SAR Report Appendix 1

FCC ID: 2AIZN-X650

RF Exposure Info

Download: PDF
FCCID_4391519

     Calibration Laboratory of
                                                                                                                        Schweizerischer Kalibrierdienst




                                                                                                              in © w
     Schmid & Partner                                                                                                   Service suisse d‘étalonnage
       Engineering AG                                                                                                   Servizio svizzero di taratura
     Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                                                                       Swiss Calibration Service


     Accredited by the Swiss Accreditation Service (SAS)                                                       Accreditation No.: SCS 0108
     The Swiss Accreditation Service is one of the signatories to the EA
     Multilateral Agreement for the recognition of calibration certificates

     Client      BACLSTWI(Auden)                                                                  Certificate No: EX3—7522_Nov18


     CALIBRATION CERTIFICATE
     Object                                      E.XSDV4 = SN:?522




     Calibration procedure(s)                    QA CAL—01.v9, QA CAL—14.v4, OA CAL—23.v5, OA CAL—25‘v6
 |                                               Calibration procedure for dosimetric E—field probes


     Calibration date                            Novamber 2, 2018


     This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1)
     The measurements and the uncertainties with confidence probatbility are given on the following pages and are part of the certificate.


     All calibrations have been conducted in the closed taboratory facility; environment temperature (22 + 3)°C and humidity < 70%


     Calibration Equipment used (M&TE critical for calibration)



     Primary Standards                      ID                           Cal Date (Certificate No.)                      Scheduled Calibration
 _P(_}g\@r meter NRP                        SN: 104778                   04—Apr—18 (No. 217—02672/02673)                 Apr—19
     Power sensor NRP—Z291                  SN: 103244                   04—Apr—18 (No. 217—02672)                       Apr—19
     Power sensor NRP—Z91                   SN: 103245                   04—Apr—18 (No. 217—02673)                       Apr—19
     Reference 20 dB Attenuator             SN: S5277 (20x)              04—Apr—18 (No. 217—02682)                       Apr—19
     Reference Probe ES3DV2                 SN: 3013                     30—Dec—17 (No. ES3—3013_Dec17)                  Dec—18
     maes_______                           | sn:seo                      21—Dec—17 (No. DAE4—660_Dec17)                  Dec—18
     SecondaryStandards            ____|    1D                           Check Date (in house)                           Scheduled Check
     Power meter E44198                     SN: GB41293874               06—Apr—16 (inhouse check Jun—18)              _ Inhouse check: Jun—20
     Power sensor E4412A                   _ SN: MY41498087             _06—Apr—16 (in house check Jun—18)               In house check; Jun—20
     Power sensor E4412A     ___            SN: 000110210                06—Apr—16 (in house check Jun—18)             | In house check: Jun—20__
     RF generator HP 8648C                  SN: US3642U01700             04—Aug—99 (in house check Jun—18)               In house check: Jun—20
     Nelwo-rk Analyzer E8358A               SN: US41080477               31—Mar—14 (in house check Oct—18)               In house check: Oct—19

                                             Name                              Function                                   Signature
 Calibrated by                               Glaudio Leubler                   Laboratory Technician                     (    J   |


Approved by:                                 Kaltja Pokovic                    Technical Manager                  M



                                                                                                                          Issued: November 6, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: EX3—7522_Nov18                                        Page 1 of 11


   cahbl_'atm“ Laboratory of
                                                                                            g   Schweizerischer Kalibrierdienst
  Schmid & Partner                                                                         _    Service suisse d‘étalonnage
     Engineering AG                                                                             Servizio svizzero di taratura
  Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                                           S    Swiss Calibration Service



  Accredited by the Swiss Accreditation Service (SAS)
                                                                                            Accreditation No.: SCS 0108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multifateral Agreement for the recognition of calibration certificates

  Glossary:
  TSL                          tissue simulating liquid
  NORMx.y.z                    sensitivity in free space
  ConvF                       sensitivity in TSL / NORMx,y,z
  DCP                         diode compression point
  CF                          crest factor (1/duty_cycle) of the RF signal
  A, B, C, D                  modulation dependent linearization parameters
  Polarization 4              ip rotation around probe axis
  Polarization 8              8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                              1e., 8 = 0 is normal to probe axis
 Connector Angle              information used in DASY system to align probe sensor X to the robot coordinate system
 Calibration is Performed According to the Following Standards:
     a)    IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
                                                                                                    Specific
           Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
           Techniques", June 2013
     b) IEC 62209—1, ", "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from hand—
           held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)", July 2016
     ©)    IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
           used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March 2010
     d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
 Methods Applied and Interpretation of Parameters:
           NORMx,y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell: f > 1800 MHz: R22 waveguide).
           NORMx.y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E*—field
           uncertainty inside TSL (see below ConvF).
           NORM(MAx,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
           implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
           in the stated uncertainty of ConvF.
           DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
          signal (no uncertainty required). DCP does not depend on frequency nor media.
          PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
          characteristics
          Ax,y.2; Bx,y,z; Cx,y.2; Dx.y,z; VRx,y,z: A, B, C, D are numerical linearization parameters assessed based on
          the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
          media. VR is the maximum calibration range expressed in RMS voltage across the diode.
          ConvrF and Boundary Effect Parameters: Assessed in flat phantomn using E—field (or Temperature Transfer
          Standard for f < 800 MHz) and inside wavequide using analytical field distributions based on power
          measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
          boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
          used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
          to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
          ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
          MHz.
          Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
          exposed by a patch antenna.
          Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
          (on probe axis). No tolerance required.
          Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
          uncertainty required).


Certificate No: EX3—7522_Novi8                                 Page 2 of 11


 EX3DV4 — SN:7522                                                               November 2, 2018




                       Probe EX3DV4

                                          SN:7522

                                  Manufactured:             February 26, 2018
                                  Calibrated:               November 2, 2018



                                 Calibrated for DASY/EASY Systems
                                    (Note: non—compatible with DASY2 system!)




Certificate No: EX3—7522_Nov18                    Page 3 of 11


        EX3DV4—   SN:7522                                                                                                                November 2, 2018




        DASY/EASY — Parameters of Probe: EX3DV4 — SN:7522
        BasicCalibration Parameters
         f              z>           31              591_15__1_:“{_ C           |              Sensgr Y¥                  Sensor Z             |       Unc (k=2)   ]
    |_Norm (uV/(V/im}")"                                0.43        |   0.43                                               0.52                ‘ #10.1 % |
    | DCP (mV)           ®                              99.1       [ _ 100.0                                     |         97.2                |         |

        Modulation Calibration Parameters
    |. UID         | Communication Syfim Name                            |           |    A       |      B    |   C    |     b     I~    VR                Une

                                                                        |_ |            __dB          dBvuaVy ' ns    ]    dB     |    mV          |      (k=2)    |
    ‘             _
        0             CW                                                |   x       |   0n             0.0       10   |   0.00         1432        |     433%      |
    |             |                                                     |           |   n n      |     0.0   t   10   1           i    1a8 1       |               |

                                                                        [(z [           oo       ~|    oo    |   15   ]           | 140.                 mew
                  C

        The reported uncertainty of measurement is stated as the standard uncertainty of measurement                                                                    |
|       nultipled by the coverage factor k=2, which for a normal distribution corresponds to a coverage                                                                J
|5 probability of approximately 95%.
                                                                                                                                                                |
l                                                                                                                    m                                         s
    " The uncertainties of Norm X.Y,Z do not affect the E*—fleld uncertaintyinside TSL (see Pages 5 and 6)
    " Numerical linearization parameter uncertainty not required
    ~ Uncertainty is determined using the max. deviation from linear response applying rectangqular distribution and is expressed for the square of the
    field value.




Certificate No: EX3—7522_Nov18                                              Page 4 of 11


 EX3DV4— SN:7522                                                                                                                 November 2, 2018




 DASY/EASY — Parameters of Probe: EX3DV4 — SN:7522


 Calibration Parameter Determined in Head Tissue Simulating Media
   |         |  Relative  Conductivity |     |       ]                                                                    | Depth® |     Unce           |
   ! f (MHz) " E      Permit_tiwil;.nIF            [S-‘m,’rr       ’ ConvF X |ConuF Y¥ {ConvF Z2 |
                                                                                                 Aipha
                                                                                                    ha °1_{@)__“@_2_]                               EI

   |_     750              41.9            |        0.89           |   9.78           9.78        |   9.78    _|   0.53   |   0.87     #12.0 %      |

   [      850              41.5                     0.92           ‘                  9.46        |       ‘        0.44   ‘0.92        + 12.0 %     !

   |— #780—|——t0d——|—tar _,!_1&. | azo |— fi‘;‘_:_.‘b}i{_;_ifif‘_ |+120%|
   ?     1900     I        40.0           |         140            |   7.91       I   7.91   |        7.91    |_0.34          0.86     + 12.0 %     |

   |     2000              40.0           _|        1.40               7.78           T78    |_       778          ors        ose      £120% |
         2300              39.5                     1.67               7.35           7.35   '!       7.35         0.33       0.90     + 12.0 % _

         2450              39.2                     1.80               6.97           6.97   _|       _6.97        0.30       1.05     + 12.0 %

         2600              39.0                     1.96               6.79      _|   6.79            6.79         0.35       0.99     +12.0 %

        _ 5250             35.9                    4.71                5.05           5.05            5.05         0.40       1.80     +13.1 % _

        5600 _| _         35.5°                L   5.07        |       448 _ |        4.48            4.48         0.40       1.80     %13.1%

         5800              35.3                    5.27                4.176          4.76            4.16         0.40       1.80     £+13.1 %



© Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (e and a) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and a) is restricted to + 5%. The uncertainty is the RSS of
Iha ConvF uncertainty for indicated target tissue parameters.
 * Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1%for frequencies below 3 GHz and below + 2%for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter fromthe boundary.




Certificate No: EX3—7522_Nov18                                           Page 5 of 11


  EX3DV4A— SN:7522                                                                                                                   November 2, 2018




  DASY/EASY — Parameters of Probe: EX3DV4 — SN:7522


 Calibration Parameter Determined in Body Tissue Simulating Media
                  — |         Relative B       | E.‘T:mductflvily |                      |            |                       M [ Depth® |       Unc      |
        _f (MHz) "          Permittivity            (S/im)"               ConvF X           ConvF¥       | ConvFZ    Aipha" ‘ _(mm)      ‘     (k=2)     ‘

        |_fs0        _| _       555                  o.96            _|    9.80              9.80          a.so     _| 044 | 090             +120% |
    |__s50                      55.2                 ose__|                as4      _| 954                 9.54     ‘ 0.A7 l _0.84           £120% ‘
    |      1750                53.4        |          1.49       !         7.88     I_?fi             1‘    7.88     ]_g_.;d_}    o.ss    _| +£12:0% |
   |__1900__|                   53.3       f' _182               ['        748      |       1ds      —|_— TA# |_0.40             0.85    _| £+120% |
   |2000             |         B23 |_ 1152                                 73e—‘{——"rs3e             "|~*~7386        0.40       0.85        +120 % I
          2300                 52.9                  im [ fa]                                im           fa        | oss        ouso | s120% I|
          2450       _ 52.7.                         1.95        ___|__7.05                  7.05 _| _ _7.05          630 |      1310—       |—@420%—|
          2600_|               52.5                  216        _ _| _6.95                   6.95          6.95       0.50       0.90        +120%
          5250                 48.9                  5.36                  4.77.            4.77__|__ sF7             0.50       1.90        #13.1%_
          5600                 48.5                  5.77                  4.27             4.27           4.27       0.50       1.90        +131 %
   | _ssoo | _                 48.2                  6.00                  4.31             4.31           4.31       0.50       1.90        +131%|]

© Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and a) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and a) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters
6 Alpha/Depth are determined during calibration, SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter fromthe boundary




Certificate No: EX3—7522_Nov18                                              Page 6 of 11


   EX3DV4— SN:7522
                                                                                                                             November 2, 2018


                                                         Frequency Response of E—Field
                                                              (TEM—Cell:ifi110 EXX, Waveguide: R22)




                      1.5—
                                         t


                      1.4—
                                     I

                                     L
                     1.3—
                o                    F
                O                    |
                N
                C      e



                     1.1

             hy
             5       19



                —   09


             5
             5 Qg.4..._.._.
             §                 —L|                                                                                                    |
            u
                    U./—]
                         t


                    0.6—
                           L

                    [']rj.i                  L_t   _    I_i'—l..      IL_iL_L_L_L |    |__L_L   .L_.i   l._q_l_;    N   L    |' |_‘
                           0                             &            1000            1500       2000        2500           3000
                                                                                      f [MHz]
                                                                 ©                                         +
                                                                TEM                                       R22


                                                       Uncertainty of Frequency Response of E—field: * 6.3% (k=2)




Certificate No: EX3—7522_Nov18                                                Page 7 of 11


   EX3DV4— SN:7522
                                                                                                                                                                         November 2, 2018




                                                            Receiving Pattern (¢), 9 = 0°

                                              t=600 MHz, TEM                                                                                f=1800 MHz,R22




                    ®                            e                    0               e                             ®                         6               e                         e
                T                                                     Y                                            To                                                                   z7




                            |C            .                      :
                                                                                     Z                         i
                                                                                                               $                        f
                                                                                                                                                       7                        T
                                                                                                                                                                                ;
                                                                                                                                                                                        —‘



        a               |             |                      §                                             §                            j                                   |                 }
        'fo_" ||...?.,.h.;__:’_,._!...‘.. '_’.__': g—~s=g=# ’ 4 t«;&gt‘ 45 ‘_‘ s# —3—>*+* ’—.—'84-':'. #=3—

        §p**}
         wl —0.5—4
                                                             :                  :                      ?                        i
                                                                                                                                    ?                                       K4|
                                                                                £                                          s                                                :                 |


                    |            L_L_j____4__u__                     £_3_1_    T           1__L   _|                    _1 _FJ__                  1         —L_1_L_IL               J    _!
                                     150                    10(                —50                     0                       50                     100                150

                                               (*#]                             [#)                                       ~&                                         &
                                              100 MHz                         600 MHz                                   1800 MHz                                  2500MH:


                                                        Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificate No: EX3—7522_Nov18                                                            Page 8 of 11


   EX3DV4—   SN:7522
                                                                                                                                            November2, 2018




                                                  Dynamic Range f(SARneaq)
                                                         (TEM cell , feys= 1900 MHz)




                       in                                                           #                                                  i|

                       C        1047


                                         |

                                         I


                                in2—
                                 0+—          i
                                     Pnd

                                     |
                                     L.
                                     I
                                     |                               j
                                     103                        0               10
                                                                          SAR [mW/icm3]
                                                        L#
                                                  not compensated                                         co
                                                                                                                   4




                                  j L.

                            ray
                            J                                                                       4
                             C  0                 Wt ioiPb i hn dn        ue   t   4 tPc   iN 4i           004 i.      oup   n
                             6                                                             :
                            d                                                              $
                                 M                                                         i.   ®



                                                                              i                      j
                                     103          102         10—              109                  101                102       103
                                                                         SAR {mW/cm3]
                                                        ¢                                                      —
                                                  ot compensated                                    compensated


                                             Uncertainty of Linearity Assessment: £ 0.6%(k=2)




Certificate No: EX3—7522_Nov18                                      Page 9 of 11


  EX3DV4— SN:7522                                                                                                             November 2, 2018


                                        Conversion Factor Assessment
                  f= 850 MHz,WGLS R9 (H_convF)                                          f= 1900 MHz,.WGLS R22 (H_convF)




                                          frrim}                        S                           a         & {min]
                                                          #]                                  a




                                      Deviation from Isotropy in Liquid
                                                           Error (6, 9), f = 900 MHz
                         Deviation
                                 s
                                 in




                           —1.0         —0.8       —0.6   —0.4   —0.2       0.0   0.2   0.4   0.6       0.8             1.0
                                     Uncertainty of Spherical Isotropy Assessment: * 2.6% (k=2)




Certificate No: EX3—7522_Nov18                                    Page 10 of 11


      EX3DV4— SN:7522
                                                                                     November 2, 2018




      DASY/EASY — Parameters of Probe: EX3DV4 — SN:7522

      Other Probe Parameters
       Sensor Arrangement
                                                                                 |        Iriangular |
       Connector Angle (°)
                                                                             |             _ ~30.9|
                                                                                                       |
       Mechanical Surface Detection Mode
                                                                                           enabled|
       Optical Surface Detection Mode                                        |             disabled |
       Probe Overall Lenath
  L
                                                                                              7 mm |
       C              Diameter
                                                                                             10 mm |
       Tip Length
                                                                                                mm
       Tip Diametey                                                                        2.5 mm |
      Drahs
       MTODé
               Tini~   o      7    Tip      r
                 ip to Sensor X Calibration Point
                                                 r
                                                                         |                    1 mm
      Probe Tip to Sensor Y Calibration Point                                                 ! mm

      Probe Tip to Sensor Z Calibration Point
                                                                                              1 mm |
      Recommended Measurement Distance from Surface                  |                     1.4 mm




Certificate No: EX3—7522_Nov18                       Page 11 of 11


                  fi“                       in Collaboration with                      § “\w’?"/             A ':F E "bkfl

                  =‘/"/‘J, _ :sw                                                                                               AFrA
                  <ag. cgemontzowor isccCONASt
              Add: No.351 Xueyuan Road. Haidian District, Beijing, 100191, China   ‘?,,/-//'-;:“\_:\\'\S-   v CALIBRATION
              Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504              "Arlyltuk®®                              CNAS LO570
              E—mail: ettl@chinattl.com        Hitp:/www.chinattl.en

       Client               BACL                                              Certificate No:               Z16—97195

 CALIBRATION CERTIFICATE

 Object                                     D835V2 — SN: 445

 Calibration Procedure(s)                   ED—211—003—04

                                            Calibration Procedures for dipole validation kits

 Calibration date:                          October 26, 2016

| This calibration Certificate documents the traceability to national standards, which realize the physical units of
  measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
‘ pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘Cc and
 humidity<70% .


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                 ID #                 Cal Date(Calibrated by, Certificate No.)                    Scheduled Calibration
   Power Meter       NRP2           101919              27—Jun—16 (CTTL, No.J16X04777)                                     Jun—17           |
   Power sensor NRP—Z91             101547              27—Jun—16 (CTTL, No.J16X04777)                                     Jun—17
   Reference Probe EX38DV4          SN 7433             26—Sep—16(SPEAG,No.EX3—7433_Sep16)                                  Sep—17
   DAE4                             SN 777              22—Aug—16(CTTL—SPEAG,No.216—97138)                                 Aug—17


  Secondary Standards               ID #                Cal Date(Calibrated by, Certificate No.)                    Scheduled Calibration
  Signal Generator E4438C           MY49071430          01—Feb—16 (CTTL, No.J16X00893)                                     Jan—17
   Network Analyzer E5071C          MY46110673          26—Jan—18 (CTTL, No.J16X00894)                                     Jan—17



                                    Name                           Function                                             Signature
 Calibrated by:                   Zhao Jing                   SAR Test Engineer                                 2        %é

| Reviewed by:                    Qi Dianyuan                 SAR Project Leader                            —       %
 Approved by:                     Liu Wei                    Deputy Director of SEM Department                       /;U] q
                                                                                             Issued: October 27, 2016
l This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



      Certificate No: Z16—97195                            Page 1 of 8


           _’.\.@                   I;Coiiaborancnewith 5

         S
         /"/   a
            °/ CALIBRATION LABORATORY




      (
        Add; No,51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hup://www.chinattl,on



Glossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y.z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
    Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
    devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e    Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
a    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
e    SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
»    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.



Certificate No: Z16—97195                              Page 2 of 8


          !.\r@                      In Collaboration with

                                    a


      (:S
                                    CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: ctt@chinattl.com        Htp://www.chinattl.en


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                         DASY52                                      52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                         15 mm                                     with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       835 MHz +1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature             Permittivity           Conductivity

     Nominal Head TSL parameters                               22.0 °C                  41.5                0.90 mho/m

     Measured Head TSL parameters                        (22.0 £0.2) °C               41.4 +6 %           0.92 mho/im +6 %

     Head TSL temperature change during test                   <1.0 °C                   =———                     ————

SAR result with Head TSL
     SAR averaged over 1 _cm"‘ _(1 g) of Head TSL                        Condition
     SAR measured                                                  250 mW input power                   2.41 mW /g

     SAR for nominal Head TSL parameters                            normalized to 1W            9.46 mW ig * 20.8 % (k=2)

     SAR averaged over 10 cnm" (10 g) of Head TSL                        Condition
     SAR measured                                                  250 mW input power                   1.58 mWw /g

     SAR for nominal Head TSL parameters                            normalized to 1W            6.23 mW /g * 20.4 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature             Permittivity           Conductivity
     Nominal Body TSL parameters                              22.0 °C                   55.2                0.97 mho/m

     Measured Body TSL parameters                        (22.0 £0.2) °C               55.8 £6 %           0.95 mho/m + 6 %

     Body TSL temperature change during test                  <1.0 °C                    ————                     ———
SAR result with Body TSL
     SAR averaged over 1 C¢m:" (1 g) of Body TSL                         Condition
     SAR measured                                                 250 mW input power                    2.36 mW / g

     SAR for nominal Body TSL parameters                            normalized to 1W            9.60 mW /g + 20.8 % (k=2)
     SAR averaged over 10 cm" (10 g) of Body TSL                         Condition
     SAR measured                                                 250 mW input power                    1.59 mW / g

     SAR for nominal Body TSL parameters                            normalized to 1W            6.44 mW /g £ 20.4 % (k=2)

Certificate No: Z16—97195                             Page 3 of 8


               r              a     in Collaboration with

                                    §       p       _e      _a@       04




(                  N
                                    CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hittp:/‘www.chinattl.en


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   45.50— 3.43j0

     Return Loss                                                              — 24.6dB


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   47.50— 7.29j0

     Return Loss                                                              —22.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.497 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.


 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data


     Manufactured by                                                             SPEAG




Certificate No: Z16—97195                                Page 4 of8


           ‘!\w                      in Callaboration with
          t                         s               e a g
          ! TTL                     CAUBRPNHON LABORATORY

        Add: No.51 Xueyuan Road, Haidian District. Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hittp://www.chinattl.cn


DASY5S Validation Report for Head TSL                                  Date: 10.26.2016
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 445
   Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
   Medium parameters used: £ = 835 MHz; 0 = 0.922 S/m: s =41.42; p = 1000 kge’m3
   Phantom section: Right Section
   Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2007)
DASYS5 Configuration:

            &    Probe: EX3DV4 — SN7433; ConvF(9.82, 9.82, 9.82); Calibrated: 9/26/2016;
            *    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAE4 Sn777; Calibrated: 2016—08—22
            &    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 58.51V/m; Power Drift =—0.02 dB
        Peak SAR (extrapolated) =3.63 W/kg
        SAR(I g) =2.41 Wi/ikg; SAR(10 g) =1.58 W/kg
        Maximum value of SAR (measured) =3.07 W/kg

           aB         0|
           o

           —2.09



           —~4.18



           —6.27



           —8.36



           —10.45

                      0 dB =3.07 W/kg = 4.87 dBW/kg




Certificate No: Z16—97195                              Page 5 of 8


                                  e     In Collaboration with




     )
              _7"7%, s_B _2 _0 _3
              _                         CALIBRATION LABORATORY
                   \igs»""
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com       Http://www.chinattl.on




           Impedance Measurement Plot for Head TSL



                    »1   835,00000 MHz —24.574 d6


          30. 00

          20. 00




         ~40. 0C

         —s0.00 !                           —                                                   3
     PD@A sii smith (@+]x) scale 1.000U0 [ri bel]
         >1   $35.00000 mHz     45.517 n —3.4308 o                      5§5.5§Z—pF _
                                                                         _-"’I                      ‘."\\
                                                                                                            %
                                                                4                                               x
                                                         £

                                                Ilf                                     r’/—\
                                                I                                        $

                                                l


                                                    \\
                                                         "\._

                                                                \\  C                                       $

                                                                                 e                  *
     >                                                                               ks _A_
     1 Start 635 MHz                                                              IFBW 100 Hz                       Stop .085 ce tm




Certificate No: Z16—97195                                                        Page 6 of 8


                              3      In Collaboration with


                                    CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hitp://www.chinattl.en




DASYS5 Validation Report for Body TSL                                    Date: 10.26.2016
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 445
   Communication System: UID 0, CW; Frequency: §35 MHz; Duty Cycle: 1:1
   Medium parameters used: £= 835 MHz; 0 = 0.952 S/m; &, = 55.79; p = 1000 kg/m*
   Phantom section: Center Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(9.5,9.5, 9.5); Calibrated: 9/26/2016;
            «    Sensor—Surface: 2mm (Mechanical Surface Detection)
            s    Electronics: DAFE4 Sn777; Calibrated: 2016—08—22
            a    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value= 56.01 V/m; Power Drift =0.01 dB
        Peak SAR (extrapolated) = 3.44 W/kg
        SAR(I g)=2.36 W/kg; SAR(10 g) = 1.59 W/ikg
        Maximum value of SAR (measured) =2.95 W/kg




            ~1.93



            —3.86



            ~5.78


            —£.21



            —9.64

                       0 dB =2.95 W/kg = 4.70 dBW/kg




Certificate No: Z16—97195                              Page 7 of 8


                                  &    In Collaboration with

            7‘7‘jJ,                    a
                                       CALIBRATION LABORATORY
            T‘/
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com      Http://www.chinattl.on


         Impedance Measurement Plot for Body TSL


      Tri S11 Log Mag 10.00da/ ref 0.000de [F1]
        50. 00
                    >1   835.00000 MHz —22.066do
        40. 00

        30. 00

        20. 60
        10. 00

        0. 000 pl
        10. 00
       —20. 00

       —30. 00

      —40, 00

      —50. 00
                                                                       a.
     PB si1 smith (R+]jx) scale 1.000Uu [fi del]
       »1   §35.00000 mHz      47.495 n —7.2889 o   26.1530—pF




    |
    (1 Btat 635 Mz                                        trew 100 e         Stop 1.095 ocz lR




Certificate No: Z16—97195                               Page 8 of 8


                       NCL CALIBRATION LABORATORIES


                                  Calibration File No: DC—1748
                                        Project Number: 5822


                                         Client.: BACL Corp.
                  Address: 6/F, the 3rd Phase of Wan Li Industrial Bldg., Shihua Rd.,
                               FuTian Free Trade Zone, Shenzhen, China




   CERTIFICATE OF CALIBRATIO N

               It is certified that the equipment identified below has been calibrated in the
           NCL CALIBRATION LABORATORIES by qualified personnel following recognized
                   procedures and using transfer standards traceable to NRC/NIST.



                                      Validation Dipole (Head & Body)


                                    Manufacturer: APREL Laboratories
                                       Part number: ALS—D—1900—5—2
                                             Frequency: 1900 MHz
                                             Serial No: 210—00710



                                      Calibrated: 20°" September 2017
                                     Released on: 27 September 2017

This Calibration Certificate is Incomplete Unless Accompanied with the Calibration Results Summary




        Released By:             /EAAA    Art Brennan, Quality Manager



                         NCL CALIBRATION LABORATORIES
                         Suite 102, 303 Terry Fox Dr           Division of APREL Lab
                             Kanata, ONTARIO                 TEL (613) 435—2300
                             CANADA K2K 3.11                  FAX: (613)495—8308


NCL Calibration Laboratories
Division of APREL Laboratories                                                                                    DC—1748




Conditions

Dipole 210—00710 was a re—calibration.

Ambient Temperature of the Laboratory:                            21 °C +/— 0.5°C
Temperature of the Tissue:                                        21 °C +/—0.5°C




    Attestation

    The below named signatories have conducted the calibration and review of the data which is presented in
    this calibration report.

    We the undersigned attest that to the best of our knowledge the calibration of this system has
    been accurately conducted and that all information contained within this report has been reviewed
    for accura




    Art Brennan QM




Primary Measurement Standards

Instrument                                     Serial Number                         Cal due     date
Tektronix USB Power Meter                      11C940                                April 13,   2019
Network Analyzer Anritsu 37347C                002106                                Jan. 26,    2019
Agilent Signal Generator                       MY45094463                            Dec. 11,    2017




                                                                                                         Dipele 5N 210. 00710

Page 2 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories                                                                                                            DC—1748




Calibration Results Summary

The following results relate the Calibrated Dipole and should be used as a quick reference
for the user.

Mechanical Dimensions

    Length                 Height                Diameter
   67.5 mm               39.5 mm                  3.6 mm

Tissue Validation

     4                                        Dielectric               Conductivity,
    Tissue             Frequency             constant, ar                 o [S/m]

      Head              1900 MHz                 39. 44                                               1.41
      Body              1900 MHz                 52.70                                                1.57

Electrical Specification

    Tissue             Frequency             Return Loss                                        SWR                      Impedance
      Head              1900 MHz              —28.662 dB                                  1.077 U                        52.368 Q
      Body              1900 MHz              —22,498 dB                                  1162 U                         55.211 Q

System Validation Results

      Tissue           Frequency            1 Gram, Wikg          10 Gram, W/kg

       Head             1900 MHz                 42.14                                         21.89
       Body             1900 MHz                 42.11                                         2212

Head                                                          Body

                                                                                                             Ares Scan
                                                                             B
                                                                  Sif fiammalrtissn3¢ &
       X Axis (mm)




                                                                                          X Aws om)
  8




                        Y Axis (mm)




                                                                                                                                    Dipele 21000740


Page 3 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


NCTL Calibration Laboratories
Division of APREL Laboratories                                                               pe—itss



Introduction

This Calibration Report has been produced in line with the SSI Dipole Calibration Procedure
SSI—TP—018—ALSAS. The results contained within this report are for Validation Dipole
210—00710. The calibration routine consisted of a three—step process. Step 1 was a
mechanical verification of the dipole to ensure that it meets the mechanical specifications.
Step 2 was an Electrical Calibration for the Validation Dipole, where the SWR, Impedance,
and the Return loss were assessed. Step 3 involved a System Validation using the ALSAS—
10U, along with APREL E—020 30 MHz to 6 GHz E—Field Probe Serial Number 225.

References

    o    IEEE Standard 1528:2013
         I|EEE Recommended Practice for Determining the Peak Spatial—Average Specific
         Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
         Measurement Techniques
         EN 62209—1:2006
         Human Exposure to RF Fields from hand—held and body—mounted wireless
         communication devices — Human models. instrumentation, and procedures — Part 1:
         Procedure to measure the Specific Absorption Rate (SAR) for hand—held mobile
         wireless devices
    o    1EC 62209—2:2010
         Human exposure to RF fields from hand—held and body—mounted wireless devices —
         Human models, instrumentation, and procedures — Part 2: specific absorption rate
         (SAR) for wireless communication devices (30 MHz — 6 GHz)
    o    D22—012—Tissue dielectric tissue calibration procedure
    o.   D28—002—Dipole procedure for validation of SAR system using a dipole
    o    IEEE 1309 Standard for Calibration of Electromagnetic Field Sensors and Probes,
         Excluding Antennas, from 9 kHz to 40 GHz

Conditions

Ambient Temperature of the Laboratory: 21 °C +/— 0.5°C
Temperature of the Tissue:             21 °C +/—0.5°C

Dipole Calibration uncertainty
The calibration uncertainty for the dipole is made up of various parameters presented below.

Mechanical                                           1%
Positioning Error                                    1.22%
Electrical                                           1.1%
Tissue                                               2.2%
Dipole Validation                                    2.2%

Combined Standard Uncertainty                        3.88% (7.76% K=2)
The Following Graphs are the results as displayed on the Vector Network Analyzer.


                                                                                      Gipule 210.007 10


Page 4 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories                                                              pc—tma



$11 Parameter Return Loss

Head

Frequency Range 1868.68 MHz to 1934.44 MHz

  S11 FORMARD REFLECTION                                         L4 5.8584
                                                                     1 + $on REF

                                                                   B.000 d8 CFFSET
  LOB MAGNTTUDE            »REF=—26.234 8   4.800 d8/DIV           8.30°    OFFSET

                                                                rHARKER 2
                                                                   1.900000 OHz
                                                                    ~280.602 dB


                                                                 aRKER TO MAX
                                                                 MaRKER TO MIN

                                                                   1.958680 6Hz
                                                                    —20.022 «B

                                                           to       934437 bHz
                                                                    —20.012 48




        4          :             —    —     :     —              MARKER READOUT
  1.700560                     GHz              2.099900         FUNC T 1ONS




Body

Frequency Range 1882.37 MHz to 1928.52 MHz

  S11 FORHARD REFLECTION                                         CH 528584
                                                                     t — #1 nm REF

                                                                   B.A00 dB DFFSET
  LUG MAGNITUDE            »REF=—20.004 8   3.800 d8/DIV           8.00°     OFFRET

                                                                PHARKER 2
                                                                   1.900080 OHz
                                                                    ~22.498 dB


                                                                 MaRKER TO MAX
                                                                 MaRKER TO MIN

                                                                   1.882373 6He
                                                           ,




                                                                    —20.012 dB

                                                                    928524 6Hz
                                                           to




                                                                    ~20.005 «B




        &          :             —    —     —     —              MaRKER READOUT
  1.700560                     GHez             2.099900         FUNC T 1ONS




                                                                                      Dipcle 21000710


Page 5 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories                                                                   DC—1748




SWR

Head

  311 FORHARD REFLECTION                                              O# 1 — §°1
                                                                        5.0584 mm REF
                                                                        ©.800 48 DFFSET
  SHR                         »REF=854.955 nU   300.000 mUZD1Y          0 .00°   OFFSET

                                                                     »NARKER 2
                                                                        1.900000 6Hz
                                                                           1.077 U


                                                                      MARKER TO MAX
                                                                      MaRKER T0 MIN

                                                                 1      1.868680 CHz
                                                                           1.223 1

                                                                 3      1.934937 bHe
                                                                           1433 H




                                                                      MaRKER READOUT
  1.700560                         GHz                2.899980        FUNCT 10M3



Body
  811   FORARD   REFLECTION                                           C# 1~— $11
                                                                        5§.0584 mm REF
                                                                        0.800 48 UFFSET
  SHR                         »REF=854.955 nU   300.000 mU/D1Y          8. 00°    OFFSET

                                                                     PHARKER 2
                                                                        1.900000 6Hz
                                                                           1.162 1


                                                                      MARKER TO MAX
                                                                      MaRKER TO MIN

                                                                 bob.                Hz
                                                                            m w
                                                                            =o
                                                                            a ol
                                                                            io c




                                                                                  6Hz
                                                                 ©




                                                                           1,.224 I




                                                                      MARKER READOUT
  1.700560                         GHz                2,099980        FUNCT 108




                                                                                           Dipcle 21000710


Page 6 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories                                                              DC—1748




Smith Chart Dipole Impedance

Head

  S11 FORMARD REFLECTION      1                           0#    1   + .811
                                                               5.0584 nm REF
  INPEDANCE
                                                               .000 48 OFFSET
                                                               0.00°    OFFSET

                                                         RNARKER 2
                                                            1.900000 GHz
                                                              52.a60 1
                                                              —2.738 j0
                                                          MARKER TD MAX
                                                          HaRKER TO WILN

                                                                .B6B68N      CHz
                                                                 §7.927      1
                                                                  7.378       j8
                                                                234437       UHz
                                                                 41.020      7
                                                                 —4.263      j9




                                                          MARKER READOUT
                                                          FUNCTIONS
                     1.700560 — 2.099980 6Hz




Body

  S11 FORMARD REFLECTION      1                           tX 1 ~ gff
  IHPEDANCE                                                 5.0584 im REF
                                                            .000 48 UFFSET
                                                            0.00°    OFFSET

                                                         »NARKER 2
                                                            1.900000 GHz
                                                              S1.211
                                                              —5.845 )9

                                                          MARKER TO MAX
                                                          MARkER TO MIN

                                                               1.882373      BHa
                                                                 61.085      9
                                                                137.319      jmf
                                                     3         1.928524      bHz
                                                                 14.594      1
                                                                  7.006      j9




                                                          MARKER BEADOUT
                                                          FUNCTIONS
                     1.700560 — 2.099980 6Hz




                                                                                      Dipcle 21000710


Page 7 of 7
This page has been reviewed for content and attested to on Page 2 of this document.


                  h                         in Collaboration with                   \a\"\ull""/o                        s

                 7"]°J       , e =\<"t
                  awllimzmyse" CALIBRATION LABORATORY
                                                                                                                    AmaL
              Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   'f,,//;“:\\\?‘   v CALIBRATION
                Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504           Ardal t                         CNAS LO570
              E—mail: ettl@chinattl.com         hitp://www.chinattl.en
       Client              BACL                                                Certificate No:     217—97192

CALIBRATION CERTIFICATE

Object                                       D2450V2 — SN: 751

Calibration Procedure(s)                     FE—211—003—01
                                            Calibration Procedures for dipole validation kits

Calibration date:                           October 12, 2017

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3yC and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                  ID #                 Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Power Meter NRVD                  102196              02—Mar—17 (CTTL, No.J17X01254)                         Mar—18
 Power sensor NRV—Z5               100596              02—Mar—17 (CTTL, No.J17X01254)                         Mar—18
 Reference Probe EX3DV4            SN 3846             13—Jan—17(CTTL—SPEAG,N0.Z16—97251)                     Jan—18
 DAE4                              SN 1331             19—Jan—17(CTTL—SPEAG,.N0.217—97015)                    Jan—18

 Secondary Standards               ID #                Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Signal Generator E4438C           MY49071430          13—Jan—17 (CTTL, No.J17X00286)                        Jan—18
 Network Analyzer E5071C           MY46110673            13—Jan—17 (CTTL, No.J17X00285)                      Jan—18



                                    Name                            Function                                Signature
Calibrated by:                    Zhao Jing                    SAR Test Engineer                           fiff W


Reviewed by:                      Lin Hao                      SAR Test Engineer                        fifi?fi%/ :

Approved by                       Qi Dianyuan                  SAR Project Leader                    é.;/fil/

                                                                                         Issued: October 15, 2017
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z17—97192                                 Page 1 of8


           im                          in Collsboration with
          ["T‘LI, a
           _                           CALIBRATION LABORATORY
         Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
         E—mail: etti@chinattl.com      hitp:/www.chinattl.en


Glossary:
TSL                       tissue simulating liquid
ConvF                      sensitivity in TSL / NORMx,y,z
N/A                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
«_ Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
a Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
   point exactly below the center marking of the flat phantom section, with the arms oriented
   paralle! to the body axis.
* Feed Point Impedance and Return Loss: These parameters are measured with the dipole
   positioned under the liquid filled phantom. The impedance stated is transformed from the
   measurement at the SMA connector to the feed point. The Return Loss ensures low
   reflected power. No uncertainty required.
* Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
   No uncertainty required.
* SAR measured: SAR measured at the stated antenna input power.
a SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
   connector.
e SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
   nominal SAR result.


   The reported uncertainty of measurement is stated as the standard uncertainty of
   Measurement multiplied by the coverage factor k=2, which for a normal distribution
   Corresponds to a coverage probability of approximately 95%.



Certificate No: Z17—97 192                            Page 2 of 8


          __r                         in Collaboration with

          /7/
          saet s p e a g
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      http://www.chinattl.on


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                        DASY52                                      52.10.0.1446

     Extrapolation                              Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL          |          2C 10 mm                                     with Spacer

     Zoom Scan Resolution                  |        dx, dy, dz =5 mm
     Frequency                             |        2450 MHz +1 MHz


Head TSL parameters
   The following parameters and calculations were applied
                                                          Temperature               Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  39.2                1.80 mho/m
     Measured Head TSL parameters                        (22.0 + 0.2) °C             39.1 +6 %           1.82 mho/im + 6 %

     Head TSL temperature change during test                  <1.0 °C                   —~—«—                    —           |
SAR result with Head TSL
     SAR averaged over 1 cm° (1 g) of Head TSL                          Condition
     SAR measured                                                  250 mW input power                  13.2 mW /g

     SAR for nominal Head TSL parameters                            normalized to 1W            §2.5 mW ig + 18.8 % (k=2)

     SAR averaged over 10 cCn(10 g) of Head TSL                         Condition
     SAR measured                                                  250 mW input power                  6.16 mWw /g

     SAR for nominal Head TSL parameters                            normalized to 1W            24.6 mW ig £ 18.7 % (k=2)
Body TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature               Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  52.7                1.95 mho/m
     Measured Body TSL parameters                        (22.0 £0.2) °C              53.3 +6 %           1.96 mho/m + 6 %

     Body TSL temperature change during test                  «1.0 °C                   ———                      ——

SAR result with Body TSL
     SAR averaged over 1 cm" (1 g) of Body TSL                          Condition
     SAR measured                                                  250 mW input power                  12.9 mW /g

     SAR for nominal Body TSL parameters                            normalized to 1W            51.7 mW ig £ 18.8 % (k=2)

     SAR averaged over 10 Cnt" (10 g) of Body TSL                       Condition
     SAR measured                                                  250 mW input power                  6.05 mW /g

     SAR for nominal Body TSL parameters                           normalized to 1W             24.2 mW ig + 18.7 % (k=2)



Certificate No: Z17—97192                            Page 3 of 8


           !\                           in Collsboration with

          @7"TLI,
          C
                  a
                  CALIBRANON LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      http://wwwchinattl.en


 Appendix (Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL

     Impedance, transformed to feed point                                   54.50+ 4.65J0

     Return Loss                                                             —24.10B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                       |           51.50+ 6.760
     Return Loss                                                |            —23.30B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                        1.265 ns




 After lang term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                            SPEAG




Certificate No: Z17—97192                             Page 4 of 8


           b                           in     Collaborstion with
          /7/ E.Ls_as_
            y                          CALIBRATION LABORATORY
         Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com      http://www.chinattl.en


DASYS Validation Report for Head TSL                                    Date: 10.12.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 751
   Communication System: UID 0, CW; Frequency: 2450 MHz: Duty Cycle: 1: 1
   Medium parameters used: f= 2450 MHz; 0 = 1.819 S/m; sr =39.06; p = 1000 kg/m3
   Phantom section: Left Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2007)
DASY5 Configuration:

             *    Probe: EX3DV4 — SN3846; ConvF(7.22,7.22,7.22); Calibrated: 1/13/2017;
             *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
             *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
             *«   Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
             *«   Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                  (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm +
        dy=5mm, dz=5mm
        Reference Value =103.5 V/m; Power Drift =—0.03 dB
        Peak SAR (extrapolated) = 26.9 W/kg
        SAR(I g)=13.2 W/kg; SAR(10 g)=6.16 W/kg
        Maximum value of SAR (measured) = 21.6 W/kg




            ~12.97


            —17.30
                         k
            —21.62
                       0 dB =21.6 W/kg = 13.34 dBW/kg




Certificate No: Z17—97192                              Page 5 of 8


            m                          In Collaboration with
          w7/"7"/,                    s p e a g
            y                         CALIBR&ATION LABORATrORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079           Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com           http://www.chinattl.on




        Impedance Measurement Plot for Head TSL




            10. 00

            0. 000 y                                                                                               A
            10. 00 _                                                                      t nreerimsanstosees
          —20.00                      x                                        §   _zze
          —30.00                                      \              //V/
          —40.00
          —50. 00                                                              *
        |PMBB 511 smith (R+jX) scale 1.000u [Fi bel]
           >»1   2.4500000 GHz   54.533 n     4.6535 0    302.30—pR
                                                                       ge




        |t Sut2mGe                                               IFBW 180 Hz                        Stho 2.65 Getr e




Certificate No: Z17—97192                                Page 6 of 8


           %’                          in Collaboration with

          o                            a
           b                           cmmm
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
         E—mail: etti@chinattl.com      hitp;//www.chinattl.en


DASY5 Validation Report for Body TSL                                         Date: 10.12.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 751
  Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1: 1
  Medium parameters used: f = 2450 MHz; 0 = 1.957 S/m; s, = 53.32; p = 1000 kg/m*
  Phantom section: Center Section
  Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

             *    Probe; EX3DV4 — SN3846; ConvF(7.31,7.31,7.31); Calibrated: 1/13/2017;
             *    Sensor—Surface; 1.4mm (Mechanical Surface Detection)
             «_   Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
             *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
             *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                  (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=Smm, dz=5mm
        Reference Value = 98.93 V/m; Power Drift =—0.03 dB
        Peak SAR (extrapolated) = 25.9 W/kg
        SAR(I g)=12.9 W/kg; SAR(10 g) = 6.05 W/kg
        Maximum value of SAR (measured) = 20.9 W/kg




            —8.77


            ~13.15


            ~17.54


            —21.92




Certificate No: Z17—97192                            Page 7 of 8


                                       In Collaboration with

           @7"[‘L, a
                \Gpgee~"               CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: cttf@chinattl.com      http://www.chinattl.on


         Impedance Measurement Plot for Body TSL

           Trl s11 Log Mag 10.00d8/ ref
                                        0.000d8 [F1]
             50.00      7.4500000 chz —23.345de




                                                                                                                   h
                                                                             §       enczie                        .


                                                                             Y————
                                                                      ue*
                                                               __./
            40. 00

           —50. 00                                       —                   %            ——   —               _
         PHR§ s11 smith (R+]X) scale 1.000U [F1 bel]
           >L   2.4500000 GHz    51.497 n   6.7573 n    438.956—98




                                                                          [ ~~




                                                                      4




                                                                      A.
        [1 Stat 2250                                           iFaW 100 He                     Stap 2.65 GHte: Rgl]i




Certificate No: Z17—97192                              Page 8 of 8



Document Created: 2019-08-07 10:43:48
Document Modified: 2019-08-07 10:43:48

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC