I18Z61354-SEM01_SAR_Rev2_4

FCC ID: 2AQRMFXMC52401

RF Exposure Info

Download: PDF
FCCID_4070310

e>
7TL                                                                                             P
e




 Appendix (Additional assessments outside the scope of SCS 0108)
 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                            54.4 2 10. jn
       Retum Loss                                                                        +275 08


 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                            4030 —3.4 n
       Retum Loss                                                                        ~20.1 4B



 General Antenna Parameters and Design

     Electrieal Delay (one direction)                                                     1.033 ns
 After long term use with 100W racieted power, only a sight warming of the dipole near the feedpoint can be meesured.

 The dipole is made of standard sermirigld coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—cirouited for DC—signals. On some ofthe dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions‘ paragragh. The SAR data are not affected by this change. The overall dipole length is stil
 according to the Standard.
 No excessive foree must be applied to the dipole arms, because they might bend or the soldered connections nearthe
 feedpoint may be damaged.

 Additional EUT Data

       Manutactured by                                                                    spaae
       Manutactured on                                                                 March 22, 2010




 Certficate No: D7S0V3—1017_Junt7                          Page 4 of 8


®
DASY5 Validation Report for Head TSL



Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 750 MHz; Type: D75OV3; Serial: D750V3 — SN:1017
                                                                                      Date: 19.07.2017




Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: £=750 MHz; o = 0.89 S/m; 6; = 41; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/EC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(10.49, 10.49, 10.49); Calibrated: 31.05.2017;

        Sensor—Surface: 1.4mm (Mechanical Surace Detection)

        Electronics: DAE4 Sn601; Calibrated: 28.03.2017

        Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

        DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.85 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.22 W/kg
SAR(L g) =2.09 W/kg; SAR(1O g) = 1.36 W/ke
Maximum valueof SAR (measured) =2.83 W/kz




        —6.60

        —8.00


        —17.00
                   0 dB =2.83 Wikg = 4.52 dBWikg




Certficate No: D7S0V3—1017_.Jult7              Page 5 of 8


omm                                                                          I
 TTL                                                                                  F
mm                                                                                          C



 Impedance Measurement Plot for Head TSL



                                                             18 Jul zeir   estarise
          ERD sa      i uis         asessse   esiore s800 pi               7s0.000 ooo ns
                                              m

          per




          e
          t°


          cne




          hy
          159




                    Sthet sse.00 ooo ns                           stoe ssa.00 ace ie




 Certficate No: D7503—1017_Jult7               Page 6 of 8


®
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017
                                                                                      Date: 19.07.2017




Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: f=750 MHz; a = 0.99 $/m; &; = 55; p = 1000 ke/m?
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEEAMEC/ANSI C63.19—2011)
DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvR(10.35, 10.35, 10.35); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (Back); Type: QD 0OR P49 AA; Serial: 1005

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.67 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 3.34 W/kg
SAR(I g) =2.22 W/kg; SAR(1O g) =
Maximum valueof SAR (measured) = 2.




        —6.60



        —11.00
                 0 dB =2.96 Wikg=4.71 dBWikg




Cortficate No: D7S03—1017_.u17                 Page 7 of 8


@7TL                                                                              I



 Impedance Measurement Plot for Body TSL



                                                             19 on zeir cessorne
          EXD sa      i ucss        maszsso —s«ntse 62 s or             7se.c00 oc htz
                                            2oues

          ber




          s
          ifl.




           cne



          ca


          hy
          Eo
          ma
                    Sthet sse.00n ooons                           stoe ssa.00 ece nne




 Certficate No: D7S0V3—1017_Jult7              Page 8 of 8


Calibration Laboratory of                                  s          >         HEWI          g. Scteizerischer Kallorierdienst
Schmid & Partner                                         ipewck              ‘"amn"          i5 Service suisse o‘#tionnage
  Engineering AG                                          Lrecs                                   Sarvislo aviezoro d taratura
Zeughausstrasse 43, 8004 Zurich, Switzeriand              ky                 Nee $)          S swins Catioration Service

Accredted by he Swiss Accrodiaion Sorvice (SAS)                                               Accreditation No: SCS 0108
The Swiss Accreditation Service is one of the signatories to he EA
MullLateral Agreoment fr the recognition of callbration certficates
Ctiem       CTTL—BJ (Auden)                                                        Certticate N: DB35V2—40069_Jul17

CALIBRATION CERTIFICATE
 Object                           DB35V2 — SN:4d069

 Cabration procedure(s)           QA CAL—05.19
                                  Calibration procedure for dipole validation kits above700 MHz



 Calbration date:                 July 19, 2017


 This cairaton certicate dacumentsthe aceablity to natlona standards,which realze the physical untso measurements 81.
 "The measurements and the uncertainieswih confddence probabity are given on thefolowing pages and are par ofthe cortfat

 Allcalbrations have beon conducted in the clsed Iaboratary fas: anvironmant tempralure(22 3)°C and hunidity< 70%.

 Calbration Equipment used (M&TE crtealfr calbraton)
 Primary Stendards                 n«                     Cal Date (Cortficate No)                     Scheduled Caltation
 Power mater NRP                   sit rowrre             dtpt7 (No. 217—cosvorse0)                    Apere
 Power sensor NRP—291              shz rosaes             otApe7 (no. 2tr—casen)                       Apere
 Poer sensor NRP—Z91               on: roams              oeApe17 (No. 2t7—0000)                       Apers
 Reterence 20 dB Attnator          sh: sose (200)         or—Apr—17 (No. 2t7—00500)                    Aoere
 Type—N mismatch combination       |SN: 0472 /O68e7       Or—Apet7 (No.217—02520)                      Apere
 Raterence Probe EOVA              snt rero               3tMay—17 (No. EXG—7340_Mayi7)                Mayts
 oaee                              s on                   26Mar—t7 (No. DAE4—601_Mart?)                Marte
 Secondary Standards               n«                     Check Date in house)                         Scheduled Chack
 Power mater EP42A                 SN: GBardeoror         0r0Oct15 in house check Oct—10)              in house chack: Oct—18
 Pover sensor HP 8481A             SN: Us8rz92708         0r—Octd$ (inhousecheck Oct16)                In house chack: Oct—18
 Power sensor HP 8101A             SN: MFétogzat?         070ett$ (nhousecheckOct18)                   in house chack: Oot—18
 AF generater nSMT—O               sn: roowre             16—Jun—15 t house chock Oct—16)              in house chack: Oct—10
 Notwork AnabyzorHP 8753E          sn: ussrasoses         1€:09.01 (n house check Oct—18)              in house chack: Oct—17
                                   Name                               Function                          Signature
 Galtrated by:                     Johannes Kurida                    Laboratoy Techrician
                                                                                                      ‘;
 Approved by:                      Katia Pokovie                      Tochrical Manager              /Z/ Zf,

                                                                                                       issuect Juty 20, 2017
 This caibration certicate shal not b reproduced exoeot n fulwthout witen approva f he aboraton

Cortficate No: DB35V2—44060_Junt7                              Page 1 of 8


®
Calibration
     4      Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzeriand

Accredted by the Swiss Accredtation Sorvice (SAS)
                                                                            s   Sehweizerischer Kallbrierdienst
                                                                            q Service suisse i‘étatonnage
                                                                              Servitio svizzoro di taratura
                                                                            5. swiss Caltbration Service

                                                                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the aignatories to the EA
Multlateral Agroomantfor the recogn‘tion ofcallration certfieatos
Glossary:
TSL                        tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x.y,z
NA                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
         300 MHz to 6 GHz)®, July 2016
    c  IEC 62209—2, "Procedureto determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)®, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
     *   SAR measured: SAR measured at the stated antenna input power.
     *   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
     *   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95% .




Certficate No: DB3V2—40060_Jut7                               Page 2 of 8


Measurement Conditions
   DASY system confiquration, as far as not given on page 1.
    DASY Version                                             Dasys                              vse.100
    Extrapolation                                    Advanced Extrapolation
    Phantom                                           Medular Fiat Phantom
    Distance Dipole Gnter — TSL                              15 mm                             with Spacer
     Zoom Sean Resolution                              t dy, dz ~5 mm
     Frequency                                         805 MHza 1 MHz



Head TSL parameters
   The following paramaters and calculations were applied.
                                                               Temperature      Permittivity        Conduetivity
     Nominal Head TSL parameters                                 220°0             415               0.90 mhoim
     Measured Head TSL parameters                             @20 02)°C         40826 %           0.91 mhoim6%
     Head TSL temperature change during test                    <o5°c                                   m
SAR result with Head TSL

     SAR averaged over 1 om(1 g) of Head TSL                    Conditin
     SAR measured                                        250 mW input power                297 Who
     SAR for nominal Head TSL parameters                  normalized to 1W          9.37 Whg = 17.0 % (ke2)

     SAR averaged over 10 em(10 g) of Head TSL                  condition
     SAR measured                                        250 mW input power                158 Wha
     SAR for nominal Head TSL paramsters                  normalized to 1W          6.06 Wikg =16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were apolied.
                                                               Temperature      Permitivity         Conductivity
     Nominal Body TSL parameters                                  22.0°C            562              0.87 mhoim
     Measured Body TSL parameters                              gao02)°C         54626%            1.01 mhoim6%
     Body TSL temperature change during test                     <05°C            ~—
SAR result with Body TSL

     SAR averaged over 1 om? (1 g) of Body TSL                  Gondition
     SAR measured                                        250 mW input power                 243 Whg
     SAR for nominal Body TSL parameters                     normalized to 1W        8.41 Whkg 2 17.0 % (k=2)
     SAR averaged over 10 cm? (10 g) of Body TSL                condition
     SAR measured                                        250 mW input power                    157 whg
     SAR for nominal Body TSL parametars                     normalized to 1W        6.12 Wikg 2 16.5 % (k=2)



Certficate No: DB35V2—40060_Jut17                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          sein—12j0
      Retum Loss                                                                      ~$2.4 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          479n—3.0 n
      Retum Loss                                                                      —25.9 0B


General Antenna Parameters and Design

    [Electrical Detay (one direction)                       |                          1302 ns
Alfter long term use with 100W radiated power, only a sight warming of the dipole near the feedpoint can be measured

The dipole is made of standard sermiigid coaxial cable. The center conductor of the feeding line is direcy connected to the
second arm ofthe dipole. The antenna is therefore shor:—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching whon loaded according to the position as explained in the
‘Measurement Condiions* paragraph. The SAR data are not affected by this change. The overall dipole length is stll
according to the Standard.
No excessive force must be applled to the dipole arms, because they might bnd or the soldered connections noar the
feedpoint may be damaged.

Additional EUT Data

      Menutactured by                                                                  sraag
      Menufactured on                                                             November 09, 2007




Certficate No: DB3SV2—44069_Ju17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                      Date: 19.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:40069

Communication System: UTD 0 — CW; Frequency: 835 MHz
Medium parameters used: (=835 MHz: a=0.91 S/m; &= 40.8; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEEAIEC/ANST C63.19—201 1

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=3mm, dy=5mm, dz=5mm
Reference Value = 62.08 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 3.65 W/kg
SAR(L g) =2.37 W/kgs SAR(IO g) = 1.53 Wikg
Maximum value of SAR (measured) = 3.21 Wke


        dB
         0

        —2.20

        ~4.40

        —6.60

        —80.00


        +11.00
                  0 dB =3.21 Wikg = 5.07 dBWike




Certficate No: De3SV2—40060_Jut7               Page 5 of 8


Impedance Measurement Plot for Head TSL



                                                        19 w 2eir. coiseres
               san     i uis           u samse rirsee d6241pr      #35.000 ac0 nz
                                             *       *




                     Sthet cas.o00 one nine                     Sto 1 aa5.000 s00 mz




Cortficate No: DB3SV2—44069.Jult7                 Page 6 of 8


(
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                   Date: 19.07,2017



DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:44069

Communication System: UID 0 — CW: Frequency: 835 MHz
Medium parameters used: {= 835 MHz; a = 1.01 S/m; e; = 54.8; p = 1000 kg/m
Phantom section: Flat Section
Measurement Standard: DASY5 (TEEE/AIEC/ANST C63,19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.2, 10.2, 10.2); Calibrated: 31.05.2017;

   +    Sensor—Surface: 1.4mm(Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (Back); Type: QD 0OR P49 AA; Scrial: 1005

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW , d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=3mm, dy=Smm, d=5mm
Reference Value = 59.35 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 3.67 Wikg
SAR(L g) =2.43 W/kg; SAR(IO g) =      .57 Wike
Maximum value of SAR (measured)        .22 whke

        dB
         0

        2.20

        ~4.40




        —11.00
                   0 dB =3.22 Wike= 5.08 dBWike




Cortficate No: DB3SV2—44060_.Jult?                Page 7 of 8


           *                                                             t
= TTL                                                                           F
mm                                                                                      ~



 Impedance Measurement Plot for Body TSL



                                                        18 aar ceir osierse
           Exn sa      i u‘rs         mamosse —assz5n «0sts pF      a35.000 o00 nuz
                                                *

           ve1                                          +7~.3
           c


           PS
           is                           \                 s



           cue



           ca




           ma
                     Stakt saz.o00 ooo nez                      sroe 1 eas.c00 s00 mz




  Certficate No: DB35V2—44060_Jult7               Page 8 of 8


Calibration Laboratory of                                   «9
                                                            sn                                         Schwoizeriachor Kallorierdienst




                                                                                               © o ®
                                                     )€E
Schmid & Partner




                                                         »,
                                                                                                       Service suisse «‘étalonnage
  Engineering AG                                                                                       Serviclo sviezoro dl taratura




                                                        o
Zeughausstrasse 43, 8004 Zurich, Suitzeriand
                                                            ihien
                                                               haad®                                   Swiss Calibration Service

Accredtad by the Siss Accreditaion Sorvice (GAS)                                                 Acereditation N   : SCs 0108
The Swiss Accreditation Service is one ofthe signatories tothe EA
MultLateral Agreemontfor the recognion ot callration corificates
Client      CTTL—BJ (Auden)                                                          Cerificate No: D1750V2—1003_Jul17

|CALIBRATION CERTIFICATE
 Object                           D1750V2 — SN:1008

 Calbration procedure(e)          QA CAL—O5.19
                                  Calibration procedure for dipole validation kits above 700 MHz



 Calbration date:                 July 21, 2017

 Thiscaibvaton cerficate documents th traceabityto natlonalstandards, which reaize the physical ts of measurements(S1.
 The measremonts and the uncrtantes wih confdence probabily are gven on the folowing pages and are pa of he corifeate.

 Allcalbrations have beanconducted in the clased aboratar faity: anvronment temperature(223)°C and hunidity <70%.
 Caltration Equipment used (MATE crtealfr catraton)
 Pimary Standards                  CB                      Gal Date (Conticate No                           Soeduled Caltration
 Poner meter NBP                   sn: rorrre              ocAp—17 (No. 217—cesevrsen)                      Aocie
 Power sensor NRP—281              sht tosoes              oeAper? (No. atz—caset)                          Apers
 Power sersor NRP—201              sn: rooms               dcApe—7 (No. 217.00520)                          Apers
 Raference 20 0B Attrator          sh: sose (209           0r—apet7 (No.217—00820)                          Aprrs
 Type—N mismatch combination       SN:5047.2/00027         Or—Ap—t7 (No.217—02509)                          Apers
 Reterence Probe EX3DV4            sn: ros                 3t—May—17 (No. EX3—7340Mayt?)                    May—1s
 page                              Sz or                   26Mar—17 (No. DAE+—001_Mart?)                    Maris
 Secondary Standards                e                      Check Date (n house)                             Soheduled Check
 Poer meter EPM—442A                SN: GBsrd8o70          0rOct18 (nhousecheck Oct18)                      in house check Oct18
 Power sensor HP 8481A              Sh: Ussrzeeres         070ct18 (n housecheck Oct18)                     In house check: Oct18
 Power sensor HP 8481A              SN: ivatoneat?         070ct18 (nhousecheck Oct16)                      In house cheok: Oct18
 AF generater nSMT—0                sht tooore             15—jun—15 (n house check Oct16)                  In house checic Oct18
 Network Anatyzer HP 8758E          sn ussrasoses          1EOct0f (inhousecheckOct18)                      In house chedc Oot17
                                    Name                               Function                             Signature
 Calbrated by:                     Michaot Wobr                        Laboretoy Technician                 1&.


 Approved by:                      Katia Pokovie                       Techrical Manager
                                                                                                        Ee
                                                                                                           Issuedt Juty 24, 2017
 Thiscaibrationcorficale shallno be repreduced excopt in ful wthout witten approval of thelaboratoy.

Certficate No: D1750¥2—1008_Jult?                              Page 1 of 8


Calibration Laboratory of                                             s   Schwelzerischer Kallrierdionst
Schmid & Partner                                                      c   Service sulsso diétalonnage
  Engineering AG                                                          Sorvizl avizzero di taratura
Zoughausstrasse 43, 0004 Zurich, Switzorland                          5. swiss Calibration Sorvice

Accredted by the Swis Accreditation Sorvice (GAG)                      Accreditation No.: SCS 0108
Te Swiss Accreditaion Service is one ofthe signatories tothe EA
Multlateral Agreementfor the recognition ofcaliration corteatos
Glossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORM xy,z
NA                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, IEEE Recommended Practice for Determining the Peak Spatial—
          Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques®, June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)®, July 2016
    ©) IEC 62209—2, "Procedure to determinethe Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)*, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in thecertificate are valid at thefrequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connectorand the antenna feedpoint.
          No uncertainty required.
    *     SAR measured: SAR measured at the stated antennainput power.
    *     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurementis stated as thestandard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certficate No: D1750¥2—1003_Jult7                      Page 2 of 8


7TL



 Measurement Conditions
    DASY system contiquration, as far as not given on page 1.
     DASY Version                                             Dasys                              vse.100
     Extrapolation                                    Advanced Extrapolation
     Phantom                                           Modular Fiat Phantom
     Distance Dipole Genter — TSL                             10 mm                             with Spacer
      Zoom Scan Resolution                              dx, dy, de ~$ mm
      Frequency                                         1750 Mz a 1 Mz

 Head TSL parameters
    The following paramaters and calculations were apolied.
                                                                Temperature      Permitivity         Condustivity
      Nominal Head TSL parameters                                  22.0°C           40.1              1.37 mhoim
      Measured Head TSL parameters                             r2002)°0          80.0£6%           138 mhoim6%
      Head TSL temperature change during test                     <as°c                                    ~—

 SAR result with Head TSL

      SAR averaged over 1 em(1 g) of Head TSL                    Condition
      SAR measured                                       250 mW input power                9.15 Whg
      SAR for nominal Head TSL paramsters                 normalized to 1W           36.7 Wha 17.0 (e=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                condition
      SAR measured                                       250 mW input power                4.84 Who
      SAF for nominal Head TSL parameters                 normalized to 1W           19.4 Whg + 16.5 % (k=2)

 Body TSL parameters
    The following paramaters and calculations were applied.
                                                                Temperature      Permittivity        Conductivity
      Nominal Body TSL parameters                                  20°C             5es               1.49 mho/m
      Measured Body TSL parameters                              g2o02)°C         §3.826%           149 mhoim«6%
      Body TSL temperature change during test                     <os*c                                  —
 SAR result with Body TSL

      SAR averaged over 1 om(1 g) of Body TSL                    Condition
      SAR measured                                        250 mW input power                0.29 Whg
      SAR for nominal Body TL parameters                      normalized to 1W       97.1 Wikg £17.0 % (k=2)
      SAR averaged over10 om* (10 g) of Body TSL               condiion
      SAR measured                                        250 mW inout power                4.94 Wihg
      SAR for nominal Body TSL parameters                  normalized to 1W           19.8 Wg # 16.5 % (k=2)


 Cortficate No: D1750V2—1003_Jult7                     Page 3 of 8


                                                                                              Eol



Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           soon+ 110
      Retum Loss                                                                       —37.1 a8

Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           «r0n+0.1jo
      Retum Loss                                                                       +202 08


General Antenna Parameters and Design

    Electreal Delay (one direction)                          |                          1213 ns
After long term use with 100W radiated power, only a sight warming of the diole near the feedpoint can be measured.

The dipole is made of standard sermirigid coaxial cable. The center conductor of the feedingline is directly connected to the
second arm ofthe dipole. The antenna is therefore short—ciroulted for DC—signals. On some of the dipoles, small end caps
are addeto the dipole arms in order to improve matching when loaded according to the positon as explained in the
‘Measurement Condiions* paragraph. The SAR data are not affected by this change. The overall dipole length is stll
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be demaged.

Additional EUT Data

      Manulactured by                                                                    srero
      Manufactured on                                                                 July 30, 2008




Certficate No: D1750V2—1008_utt7                          Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                  Date: 21.07.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used:   f= 1750 MHz; 0 = 1.35 S/m; e, = 39; p = 1000 kg/m?
Phantomsection: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:
   *   Probe: EX3DV4 — SN7349; ConvB(8.73, 8.73, 8.73); Calibrated: 31.05.2017;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAE4 $n601; Calibrated: 28.03.2017

   *   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   *   DASY52.52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.4 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 17.0 Wikg
SAR(T g) = 9.15 W/kg; SAR(1O g) = 4.84 W/kg
Maximum value of SAR (measured) = 14.0 Wke

       dB
        0
       —3.60
       ~7.20
       —10.00
       +14.40

       —18.00
                  0 dB = 14.0 Wikg= 1146 dBW/kg




Certficate No: D1750V2—1008_Jult7              Page 5 of 8


          *                                                                     t
= TTL                                                                                  F
mm                                                                                             ~



  Impeciance MeasurementPlot for Head TSL



                                                                 m aur cerr snsomiar
           gHp sa     a us            is se.oos o     10566 o 96,097 ph     1 7se.00e ccemtz
                                                    ffh—.

           ver




           ma
           cne



           ca




           ma
                    Stamt 1 ssa.000 aoo mie                            stoe 1 ssn.000 ooo mz




  Cortficate No: D1750V2—1008.u17                       Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 20.07.2017
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003
Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: £=1750 MHz; 0 = 1.49 8/m; 6: = 53.3; p = 1000 kg/m®
Phantom section: Flat Section
Measurement Standard: DASY5 (TEEE/TEC/ANSI C63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(8.46, 8.46, 8.46); Calibrated: 31.05.2017;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronies: DAE4 $n601; Calibrated: 28.03.2017

   *   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   +   DASY52.52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=3mm, dy=5mm, dz=5mm
Reference Value = 99.34 V/m; Power Drift —0.02 dB
Peak SAR (extrapolated) = 16.4 W/kg
SAR(L g) =9.29 W/kg: SAR(1O g) =4.94 Whq
Maximum value of SAR (measured)   13.4 W/kg




       10.20

       +13.60


       <17.00
                  0 dB = 134 Wikg= 11.27 dBW/kg




Certficate No: D1750¥2—1003_Jult?             Page 7 of 8


Impedance Measurement Plot for Body TSL



                                                          20 aar nesr| cerarior
         EBp sa     i us           in s6.9n« a e0r09 0. snn pi         1 rse.000 se mz
                                              t

         ner
         ca


         hy
         ass


         ma
         cne




         me
                  Sthe1 sea.000 ooo mis                         Stoe 1 s5e.000 o0 mmz




Gertficate No: D1750V2—1008_ult7                Page 8 of 8


..>                                                                                              e
@7 TL                                                                                                           o
tw..corf




  Calibration Laboratory of                                 «9n                                       Schwelzerischer Kallbrerdienst
  Schmid & Partner                                          ii%                                       Service sulase détalonnage
    Engineering AG                                                                                    Serviclo svizzero ditarature
  Zeughausstrasse 49, 8004 Zurich, Suitzerland                                                        Swiss Calibration Service

  Accredtea by he Sviss Accroditaion Service (GAS)                                                           ion No: SCS 0108
  The Swiss Accreditation Service is one of the signatoriesto he EA
  Mulltateral Agreement for the recognition of calloration certficates
  Cliont      CTTL—BJ (Auden)                                                          Certficate No: D1900V2—5d101_Jul17

  CALIBRATION CERTIFICATE
   Object                            D1900V2 — SN:sd101


   Calibration procedure(s)         QA CAL—05.9
                                     Calibration procedure for dipole validation kits above 700 MHz



   Calbration date:                  July 26, 2017


   This caibratoncerficate documents the raceabilyto ntional standards,vich reaizethe hysical uns of measurements (S1.
   "The measurements and the uncertaintes wih confdence probabiity are given on the folowing pages and are part ofthe cortcale

   Allealbrations have beenconducted in the clased laboratay facity: envrorment temperatire (22 + 3)°C and humiity < 700.
   Callbrtion Equipment used (METE ertealfor calbation)
   Primary Standards                  n#                     Cal Date (Gortlcate No)                        Scheduled Caltation
   Pover meter NRP                    stt torre              ot—Apr—17 (Ne. 217—cosetiase0)                 Apeie
   Poer sensor NRP—231                Sh: rosacs             O—Ape17 (No arz—cesen                          Apere
   Power sensor NRP—z91               sit roms               o—Apet (No. 217—02522)                         Apere
   Referonce 20 dB Attonator          sh:sose (200           0—Apr—t7 (No.217.02520)                        Apere
   TypeN mismatch comninaton          SN:soar.a /osser       or—Apr—t7 (No. 2t7—02se9)                      Apers
   Raference Probe EXIDV4             snc ro                 3l—May—17 (No. EX3—7949. Mayt?)                Mayte
   paee                               sn eor                 28Mar—17 (No. DAEA—601_Mart?)                  Marie
   Secondary Standards                n«                     Check Date (n house)                           Scheduled Check
   Poar mater EPM—442N                SN: Gasramoros         0rOct18 (nhouse check Oct16)                   In house chacic Oct—18
   Power sensor HP 881A               sh: ussreszres         07—0st—15 (n house check Oct—16)               In house chack: Oct—18
   Poer sensor HP 881A                shmverocest7           07—0ct—15 in house check Oct—16)               In house chack: Oct—18
   AF generater NS SMT—08             sn: roowra             15—Jun—15 (n house check Oct10)                In house check: Oct18
   Natwork Analyzer HP 875E           sn ussrasoses          180001 (n house check Oct16)                   In house chack: Oct—17
                                      Name                               Functon                             Signature
   Calbrated by:                      JohannesKuridca                    Laboratoy Technican
                                                                                                                         low—
   Approved by:                       Kata Porovie                       Techrical Manager
                                                                                                        kT  Issued: Juty 28, 2017
   "This caltrtion corticate shall not be reproduced exceptin fulwthout witen approvalo e aboratory

   Certficate No: D1900V2—50101_Jult7                            Page 1 or8


Calibration
     r      Laboratory of                               \i\‘lé/          «. Sotwreizerischer Kallbrerdionst
Schmid & Partner                                       m                 c   Service sulsse détatonnage
  Engineering AG                                       7 3                   Serviclo sviezero di taratura
Zeughausstrasse 49, 8004 Zurich, Switzorand             y                5. swise Callpration Srvice

Accredted by he Sniss Accredtation Srvice (GAG)                           Accreditation No.: SCS 0108
The Suis Accreditation Sorviceis one of the signatoriesto the EA
Multiatoral Agroomant forthe recognition ofcallration corttatos
Glossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORM x,y,z
NA                        not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62208—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)*, July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
          communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)®, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filed phantomn. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antennafeed point.
          No uncertainty required.
     *«   SAR measured: SAR measured at the stated antenna input power.
     *«   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     +    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certficate No: D1900V2—5d101_Jult?                         Page 2 of 8



Document Created: 2018-11-13 16:02:49
Document Modified: 2018-11-13 16:02:49

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC