I18Z62361-SEM02_SAR_Rev0_208-221

FCC ID: 2AJOTTA-1149

RF Exposure Info

Download: PDF
FCCID_4178884

n
S
(I
     DASY5 Validation Report for Body TSL

                                                                                          Date: 16.07.2018
     Test Laboratory: SPEAG, Zurich, Switzerland

     DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

     Communication System: UID 0 — CW; Frequency: 2450 MHz
     Medium parameters used: f = 2450 MHz; 0 = 2.02 S/m; &, = 51.9; p = 1000 kg/m‘
     Phantom section: Flat Section
     Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

     DASY52 Configuration:

         *    Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

         e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

         *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

         «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


     Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
     Measurement grid: dx=5mm, dy=5mm, dz=5mm
     Reference Value = 108.0 V/m; Power Drift = —0.03 dB
     Peak SAR (extrapolated) = 25.6 W/kg
     SAR(I g) = 13.1 W/kg; SAR(10 g) = 6.1 W/kg
     Maximum value of SAR (measured) = 21.0 W/kg


              dB
              0


              ~4.00


              —8.00


              —12.00


              ~16.00


              —20.00
                        0 dB = 21.0 W/kg = 13.22 dBW/kg




     Certificate No: D2450V2—853_Jul18               Page 7 of 8


Impedance Measurement Plot for Body TSL




     Eile View Channel Sweep Calibration Trace Scale Marker System Window Help .
                                                                          2450000 GHz       48.464 O
                                                                             389.27 pH      5.9924 Q
                                                                          2450000 GHz     62.710 mU
                                                                                             100.89 °




                   Ch1Avg= 20
          Ch1: Start 2.25000 GHz   ——                                                     Stop 2.65000 GHz

         .00
        00
                                                                          2450000   GHz   —24.053 dB
        .00
        .00




         00

         00
         00
         00
         00       Ch 1     =
          Chi: Start 2.25000 GHz   ——

       Status    _ CH 1:




Certificate No: D2450V2—853_Jul18                 Page 8 of 8


. (El)
             .       5                                                         w1 0j
         Calibration Laboratory of                                         Q\\“\\\/\\\_/’//"/,;                     g        Schweizerischer Kalibrierdienst
         Schmid & Partner                                                 i‘afi&%                                    c       Service suisse d‘étalonnage
           Engineering AG                                                  ud                                               Servizio svizzero di taratura
         Zeughausstrasse 43, 8004 Zurich, Switzerland                      *g/I//\\\\\\\“                           S       swiss Calibration Service
                                                                               "dhilals


         Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
         The Swiss Accreditation Service is one of the signatories to the EA
         Multilateral Agreementfor the recognition of calibration certificates

         Client       CTTL (Auden)                                                                        Certificate No: D2600V2—1012_Jul18

         CALIBRATION CERTIFICATE
          Object                                D2600V2 — SN:1012


          Calibration procedure(s)              QA CAL—05.v10
                                                Calibration procedure for dipole validation kits above 700 MHz



          Calibration date:                     July 26, 2018


          This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
          The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


          All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


          Calibration Equipment used (M&TE critical for calibration)


          Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
          Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
          Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
          Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
          Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
          Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
          Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
          DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


          Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
          Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
          Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                                 Name                                       Function                              Signature
          Calibrated by:                         Michael Weber                              Laboratory Technician                         H-



          Approved by:                           Katia Pokovic                              Technical Manager                   ///%’C




                                                                                                                                 Issued: July 26, 2018
          This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


         Certificate No: D2600V2—1012_Jul18                                     Page 1 of 8


(IEI)
            P      s                                                    w4 uj
        Callbl.'atlon Laboratory of                                  Q‘QM?Z           S   Schweizerischer Kalibrierdienst
        Schmid & Partner                                            m                 c   Service suisse d‘étalonnage
          Engineering AG                                            s o amls              Servizio svizzero di taratura
                                     ;      M
        Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                    x4rosng
                                                                       fan &
                                                                                      S      i      iFSH          M
                                                                                          Swiss Calibration Service
                                                                      Hlilals

        Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
        The Swiss Accreditation Service is one of the signatories to the EA
        Multilateral Agreement for the recognition of calibration certificates

        Glossary:
        TSL                          tissue simulating liquid
        ConvF                        sensitivity in TSL / NORM x,y,z
        N/A                          not applicable or not measured

        Calibration is Performed According to the Following Standards:
           a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
              Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
              Communications Devices: Measurement Techniques", June 2013
           b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
              (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
              300 MHz to 6 GHz)", July 2016
           c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
              communication devices used in close proximity to the human body (frequency range of 30
              MHz to 6 GHz)", March 2010
           d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

        Additional Documentation:
           e) DASY4/5 System Handbook

        Methods Applied and Interpretation of Parameters:
           e Measurement Conditions: Further details are available from the Validation Report at the end
             of the certificate. All figures stated in the certificate are valid at the frequency indicated.
           e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
             point exactly below the center marking of the flat phantom section, with the arms oriented
             paralle! to the body axis.
           e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
             positioned under the liquid filled phantom. The impedance stated is transformed from the
             measurement at the SMA connector to the feed point. The Return Loss ensures low
             reflected power. No uncertainty required.
           e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
             No uncertainty required.
           e SAR measured: SAR measured at the stated antenna input power.
           e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
                 connector.
            e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
                 nominal SAR result.
          The reported uncertainty of measurement is stated as the standard uncertainty of measurement
          multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
          probability of approximately 95%.




        Certificate No: D2600V2—1012_Jul18                              Page 2 of 8


3
l
                                                                                                   oo —              D2   _




    Measurement Conditions
       DASY system configuration, as far as not given on page 1.
         DASY Version                                              DASY5                             V52.10.1
         Extrapolation                                   Advanced Extrapolation
         Phantom                                         Modular Flat Phantom
         Distance Dipole Center — TSL                              10 mm                            with Spacer
         Zoom Scan Resolution                              dx, dy, dz =5 mm
         Frequency                                         2600 MHz + 1 MHz



    Head TSL parameters
       The following parameters and calculations were applied.
                                                                   Temperature      Permittivity          Conductivity
         Nominal Head TSL parameters                                  22.0 °C          39.0                1.96 mho/m
         Measured Head TSL parameters                              (22.0 £0.2) °C   37.2 26 %           2.02 mho/m +6 %
         Head TSL temperature change during test                      <0.5 °C           ——                      ——


    SAR result with Head TSL

         SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
         SAR measured                                        250 mW input power                    14.2 W/ikg
         SAR for nominal Head TSL parameters                     normalized to 1W       55.4 Wikg + 17.0 % (k=2)

         SAR averaged over 10 cm* (10 g) of Head TSL                condition
          SAR measured                                       250 mW input power                    6.33 W/kg
          SAR for nominal Head TSL parameters                    normalized to 1W       24.9 Wikg x 16.5 % (k=2)


    Body TSL parameters
       The following parameters and calculations were applied.                                                            .
                                                                   Temperature      Permittivity          Conductivity
          Nominal Body TSL parameters                                 22.0 °C          52.5                2.16 mho/m
          Measured Body TSL parameters                             (22.0 £0.2) °C   51.5 26 %           2.20 mho/m £ 6 %
          Body TSL temperature change during test                     <0.5 °C           =—

    SAR result with Body TSL

          SAR averaged over 1 cm‘ (1 g) of Body TSL                 Condition
          SAR measured                                       250 mW input power                    13.7 W/kg
          SAR for nominal Body TSL parameters                    normalized to 1W        54.1 Wikg #17.0 % (k=2)


          SAR averaged over 10 em* (10 g) of Body TSL                condition
          SAR measured                                       250 mW input power                    6.17 W/kg
          SAR for nominal Body TSL parameters                    normalized to 1W        24.5 Wikg # 16.5 % (k=2)




    Certificate No: D2600V2—1012_Jul18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 01 08)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           474 Q —7.4 jQ
      Return Loss                                                                       —21.9 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           44.1 Q — 4.9 jQ
      Return Loss                                                                      —21.8 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.154 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               October 30, 2007




Certificate No: D2600V2—1012_Jul18                        Page 4 of 8


m_
/7/
y                                                                               C




 DASY5 Validation Report for Head TSL

                                                                                     Date: 26.07.2018

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

 Communication System: UID 0 — CW; Frequency: 2600 MHz
 Medium parameters used: f = 2600 MHz; 0 = 2.02 $/m; & = 37.2; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    *    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 118.3 V/m; Power Drift = —0.04 dB
 Peak SAR (extrapolated) = 28.3 W/kg
 SAR(1 g) = 14.2 W/kg; SAR(10 g) = 6.33 W/kg
 Maximum value of SAR (measured) = 23.7 W/kg


        dB
        0

         —5.20

         —10.40

         —15.60

         —20.80

         —26.00
                    0 dB = 23.7 W/kg = 13.75 dBW/kg




 Certificate No: D2600V2—1012_Jul18              Page 5 of 8


Impedance Measurement Plot for Head TSL




File View         Channel Sweep         Calibration Trace Scale   Marker   System   Window   Help

                                                                                            2600000 GHz       47.371 Q
                                                                                                8.2358 pF    —7.4326 Q
                                                                                          , 2600000 GHz     80.734 mU
                                                                                                              ~105.12 °




              Ch1Avg= 20
     Ch1: Start 2.40000 GHz       ——                                                                        Stop 2.90000 GHz


                                                                                                            —2).859 dB




     0       Ch         =
     Ch1: Start    2.40000 GHa    ———                                                                       Stop 2.20000 GHz

  Status      CH 1:         511         ac        C* 1—Port                Avg=20 Delay                          LCL




Certificate No: D2600V2—1012_Jul18                                  Page 6 of 8


m*
/7/
y                                                                               C




 DASY5 Validation Report for Body TSL
                                                                                    Date: 26.07.2018

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

 Communication System: UID 0 — CW; Frequency: 2600 MHz
 Medium parameters used: f = 2600 MHz; 0 = 2.2 S/m; & = 51.5; p = 1000 kg/m3
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

    e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    e   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

    e   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 107.5 V/m; Power Drift = —0.07 dB
 Peak SAR (extrapolated) = 27.7 W/kg
 SAR(I g) = 13.7 W/kg; SAR(10 g) = 6.17 W/kg
 Maximum value of SAR (measured) = 22.6 W/kg


        dB
        0

        —5.20

        —10.40

        —15.60

        —20.80

         —26.00
                    0 dB = 22.6 W/kg = 13.54 dBW/kg




 Certificate No: D2600V2—1012_Jul18             Page 7 of 8


(ISI,°
         Impedance Measurement Plot for Body TSL




          File         View    Channel    Sweep   Calibration    Trace    Scale   Marker   System   Window    Help

                                                                                                             2600000 GHz              44.132 O
                                                                                                                     12.374 pF       —4.9469 C
                                                                                                          y 2600000 GHz            81.418 mU
                                                                                                                                     —136.86 °




                           Ch1Avg= 20
                  Ch1: Start 2.40000 GHz       ——                                                                                  Stop 2.80000 GHz

                  00                                                                                                 000     GHz   —21.786 dB
                 00
            0.00




                  00       Ch1
                  Ch1: Start 2.                                                                                                    Stop 2.90000 GHz

            Status            CH 1:      E11                    | C" 1—Port                Avg=20 Delay                                 LCL




         Certificate No: D2600V2—1012_Jul18                                         Page 8 of 8


                                                               No.I18Z62361-SEM02
                                                                  Page 218 of 221


ANNEX I          Sensor Triggering Data Summary
1, Maximum transmit power reduce process follow below strategy when mobile connect network.
Headset               P-sensor           SAR sensor      TX Power reduce
Insert                     Near                  Near               Yes
Insert                     Near                  Far                 No
Insert                      Far                  Near               Yes
Insert                      Far                  Far                 No
Pull out                   Near                  Near                No
Pull out                   Near                  Far                 No
Pull out                    Far                  Near               Yes
Pull out                    Far                  Far                 No


2, Distance definition
         P-sensor Detect                      Near                   Far
       Distance Detected                   <3cm                    >=5cm


     SAR Sensor Detect                        Near                   Far
              black                       <=16mm                  >16mm
              front                       <=12mm                  >12mm
             bottom                       <=16mm                  >16mm
               top                       Not Detect              Not Detect
              right                      Not Detect              Not Detect
               left                      Not Detect              Not Detect


3, Reduction and Bands
                                  Maximum conduct
 Band         Requirement          power reduction
 DCS          2、3、4 Slots              2.5dB
 PCS          2、3、4 Slots              2.5dB
  W1                                    3dB
  W2                                   4.5dB
  W4                                    4dB
  L1                                   2.5dB
  L2                                   2.5dB
  L3                                   2.5dB
  L4                                    1dB
  L7                                    1dB
  L66                                   2dB




                                                            ©Copyright. All rights reserved by CTTL.


                                                                                          No.I18Z62361-SEM02
                                                                                             Page 219 of 221

      According to the above description, this device was tested by the manufacturer to determine the
      SAR sensor triggering distances for the rear and bottom edge of the device. The measured power
      state within ±5mm of the triggering points (or until touching the phantom) is included for rear and
      each applicable edge.

      To ensure all production units are compliant it is necessary to test SAR at a distance 1mm less than
      the smallest distance from the device and SAR phantom with the device at maximum output power
      without power reduction.

      We tested the power and got the different proximity sensor triggering distances for rear and bottom
      edge. But the manufacturer has declared 16mm is the most conservative triggering distance for
      main antenna. So base on the most conservative triggering distance of 16mm, additional SAR
      measurements were required at 15mm from the highest SAR position between rear and bottom
      edge of main antenna.

      Rear
      Moving device toward the phantom:
                                                       The power state
Distance [mm]      21         20             19          18          17         16        15         14          13       12         11
Main antenna      Normal     Normal     Normal         Normal       Normal      Low       Low        Low        Low      Low        Low


      Moving device away from the phantom:
                                                        The power state
 Distance [mm]      11       12         13        14       15        16         17          18             19           20          21
  Main antenna     Low       Low      Low         Low     Low        Low       Normal     Normal      Normal          Normal       Normal


      Bottom Edge
      Moving device toward the phantom:
                                                         The power state
  Distance [mm]         21         20             19          18          17         16        15         14       13         12         11
   Main antenna     Normal     Normal         Normal     Normal       Normal      Low       Low        Low        Low        Low      Low


      Moving device away from the phantom:
                                                         The power state
  Distance [mm]      11       12         13        14         15       16         17            18          19           20           21
   Main antenna     Low       Low       Low        Low        Low     Low       Normal     Normal         Normal       Normal       Normal




                                                                                       ©Copyright. All rights reserved by CTTL.


                                                                    No.I18Z62361-SEM02
                                                                       Page 220 of 221

The influence of table tilt angles to proximity sensor triggering is determined by positioning each
edge that contains a transmitting antenna, perpendicular to the flat phantom, at the smallest sensor
triggering test distance by rotating the device around the edge next to the phantom in ≤ 10°
increments until the tablet is ±45°or more from the vertical position at 0°.


                           Re
                             ar


                                              Body Phantom
                            45°
                                                  16mm
                 0°

                            45°
                      ar
                   Re




                                The rear evaluation for main antenna




                                  Body Phantom
                                                    16mm




                                          45°         45°



                           Top edge             0°             Top edge

                          The bottom edge evaluation for main antenna
Based on the above evaluation, we come to the conclusion that the sensor triggering is not released
and normal maximum output power is not restored within the ±45°range at the smallest sensor
triggering test distance declared by manufacturer.


                                                                 ©Copyright. All rights reserved by CTTL.


                                         No.I18Z62361-SEM02
                                            Page 221 of 221


ANNEX J   Accreditation Certificate




                                      ©Copyright. All rights reserved by CTTL.



Document Created: 2019-02-06 10:23:06
Document Modified: 2019-02-06 10:23:06

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC