Test Report_DTS

FCC ID: 2ANB8-T1

Test Report

Download: PDF
FCCID_4078477

                                      Test Report

                                        FCC ID: 2ANB8-T1


                                      Date of issue: Oct. 22, 2018



       Report Number:                              MTi181018E052


       Sample Description:                         GETWELL Aerobic capacity assessment and muscle oxygen
                                                   monitoring


       Model(s):                                   T1


       Applicant:                                  Getwell Health Technology(Wuhu)Co., Ltd.


       Address:                                    4th Building, Service outsourcing park, new and high-tech
                                                   zone, Yijiang District, Wuhu, Anhui


       Date of Test:                               Oct. 11, 2018 - Oct. 22, 2018




                                          Shenzhen Microtest Co., Ltd.
                                            http://www.mtitest.com




This test report is valid for the tested samples only. It cannot be reproduced except in full without prior written consent of Shenzhen Microtest
Co., Ltd.
  Tel:(86-755)88850135                Fax: (86-755) 88850136                 Web: http://www.mtitest.com              E-mail: mti@51mti.com
  Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                                               - Page 2 of 27 -                                        Report No.: MTi181018E052

                                                                      Table of Contents

1.       GENERAL INFORMATION ................................................................................................................................... 4
     1.1.        DESCRIPTION OF EUT ................................................................................................................................................... 4
     1.2.        OPERATION CHANNEL LIST .............................................................................................................................................. 4
     1.3.        TEST CHANNEL LIST ....................................................................................................................................................... 5
     1.4.        ANCILLARY EQUIPMENT LIST ........................................................................................................................................... 5
     1.5.        DESCRIPTION OF SUPPORT UNITS .................................................................................................................................... 5
2.       SUMMARY OF TEST RESULTS .......................................................................................................................... 6
3.       TEST FACILITIES AND ACCREDITATIONS ...................................................................................................... 7
     3.1.        TEST LABORATORY ........................................................................................................................................................ 7
     3.2.        ENVIRONMENTAL CONDITIONS ........................................................................................................................................ 7
     3.3.        MEASUREMENT UNCERTAINTY ........................................................................................................................................ 7
     3.4.        TEST SOFTWARE ........................................................................................................................................................... 7
4.       EQUIPMENT LIST ................................................................................................................................................. 8
5.       TEST RESULT ....................................................................................................................................................... 9
     5.1.     ANTENNA REQUIREMENT ............................................................................................................................................... 9
        5.1.1 Standard requirement ......................................................................................................................................... 9
        5.1.2 EUT Antenna ....................................................................................................................................................... 9
     5.2.     CONDUCTED EMISSION ................................................................................................................................................ 10
        5.2.1 Limits ................................................................................................................................................................. 10
        5.2.2 Test setup .......................................................................................................................................................... 10
        5.2.3 Test procedure ................................................................................................................................................... 11
        5.2.4 Test results......................................................................................................................................................... 11
     5.3      RADIATED SPURIOUS EMISSION...................................................................................................................................... 16
        5.3.1 Limits ................................................................................................................................................................. 16
        5.3.2 Test setup .......................................................................................................................................................... 17
        5.3.3 Test procedure ................................................................................................................................................... 18
        5.3.4 Test results......................................................................................................................................................... 19
        5.3.4.1    Radiation emission ....................................................................................................................................... 19
        5.3.4.2    Bandedge-radiated ....................................................................................................................................... 23
        5.3.4.3    Spurious Emission in Restricted Band 3260MMHz-18000MHz ..................................................................... 24
PHOTOGRAPHS OF THE TEST SETUP ................................................................................................................................ 25
PHOTOGRAPHS OF THE EUT ............................................................................................................................................ 27




  Tel:(86-755)88850135                       Fax: (86-755) 88850136                                Web: http://www.mtitest.com                       E-mail: mti@51mti.com
  Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 3 of 27 -                       Report No.: MTi181018E052




                             PRODUCT INFORMATION

  Applicant’s name:                Getwell Health Technology(Wuhu)Co., Ltd.

                                   4th Building, Service outsourcing park, new and high-tech zone, Yijiang
  Address:                         District, Wuhu, Anhui


  Manufacture's Name:              Getwell Health Technology(Wuhu)Co., Ltd.

                                   4th Building, Service outsourcing park, new and high-tech zone, Yijiang
  Address:                         District, Wuhu, Anhui


  Product name:                    GETWELL Aerobic capacity assessment and muscle oxygen monitoring


  Trademark:                       getwell


  Model name:                      T1


  Standards:                       FCC Part15 Subpat C
                                   ANSI C63.10:2013
                                   KDB 558074 D01 DTS Meas Guidance v05
  Test Procedure:                  KDB 174176 D01 Line Conducted FAQ v01r01

This device described above has been tested by Shenzhen Microtest Co., Ltd and the test results show that the
equipment under test (EUT) is in compliance with the FCC requirements. And it is applicable only to the tested
sample identified in the report.

Note: The RF component test data for this report is based on the original test report MTi170811E069 with a date
of 2017-08-11. This report only reflects the retesting part, the conduction and radiation data are updated, and the
rest of the data remains unchanged.

                  Tested by:s

                                                               Leo Su                              Oct. 22, 2018


                Reviewed by:


                                                               Blue Zheng                          Oct. 22, 2018

               Approved by:

                                                               Smith Chen                          Oct. 22, 2018


 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                        - Page 4 of 27 -                      Report No.: MTi181018E052


1. General Information
1.1. Description of EUT

                                             GETWELL Aerobic capacity assessment and muscle
      Product name:                           oxygen monitoring

      Model name:                            T1

      Serial model:                          N/A

      Difference in series
                                             N/A
      models:

      Operation frequency:                   2402-2480MHz

      Modulation type:                       GFSK

      Bit Rate of transmitter:               1 Mbps

      Antenna type:                          FPCB Antenna

      Antenna gain:                          2dBi

      Max. output power:                     1.348dBm

      Hardware version:                      V01

      Software version:                      V218

      Power supply:                          DC 3.7V by Battery or DC 5V from adapter AC 120V/60Hz

                                             Model:SK12G-0500100U
      Adapter information:                   Input:AC100-240V~ 50/60Hz 0.2A
                                             Output:5V 1A

      Battery:                               DC 3.7V 205mA



1.2. Operation channel list

     Channel Frequency Channel Frequency Channel Frequency Channel Frequency
       No.     (MHz)     No.     (MHz)     No.     (MHz)     No.     (MHz)
          0             2402            10           2422             20            2442             30            2462
          1             2404            11           2424             21            2444             31            2464
          2             2406            12           2426             22            2446             32            2466
          3             2408            13           2428             23            2448             33            2468
          4             2410            14           2430             24            2450             34            2470
          5             2412            15           2432             25            2452             35            2472
          6             2414            16           2434             26            2454             36            2474
          7             2416            17           2436             27            2456             37            2476
          8             2418            18           2438             28            2458             38            2478
          9             2420            19           2440             29            2460             39            2480

 Tel:(86-755)88850135          Fax: (86-755) 88850136               Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 5 of 27 -                        Report No.: MTi181018E052

1.3. Test channel list

                Channel                             Channel                                 Frequency (MHz)
                   Low                                  00                                         2402
                  Middle                                19                                         2440
                   High                                 39                                         2480

1.4. Ancillary equipment list

          Equipment                   Model                    S/N             Manufacturer            Certificate type
           Adapter            SK12G-0500100U                     /                      /                        /

1.5. Description of Support Units
The EUT has been tested as an independent unit together with other necessary accessories or support
units. The following support units or accessories were used to form a representative test configuration
during the tests.

           Item          Equipment                Brand              Model/Type No.           Series No.         Note
              /              /                      /                       /                      /
              /              /                      /                       /                      /
Note:
  (1)The support equipment was authorized by Declaration of Confirmation.
  (2)For detachable type I/O cable should be specified the length in cm in『Length』column.




 Tel:(86-755)88850135         Fax: (86-755) 88850136                 Web: http://www.mtitest.com          E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 6 of 27 -                       Report No.: MTi181018E052


2. Summary of Test Results
Test procedures according to the technical standards:

      No.               Standard Section                        Test Item                         Result             Remark

       1                      15.207                     Conducted Emission                       Pass

       2              15.247 (d) & 15.209           Radiated Spurious Emission                    Pass

       3                      15.203                    Antenna Requirement                       Pass




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com          E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 7 of 27 -                       Report No.: MTi181018E052


3. Test Facilities and Accreditations
3.1. Test laboratory

      Test Laboratory                   Shenzhen Microtest Co., Ltd
                                        No.102A & 302A, East Block, Hengfang Industrial Park, Xingye
      Location
                                        Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China
      FCC Registration No.:             448573

3.2. Environmental conditions

      Temperature:                      20ºC~30ºC
      Humidity                          30%~70%
      Atmospheric pressure              98kPa~101kPa

3.3. Measurement uncertainty

The reported uncertainty of measurement y ± U,where expended uncertainty U is based on a standard
uncertainty multiplied by a coverage factor of k=2,providing a level of confidence of approximately 95
%

              No.                                 Item                                        Uncertainty
               1                    Conducted Emission Test                                       ±1.38dB
               2                       RF power, conducted                                        ±0.16dB
               3                Spurious emissions, conducted                                     ±0.21dB
               4                  All emissions, radiated(<1G)                                    ±4.68dB
               5                  All emissions, radiated(>1G)                                    ±4.89dB
               6                            Temperature                                           ±0.5°C
               7                               Humidity                                            ±2%

3.4. Test software

          Software
                                       Manufacturer                            Model                            Version
            Name
       RF Test System                       Farad                              LZ-RF                          Lz_Rf 3A3




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 8 of 27 -                       Report No.: MTi181018E052


4. Equipment list
 Equipment              Equipment           Manufactur                                            Calibration
                                                                 Model          Serial No.                          Due date
    No.                   Name                 er                                                    date
                       Spectrum
   MTI-E001                                     Agilent         E4407B        MY41441082          2017/11/18       2018/11/17
                        Analyzer
                       CMU 200
                     universal radio        Rohde&schw
   MTI-E002                                                    CMU 200            114587          2017/11/18       2018/11/17
                     communication              arz
                         tester
                                            Rohde&schw
   MTI-E004        EMI Test Receiver                              ESPI           1000314          2017/11/18       2018/11/17
                                                arz
                         Broadband                             VULB916
   MTI-E006                                 schwarabeck                             872           2017/11/18       2018/11/17
                          antenna                                 3
                                                               BBHA912
   MTI-E007             Horn antenna        schwarabeck                            1201           2017/11/18       2018/11/17
                                                                 0D
   MTI-E014               amplifier            America           8447D         3113A06150         2017/11/18       2018/11/17
                       Conduction
                                                                              126A1343/20
   MTI-E015         Immunity Signal           Schloder         CDG6000                            2017/11/18       2018/11/17
                                                                                  15
                       Generator
                        Coupled
                                                                  CDA         A2210332/20
   MTI-E016            decoupling             Schloder                                            2017/11/18       2018/11/17
                                                                 M2/M3            15
                         network
                    Comprehensive           Rohde&schw
   MTI-E032                                                    CMW500             124192          2017/11/18       2018/11/17
                     test instrument            arz
   MTI-E034               amplifier             Agilent          8449B         3008A02400         2017/11/22       2018/11/21
                     Artificial power                          NSLK812
   MTI-E037                                 Schwarzbeck                            #841           2017/12/26       2018/12/25
                         network                                  7
                       Spectrum
   MTI-E040                                     Agilent         N9020A        MY49100060          2017/11/05       2018/11/04
                        analyzer
   MTI-E041         Signal generator            Agilent         N5182A        MY49060455          2017/11/23       2018/11/22
                        Analog signal
   MTI-E042                                     Agilent         E4421B        GB40051240          2017/11/23       2018/11/22
                         generator
                                                 Dare          RPR3006         16I00054SN
   MTI-E043             Power sensor                                                              2017/12/29       2018/12/28
                                             Instruments         W                 O16
   MTI-E047         10dB attenuator         Mini-Circuits      UNAT-10+           15542           2017/11/24       2018/11/23
                        spectrum            Rohde&schw
   MTI-E049                                                     FSP-38            100019          2017/11/18       2018/11/17
                        analyzer                arz
                      PSG Signal
   MTI-E050                                     Agilent         E8257D        MY46520873          2017/11/24       2018/11/23
                       generator
                      Active Loop
                                                                 FMZB
   MTI-E051          Antenna 9kHz -         Schwarzbeek                           00044           2017/11/26       2018/11/25
                                                                 1519 B
                         30MHz
                                              Chengdu
                         18-40GHz                              ZLNA-18-
   MTI-E052                                  step Micro                          1608001          2017/11/18       2018/11/17
                          amplifier                             40G-21
                                             Technology
                                                        BBHA917 BBHA91705
   MTI-E053         15-40G Antenna          Schwarzbeek                             2017/11/18      2018/11/17
                                                            0            82
                   Artificial power                     NSLK812
   MTI-E058                             Schwarzbeck                     #841        2017/11/05      2018/11/04
                       network                              7
 Note: the calibration interval of the above test instruments is 12 months and the calibrations are traceable to
 international system unit (SI).




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 9 of 27 -                       Report No.: MTi181018E052


5. Test Result
5.1. Antenna requirement
5.1.1 Standard requirement
     15.203 requirement: For intentional device, according to 15.203: an intentional radiator shall be
designed to ensure that no antenna other than that furnished by the responsible party shall be used
with the device
5.1.2 EUT Antenna
     The EUT antenna is FPCB antenna (2dBi). It comply with the standard requirement. In case of
replacement of broken antenna the same antenna type must be used.




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 10 of 27 -                      Report No.: MTi181018E052


5.2. Conducted emission
5.2.1 Limits

                                                                                  Class B (dBuV)
                     FREQUENCY (MHz)
                                                                    Quasi-peak                         Average
                           0.15 -0.5                                  66 - 56 *                        56 - 46 *
                           0.50 -5.0                                    56.00                            46.00
                           5.0 -30.0                                    60.00                            50.00
Note
       (1)The tighter limit applies at the band edges.
       (2)The limit of " * " marked band means the limitation decreases linearly with the logarithm of the frequency
       in the range.

5.2.2 Test setup

                              V e r tic a l R e fe r e n c e

                              G r o u n d P la n e                                                    T e s t R e c e iv e r



                              EUT
            40cm




                                                               80cm

            L IS N




                                                                                 H o r iz o n ta l R e f e r e n c e
                                                                                  G r o u n d P la n e


              N o t e : 1 .S u p p o r t u n its w e r e c o n n e c t e d t o s e c o n d L IS N .

                        2 .B o th o f L IS N s ( A M N ) a r e 8 0 c m fr o m E U T a n d a t le a s t 8 0

               cm         f r o m o t h e r u n its a n d o t h e r m e ta l p la n e s

                 s u p p o rt.   u n i ts .




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                           - Page 11 of 27 -                      Report No.: MTi181018E052



5.2.3 Test procedure
a. EUT Operating Conditions
       The EUT was configured for testing in a typical fashion (as a customer would normally use it). The
       EUT has been programmed to continuously transmit during test. This operating condition was
       tested and used to collect the included data.
b. The following table is the setting of the receiver

                      Receiver Parameters                                                      Setting
                          Attenuation                                                           10 dB
                        Start Frequency                                                       0.15 MHz
                        Stop Frequency                                                         30 MHz
                         IF Bandwidth                                                           9 kHz

c. The EUT was placed 0.8 meters from the horizontal ground plane with EUT being connected to the
   power mains through a line impedance stabilization network (LISN). All other support equipments
   powered from additional LISN(s). The LISN provide 50 Ohm/ 50uH of coupling impedance for the
   measuring instrument.
d. Interconnecting cables that hang closer than 40 cm to the ground plane shall be folded back and
   forth in the center forming a bundle 30 to 40 cm long.
e. I/O cables that are not connected to a peripheral shall be bundled in the center. The end of the
   cable may be terminated, if required, using the correct terminating impedance. The overall length
   shall not exceed 1 m.
f.     LISN at least 80 cm from nearest part of EUT chassis.
       For the actual test configuration, please refer to the related Item –EUT Test Photos.
5.2.4 Test results




     Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
     Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 12 of 27 -                      Report No.: MTi181018E052

Test data
                        GETWELL Aerobic capacity
  EUT:                  assessment and muscle oxygen                   Model Name. :              T1
                        monitoring
  Temperature: 26 ℃                                                    Relative Humidity: 54%
  Pressure:             1010hPa                                        Phase :                    L
  Test Voltage : DC 5V from adapter AC 120V/60Hz Test Mode:                                       Normal link




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                      - Page 13 of 27 -                      Report No.: MTi181018E052


                       GETWELL Aerobic capacity
EUT:                   assessment and muscle oxygen                   Model Name:                T1
                       monitoring
Temperature: 26 ℃                                                     Relative Humidity: 54%
Pressure:              1010hPa                                        Phase :                    N
Test Voltage : DC 5V from adapter AC 120V/60Hz Test Mode:                                        Normal link




Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                      - Page 14 of 27 -                      Report No.: MTi181018E052


                       GETWELL Aerobic capacity
EUT:                   assessment and muscle oxygen                   Model Name. :              T1
                       monitoring
Temperature: 26 ℃                                                     Relative Humidity: 54%
Pressure:              1010hPa                                        Phase :                    L
Test Voltage : DC 5V from adapter AC 240V/60Hz Test Mode:                                        Normal link




Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                      - Page 15 of 27 -                      Report No.: MTi181018E052


                       GETWELL Aerobic capacity
EUT:                   assessment and muscle oxygen                   Model Name. :              T1
                       monitoring
Temperature: 26 ℃                                                     Relative Humidity: 54%
Pressure:              1010hPa                                        Phase :                    N
Test Voltage : DC 5V from adapter AC 240V/60Hz Test Mode:                                        Normal link




Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 16 of 27 -                      Report No.: MTi181018E052



5.3 Radiated spurious emission
5.3.1 Limits

                    Frequency                           Field Strength                  Measurement Distance
                        (MHz)                        (micorvolts/meter)                            (meters)
                   0.009~0.490                           2400/F(KHz)                                 300
                   0.490~1.705                          24000/F(KHz)                                 30
                    1.705~30.0                                 30                                    30
                        30~88                                  100                                    3
                        88~216                                 150                                    3
                     216~960                                   200                                    3
                    Above 960                                  500                                    3

                    Spectrum Parameter                                                Setting
                          Attenuation                                                  Auto
                        Start Frequency                                             1000 MHz
                        Stop Frequency                                      10th carrier harmonic
             RB / VB (emission in restricted                   1 MHz / 1 MHz for Peak, 1 MHz / 10Hz for
                              band)                                                  Average

                     Receiver Parameter                                               Setting
                          Attenuation                                                  Auto
                   Start ~ Stop Frequency                            9kHz~150kHz / RB 200Hz for QP
                   Start ~ Stop Frequency                            150kHz~30MHz / RB 9kHz for QP
                   Start ~ Stop Frequency                        30MHz~1000MHz / RB 120kHz for QP




 Tel:(86-755)88850135         Fax: (86-755) 88850136                 Web: http://www.mtitest.com          E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 17 of 27 -                      Report No.: MTi181018E052



5.3.2 Test setup
Radiated emission test-up frequency below 30MHz




Radiated emission test-up frequency 30MHz~1GHz




Radiated emission test-up frequency above 1GHz




                              3m

Turntabl                                                                                     4m
   e
               EUT
                                               1.5
       1.5                                     m
       m
                                                                                     Amplifie          Test Receiver
                                                                                       r

 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                           - Page 18 of 27 -                      Report No.: MTi181018E052

5.3.3 Test procedure
a. EUT operating conditions. The EUT tested system was configured as the statements of 2.4 Unless
   otherwise a special operating condition is specified in the follows during the testing.
b. The measuring distance of at 3 m shall be used for measurements at frequency up to 1GHz. For
   frequencies above 1GHz, any suitable measuring distance may be used.
c. The EUT was placed on the top of a rotating table 0.8 meters above the ground at a 3 meter shield
   area test site. The table was rotated 360 degrees to determine the position of the highest radiation.
d. The height of the equipment or of the substitution antenna shall be 0.8 m; the height of the test
   antenna shall vary between 1 m to 4 m. Both horizontal and vertical polarizations of the antenna
   are set to make the measurement.
e. The initial step in collecting conducted emission data is a spectrum analyzer peak detector mode
   pre-scanning the measurement frequency range. Significant peaks are then marked and then
   Quasi Peak detector mode re-measured.
f.     If the Peak Mode measured value compliance with and lower than Quasi Peak Mode Limit, the
       EUT shall be deemed to meet QP Limits and then no additional QP Mode measurement
       performed.
g. For the actual test configuration, please refer to the related Item –EUT Test photos.
Note: Both horizontal and vertical antenna polarities were tested and performed pretest to three orthogonal axis.
The worst case emissions were reported.

During the radiated emission test, the Spectrum Analyzer was set with the following configurations:
      Frequency Band (MHz)                   Function              Resolution bandwidth                Video Bandwidth

              30 to 1000                         QP                        120 kHz                           300 kHz

                                               Peak                         1 MHz                             1 MHz
             Above 1000
                                              Average                       1 MHz                             10 Hz
Note: for the frequency ranges below 30 MHz, a narrower RBW is used for these ranges but the measured value
should add a RBW correction factor (RBWCF) where RBWCF [dB] =10*lg(100 [kHz]/narrower RBW [kHz]). , the
narrower RBW is 1 kHz and RBWCF is 20 dB for the frequency 9 kHz to 150 kHz, and the narrower RBW is 10
kHz and RBWCF is 10 dB for the frequency 150 kHz to 30 MHz.




     Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
     Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 19 of 27 -                       Report No.: MTi181018E052



5.3.4 Test results
5.3.4.1 Radiation emission
Below 30MHz
                             GETWELL Aerobic
      EUT:                   capacity assessment and             Model Name:                  T1
                             muscle oxygen monitoring
      Temperature:           20 ℃                                Relative Humidtity:48%
      Pressure:              1010 hPa                            Test Voltage:      DC 5V from adapter
      Test Mode:             TX                                  Polarization :     --

              Freq.                Reading                    Limit                   Margin                  State
             (MHz)                 (dBuV/m)                (dBuV/m)                    (dB)                    P/F
                --                      --                      --                       --                   Pass
                --                      --                      --                       --                   Pass
Note:
The amplitude of spurious emissions which are attenuated by more than 20dB below the
permissible value has no need to be reported.
Distance extrapolation factor =40 log (specific distance/test distance)(dB);
Limit line = specific limits(dBuV) + distance extrapolation factor.




 Tel:(86-755)88850135         Fax: (86-755) 88850136                  Web: http://www.mtitest.com       E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 20 of 27 -                      Report No.: MTi181018E052


Between 30MHz – 1GHz:
    All the modulation modes have been tested, and the worst result was report as below:
                         GETWELL Aerobic capacity
 EUT:                    assessment and muscle oxygen                  Model Name : T1
                         monitoring
 Relative
                         52%                                           Phase:                H
 Humidity:
                                                                                             DC 5V from adapter AC
 Pressure:               1010 hPa                                      Test Voltage :
                                                                                             120V/60Hz
 Test Mode :             Normal link




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                      - Page 21 of 27 -                      Report No.: MTi181018E052


                       GETWELL Aerobic capacity
EUT:                   assessment and muscle oxygen Model Name :                            T1
                       monitoring
Relative
                       52%                                       Phase:                     V
Humidity:
                                                                                            DC 5V from adapter AC
Pressure:              1010 hPa                                  Test Voltage :
                                                                                            120V/60Hz
Test Mode :            Normal link




Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 22 of 27 -                      Report No.: MTi181018E052

1G-25GHz
Note:   (1) All Readings are Peak Value (VBW=3MHz) and AV Value (VBW=10Hz).
        (2) Emission Level= Antenna Factor + Cable Loss + Read Level - Preamp Factor
        (3) All other emissions more than 20dB below the limit.

All the modulation modes have been tested, and the worst result was report as below:
Frequenc       Read        Cable      Antenna   Preamp Emission
                                                                     Limits                    Margin
    y          Level        loss       Factor   Factor      Level                                             Remark       Comment
  (MHz)       (dBμV)        (dB)       dB/m      (dB)     (dBμV/m) (dBμV/m)                       (dB)
                                           Low Channel (2402 MHz)-Above 1G
 4804.34       62.80        5.21        35.59          44.30        59.30         74.00        -14.70            Pk          Vertical
 4804.34       41.49        5.21        35.59          44.30        37.99         54.00        -16.01            AV          Vertical
 7206.11       60.19        6.48        36.27          44.60        58.34         74.00        -15.66            Pk          Vertical
 7206.11       41.05        6.48        36.27          44.60        39.20         54.00        -14.80            AV          Vertical
 4804.17       62.99        5.21        35.55          44.30        59.45         74.00        -14.55            Pk        Horizontal
 4804.17       42.21        5.21        35.55          44.30        38.67         54.00        -15.33            AV        Horizontal
 7206.21       62.24        6.48        36.27          44.52        60.47         74.00        -13.53            Pk        Horizontal
 7206.21       41.04        6.48        36.27     44.52     39.27    54.00                     -14.73            AV        Horizontal
                                            Mid Channel (2440 MHz)-Above 1G
 4880.47       62.46        5.21        35.66          44.20        59.13         74.00        -14.87            Pk          Vertical
 4880.47       44.26        5.21        35.66          44.20        40.93         54.00        -13.07            AV          Vertical
 7320.27       65.77        7.10        36.50          44.43        64.94         74.00           -9.06          Pk          Vertical
 7320.27       42.30        7.10        36.50          44.43        41.47         54.00        -12.53            AV          Vertical
 4880.37       63.05        5.21        35.66          44.20        59.72         74.00        -14.28            Pk        Horizontal
 4880.37       41.56        5.21        35.66          44.20        38.23         54.00        -15.77            AV        Horizontal
 7320.23       61.03        7.10        36.50          44.43        60.20         74.00        -13.80            Pk        Horizontal
 7320.23       44.04        7.10        36.50     44.43     43.21     54.00                    -10.79            AV        Horizontal
                                           High Channel (2480 MHz)- Above 1G
 4960.48       64.30        5.21        35.52          44.21        60.82         74.00        -13.18            Pk          Vertical
 4960.48       43.04        5.21        35.52          44.21        39.56         54.00        -14.44            AV          Vertical
 7440.13       65.40        7.10        36.53          44.60        64.43         74.00           -9.57          Pk          Vertical
 7440.13       49.24        7.10        36.53          44.60        48.27         54.00           -5.73          AV          Vertical
 4960.33       62.96        5.21        35.52          44.21        59.48         74.00        -14.52            Pk        Horizontal
 4960.33       43.77        5.21        35.52          44.21        40.29         54.00        -13.71            AV        Horizontal
  7440.2       63.53        7.10        36.53          44.60        62.56         74.00        -11.44            Pk        Horizontal
  7440.2       45.89        7.10        36.53          44.60        44.92         54.00           -9.08          AV        Horizontal




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com           E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 23 of 27 -                      Report No.: MTi181018E052




5.3.4.2 Bandedge-radiated
Note:   (1) All Readings are Peak Value (VBW=3MHz) and AV Value (VBW=10Hz).
         (2)Emission Level= Antenna Factor + Cable Loss + Read Level - Preamp Factor
         (3) All other emissions more than 20dB below the limit.

All the modulation modes have been tested, and the worst result was report as below:
 Frequenc     Meter     Cable    Antenna Preamp Emission
                                                                   Limits     Margin                     Detector
      y     Reading     Loss      Factor     Factor     Level                                                        Comment
   (MHz)     (dBμV)      (dB)     dB/m        (dB)    (dBμV/m) (dBμV/m)         (dB)                       Type
                                                   GFSK
  2310.00     64.17     2.97      27.80      43.80      51.14        74        -22.86                       Pk       Horizontal
  2310.00     42.92     2.97      27.80      43.80      29.89        54        -24.11                       AV       Horizontal
  2310.00     61.69     2.97      27.80      43.80      48.66        74        -25.34                       Pk        Vertical
  2310.00     41.72     2.97      27.80      43.80      28.69        54        -25.31                       AV        Vertical
  2390.00     63.70     3.14      27.21      43.80      50.25        74        -23.75                       Pk        Vertical
  2390.00     43.29     3.14      27.21      43.80      29.84        54        -24.16                       AV        Vertical
  2390.00     63.93     3.14      27.21      43.80      50.48        74        -23.52                       Pk       Horizontal
  2390.00     42.37     3.14      27.21      43.80      28.92        54        -25.08                       AV       Horizontal
  2483.50     62.43     3.58      27.70      44.00      49.71        74        -24.29                       Pk        Vertical
  2483.50     42.26     3.58      27.70      44.00      29.54        54        -24.46                       AV        Vertical
  2483.50     65.73     3.58      27.70      44.00      53.01        74        -20.99                       Pk       Horizontal
  2483.50     43.51     3.58      27.70      44.00      30.79        54        -23.21                       AV       Horizontal




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 24 of 27 -                      Report No.: MTi181018E052


5.3.4.3 Spurious Emission in Restricted Band 3260MMHz-18000MHz
All the modulation modes have been tested, and the worst result was report as below:
                                        Antenn
 Frequenc Readin            Cable                   Preamp Emission               Detecto
                                           a                        Limits Margin
     y    g Level           Loss                     Factor Level                    r
                                        Factor                                            Comment
                                                            (dBμ    (dBμ
   (MHz)       (dBμV)        (dB)        dB/m         (dB)                  (dB)   Type
                                                            V/m)    V/m)
    3260         62.70        4.04       29.57         44.70        51.61          74      -22.39        Pk         Vertical
    3260         55.78        4.04       29.57         44.70        44.69          54       -9.31        AV         Vertical
    3260         65.42        4.04       29.57         44.70        54.33          74      -19.67        Pk       Horizontal
    3260         56.86        4.04       29.57         44.70        45.77          54       -8.23        AV       Horizontal
    3332         64.81        4.26       29.87         44.40        54.54          74      -19.46        Pk         Vertical
    3332         56.98        4.26       29.87         44.40        46.71          54       -7.29        AV         Vertical
    3332         63.84        4.26       29.87         44.40        53.57          74      -20.43        Pk       Horizontal
    3332         50.35        4.26       29.87         44.40        40.08          54      -13.92        AV       Horizontal
   17797         44.50       10.99       43.95         43.50        55.94          74      -18.06        Pk         Vertical
   17797         34.00       10.99       43.95         43.50        45.44          54       -8.56        AV         Vertical
   17788         42.95       11.81       43.69         44.60        53.85          74      -20.15        Pk       Horizontal
   17788         34.88       11.81       43.69         44.60        45.78          54       -8.22        AV       Horizontal




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 25 of 27 -                      Report No.: MTi181018E052

Photographs of the Test Setup
                                                     Radiated emission




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                      - Page 26 of 27 -                      Report No.: MTi181018E052

                                                   Conducted emission




Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China


                                                       - Page 27 of 27 -                      Report No.: MTi181018E052



Photographs of the EUT

See the APPENDIX 1: EUT PHOTO in the report No.: MTi181018E051-1.



                                                 ----END OF REPORT----




 Tel:(86-755)88850135         Fax: (86-755) 88850136                Web: http://www.mtitest.com         E-mail: mti@51mti.com
 Address: No.102A & 302A, East Block, Hengfang Industrial Park, Xingye Road, Xixiang, Bao'an District, Shenzhen, Guangdong, China



Document Created: 2018-11-16 14:23:12
Document Modified: 2018-11-16 14:23:12

© 2024 FCC.report