SAR_Test_Report_Part2

FCC ID: 2AFZZC3JH

RF Exposure Info

Download: PDF
FCCID_4403713

                                 ®      In Collsborationwith
           -iTTL a
                 e a
                                                       LABORATORY
         Add: No.5 1 Xueyuan
                             Road, Haidian District, Beijing, 100191, China
         Tel: +8
               6. 10—6230463
                               3—2079       Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com          http//www.chinatt.en
 Measurement Condit
                                ions
    Dasy System configuration,
                               as far as not given on page 1.
      DASY Version                                             DasY52                                   52.10.0.1446
      Extrapolation                                Advanced Extrapolation
       Phantom                                      Triple Flat Phantom 5.1C
       Distance Dipole Center — TSL                            10 mm                                  with Spacer
      Zoom Scan Resolution                             dx, dy, dz = 5 mm
       Frequency                                       1900 MHz & 1 MHz


 Head TSL parameters
    The following parameters and calculations were applied.
                                                               Temperature             Permittivity           Conductivity
       Nominal Head TSL parameters                               22.0 °C                  40.0                1.40 mho/m
       Measured Head TSL parameters                         (22.0 £0.2) °C              39.9 £ 6 %          141 mho/m £ 6 %

       Head TSL temperature change during test                   <1.0 °C                   —                        —
 SAR result with Head TSL
    SAR averaged over 1_Cm2°_(1 g) of Head TSL                             Condition
      SAR measured                                                     250 mW input power                 10.1 mW/g
      SAR for nominal Head TSL parameters                               normalized to 1W         40,1 mW /g £ 18.8 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                          Condition

      SAR measured                                                     250 mW input power                 5.19 mW /g

      SAR for nominal Head TSL parameters                               normalized to 1W         20.7 mW Ig £ 18.7 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                               Temperature             Permittivity           Conductivity

      Nominal Body TSL parameters                               22.0 °C                   53.3                1.52 mho/m
      Measured Body TSL parameters                         (22.0 £ 0.2) °C              53.6 46 %           1.53 mho/m + 6 %
      Body TSL temperature change during test                   <1.0 °C                    ——                       —
SAR result with Body TSL
   SAR averaged over 1 cm* _(1 g) of Body TSL                              Condition
     SAR measured                                                   250 mW input power                    9.90 mW /g
     SAR for nominal Body TSL parameters                               normalized to 1W          39.5 mW /g + 18.8 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                           Condition
     SAR measured                                                   250 mW input power                    5.21 mw/g
     SAR for nominal Body TSL parameters                               normalized to 1W          20.8 mW 1g £ 18.7 % (k=2)
Certificate No: Z17—97115                               Page 3 of 8


                                  ®    in Collaborationwith
                   TTL a
                                       CALIERATION LABORATORY
          Add: Iy Xueyuan Road, Haidian
          Tek +3g]s|;                        District, Beijing, 100191, China
          Bomghq §2304633—2079             Pug:+86—10—02308633—2504
              all: ettl@chinattl.com       hitps//wwawchinatten

   Appendix (Additional assessments outside the scope of CNAS L0570)
  Antenna Parameters with Head
                               TSL

       Impadance, transformed to feed point                                     52.00+ 6.5910
       Return Loss                                                                —23.408

  Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                     §52.70+ 8.3510
       Return Loss                                                                —21.408

  General Antenna Parameters and Design

       Elactrical Delay (one direction)                           J               1.302 ns




  After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
  be measured.

  The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
  connected to the second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some
  of the dipoles, small end caps are added to the dipole arms in orderto improve matching when loaded
  according to the position as explained in the "Measurement Conditions® paragraph. The SAR data are not
  affected by this change. The overall dipole lengthis still according to the Standard
  No excessive force must be applied to the dipole arms, because they might bend or the soldered
  connections near the feedpoint may be damaged.



  Additional EUT Data

      Manufactured by                                                                SPEAG




Certificate No: Z17—97115                              Page 4 of 8


           b’
                                     in Collaboration with
           ©77TJI a
           1\-/                      causranon tasor,
         Add: No.51 Xueyuan Road
                                . Haidian District, Beijing.
         Eel: f!{s- mfz:_om:.                                100191, China
                                mw        Fax: +86—10—62304633—2504
          »mail: ¢H @chinaitl.com         bups//wwchinanl.en
DASY5 Validation Report for    Head TSL
Test Laboratory: CTTL,
                       Beijing, China                                          Bate: 08.26.2017
   UT: Dipole 1900 MHz; Typ
                            e: D1900V2; Serial: D1900V
                                                       2 — SN: 54060
    Communication System: UID 0, CW; Fre
                                         quency: 190 0 MHz: Duty Cycle: 1:1
    Medium parameters used: f= 190
                                   0 MHz; 0 = 1.413.8/m; er=39.85;
    Phantom section: Left Section                                  p= 1000 kg/m3
    Mcasurement Standard: DASYS (IEE
                                       EIEC/ANSI ©63.19—2007)
DASY3 Configuration:

             »—   Probe: EX3DV4— SN3617; ConvF(8.26, 8.26, 8.26);
                                                                     Calibrated: 1/23/2017;
             *    Sensor—Surface: 1.4mm (Mechanical Surface
                                                             Detection)
             *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2
                                                               017
             *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial:
                                                                                  1161/1
             +    Measurement SW: DASY52, Version 52.10 (0); SEMCA
                                                                   D X Version 14.6.10
                  (7417)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measure
                                                                           ment grid:
        dx=5mm, dy=Smm, dz=Smm
        Reference Value =94.94 V/m; Power Drift=0.01 dB
         Peak SAR (extrapolated)=19.5 W/kg
        SAR(I g) =10.1 W/kg; SAR(10 g)=5.19 W/kg
        Maximum value of SAR (measured)= 15.9 W/kg




            ~3.64


            ~1.28

            10.93

            14.57


            18.21
                       _

                       0 dB =15.9 W/kg=12.01 dBW/kg



Certificate No: Z17—97115                              Page 5 of 8


         #dd: No.51 Xueyuan Road, Haidis
          T: 86—10—62304633—2079           ian District, Befjing, 100191, China
         E—mail: cri@chinattl.com
                                            ray, +ss-|o-5536'3533-2504
                                           hitp/fwww.chinatten




         Impedance Measurement Plot for Head TSL

          Tri si1 Log Hag 10.00087 nef 0.000d0 [F2]
             50.
                   * i—asvtor en crranr as
            «0. 00      ——                     >
            30.00        _          _1

            20.00     _             —_.

            20,00       —                                          I     —             &       +       —
            0. 0000            —           >           :           +         abe.__1               —       (f
           —20.00                                                                          ‘
           ~20.00

           ~30.00
           ~40.00

           ~50.00
         IMIBRM s11 snith (a+p) scale 1,0000 [FL oe1]
            »2. 2,9000000 ahz 31.991 0 6.5866 o. 551.7,




Certificate No: 217—97115                                  Page 6 of 8


        Add:y No.51 Xueyyuan Roud,       idion n Distric
                                                   Di t, Beljing
                                                             i , 100191, China
                 j
        TEI» *_85-I0-623I]4633-2fl79
                                      Haidia                                i
                                           Fax: +86—10—62304633—2504
          —mall: ctll@chinattl.com         hipolinwwchinattl.en



DASYS Validation Report for Body TSL                                             Date: 08.26.20 17
Test Laboratory: CTTL, Beijing,
                                China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial; D1900V2 —
                                                       SN: 50060
   Communication System: UID 0. CW; Frequency: 1900 MHz;
                                                          Duty Cycle: 1:1
   Medium parameters used: = 1900 MHz; a= 1.528 S/im; ; = 53.55; p=1000 kg/m!
  Phantom section: Center Section
  Measurement Standard: DASY5 (IEEEIEC/ANSI C63.19—2007)
DASYS5 Configuration:

             *   Probe: EX3DV4 — SN3617; ConvF(7.95, 7.95, 7.95); Calibrated: 1/23/2017;
             *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
             *   Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
             +   Phantom: Triple Plat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
             +   Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        System Performance Check/Zoom Scan (7x787) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=Smm, dz=5mm
        Reference Value =91.19 V/m; Power Drif=0.01 dB
        Peak SAR (extrapolated)= 18.1 W/kg
        SAR(I g) = 9.9 W/kg; SAR(LO g)=5.21 Wikg
        Maximum value of SAR (measured) =15.3 W/kg




                       0 dB=15.3 W/kg = 11.85 dBW/kg




Certificate No: Z17—97115                                  Page 7 of 8


C
               !-\-                        in Collaboration with
               2 7TI a
               ‘, CAUB RANON LABOR                                 ATORY
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             1:: : +86—1 fl~623_0463
                                  3<1079       Fax: +86—10—62304633—2504
             E—mail: cttl@chinattl.com         hups//wwachinaltlen

             Impedance Measurement Plot for Bod
                                                                        y TSL

               Tri SX Log nag 15. 09087 nar 0.00008 iT
                 50.0 * pr—
                            romumrec—amore
                 40.00        ———
                 30.00       ————
                 20.00

                10.00
                                                                                      %       F
                 0.000p]
                                                                                 +4       f       1
               ~20.00                                                                                 —f
               ~20.00
               ~30.00
               ~40.00

                 c
              m sia smith (raefx) seate 1.0000 [ri oel]
                s1   2.9000000 GHz   $2.660 0.   s.3483 0   699. 39.   om




    Certificate No: 217—97115                               Pagesofs


                          TL sop—0—a_o _ *\ tikky
                                 ®     in Collaboration with                    C uy




             Add: N        n
                                       epuennion taqnstont                  l‘CNASMWN
                                    oad, Haidian District, Beijing, 100191, Chis     ‘f, f
             Tel +85.ié.’éz;oy:;l;zr%           Tec "B6.To—2304633 2s _                 *4
                                                                                         "h,
                                                                                           *                     CNAS LO570
             E—mail: cttl@chinattl.com          hitp://wwwchinattl.en            ~diachas          7       s zen S
      Client           [TA(Shanghai)            _____           |      [Gertificate No:\ 217—97116


Object                                 (b2asova®su~7a§ es ul 4P l
Calibration Procedure(s)               ee m mm

                                       | Calibration Procedures for dipole validation kits_
Calibration date:                      (August 29,2017 _

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility; environment temperature(2223)c and
humidity<70%.


Callbration Equipment used (M&TE critical for calibration)


Primary Standards               ID #               Cal Date(Calibrated by, Certificate No.)            Scheduled Calibration
 Power Meter   NRVD             102083            22—Sep—16 (CTTL, No.J16X06309)                              Sep—17
 Power sensor NRV—Z5            100595            22—Sep—16 (CTTL, No.J16X06809)                              Sep—17
 Reference Probe EX3DV4         SN 3617           23—Jan—17(SPEAG,No.EX3—3617_Jan17)                         Jan—18
 DAE4                           SN 1331           18—Jan—17(CTTL—SPEAG,No.217—97015)                         Jan—18

 Secondary Standards            ID #              Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 Signal Generator E4438C        MY49071430         13—Jan—17 (CTTL, No.J17X00286)                            Jan—18
 Network Analyzer E5071C        MY46110673         13—Jan—17 (CTTL, No.J17X00285)                            Jan—18



                                 N_amg              _          Functlon —     es               —            Signature
Calibrated by:                |Zhao Jing                SAR Test Engineer
Reviewed by:                  | Lin Hao                  SAR TestEnémeér

Approved by:                  {Q[Dianyuan                SAR Project Leader
                                                                                 Issued: September 1, 2017
This calibration certificate shall not be reproduced except in full without written approvalof the laboratory.



    Certificate No: Z17—97116                            Page 1 of 8


           !\'                         in Collaboration with

          %TTL a
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com       http://wwwchinattl.en

Glossary:
TSL                        tissue simulating liquid
ConvF                       sensitivity in TSL / NORMx,y,z
N/A                        not applicable or not measured

Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
* Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
& Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
   point exactly below the center marking of the flat phantom section, with the arms oriented
   parallel to the body axis.
* Feed Point Impedance and Return Loss: These parameters are measured with the dipole
   positioned under the liquid filled phantom. The impedance stated is transformed from the
   measurement at the SMA connector to the feed point. The Return Loss ensures low
   reflected power. No uncertainty required.
* Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
   No uncertainty required.
   SAR measured: SAR measured at the stated antenna input power.
   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
   connector.
* SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


   The reported uncertainty of measurement is stated as the standard uncertainty of
   Measurement multiplied by the coverage factor k=2, which for a normal distribution
   Corresponds to a coverage probability of approximately 95%.



Certificate No: Z17—97116                            Page 2 of 8


                          top—.>




                                         *       in Collaboration with

                           TTL a
                               CALIBRATION LABORArory
                   Add: No.S1 Xucyuan Road, Haidian District, Neljing, 100191, China
                   Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
                   E—mail: ctt!@chinattl.com       hitps//www.chinattl.en

           Measurement Conditions
              DASY system configuration, as far as not given on page 1
                DASY Version                                         DASY52                                        52.10.0.1446
                Extrapolation                              Advanced Extrapolation

                Phantom                                     Triple Flat Phantom 5.1C

                Distance Dipole Center — TSL                         10 mm                                       with Spacer

                Zoom Scan Resolution                            dx, dy, dz =5 mm
                Frequency                                       2450 MHz £ 1 MHz


           Head TSL parameters
              The followingparameters and calculations were applied.
                                                                      Temperature                 Permittivity           Conductivity
                Nominal Head TSL parameters                              22.0°C                      39.2                1.80 mho/m
                Measured Head TSL paramaters                         (22.0 £0.2) °C                39.7 26 %           1,82 mho/m £ 6 %

                Head TSL temperature change during test                  <1.0 °C                      —                        hane
           SAR result with Head TSL
              SAR averaged over 1_Cm"_(1g) of Head TSL                                Condition
                SAR measured                                                   250 mW input power                    13.2 mW/g

                SAR for nominal Head TSL parameters                              normalized to 1W           52.6 mW ig 4 18.8 % (k=2)

                SAR averaged over 10 cm° (10 g) of Head TSL                           Condition
                SAR measured                                                   250 mW input power                    6.16 mW /g

                SAR for nominal Head TSL parameters                              normalized to 1W           24.6 mW ig £ 18.7 % (k=2)
           Body TSL parameters
              The following parameters and calculations were applied.
                                                                      Temperature                 Permitivity            Conductivity

                Nominal Body TSL parameters                                 22.0 °C                  §2.7                1.95 mho/m

                Measured Body TSL parameters                         (22.0 £0.2) °C                525146 %            1.94 mho/m £ 6 %
                Body TSL temperature change during test                     <1.0°C                    us                       se
           SAR result with Body TSL
                SAR averaged over 1 cm‘ (1 g) of Body TSL                             Condition
                 SAR measured                                                   250 mW input power                    12.7 mW /g

                SAR for nominal Body TSL parameters                              normalized to 1W           50.8 mW Ig + 18.8 % (k=2)

                SAR averaged over 10 _cm> _(10 g) of Body TSL                         Condition
                 SAR measured                                                   250 mW input power                   587 mW 1g

                 SAR for nominal Body TSL parameters                             normalized to 1W           23.5 mW 1g 2 18.7 % (k=2)

           Certificate No: Z17—97116                             Page 3 of 8




vems eopores ornan e we we roprenracen cnnupe mm gony c


           ’                            in Collaboration with
          g
          L                           __MLL_
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com         http:/www.chinattl.en

 Appendix (Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL

     Impedance, transformed to feed point                                   53.40+ 4.29)0

     Returm Loss                                                              —25.50B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   51.00+ 6.6110

     Return Loss                                                             —23.6dB


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                        1,265 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in orderto improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

    [ Manufactured by                                                           SPEAG                    ]




Certificate No: Z17—97116                             Page 4 of 8


                                      Caueranon Lagorarory
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ctt!@chinattl.com       http/frww.chinatth.en


DASYS Validation Report for Head TSL                                        Date: 08.29.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial; D2450V2 — SN: 786
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
  Medium parameters used: £= 2450 MHz; 0 = 1.822 S/m; er = 39.65; p = 1000 kg/m3
  Phantom section; Left Section
  Measurement Standard: DASYS (IEEE/AIEC/ANSI €63.19—2007)
DASYS Configuration:

            +    Probe: EX3DV4 — SN3617; ConvF(7.74, 7.74, 7.74); Calibrated: 1/23/2017;
            *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
            +    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurementgrid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value=105.1 V/m; Power Drift= —0.03 dB
        Peak SAR (extrapolated)=27.5 W/kg
        SAR(I g)=13.2 W/kg; SAR(10 g)=6.16 W/kg
        Maximum value of SAR (measured) =22.2 W/kg


           dB
            0

            ~4.33

            —8.66


            12.98

            "17.31

                         C
            —21.64
                       0 dB =22.2 W/kg=13.46 dBW/kg




Certificate No: Z17—97116                           Page 5 of8


                              t       in Collaboration with

                                      CALIBRATION LABORKTORy
        Add: No.51 Xueyuan Road, Haidian District, Beljing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +6—1 M?)fl‘fl]—lim
        E—mail: e@chinatt.com           hitp/wwwchinatilen




        Impedance Measurement Plot for Head TSL


         Tri iX Log mag 10.00007 wer 5. 00008 [FI]
           00(51 2322300000 one —33.30. ds
           40.00
           30.c0
           20.00
           10.60
           0. 000 B                                                         4
          —10.00
          —20.00

          —30.00
          ~40.c0
          —50.c0
        NIRBJ 511 seith (ReJX) Scale 1.000U [¢1 bel]
          »2. 2.4500000 cz 83.436 0. 4.2806 0. 278.3




Certificate No: 217—97116                            Page 6 ofs


          @TTL s‘p.e—a
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel:+86—10—62304833—2079        Fax: +86—10—62304633—2504
        E—mail: ctti@chinattl.com       hitps/wwchinattl.en

DASY5 Validation Report for Body TSL                                        Date: 08.29.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial; D2450V2 — SN: 786
  Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
  Medium parameters used: {=2450 MHz; 0= 1.943 S/m; &,= 52.45; p= 1000 kg/m*
  Phantom section: Right Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)
DASY5 Configuration:

            +    Probe: EX3DV4 — SN3617; ConvF(7.8, 7.8, 7.8); Calibrated: 1/23/2017;
            +    Sensor—Surface: 1Amm (Mechanical Surface Detection)
            *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
            *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=Smm,
        dy=5mm, dz=5mm
        Reference Value =92.28 V/m; Power Drift =—0.01 dB
        Peak SAR (extrapolated) = 27.0 W/kg
        SAR(L g)= 12.7 W/kg; SAR(1O g)= 5.87 W/kg
        Maximum value of SAR (measured) = 21.5 W/kg


           dB
           0

            ~4.38

            —8.75


            13.13

            —17.50
                         L
           —21.80
                      0 dB =21.5 W/kg =13.32 dBW/kg




Certificate No: Z17—97116                            Page 7 of 8


                      TL              :      p -ema                    q
                                      CALIBRATION LABORArory

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       hitps//wwwchinatten


        Impedance Measurement Plot for Body TSL


          Tri siL Log nay 10.00087 ner 0,000d8 [F1]
           s0.00    31 —7.2500005 onz —23, 801. de
           40.00                                     T

           30. 06

           20. 00

           20. 00            ;




          ~40. oo
          ~50. 00
        M si1 snith (RJ>) scale 1.000u [ea oel]
         »1 2.4500000 anz S1.046 0 6.6076 0. 420.2




Certificate No: 217—97116                                Page 8 of 8


                        es as"wil a «l7 M PRWH
                     TTL s
                         a j\\—_,/;’?CNASggfl
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, Chinn                S v CALIBRATION
             Tel: +86—10—62304033—2079       Fux: +86—10—62304637—2504           "m“                                        CNAS LO570
             E—mail ettl@chinatiLoom         hitpetfwenwchinattlen
                        TA(Shanghal)                                        Certificate No:     218—60094
                    en‘                            '“;‘F                                >                                   =———l
                 TJON CERTIFICATE______                                           e                                            e
Object                                     D2600V2 — SN 1025
Calibration Procedure(s)                   FP211—003—01

                                           Calibration Procedures for dipole validation kits
Calibration date:                          May 2, 2018


This calibration Cerfificate documents the traceability to national standards, which realize the physical units of
measurements(S)), The messurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperature(2223)C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #               Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Power Meter    NRVD              102083            O1—Nov—17 (CTTL, No.J17X08756)                       Oct—18
 Power sensor       NRV—Z5        100542            01—Nov—17 (CTTL, No.J17X08766)                                  Oct—18
 Reference Probe EX3DV4           SN 7464            12—Sep—17(SPEAG No.EX3—7464_Sep17)                              Sep—18
 DaE4                             sN1525            02—0ct—17(SPEAG No,DAE4—1525_Oct17)                              Oct—18
 Secondary Standards              1D #              Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Signal Generator E4438C          MY49071430        23—Jan—18 (CTTL, No.J18X00560)                                  Jan—19
 Network Analyzer E5071C          MY46110673.       24—Jan—18 (CTTL, No.J18X00561)                                  Jan—19



                                  Name                         Function                                        Sigqature
Calibrated by:                 ‘Zhao Jing                  SAR Test Engineer                               q&@f4%.
   ¢
Reviewed by:e                    Lin Hao
                                                                                                           1                / ~ubo
                                                           SAR Test Engineer                               Lfi‘fi%} {\:'l \
                                                                                                       7                        1'5‘:‘

Appraved by:                     Ql Dianyuan               SAR Project Leader                                  ;              ,"."e/
                                                                                                   %               /;/,'--;       ,
                                                                                    lssued: May 5, 20418               *6
This calibration certificate shall not be reproduced except in full without written
                                                                                    approvalof the laboratory,



     Certificate No: Z18—60094                             Page 1 of 8


C

              !TTL ;_Leiar_s._
                                           CALIRZATION LABORATORY
             Add: No,51 Xuey uan Road, Haldian District, Heijing, 100191, China
             Tet +86—10—62304633—2070       Fux: +8be10—62304633—2504
             Exmail: ettl@chinatt.com        httpeiriwceliintther

    Glossary:
    TSL                       tissue simulating liquid
    ConvF                      sensitivity in TSL / NORMx.y.a
    NA                        not applicable or not measured


    Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, *IEEE Recommended Practice for Determining the Peak
         Spatia—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices; Measurement Techniques®, June 2013
    b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
       exposure to radio frequency fields from hand—held and body—mounted wireless
       communication devices— Part 1: Device used nextto the ear (Frequency range of 300MHz to
       6GHz)", July 2016
    c) IEC 62209—2, *Procedure to measure the Specific Absorption Rate (SAR) For wireless
       communication devices used in close proximity to the human body (frequency range of
       30MHz to 6GH2z)", March 2010
    d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

    Additional Documentation:
    e) DASY4/5 System Handbook

    Methods Applied and Interpretation of Parameters:
    «_ Measurement Conditions: Further details are available from the Validation Report at the end
       ofthe certificate. All figures stated in the certificate are valid at the freq uency indicated.
    a     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the fiat phantom section, with the arms oriented
          parallel to the body axis.
    «_    Feed Point Impedance and Retum Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
    «_    Electical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required..
    «_    SAR measured: SAR measured at the stated antenna input power.
    «_    SAR normalized: SAR as measured, normalized to an input powerof 1 W at the antenna
          connector.
    «_    SAR fornominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR res ult.


         The reporied uncertainty of measurement is stated as the standard uncertainty of
         Measurement multipied by the coverage factor k=2, which for a normal distribution
         Corrasponds to a coverage probability of approximately 95%.



    Certificate No: Z18—60094                               Page 2 of 8


                            «


               TTL ;_um_a.s_
                                    CALIERATION LABORATORY
       Add: No,$1 Xuey an Road, Haklian District, Beljing. 10019 China
       Tel:+86—10—62304632—2079      Faxt +86—10—62304633—2504
       E—muil ctt}zchinattl.com      hitpsthnin;chinstthen


Measurement Conditions
   DASYsystem configuration, as far as not gven on page 1.
     DASY Vorsion                                      DASY52                                     $2,10.0.1446

     Extrapolation                            Advanced Extrapotation
     Phantom                                   Triple Flat Phantom 5.1C
    Distance Dipole Center —TSL                        10 mm                                    with Spacer

    Zoom Scan Resolution                          dx, dy, ce = 5 mm
    Frequancy                                     2600 MHz + 1 MHz

Head TSL paramaters
  The following parameters and calculations were applied.
                                                        Temporature              Parmittivity          Conductivity

    Nominal Head TSL parameters                           22.0 °C                   39.0                1.98 mhaim
     Measured Hoad TSL para meters                     (22.0 £0.2) °C             40.1 £6%           2.01 mho‘m # 6 %

     Head TSL tomperature change during test              <1.0 °C                     —                       a
SAR result with Head TSL
   SAR averagedover 1_cnt"_(1 g) of Head TSL                         Condition
     SAR measured                                              250 mW input power                   136 mW/g

    SAR for nominal Head TSL parameters                         normallzed to 1W           54.1 mW (g # 18.8 % (k=2)
    SAR avoraged over 10 cm? (10 g) of Head TSL                      Condition
     SAR measured                                              250 mW input power                   6.03 mW /g
     SAR for nominal Head TSL parameters                        normalized to 1W           24,1 mW /g 4 18.7 % (k=2)
Body TSL parameters
  The following parameters and calculations were applied.
                                                        Temperature              Pormittivity          Conductivity

     Nomiral Body TSL paramoters                          22.0°C                    §2.6               2.16 mhaim
     Measurad Body TSL parameters                      (22.020.2) °C              §2116%             2.15 mhoim + 6.%
   Lawd-y TSL temperature change during test              <1.0°C                      —                       n
SAR result with Body TSL
     SAR avoraged over 1 cm> (1 g) of Body TSL                       Condition
     SAR measured                                              250 mW input power                   13.6 mW/g
     SAR for nominal Body TSL parameters                        normalized to 1W           54.5 mW /g +#18.8 % (k=2)

     SAR averaged over 10. Cm* {10 g) of Body TSL                    Condition
     SAR measured                                              250 rrW input power                  6.05 mW /g

     SAR for nominal Body TSL parameters                           normalized to 1W        24.3 mW Ig & 18.7 % (ke2)

Certificate No: Z18—60094                            Page 3 of #


          é_TTL                         ;—E_?:l_fl_—-
                                        CALIBRATION LABORATORY
         Add: No.S1 Xury uan Road, Haiian District, Beljing, 100 191, Chine
         Tel: +86—10—62304633—2079      Fiee: +86—10—62304633—2504
         E—mail: atk@chinattLcom          hitphwwchinattLen

  Appendix(Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL

      Impadance, transformed to feed point                                    48.10—7.55]0
      Retum Loss                                                                —22.008


 Antenna Parameters with Body TSL

     impadance, transformed to feed point                                     46.60—7.0510
     Retum Loss                                                                 —21.908
 General Antenna Parameters and Design

     Electrical Delay (one direction)                                           1,014 ns



 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured,

 The dipole is made ofstandard semingid coaxial cable. The center conductor of the feeding line is direcily
 connected to the second arm ofthe dipole. The antenina is therefore short—cirouited for DC—signals, On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend orthe soldered
 connections nearthe feedpoint may be damaged,



 Additional EUT Data

     Manufactured by                                            |                 sFreac            _




Certificate No: Z18—60094                              Page 4 of n


C
               BP3                         inColsborsionwith
              7T‘J,
              <,    s
                    a e a
             Add: No.$1 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tek +86—10—62304633—2079        Fax: +86—10—02304633—2504
             E—mail: ctt¥@ichinattl.coms     hitp oAwww chinattl en


    DASYS Validation Report for Head TSL                                          Date: 05.02.2018
    Test Laboratory: CTTL, Beijing, China
    DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN; 1025
        Communication System: UID 0, CW; Frequency: 2600 MHz; Duty Cyele: 1 :1
        Medium parameters used: £= 2600 MHz; 0 = 2.014 S/m; er = 40.09; p = 1000 kg/m3
        Phantom section: F it Section
        Measurement Standard: DAS Y5 (IEEE/IEC/ANSI C63.19—2007)
    DASY5 Configuration:

                 +     Probe: EX3DV4 — SN7464; ConvF(7.76, 7.76, 7.76); Calibrated: 9/12/2017;
                 *     Sensor—Surface: 1 4mm (Mechanical Surface Detection)
                 *     Electronies: DAE4 Sn1525; Calibrated: 10/2/2017
                 *     Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
                 *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                      (7417)

            Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0; Measurement grid: dx=3 mm,
            dy=Smm, dz=5mm
            Reference Value = 98.50 V/m; Power Drift= 0.00 dB
            Peak SAR (extrapolated) = 29.6 W/kg
            SAR(I g)=13.6 Wikg; SAR(10 g)=6.03 W/kg
            Maximum value of SAR (measured)= 23.5 W/kg




                ~4.80

                —1.59

                ~14.39

                ~19.18

                ~23.98
                                L.
                             0 dB =23.5 Wikg =13.71 dBWkg




    Certificate No: Z18—60094                              Page 5 of &


                                     In Collaberation with
                 TTL a               CAUIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191,China
        Tel: +86—10—62304633—2079      Fix: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      hitprtivinchinatLen




        Impedance Measurement Plot for Head TSL


      Tri siT Log nag 10.00d87 ref 0.00000 [F]
        $0.00
                 H 2w          ene <22.00 0s
       10.60
       30.00
       20.00
       10.00
       a.coo y
      —10. 00
                                                                           i
      ~20. 00
      ~30. 00
      40. 00
      ~40.c0
    e sin satch (rofx) scale 1.0000 [52 oel]
     s1 2.6000000 onz 48.007 o —7.8%0r 8. .1




Certificate No: Z18—60094                            Page 6 of 8


C
              !\                          In Collaboration with
             @7"7"J,
             un                           sa
            Add: No.51 Xuey uan Road, Haidian District, Beljing. 100191, China
            Tel +86—10—62304633—2070       Fax: +86—10—62304633—2504
            E—mail: cttl@chinatiL.com       bitpriwwwchinatthen

    DASYS Validation Report for Body TSL                                         Date: 05.02.2018
    Test Laboratory: CTTL, Beijing. China
    DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600 V2 — SN: 1025
       Communication System: UID 0, CW; Frequency: 2600 MHz; Duty Cycle: 1 :1
       Medium parameters used: {= 2600 MHz; 0 = 2.146 S/m; & = 52.09; p = 1000 kg/m*
       Phantom section: Right Section
       Mcasurement Standard: DASY3 (IEEE/IEC/ANSI €63.19—2007)
    DASYS Configuration:

                «_   Probe: EX3DV4 — SN7464; ConvF(7.$4, 7.84,7.84); Calibrated : 9/12/2017;
                *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
                *    Electronics: DAE4 $n1525; Calibrated: 10/2/2017
                *    Phantom: Tripk Flat Phantom 5.1C; Type: QD 000 PS1 CA; Serial: 1161/1
                *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                     (7417)

            Dipole Calibration/Zoom Scan (7x7x7) (7x7x7 )Cube 0: Measurement grid : dx=5 mm,
            dy=5mm, dz=5mm
            Reference Value = 83.79 V/im; Power Drift= 0.02 dB
            Peak SAR (exirapolated)=29.7 W/kg
            SAR(L g)= 13.6 Wikg; SAR(IO g)=6.06 W/kg
            Maximum value of SAR (measured)= 23.6 W/kg


               dB
                o


               ~4.69

               —9.38

               14.07

               18.76
                              .
               —23.45
                           0 dB =23.6 Wikg =13.73 dB Wikg




    Certificate No: £18—60094                              Page 7 of 8


                 TTi, ;_e_éi_.q_
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattLcom      hitpoiwww.chinattLen


        Impedance Measurement Plot for Body TSL


      Tek si% Lo nay 10.00007 nef 0.90000 [Fi]
        **+ 09. 223000000
                5t   ne —#1. ad
        £0.00

       30.c0
       20.c0
       10.c0
       o.oco9)                                                              i
       —10.00
       —20.00
       ~s0.00
       ~s0.00
       ~$0,00                                                       a
     IMIRM 511 smith (iefx) Scale 1.000u {e1 bel)
       3%. 2,6000000 ohz d6.642 6 —7.056% n 8.67




Certificate No: Z18—60094                             Page 8 of 8


                                                                                                                  EE‘A
                 zs                      CALIBRATION LABoRArory                  Mfi CNA
                                                                                 ‘            &
                                                                                                                  Rag
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   ’4,,/—\\\\‘ v CALIBRATION
             Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504               /,,m\.\\    CNAS LO570
             E—mail: cttl@chinattl.com       http://www.chinattl.en
      Client               TA(Shangha|)                                     Certificate No:       217—97001




Object                                     DSGHzV2 — SN: 1151

Calibration Procedure(s)                   FD—Z11—003—01
                                           Calibration Procedures for dipole validation kits

Calibration date:                          January 5, 2017


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #                Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Power Meter        NRP2          101919             27—Jun—16 (CTTL, No.J16X04777)                           Jun=17
 Power sensor       NRP—Z291      101547             27—Jun—16 (CTTL, No.J16X04777)                           Jun—17
 ReferenceProbe EX3DV4 SN 7433                       26—Sep—16(SPEAG,No.EX3—7433_Sep16)                       Sep—17
 DAE4                             SN 771             02—Feb—16(CTTL—SPEAG,N0.Z16—97011)                       Feb—17

 Secondary Standards              ID #               Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Signal Generator E4438C          MY49071430          01—Feb—16 (CTTL, No.J16X00893)                        Jan—17
 NetworkAnalyzer E5071C           MY46110673          26—Jan—16 (CTTL, No.J16X00894)                        Jan—17



                                   Name                         Function                                   Signature
Calibrated by:                   Zhao Jing                  SAR Test Engineer

Reviewed by:                     Qi Dianyuan                SAR Project Leader

Approved by:                     Lu Bingsong                 Deputy Director of the laboratory         1


                                                                                     Issued: January 7{42017
This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.



     Certificate No: Z17—97001                              Page 1 of 14


           *®                       In Collaboration with

           m_‘7/"]"J,
           -““/
                      a
                      CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      hitp://www.chinattl.en


Glossary:
TSL                      tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
   devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
    30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
«_ Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
   point exactly below the center marking of the flat phantom section, with the arms oriented
   parallel to the body axis.
* Feed Point Impedance and Return Loss: These parameters are measured with the dipole
   positioned under the liquid filled phantom. The impedance stated is transformed from the
   measurement at the SMA connector to the feed point. The Return Loss ensures low
   reflected power. No uncertainty required.
* Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
   No uncertainty required.
   SAR measured: SAR measured at the stated antenna input power.
   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
*    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.



Certificate No: Z17—97001                             Page 2 of 14



Document Created: 2019-08-16 16:51:09
Document Modified: 2019-08-16 16:51:09

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC