SAR Probe & Dipole Cal Cert

FCC ID: 2AM6U-SP630

RF Exposure Info

Download: PDF
FCCID_4375150

PROBE CALIBRATION CERTIFICATES


           omm                       in Collaboration with

           @=7TJs       a
            Wimaaeee>"" CALIEBRATION LABORATORY
         Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
         Tel      0—62304633—2512       Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com        Hitpy//www.chinattl.on


Glossary:
TSL                    tissue simulating liquid
 NORMx,y,z             sensitivity in free space
ConvF                  sensitivity in TSL / NORMx,y,z
DCP                    diode compression point
CF                     crest factor (1/duty_cycle) of the RF signal
A,B,C,D                modulation dependent linearization parameters
Polarization ®         ® rotation around probe axis
Polarization 8         9 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                       8=0 is normal to probe axis
Connector Angle        information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
   Measurement Techniques®, June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)*,
     July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e     NORMx,y,z: Assessed for E—field polarization 8=0 (fs900MHz in TEM—cell; f>1800MHz: waveguide).
      NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
      E" —field uncertainty inside TSL (see below ConvF).
e     NORM(Mx,y,2 = NORMx,y,2* frequency_response (see Frequency Response Chart). This
      linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
      frequency response is included in the stated uncertainty of ConVF.
e     DCPxX,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
      (no uncertainty required). DCP does not depend on frequency nor media.
e     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
      characteristics.
e     Axy2 Bxyz; Coy,z:VRxy,z:A,B,C are numerical linearization parameters assessed based on the
      data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
      media. VR is the maximum calibration range expressed in RMS voltage across the diode.
«_    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
      Transfer Standard for fs800MHz) and inside waveguide using analytical field distributions based on
      power measurements for f >800MHz. The same setups are used for assessment of the parameters
      applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
      These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
      The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
      that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
      allows extending the validity from+50MHz to+100MHz.
e     Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
      phantom exposed by a patch antenna.
e     Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
      probetip (on probe axis). No tolerance required.
e     Connector Angle: The angle is assessed using the information gained by determining the NORMx
      (no uncertainty required).

Certificate No: Z18—60353                               Page 2 of11


                                    In Collsboration with
                            TL s           p_     e         a    q
                                    CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
        A               1633—2512                      304633—2504
                        attl.com         Hitp://wwwchinattl.en




                       Probe EX3DV4


                                                SN: 7329


                                    Calibrated: September 30, 2018

                              Calibrated for DASY/EASY Systems
                                       (Nete: non—compatible with DASY2 system!)




Certificate No: Z18—60353                              Page 3 of11




              se© ; Collsboration with
             @77]J,    a
             CWimaeee" CALIBRATION LABORATORY
           Add:   No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel        0—62304633—2512       Fax: +86—10—62304633—2504
           E—mail: cttl@chinatt.com         Hittpy//www.chinattl.en



       DASY/EASY —— Parameters of Probe: EX3DV4 — SN: 7329

Calibration Parameter Determined in Head Tissue Simulating Media
                        j                        i                                                          6
 f MHz]° P::;l""’;y § C°"‘:;/°;:‘)":y ConvF X ConvF Y ConvF Z Alpha$ E;::) :‘:(:‘;t)
    750            419                  0.89              i0.01_|         i0.01_|   _i0.01_|   040   0.80       +121%
    900            415                  0.97               966             g65       9656      o16   139        +121%
   1750            40.1                 1.37               8.35            835       835       024   1.05       +121%
   1900            40.0                 1.40               810             810       810       030   095        |+121%
   2450            392                  1.80               762             162       762       o64   070        +121%
   2600            39.0                 1.96               738             138       138       o60   072        +121%
   5200            36.0                 4.66               552             552       552       045   120        +133%
   5300            350                  4.76               528             Ss28      528       045   120        +133%
   5600            355                  5.07               am              4am       47        050   120        +133%
   5800            35.3                 527                468             468       468       050   150        £133%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, thevalidity of tissue parameters(€ and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




  Certificate No: Z18—60353                                Page 5 of 11


                                       In Collaboration with
             77/                       s_     p      _e_       a    q
              i                        CALIBRATION LABORATORY
           Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512         Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com         Hup:/wwas.chinattl.en


       DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7329

Calibration Parameter Determined in Body Tissue Simulating Media
                 Relative          Conductivi                                                               C              ;
 PMEA pormngvity ®                          (om) Y     conyF X ConvF Y ConvF 2 Alpha® ':;:1":)                  (li':;t)
    750            555                  0.96               1023          1023   1023      040        0.80       £121%
    900            55.0                 1.05               a.79          a.79   979       023        125        £121%
   1750            53.4                 1.49               8.05          8.05   8.05      025        1.05       +121%
   1900            CFF                  1.52               7.70          7.70   T70       om         115        £121%
   2450            52.7                 1.95               T47           7.47   7ar       056        083        £121%
   2600            825                  216                712           7.12   712       065        0.72       £121%
   5200            49.0                 5.30               4.92          4.92   492       050        150        £133%
   5300            48.9                 5.42               4.79          4.79   47g9      0.50       1.50       £13.3%
   5600            48.5                 5.77               4.14          474    414       0.60       140        £133%
   5800            48.2                 6.00               4.37          437    437       0.60       1.35       £133%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to £ 110 MHz.
F At frequency below 3 GHz, the validity of tissue parameters (e and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (€ and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvrF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than x 1% for frequencies below 3 GHz and below + 2%for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




  Certificate No: Z18—60353                                Page 6 of11


                                         _                   In Collaboration with
                                x7"7J, a
                                         ~Wimzeee>""         CALIBRATION LABORATORY
              Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
              Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
               E—mail: ctl@chinatt.com                            Hitps/www.chinattl.en



                                                       Frequency Response of E—Field
                                                       (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normalized)




                                          0.5                T              T                      iO
                                                0           500           1000            15I00    2000         2500   3000

                                                                  ces                     f [MHz          =%=

                                                    Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z18—60353                                                       Page 7 of 11


               sw\© 1. Collsboration with
              ©7‘77/,s_ p e a q
               s                            CALIBRATION LABORATORY
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2512         Fax: +86—10—62304633—2504
            E—mail: cttl@chinatt.com           Hitpy//www.chinattl.on



                                   Receiving Pattern (®), 0=0°

                  f=600 MHz, TEM                                                    f=1800 MHz, R22
Eror{dB]




                     t        f         t                t        r         .       +     y           t           .
                   —150                —100             —50                 0            50          100         150
                                                                        Rollf*]
                             [—«*—1100MHz             __—+—600MHz               __—+— 1800MHz   __—«— 2500MHz]
                                  Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




    Certificate No: Z18—60353                                 Page 8 of11


                                           In Collaboration with




        Tel:   +86—10—62304633—2512                        +86—1 0-62]6J633-25()4
        E—mail: c           hinattl.com              Hitp//www.chinattl.en



                                          Dynamic Range f(SARneaq)
                                                        (TEM cell, f= 900 MHz)
                      10° 4               :



                      10° 4



               E 10° 4
               C
               m
               2
               in
               § 1054
               2.
               £

                      104



                      10°                                                                      f
                            10°                      10"                     10°           10‘            10°          10°
                                                                             SAR[mW/cm‘]
                                                              EHX}not compensated      —@— compensated
                 2




           <
           £     0

           8
           i
                a1


                %#4                            fo—      tfi           dR            p   LRB ...b           —          hk
                     10°                      10°                      10°              , 10             10°          10°
                                                                          SAR[mW/cm‘]
                                   [___C®—Inot compensated             —#— compensated                          _]
                                     Uncertainty of Linearity Assessment: £0.9% (k=2)
Certificate No: Z18—60353                                           Page 9 of 11


             se© 1. Collsboration with
          @77J,s     a
           \\Wimzee" CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512                        Fax: +86—10—62304633—2504
        E—mail: cttl@chinatt.com                          Hitp:/fwww.chinattl.en



                                      Conversion Factor Assessment
       f=750 MHz, WGLS R9(H_convF)                                                           f=1750 MHz, WGLS R22(H_convF)
                       350                                               ‘                                3000                I


                                                                    |    ‘                                2500
                                           |                    |                                                     t |
                   B
                             —*7




                                                                                                          2000                                        ‘




                                                                                                                      k***
           SARWikg/W




                                                                                              SARIWkgDW
                   §




                                                                                                          8

                                                                                                                       La»**"
                                   _




                       aso

                       aco   oeH Sss
                             a             2        «
                                                        {mm}
                                                               co       so         100                           o0          dn    m     n
                                                                                                                                             ztum}
                                                                                                                                                     4n   m   sn   o
                                           _—anatical


                                   Deviation from Isotropy in Liquid
                                   2 Iis




                                                     10     .00 —o%0 .040 .020           0        20             o40     080      080   10
                                               Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z18—60353                                               Page 10 of 11


          w® in Collsboration with
          ©7/"7‘J,s p e a g
           SWmm                      CALIBRATION LABORATORY
            :   No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                   10—62304633—2512             86—10—62304633—2504
            ail: ettl@chinattl.com        Hitpo//www.chinattl.en


    DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7329


     Other Probe Parameters
     Sensor Arrangement                                                        Triangular

     Connector Angle (°)                                                            44.7

     Mechanical Surface Detection Mode                                           enabled

     Optical Surface Detection Mode                                              disable

     Probe Overall Length                                                        337mm

     Probe Body Diameter                                                           10mm

    Tip Length                                                                      9mm

    Tip Diameter                                                                  2.5mm

    Probe Tip to Sensor X Calibration Point                                         1mm

    Probe Tip to Sensor Y Calibration Point                                         1mm

    Probe Tip to Sensor Z Calibration Point                                         1mm

    Recommended Measurement Distance from Surface                                 1.4mm




Certificate No: Z18—60353                               Page 11 of 11


DIPOLE CALIBRATION CERTIFICATES


                                    In Coaboration with
                TTL 2.
        Add: No.S1 Xueywsn Rond, nidian Distict Being. 100191, China
        Tel +B6—0—623046—200          ravc«R—2304e3.250¢
        E—mail: ctl@rchinat.com       bups hvwchinatt en

Glossary:
TSL                     tissue simulating liquid
ConvF                       sensitivity in TSL / NORMxy,z
N/A                     not applicable or not measured


Calibrationis Performed According to the Following Standards:
a) IEEE Sid 1528—2013, ‘IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
    Communications Devices: Measurement Techniques®, June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
    exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)*, July 2016
c) IEC 62209—2, *Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)*, March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
€) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
*    Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. Allfigures stated in thecertificate are valid at the frequency indicated.
*    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
*    Feed Point Impedance and Retur Loss: These parameters are measured with the dipole
     positioned under the liquid filed phantom. The impedance stated is transformed from the
     measurement at the SMA connectorto the feed point. The Return Loss ensureslow
     reflected power. No uncertainty required
*    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
*    SAR measured: SAR measured at the stated antenna input power.
*    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector
*    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurementis stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.




Certificate No: 218—60218                        Page 2 of%


                                   in Colaboration with
               TTL a
       Add: No 1 Xueyuan Rond, Haidian Distct Beling. 100191; China
       Tel +86—10—2304633—2079      Fac «i6—10420433—2504
        E—mal: t@chinatcom           hipohivowctinaton
Measurement Conditions
   DASY system confiquration, as far as not given onpage 1.
    DASY Varsion                                     pasvse                                   s2.10.1.1478
    Extrapolation                            Advanced Extrapolation
    Phantom                                  Triple Fiat Phantom §.1C
    Distance Dipole Center — TSL                     10 mm                                  with Spacer
    Zoom Scan Resolution                        d dy, oz = mm
    Frequency                                   2480 MBz x 1 Miiz

Head TSL parameters
   The folowing parameters and calculations were applied
                                                    Temporature               Pormitivity           Conguctivity
     Nominal Head TSL parameters                      220°C                      s02                1.80 mhoim
     Measured Head TSL paramotors                  @aoz02)‘c                   10326 %            1.84 mhoim £6 %
     Head TSL temperature changeduring test)          <1.0 °C                     —                     —
SAR result with Head TSL
    SAR averaged over 1 _cm"_(t g) of Head TSL                    Congition
    SAR measured                                              250 mW input power                134 mW/g
    SAR for nominal Head TSL parameters                        normalized to 1W        83.3 mW /g £18.8 % (Ke2)
    SAR averaged over 10 cn:"(10 g) of Head TSL                   Condiion
    SAR measured                                              250 mW input power             828 mW /g
    SAR for nominal Head TSL parameters                        normalized to 1W        26.0 mW /g £ 18.7 % (Ke2)
Body TSL parameters
   Te folowing parameters and calculations were applied
                                                   Temperature                Pormitivity           Conductivity
    Nominal Body TSL parameters                      220°C                       s27                1.95 mhoim
    Measured Body TSL parameters                  @aoz02°c                     54128%             1.62 mhoim +6 %
    Body TSL temperature change during test)         <1.0 °C                      —                     —
SAR result with Body TSL
    SAR averaged over 1_ctm_(1 g) of Body TSL                     Congtion
    SAR measured                                              250 mW input power             122 mW1g
    SAR for nominal Body TSL parameters                        normalized to 1W        49.8 mW /iq £ 18.% (K=2)
    SAR averaged over 10 m‘ (10.g)of Body TSL                     Condition
    SAR measured                                              250 mW input powe:             5.68 mW /g
    SAR for nominal Body TSL parameters                        normalized to 1W        22.9 mW /g £ 18.7 % (k=2)

Certiicate No: 18—60218                          Page s ofs


        Add: No.51 Xueyuan Road, Haidan DistictBeling. 100191, China
        Tok +86—10—62304003—2079..    P «Be—1O—23043—2504
        E—mail: ctl@rchinatt.com      hipelhnchinatten

  Appendix (Additional assessments outside the scope of CNAS L0570)

 Antenna Parameters with Head TSL

     Impedance,transformed to feed point                               se10+e.3tn
     Retun Loss                                                          «22008

 Antenna Parameters with Body TSL

     Impedance,tansformed to feed point                                soso+7.s3n
     Retum Loss                                                         —22408

 General Antenna Parameters and Design

     Electical Delay (one cirection)                                    1020 ns


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is direcily
 connected to the second arm of the dipole. The antennais therefore short—cirouited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the *Measurement Conditons" paragraph. The SAR data are not
 affected by this change. The overall dipole length is stil according to the Standard
 No excessive force must be applied to the dipole arms, becausethey might bend or the soldered
 connections near the feedpoint may be damaged.

 Additional EUT Data

     Manufactured by                                      3                speac




Certificate No: Z18—60218                        Page 4 of s


          r\'                        in Colsboration with
          e= 7TL a
           \/                        causranon Lasorarory
        Add: No 1 Xueyuan Road, Haidian Distict Beling100191; China
        Tel +86—10—62304003—2070..     PaxceBe—l0.23046832504
        E—mail ctl@chinat.com          hipAwnechinati.en

DASYS Validation Report for Head TSL                                              Date: 06.26,2018
Test Laboratory: CTTL, Beifing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 971
    Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
    Medium parameters used:      2450 MHz; 0 = 1.844 $/m; e,          40.25; p = 1000 ke/m?
    Phantom section: Center Section
DASYS Configuration:

            *   Probe: EX3DV4 — SN7464; ConvF(7.89, 7.89, 7.89) @ 2450 MHz; Calibrated:
                9122017
            +   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +   Electronies: DAE4 Sn1524; Calibrated: 9/13/2017
            +   Phantom: MEP_V5.1C ; Type: QD 000 PSICA; Serial: 1062
            *   Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11
                (7439)

        Dipole Calibration/ZoomScan (7x7x7) (7x7x7)/Cube 0: Measurement grid
         y—Smm, dz=Smm
        Reference Value = 94.13 V/m; Power Drift = 0.01 dB
        Peak SAR (extrapolated) = 27.4 Wikz
        SARCI g)= 134 Wikg: SAR(IO g)=6.26 Wike
        Maximumvalue of SAR (measured) = 22.0 We


           «8
            8         |

            ~4.34


            —a.co

            43.02


           47.36

                          C:
           21.70
                      0 dB =22.0 W/kg = 13.42 dBW/g




Certificate No: Z18—60218                         Page s ofs


            fl‘TTL M
          Add: No 51 Xueyuan Road. Haidian Distict, Beijng, 100191, China
           Tok +86—10—62304033—2079      racceBe—lOaz30i6n
          E—mail:clchinacom              hupohwwchinatl.en




          Impedance Measurement Plot for Head TSL


        Te stttoyrap 10. 00007 ner a.oonde T
                x —mamus on cmmar e




        BB sis sntch (@efo) scate 2.0000 (r2 o1]
         on acaso0000 one. se.ore n a.3000 o. 400,70—p07




    I                                                       %
    Iremessow                                            L




Certificate No: Z18—60218                            Page 6 of s


          ‘r\‘                      in Collbortion with
                TTL a              causranon Lasorarory

        Add: No.51 Xueyuan Road, Haidian Distic, Belng 100191, China
        Tel +86—10—62304603.2079      P «8e—lO—62304e83.2504
        E—mail cul@@chinatcom         hipe/hncwchinatl.en

DASYS Validation Report for Body TSL                                     Date: 06.
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 971
  Communication System: UID 0, CW; Frequency: 2450 MHz; DutyCycle: 1:1
  Medium parameters used: £= 2450 MHz; 0 = 1.92 S/m; e, = 54.06; p = 1000 kg/m?
   Phantom section: Right Section
DASY5 Configuration:

            *   Probe: EX3DV4 — SN7464; ConvF(8.09, $.09, 8.09) @ 2450 MHz; Calibrated:
                971272017
            +   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +   Electronies: DAE4 Sn1524; Calibrated: 9/13/2017
            *   Phantom: MFP_V5.1C ; Type: QD 000 PSICA; Serial: 1062
            +   Measurement SW; DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11
                (7439)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=Smm,
        dy=Smm, dz=Smm
        Reference Value = 90.43 V/m; Power Drift=—0.01 dB
        Peak SAR (extrapolated     5.1 Wkg
        SARCI g) = 12.2 W/kgs SAR(1O g)= 5.68 We
        Maximum value of SAR (measured) = 20.1 Wike

           «8        [
           0


           4. |
           —8.94


           13.40


           417.07

                         C
           2234
                     0 dB =20.1 Wikg = 13.03 dBW/kg




Certifiate No: 218—60218                         Page 7 ofs


            “TTL p
                 s _e_a
                 causranion Lagorerory
          Add: No.51 Xueyuan Road, Hoidian Disvict, BJing, 100191, China
          Tok +86—1062304803—2070        P ste—10420IG32sO
          E—mal: ctl@tchinat.com         hipohivewchinatln

          Impedance Measurement Plot for Body TSL


                                  rer s.cooge Te
                                          ras




                                                                                     x
                                                                    4


          ce                                                        +
         im 5is setth (Rp9 Scate 1.0000 (Fs oel]
          m ncasoooso one. So.ser 0 7eo. ass.giop




     f                                                       *«
     reutnzson                                           weuntore          se aas nioi




Certificate No: 218—60218                            Page 8 ofs



Document Created: 2019-07-26 17:49:56
Document Modified: 2019-07-26 17:49:56

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC