I17Z60313-SEM01_SAR_Rev1_part4

FCC ID: 2AJOTTA-1039

RF Exposure Info

Download: PDF
FCCID_3383627

                                                  No. I17Z60313-SEM01
                                                      Page 151 of 212


ANNEX H        Dipole Calibration Certificate
750 MHz Dipole Calibration Certificate




                                                ©Copyright. All rights reserved by CTTL.


(IEHQ
            s      x                                                  wl 199
        Callbl_'atlon Laboratory of                                S\\\\\\/__/////’;          G    Schweizerischer Kalibrierdienst
        Schmid & Partner                                           M                          c    Service suisse d‘étalonnage
           Engineering AG                                          T mm s                          Servizio svizzero di taratura
                                     ;      F                      * ty
        Zeughausstrasse 43, 8004 Zurich, Switzerland                *g/7/-‘\?\\\\«‘           S    Swiss Calibration Service
                                                                       tlolale

        Accredited by the Swiss Accreditation Service (SAS)                                    Accreditation No.: SCS 0108
        The Swiss Accreditation Service is one of the signatories to the EA
        Multilateral Agreementfor the recognition of calibration certificates

        Glossary:
        TSL                          tissue simulating liquid
        ConvF                        sensitivity in TSL / NORM x,y,z
        N/A                          not applicable or not measured

        Calibration is Performed According to the Following Standards:
           a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
              Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
                 Communications Devices: Measurement Techniques", June 2013
                 IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
                 devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
                 February 2005
            c)   IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
                 communication devices used in close proximity to the human body (frequency range of 30
                 MHz to 6 GHz)", March 2010
            d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

        Additional Documentation:
            e)   DASY4/5 System Handbook

        Methods Applied and Interpretation of Parameters:
             Measurement Conditions: Further details are available from the Validation Report at the end
             of the certificate. All figures stated in the certificate are valid at the frequency indicated.
             Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
             point exactly below the center marking of the flat phantom section, with the arms oriented
             parallel to the body axis.
             Feed Point Impedance and Return Loss: These parameters are measured with the dipole
             positioned under the liquid filled phantom. The impedance stated is transformed from the
             measurement at the SMA connector to the feed point. The Return Loss ensures low
             reflected power. No uncertainty required.
             Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
             No uncertainty required.
             SAR measured: SAR measured at the stated antenna input power.
             SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
                 connector.
                 SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
                 nominal SAR result.
          The reported uncertainty of measurement is stated as the standard uncertainty of measurement
          multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
          probability of approximately 95%.




        Certificate No: D750V3—1017_Jul16                               Page 2 of 8




                                                                                       No Nnn rage men n m rrgge nee n nn se n n ue n n n c   en


<|I§H|}
                                                                                                           ~o          D       s




          Measurement Conditions
             DASY system configuration, as far as not given on page 1.
               DASY Version                                              DASY5                                  V52.8.8
               Extrapolation                                   Advanced Extrapolation
               Phantom                                          Modular Flat Phantom
               Distance Dipole Center — TSL                              15 mm                              with Spacer
               Zoom Scan Resolution                              dx, dy, dz =5 mm
               Frequency                                          750 MHz &1 MHz



          Head TSL parameters
             The following parameters and calculations were applied.
                                                                         Temperature        Permittivity           Conductivity
               Nominal Head TSL parameters                                  22.0 °C            41.9                 0.89 mho/m
                Measured Head TSL parameters                             (22.0 £ 0.2) °C     40.9 x 6 %          0.91 mho/m + 6 %
                Head TSL temperature change during test                     <0.5 °C


          SAR result with Head TSL

               SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition
               SAR measured                                        250 mW input power                      2.13 W/kg
               SAR for nominal Head TSL parameters                     normalized to 1W          8.33 Wikg £ 17.0 % (k=2)

               SAR averaged over 10 cm* (10 g) of Head TSL                 condition
                SAR measured                                       250 mW input power                   1.39 W/kg
                SAR for nominal Head TSL parameters                    normalized to 1W          5.46 Wikg + 16.5 % (k=2)


          Body TSL parameters
             The following parameters and calculations were applied.
                                                                         Temperature        Permittivity           Conductivity
                Nominal Body TSL parameters                                 22.0 °C             $8.5                0.96 mho/m
                Measured Body TSL parameters                             (22.0 £ 0.2) °C     55.1 +6 %           0.99 mho/m + 6 %
                Body TSL temperature change during test                     <0.5 °C             ————                      «e


          SAR result with Body TSL

                SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
                SAR measured                                       250 mW input power                      2.25 W/kg
                SAR for nominal Body TSL parameters                    normalized to 1W          8.78 Wikg x 17.0 % (k=2)

                SAR averaged over 10 cm* (10 g) of Body TSL                condition
                SAR measured                                       250 mW input power                      1.47 Wikg
                SAR for nominal Body TSL parameters                    normalized to 1W          5.76 Wikg #16.5 % (k=2)




          Certificate No: D750V3—1017_Jul16                     Page 3 of 8




                                                                                           Nnn rage en n m en e


®
                                                                                                       Dt e e e en e




Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                             53.8 Q — 0.2 jQ
      Return Loss                                                                          — 28.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                             49.2 0 — 2.3 jQ
      Retum Loss                                                                          — 32.3 dB




General Antenna Parameters and Design

    l Electrical Delay (one direction)                                           L         1.034 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                       SPEAG
      Manufactured on                                                                  March 22, 2010




Certificate No: D750V3—1017_Jul16                        Page 4 of 8




                                                                                     w n vepeg n nage e e n orage isw n nninr n vevr w g n n e es


—m_>
&777
 Tagame"
                                                                                         [ 0C


   DASY5 Validation Report for Head TSL

                                                                                                  Date: 20.07.2016

   Test Laboratory: SPEAG, Zurich, Switzerland

   DUT: Dipole 750 MHz D750V3; Type: D750V3; Serial: D750V3 — SN:1017

   Communication System: UID 0 — CW; Frequency: 750 MHz
   Medium parameters used: £ = 750 MHz; 0 = 0.91 S/m; & = 40.9; p = 1000 kg/m*
   Phantom section: Flat Section
   Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

   DASY52 Configuration:

      e     Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07); Calibrated: 15.06.2016;

      e     Sensor—Surface: 1.4mm (Mechanical Surface Detection)

      e     Electronics: DAE4 Sn601; Calibrated: 30.12.2015

      e     Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

       «_   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


   Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
   Measurement grid: dx=5mm, dy=5mm, dz=5mm
   Reference Value = 58.54 V/m; Power Drift = 0.03 dB
   Peak SAR (extrapolated) =3.21 W/kg
   SAR(I g) = 2.13 W/kg; SAR(10 g) = 1.39 W/kg
   Maximum value of SAR (measured) = 2.84 W/kg




            —2.00

            —4.00

            —6.00

            —8.00

            —10.00
                      0 dB = 2.84 W/kg = 4.53 dBW/kg




   Certificate No: D750V3—1017_Jul16               Page 5 of 8




                                                                         No Nnn rage men n m rrgge nee n nn se n n ue n n n c   en


TTL                                                                                                 ~3 —
z>




Impedance Measurement Plot for Head TSL




                                                                     20 Jul 2016     13:12:13
          CHI) si1     1 U FS          1: 53.848 a   -Zgfl.vgf_mn   890.57 pF        750.000 000 MHz




          fAvg
          16


          H1d

          CH2




          Ca




          fivg
          16


          H1d

                     START 550.000 000 MHz                                 SToP 950.000 a0o MHz




Certificate No: D750V3—1017_Jul16                      Page 6 of 8




                                                                                   sn vpeg n rag in +   rrggn ce   naer n vive w n n es


mm©                                                                     e
///
_                                                                                    ~a >        —      —




 DASY5 Validation Report for Body TSL

                                                                                              Date: 20.07.2016

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 750 MHz D750V3; Type: D750V3; Serial: D750V3 — SN: 1017

 Communication System: UID 0 — CW; Frequency: 750 MHz
 Medium parameters used: f = 750 MHz; 0 = 0.99 $/m; & = 55.1; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

         Probe: EX3DV4 — SN7349; ConvF(9.99, 9.99, 9.99); Calibrated: 15.06.2016;

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.12.2015

    &    Phantom: Flat Phantom 5.0 (back); Type: QDOOOP5OAA; Serial: 1002

    e    DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 57.82 V/m; Power Drift =—0.03 dB
 Peak SAR (extrapolated) = 3.37 W/kg
 SAR(1 g) = 2.25 W/kg; SAR(10 g) = 1.47 W/kg
 Maximum value of SAR (measured) = 3.00 W/kg




         —2.00

         —4.00

         —6.00

         —8.00

         —10.00
                    0 dB = 3.00 W/kg = 4.77 dBW/kg




 Certificate No: D750V3—1017_Jul16              Page 7 of 8




                                                                     No Nnn rage men n m rrgge nee n nn se n n ue n n n c   en


TTL
V




Impedance Measurement Plot for Body TSL




                                                                    20   Jul   2016    @9:12:31
          (CHI sii     4 uors       m49.2320   —23125 n 94.765 pF                     750.000 000 MHz
                                               *%,
          x                                              hig   se

          Del                            4

          Ca
                                    |
                                    cidlel



          fAvq
          16


          H1d                                  .     j

          CH2                                                                         750.000000MHz
                                                                                       pssees se
                                                                                                    _1
                                                                                              +777 —.|




          Ca




          H1d
                                     |
                     6e 550.000 00B MHz                                   ~STOP 950.000000 MHz




Certificate No: D750V3—1017_Jul16                  Page 8 of 8


mm_:
‘/"/‘/
uy




      &     a                                                         wl1w
  Callbl.'atlon Laboratory of                                       i‘i\\_/é‘b}                               S      Schweizerischer Kalibrierdienst
  Schmid & Partner                                                 M                                          c      Service suisse d‘étalonnage
    Engineering AG                                                  prq                                              Servizio svizzero di taratura
  Zeughausstrasse 43, 8004 Zurich, Switzerland                      {',,/7/\\\\\\@                            §      Swiss Calibration Service
                                                                        tlotale

  Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multilateral Agreement for the recognition of calibration certificates

  Client       CTTL—BJ (Auden)                                                                    Certificate No: D835V2—40069_Jul16

  [CALIBRATION CERTIFICATE
   Object                                D835V2 — SN:4d069


   Calibration procedure(s)              QA CAL—05.v9
                                         Calibration procedure for dipole validation kits above 700 MHz



   Calibration date:                     July 20, 2016


   This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
   The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


   All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


   Calibration Equipment used (M&TE critical for calibration)


   Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
   Power meter NRP                        SN: 104778                06—Apr—16 (No. 217—02288/02289)                       Apr—17
   Power sensor NRP—Z91                   SN: 103244                06—Apr—16 (No. 217—02288)                             Apr—17
   Power sensor NRP—Z91                   SN: 103245                06—Apr—16 (No. 217—02289)                             Apr—17
   Reference 20 dB Attenuator             SN: 5058 (20k)            05—Apr—16 (No. 217—02292)                             Apr—17
   Type—N mismatch combination            SN: 5047.2 / 06827        05—Apr—16 (No. 217—02295)                             Apr—17
   Reference Probe EX3DV4                 SN: 7349                  15—Jun—16 (No. EX3—7349_Jun16)                        Jun—17
   DAE4                                   SN: 601                   30—Dec—15 (No. DAE4—601_Dec15)                        Dec—16


   Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
   Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (No. 217—02222)                             In house check: Oct—16
   Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (No. 217—02222)                             In house check: Oct—16
   Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (No. 217—02223)                             In house check: Oct—16
   RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Jun—15)                     In house check: Oct—16
   Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—15)                     In house check: Oct—16


                                          Name                                    Function                                Signature
   Calibrated by:                         Michael Weber                           Laboratory Technician                   M_.
                                                                                                                             e

   Approved by:                           Katia Pokovic                           Technical Manager                  /W;/_



                                                                                                                         Issued: July 22, 2016
   This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


  Certificate No: D835V2—40069_Jul16                                    Page 1 of 8




                                                                                                        No Nnn rage men n m rrgge nee n nn se n n ue n se i


TTL                                                                                |



Calibration Laboratory of                                                     S        Schweizerischer Kalibrierdienst
Schmid & Partner                                                              Cc       Service suisse d‘étalonnage
  Engineering AG                                                                       Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                  S        Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
   e SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D835V2—40069_Jul16                              Page 2 of 8


(lléll|>
           Measurement Conditions
               DASY system configuration, as far as not g iven on page 1.
                 DASY Version                                               DASYS                                    V52.8.8
                 Extrapolation                                   Advanced Extrapolation
                 Phantom                                          Modular Flat Phantom
                 Distance Dipole Center — TSL                               15 mm                                  with Spacer
                 Zoom Scan Resolution                               dx, dy, dz =5 mm
                 Frequency                                          835 MHz + 1 MHz



           Head TSL parameters
               The following parameters and calculations were applied.
                                                                            Temperature            Permittivity         Conductivity
                 Nominal Head TSL parameters                                   22.0 °C                41.5               0.90 mho/m
                 Measured Head TSL parameters                               (22.0 £0.2) °C         40.6 £ 6 %         0.94 mho/m + 6 %
                 Head TSL temperature change during test                       <0.5 °C                 ~——                     _

           SAR result with Head TSL

                 SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
                 SAR measured                                        250 mW input power                           2.45 W/kg
                 SAR for nominal Head TSL parameters                     normalized to 1W              9.44 Wikg £ 17.0 % (k=2)

                 SAR averaged over 10 cm* (10 g) of Head TSL                 condition
                 SAR measured                                        250 mW input power                           1.59 W/kg
                 SAR for nominal Head TSL parameters                     normalized to 1W              6.18 W/kg + 16.5 % (k=2)



           Body TSL parameters
               The following parameters and calculations were applied.
                                                                            Temperature            Permittivity         Conductivity
                 Nominal Body TSL parameters                                   22.0 °C                55.2               0.97 mho/m
                 Measured Body TSL parameters                               (22.0 + 0.2) °C        54.9 # 6 %         1.01 mho/m £ 6 %
                 Body TSL temperature change during test                       <0.5 °C                 ~~—                     ————


           SAR result with Body TSL

                 SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
                 SAR measured                                         250 mW input power                          2.50 W/kg
                 SAR for nominal Body TSL parameters                     normalized to 1W              9.69 W/kg + 17.0 % (k=2)


                 SAR averaged over 10 cm* (10 g) of Body TSL—|—               condition       ——             —
                 SAR measured                                         250 mW input power                          1.63 W/kg
                 SAR for nominal Body TSL parameters                     normalized to 1W              6.36 Wikg 16.5 % (k=2)



           Certificate No: D835V2—40069_Jul16                      Page 3 of 8


®
Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point
      Return Loss
                                                                                    51.9 Q — 21 jQ
                                                                                       —31.1 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          48.8 Q — 2.5 jQ
      Return Loss                                                                      — 31.0 dB



General Antenna Parameters and Design

    IEectrical Delay (one direction)                                                    1.394 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   SPEAG
      Manufactured on                                                             November 09, 2007




Certificate No: D835V2—4d069_Jul16                       Page 4 of 8




                                                                                    ~ c cra   &         v


()
 S
 DASY5 Validation Report for Head TSL



 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 835 MHz D835V2; Type: D835V2; Serial: D835V2 — SN:4d069
                                                                                    Date: 20.07.2016




 Communication System: UID 0 — CW; Frequency: 835 MHz
 Medium parameters used: f= 835 MHz; 0 = 0.94 $/m; s = 40.6; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

         Probe: EX3DV4 — SN7349; ConvF(9.72, 9.72, 9.72); Calibrated: 15.06.2016;

         Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         Electronics: DAE4 Sn601; Calibrated: 30.12.2015

         Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

         DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 62.09 V/m; Power Drift = 0.02 dB
 Peak SAR (extrapolated) = 3.70 W/kg
 SAR(L g) = 245 W/kg; SAR(10 g) = 1.59 W/kg
 Maximum value of SAR (measured) = 3.26 W/kg


        dB



         —2.20

         —4.40

         —6.60




         —11.00
                    0 dB = 3.26 W/kg = 5.13 dBW/kg




 Certificate No: D835V2—40069_Jul16             Page 5 of 8


®
Impedance Measurement Plot for Head TSL




                                                                      20 Jul 2016     13:37:09
          CHI)   sii     1 U   F§       1: 51.943 n   —2.0898 0   91.205 pF          835.000 000 MHz
                                                 _.




          Av
          16 *




                       START 635.000 00 MHz                                   STOP 1 035.000 000 MHz




Certificate No: D835V2—40069_Jul16                      Page 6 of 8


G
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                   Date: 20.07.2016



DUT: Dipole 835 MHz D835V2; Type: D835V2; Serial: D835V2 — SN: 40069

Communication System: UID 0 — CW; Frequency: $35 MHz
Medium parameters used: f = 835 MHz; 0 = 1.01 S/m; & = 54.9; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(9.73, 9.73, 9.73); Calibrated: 15.06.2016;

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 S$Sn601; Calibrated: 30.12.2015

        Phantom: Flat Phantom 5.0 (back); Type: QDOOOPS5OAA; Serial: 1002

        DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 60.64 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 3.68 W/kg
SAR(I g) = 2.5 W/kg; SAR(10 g) = 1.63 W/kg
Maximum value of SAR (measured) = 3.31 W/kg




        —2.00

        ~4.00

        —6.00

        —8.00

        —10.00
                   0 dB = 3.31 W/kg = 5.20 dBW/kg




Certificate No: D835V2—40069_Jul16               Page 7 of 8



Document Created: 2017-05-02 17:32:10
Document Modified: 2017-05-02 17:32:10

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC