid: dx=10mm, dy=10mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                      Power Drift = 0.06 dB
                              Peak SAR (extrapolated) = 7.53 W/kg
                          SAR(1 g) = 3.95 W/kg; SAR(10 g) = 2.05 W/kg




              0 dB = 5.59 W/kg


                                    DT&C Co., Ltd.
         DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 - SN:920
            Communication System: UID 0, CW; Frequency: 2450 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 2450 MHz; σ = 1.957 S/m; εr = 54.441; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(7.75, 7.75, 7.75); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-08; Ambient Temp: 21.1; Tissue Temp: 21.3

                      2450 MHz System Body Verification (100mW)

                Area Scan (7x9x1): Measurement grid: dx=12mm, dy=12mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                      Power Drift = 0.07 dB
                             Peak SAR (extrapolated) = 10.98 W/kg
                          SAR(1 g) = 5.31 W/kg; SAR(10 g) = 2.45 W/kg




              0 dB = 6.95 W/kg


                                    DT&C Co., Ltd.
         DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 - SN:1016
          Communication System: UID 0, CW (0); Frequency: 2600 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 2600 MHz; σ = 2.133 S/m; εr = 51.434; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(7.63, 7.63, 7.63); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-04; Ambient Temp: 21.2; Tissue Temp: 21.3

                      2600 MHz System Body Verification (100mW)

                Area Scan (6x8x1): Measurement grid: dx=12mm, dy=12mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                      Power Drift = 0.00 dB
                              Peak SAR (extrapolated) = 12.1 W/kg
                          SAR(1 g) = 5.49 W/kg; SAR(10 g) = 2.46 W/kg




              0 dB = 10.1 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 5000 MHz; Type: D5GHzV2; Serial: D5GHzV2 - SN:1103
          Communication System: UID 0, CW (0); Frequency: 5300 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 5300 MHz; σ = 5.516 S/m; εr = 48.39; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(4.51, 4.51, 4.51); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-09; Ambient Temp: 21.5; Tissue Temp: 21.8

                      5300 MHz System Body Verification (100mW)

              Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                    Power Drift = -0.06 dB
                             Peak SAR (extrapolated) = 30.2 W/kg
                         SAR(1 g) = 7.37 W/kg; SAR(10 g) = 2.03 W/kg




              0 dB = 17.5 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 5000 MHz; Type: D5GHzV2; Serial: D5GHzV2 - SN:1103
          Communication System: UID 0, CW (0); Frequency: 5600 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 5600 MHz; σ = 5.932 S/m; εr = 47.398; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(4.01, 4.01, 4.01); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-10; Ambient Temp: 21.2; Tissue Temp: 21.5

                      5600 MHz System Body Verification (100mW)

              Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                      Power Drift = 0.03 dB
                              Peak SAR (extrapolated) = 34.7 W/kg
                          SAR(1 g) = 7.86 W/kg; SAR(10 g) = 2.19 W/kg




              0 dB = 19.6 W/kg


                                   DT&C Co., Ltd.
       DUT: Dipole 5000 MHz; Type: D5GHzV2; Serial: D5GHzV2 - SN:1103
         Communication System: UID 0, CW (0); Frequency: 5800 MHz;Duty Cycle: 1:1
        Medium parameters used: f = 5800 MHz; σ = 6.222 S/m; εr = 46.984; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                    DASY5 Configuration:
 Probe: EX3DV4 - SN3933; ConvF(4.1, 4.1, 4.1); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-10; Ambient Temp: 21.2; Tissue Temp: 21.5

                      5800 MHz System Body Verification (100mW)

              Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                     Power Drift = 0.00 dB
                             Peak SAR (extrapolated) = 34.6 W/kg
                         SAR(1 g) = 7.57 W/kg; SAR(10 g) = 2.08 W/kg




             0 dB = 18.1 W/kg


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
     Communication System: UID 0, GSM 850_4Tx (0); Frequency: 836.6 MHz;Duty Cycle: 1:2.075
        Medium parameters used: f = 836.6 MHz; σ = 1.009 S/m; εr = 56.583; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.27, 10.27, 10.27); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-01; Ambient Temp: 21.2 Tissue Temp: 21.5

           Touch from Body, Bottom, GSM850 GPRS 4 Tx Ch. 190, Ant Internal

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = -0.06 dB
                              Peak SAR (extrapolated) = 0.617 W/kg
                          SAR(1 g) = 0.481 W/kg; SAR(10 g) = 0.359 W/kg




               0 dB = 0.559 W/kg
                                                                       A1


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
    Communication System: UID 0, PCS1900_4Tx (0); Frequency: 1880 MHz;Duty Cycle: 1:2.075
       Medium parameters used: f = 1880 MHz; σ = 1.521 S/m; εr = 52.075; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.21, 8.21, 8.21); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-02; Ambient Temp: 21.4 Tissue Temp: 21.6

         Touch from Body, Bottom, PCS1900 GPRS 4 Tx Ch. 661, Ant Internal

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = -0.03 dB
                            Peak SAR (extrapolated) = 0.365 W/kg
                        SAR(1 g) = 0.233 W/kg; SAR(10 g) = 0.136 W/kg




              0 dB = 0.307 W/kg
                                                                   A2


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
       Communication System: UID 0, WCDMA 850 (0); Frequency: 836.6 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 836.6 MHz; σ = 1.009 S/m; εr = 56.583; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.27, 10.27, 10.27); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-01; Ambient Temp: 21.2; Tissue Temp: 21.5

            Touch from Body, Bottom, WCDMA Band 5, Ch. 4183, Ant Internal

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = -0.04 dB
                              Peak SAR (extrapolated) = 0.400 W/kg
                          SAR(1 g) = 0.306 W/kg; SAR(10 g) = 0.229 W/kg




               0 dB = 0.358 W/kg
                                                                        A3


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
    Communication System: UID 0, WCDMA 1700 (0); Frequency: 1732.4 MHz;Duty Cycle: 1:1
      Medium parameters used: f = 1732.4 MHz; σ = 1.509 S/m; εr = 52.426; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.62, 8.62, 8.62); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-03; Ambient Temp: 21.3; Tissue Temp: 21.5

          Touch from Body, Bottom, WCDMA Band 4, Ch. 1412, Ant Internal

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = -0.04 dB
                            Peak SAR (extrapolated) = 0.299 W/kg
                        SAR(1 g) = 0.191 W/kg; SAR(10 g) = 0.119 W/kg




              0 dB = 0.248 W/kg
                                                                    A4


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
     Communication System: UID 0, WCDMA 1900 (0); Frequency: 1880 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 1880 MHz; σ = 1.521 S/m; εr = 52.075; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.21, 8.21, 8.21); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-02; Ambient Temp: 21.4; Tissue Temp: 21.6

          Touch from Body, Bottom, WCDMA Band 2, Ch. 9400, Ant Internal

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = 0.03 dB
                            Peak SAR (extrapolated) = 0.469 W/kg
                        SAR(1 g) = 0.286 W/kg; SAR(10 g) = 0.171 W/kg




              0 dB = 0.379 W/kg
                                                                   A5


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
    Communication System: UID 0, LTE Band 12(FCC) (0); Frequency: 707.5 MHz; Duty Cycle: 1:1
        Medium parameters used: f = 707.5 MHz; σ = 0.925 S/m; εr = 56.964; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.43, 10.43, 10.43); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-05; Ambient Temp: 21.5; Tissue Temp: 21.7

             Touch from Body, Bottom, LTE Band 12 Ch. 23095, Ant Internal

                        Mode : BandWidth 10 MHz, QPSK, RB Size: 1

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = -0.19 dB
                              Peak SAR (extrapolated) = 0.261 W/kg
                          SAR(1 g) = 0.211 W/kg; SAR(10 g) = 0.164 W/kg




               0 dB = 0.239 W/kg
                                                                      A6


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
        Communication System: UID 0, LTE Band 13 (0); Frequency: 782 MHz; Duty Cycle: 1:1
          Medium parameters used: f = 782 MHz; σ = 0.994 S/m; εr = 56.295; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.43, 10.43, 10.43); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-05; Ambient Temp: 21.5; Tissue Temp: 21.7

             Touch from Body, Bottom, LTE Band 13 Ch. 23230, Ant Internal

                        Mode : BandWidth 10 MHz, QPSK, RB Size: 1

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = 0.16 dB
                              Peak SAR (extrapolated) = 0.457 W/kg
                          SAR(1 g) = 0.358 W/kg; SAR(10 g) = 0.275 W/kg




               0 dB = 0.412 W/kg
                                                                     A7


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
        Communication System: UID 0, LTE Band 14 (0); Frequency: 793 MHz; Duty Cycle: 1:1
          Medium parameters used: f = 793 MHz; σ = 1.005 S/m; εr = 56.197; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.43, 10.43, 10.43); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-05; Ambient Temp: 21.5; Tissue Temp: 21.7

             Touch from Body, Bottom, LTE Band 14 Ch. 23330, Ant Internal

                        Mode : BandWidth 10 MHz, QPSK, RB Size: 1

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = 0.02 dB
                              Peak SAR (extrapolated) = 0.343 W/kg
                          SAR(1 g) = 0.267 W/kg; SAR(10 g) = 0.201 W/kg




               0 dB = 0.310 W/kg
                                                                     A8


                                      DT&C Co., Ltd.
                                DUT: DOTR-3000; Type: Gun
     Communication System: UID 0, LTE Band 5(FCC) (0); Frequency: 836.5 MHz; Duty Cycle: 1:1
        Medium parameters used: f = 836.5 MHz; σ = 1.009 S/m; εr = 56.585; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.27, 10.27, 10.27); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-01; Ambient Temp: 21.2; Tissue Temp: 21.5

              Touch from Body, Bottom, LTE Band 5 Ch. 20525, Ant Internal

                        Mode : BandWidth 10 MHz, QPSK, RB Size: 1

                Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
           Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                       Power Drift = -0.17 dB
                              Peak SAR (extrapolated) = 0.396 W/kg
                          SAR(1 g) = 0.300 W/kg; SAR(10 g) = 0.223 W/kg




               0 dB = 0.354 W/kg
                                                                       A9


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
   Communication System: UID 0, LTE Band 4(FCC) (0); Frequency: 1732.5 MHz; Duty Cycle: 1:1
      Medium parameters used: f = 1732.5 MHz; σ = 1.509 S/m; εr = 52.426; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.62, 8.62, 8.62); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-03; Ambient Temp: 21.3; Tissue Temp: 21.5

            Touch from Body, Bottom, LTE Band 4 Ch. 20175, Ant Internal

                      Mode : BandWidth 20 MHz, QPSK, RB Size: 1

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = -0.00 dB
                            Peak SAR (extrapolated) = 0.316 W/kg
                        SAR(1 g) = 0.204 W/kg; SAR(10 g) = 0.127 W/kg




              0 dB = 0.264 W/kg
                                                                   A10


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
    Communication System: UID 0, LTE Band 2(FCC) (0); Frequency: 1900 MHz; Duty Cycle: 1:1
       Medium parameters used: f = 1900 MHz; σ = 1.537 S/m; εr = 52.039; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(8.21, 8.21, 8.21); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-02; Ambient Temp: 21.4; Tissue Temp: 21.6

            Touch from Body, Bottom, LTE Band 2 Ch. 19100, Ant Internal

                      Mode : BandWidth 20 MHz, QPSK, RB Size: 1

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = -0.07 dB
                            Peak SAR (extrapolated) = 0.511 W/kg
                        SAR(1 g) = 0.318 W/kg; SAR(10 g) = 0.191 W/kg




              0 dB = 0.418 W/kg
                                                                 A11


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
    Communication System: UID 0, LTE Band 7(FCC) (0); Frequency: 2535 MHz; Duty Cycle: 1:1
       Medium parameters used: f = 2535 MHz; σ = 2.055 S/m; εr = 51.636; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(7.63, 7.63, 7.63); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-04; Ambient Temp: 21.2; Tissue Temp: 21.3

            Touch from Body, Bottom, LTE Band 7 Ch. 21100, Ant Internal

                      Mode : BandWidth 20 MHz, QPSK, RB Size: 1

               Area Scan (11x21x1): Measurement grid: dx=12mm, dy=12mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = -0.00 dB
                            Peak SAR (extrapolated) = 0.584 W/kg
                        SAR(1 g) = 0.324 W/kg; SAR(10 g) = 0.185 W/kg




              0 dB = 0.450 W/kg
                                                                      A12


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
 Communication System: UID 0, LTE Band 41(TDD) (0); Frequency: 2593 MHz; Duty Cycle: 1:1.58
      Medium parameters used: f = 2593 MHz; σ = 2.124 S/m; εr = 51.453; ρ = 1000 kg/m3
                                Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(7.63, 7.63, 7.63); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-04; Ambient Temp: 21.2; Tissue Temp: 21.3

            Touch from Body, Bottom, LTE Band 41 Ch. 40620, Ant Internal

                      Mode : BandWidth 20 MHz, QPSK, RB Size: 1

               Area Scan (11x21x1): Measurement grid: dx=12mm, dy=12mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.14 dB
                            Peak SAR (extrapolated) = 0.482 W/kg
                        SAR(1 g) = 0.264 W/kg; SAR(10 g) = 0.149 W/kg




              0 dB = 0.368 W/kg
                                                                      A13


                                      DT&C Co., Ltd.
                                 DUT: DOTR-3000; Type: Gun
Communication System: UID 0, W-LAN 2.4G(802.11b/g/n20, 40) (0); Frequency: 2462 MHz;Duty Cycle: 1:1
          Medium parameters used: f = 2462 MHz; σ = 1.973 S/m; εr = 54.409; ρ = 1000 kg/m3
                                    Phantom section: Flat Section

                                      DASY5 Configuration:
  Probe: EX3DV4 - SN3933; ConvF(7.75, 7.75, 7.75); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                        Sensor-Surface: 2mm (Mechanical Surface Detection)
               Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
           Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                    Test Date: 2019-07-08; Ambient Temp: 21.1; Tissue Temp: 21.3

              Touch from Body, Bottom, W-LAN(802.11b) Ch. 11, Ant Internal

                 Area Scan (11x21x1): Measurement grid: dx=12mm, dy=12mm
           Zoom Scan (10x10x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                       Power Drift = 0.15 dB
                              Peak SAR (extrapolated) = 0.0420 W/kg
                         SAR(1 g) = 0.016 W/kg; SAR(10 g) = 0.00568 W/kg




                0 dB = 0.0258 W/kg
                                                                    A14


                                    DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
Communication System: UID 0, W-LAN 5.3G(802.11a/n/ac) (0); Frequency: 5300 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 5300 MHz; σ = 5.516 S/m; εr = 48.39; ρ = 1000 kg/m3
                                Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(4.51, 4.51, 4.51); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-09; Ambient Temp: 21.5; Tissue Temp: 21.8

            Touch from Body, Bottom, W-LAN(802.11a) Ch. 60, Ant Internal

             Area Scan (13x25x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                     Power Drift = 0.00 dB
                            Peak SAR (extrapolated) = 0.345 W/kg
                        SAR(1 g) = 0.073 W/kg; SAR(10 g) = 0.026 W/kg




              0 dB = 0.202 W/kg
                                                                      A15


                                   DT&C Co., Ltd.
                              DUT: DOTR-3000; Type: Gun
Communication System: UID 0, W-LAN 5.6G(802.11a/n/ac) (0); Frequency: 5700 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 5700 MHz; σ = 6.074 S/m; εr = 47.184; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
 Probe: EX3DV4 - SN3933; ConvF(4.1, 4.1, 4.1); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-10; Ambient Temp: 21.2; Tissue Temp: 21.5

           Touch from Body, Bottom, W-LAN(802.11a) Ch. 140, Ant Internal

             Area Scan (13x25x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                     Power Drift = -0.17 dB
                            Peak SAR (extrapolated) = 0.624 W/kg
                        SAR(1 g) = 0.131 W/kg; SAR(10 g) = 0.043 W/kg




             0 dB = 0.344 W/kg
                                                                     A16


                                   DT&C Co., Ltd.
                              DUT: DOTR-3000; Type: Gun
Communication System: UID 0, W-LAN 5.8G(802.11a/n/ac) (0); Frequency: 5745 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 5745 MHz; σ = 6.142 S/m; εr = 47.099; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                    DASY5 Configuration:
 Probe: EX3DV4 - SN3933; ConvF(4.1, 4.1, 4.1); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                     Sensor-Surface: 1.4mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-07-10; Ambient Temp: 21.2; Tissue Temp: 21.5

           Touch from Body, Bottom, W-LAN(802.11a) Ch. 149, Ant Internal

             Area Scan (13x25x1): Measurement grid: dx=10mm, dy=10mm
Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm, Graded Ratio: 1.4
                                     Power Drift = 0.08 dB
                            Peak SAR (extrapolated) = 0.457 W/kg
                        SAR(1 g) = 0.098 W/kg; SAR(10 g) = 0.033 W/kg




             0 dB = 0.248 W/kg
                                                                   A17


                                     DT&C Co., Ltd.
                               DUT: DOTR-3000; Type: Gun
   Communication System: UID 0, RFID(FCC) (0); Frequency: 914.75 MHz;Duty Cycle: 1:4.903
  Medium parameters used (interpolated): f = 914.75 MHz; σ = 1.089 S/m; εr = 56.92; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3933; ConvF(10.2, 10.2, 10.2); Calibrated: 9/25/2018; Electronics: DAE4 Sn1396
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-07-22; Ambient Temp: 21.1; Tissue Temp: 21.5

                Touch from Body, Pistol Grip, RFID Ch. 25, Ant Internal

               Area Scan (9x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (8x7x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = -0.06 dB
                             Peak SAR (extrapolated) = 0.527 W/kg
                         SAR(1 g) = 0.332 W/kg; SAR(10 g) = 0.234 W/kg




              0 dB = 0.423 W/kg
                                                                      A18