FCC SAR_Appendix C_part 2

FCC ID: 2AIRN-002

RF Exposure Info

Download: PDF
FCCID_3974936

                               ®                         .
          A                         In Collaboration with

         =_‘"/"[‘/, a
          ues                        CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http://www.chinattl.cn



                               Receiving Pattern (®), 0=0°

              f=600 MHz, TEM                                                    f=1800 MHz, R22




  1.0     —        j       :        j        A       {           :          j   i   j   :     f     :    j

                                                            Rollfo]
                           {[—«<7100MHz           __—+—600MHz __—+—1800MHz              __—+— 2500MHz]
                               Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z18—60104                                    Page 8 of 11


                                       ®
                                            In Collal    ion with

        —                                  S
         y
       Add No.51 Xueyuan Road, Hai ian District, Beijing, 100191, Chima
       Tel +86—10—62304633—2512     Fax +86—10—62304633—2504
       E —mal 1: cttl@chimnattl.com Http://www.chinattl.cn



                                           Dynamic Range f(SARneaq)
                                                        (TEM cell, f 900 MHz)

                                                                                      EELbScccclss




                                 «442242004242   stt                                                   L 2424 EAdd44.....
             [aiJjreubis induy




                                                                     144525044




                                                                                          Abwss=sles       +¥st sA 4d d ues sn



                                                                               T                       T    Podd      l          T       P P   Prd III

                                                                                            1
                                                                     10               10                             10"                           10
                                                                      SAR[mW/cm*
                                                          AM—)not compensated      —@— compensated
         [aplhou3
                           o




                                                 10°                                10‘
                                                                     SAR[mW/cm*]
                                      |___ E®=\not compensated                              —#&— compensated                         |
                                           Uncertainty of L inearli ty Assessment: £0.9% (k=2)
Certificate No 718—60104                                        Page 9 of 11


         4WC \n Collsboration with
        _‘"/"[‘[,
        ie
                  a
                  CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com                                Http://www.chinattl.cn



                                    Conversion Factor Assessment
      f=750 MHz, WGLS R9(H_convPF)                                                             f=1750 MHz, WGLS R22(H_convF)
               3.50                                                                                     30.00

               3.00                                                                                     25.00
               2.50
                                                                                                        20.00
           %z.oo                                                                                   %
           g                                                                                       §15.oo
           & 1.50                                                                                  KE
           6                                                                                       6
                                                                                                        10.00
               1.00

               0.50                                                                                      $99
               §i6g       Es scu                 yz ,                                                    §1g Euiss Li ns Ecce Lacl
                      0                     20            40            60      80       100                @0    4000  200  300   400     5p0   &   o
                                                                2{mm]                                                              2{mm]




                               Deviation from Isotropyin Liquid
                                                 1.0

                                                 0.8

                                                 0.6
                                                 0.4

                                                  0.2
                                   7 Axis




                                                  0.0
                                                  —0.2
                                                  —0.4
                                                  —0.6
                                                   —0.8
                                                   -1.8


                                                                   150
                                                                  *    200
                                                                      4*’8       250




                                                               —1.0   —0.80 —060 —0.40 —0.20   0        020     040   060   080   10
                                                 Uncertainty of Spherical Isotropy Assessment: +3.2% (K=2)



Certificate No: Z18—60104                                                        Page 10 of 11


         —MmiWeeA® n Collsboration with
        =_‘/"/‘/,  a
        ~iiaaagys" CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com       Http://www.chinattl.cn


  DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3578


    Other Probe Parameters
   Sensor Arrangement                                                      Triangular

   Connector Angle (°)                                                         169.7

   Mechanical Surface Detection Mode                                         enabled

   Optical Surface Detection Mode                                            disable

    Probe Overall Length                                                      337mm

    Probe Body Diameter                                                        10mm

    Tip Length                                                                  9mm

    Tip Diameter                                                              2.5mm

    Probe Tip to Sensor X Calibration Point                                     1imm

    Probe Tip to Sensor Y Calibration Point                                     1mm

    Probe Tip to Sensor Z Calibration Point                                     1imm

    Recommended Measurement Distance from Surface                             1.4mm




Certificate No: Z18—60104                              Page 11 of 11


                 im®                     In Collaboration with                       wl
                                                                                      :,/                            Bm




                                                                                               KI’
                                                                                     /,




                                                                               /,,
                 @=®‘7/"/"‘J, s_p_
                              e    a                                                                                 AhkaA
                 amzemyy7                CALIBRATION LABORATORY                                CNAs
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China ’?
                                                                               r/ $ v CAUBRATION
             Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504           /,,/,,h‘\“\                         CNAS LO570
             E—mail: cttl@chinattl.com       http:/www.chinattl.en
      Client            UnionTrust                                          Certificate No:          2Z18—60057

CALIBRATION CERTIFICATE

Object                                     D2450V2 — SN: 809

Calibration Procedure(s)                   FP—211—003—01

                                           Calibration Procedures for dipole validation kits

Calibration date:                          March 22, 2018


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22#3)c and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #               Cal Date(Calibrated by, Certificate No.)            Scheduled Calibration
 Power Meter        NRVD          102083            01—Nov—17 (CTTL, No.J17X08756)                                Oct—18
 Power sensor       NRV—Z5        100542            01—Nov—17 (CTTL, No.J17X08756)                                Oct—18
 Reference Probe EX3DV4           SN 7464           12—Sep—17(SPEAG,No.EX3—7464_Sep17)                            Sep—18
 DAE4                             SN 1525           02—Oct—17(SPEAG,No.DAE4—1525_Oct17)                           Oct—18

 Secondary Standards              ID #              Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 Signal Generator E4438C          MY49071430        23—Jan—18 (CTTL, No.J18X00560)                                Jan—19
 NetworkAnalyzer E5239A           MY55491241         29—Jun—17 (CTTL, No.J18X00561)                               Jun—18



                                   Name                          Function                                    Signature
Calibrated by:                   Zhao Jing                 SAR Test Engineer                                  é f/

Reviewed by:                     Lin Hao                   SAR Test Engineer                               ‘fifi%

Approved by:                     Qi Dianyuan               SAR Project Leader                         _éqfl/

                                                                                      Issued: March 25, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z18—60057                             Page 1 of 8


           Ag                         In Collaboration with

          x7"7J,
          Wimaaee>""
                     a
                     CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com       http:/Awww.chinattl.en


Glossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
e Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
   point exactly below the center marking of the flat phantom section, with the arms oriented
   parallel to the body axis.
e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
   positioned under the liquid filled phantom. The impedance stated is transformed from the
   measurement at the SMA connector to the feed point. The Return Loss ensures low
   reflected power. No uncertainty required.
e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
   No uncertainty required.
   SAR measured: SAR measured at the stated antenna input power.
   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
e     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.




Certificate No: Z18—60057                            Page 2 of 8


                                      In Collaboration with

          u@=‘/"/L                    $_       p__e_          _a       _0
           jz                         CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079          Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com          http:/www.chinattl .Cn


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                         DASY52                                     52.10.0.1446

     Extrapolation                                Advanced Extrapolation

     Phantom                                       Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                          10 mm                                   with Spacer

     Zoom Scan Resolution                             dx, dy, dz = 5 mm

     Frequency                                        2450 MHz +1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature              Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  39.2                1.80 mho/m

     Measured Head TSL parameters                         (22.0 £0.2) °C             40.3 + 6 %          1.84 mho/m + 6 %

     Head TSL temperature change during test                  <1.0 °C
SAR result with Head TSL
     SAR averaged over 1 Cm" (1 g) of Head TSL                          Condition
     SAR measured                                                   250 mW input power                 13.1 mW/g

     SAR for nominal Head TSL parameters                             normalized to 1W         52.2 mW /g £ 18.8 % (k=2)

     SAR averaged over 10 cm (10 g) of Head TSL                         Condition

     SAR measured                                                   250 mW input power                 6.09 mW /g

     SAR for nominal Head TSL parameters                             normalized to 1W         24.3 mW ig £ 18.7 % (k=2)
Body TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature              Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  52.7                1.95 mho/m

     Measured Body TSL parameters                         (22.0 £ 0.2) °C            51.3 £ 6 %          2.00 mho/m + 6 %

     Body TSL temperature change during test                  <1.0 °C

SA R result with Body TSL
     SAR averaged over 1_cnm" _(1 g) of Body TSL                        Condition
     SAR measured                                                   250 mW input power                 12.9 mW/g

     SAR for nominal Body TSL parameters                             normalized to 1W         50.7 mW ig + 18.8 % (k=2)
     SAR averaged over 10 cm (10 g) of Body TSL                         Condition

     SAR measured                                                   250 mW input power                 5.93 mW /g
     SAR for nominal Body TSL parameters                            normalized to 1W          23.5 mW ig + 18.7 % (k=2)


Certificate No: Z18—60057                             Page 3 of 8


           Ag                           In Collaboration with

          @77‘J, a
          *\                            CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      http:/www.chinattl.en


 Appendix (Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   49.50+ 5.650
     Return Loss                                                             —24.90B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   50.00+ 71990
     Return Loss                                                             —22.9d0B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                        1.067 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                            SPEAG




Certificate No: Z18—60057                             Page 4 of 8


          !®                          In Collaboration with
         e=x‘/"[‘]J,
         .
                     a
                     CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       http://www.chinattl.en


DASYS5 Validation Report for Head TSL                                       Date: 03.22.2018
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 809
  Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
  Medium parameters used: £= 2450 MHz; 0 = 1.841 S$/m; sr =40.32; p = 1000 kg/m3
  Phantom section: Center Section
  Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7464; ConvF(7.89, 7.89, 7.89); Calibrated: 9/12/2017;
            e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            e    Electronics: DAEB4 Sn1525; Calibrated: 10/2/2017
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 102.7 V/m; Power Drift = 0.01 dB
        Peak SAR (extrapolated) = 27.4 W/kg
        SAR(I g) =13.1 W/kg; SAR(10 g) = 6.09 W/kg
        Maximum value of SAR (measured) = 21.9 W/kg


           dB
           0


            ~4.A6


            —8.92


            13.39


            —17.85
                          >
                         L«
            —22.31

                       0 dB =21.9 W/kg =13.40 dBW/kg




Certificate No: Z18—60057                            Page 5 of 8


          A@                           In Collaboration with

          e=‘[/"]"/,
          C
                     a
                     CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        http://www.chinattl.en




        Impedance Measurement Plot for Head TSL


                      Tr 1 _S11 LogM 10.00dB/ 0.000B
          50.00                                                                   2450 GHz    |—24.91 dB




                                                                            v
          40.00
           30.00
          20.00
           10.00
               0.00
          —10.00
          —20.00
          —30.00
          ~40.00
          —50.00
           1          Ch1: Start_2.25000 GHz —                                        Stop 2.65000 GHz
                  _TR s11 smitn 1.0000/ 1.000
                                                                            »11   2450GHz      49540
                                                                                  366.91 pH     5.650




         "2       _ >Ch1: Start 225000 GHz —                                          Stop 2.65000 GHz




Certificate No: Z18—60057                            Page 6 of 8


           A®                         In Collaboration with
         ©‘/"/‘]J, s p e a q
           es                         CALIBRATION LABORATORY
        Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com       hitp://www.chinattl.en


DASY5 Validation Report for Body TSL                                     Date: 03.22.2018
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 809
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1: 1
   Medium parameters used: £= 2450 MHz; 0 = 1.998 S/m; s, = 51.28; p = 1000 kg/m3
   Phantom section: Left Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2007)
DASYS5 Configuration:

            e    Probe: EX3DV4 — SN7464; ConvF(8.09, 8.09, 8.09); Calibrated: 9/12/2017;
            e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            e    Electronics: DAE4 Sn1525; Calibrated: 10/2/2017
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 97.63 V/m; Power Drift = 0.01 dB
        Peak SAR (extrapolated) = 27.2 W/kg
        SAR(I g) =12.9 W/kg; SAR(10 g) = 5.93 W/kg
        Maximum value of SAR (measured) = 21.6 W/kg


           dB
           0


            ~4.55


            —9.09


            ~13.64


            18.18


            22.13
                          £—
                       0 dB =21.6 W/kg = 13.34 dBW/kg




Certificate No: Z18—60057                            Page 7 of 8


          i’TfL                       ;Collaborationewith 3

          *\                          CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      http://www.chinattl.en


        Impedance Measurement Plot for Body TSL


                   Tr 1_S11 LogM 10.004B/ 0.00dB
          50.00                                                                 2450 GHz    |—22.88 dB




                                                                            v
          40.00
           30.00
           20.00
           10.00
            0.00
          —10.00
          —20.00                                               |——yyge—=—
          —30.00
         ~40.00
         —50.00
           1        Ch1: Start_2.25000 GHz —                                        Stop 2.65000 GHz
                   W511 smitn 1.000u7 1.000
                                                                                2450 GHz     49.97 0
                                                                                467.25 pH     7190




          2        »Ch1: Start 225000 GHz —                                         Stop 2.65000 GHz |




Certificate No: Z18—60057                            Page 8 of 8


   ¢    .                                                              w¥ uy
Calibration Laboratory of                                           \“\\\\_//I/4
          a
Schmid & Partner
                                                                                   >,                       S Schweizerischer Kalibrierdienst
                                                                 i\'a\\'eg.—%fi                              c          Service suisse d‘étalonnage
   Engineering AG                                                 2y                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      * n                                       S          Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client        Auden                                                                             Certificate No: D5GHzV2—1040_Jun18

CALIBRATION CERTIFICATE
 Object                                D5GHzV2 — SN:1040


 Calibration procedure(s)             QA CAL—22.v3
                                      Calibration procedure for dipole validation kits between 3—6 GHz



 Calibration date:                    June 28, 2018



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                                Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                           Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                                 Apr—19
 Power sensor NRP—Z91                   SN:    103245             04—Apr—18 (No. 217—026783)                                Apr—19
 Reference 20 dB Attenuator             SN:    5058 (20k)         04—Apr—18 (No. 217—02682)                                 Apr—19
 Type—N mismatch combination            SN:    5047.2 / 06827     04—Apr—18 (No. 217—02683)                                 Apr—19
 Reference Probe EX3DV4                 SN:    3503               30—Dec—17 (No. EX3—3503_Dec17)                            Dec—18
 DAE4                                   SN:    601                26—0ct—17 (No. DAE4—601_Oct17)                            Oct—18

 Secondary Standards                    ID #                      Check Date (in house)                                     Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                         In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                         In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                         In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                         In house check: Oct—18
 Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—17)                         In house check: Oct—18


                                        Name                                       Function                                 Signature
 Calibrated by:                        Jeton Kastrati                          Laboratory Technician

                                                                                                                   |
 Approved by:                           KatJa Pokovic                          Technical Manager                   %g




                                                                                                                           Issued: June 28, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D5GHzV2—1040_Jun18                                   Page 1 of 16


    s      1                                                   w9 1991
Calibration LabOratOry of
      i                                                    s\s\\\\L\_/L/,I/>   S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           i\hgmfi              c   Service suisse d‘étalonnage
  Engineering AG                                           ze u.                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                {'/,1//A\\\\\}     S   swiss Calibration Service
                                                               uluilibs

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    *    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D5GHzV2—1040_Jun18                             Page 2 of 16


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.1

     Extrapolation                                   Advanced Extrapolation
     Phantom                                       Modular Flat Phantom V5.0
     Distance Dipole Center — TSL                              10 mm                             with Spacer

     Zoom Scan Resolution                         dx, dy = 4.0 mm, dz =1.4 mm        Graded Ratio = 1.4 (Z direction)
                                                       5200    MHz   + 1   MHz
                                                       5300    MHz   +1    MHz
     Frequency                                         5500    MHz   +1    MHz
                                                       5600    MHz   +1    MHz
                                                       5800    MHz   +1    MHz



Head TSL parameters at 5200 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           36.0               4.66 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   36.1 + 6 %         4.55 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C            se«                     ar+


SAR result with Head TSL at 5200 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     7.88 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        78.8 Wikg * 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                  2.27 W/ikg
     SAR for nominal Head TSL parameters                     normalized to 1W         22.7 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                    Page 3 of 16


Head TSL parameters at 5300 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.9                4.76 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   36.0 + 6 %          4.66 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            ~——>                    ««»—

SAR result with Head TSL at 5300 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        100 mW input power                     8.22 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        82.2 W / kg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                     2.36 Wi/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        23.6 W/kg + 19.5 % (k=2)

Head TSL parameters at 5500 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.6                4.96 mho/m
     Measured Head TSL parameters                              (22.0 £ 0.2) °C   35.7 +6 %           4.86 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            «>                      s«=>


SAR result with Head TSL at 5500 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        100 mW input power                     8.34 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        83.4 Wikg x 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                     2.38 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        23.8 W/kg + 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                   Page 4 of 16


Head TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.5                5.07 mho/m

     Measured Head TSL parameters                              (22.0 + 0.2) °C   35.5 + 6 %          4.97 mho/m + 6 %

     Head TSL temperature change during test                      <0.5°C             ~——                     ~———

SAR result with Head TSL at 5600 MHz

     SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition

     SAR measured                                        100 mW input power                     8.54 Wi/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        85.3 W/kg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                  2.45 Wi/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.5 Wikg * 19.5 % (k=2)

Head TSL parameters at 5800 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.3                5.27 mho/m

     Measured Head TSL parameters                              (22.0 + 0.2) °C   35.2 + 6 %          5.18 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            =———                    ose—


SAR result with Head TSL at 5800 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     8.07 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         80.6 W/kg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                  2.30 Wi/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         23.0 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                    Page 5 of 16


Body TSL parameters at 5200 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           49.0               5.30 mho/m
     Measured Body TSL parameters                              (22.0 £ 0.2) °C   47.0 x6 %          5.45 mho/m + 6 %

     Body TSL temperature change during test                      <0.5 °C            ««+~                   «ons

SAR result with Body TSL at 5200 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition

     SAR measured                                        100 mW input power                     7.58 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         75.2 Wikg # 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                     211 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         20.9 W/kg x 19.5 % (k=2)



Body TSL parameters at 5300 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.9               5.42 mho/m

     Measured Body TSL parameters                              (22.0 £0.2) °C    46.8 +6 %          5.58 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C                                   ———


SAR result with Body TSL at 5300 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        100 mW input power                     7.70 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W         76.4 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                     216 Wkg
     SAR for nominal Body TSL parameters                     normalized to 1W         21.4 W/kg * 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                    Page 6 of 16


Body TSL parameters at 5500 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.6               5.65 mho/m

     Measured Body TSL parameters                              (22.0 £ 0.2) °C   46.5 + 6 %         5.85 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C            ~——                    xee«

SAR result with Body TSL at 5500 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition

     SAR measured                                        100 mW input power                     8.25 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         81.9 Wikg £ 19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  2.28 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         22.6 Wikg + 19.5 % (k=2)

Body TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.5               5.77 mho/m

     Measured Body TSL parameters                              (22.0 + 0.2) °C   46.3 +6 %          5.99 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C            ————                   ————


SAR result with Body TSL at 5600 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        100 mW input power                     8.21 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         81.5 W/kg + 19.9 %(k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  2.29 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         22.7 Wikg £ 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                    Page 7 of 16


Body TSL parameters at 5800 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.2               6.00 mho/m

     Measured Body TSL parameters                              (22.0 £ 0.2) °C   46.0 + 6 %         6.26 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C

SAR result with Body TSL at 5800 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition

     SAR measured                                        100 mW input power                     7.79 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        77.3 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                     2.15 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         21.3 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1040_Jun18                    Page 8 of 16


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL at 5200 MHz

     Impedance, transformed to feed point                  50.5 Q — 8.6 jQ

     Return Loss                                             —21.3 dB


Antenna Parameters with Head TSL at 5300 MHz

     Impedance, transformed to feed point                  49.3 Q — 2.8 jQ

     Return Loss                                             — 30.8 dB


Antenna Parameters with Head TSL at 5500 MHz

     Impedance, transformed to feed point                  50.2 Q — 6.1 jQ

     Return Loss                                             — 24.4 dB


Antenna Parameters with Head TSL at 5600 MHz

     Impedance, transformed to feed point                  56.8 Q — 2.0 jQ
     Return Loss                                             —23.6 dB


Antenna Parameters with Head TSL at 5800 MHz

     Impedance, transformed to feed point                  54.4 Q — 1.0 jQ
     Return Loss                                              — 27.3 dB



Antenna Parameters with Body TSL at 5200 MHz

     Impedance, transformed to feed point                  49.8 Q — 7.7 jQ
     Return Loss                                              —22.3 dB


Antenna Parameters with Body TSL at 5300 MHz

     Impedance, transformed to feed point                  49.1 Q —1.8jQ
     Return Loss                                             —34.1 dB

Antenna Parameters with Body TSL at 5500 MHz

     Impedance, transformed to feed point                  50.5 Q — 4.8 jQ
     Return Loss                                             —26.3 dB




Certificate No: D5GHzV2—1040_Jun18          Page 9 of 16


Antenna Parameters with Body TSL at 5600 MHz

      Impedance, transformed to feed point                                           57.6 Q — 24 jQ
      Return Loss                                                                       — 22.6 dB


Antenna Parameters with Body TSL at 5800 MHz

      Impedance, transformed to feed point                                            55.1 Q — 1.3 jQ
      Return Loss                                                                       — 26.0 dB



General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.203 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.

Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                              December 30, 2005




Certificate No: D5GHzV2—1040_Jun18                       Page 10 of 16


DASY5 Validation Report for Head TSL

                                                                                                         Date: 28.06,2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole DSGHzV2; Type: DSGHzV2; Serial: DSGHzV2 — SN:1040

Communication System: UID 0 — CW:; Frequency: 5200 MHz, Frequency: 5300 MHz, Frequency: 5500 MHz,
Frequency: 5600 MHz, Frequency: 5800 MHz
Medium   parameters   used:   f=   5200   MHz;   0   =   4.55   S/m:   &, =   36.1;   p = 1000 kg/m‘ .
Medium   parameters   used:   f=   5300   MHz;   0   =   4.66   S/m;   &, =   36; p   = 1000 kg/m‘ ,
Medium   parameters   used:   f=   5500   MHz:   o   =   4.86   S/m;   &, =   35.7;   p = 1000 kg/m* ,
Medium   parameters   used:   f=   5600   MHz;   0   =   4.97   $/m;   &=     35.5;   p = 1000 kg/m‘ ,
Medium   parameters   used:   f=   5800   MHz;   0   =   5.18   S/m;   £, =   35.2;   p = 1000 kg/m*
Phantomsection: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY532 Configuration:

   *    Probe: EX3DV4 — SN3503; ConvF(5.75, 5.75, 5.75) @ 5200 MHz,
        ConvF(5.5, 5.5, 5.5) @ 5300 MHz, ConvF(5.2, 5.2, 5.2) @ 5500 MHz,
        ConvF(5.05, 5.05, 5.05) @ 5600 MHz, ConvF(4.96, 4.96, 4.96) @ 5800 MHz; Calibrated: 30.12.2017

   *     Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601 (5GHz); Calibrated: 26.10.2017

    *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    e    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5200 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 74.83 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 27.8 W/kg
SAR(I g) =7.88 W/kg; SAR(10 g) = 2.27 W/kg
Maximum value of SAR (measured) = 17.7 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5300 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 75.69 V/m; Power Drift =—0.09 dB
Peak SAR (extrapolated) = 29.8 W/kg
SAR(1 g) = 8.22 W/kg; SAR(10 g) = 2.36 W/kg
Maximum value of SAR (measured) = 18.6 W/kg


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5500 MHz/Zoom Scan, dist=1.Amm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 75.52 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 31.8 W/kg
SAR(1 g) = 8.34 W/kg; SAR(10 g) = 2.38 W/kg
Maximum value of SAR (measured) = 19.3 W/kg




Certificate No: D5GHzV2—1040_Jun18                               Page 11 of 16


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 76.52 V/m; Power Drift =—0.01 dB
Peak SAR (extrapolated) = 31.9 W/kg
SAR(1 g) = 8.54 W/kg; SAR(10 g) = 2.45 W/kg
Maximum value of SAR (measured) = 19.8 W/kg


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5800 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 73.34 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 31.6 W/kg
SAR(I g) = 8.07 W/kg; SAR(10 g) = 2.3 W/kg
Maximum value of SAR (measured) = 19.1 W/kg


       dB
       0

       —6.00
       —12.00
       —18.00
       —24.00
       —30.00
                  0 dB = 17.7 W/kg = 12.48 dBW/kg




Certificate No: D5GHzV2—1040_Jun18              Page 12 of 16


Impedance Measurement Plot for Head TSL




                                                             28 Jun 2018      @7:29:36
         (CHZ sii     4 U ors         150477 a   —8.6300
                                                  "Rs22  0 3.5462 pF         5 200.000 000 MHz


         *
                                                                                                 CH1 Markers
         Del                                                                                      2: 49.344 a
                                                                                                    —2.7595 a
         Ser                                                                                     5.30000 GHz
                                                                                                  3: $0,234 a
                                                                                                    —6.0545 a
                                                                                                 5.50000 GHz
                                                                                                  41 56.771 a
                                                                                                    ~1.9598 a
         five                                                                                    5.G0000 GHz
         16
                                                                                                  St 54.406 a
                                                                                                    —1.0254 A
                                                                                                 5.80000 GHz
         H1d

         CH2



                                                                                                 CH2 Markers

                                                                                                 2:—30,949 d8
         tor                                                                                     S.30000 GHz
                                                                                                 3t—24.968 dB
                                                                                                  5.50000 GHz
                                                                                                 41—23.609 dB
         ee                                                                                      5.60000 GHz
         16
                                                                                                 5:—27.260 dB
                                                                                                  5.90000 GHz
         H1d

                    START 5 000.000 000 MHz                            STOP 6 000.000 000 MHz




Certificate No: D5GHzV2—1040_Jun18                Page 13 of 16


DASY5 Validation Report for Body TSL
                                                                                               Date: 27.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole DSGHzV2; Type: D5GHzV2; Serial: D5SGHzV2 — SN:1040

Communication System: UID 0 — CW; Frequency: 5200 MHz, Frequency: 5300 MHz, Frequency: 5500 MHz,
Frequency: 5600 MHz, Frequency: 5800 MHz
Medium parameters used: f = 5200 MHz; 0 = 5.45 $/m; &, = 47; p = 1000 kg/m‘ ,
Medium parameters used: f = 5300 MHz; 0 = 5.58 $/m; &, = 46.8; p = 1000 kg/m‘,
Medium parameters used: f = 5500 MHz: 0 = 5.85 S/m; &, = 46.5; p = 1000 kg/m‘ ,
Medium parameters used: £ = 5600 MHz; o = 5.99 S/m; &, = 46.3; p = 1000 kg/m‘ ,
Medium parameters used: f = 5800 MHz; 0 = 6.26 S/m; &, = 46; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e     Probe: EX3DV4 — SN3503; ConvF(5.35, 5.35, 5.35) @ 5200 MHz,
         ConvF(5.15, 5.15, 5.15) @ 5300 MHz, ConvF(4.7, 4.7, 4.7) @ 5500 MHz,
         ConvF(4.65, 4.65, 4.65) @ 5600 MHz, ConvF(4.53, 4.53, 4.53) @ 5800 MHz; Calibrated: 30.12.2017

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601 (5GHz); Calibrated: 26.10.2017

    *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

    «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5200 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.84 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 28.1 W/kg
SAR(I g) = 7.58 W/kg; SAR(10 g) =2.11 W/kg
Maximum value of SAR (measured)= 17.1 W/kg


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5300 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.55 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 29.6 W/kg
SAR(1 g) =7.7 W/kg; SAR(10 g) = 2.16 W/kg
Maximum value of SAR (measured) = 17.6 W/kg


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5500 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value =69.16 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 33.4 W/kg
SAR(1 g) = 8.25 W/kg; SAR(10 g) = 2.28 W/kg
Maximum value of SAR (measured) = 19.3 W/kg




Certificate No: D5GHzV2—1040_Jun18                Page 14 of 16


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 68.09 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 34.0 W/kg
SAR(I g) = 8.21 W/kg; SAR(10 g) = 2.29 W/kg
Maximum value of SAR (measured) = 19.5 W/kg


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5800 MHz/Zoom Scan, dist=1.4mm
(8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 63.60 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 33.5 W/kg
SAR(I g) =7.79 W/kg; SAR(10 g) = 2.15 W/kg
Maximum value of SAR (measured) = 18.8 W/kg




                  0 dB = 18.8 W/kg = 12.74 dBW/kg




Certificate No: D5GHzV2—1040_Jun18               Page 15 of 16


Impedance Measurement Plot for Body TSL




                                                                27 Jun 2018     13:35:54
         CHZ sii     4o uors         {149.781 8    —7.7109a ©.96393 pF         5 200.000 dad MHz
                                                  * f.

                                                                                                   CH1 Markers
         Del
                                                                                                    2: 49.148 a
                                                                                                        .7675 a
         Cor                                                                                       5.30000 GHz




         Avg
         16


         H1d


          CH2




                                                                                                   CH2 Markers

         Cor
                                                                                                   2:—34.067 dB
                                                                                                    5.30000 GHz
                                                                                                   3:—26.294 dB
                                                                                                    5.50000 GHz
                                                                                                   41—22.632 dB
         Av                                                                                         5.G0000 GHz
         16°
                                                                                                   §:—25.953 dB
                                                                                                    5.80000 GHz
         H1d

                   START 5 000.000 000 MHz                               STOP 6 000.000 000 MHz




Certificate No: D5GHzV2—1040_Jun18                   Page 16 of 16



Document Created: 2018-08-21 16:37:00
Document Modified: 2018-08-21 16:37:00

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC