Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/10/18


#01_HAC_E_GSM850_GSM Voice_Ch128


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 824.2 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 50.49 V/m; Power Drift =—0.11 dB
Applied MIF = 3.63 dB
RF audio interference level = 35.30 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4  Grid 3 M4
                                   33.02 dBV/m|34.88 dBV/m |34.85 dBV/m
                                   Grid 4 M4    Grid 5 M4             Grid 6 M4
                                   33.25 dBV/m|35.3 dBV/m             |35.25 dBV/m
                                   Grid 7 M4    Grid 8 M4   Grid 9 M4
                                   33.12 dBV/m|34.99 dBV/m|34.98 dBV/m
                                                          Cursor:
                                                   Total = 35.30 dBV/m
                                                     E Category: M4
                                               Location: —6.5, —0.5, 8.7 mm




        —1.08


        —2.16


        —3.24


        —4.32


        —5.40       T
                    0 dB = 58.24 V/m = 35.30 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/10/18


#02_HAC_E_GSM850_GSM Voice_Ch189


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 836.4 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 52.51 V/m; Power Drift = —0.04 dB
Applied MIF = 3.63 dB
RF audio interference level = 35.72 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4   Grid 3 M4
                                   33.76 dBV/m|35.42 dBV/m|35.4 dBV/m
                                   Grid 4 M4    Grid 5 M4   Grid 6 M4
                                   33.75 dBV/m|35.72 dBV/m|35.7 dBV/m
                                   Grid 7 M4           Grid 8 M4      Grid 9 M4
                                   33.3 dBV/m          |35.35 dBV/m|35.35 dBV/m
                                                          Cursor:
                                                   Total =35.72 dBV/m
                                                     E Category: M4
                                                 Location: —7, —2, 8.7 mm




        —1.06


        —2.12


        —3.19


        —4.25


        —5.31       T
                    0 dB = 61.06 V/m =35.72 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/10/18


#03_HAC_E_GSM850_GSM Voice_Ch251


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 848.8 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 49.86 V/m; Power Drift =—0.03 dB
Applied MIF = 3.63 dB
RF audio interference level = 35.34 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4  Grid 3 M4
                                   33.16 dBV/m|35.04 dBV/m |35.04 dBV/m
                                   Grid 4 M4    Grid 5 M4   Grid 6 M4
                                   32.81 dBV/m|35.34 dBV/m|35.34 dBV/m
                                   Grid 7 M4           Grid 8 M4      Grid 9 M4
                                   32.2 dBV/m          |34.92 dBV/m|34.98 dBV/m
                                                          Cursor:
                                                   Total = 35.34 dBV/m
                                                     E Category: M4
                                                 Location: —9, —2, 8.7 mm




        —1.39


        —2.171


        —4.16


        —5.54


        —6.93       T
                    0 dB = 58.49 V/m = 35.34 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#04_HAC_E_GSM1900_GSM Voice_Ch512


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 1850.2 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 28.13 V/m; Power Drift = 0.03 dB
Applied MIF = 3.63 dB
RF audio interference level = 28.59 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4   Grid 3 M4
                                   26.39 dBV/m|28.06 dBV/m|27.96 dBV/m
                                   Grid 4 M4    Grid 5 M4   Grid 6 M4
                                   24.35 dBV/m|28.25 dBV/m|28.02 dBV/m
                                   Grid 7 M4    Grid 8 M4   Grid 9 M4
                                   27.76 dBV/m|28.59 dBV/m|28.59 dBV/m
                                                          Cursor:
                                                   Total = 28.59 dBV/m
                                                     E Category: M4
                                                 Location: —8, 25, 8.7 mm




        —2.20


        —4.39


        —6.59


        —8.78


        —10.38      [T
                    0 dB = 26.87 V/m = 28.59 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#05_HAC_E_GSM1900_GSM Voice_Ch661l


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 1880 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 27.44 V/m; Power Drift = —0.04 dB
Applied MIF = 3.63 dB
RF audio interference level = 28.54 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4   Grid 3 M4
                                   26.08 dBV/m|27.87 dBV/m|27.81 dBV/m
                                   Grid 4 M4    Grid 5 M4   Grid 6 M4
                                   23.65 dBV/m|27.67 dBV/m|27.26 dBV/m
                                   Grid 7 M4           Grid 8 M4      Grid 9 M4
                                   27.63 dBV/m|28.54 dBV/m|28.53 dBV/m
                                                          Cursor:
                                                   Total = 28.54 dBV/m
                                                     E Category: M4
                                                Location: —7.5, 25, 8.7 mm




        —2.07


        —4.14


        —6.22


        —8.29


        410.36      [T
                    0 dB = 26.72 V/m = 28.54 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#06_HAC_E_GSM1900_GSM Voice_Ch810


Communication System: GSM—FDD (TDMA, GMSK); Frequency: 1909.8 MHz;Duty Cycle: 1:8.6896
Medium: Air Medium parameters used: 0 = 0 S/m, g, = 1; p = 0 kg/m3
Ambient Temperature : 23.5 °C

DASY5 Configuration:
— Probe: ER3DV6 — SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
— Sensor—Surface: (Fix Surface)
— Electronics: DAE4 Sn854; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan — ER3D:; 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 27.33 V/m; Power Drift =—0.01 dB
Applied MIF = 3.63 dB
RF audio interference level = 28.48 dBV/m
Emission category: M4

                                   MIF scaled E—field

                                   Grid 1 M4    Grid 2 M4   Grid 3 M4
                                   26.14 dBV/m|27.73 dBV/m|27.49 dBV/m
                                   Grid 4 M4    Grid 5 M4   Grid 6 M4
                                   23.99 dBV/m|27.75 dBV/m|27.3 dBV/m
                                   Grid 7 M4    Grid 8 M4   Grid 9 M4
                                   26.99 dBV/m|28.47 dBV/m|28.48 dBV/m
                                                          Cursor:
                                                   Total = 28.48 dBV/m
                                                     E Category: M4
                                                 Location: —9, 25, 8.7 mm




        —2.00


        —4.00


        —5.99


        —£.99


        —9.99       T
                    0 dB = 26.54 V/m = 28.48 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#07_HAC_E_LTE Band 38_20M_QPSK_1_0_Ch37850


Communication System: LTE-TDD (SC-FDMA, 1 RB, 20 MHz, QPSK); Frequency: 2580 MHz;Duty
Cycle: 1:8.33681
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - ER3D: 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 11.51 V/m; Power Drift = -0.16 dB
Applied MIF = -1.62 dB
RF audio interference level = 23.59 dBV/m
Emission category: M4

                                   MIF scaled E-field
                                   Grid 1 M4           Grid 2 M4     Grid 3 M4
                                   21.33 dBV/m 21.17 dBV/m 22.16 dBV/m
                                   Grid 4 M4           Grid 5 M4     Grid 6 M4
                                   17.98 dBV/m 20.7 dBV/m 21.52 dBV/m
                                   Grid 7 M4   Grid 8 M4   Grid 9 M4
                                   23.21 dBV/m 23.59 dBV/m 22.41 dBV/m
                                                        Cursor:
                                                  Total = 23.59 dBV/m
                                                    E Category: M4
                                                Location: 3.5, 25, 8.7 mm




                    0 dB = 15.13 V/m = 23.60 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#08_HAC_E_LTE Band 38_20M_QPSK_1_0_Ch38000


Communication System: LTE-TDD (SC-FDMA, 1 RB, 20 MHz, QPSK); Frequency: 2575 MHz;Duty
Cycle: 1:8.33681
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - ER3D: 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 8.565 V/m; Power Drift = 0.01 dB
Applied MIF = -1.62 dB
RF audio interference level = 22.15 dBV/m
Emission category: M4

                                   MIF scaled E-field
                                   Grid 1 M4           Grid 2 M4     Grid 3 M4
                                   21.94 dBV/m         19.99 dBV/m   21.15 dBV/m
                                   Grid 4 M4           Grid 5 M4     Grid 6 M4
                                   20.93 dBV/m         19.53 dBV/m   20.8 dBV/m
                                   Grid 7 M4           Grid 8 M4     Grid 9 M4
                                   22.15 dBV/m         21.47 dBV/m   20.26 dBV/m
                                                       Cursor:
                                                 Total = 22.15 dBV/m
                                                    E Category: M4
                                                Location: 25, 25, 8.7 mm




                    0 dB = 12.81 V/m = 22.15 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2018/10/18


#09_HAC_E_LTE Band 38_20M_QPSK_1_0_Ch38150


Communication System: LTE-TDD (SC-FDMA, 1 RB, 20 MHz, QPSK); Frequency: 2610 MHz;Duty
Cycle: 1:8.33681
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - ER3D: 15 mm from Probe Center to the Device/Hearing Aid Compatibility Test
(101x101x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 6.489 V/m; Power Drift = 0.17 dB
Applied MIF = -1.62 dB
RF audio interference level = 21.29 dBV/m
Emission category: M4

                                   MIF scaled E-field
                                   Grid 1 M4           Grid 2 M4     Grid 3 M4
                                   19.98 dBV/m         18.21 dBV/m   19.54 dBV/m
                                   Grid 4 M4           Grid 5 M4     Grid 6 M4
                                   18.14 dBV/m         17.58 dBV/m   19.39 dBV/m
                                   Grid 7 M4           Grid 8 M4     Grid 9 M4
                                   21.29 dBV/m         20.96 dBV/m   18.45 dBV/m
                                                       Cursor:
                                                 Total = 21.29 dBV/m
                                                    E Category: M4
                                                Location: 17, 25, 8.7 mm




                    0 dB = 11.60 V/m = 21.29 dBV/m



Document Created: 2019-03-28 12:35:49