SAR Calibration Report-3

FCC ID: 2AOPD-UT10

RF Exposure Info

Download: PDF
FCCID_4214141

C‘                           mm
                                    in Collabaration with

         ETTL 2
          \@@ggzgger~               CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
        Tel: +86—10—02304033—2079       Fax: +86—10—62304633—2504
        E—mail: citl@chinattl.com       hitps//wwwchinattl.en


Glossary:
TSL                      tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
«_ Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
«_   Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
&     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connectorto the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
&     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
      SAR measured: SAR measured at the stated antenna input power.
      SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
&     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.



Certificate No: Z17—9708 1                             Page 2 of8


 U                            m
                               *    In Collaboration with

                TTZL, _
                                    CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        tel: +86—10—02304033—2079       Fax: +86—10—02304033—2504
        F—mail: ettl@chinattl.com       hitp:/iwwwchinattl.cn


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                           DASY52                                     52.10.0.1446

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                           10 mm                                    with Spacer

     Zoom Scan Resolution                           dx, dy, de = 5 mm

     Frequency                                       1900 MHz & 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature              Permittivity           Conductivity
     Nominal Head TSL parameters                               22.0 °C                   40.0                1.40 mho/m
     Measured Head TSL parameters                           (22.0 £ 0.2) °C            40.5 + 6 %          1.39 mhoim + 6 %

     Head TSL temperature change during test                    <1.0 °C                   ——                       ———
SAR result with Head TSL
     SAR averaged over 1_cm" _(1 g) of Head TSL                           Condition
     SAR measured                                                   250 mW input power                   9.88 mW /g
     SAR for nominal Head TSL parameters                             normalized to 1W           39.9 mW 1g + 18.8 % (k=2)

     SAR averaged over 10 cn" (10 g) of Head TSL                          Condition
     SAR measured                                                   250 mW input power                   5.10 mW /g
     SAR for nominal Head TSL parameters                             normalized to 1W           20.5 mW /g 2 18.7 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature              Permittivity           Conductivity

     Nominal Body TSL parameters                               22.0 °C                   53.3                1.52 mhoim

     Measured Body TSL parameters                           (22.0 £0.2) °C             53.0 £ 6 %          1.51 mhoim £ 6 %

     Body TSL temperature change during test                   <1.0 °C                                             ——
SAR result with Body TSL
     SAR averaged over 1_cm"_(1 g) of Body TSL                            Condition
     SAR measured                                                   250 mW input power                   9.95 mW /g

     SAR for nominal Body TSL parameters                             normalized to 1W           39.9 mW /g £ 18.8 % (k=2)

     SAR averaged over 10 cn° (10 g) of Body TSL                          Condition
     SAR measured                                                   250 mW input power                   5.24 mW /g

     SAR for nominal Body TSL parameters                             normalized to 1W           21.0 mW /g £ 18.7 % (k=2)

Certificate No: 717—9708 1                             Page 3 of 8


.@@‘fl                          m


           @77TL s_p_e_a_gq
            .'\'                     in Collaboration with
           §                        S              e          a

         Add: No.51 Xueyuan RoadHaidian District, Beijing, 100191, China
                                           c                 4633—2504




  Appendix (Additional assessments outside the scope of CNAS LO0570)

  Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                 49.20+ 5.27;0

      Return Loss                                                            —2540B


  Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                 46.60+ 4.3210

      Return Loss                                                            —24.9 dB


  General Antenna Parameters and Design

       Electrical Delay (one direction)                                      1.308 ns




  After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
  be measured.

  The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding lineis directly
  connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
  of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
  according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
  affected by this change. The overall dipole lengthis still according to the Standard.
   No excessive force must be applied to the dipole arms, because they might bend or the soldered
  connections near the feedpoint may be damaged.



  Additional EUT Data

       Manufactured by                                                          SPEAG




  Certificate No: Z17—97081                             Page 4 of 8


C                            mm
         fi                           ;Collnborafionewnh 8


          l                          CALIBRATION LABORATORY

       Add: No.51 Xucyuan Road, Haidian District, I       00191, China
       Tel:   +86—10—62304633—2070      Fax: +86—10—62304633—2504
       E—mail: cttt@chinattl.com        hitps/wwwchinattlen


DASYS Validation Report for Head TSL                                          Date: 06.30.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50193
  Communication System: UID 0, CW: Frequency: 1900 MHz; Duty Cyele: 1:1
  Medium parameters used: {= 1900 MHz: 0 = 1.385 S/m; er = 40.51; p = 1000 kg/m3
   Phantom section: Left Section
   Measurement Standard: DASYS (IEEE/IEC/ANS1 €63.19—2007)
DASYS Configuration:

              +    Probe: EX3DV4 — SN3617: ConyF(8.26, 8.26, 8.26); Calibrated: 1/23/2017;
              +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
              +    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
              «+   Phantom: Triple Flat Phantom5.1C; Type: QD 000 P51 CA; Serial: 1161/1
              +    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                   (7417)
        System Performance Check/Zoom Scan(7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 98.36 V/m; Power Drift = 0.00 dB
        Peak SAR (extrapolated) = 18.9 W/kg
        SAR(I g)=9.88 W/kg; SAR(1O g) = 5.1 W/kg
        Maximum value of SAR (measured) = 15.6 W/ke




              —3.58


               447


               —10.75


               14.34


               17.92

                        0 dB =15.6 W/kg= 11.93 dBW/kg




 Certificate No: Z17—97081                             Page 5 of 8


MH_—

         fi in Collaboration with
          \lizzzmyyy»7              CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191 China
       Tel: +86—10—62304633—2079     Faxc +86—10—62504633—2504
       E—mail: etl@chinatt.com        hitp/fwwwchinattl.en




        Impedance Measurement Plot for Head TSL


         Tri Sif tog wag 10.00de/    mef 0.000de F]
           59: °9 rst—1,9000000 onz =25—408 ds
           10.00
           30.00
           20.00
           10.00
           0.000p
          —10.00

          ~20,00

          —30.c0
          —40.00
         —s0.00                                                       x
        NiR siz snith (ReJx) scate 1.0000 [Fi oe1]
           s1 1.9000000 ue 49.102 0. 5.2677 0. 441.26—pH
                                                        /2                ~




Certificate No: Z17—9708 1                            Page 6 of 8


‘@@n_—
          _i_8                      in Collsboration with
          =7"7}J, a
          v                         CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: citl@chinattl.com       hitp/www.chinattl.en




DASY5 Validation Report for Body TSL                                        Date: 06.30.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50193
  Communication System: UID 0, CW: Frequency: 1900 MHz: Duty Cycle: 1:1
   Medium parameters used: £= 1900 MHz: a = 1.509 $/m: &, = 53.02; p = 1000 kg/m‘
   Phantom section: Center Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)
DASY5 Configuration:

            *    Probe: EX3DV4 — SN3617; ConyF(7.95, 7.95, 7.95); Calibrated: 1/23/2017;
            +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            *«   Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
            *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *    Measurement SW: DASYS2, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Mcasurement grid:
        dx=5mm, dy=3mm, dz=5mm
        Reference Value = 97.05 V/m; Power Drift =—0.02 dB
        Peak SAR (extrapolated) = 18.0 W/kg
        SAR(I g) =9.95 W/kg; SAR(10 g) = 5.24 W/kg
        Maximum value of SAR (measured)= 15.3 W/kg


           dB
           0


            —3.38


            —6.76


            410.15


            "13.53


            416.91
                       0 dB = 15.3 W/kg = 11.85 dBW/kg




Certificate No: 717—9708 1                             Page 7 of 8


Report No.: BL-EC1930222-701


                                   ®      in Collaboration with                w UI"’                           HBLAH

                     T7                   CcaLisraNon LAporatory
                                                                                    _aa¥c
                                                                               STY& CNAs Hot
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China ’@‘                              CALIBRATION
             Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504           falistubs®                    CNAS LO570
             E—mail: cttl@chinattl.com        hitp/fwww.chinattl.en

      Client             baluntek                                          Certificate No:       217—97036



Object                                     D2450V2 — SN: 952

Calibration Procedure(s)                   FD—211—003—0

                                           CalibrationProcedures for dipole validationkits
Calibration date:                          March 21, 2017

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperature(22«3)C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                  ID #               Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Power Meter NRP2                  101919             27—Jun—16 (CTTL, No.J16X04777)                       Jun—17
 Power sensor       NRP—Z91        101547            27—Jun—16 (CTTL, No.J16X04777)                        Jun—17
 Reference Probe EX3DV4            SN 3617            23—Jan—17(SPEAG,No.EX3—3617_Jan17)                  Jan—18
 DAE4                             SN 777             22—Aug—16(CTTL—SPEAG.No.216—97138)                   Aug—17

 Secondary Standards               ID #               Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Signal Generator E4438C           MY49071430          13—Jan—17 (CTTL, No.J17X00286)                     Jan—18
 Network Analyzer ES071C           MY46110673         13—Jan—17 (CTTL, No.J17X00285)                      Jan—18



                                  Name                         Function                                  Signature
Calibrated by:                   Zhao Jing                   SAR Test Engineer


Reviewed by:                     iDianyuan                   SAR Project Leader         _           .:"%> 92           >
Approved by:                     LuBingsong                  Deputy Director of the laboratory       {»T_?Q A              '
                                                                                 Issued: March 25, 2 1-
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z17—97036                                Page 1 of 8


%_7



         Add: NS Xueyuan Road,Hhidan Dinic, Bling, 10019 China
         Tek e ieasieeisa070       Facoteoansoi6s—2s04
         Eomaltcttchinat om        npshawwchinatien
Glossary:
TSL                    tissue simulating liquid
ConvF                   sensitivity in TSL / NORMxy,z
N/A                    not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
     Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
     Communications Devices: Measurement Techniques, June2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
   devices used in close proximily to the ear (frequency range of 300MHz to 3GHz)*, February
     2005
) IEC 62200—2, *Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   3OMHz to GGHz)®, March 2010
d) KDBB65G64, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
+     Measurement Conditions: Further details are available from the Validation Report at the end
      of the certficate. All figures stated in the certficate are valld at the frequency indicated.
+.    Anterna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the tfat phantom section, with the arms oriented
      paraltel to the body axis.
+.    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filed phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
+     Electrical Delay: One—way delay between the SMA connector and the antenna feed point,
      No uncertainty required
*     SAR measured: SAR measured at the stated antenna input power.
*     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector
*     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probabiliy of approximately 95% .


Centificare No: 217—97036                      Page2 ofs


C

       Add: N1 Xocpun Road, Haidan Disric. Heting 100191; Chioa
       Tt «4incasomana02         rec ue ioamoiemastt
       Eomalettchinaleom         hophnvoactiation

Measurement Conditions
   DASY system confiuration, as faras notgiven on page 1
    DaSY version                                  oasyse                                  sess 12se
    Extrapolation                          Advanced Extapotaton
    Phantom                                TslFlatPhantom $.1C
    Distance Dipole Conter— TSL                   10 mm                                 wih Spacer
     Zoom Scan Resolution                    d«, dy. dz= $ mm
     Frequency                               2450 hes 1 fike

Head TSL parameters
   Te foloning parameters and calculatons were applied
                                                  Tomporature             Pormittvity          Conductivity
     Nominal Head TSL paramotors                     220°C                   302               180 mhoim

     Measured Head TSL paramotors                 @20+02)°C                300 26 %          177 mhoim x 6 %

     Head TSL temperature change during tost         <10°C                    ~=                     —
SAR result with Head TSL
   SAR averaged over 1 _cm"_(t g ofHead TSL                   Condition
     SAR measured                                         250 mW nput power                 130 mW (g

     SAR fornorminal Head TSL parametors                   normalzod to TW         $2.4 mW 1g2208 % (ket)

     SAR averaged over 10 c‘ (109)of Head TS                  Condition
     SAR measured                                         250 mW inut power                 5.04 mW ia

     SAR fornorminaHead TSL parameters                     normalzed to 1W         24.3 mW 19 204 % (het)
Body TSL parameters
   The folowing parameters and calculaions wore appled
                                                  Temperature             Pormitivity          Cenductivity
     Nominal Body TSL paramotors                     220°C                   527               195 mhoin
     Measured Body TL paramotors                  @20202)°C                52326%            198 mhoim 6 %

     Body TSL temperature change during test         sto°c                    ~                      we
SAR result with Body TSL
   SAR averaged over 1 _cin" (1 g)of Rody TSL                 Congtion
     SAR measured                                         250 mW input power                128 mw/g
     SAR for nominal Body TSL paramaters                   normatiznd to 1W        50.5 mW 1g 2 20.% (Ket)
     SAR averaged over 10 on‘ (10g) of Body TSL               Coniton
     SAR measured                                         250 mW input power                582 mW io

     SAR for nominal Body TSparameters                     nomallzed to TW         23.3 mW i9 204 % (ke)


Cenificate No: 217—97036                      Pages ofs


C‘                          esmm

         Add: No.31 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com      httpy/wwwchinattl.en


  Appendix (Additional assessments outside the scope of CNAS LO0570)

  Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                   52.00+ 5.940

      Return Loss                                                              —24.30B


  Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                   48.70+ 6.25]0

      Return Loss                                                              —23.80B

  General Antenna Parameters and Design

      Electrical Delay (one direction)                                         1.257 ns




  After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
  be measured.

  The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
  connected to the second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some
  of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
  according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
  affected by this change. The overall dipole length is still according to the Standard.
  No excessive force must be applied to the dipole arms, because they might bend orthe soldered
  connections near the feedpoint may be damaged.



  Additional EUT Data

      Manufactured by                                                             SPEAG




 Certificate No: Z17—97036                            Page 4 of 8


U                          m
                                      in Collaboration with


                                      CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2070      Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com      htpsfwwwchinattl.en

DASY5 Validation Report for Head TSL                                        Date: 03.21.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 952
  Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
  Medium parameters used: {= 2450 MHz; a = 1.768 $/m; er= 39.02; p = 1000 kg/m3
  Phantom section: Right Section
  Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)
DASY5 Configuration:

            «+   Probe: EX3DV4 — SN3617; ConvF(7.74, 7.74, 7.74); Calibrated: 1/23/2017;
            +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 $n777; Calibrated: $/22/2016
            +    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            +    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 104.6 V/m; Power Drift = 0.02 dB
        Peak SAR (extrapolated) = 27.0 W/kg
        SAR(I g) = 13 W/kg:; SAR(10 g)=6.04 W/kg
        Maximum value of SAR (measured) =21.7 W/kg




            ~4.42


            —8.84


            13.26


            —17.68

                          E.
            2210

                       0 dB =21.7 W/kg =13.36 dBW/kg




Certificate No: Z17—97036                             Page 5 of 8


        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       hitp=//www.chinatl.en




        Impedance Measurement Plot for Head TSL


      Tri si1 Log Mag 10.00d8/   ref 0.000d8 Fi]
        50.00
                >12.4500000 ne —24.261 do
       «0.00
       30. co
       20. 00

       10.00

       0.ccop
      —20. 00

      ~20.00
      ~30.00

      —40. 00
      ~50. 00
     MBR S11 snith (ReJ>) scale 1.000u [ ve1]
       »1. 2.4500000 ane st.9s4 o. s.9372 n 3e5.6




Certificate No: Z17—97036                            Page 6 of 8


U                           m

            a 86—10—62304633—2079   Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com   hitps/wwwchinatl.en


DASYS Validation Report for Body TSL                                       Date: 03.21.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 952
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle:1:1
   Medium parameters used: {= 2450 MHz; a= 1.931 S/m; s, = 52.27; p= 1000 kg/im‘
   Phantom section: Center Section
  Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
DASY3 Configuration:

            *   Probe: EX3DV4 — SN3617; ConvF(7.8, 7.8, 7.8); Calibrated: 1/23/2017;
            +   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +   Electronics: DAE4 Sn777; Calibrated: 8/22/2016
            +   Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            +    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 96.07 V/m; Power Drift =—0.04 dB
        Peak SAR (extrapolated) = 26.4 W/kg
        SAR(I g) = 12.6 W/kg; SAR(1O g) = 5.82 Wkg
        Maximum value of SAR (measured) =20.9 W/kg


           dB                             —                     —      _    Te
           0


            ~4.41


            —8.82


            "13.22


            417.63

                         k.
            —22.04

                      0 dB =20.9 W/kg =13.20 dBW/kg




Certificate No: Z17—97036                       Page 7 of 8


Report No.: BL-EC1930222-701


                      Report No.: BL-EC1930222-701

F.8 2600 MHz Dipole


‘.%_—


      r
Calibration Laboratory of                                                    Schweizerischer Kaltertonst
Schmid & Partner                                                         g Sorvice       étatonnage
  Engineering AG                                                           Sorvito viezero i trature
Zooghausstrasse 43, 0004 2urich, Stzeriand                               S Swise Gatbration Sevice

Accrted by tha Swiss Accredtaton Sovics (8A5)                            Accrediution .: SCS 0108
The Suiss Aceredation Serice is one ofthesignatoriosto the EA
Moitlateral Agreementfor the recogntion of allraioncerilcates
Glossary:
TSL                      tissue simulating liquid
ConvF                    sensitivity in TSL / NORM xy,z
N/A                      not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2019, ‘IEEE Recommended Practice for Determining the Peak Spatial«
        Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques®, June 2013
    b  IEC 62209—1, "Measurement procedure for the assessmentof Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c IEC 62209—2, "Procedure to determine the Specific Absorption Rate (GAR) for wireless
       communication devices used in closeproximity to the human body (frequency range of 30
       MHz to 6 GHz)®, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    *   Measurement Conditions: Further details are available from the Validation Report at the end
        of the certficate. All figures stated in the certificate are valid at the frequency indicated.
    *   Antenna Parameters with TSL: The dipole is mounted with the spacer to position ite feed
        point exactly below the center marking of the flat phantom section, with the arms oriented
        parallel to the body axis.
    *   FeedPoint Impedance and Return Loss: These parameters are measured with the dipole
        positioned under the liquid fled phantom. The impedancestated is transformed from the
        measurement at the SMA connector to the feed point. The Retun Loss ensures low
        reflected power. No uncertainty required.
    * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
        No uncertainty required.
    * SAR measured: SAR measured at the stated antenna input power.
    *   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
    +   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
        nominal SAR result.
 The reported uncertainty of measurementis stated as the standard Uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.



Corticate No: D26002—1085.Jut?                          Page 2 0f8


-@@n_
Measurement Conditions
   DASY system configuration, asfar as not given on page 1.
     DASY Version                                              DASY5                              V52.10.0

     Extrapolation                                   Advanced Extrapolation

     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm

     Frequency                                         2600 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.0               1.96 mho/m
     Measured Head TSL parameters                              (22.0 £ 0.2) °C   87.2 26 %          2.04 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            «+                      sez

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        250 mW input power                     14.5 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        56.4 W/kg 2 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.40 Wikg

     SAR for nominal Head TSL parameters                     normalized to 1W        25.2 Wikg 2 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                  22.0 °C            52.5              2.16 mho/m
     Measured Body TSL parameters                              (22.0 £ 0.2) °C   51.6 + 6 %         2.22 mho/m x 6 %

     Body TSL temperature change during test                      <0.5°C             ——                      ———

SAR result with Body TSL

     SAR averaged over 1 cm‘ (1 g) of Body TSL                   Condition
      SAR measured                                       250 mW input power                     13.8 Wikg

      SAR for nominal Body TSL parameters                    normalized to 1W         54.3 W/kg 2 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     6.15 Wig

      SAR for nominal Body TSL parameters                    normalized to 1W         24.4 Wikg 2 16.5 % (k=2)




Certificate No: D2600V2—1095_Jul17                    Page 3 of 8


&&&
 Appendix (Additional assessments outside the scope of SCS 0108)

 Antenna Parameters with Head TSL

       Impedance,transtormas to faed point                                         ar20—74 j
       Return Loss                                                                   —21.008

 Antenna Parameters with Body TSL

       Impedance,transformed to feed point                                         s160—50 j
       Retur Loss                                                                   —21508


 General Antenna Parameters and Design

     [ Electrical Detay (one girection}                    |                          1.180 ns
 Alterlong tarm use with 100W radiated power, only a slght warming of the dipole near the feedpointcan be measured.
 The dipol is made of standard semrigid coaxial eable. The center condluctor of the feeding e is drectly connected to the
 second arm ofthe dipole, The antenn is therefore shortcrcuited for DC—signals. On some ofthe cipoles, small end caps
 are added to the dipole arms in orderto improve matching when loaded according tothe postion as explained in the
 ‘Measuremont Conditions* paragraph. The SAR data are not affected by ths change. The overallipole length is stll
 according to the Standard.
 No excessive force must be applied to the dipole arms, because thay might bend or the soldered connections near the
 feedpoint may bo damaged.


 Additional EUT Data

       Manutactured by                                                                sreao
       Manutactured on                                                            October 12, 2016




 Gerificare No: Dzs00¥2—1095_Jut?                       Page 4 ot8


C                          mm

DASYS Validation Report for Head TSL

                                                                                Date: 10.07.2017
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 2600 MHz: Type: D2600V2; Serial: D260OV2 — SN:1095
Communication System; UID 0 — CW; Frequency: 2600 MHz
Mediumparameters used: I‘= 2600 MHz: a =2.04 $/m; £, = 37.2; p = 1000 kg/m‘
Phantomsection: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—201 1)
DASY52 Configuration:

   *   Probe: EX3DV4 SN7349; ConvF(7.96. 7.96, 7.96); Calibrated: 31.05.2017;
   +   Sensor Surface: 1 Amm (Mechanical Surface Detection)

   +   Electronics: DAE4 Sn601; Calibrated: 28.03.2017
   +   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   +   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pi         250 mW, =10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=Smm, dy=5mm, de=Smim
Reference Value= 114.1 V/m; Power Drift = —0.08 dB
Peak SAR (extripolated)       31.2 W/kg
SAR(T g) =  14.5 W/kz: SAR(IOg) 6.4 We
Maximum value of SAR (measured) = 24.2 We

       4n
       a

       .00

       10.00

       +15.00

       —20.00

       —25.00
                  0 dB =24.2 W/kg = 13.84 dBW/kg




Gerticate No: D2600V2—1095 _Jut?              Page S of 8


C                          mm
Impedance Measurement Plot for Head TSL




                                                            a0 aul meur    asisoine
         tBD sn      io rs          menaree —rossse anauer                2 go0.000 o0n mz

         te




         cne sin     uos      5 cerngr ~a0 e            m—2u001 on 2 sooe see mz



         ca




         ma
                   Stait 2 sonooo o00 mz                         Stoe 2 son.000 ooo me




Conticate No: 2s00v2—1088.lut?                 Page 6ot 8


C                        esmm
DASY5 Validation Report for Body TSL
                                                                                Date: 10.07.2017

Test Luboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D260OV2; Serial: D2600V2 — SN:1095

Communication System: UID 0 — CW; Frequen  2600 MHz
Medium parameters used:      2600 MHz:     S/m; £, = 51.6; p =   1000 ke/m‘
Phantomsection: Flat Section
Mcasurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)
DASY52 Configuration:
   * Probe: EX3DV4 — SN7349; ConvB(7.94, 7.94, 7.94); Calibrated: 31.05.2017;
   *   Sensor—Surfice: 1.4mm (Mechanical Surface Detection)
   * Electronics: DAE4 Sn601; Calibrated: 28.03.2017
   + Phantom: Flat Phantom 5.0(back);Type: QD 000 PSO AA; Serial: 1002
   +   DASY32 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for BodyTissue/Pin=250 mW, d=10mm/ZoomScan (7x7x7)/Cube 0:
Measurement grid: dx=Smm, dy=5mm, de
Reference Value= 103.9 ¥/m; Power Drif
Peak SAR (extrapolated) = 28.9 W/kg
SARC g)           /

       an
        o




       12.00

       16.00

       20.00
                 0 4B =22.1 Whkg= 13444 dBWhke




Cortfcate No: D26002—1095_Jutt?               Page 7 of8


               Report No.: BL-EC1930222-701




‐‐END OF REPORT‐‐



Document Created: 2019-03-18 12:48:47
Document Modified: 2019-03-18 12:48:47

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC