SAR_Test_Report_ANNEX E3

FCC ID: 2AFZZC3IG

RF Exposure Info

Download: PDF
FCCID_4440699

             ETT[:: p Aema q
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
            E—mail: ttl@chinatt.com        Iitpo/wwschinattlen

      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3677

Calibration Parameter Determined in Body Tissue Simulating Media
                                                                                                              6
 f [MHz]®      Pe::l:t’tulv::y s   Cond(;;:'l’:\)ll:y   ConvF X       ConvF Y    ConvF Z   Alpha®   D(:::;        :'::::)'

    750             55.5                 0.96             9.75            9.75    9.75      0.40     0.75         £12.1%
    835             65.2                 0.97             9.40            9.40    9.40      0.18     1.38         £12.1%
   1750             53.4                 1.49             7.86            7.86    7.86      0.23     1.09         £121%
   1900             53.3                 1.52             7.62            7.62    7.62      0.22     115          £121%
   2300             52.9                 1.81             7.67            7.67    7.67      0.55     0.81_|       £12.1%
   2450             52.7                 1.95             7.57            7.57    7.57      0.59     0.75         £12.1%
   2600             52.5                 2.16             7.33            738     7.33      0.74     0.65         £12.1%
   5250             48.9                 5.36             4.93            4.93    4.93      0.45     155          £13.3%
   5600             48.5                 §.77             4.24            4.24    4.24      0.50     145|         £13.3%
   5750             48.3                 5.94             4.35            4.35    4.35      0.50     1.50         £13.3%



© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), elseit is restricted to
50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
F At frequency below 3 GHz, thevalidity of tissue parameters (e and 0) can be relaxed to £10% if iquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (c and 0) is
restricted to £5%. The uncertainty is the RSS ofthe ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below : 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z19—60169                               Page 6 of 11


       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com      Http://www.chinattl.en



                                                   Frequency Response of E—Field
                                                   (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normallzed)




                                         0.5        i—f       +——f       4+———|       4    {      i     f     +     T
                                               0      500         1000        1500        2000         2500       3000
                                                            Eeq              f[MHz]
                                                            TEM                                  R22


                                               Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z19—60169                                            Page 7 of 11


                                             In Colleborationwith
                    =7/‘7‘/,
                    m                               p                      q
                    y                        csnummoueunonanow
                  Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                  Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
                  E—mail: cttl@chinattl.com     Hitpy/wwwchinattl.en



                                         Receiving Pattern (®), 0=0°

                        f=600 MHz, TEM                                                f=1800 MHz, R22
Error{dB]




            1.0                                                               T
                          —150            —100              50                0          so       100       150
                                                                          Rollf]
                                    [———]100MHz__—+—600MHz                        —+—1800MHz__—+—2500MHz]
                                       Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




    Certificate No: Z19—60169                                    Page 8 of 11


           _q\n_: in Cotaboretonwith

       Add: No51 Xoeyuan Roat, Taidan Ditric, Bejing, 100191; China
       T «te—1Oendorasta            Ts +16—10—62300633—
       E—mail:etlchinaticom            M/ownchintlon
                                   Dynamic Range f(SARnead)
                                          (TEM cell, £= 900 MHz)
          Input Signal[aV)




                                                            r              t                   h
                             10°        10                 t0°             10            10   10
                                                           SARImW/cm‘]
                                              EJnt conpensated         ~#~ compensated
       ErrortaB]




                                                     10                                       10
                                                                ARImWiem
                                    I—a—inot compensated                       —e—c
                                   Uncertainty of Linearity Assessment: £0.9% (k=2)
Cenifieate No: Z19—60169                           Pages of 11


       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com      Hitpy//www.chinattl.en



                         Conversion Factor Assessment
      f=750 MHz, WGLS R9(H_convF)                                              f=1750 MHz, WGLS R22(H_convF)
                                                                                               30.00


                                                                                               25.00




                                                                                          8
                                                                                                8
                                                                                   SARIWAg/W
                                                                                          a
                                                                                                8
                                                                                               1000          A
                                                                                                5.00

                                                                                                0.00
                 00             m           40           «o   sn       m                               00    q0    ze    se       w0   so   eo   To
                                                 zimm}                                                                        z{mm]




                       Deviation from Isotropy in Liquid
                                            sccne ie                       E                                             m
                                    1.0
                                                                           |
                                    0.8

                                    0.6

                                    0.4
                                     0.2
                       7 Axis




                                     0.0
                                    —0.2
                                     —0.4
                                     —0.6




                                             —1.0    —080 —060 —040 —020       0         020           040   060   080   10
                                    Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z19—60169                                      Page 10 of 11






                             *     in Colaborationwith

                    a
                TTL CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District,Beiing, 100191, China
        Tels +86—10—42304633—2070      Pac +86—10—2304633.2504
        E—mail: ctl@chinatl.com        hitpoIhwwewcchinatl.en

Glossary:
TSL                      tissue simulating liquid
ConvF                     sensitivity in TSL / NORMxy,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Sid 1528—2013, "IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
   devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)*, February
   2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
     communication devices used in close proximity to the human body (frequency range of
     30MHz to 6GHz)®, March 2010
d) KDBB65664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
*     Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
*     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
*     Feed Point Impedance and Retun Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
*     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
*     SAR measured: SAR measured at the stated antennainput power.
&     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
«_    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


Certificate No: Z17—97002                             Page 2 of 8


                TT-l:              in Callaberation with




        Add:No.S1 Xueyuan Road. Haidian District Beiing, 100191, China
        Tels +86—10—402304033—2070   nc +86—10—6230463—2504
        E—mail ctl@chinathcom          hitps wwchinatl.cn

Measurement Conditions
   DASY system confiquration, as faras not given on page 1.
    DASY Version                                    pasyse                                       s2.8.8.1258
    Extrapolation                           Advanced Extrapolation
    Phantom                                 Triple Fiat Phantom 5.1C
    Distance Dipole Center — TSL                    10 mm                                      with Spacer
    Zoom Scan Resolution                       de, dy, dz =5 mm
    Frequency                                  1750 Miza 1 Niz

Head TSL parameters
   The following parameters and calculations were applied.
                                                           Temporature           Permittvity           CGonductivity

     Nominal Head TSL parameters                             22.0°C                40.1                1.37 mhoim
     Measured Head TSL parameters                          r20202)°C              30416 %            1.35 mhoim £ 6 %
     Head TSL temperature change during test                 «10°c                  —                       ——
SAR result with Head TSL
   SAR averaged over 1_c1"_(1 g) of Hoad TSL                         Condition
     SAR measured                                                250 mW input power                9.27 mW /g
     SAR for nominal Head TSL parameters                           normalized to 1W       37.2 mW 1g 20.8 % (ke2)
     SAR averaged over 10 Cm". (10 g) of Head TSL                    Condition
     SAR measured                                                250 mW input    power          4.90 mW/g
     SAR for nominal Head TSL parameters                           nomalized to 1W        19.7 mW i9 # 204 % (ke2)

Body TSL parameters
  The following parameters and calculations were applied.
                                                            Temperature          Pormittvity           Conductvity
     Nominal Body TSL parameters                              220°C                 sea                1.49 mho/m
     Measured Body TSL parameters                          @a0202)°C              58126 %            1.48 mhoim £6 %
     Body TSL temperature change during test                  «10°C                  —                     m
SAR result with Body TSL
     SAR averaged over 1 _cm"_(1 g) of Body TSL                      Condition
     SAR measured                                                250 mW input power                9.40 mW /g
     SAR for nominal Body TL parameters                            nomalized to 1W        37.6 mW 1g £ 20.8 % (k=2)
     SAR averaged over 10 cm> (10 ) of Body TSL                      Condition
     SAR measured                                                250 mW input power            5.03 mW /g
     SAR for nominal Body TSparameters                            normalized to 1W        20. mW 1g # 20.4 % (k=2)
Certificate No: Z17—97002                            Page s of $


       Add:No.51 Xueyuan Road, Haidian District, Beiing, 100191, China
       Tel: +86—10—62304633—2070     Faxc+86—10—62304633—2504
       E—mail: ctl@chinatl.com       htpowwwchinatt.cn

 Appendix

Antenna Parameters with Head TSL

     Impedance, transformed to feed point                                49.00+ 0.930
     Retum Loss                                                           —40.308


 Antenna Parameters with Body TSL

     impedance, transformed to feed point                                44.70—0.100.
     Retur Loss                                                           —25.008


 General Antenna Parameters and Design

     Electrcal Delay (one direction)                                       1.327 ns


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is direcly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in orderto improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is stil according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                          SPEAG




Certificate No: Z17—97002                            Page 4 of %


          2l5                      in Collboration with

           ‘TTL’                  CauBRAnION LABORATORY

        Add:No.51 Xueyuan Rood. Haidian Distict, Beiing, 100191, China
        Tels +86—1042304033—2070     Paxc +86—10.62304603.2504
        E—mail: ct@chinatt.com        hupswwwchinatlcn

DASYS Validation Report for Head TS                                          Date: 01.10.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1033
  Communication System: UID 0, CW; Frequency: 1750 MHz; Duty Cycle: 1:1
  Medium parameters used: {= 1750 MHz; & = 1.352 $/m; er = 39.36; p = 1000 kg/m3
  Phantom section: Center Section
  Measurement Standard: DASYS (IEEEAEC/ANSI C63.19—2007)
DASY5 Configuration:

            *    Probe: EX3DV4 — SN7307; ConvF(8.37, $.37, 8.37); Calibrated: 2/19/2016;
            +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            *    Electronics: DAE4 Sn771; Calibrated: 222016
            +    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            +    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Sean (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=Smm, dz=5mm
        Reference Value = 98.21V/m; PowerDrift = 0.03 dB
        Peak SAR (extrapolated) = 17.1 Wke
        SAR(L g) =9.27 W/kg; SAR(IO g) = 4.9 Wig
        Maximum value of SAR (measured) = 14.4 W/kg

           dB
            0


            3139


            5.78




                                                                         —
            10.17


            +13.56


            +16.95

                      0 dB = 14.4 Wikg=11.58 dBW/hg




Certificate No: Z17—97002                            Page 5 of 8


                              ®     In Collaboration with


                                    CauRAnN LABORATORY

        Add: No.31 Xueyuan Road. Haidian District, Beiing, 100191, Ctina
        Tel: +86—10—42304633—2070       Paxc +86—10—62304633—2504
        E—mail: ctl@chinatt.com          hitpohwwwchinatten




        Impedance Measurement Plot for Head TSL


     Ti TT toy say 10. 00007 wet o. oode Ti
       5900.    rex—screoeons ons    cepses o
       so.00
       s0.00
       20.00
       s0.00
       o.0cop
      x0.00
      «20.00
      —30.00
      —40.c0
      s0.00
     BB 5i1 setth (i+p9 seate 2.0000 (et o1
      on alrsoo0 o. anurss o saras m oi3




Certificate No: Z17—97002                             Page 6 of 8


                             *    in Collaborationwith
                TTLf‘_uLLfl_S_
                                       leramion tasorarory
        Add:No.S1 Xueyuan Road. Haidian Distict,Beiing, 100191, China
        Tels +86—10—62304633—2070     Faxc +86—10—62304633—2504
        E—mail: etl@chinatl.com       hitpshwwwchinaten


DASY3 Validation Report for Body TSL                                   Date: 01.10.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1033
   Communication System: UID 0, CW; Frequency: 1750 MHz; Duty Cycle: 1:1
   Medium parameters used: {= 1750 MHz; & = 1.484 S/m; s= $3.05; p = 1000 ke/m‘
   Phantom section: Left Section
   Measurement Standard: DASY5 (IEE               TEC/ANSI C63.19—2007)
DASY5 Configuration:
            *    Probe: EX3DV4 — SN7307; ConvF(8.18, 8.18, 8.18); Calibrated: 2719/2016;
            +   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +   Electronics: DAE4 Sn771; Calibrated: 2/22016
            +   Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            *   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Sean (7x7x7) (7x7x7)/Cube 0: Measurementgrid:
        dx=5mm, dy=Smm, dz=5mm
        Reference Value = 86.52 V/m; Power Drift = 0.04 dB
        Peak SAR (extrapolated) = 16.6 W/kg
        SAR(L g) = 9.4 Wikz: SAR(IO g) =
        Maximum value of SAR (measured


           dB
            0


            3.26


            5.52


            —8.78


            +13.04


            16.30

                      0 dB = 14.1 W/kg= 11.49 dBW/kg



Certificate No: Z17—97002                           Page 7 of8




                                                                                                           shBRAS
             A
                ZArarse > "\um         Causranon LaporArory                3             § cNAs
                                                                                              cmaRATIDOH
              ueneil nemrtetepee
             iE—mail:ct@chinatl.com         pavemonlon 5oys w on
                  No.51 Xueywan Road, Haition Disvct Bing. 100191, China
                                    hitpifwwwchinatilen
                                                                           %           §                         s LoS7(

     Client                                           Cal              [GertificateNo:


Object

Calibration Procedure(s)

Callbration date:

This callbration Certiicate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory faciity: environment temperature@2:3)‘c. and
humidity<70%.

Calibration Equipment used (M&TE critical for callbration)

Primery Standards                  D#                 Cal Date(Calibrated by, Certficate No.)   Scheduled Calibration
  Power Meter   NRVD                  102083         22—8ep—16 (CTTL, No.J16X06809)                   Sep—17
  Power sensor NRV—Z5                 100595         22—Sep—16 (CTTL, No.J16X06809)                   Sep—17
  Reference Probe EX3DV4              SN 3617        23—Jan—17(SPEAG No.EX3—3617_Jan17)               Jan—18
  DaE4                                SN 1331        19—lan—17(CTTL—SPEAGNo.217—07015)                Jan—18

  Secondary Standards                 D#             Cal Date(Calibrated by, Certiicate No.)    Scheduled Calibration
  Signal Generator E4438C             MY49071430     13—Jan—17 (CTTL, NoJ17X00286)                    Jan—18
  Network Analyzer ESO71G             MY46110673     18—Jan—17 (CTTL, No.J17X00285)                   Jan—18



                                      Name                     Function _                            Signature
 Calibrated by:                         s

 Reviewed by:

 Approved by:                     6QlDlnnyuan             «SAR Project Leider


  This calibration certiicate shall not be reproduced except in full without written approval of meIaboralcry


         Certificate No: Z17—971 15                          Page 1 of 8


         u*     stion vith         in Collsbor
         s T°TL a
         "                         CALIBER&ANON LABORATORY
       Add: No.51 Xueyuan Re        idh      lad
       Tel: +86—10—
             s
                    6230403
                      7     3.3079°onFax:Dstlct,
                                           +86— Beijing, 100191,
                                                            s
                                                                   R
                                                                 China
       E—mail: cttl@chinattl.com          hIn;://m ]gsfibg‘? 2504

Glossary:
‘(I;SL                   tissug simulating liquid
N?AWF                    sensitivity in TSL / NORMx.y,z
                         not applicable or not measured


Calibration is Performed According to the Following Standa
                                                           rds:
a) |EEI§ Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
     6GHz)", July 2016
 ©) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
    communication devices used in close proximity to the human body (frequency range of
     30MHz to 6GHz)", March 2010
 d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

 Additional Documentation:
 e) DASY4/5 System Handbook


 Methods Applied and Interpretation of Parameters:
 * Measurement Conditions: Further details are available from the Validation Report at the end
    of the certificate. All figures stated in the certificate are valid at the frequency indicated.
 e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
    point exactly below the center marking of the flat phantom section, with the arms oriented
    parallel to the body axis.
 e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
       positioned under the liquid filled phantom. The impedance stated is transformed from the
       measurement at the SMA connector to the feed point. The Return Loss ensures low
       reflected power. No uncertainty required.
 e     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
       No uncertainty required.
 e     SAR measured: SAR measured at the stated antenna input power.
 e     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
 e     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
       nominal SAR result.


      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement multiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probability of approximately 95%.



  Certificate No: Z17—97115                             Page 2 of 8


                                    ®   In Collaboration with
           iTTL s                                     e
         Add: No.5
                     1 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: *‘.86-10-623046
                              33-2079
           »mail: ett!@chinattl.com       Fax: +86—10—62304633—2504
                                            http:!/www.chinatt.en

 MeDasurement Conditions
      ASY system configuration, as far
                                       as not given on page 1.
       DASY Version                                             DaASY52                                  52.10.0.1446
       Extrapolation                                Advanced Extrapolation
       Phantom                                       Triple Flat Phantom 5.1C
       Distance Dipole Center — TSL                             10 mm                                  with Spacer
      Zoom Scan Resolution                              dx, dy, dz = 5 mm

       Frequency                                        1900 MHz + 1 MHz

 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature             Permittivity           Conductivity

       Nominal Head TSL parameters                                22.0 °C                  40.0                1.40 mho/m
       Measured Head TSL parameters                          (22.0 £ 0.2) °C             39.9 £ 6 %          141 mho/m £ 6 %

       Head TSL temperature change during test                    <1.0 °C                   ——                       x
 SAR result with Head TSL
      SAR averaged over 1_Cnm:°_(1 g) of Head TSL                           Condition
       SAR measured                                                     250 mW input power                 10.1 mW /g

      SAR for nominal Head TSL parameters                                normalized to 1W         40.1 mW /g £ 18.8 % (k=2)

      SAR averaged over 10 em‘ (10 g) of Head TSL                           Condition
      SAR measured                                                      250 mW input power                 5.19 mW /g

      SAR for nominal Head TSL parameters                               normalized to 1W          20.7 mW ig £ 18.7 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature             Permittivity           Conductivity

      Nominal Body TSL parameters                                22.0 °C                   53.3                1.52 mho/m
      Measured Body TSL parameters                          (22.0 £ 0.2) °C              53.6 16 %           1.53 mho/m £ 6 %
      Body TSL temperature change during test                    <1.0 °C                    —                        ——
SAR result with Body TSL
   SAR averaged over 1_cm> _(1 g) of Body TSL                               Condition
     SAR measured                                                    250 mW input power                    9.90 mW /g
     SAR for nominal Body TSL parameters                                normalized to 1W          39.5 mW /g + 18.8 % (k=2)
     SAR averaged over 10 cm‘ (10 g) of Body TSL                            Condition
     SAR measured                                                    250 mW input power                    5.21 mW /g

     SAR for nominal Body TSL parameters                                normalized to 1W          20.8 mW /g & 18.7 % (k=2)

Certificate No: Z17—97115                                Page 3 of 8


                              5     inCotaboredonwith

                                    causeanon LasorATORY
         Ada:; No.51 Xueyuan Rond. Haidian Disvict Be/ 3. 100191, Chinn
         Tel: e06—10—00004
               :           ©32000       Pacone—
         E—mail:ctl@chinatcom           nfi/‘mfifmfif as
                                                   S
  Appendix {Additional assessments outside the scope of CNAS L0570)

  Antenna Parameters with Head
                               TSL

      Impedance, transtormed to feed point                                s2.00+ 6.590
      Retum Loss                                                            —23408

  Antenna Parameters with Body TSL

      Impedance, ransformed to feed point                                 s2.70+ 8.350
      Retum Loss                                                           —21408

  General Antenna Parameters and Design

      Electrcal Delay (one direction)                                       1.302 ns


  After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
  be measured.

   The dipole is made of standard semirigid coaxtal cable. The center conductor of the feeding lineis directly
  connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
  of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
  according to the position as explained in the "Measurement Condiions" paragraph. The SAR data are not
  affected by this change. The overall dipole length is stil according to the Standard.
  No excessive force must be applied to the dipole arms, becausethay might bend or the soldered
  connections near the feedpoint may be damaged.



  Additional EUT Data

      Manufactured by                                        l                 SPEAG




Certificate No: Z17—97115                            Page 4 of8


           pemmmeS in Colibomionwith
           w
           @TTL a              e _a
         Add: No1
         Toh rse—10x1.eey
                       03
                         uan Road, Haidian Distict Beling., 100191, China
         Email FEX@chin 0003.0070        Fax: +86—10.e2304633—2504
                         aL.com          biphewchinatlen
 DASYS Validation Report    for Head TSL
Test Laboratory: CTT
                     L, Beijing, China                                          Date: 08262017
 DUT: Dipole 1900 MHz; Type D1900V2;
                                     Serial:                    D19
                                                    00V2 — SN: 54060
    Communication System: UID 0, CW; Fre
                                         quency: 1900 MHz; Duty Cycle: 1:1
    Medium parameter       s used: £= 1990 MHz; 0 = 1.413 S/m; er = 39.8
    Phantom section: Left Sectio                                         5; p= 1000 kg/m3
                                n
    Measurement Standard: DASYS (IEE
                                     ETEC/ANSI ©63.19—2007)
DASY5 Configuration:

                  Probe: EX3DV4 — SN3617; ConvF(8.26, 8.26, 8.26); Calibra
                                                                           ted: 1/23/2017;
                  Sensor—Surface: 1.4mm (Mechanical Surface Detect
                                                                   ion)
             *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
                  Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA;
                                                                          Scrial: 1161/1
                  Mcasurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                  (a17)
         System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
         dx=5mm, dy=Smm, dz=Smm
         Reference Value = 94.94 V/im; Power Drit=0.01 dB
         Peak SAR (extrapolated) = 19.5 Wikg
         SAR(I g) = 10.1 W/kg: SAR(IO g)=
         Maximum value of SAR (measured) =




             3.64

             128

            +10.93

            1457


            "18.21
                       J

                      0 dB=15.9 W/kg=12.01 dBW/kg



Certificate No: Z17—971 15                            Page S of $


                             °in Callborrtiomnith

                                 CAuszknont Lagorarory
       Add:No.31 Xueyuan Road, Haidian Disti, Bejing, 100191 China
       Tel:86—10.62300600—2079      pacoge104300032501
       E—mail: enl@chinat.com       Witpiiwcctinattlen



        Impedance Measurement Plot for Head TSL


          Ti s1 toy wag 10.c0007 mer o.soice NJ
           59090 mromomons en cmmar c
           «0.00
           10.c0      —
           20.00 |.—           ——         maj z_~doz _
           10.00
            s.coo9
           —10.00
           20.00
           —30.00
           —40.00
            o.co
          TB sis sntch (rep) seate 1.0000 trz o)
           n acsooono on se ecsmee c 5s




Certificate No: Z17—971 15                        Page 6 of8


       Add: NoS. 1 Xueyywan Road, Haidiai n
       Tiasinesnesan ; Di: ic +ne—10.62304633—2504   Beljng. 100191, China
       E—mail: il@chinatlcom                hipslhwwchinattien


DASYS Validation Report for Body TSL                                         Date: 08.26.2017
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz: Type: D1900V2; Serial: D1900V2 — SN: 54060
  Communication System: UID 0, CW; Frequency: 1900 MHz; DutyCycle: 1:1
   Medium parameters used: f= 1900 MHz; 0 = 1.528 S/m; c, = 53.55; p = 1000 kg/m‘
   Phantom section: Center Section
   Measurement Standard: DASY5 (IEEEEC/ANSI C63.19—2007)
DASYS Configuration:

            *    Probe: EX3DV4 — SN3617; ConvF(7.95, 7.95, 7.95); Calibrated: 1/23/2017;
            *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
            *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 PS1 CA; Serial: 1161/1
            *    Measurement SW: DASY$2, Version 52.10 (0); SEMCAD X Version 14.6.10
                 (7417)

       System Performance Check/Zoom Sean (7x7x7) (7x7x7)/Cube 0: Measurement grid:
       dx=Smm, dy=3mm, dz=Smm
       Reference Value = 91.19 V/im; Power Drit=0.01 dB
       Peak SAR (extrapolated)= 18.1 Wikg
       SAR(L g) = 9.9 W/kg; SAR(1O g)= 5.21 Wkg
       Maximum value oSAR (measured) = 15.3 Wikg

           d8
           0

            348

            5.96

            40.44

            413.92


            17.40

                       0 dB =15.3 W/kg = 11.85 dBW/kg




Certificate No: Z17—97115                                Page 7 of 8


                            *      in Collaborrtionwith
               TTIL a
                    causrAnon LABORATORY
       Add: No.S1 Xueyuan Road, Haidian Di wict Bciin, 100191, China
       Tel: aso—10—62300533—2075            ne—io.aasou332504
       E—mail: tl@chinatl.com          F
                                       hipiwnwchinatien
       Impedance Measurement Plot for Body TSL


                 Tey rap ze
           so.00 m Eromns on —moor ce
           «0.00
           30.¢0
           20.00
           10.00
           .000                    T                              m      >
         —20.¢0                                               —        //'
         10.¢0                                   —                     i
         —30.00 |—               f=—                          21        bevernacelea o mccl
         —10.00                  {
       L —s0.00                                                    x
        TR six sith (riga) seale 3.0000 [Fi oet)
         o csoome oo snsto n mm o eonge




Centificate No: 7177115                              Page 8 of8


Calibration Laboratory of                                                                           Sctwelzerischer Kallorerdinst
Schmid & Partner                                                                                    Service sutsso o‘étalonnage
  Engineering AG                                                                                    Servitlo svizzero d taratura
Zeughausstrasse 43, 6004 Zurich, Switzeriand                                                        Swiss Calbration Service

Accredted y tna Swiss Accredtaton Senvce (SAS)
The Swiss Accreditation Service is one of the signatories to the EA
Mutitateral Agreementfor the recognition ofcalibration cerificates
Cent       TASH (Auden)                                                             Gevificsis No: DAE4—1201_Dec18
|CALIBRATION CERTIFICATE
 Otject                           DAE4 — SD 000 D04 BM — SN: 1291

 Cattraton procedures)            OA CAL—08.v29
                                  Calibration procedure for the data acquisition electronics (DAE)



                                  December 04, 2018




 This caltraton certfcate dacumentsthe traceabity to natlonal standards, which realz the physicaluntso measurements (6).
 "The measurements and the unceraintes wih contdnce probabity are given on th foloning pages and are parof tconticate.

 Allcaltvations have been conducte in the closed laboratoy facity: envronmont temporature(22 + 3)°C and hurty < 70%.
 Caitvaton Equpment used (M&TE ertcafor caltraton)
 Primary Standards                 o«                     Gal ate (Contfcato No)                        Scheduted Galbation
 Kethiey Matimeter Type 2001       |SN:o810ere            tS0p18 (Nozsien)                              Sepro
 Secontary Standarde               B                  Check Date (n house)                              Schocuted Chack
 Auto DAE Caltration Unt           SE Lws 053 AA 1001 Ofln—18 (in house checi)                          in house cheeks Jan—10
 Caltrator Box ¥2.1                S uis 006 AA 1002 Otlan—18 (in house check)                          In house chocks Jan—t0




                                                                      Function                           Siatire
 Caltratodby:                      Dominique Stotten                  Laboratoy Technician
                                                                                                      Z2
                                  (SvenKitn                           Deputy Manager

                                                                                                        tesued: December 4, 2018
 "This attation corticate shallnot be reproduced exceptful wihout witen approval f the aboratory.


Certficate No: DAE4—1291_Dect8                                Page 1 of 5


Calibration
      E     Laboratory of                               «um
                                                        «/x                              cher Katbriertionst
Schmid & Partner                                        inss‘             2              se étalonnage
  Engineering AG                                       T      3              Servilo sizzero ol taatura
Zeughausstasse 43, 6004 Zurich, Swterand               i@'fil\w            5| duiss Calbration Service

Aceredted ty me Swss Accrtaton Servce (SAS)                                Aecreditationo:: SCS 0108
‘The Suiss Acerediation Service is one of thesignatories to he EA
Motslateral Agreement for the recogntion o caaton cericates

Glossary
DAE                       data acquisition electronics
Connector angle           information used in DASY system to align probe sensor X to the robot
                          coordinate system.

Methods Applied and Interpretation of Parameters
   * DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
     comparison with a calibrated instrument traceable to national standards. The figure given
        corresponds to the full scale range of the voltmeter in the respective range.
    *   Connector angle: The angle of the connector is assessed measuring the angle
        mechanically by a toolinserted. Uncertainty is not required.
    *   The following parameters as documented in the Appendix contain technical information as a
        result from the performance test and require no uncertainty.
        *     DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
              the nominal calibration voltage. Influence of offset voltage is included in this
              measurement.
        *     Common mode sensitivity: Influence of a positive or negative common mode voltage on
              the differential measurement.
        *     Channel separation: Influence of a voltageon the neighbor channels not subject to an
              input voltage.
        *     AD Converter Values with inputs shorted: Values on the intenal AD converter
              corresponding to zero input voltage
        *     Input Offset Measurement Output voltage and statistical results over a large number of
              zero voltage measurements.
        *     Input Offset Current: Typical value for information; Maximum channel input offset
              current, not considering the input resistance.
        *     Input resistance: Typical value for information: DAE input resistance at the connector,
              during internal auto—zeroing and during measurement.
         *    Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
              alarm signal is generated.
         *    Power consumption: Typical value for information. Supply currents in various operating
              modes.




Corificate No: DAEA—1291_Dects                              Page 20(5


DC Voltage Measurement
   ATD — Converter Resolution nominal
       High Range:          1so«        otay.       fullrange = —100...+300 mV
       Low Range:          wso«         stav.       tullrango =   <     aany
   DASY measurement parameters: Auto Zero Time: 3 sec; Maasuring time: sec
     Calibration Factors            x                     v                      z
     High Range          402 .580 + 0.02%s(ke2) 403.240 + 0.02%s (ke2) 409.163 2 0.02% (k=2)
     Low Range           sorseo + 1.50% (ke2) 897000 + 1.50% (ke2) 397550 2 1.50% (k=2)

Connector Angle

     Connector Angle to be used in DASY system                                   1easte1®




Ceniicate No: DAE#—1291_Dect8                     Page a of$


Appendix (Additionalassessments outside the scope of SCS0108)
1. DC Voltage Linearity
    High Range                                Reading (xV)            Difference (1¥)              Ervor (%)
     Channel x      + Input                    2ocosest                      105                    000
     Channel X      + Input                     2000661                      129                    001
     Channel X      + Input                    —20003.34                    204                     001
     Channel Y      + Input                    20003027                     005                     0.00
     Ghannel ¥      + input                     200038                      "184                    001
     ChannelY       _ _— Input                 —20006.11                    oze                     000
     Channel 2      + input                    20003500                                             000
     Channel 2      + input                    200022                                               a01
     Channel 2      _ _— input                 2000540                                              oco
     Low Range                                Reading (w¥)            __|_Ditterence (iV)          Ervor (%)
     Channel x      + input                     200129                      031                     oce
     Channel X      + input                      20113                      ose                     o1
     Channel X      _ _— input                  8850                        oso                     o15
     Ghannel ¥       + input                    2000.40                     4s                      oce
     Channel ¥      + input                      2021                       66                      s3
     Channel Y      — Input                      18989                      89                      050
     ChannelZ       + Input                     200044                      041                     a02
     Channel Z      + Input                      199.70                     +105                    ase

     Channel 2      —input                       20088                      170                     oss

2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 se: Measuring time: 3 sec
                    Gommon mode                     High Range                                  Low Range
                       Input Voltage (m¥)            Average Reading (1V)                   Average Reading (2¥)
    Channel X                           200                        10e                            zou
                                       +200                        «52                            820

    Channel ¥                           200                        t4.18                          1358
                                       ~20                         a1s.t0                        1sse
    Channel Z                           200                        1707                          «1723

                                       —200                        1474                           14.83



3. Channel separation
   DASY measurement parameters: Auto Zero Time: 9 se: Measuring tme: 3 see
                       Input Voltage (mV)__|      ChannelX (xV) |_Channei¥ (s¥)                 ChannelZ (@¥)
    Channel X                    200                       —                    01                   ma7

    Channel ¥                    200                    758                        £                 048
    Channelz                     200                  maz                       aar                       —
Corificate No: DAE4—1291_Dect8                                 Page 4 of5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 see
                                                   High Range (LSB)                  Low Range (LSB)
     Channel X                                            16117                            16241
     Channel Y                                            15930                            16718
     Channel Z                                            16177                            128

5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 see
    Input 10M
                            Average (1V)       min. Offset(V)        max. Offset (uV)      fid.: [():“;;afio"

     Channel X                  —0.59                181                   o.89                    0.47
     Channel Y                   117                 004                   205                     0.45
     Channel 2                  412                  270                   ost                     0.57


6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <251A

7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                    Measuring (MOhm)

     Channel X                                          200                                200
     Channel Y                                          200                                200
     Channel 2                                          200                                200


8. Low Battery Alarm Voltage (Typical values for information)
     Typical values                               Alarm Level (VDC)
     Supply (+ Vee)                                                          +7.9
     Supply (— Vee)                                                          7.6



9. Power Consumption (Typical values for information)
     Typical values                         Switchedoff (mA)        Stand by (mA)       Transmitting (mA)
     Supply (+ Vee)                               +0.01                  16                    +14
     Supply (— Vee)                               —0.01                   —8                    —9




Certiicate No: DAE4—1291_Dect8                           Page 5 of 5



Document Created: 2019-08-27 17:06:53
Document Modified: 2019-08-27 17:06:53

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC