Test Setup Photos (inside report)

FCC ID: 2AR4G-HL7618RD

Test Setup Photos

Download: PDF
FCCID_4197310

Intertek
731 Enterprise Drive
Lexington, KY 40510

Tel 859 226 1000
Fax 859 226 1040

www.intertek.com




eNeura, Inc.
TEST REPORT
SCOPE OF WORK
EMC TESTING – STMS MINI TRANSCRANIAL MAGNETIC STIMULATOR

REPORT NUMBER
103714598LEX-006

ISSUE DATE

2/1/2019

PAGES
31

DOCUMENT CONTROL NUMBER
Non-Specific EMC Report Shell Rev. December 2017
© 2017 INTERTEK


                                                       EMC TEST REPORT
                                                            (FULL COMPLIANCE)

                                                  Report Number:            103714598LEX-006
                                                  Project Number:           G103714598

                                                Report Issue Date:          2/1/2019

                                                  Model(s) Tested:          sTMS Mini Transcranial Magnetic
                                                                            Stimulator



                                                          Standards:        FCC Part 15B
                                                                            ICES-003 Issue 6
                                                                            FCC Part 27
                                                                            (Radiated Spurious Emissions)




                         Tested by:                                                                    Client:
             Intertek Testing Services NA, Inc.                                                      eNeura, Inc.
                     731 Enterprise Dr.                                                         715 N Pastoria Avenue
                    Lexington, KY 40510                                                          Sunnyvale, CA 94085
                            USA                                                                         USA




                       Report prepared by                                                            Report reviewed by




                   Bryan Taylor, Team Leader                                                        Brian Daffin, Engineer




This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement between Intertek and its Client. Intertek's
responsibility and liability are limited to the terms and conditions of the agreement. Intertek assumes no liability to any party, other than to the
Client in accordance with the agreement, for any loss, expense or damage occasioned by the use of this report. Only the Client is authorized to
copy or distribute this report and then only in its entirety. Any use of the Intertek name or one of its marks for the sale or advertisement of the
tested material, product or service must first be approved in writing by Intertek. The observations and test results in this report are relevant only
to the sample tested. This report by itself does not imply that the material, product, or service is or has ever been under an Intertek certification
program.



                                                               Intertek Testing Services NA, Inc.


                                                                                                           Evaluation For: eNeura, Inc.
                                                                                   Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                                        Date: 2/1/2019

                                                                Table of Contents


1   Introduction and Conclusion ..................................................................................................................... 4
2   Test Summary .......................................................................................................................................... 4
3   Client Information .................................................................................................................................... 5
4   Description of Equipment under Test and Variant Models .......................................................................... 6
5   System Setup and Method ........................................................................................................................ 7
6   Radiated Emissions ................................................................................................................................ 11
7   Conducted Emissions .............................................................................................................................. 27
8   Revision History ..................................................................................................................................... 31




Non-Specific EMC Report Shell Rev. December 2017                                                                                            Page 3 of 31
Report Number: 103714598LEX-006


                                                                                       Evaluation For: eNeura, Inc.
                                                               Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                    Date: 2/1/2019


1    Introduction and Conclusion

The tests indicated in section 2.0 were performed on the product constructed as described in section 4.0. The
remaining test sections are the verbatim text from the actual data sheets used during the investigation. These test
sections include the test name, the specified test Method, a list of the actual Test Equipment Used, documentation
Photos, Results and raw Data. No additions, deviations, or exclusions have been made from the standard(s) unless
specifically noted.

Based on the results of our investigation, we have concluded the product tested complies with the requirements of
the standard(s) indicated. The results obtained in this test report pertain only to the item(s) tested. Intertek does
not make any claims of compliance for samples or variants which were not tested.

2    Test Summary

        Section     Test full name                                                                  Result

                    Radiated Emissions (Transmitters Idle)
           6                                                                                         Pass
                    (ANSI C63.4:2014)
                    Radiated Spurious Emissions (Transmitters Active)
           6                                                                                         Pass
                    (ANSI C63.26:2015)
                    Conducted Emissions
           7                                                                                         Pass
                    (ANSI C63.4:2014)




Non-Specific EMC Report Shell Rev. December 2017                                                        Page 4 of 31
Report Number: 103714598LEX-006


                                                                                    Evaluation For: eNeura, Inc.
                                                            Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                 Date: 2/1/2019

3    Client Information

This product was tested at the request of the following:

                                               Client Information
                     Client Name:   eNeura, Inc.
                         Address:   715 N Pastoria Avenue
                                    Sunnyvale, CA 94085
                                    USA

                         Contact:   Cynthia Merrell
                      Telephone:    (408) 245-6400 x126
                           Email:   cmerell@eneura.com
                                           Manufacturer Information
           Manufacturer Name:       eNeura, Inc.
          Manufacturer Address:     715 N Pastoria Avenue
                                    Sunnyvale, CA 94085
                                    USA




Non-Specific EMC Report Shell Rev. December 2017                                                    Page 5 of 31
Report Number: 103714598LEX-006


                                                                                   Evaluation For: eNeura, Inc.
                                                           Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                Date: 2/1/2019

4      Description of Equipment under Test and Variant Models

                                                Equipment Under Test
                      Product Name sTMS Mini Transcranial Magnetic Stimulator
                     Model Number sTMS Mini
                       Serial Number 20001152
                        Receive Date 11/20/2018
                      Test Start Date 11/21/2018
                       Test End Date 12/4/2018
           Device Received Condition Good
                   Test Sample Type Production
                       Rated Voltage 12VDC
                       Rated Current 1.25A
                    Rated Frequency DC
                   Number of Phases DC
                               Description of Equipment Under Test (provided by client)
    The sTMS Mini Transcranial Magnetic Stimulator is a therapeutic device used in the treatment of
    migraine headaches. It applies magnetic stimulation therapy and then transmits data back to the
    supervising clinician via an embedded cellular module (Sierra Wireless Model HL7618RD). The data
    transmission occurs after the therapy has completed and the device could then be lying in the users lap
    or in a purse or bag. The data transmission does not occur whilst the device is being held to the head.


4.1      Variant Models:

There were no variant models covered by this evaluation.




Non-Specific EMC Report Shell Rev. December 2017                                                   Page 6 of 31
Report Number: 103714598LEX-006


                                                                                      Evaluation For: eNeura, Inc.
                                                              Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                   Date: 2/1/2019

5       System Setup and Method

5.1      Method:

Configuration as required by ANSI C63.4:2014 and ANSI C63.26



 No. Descriptions of EUT Exercising
  1 During the FCC Part 15B testing the transmitters were in idle mode.
  2 During the radiated spurious emissions testing the cellular transmitter was outputting on low, mid, or high
     channels at maximum output power in either band 4 or band 13.
  3 During the conducted emission testing the device was actively charging and the test was performed at the
     connection point to the AC mains.


                                                     Cables

 Qty.                Description                 Length (m)          Shielding         Ferrites     Termination
                                                                                                       AC/DC
    1                12VDC Cable                     2m                None              Yes           Power
                                                                                                      Adapter


5.2      EUT Block Diagram:




      Radio Base Station                        Test Sample                              AC Power Source
                                                                           DC
                              Cellular                                     Power
                              Link




Non-Specific EMC Report Shell Rev. December 2017                                                      Page 7 of 31
Report Number: 103714598LEX-006




                                                                           Evaluation For: eNeura, Inc.
                                                   Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                        Date: 2/1/2019


5.4   EUT Photo (Back):




Non-Specific EMC Report Shell Rev. December 2017                                           Page 9 of 31
Report Number: 103714598LEX-006


                                                                           Evaluation For: eNeura, Inc.
                                                   Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                        Date: 2/1/2019

5.5   EUT Photo (Power Adapter):




Non-Specific EMC Report Shell Rev. December 2017                                         Page 10 of 31
Report Number: 103714598LEX-006


                                                                                      Evaluation For: eNeura, Inc.
                                                              Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                   Date: 2/1/2019

6     Radiated Emissions

6.1    Method

Tests are performed in accordance with ANSI C63.4:2014 and ANSI C63.26

TEST SITE: 10m ALSE

Site Designation: 10m Chamber

Measurement Uncertainty

                                                                            Expanded Uncertainty
              Measurement                       Frequency Range                    (k=2)                 Ucispr
         Radiated Emissions, 10m                  30-1000 MHz                      3.9dB                 6.3 dB
         Radiated Emissions, 3m                   30-1000 MHz                      4.0dB                 6.3 dB
         Radiated Emissions, 3m                      1-6 GHz                       4.7dB                 5.2 dB
         Radiated Emissions, 3m                     6-15 GHz                       4.7dB                 5.5 dB
         Radiated Emissions, 3m                    15-18 GHz                       4.7dB                 5.5 dB
         Radiated Emissions, 3m                    18-40 GHz                       4.7dB                 5.5 dB

As shown in the table above our radiated emissions   U lab is less than the corresponding U CISPR reference value in
CISPR 16-4-2 Table 1, hence the compliance of the product is only based on the measured value, and no
measurement uncertainty correction is required.




Non-Specific EMC Report Shell Rev. December 2017                                                      Page 11 of 31
Report Number: 103714598LEX-006


                                                                                      Evaluation For: eNeura, Inc.
                                                              Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                   Date: 2/1/2019

6.2    Sample Calculation

The field strength is calculated by adding the Antenna Factor and Cable Factor, and subtracting the Amplifier Gain
(if any) from the measured reading. The basic equation with a sample calculation is as follows:

FS = RA + AF + CF - AG
Where              FS = Field Strength in dBV/m
                   RA = Receiver Amplitude (including preamplifier) in dBV
                   CF = Cable Attenuation Factor in dB
                   AF = Antenna Factor in dB
                   AG = Amplifier Gain in dB

In the following table(s), the reading shown on the data table reflects the preamplifier gain. An example for the
calculations in the following table is as follows.

Assume a receiver reading of 52.0 dBV is obtained. The antenna factor of 7.4 dB and cable factor of 1.6 dB is
added. The amplifier gain of 29 dB is subtracted, giving a field strength of 32 dBV/m. This value in dBV/m was
converted to its corresponding level in V/m.

RA = 52.0 dBV
AF = 7.4 dB/m
CF = 1.6 dB
AG = 29.0 dB
FS = 32 dBV/m

To convert from dBV to V or mV the following was used:

               where UF = Net Reading in V
       (NF / 20)
UF = 10
          NF = Net Reading in dBV

Example:

FS = RA + AF + CF – AG = 52.0 + 7.4 + 1.6 – 29.0 = 32.0
                      = 39.8 V/m
       (32 dBV / 20)
UF = 10



6.3    Field Strength to Power Calculation
As allowable by ANSI C63.26: 2015 section 5.2.7, the output power of unwanted emissions can be
calculated from a field strength measurement. The transmitter measurements that follow in this report
have applied the following calculation to the -13dBm limit to arrive an equivalent field strength limit at 3
meters as follows:

E (dBμV/m) = EIRP (dBm) − 20log(D) + 104.8; where D is the measurement distance (in the far
field region) in m.

Example:
 Limit (dBuV/m) = -13 -20log(3) + 104.8 = 82.25dBuV/m




Non-Specific EMC Report Shell Rev. December 2017                                                       Page 12 of 31
Report Number: 103714598LEX-006


                                                                                    Evaluation For: eNeura, Inc.
                                                            Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                 Date: 2/1/2019

6.4    Test Equipment Used:

         Description         Asset         Manufacturer                Model          Cal Date       Cal Due
      EMI Test Receiver      3900         Rohde & Schwarz              ESU40         9/18/2018      9/18/2019
        Bilog Antenna        7088              SunAR                     JB6         7/24/2018      7/24/2019
        Horn Antenna         3780           ETS Lindgren                3117         6/11/2018      6/11/2019
      System Controller      4096           ETS Lindgren                2090          Verify at      Verify at
                                                                                    Time of Use    Time of Use
      System Controller      3957          Sunol Sciences              SC99V          Verify at      Verify at
                                                                                    Time of Use    Time of Use
         3m Cable            3074                                                   11/29/2017     11/29/2018
    AntennaPreamp
         3m Cable            3918         Rohde & Schwarz             TS-PR18       11/29/2017     11/29/2018
       Preamplifier
         3m Cable            2588                                                   11/29/2017     11/29/2018
   PreampChamber
         3m Cable            2593                                                   11/29/2017     11/29/2018
 ChamberControl Room
         3m Cable            2592                                                   11/29/2017     11/29/2018
 Control RoomReceiver


6.5    Software Utilized:

                  Name                              Manufacturer                           Version
                  EMC32                            Rohde & Schwarz                      Version 9.15.02



6.6    Results:

The sample tested was found to Comply.




Non-Specific EMC Report Shell Rev. December 2017                                                   Page 13 of 31
Report Number: 103714598LEX-006


EB


     Db




          t=p—~>


                                      ruge <ovc—~—
enepurs ramarens averra rovusen vee




                                                                                                  Evaluation For: eNeura, Inc.
                                                                          Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                               Date: 2/1/2019

6.10 Plots/Data: Radiated Emissions, 30MHz – 1GHz (Transmitters Idle)




    Frequency             QuasiPeak               Limit          Margin          Bandwidth             Height        Pol     Azimuth
      (MHz)               (dBµV/m)              (dBµV/m)          (dB)             (kHz)                (cm)                  (deg)

        30.217000                   21.17                40.00        18.83            120.000            142.1   V               145.0
        43.657445                   17.79                40.00        22.21            120.000            102.7   V               348.0
        87.726556                   15.18                40.00        24.82            120.000             97.9   V                10.0
        96.395778                   28.50                43.52        15.02            120.000            103.2   V               312.0
       156.081667                   21.27                43.52        22.25            120.000            103.4   V                74.0
       191.994000                   31.55                43.52        11.97            120.000             98.1   V               295.0
       346.958889                   24.86                46.02        21.16            120.000            127.8   V                74.0
       471.821333                   26.68                46.02        19.34            120.000            222.8   V               266.0
       701.849222                   33.36                46.02        12.66            120.000            187.1   H               174.0
       959.941778                   36.66                46.02         9.36            120.000            103.0   V               137.0




                      Test Personnel:     Brian Lackey                                   Test Date:     11/21/2018
      Supervising/Reviewing Engineer:
                   (Where Applicable)     NA                                          Limit Applied:    Class B
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6                    Ambient Temperature:      23.2oC
                                          120VAC / 60Hz (Into Power
                         Input Voltage:   Adapter)                               Relative Humidity:     44.5%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                                 Atmospheric Pressure:     985.4mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                           Page 17 of 31
Report Number: 103714598LEX-006


                                                                                              Evaluation For: eNeura, Inc.
                                                                      Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                           Date: 2/1/2019

6.11 Plots/Data: Radiated Emissions, 1GHz – 18GHz (Transmitters Idle)




    Frequency            MaxPeak             Limit         Margin     Bandwidth       Height       Pol        Azimuth     Corr.
      (MHz)              (dBµV/m)          (dBµV/m)         (dB)        (kHz)          (cm)                    (deg)      (dB)
     1971.500000              37.21             74.00         36.79      1000.000        211.0    H               239.0      1.9
     2073.500000              38.12             74.00         35.88      1000.000        193.0    H               300.0      2.0
     2310.500000              38.39             74.00         35.61      1000.000        245.0    H               166.0      2.7
     3048.000000              39.43             74.00         34.57      1000.000        171.0    H               201.0      4.6
     4507.000000              42.62             74.00         31.38      1000.000        410.0    V               100.0      7.6
     5994.000000              43.88             74.00         30.12      1000.000        410.0    V               284.0      8.8


    Frequency             Average            Limit         Margin     Bandwidth       Height       Pol        Azimuth     Corr.
      (MHz)              (dBµV/m)          (dBµV/m)         (dB)        (kHz)          (cm)                    (deg)      (dB)
     1971.500000               24.09            54.00         29.91      1000.000        211.0    H               239.0      1.9
     2073.500000               24.56            54.00         29.44      1000.000        193.0    H               300.0      2.0
     2310.500000               24.87            54.00         29.13      1000.000        245.0    H               166.0      2.7
     3048.000000               26.28            54.00         27.72      1000.000        171.0    H               201.0      4.6
     4507.000000               29.49            54.00         24.51      1000.000        410.0    V               100.0      7.6
     5994.000000               30.74            54.00         23.26      1000.000        410.0    V               284.0      8.8


                      Test Personnel:     Brian Lackey                              Test Date:   11/21/2018
      Supervising/Reviewing Engineer:
                   (Where Applicable)     NA                                    Limit Applied:   Class B
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6              Ambient Temperature:     23.2oC
                                          120VAC / 60Hz (Into Power
                         Input Voltage:   Adapter)                          Relative Humidity:   44.5%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                           Atmospheric Pressure:    985.4mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                    Page 18 of 31
Report Number: 103714598LEX-006


                                                                                              Evaluation For: eNeura, Inc.
                                                                      Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                           Date: 2/1/2019

6.12 Radiated Spurious Emissions (LTE Band 4, Mid Channel)




  Frequency           QuasiPeak             Limit           Margin    Bandwidth       Height        Pol       Azimuth          Corr.
    (MHz)             (dBµV/m)            (dBµV/m)           (dB)       (kHz)          (cm)                    (deg)           (dB)
    59.323000              24.23              82.25           58.02      120.000        106.4      H             203.0          15.0
   110.680000              33.31              82.25           48.94      120.000        105.0      V             184.0          20.9
   197.780000              32.42              82.25           49.83      120.000        110.6      H             164.0          21.3
   296.700000              35.14              82.25           47.11      120.000        400.0      H             202.0          23.0
   494.540000              35.90              82.25           46.35      120.000        213.1      V             192.0          27.6
   534.080000              36.34              82.25           45.91      120.000        201.2      H             350.0          28.8
   593.420000              41.44              82.25           40.81      120.000         99.9      V             268.0          29.1
   642.830000              39.50              82.25           42.75      120.000        212.3      V              64.0          30.1
   741.770000              44.93              82.25           37.32      120.000        295.9      H             109.0          31.8
   933.280000              43.25              82.25           39.00      120.000        236.5      H               8.0          34.6


                      Test Personnel:     Michael Carlson                           Test Date:    12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                    Relative Humidity:    26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: The middle channel for the bilog range is representative of the low and high
channels as well.




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 19 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019

6.13 Radiated Spurious Emissions (LTE Band 4, Low Channel)




*Note: Carrier is suppressed with the use of a band reject filter.

    Frequency             Average            Limit          Margin     Bandwidth       Height       Pol       Azimuth          Corr.
      (MHz)              (dBµV/m)          (dBµV/m)          (dB)        (kHz)          (cm)                   (deg)           (dB)
     1224.600000               19.71            82.25          62.54      1000.000        216.0    V              232.0          -2.1
     3429.400000               27.64            82.25          54.61      1000.000        410.0    H              232.0           4.5
     5145.000000               30.38            82.25          51.87      1000.000        236.0    V              195.0           7.4
     6860.000000               35.51            82.25          46.74      1000.000        314.0    H              210.0          10.6
     8575.000000               32.17            82.25          50.08      1000.000        332.0    V              256.0          11.9
    11053.200000               35.66            82.25          46.59      1000.000        410.0    H                 0.0         15.9
    13720.000000               39.94            82.25          42.31      1000.000        287.0    V              226.0          18.8
    17999.400000               44.91            82.25          37.34      1000.000        220.0    H              338.0          26.6




                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 20 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019


6.14 Radiated Spurious Emissions (LTE Band 4, Middle Channel)




*Note: Carrier is suppressed with the use of a band reject filter.


    Frequency             Average            Limit          Margin     Bandwidth       Height       Pol       Azimuth          Corr.
      (MHz)              (dBµV/m)          (dBµV/m)          (dB)        (kHz)          (cm)                   (deg)           (dB)
     3466.000000               27.29            82.25          54.96      1000.000        410.0    H              234.0           4.8
     5197.500000               30.11            82.25          52.14      1000.000        379.0    V              240.0           7.7
     6819.200000               32.23            82.25          50.02      1000.000        410.0    H              252.0          10.5
     6930.000000               33.77            82.25          48.48      1000.000        410.0    V              254.0          10.7
     8662.500000               32.25            82.25          50.00      1000.000        221.0    V              250.0          11.8
     9975.600000               34.01            82.25          48.24      1000.000        410.0    V              202.0          14.0
    17350.800000               43.92            82.25          38.33      1000.000        243.0    V              286.0          24.8



                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 21 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019

6.15 Radiated Spurious Emissions (LTE Band 4, High Channel)




*Note: Carrier is suppressed with the use of a band reject filter.

    Frequency             Average            Limit          Margin     Bandwidth       Height       Pol       Azimuth          Corr.
      (MHz)              (dBµV/m)          (dBµV/m)          (dB)        (kHz)          (cm)                   (deg)           (dB)
     1247.300000               19.82            82.25          62.43      1000.000        250.0    H              169.0          -1.8
     3500.200000               30.96            82.25          51.29      1000.000        241.0    H              192.0           5.1
     5250.800000               32.46            82.25          49.79      1000.000        227.0    H              193.0           7.7
     7000.000000               38.93            82.25          43.32      1000.000        299.0    H              226.0          11.1
     8750.000000               32.45            82.25          49.80      1000.000        236.0    H              157.0          12.2
    10500.000000               34.50            82.25          47.75      1000.000        410.0    H                 0.0         14.8
    17983.800000               44.64            82.25          37.61      1000.000        410.0    V              231.0          26.3




                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 22 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019

6.16 Radiated Spurious Emissions (LTE Band 13, Mid Channel)




*The large peak at approximately 780MHz is the fundamental transmission and does not indicate a failing result.


    Frequency           QuasiPeak            Limit          Margin     Bandwidth        Height       Pol       Azimuth          Corr.
      (MHz)             (dBµV/m)           (dBµV/m)          (dB)        (kHz)           (cm)                   (deg)           (dB)
       30.735000              38.58             82.25          43.67       120.000         100.1    H              148.0          28.2
       34.613000              36.11             82.25          46.14       120.000         141.7    H              194.0          25.2
       59.358000              26.26             82.25          55.99       120.000         343.6    H               64.0          15.1
      110.600000              30.21             82.25          52.04       120.000          99.9    V              340.0          20.9
      197.820000              34.14             82.25          48.11       120.000         107.9    H              313.0          21.3
      296.700000              34.46             82.25          47.79       120.000         165.5    V              210.0          22.7
      494.520000              37.78             82.25          44.47       120.000         307.8    V               92.0          27.6
      593.390000              44.10             82.25          38.15       120.000          99.9    V               18.0          29.1
      752.450000              48.91             82.25          33.34       120.000         141.7    V              111.0          31.4


                      Test Personnel:     Michael Carlson                            Test Date:    12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                              -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:    Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:      21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:    26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:     978.7mbar




Deviations, Additions, or Exclusions: The middle channel for the bilog range is representative of the low and high
channels as well.




Non-Specific EMC Report Shell Rev. December 2017                                                                      Page 23 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019

6.17 Radiated Spurious Emissions (LTE Band 13, Low Channel)




    Frequency             Average            Limit          Margin     Bandwidth       Height       Pol       Azimuth          Corr.
      (MHz)              (dBµV/m)          (dBµV/m)          (dB)        (kHz)          (cm)                   (deg)           (dB)
     1561.600000               40.61            82.25          41.64      1000.000        208.0    V              238.0          -2.2
     2340.800000               32.55            82.25          49.70      1000.000        410.0    V              233.0           2.6
     3123.000000               26.42            82.25          55.83      1000.000        344.0    V              225.0           4.5




                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 24 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019


6.18 Radiated Spurious Emissions (LTE Band 13, Middle Channel)




  Frequency              Average            Limit           Margin    Bandwidth       Height       Pol        Azimuth          Corr.
    (MHz)               (dBµV/m)          (dBµV/m)           (dB)       (kHz)          (cm)                    (deg)           (dB)
  1564.200000               39.63             82.25           42.62     1000.000        100.0      V             241.0           -2.1
  2349.800000               25.06             82.25           57.19     1000.000        348.0      H             204.0            2.6
  7125.200000               32.90             82.25           49.35     1000.000        410.0      V             272.0          11.0
 17660.400000               44.51             82.25           37.74     1000.000        379.0      H             146.0          25.1


                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 25 of 31
Report Number: 103714598LEX-006


                                                                                               Evaluation For: eNeura, Inc.
                                                                       Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                            Date: 2/1/2019

6.19 Radiated Spurious Emissions (LTE Band 13, High Channel)




    Frequency             Average            Limit          Margin     Bandwidth       Height       Pol       Azimuth          Corr.
      (MHz)              (dBµV/m)          (dBµV/m)          (dB)        (kHz)          (cm)                   (deg)           (dB)
     1562.600000               38.78            82.25          43.47      1000.000        410.0    V              327.0          -2.1
     2345.800000               30.20            82.25          52.05      1000.000        410.0    V              236.0           2.7
     3128.400000               26.08            82.25          56.17      1000.000        285.0    H              164.0           4.5




                      Test Personnel:     Michael Carlson                            Test Date:   12/3/2018 – 12/4/2018
      Supervising/Reviewing Engineer:                                                             -13dBm (Converted to Field
                   (Where Applicable)     NA                                     Limit Applied:   Strength at 3m)
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6               Ambient Temperature:     21.9oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                     Relative Humidity:   26.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                            Atmospheric Pressure:    978.7mbar




Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                     Page 26 of 31
Report Number: 103714598LEX-006


                                                                                       Evaluation For: eNeura, Inc.
                                                               Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                    Date: 2/1/2019


7     Conducted Emissions

7.1    Method

Tests are performed in accordance with ANSI C63.4:2014.

TEST SITE: Ground Plane

Site Designation: Ground Plane

Measurement Uncertainty

                                                                            Expanded Uncertainty
              Measurement                           Frequency Range                (k=2)                 Ucispr
       AC Line Conducted Emissions                  150 kHz - 30 MHz               3.1dB                 3.4dB

As shown in the table above our conducted emissions     U lab is less than the corresponding U CISPR reference value
in CISPR 16-4-2 Table 1, hence the compliance of the product is only based on the measured value, and no
measurement uncertainty correction is required.


7.2    Sample Calculations

The following is how net line-conducted readings were determined:

NF = RF + LF + CF + AF
Where NF = Net Reading in dBV
         RF = Reading from receiver in dBV
         LF = LISN or ISN Correction Factor in dB
         CF = Cable Correction Factor in dB
         AF = Attenuator Loss Factor in dB

To convert from dBV to V or mV the following was used:

               where UF = Net Reading in V
       (NF / 20)
UF = 10
          NF = Net Reading in dBV

Example:

NF = RF + LF + CF + AF = 28.5 + 0.2 + 0.4 + 20.0 = 49.1 dBV
                        = 285.1 V/m
       (49.1 dBV / 20)
UF = 10




Non-Specific EMC Report Shell Rev. December 2017                                                       Page 27 of 31
Report Number: 103714598LEX-006


                                                                                 Evaluation For: eNeura, Inc.
                                                         Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                              Date: 2/1/2019

7.3    Test Equipment Used:

        Description          Asset          Manufacturer           Model           Cal Date        Cal Due
      EMI Test Receiver      3900         Rohde & Schwarz           ESU40         9/18/2018       9/18/2019
            LISN             2509          Fischer Custom        FCC-LISN-50-     4/10/2018       4/10/2019
                                           Communication            50-2M
  Coaxial Cable (COND 3)     6026                                                11/26/2018       11/26/2019


7.4    Software Utilized:

                  Name                         Manufacturer                           Version
                   TILE                        ETS Lindgren                          V7.0.6.545



7.5    Results:

The sample tested was found to Comply.




Non-Specific EMC Report Shell Rev. December 2017                                                  Page 28 of 31
Report Number: 103714598LEX-006


                                                                                              Evaluation For: eNeura, Inc.
                                                                      Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                           Date: 2/1/2019


7.6    Plots/Data: Conducted Emissions (Line)




 Frequency          Quasi-Peak            Quasi-Peak         Quasi-Peak      Average                Average       Average
 (MHz)              (dBuV)                Limit              Margin (dB)     (dBuV)                 Limit         Margin (dB)
                                          (dBuV)                                                    (dBuV)
 0.159              50.038                65.743             15.705          23.490                 55.743        32.253
 0.285              42.142                62.143             20.001          26.294                 52.143        25.848
 0.366              32.994                59.829             26.834          14.174                 49.829        35.655
 0.474              25.328                56.743             31.415          13.754                 46.743        32.989
 0.587              27.678                56.000             28.322          14.032                 46.000        31.968
 1.878              26.586                56.000             29.414          16.360                 46.000        29.640
 2.711              28.441                56.000             27.559          18.225                 46.000        27.775
 3.363              26.977                56.000             29.023          16.128                 46.000        29.872
 10.124             32.493                60.000             27.507          26.892                 50.000        23.108
 13.184             29.315                60.000             30.685          20.947                 50.000        29.053



                      Test Personnel:     Bryan Taylor                                Test Date:     11/27/2018
      Supervising/Reviewing Engineer:
                   (Where Applicable)     NA                                       Limit Applied:    Class B
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6                 Ambient Temperature:      22.7oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                      Relative Humidity:     42.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                              Atmospheric Pressure:     989.2mbar


Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                   Page 29 of 31
Report Number: 103714598LEX-006


                                                                                              Evaluation For: eNeura, Inc.
                                                                      Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                                           Date: 2/1/2019

7.7    Plots/Data: Conducted Emissions (Neutral)




 Frequency          Quasi-Peak            Quasi-Peak         Quasi-Peak      Average                Average       Average
 (MHz)              (dBuV)                Limit              Margin (dB)     (dBuV)                 Limit         Margin (dB)
                                          (dBuV)                                                    (dBuV)
 0.159              50.038                65.743             15.705          23.490                 55.743        32.253
 0.285              42.142                62.143             20.001          26.294                 52.143        25.848
 0.366              32.994                59.829             26.834          14.174                 49.829        35.655
 0.474              25.328                56.743             31.415          13.754                 46.743        32.989
 0.587              27.678                56.000             28.322          14.032                 46.000        31.968
 1.878              26.586                56.000             29.414          16.360                 46.000        29.640
 2.711              28.441                56.000             27.559          18.225                 46.000        27.775
 3.363              26.977                56.000             29.023          16.128                 46.000        29.872
 10.124             32.493                60.000             27.507          26.892                 50.000        23.108
 13.184             29.315                60.000             30.685          20.947                 50.000        29.053


                      Test Personnel:     Bryan Taylor                                Test Date:     11/27/2018
      Supervising/Reviewing Engineer:
                   (Where Applicable)     NA                                       Limit Applied:    Class B
                                          FCC Part 15B
                    Product Standard:     ICES-003 Issue 6                 Ambient Temperature:      22.7oC
                                          120VAC / 60Hz (Into AC
                         Input Voltage:   Power Adapter)                      Relative Humidity:     42.3%
      Pretest Verification w / Ambient
                 Signals or BB Source:    Yes                              Atmospheric Pressure:     989.2mbar


Deviations, Additions, or Exclusions: None




Non-Specific EMC Report Shell Rev. December 2017                                                                   Page 30 of 31
Report Number: 103714598LEX-006


                                                                             Evaluation For: eNeura, Inc.
                                                     Product: sTMS Mini Transcranial Magnetic Stimulator
EMC Test Report                                                                          Date: 2/1/2019

8     Revision History

    Revision       Date         Report Number      Prepared      Reviewed               Notes
     Level                                            By            By
       0          2/1/2019     103714598LEX-006                                      Original Issue




Non-Specific EMC Report Shell Rev. December 2017                                            Page 31 of 31
Report Number: 103714598LEX-006



Document Created: 2019-02-01 15:56:12
Document Modified: 2019-02-01 15:56:12

© 2024 FCC.report