Appendix_B

FCC ID:

RF Exposure Info

Download: PDF

Test Laboratory: UnionTrust                                                  Date: 2019/5/23


PO1_GSMS850_GPRS12_Left Check_190
DUT: EUT
Communication System: GPRS 850—4solt; Frequency: 836.6 MHz;Duty Cycle: 1:2
Medium: H850 Medium parameters used: £=837 MHz; 0 = 0.899 mho/m; & = 41.9; p = 1000 kg/m3

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.12, 6.12, 6.12); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.400 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=7.72 V/m; Power Drift=0.158 dB
Peak SAR (extrapolated)= 0.564 W/kg
SAR(1 g) = 0.382 mW/g; SAR(10 g) = 0.263 mW/g
Maximum value of SAR (measured) =0.431 mW/g

   dB
    0.000

    —1.86

    —3.72

    —5.58

    —7.44

    —9.30
               0 dB =0.431mW/g


Test Laboratory: UnionTrust                                                  Date: 2019/5726


PO2_GSM1900_GPRS12_Left Cheek_512
DUT: EUT

Communication System: GPRS1900—4slots; Frequency: 1850.2 MHz;Duty Cycle: 1:2
Medium: H1900 Medium parameters used (interpolated): {= 1850.2 MHz; 0 = 1.39 mho/m; 8. =412;
p = 1000 kg/m}

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(5.06, 5.06, 5.06); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 0.283 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=1.66 V/m; Power Drift=0.02 dB
Peak SAR (extrapolated) = 0.355 W/kg
SAR(1 g) = 0.224 mW/g; SAR(10 g) = 0.134 mW/g
Maximum value of SAR (measured) = 0.273 mW/g

   dB
    0.000

    —3.08

    —1.16

    —11.6

    —15.5

    —19.4
               0 dB =0.273mW/g


Test Laboratory: UnionTrust                                                  Date: 2019/5/26


PO3_WCDMA II_RMC12.2K_Left Cheek_9262

DUT: EUT

Communication System: WCDMA Band II; Frequency: 1852.4 MHz:Duty Cycle: 1:1
Medium: H1900 Medium parameters used (interpolated): £= 1852.4 MHz; 0 = 1.39 mho/m; 8. =412;
p = 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(3.06, 5.06, 5.06); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAEA4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.428 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=2.60 V/m; Power Drift=0.031 dB
Peak SAR (extrapolated) = 0.516 W/kg
SAR(1 g) = 0.334 mW/g; SAR(10 g) = 0.199 mW/g
Maximum value of SAR (measured) = 0.397 mW/g

   dB
    0.000

    ~4.16

    —8.32

    "12.5

    16.6

    —20.8
               0 dB =0.397mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/26


PO4_WCDMA IV_RMC12.2K_Left Cheek_1513

DUT: EUT

Communication System: WCDMA Band IV; Frequency: 1752.6 MHz;Duty Cycle: 1:1
Medium: H1750 Medium parameters used: £{=1753 MHz; 0 = 1.39 mho/m; 8.= 39.4; p= 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(5.36, 5.36, 5.36); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 0.553 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=1.69 V/m; Power Drift=0.041 dB
Peak SAR (extrapolated)= 0.660 W/kg
SAR(1 g) = 0.456 mW/g; SAR(10 g) = 0.288 mW/g
Maximum value of SAR (measured) = 0.532 mW/g

   dB
    0.000

    —3.80

    —7.60

    <11.4

    ~15.2

    "19.0
               0 dB =0.532mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/23


PO5_WCDMA V_RMC12.2K_Left Cheek_4233
DUT: EUT

Communication System: WCDMA Band V; Frequency: 846.6 MHz;Duty Cycle: 1:1
Medium: H850 Medium parameters used: £=847 MHz; 0 = 0.909 mho/m; & = 41.8; p= 1000 kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.12, 6.12, 6.12); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.222 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=7.76 V/m; Power Drift=0.055 dB
Peak SAR (extrapolated)= 0.280 W/kg
SAR(1 g) = 0.195 mW/g; SAR(10 g) = 0.145 mW/g
Maximum value of SAR (measured) = 0.225 mW/g

   dB
    0.000

    —1.87

    —3.15

    —5.62

    —7.50

    —9.37
               0 dB =0.225mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5726


PO6_LTE 2_QPSK20M_Left Cheek_18900_1RB_50 Offset
DUT: EUT

Communication System: LTE Band 2; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: H1900 Medium parameters used: £= 1880 MHz; a= 1.42 mho/m; 8.= 41.1; p=1000
kg/m*

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(5.06, 5.06, 5.06); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 0.369 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, daz=5mm
Reference Value=3.02 V/m; Power Drift=0.02 dB
Peak SAR (extrapolated) = 0.490 W/kg
SAR(1 g) = 0.299 mW/g; SAR(10 g) = 0.174 mW/g
Maximum value of SAR (measured) = 0.364 mW/g

   dB
    0.000

    —3.66

    —7.32

    —11.0

    —14.6

    —18.3
               0 dB =0.364mW/g


Test Laboratory: UnionTrust                                                   Date: 2019/5/26


PO7_LTE 4_QPSK20M_Left Cheek_20175_1RB_50 Offset

DUT: EUT

Communication System: LTE Band 4&20M; Frequency: 1732.5 MHz;Duty Cycle: 1:1
Medium: H1750 Medium parameters used (interpolated): £= 1732.5 MHz; 0 = 1.37 mho/m; 8, =39.5;
p = 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(5.36, 5.36, 5.36); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 0.455 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value= 1.79 V/m; Power Drift=—0.04 dB
Peak SAR (extrapolated)= 0.544 W/kg
SAR(1 g) = 0.368 mW/g; SAR(10 g) = 0.231 mW/g
Maximum value of SAR (measured) =0.431 mW/g

   dB
    0.000

    —3.82

    —7.64

    11.5

    15.3

    "19.1
               0 dB =0.431mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/23

PO8_LTE 5_QPSK10M_Left Cheek_20600_1RB_24 Offset
DUT: EUT

Communication System: LTE BandS; Frequency: 844 MHz;Duty Cycle: 1:1
Medium: H900 Medium parameters used: £= 844 MHz; 0 = 0.906 mho/m; & = 41.8; p= 1000 kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.12, 6.12, 6.12); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 0.201 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value= 4.37 V/m; Power Drift=0.091 dB
Peak SAR (extrapolated)= 0.231 W/kg
SAR(1 g) = 0.180 mW/g; SAR(10 g) = 0.138 mW/g
Maximum value of SAR (measured) =0.199 mW/g

   dB
    0.000

    —1.99

    —3.98

    —5.96

    —7.95

    —9.94
               0 dB =0.199mW/g


Test Laboratory: UnionTrust                                                 Date: 2019/931


PO9_LTE 7_QPSK20M_Right Check_21350_1RB_50 Offset
DUT: EUT

Communication System: LTE Band 7; Frequency: 2560 MHz;Duty Cycle: 1:1
Medium: H2600 Medium parameters used: £=2560 MHz; a=2.01 mho/m; 8.= 37.8; p= 1000
kg/m*

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvyF(4.48, 4.48, 4.48); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (interpolated)= 0.064 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=1.17 V/m; Power Drift=0.02 dB
Peak SAR (extrapolated) = 0.097 W/kg
SAR(1 g) = 0.052 mW/g; SAR(10 g) = 0.027 mW/g
Maximum value of SAR (measured) = 0.063 mW/g

   dB
    0.000

    ~4.30

    ~8.60

    —12.9

    —17.2

    —21.5
               0 dB =0.063mW/g


Test Laboratory: UnionTrust                                                 Date: 2019/5/25


P10_LTE 12_QPSK10M_Left Cheek_23095_1RB_24 Offset
DUT: EUT

Communication System: LTE Band 12; Frequency: 707.5 MHz;Duty Cycle: 1: 1
Medium: H750 Medium parameters used (interpolated): £=707.5 MHz; 5 =0.861 mho/m; 8 = 40.8;
p = 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.22, 6.22, 6.22); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 0.126 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=2.88 V/m; Power Drift=—0.02 dB
Peak SAR (extrapolated)= 0.144 W/kg
SAR(1 g) = 0.114 mW/g; SAR(10 g) = 0.088 mW/g
Maximum value of SAR (measured) = 0.125 mW/g

   dB
    0.000

    212

    ~4.24

    —6.36

    —8.48

    10.6
               0 dB =0.125mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/30


P24_802.11b_Left Cheek_11

DUT: EUT

Communication System: Wlan 802.11b; Frequency: 2462 MHz;Duty Cycle: 1:1
Medium: H2450 Medium parameters used: £=2462 MHz; 0 = 1.87 mho/m; 8.= 37.9; p= 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.57, 4.57, 4.57); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (81x101x1): Measurement grid: dx=1 2mm, dy=12mm
Maximum value of SAR (interpolated) = 1.03 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=14.2 V/m; Power Drift=—0.04 dB
Peak SAR (extrapolated)= 1.82 W/kg
SAR(1 g) = 0.809 mW/g; SAR(10 g) = 0.384 mW/g
Maximum value of SAR (measured) = 1.01 mW/g

   dB
    0.000

    ~4.32

    —8.64

    "13.0

    17.3

    —21.6
               0 dB =1.01mW/g


Test Laboratory: UnionTrust                                                  Date: 2019/5722


P11_GSMS850_GPRS12_Rear Face_1cm_128
DUT: EUT

Communication System: GPRS 850—4solt; Frequency: 824.2 MHz;Duty Cycle: 1:2
Medium: B835 Medium parameters used (interpolated): {= 824.2 MHz; 0 = 0.986 mho/m; 8 <56.5;
p = 1000 kg/m}

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.18, 6.18, 6.18); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 0.961 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=16.0 V/m; Power Drift=—0.018 dB
Peak SAR (extrapolated) = 1.21 Wikg
SAR(1 g) = 0.794 mW/g; SAR(10 g) = 0.514 mW/g
Maximum value of SAR (measured) = 0.932 mW/g

   dB
    0.000

    —2.98

    ~5.96

    ~8.94

    —11.9

    —14.9
               0 dB =0.932mW/g


Test Laboratory: UnionTrust                                                    Date: 2019/5/21


P12_GSM1900_GPRS12_Rear Face_1cm_512
DUT: EUT
Communication System: GPRS1900—4slots; Frequency: 1850.2 MHz;Duty Cycle: 1:2
Medium: B1900 Medium parameters used (interpolated): {= 1850.2 MHz; 0 = 1.49 mho/m; £, = 52.7;
p = 1000 ke/m*
DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.79, 4.79, 4.79); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.444 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=8.38 V/m; Power Drift=—0.036 dB
Peak SAR (extrapolated)= 0.624 W/kg
SAR(1 g) = 0.358 mW/g; SAR(10 g) = 0.210 mW/g
Maximum value of SAR (measured) =0.431 mW/g

   dB
    0.000

    —3.24

    —6.48

    ~1.72

    "13.0

    —16.2
               0 dB =0.431mW/g


Test Laboratory: UnionTrust                                                   Date: 2019/5/21


P13_WCDMA II_RMC12.2K._Rear Face_1om_9262

DUT: EUT

Communication System: WCDMA Band II; Frequency: 1852.4 MHz:Duty Cycle: 1:1
Medium: B1900 Medium parameters used (interpolated): {= 1852.4 MHz; 0 = 1.49 mho/m; 8. =52.7;
p = 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvyF(4.79, 4.79, 4.79); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAEA4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.691 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 9.94 V/m; Power Drift=—0.052 dB
Peak SAR (extrapolated) = 0.943 W/kg
SAR(1 g) = 0.547 mW/g; SAR(10 g) = 0.326 mW/g
Maximum value of SAR (measured) = 0.664 mW/g

   dB
    0.000

    —3.34

    —6.68

    —10.0

    134

    16.7
               0 dB =0.664mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/20


P14_WCDMA IV_RMC12.2K_Rear Side_1em_1513

DUT: EUT

Communication System: WCDMA Band IV; Frequency: 1752.6 MHz;Duty Cycle: 1:1
Medium: B1750 Medium parameters used: £= 1753 MHz; 0 = 1.47 mho/m; &.= 54.8; p= 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.95, 4.95, 4.95); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.751 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=9.67 V/m; Power Drift=—0.014 dB
Peak SAR (extrapolated)= 1.06 W/kg
SAR(1 g) = 0.612 mW/g; SAR(10 g) = 0.354 mW/g
Maximum value of SAR (measured) = 0.738 mW/g

   dB
    0.000

    —3.17

    —6.33

    —9.50

    12.7

    15.8
               0 dB =0.738mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/22


P15_WCDMA V_RMC12.2K_Rear Face_1om_4233
DUT: EUT

Communication System: WCDMA Band V; Frequency: 846.6 MHz;Duty Cycle: 1:1
Medium: B835 Medium parameters used: {=847 MHz; 0 = 1.01 mho/m; & = 56.2; p= 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.18, 6.18, 6.18); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 0.483 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, daz=5mm
Reference Value=11.1 V/m; Power Drift=—0.012 dB
Peak SAR (extrapolated) = 0.614 W/kg
SAR(1 g) = 0.404 mW/g; SAR(10 g) = 0.260 mW/g
Maximum value of SAR (measured) = 0.468 mW/g

   dB
    0.000

    ~2.92

    ~5.84

    —8.76

    <11.7

    —14.6
               0 dB =0.468mW/g


Test Laboratory: UnionTrust                                                 Date: 2019/5/21


P16_LTE 2_QPSK20M_Rear Face_1cm_18900_1 RB_50 offset

DUT: EUT

Communication System: LTE Band 2; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: B1900 Medium parameters used: £= 1880 MHz; 0 = 1.52 mho/m; &.= 52.6; p = 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.79, 4.79, 4.79); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 Smm
Maximum value of SAR (interpolated) = 0.424 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=8.05 V/m; Power Drift=—0.070 dB
Peak SAR (extrapolated)= 0.564 W/kg
SAR(1 g) = 0.417 mW/g; SAR(10 g) = 0.247 mW/g
Maximum value of SAR (measured) = 0.395 mW/g

   dB
    0.000

    —3.56

    T12

    ~10.7

    "14.2

    —17.8
               0 dB =0.395mW/g


Test Laboratory: UnionTrust                                                  Date: 2019/5/20


P17_LTE 4_QPSK20M_Rear Face_lem_20175_1 RB_50 offset

DUT: EUT

Communication System: LTE Band 4&20M; Frequency: 1732.5 MHz;Duty Cycle: 1:1
Medium: B1750 Medium parameters used (interpolated): £=1732.5 MHz; 0 = 1.44 mho/m; 8, =34.8;
p = 1000 kg/m}

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvyF(4.95, 4.95, 4.95); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x71x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 1.03 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=12.2 V/m; Power Drift=—0.05 dB
Peak SAR (extrapolated) = 1.21 Wikg
SAR(1 g) = 0.694 mW/g; SAR(10 g) = 0.401 mW/g
Maximum value of SAR (measured) = 0.860 mW/g

   dB
    0.000

    —3.36

    ~6.72

    —10.1

    —13.4

    —16.8
               0 dB =0.860mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5722


P18_LTE 5_QPSK10M_Rear Face_lem_20600_1 RB_24 offset
DUT: EUT

Communication System: LTE Band5; Frequency: 844 MHz;Duty Cycle: 1:1
Medium: B835 Medium parameters used: {=844 MHz; 0 = 1 mho/m; & = 56.3; p= 1000 kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.18, 6.18, 6.18); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated)= 0.465 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, daz=5mm
Reference Value=11.2 V/m; Power Drift=0.001 dB
Peak SAR (extrapolated) = 0.578 W/kg
SAR(1 g) = 0.384 mW/g; SAR(10 g) = 0.246 mW/g
Maximum value of SAR (measured) = 0.444 mW/g

   dB
    0.000

    ~3.02

    ~6.04

    —9.06

    —12.1

    —15.1
               0 dB =0.444mWig


Test Laboratory: UnionTrust                                                Date: 2019/5/29


P19_LTE 7_QPSK20M_Bottom Side_lom_21350_1 RB_50 offset
DUT: EUT

Communication System: LTE Band 7; Frequency: 2560 MHz;Duty Cycle: 1:1
Medium: B2600 Medium parameters used: £=2560 MHz; 0 =2.15 mho/m; &.= 52.5; p = 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.24, 4.24, 4.24); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1125
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (71x101x1): Measurement grid: dx=1 2mm, dy=12mm
Maximum value of SAR (interpolated) = 1.55 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=8.15 V/m; Power Drift=0.04 dB
Peak SAR (extrapolated)= 2.53 W/kg
SAR(1 g) = 1.2 mW/g; SAR(10 g) = 0.532 mW/g
Maximum value of SAR (measured) = 1.62 mW/g

   mWilg
    1.55

    1.24

    0.932

    0.622

    0.313

    0.004


Test Laboratory: UnionTrust                                                   Date: 2019/5/26


P20_LTE 12_QPSK10M_Rear Face_lem_23095_1RB_24 offset
DUT: EUT

Communication System: LTE Band 12; Frequency: 707.5 MHz;Duty Cycle: 1: 1
Medium: B750 Medium parameters used (interpolated): {= 707.5 MHz; 0 = 0.932 mho/m; 8. =55.8;
p = 1000 kg/m®

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(6.4, 6.4, 6.4); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: SAM 2; Type: QD 000 P40 CB; Serial: TP—1376
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (61x81x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 0.213 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value=11.8 V/m; Power Drift=—0.012 dB
Peak SAR (extrapolated)= 0.278 W/kg
SAR(1 g) = 0.173 mW/g; SAR(10 g) = 0.113 mW/g
Maximum value of SAR (measured) =0.201 mW/g

   mWilg
    0.213

    0.171

    0.129

    0.088

    0.046

    0.004


Test Laboratory: UnionTrust                                                  Date: 2019/5/30


P22_802.11b_Rear Face_lem_11

DUT: EUT

Communication System: Wlan 802.11b; Frequency: 2462 MHz;Duty Cycle: 1:1
Medium: B2450 Medium parameters used: £= 2462 MHz; 0 = 2.03 mho/m; &, = 53; p = 1000 kg/m*

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.47, 4.47, 4.47); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn662; Calibrated: 2019/4/11
— Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1125
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (91x81x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (interpolated) = 0.221 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value= 4.14 V/m; Power Drift=—0.007 dB
Peak SAR (extrapolated)= 0.331 W/kg
SAR(1 g) = 0.165 mW/g; SAR(10 g) = 0.083 mW/g
Maximum value of SAR (measured) =0.213 mW/g

   dB
    0.000

    —3.78

    —7.56

    11.3

    —15.1

    18.9
               0 dB =0.213mW/g


Test Laboratory: UnionTrust                                                Date: 2019/5/29


P19_LTE 7_QPSK20M_Rear Face_1cm_21350_1 RB_50 offset

DUT: EUT

Communication System: LTE Band 7; Frequency: 2560 MHz;Duty Cycle: 1:1
Medium: B2600 Medium parameters used: £=2560 MHz; 0 =2.15 mho/m; &.= 52.5; p = 1000
kg/m>

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvF(4.24, 4.24, 4.24); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1125
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (81x71x1): Measurement grid: dx=1 2mm, dy=12mm
Maximum value of SAR (interpolated) = 1.60 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value= 1.86 V/m; Power Drift=0.09 dB
Peak SAR (extrapolated)= 2.51 W/kg
SAR(1 g) = 1.14 mW/g; SAR(10 g) = 0.651 mW/g
Maximum value of SAR (measured) = 1.53 mW/g

   mWilq
    1.60

    1.28

    0.960

    0.640

    0.320

    0.000


Test Laboratory: UnionTrust                                                 Date: 2019/5729


P19_LTE 7_QPSK20M_Rear Face_Ocm_21350_1 RB_50 offset

DUT: EUT

Communication System: LTE Band 7; Frequency: 2560 MHz;Duty Cycle: 1:1
Medium: B2600 Medium parameters used: {=2560 MHz; 0 =2.15 mho/m; &.= 52.5; p = 1000
kg/m*

DASY4 Configuration:
— Probe: ES3DV3 — SN3090; ConvyF(4.24, 4.24, 4.24); Calibrated: 2019/4/12
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n662; Calibrated: 2019/4/11
— Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1125
— ; Postprocessing SW: SEMCAD, V1.8 Build 186


Test/Area Scan (81x71x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (interpolated)= 4.59 mW/g

Test/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=Smm
Reference Value=3.38 V/m; Power Drift=0.04 dB
Peak SAR (extrapolated) = 10.5 Wikg
SAR(1 g) = 4.19 mW/g; SAR(10 g) = 1.56 mW/g
Maximum value of SAR (measured) = 5.77 mW/g

   dB
    0.000

    —5.865

    <11.7

    —17.6

    ~23.4

    29.3
               0 dB =5.77mW/g


Test Laboratory: UnionTrust                                                   Date: 6/24/2019

P01 WLAN 5G_802.11a_Left Cheek_Ch36

DUT: X150

Communication System: 802.11a; Frequency: 5180 MHz;Duty Cycle: 1:1
Medium: H5G Medium parameters used: f = 5180 MHz; σ = 4.659 S/m; εr = 36.501; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(5.4, 5.4, 5.4); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (121x91x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 3.07 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 9.175 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 4.24 W/kg
SAR(1 g) = 0.941 W/kg; SAR(10 g) = 0.259 W/kg
Maximum value of SAR (measured) = 3.03 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/24/2019

P02 WLAN 5G_802.11a_Left Cheek_Ch140

DUT: X150

Communication System: 802.11a; Frequency: 5700 MHz;Duty Cycle: 1:1
Medium: H5G Medium parameters used: f = 5700 MHz; σ = 5.131 S/m; εr = 35.761; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.77, 4.77, 4.77); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (121x91x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.848 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 9.877 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 2.15 W/kg
SAR(1 g) = 0.529 W/kg; SAR(10 g) = 0.174 W/kg
Maximum value of SAR (measured) = 0.831 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/24/2019

P03 WLAN 5G_802.11a_Left Cheek_Ch149

DUT: X150

Communication System: 802.11a; Frequency: 5745 MHz;Duty Cycle: 1:1
Medium: H5G Medium parameters used: f = 5745 MHz; σ = 5.221 S/m; εr = 35.652; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.9, 4.9, 4.9); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (121x91x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.88 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 14.27 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.22 W/kg
SAR(1 g) = 0.780 W/kg; SAR(10 g) = 0.241 W/kg
Maximum value of SAR (measured) = 1.88 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P04 WLAN 5G_802.11a_Front Face_1cm_Ch36

DUT: X150

Communication System: 802.11a; Frequency: 5180 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5180 MHz; σ = 5.413 S/m; εr = 48.443; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.49, 4.49, 4.49); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (111x201x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.44 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 2.152 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 1.96 W/kg
SAR(1 g) = 0.595 W/kg; SAR(10 g) = 0.252 W/kg
Maximum value of SAR (measured) = 1.43 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P05 WLAN 5G_802.11a_Rear Face_1cm_Ch140

DUT: X150

Communication System: 802.11a; Frequency: 5700 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5700 MHz; σ = 6.033 S/m; εr = 47.629; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(3.91, 3.91, 3.91); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (111x201x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.504 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 1.134 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 0.973 W/kg
SAR(1 g) = 0.269 W/kg; SAR(10 g) = 0.084 W/kg
Maximum value of SAR (measured) = 0.504 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P06 WLAN 5G_802.11a_Rear Face_1cm_Ch149

DUT: X150

Communication System: 802.11a; Frequency: 5745 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5745 MHz; σ = 6.162 S/m; εr = 47.503; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.1, 4.1, 4.1); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (111x201x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.21 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 1.544 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 1.89 W/kg
SAR(1 g) = 0.549 W/kg; SAR(10 g) = 0.183 W/kg
Maximum value of SAR (measured) = 1.07 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P07 WLAN 5G_802.11a_Top_0cm_Ch36

DUT: X150

Communication System: 802.11a; Frequency: 5180 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5180 MHz; σ = 5.413 S/m; εr = 48.443; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.49, 4.49, 4.49); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (61x101x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 15.9 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 21.07 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 25.6 W/kg
SAR(1 g) = 5.48 W/kg; SAR(10 g) = 1.49 W/kg
Maximum value of SAR (measured) = 15.6 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P08 WLAN 5G_802.11a_Top_0cm_Ch140

DUT: X150

Communication System: 802.11a; Frequency: 5700 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5700 MHz; σ = 6.033 S/m; εr = 47.629; ρ = 1000 kg/m3

DASY4 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(3.91, 3.91, 3.91); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (61x101x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 3.41 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 20.39 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 7.90 W/kg
SAR(1 g) = 1.23 W/kg; SAR(10 g) = 0.357 W/kg
Maximum value of SAR (measured) = 4.10 W/kg


Test Laboratory: UnionTrust                                                   Date: 6/25/2019

P09 WLAN 5G_802.11a_Top_0cm_Ch149

DUT: X150

Communication System: 802.11a; Frequency: 5745 MHz;Duty Cycle: 1:1
Medium: B5G Medium parameters used: f = 5745 MHz; σ = 6.162 S/m; εr = 47.503; ρ = 1000 kg/m3

DASY5 Configuration:
- Probe: EX3DV4 - SN7351; ConvF(4.1, 4.1, 4.1); Calibrated: 12/14/2018
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn679; Calibrated: 4/23/2019
- Phantom: SAM 1; Type: QD 000 P40 CB; Serial: TP/1378
- ; Postprocessing SW: SEMCAD, V1.8 Build 186


- Area Scan (61x101x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 11.7 W/kg

- Zoom Scan (7x7x6)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 34.28 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 21.7 W/kg
SAR(1 g) = 3.87 W/kg; SAR(10 g) = 1.04 W/kg
Maximum value of SAR (measured) = 11.3 W/kg