rf exposure

FCC ID: 2AMRM-YS6060

RF Exposure Info

Download: PDF
FCCID_4102700

RF Exposure Requirements

Product Description: Benchtop Spectrophotometer
Model No.: YS6060, YS6010, YS6020, YS6080, YS6090, YS6002, YS6003, YS6022, YS6023
FCC ID: 2AMRM-YS6060

According to the KDB 447498 D01 v06 section 4.3.1, for 100 MHz to 6 GHz and test separation
distances ≤ 50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the
following:
[(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance,
mm)] ·[√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR, where
- f(GHz) is the RF channel transmit frequency in GHz
- Power and distance are rounded to the nearest mW and mm before calculation17
- The result is rounded to one decimal place for comparison

Calculation Result:

Tx frequency range: 2402-2480MHz
Min. test separation distance: 5mm
Maximum Conducted Output Power: 2.92dBm
Tune-Up output power: 3dBm
RF channel transmit frequency: 2402MHz
Result: 0.6
Limit: 3.0

The exclusion thresholds is 0.6 < 3, so the transmitter complies with the RF exposure
requirements and the SAR is not required.



Document Created: 2018-12-11 14:05:36
Document Modified: 2018-12-11 14:05:36

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC