SAR report part 3

FCC ID: 2AFZZ-RMSF6LH

RF Exposure Info

Download: PDF
FCCID_4158508

Calibration
      .     Laboratory of                                    en
                                                            #ar/                 S   Schweizerischer Kalibrierdienst
                                                            $                ~
Schmid & Partner                                           ib\g@                 c   Service suisse d‘étalonnage
  Engineering AG                                           T 7\\ 3                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                {'/,,///_\\\\‘\3     S   Swiss Calibration Service
                                                                frliluals®


Accredited by the Swiss Accreditation Service (SAS)                              Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
         (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
         300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
         paralle! to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2600V2—1058_Jun18                              Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASYS5                             V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom

     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm

     Frequency                                         2600 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.0               1.96 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   37.4 +6 %          2.03 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     14.3 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         55.8 W/kg + 17.0 %(k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                  6.38 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         25.1 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           52.5               2.16 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   51.8 +6 %          2.22 mho/m + 6 %

     Body TSL temperature change during test                      <0.5 °C            ————                    ~——

SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                       250 mW input power                     13.8 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         54.4 Wikg + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                  6.15 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         24.4 Wikg + 16.5 % (k=2)



Certificate No: D2600V2—1058_Jun18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.7 Q — 7.5 jQ

      Return Loss                                                                       —22.4 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           45.3 Q — 6.9 jQ

      Return Loss                                                                       —21.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.149 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                August 14, 2012




Certificate No: D2600V2—1058_Jun18                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                     Date: 19.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1058

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.03 S/m:; &, = 37.4; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e     Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

         Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         Electronics: DAE4 $n601; Calibrated: 26.10.2017

         Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

         DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 117.8 V/m; Power Drift =—0.05 dB
Peak SAR (extrapolated) = 28.5 W/kg
SAR(I g) = 14.3 W/kg; SAR(10 g) = 6.38 W/kg
Maximum value of SAR (measured) = 23.8 W/kg


        dB
       F 0


         —4.00


         —8.00


         —12.00


         —16.00


         —20.00
                   0 dB = 23.8 W/kg = 13.77 dBW/kg




Certificate No: D2600V2—1058_Jun18               Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   19 Jun 2018    14106147
         CHI) sii     1 U Fs          1149.689 0   —7.5430a   81153 pF           2 600.000 000 MHz




         fAvq
         16


         H1d


          CH2




         Ca




         Ay
         16°

         H1d

                    START 2 400.000 000 MHz                              STOP 2 ©00.000 000 MHz




Certificate No: D2600V2—1058_Jun18                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                     Date: 19.06.2018
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1058

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.22 $/m; & = 51.8; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   e     Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

   e     Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         Electronics: DAE4 Sn601; Calibrated: 26.10.2017

         Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

         DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.8 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 28.0 W/kg
SAR(I g) = 13.8 W/kg; SAR(10 g) = 6.15 W/kg
Maximum value of SAR (measured) =23.1 W/kg


        dB
       ! 0


       —| —5.00


         —10.00


         —15.00


         —20.00


         —25.00
                   0 dB = 23.1 W/kg = 13.64 dBW/kg




Certificate No: D2600V2—1058_Jun18              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                      19 Jun 2018    10:09:42
          CHI) sii       4 uors          41 45.258 a   —£.9160 0 ©.9510 pF          2 e00.000 ooo MHz
                                                        T

          Del

          Ca




          fivg
          16


          H1d

          CH2    S11




          Ca




          Av
          16°

          H1d

                       START 2 400.000 000 MHz                               STOP 2 800.000 000 MHz




Certificate No: D2600V2—1058_Jun18                      Page 8 of 8


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       Auden                                                                              Certificate No: DAE4—915_Jun18

|' CALIBRATION CERTIFICATE                                                                                                                           |
 Object                               DAE4 — SD 000 DO4 BK — SN: 915


 Calibration procedure(s)             QA CAL—06.v29
                                      Calibration procedure for the data acquisition electronics (DAE)




 Calibration date:                    June 20, 2018




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               31—Aug—17 (No:21092)                                  Aug—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        04—Jan—18 (in house check)                            In house check: Jan—19
 Calibrator Box V2.1                    SE UMS 006 AA 1002        04—Jan—18 (in house check)                            In house check: Jan—19




                                        Name                                   Function                                 Signature
 Calibrated by:                         Eric Hainfeld                          Laboratory Technician



 Approved by:                           Sven KGhn                              Deputy Manager                             \ \/ (’& UMM f
                                                                                                                          t
                                                                                                                              —

                                                                                                                        Issued: June 20, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—915_Jun18                                        Page 1 of 5


    harati                                                     l9
Calibration Laboratory of                                   x‘                   Schweizerischer Kalibrierdienst
      .                                                    s       >         S
Schmid & Partner                                           iia\éw{/&         c   Service suisse d‘étalonnage
  Englneermg AG                                            I resir               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                4/4@@            S   Swiss Calibration Service
                                                               Autits

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
    e     DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    *    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         e     DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
         e     Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
         e     Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
         e     AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
         e     Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
         e     Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
         e     Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
         e     Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
         «_    Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—915_Jun18                                 Page 2 of 5


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB =     6AuV ,        full range =  —~100...+300 mV
       Low Range:           1LSB =     61nv ,        full range =  *A ssree +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


     Calibration Factors             X                            Y                     Z
     High Range             404.270 + 0.02%(k=2)    404.392 + 0.02% (k=2)   404.745 + 0.02% (k=2)
     Low Range              3.97960 + 1.50% (k=2)   3.99573 + 1.50% (k=2)   3.98894 + 1.50% (k=2)

Connector Angle

     Connector Angle to be used in DASY system                                       1155 °%1°




Certificate No: DAE4—915_Jun18                      Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
     High Range                                Reading (V)          Difference (uV)             Error (%)
     Channel X       + Input                    200034.18                   0.02                0.00
     Channel X       + Input                    20005.72                    0.36                0.00
     Channel X       — Input                    —20001.83                   3.19                —0.02
     Channel Y       + Input                    200032.03                   210                 —0.00
     Channel Y       + Input                    20003.80                    ~1.56               —0.01
     Channel Y       — Input                    —20005.18                   —0.03               0.00
     Channel Z       + Input                    200032.88                   ~1.19               —0.00
     Channel Z       + Input                    20004.47                    —0.69               —0.00
     Channel 2       — Input                    —20006.56                   ~1.42               0.01


     Low Range                                 Reading (uV)         Difference (uV)             Error (%)
     Channel X       + Input                     2001.29                    —0.05               —0.00
     Channel X       + Input                      201.57                    0.05                0.02
     Channel X       — Input                     —197.95                    0.68                —0.34
     Channel Y       + Input                     2001.65                    0.39                0.02
     Channel Y       + Input                      201.72                    0.30                0.15
     Channel Y       — Input                     —~198.98                   —0.32               0.16
     Channel Z       + Input                     2001.02                    —0.33               —0.02
     Channel Z       + Input                      200.26                    ~1,20               —0.60

     Channel Z       — Input                     —199.86                    ~1.22               0.62



2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                        High Range                        Low Range
                        Input Voltage (mV)             Average Reading (V)              Average Reading (uV)

    Channel X                           200                      —10.65                      ~12.47
                                       —200                      14.97                        13.31
    Channel Y                           200                      ~4.02                        ~4.53
                                       200                        3.45                        3.51
    Channel Z                           200                      —0.67                        ~1.07
                                       ~ 200                     ~1.43                        ~1.13


3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)         Channel X (uV)          Channel Y (uV)    Channel Z (uV)

    Channel X                    200                        —                  3.56              —3.86
    Channel Y                    200                    7.89                        —            4.44
    Channel Z                    200                    9.06                   6.41                   —


Certificate No: DAE4—915_Jun18                              Page 4 of 5


4. AD—Converter Values with inputs shorted
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                   High Range (LSB)                   Low Range (LSB)

    Channel X                                              16103                           16654

    Channel Y                                              15971                           16362

    Channel Z                                              15895                           17125



5. Input Offset Measurement
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
   Input 10MQ

                           Average (uV)        min. Offset (uV)      max. Offset (V)       sid. l():\\ll;ation

    Channel X                    0.81                —0.60                2.59                    0.52
    Channel Y                    0.21                —0.79                 1.72                    0.49
    Channel Z                    —0.89               —1.92                0.09                     0.43



6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                 Zeroing (kOhm)                      Measuring (MOhm)

    Channel X                                             200                              200
    Channel Y                                             200                              200
    Channel Z                                             200                              200


8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
    Supply (+ Vec)                                                            +7.9
    Supply (— Vee)                                                            —7.6



9. Power Consumption (Typical values for information)
    Typical values                         Switched off (mA)        Stand by (mA)       Transmitting (mA)
    Supply (+ Vec)                                +0.01                  +6                      +14

    Supply (— Vee)                                —0.01                  —8                      —9




Certificate No: DAE4—915_Jun18                            Page 5 of 5


                                                                                /’l/”
               !\” in Collaboration with                                                \\“l"/“/A *;g:l
                    TTL a                                                                               CNAS




                                                                                 »
                                        CALIBRATION LABORATORY                          4\\ § v cALIBRATION




                                                                                wl



                                                                                                 ',/“
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China           ;/,,I"h‘\“\                       CNAS LO570




                                                                                   &
            Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504
            E—mail: cttl@chinattl.com       Hitp:/www.chinattl.on
      Client :        Sporton                                                   Certificate No: Z218—60487

| CALIBRATION CERTIFICATE

Object                                    DAE4 — SN: 1338

Calibration Procedure(s)                  FF—211—002—01

                                          Calibration Procedure for the Data Acquisition Electronics
                                          (DAEx)
Calibration date:                         December 03, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(2213)°C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards               ID #          Cal Date(Calibrated by, Certificate No.)                  Scheduled Calibration



Process Calibrator 753        1971018              20—Jun—18 (CTTL, No.J18X05034)                               June—19




                                Name                      Function                                         Slgnature
Calibrated by:                   Yu Zongying             SAR Test Engineer                              }%

Reviewed by:                     Lin Hao                 SAR Test Engineer                                   fi&i% ~/"

Approved by:                     Qi Dianyuan             SAR Project Leader                             e_
                                                                                                  Issued: December 05, 2018
This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.




     Certificate No: Z18—60487                               Page 1 of 3


                              ®     in Collaboration with
         éTTL a
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Hittp://www.chinattl.en

Glossary:
DAE                               data acquisition electronics
Connector angle                   information used in DASY system to align probe sensor X
                                  to the robot coordinate system.


Methods Applied and interpretation of Parameters:
«_   DC Voltage Measurement: Calibration Factor assessed for use in DASY
     system by comparison with a calibrated instrument traceable to national
     standards. The figure given corresponds to the full scale range of the
     voltmeter in the respective range.

&    Connector angle: The angle of the connector is assessed measuring the
     angle mechanically by a tool inserted. Uncertainty is not required.

&    The report provide only calibration results for DAE, it does not contain other
     performance test results.




 Certificate No: Z18—60487                                  Page 2 of 3


        E‘ZTATL l;Collaborafionewim a

           @imzzae~~               CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com        Hitp://ww.chinattl.cn


DC Voltage Measurement
   AID — Converter Resolution nominal
      High Range:     1LSB=       6AuV ,      full range =    ~100...+300 mV
      Low Range:     1LSB =        61nV ,     full range =    41     +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


    Calibration Factors                     X                   ‘          Y               Z              |
    High Range                     403.662 + 0.15% (k=2) ’ 404.235 + 0,15% (k=2)   404.194 + 0.15% (k=2) |

   | Low Range                     3.97131 + 0.7% (k=2) ]’3‘97736 £0.7% (k=2)      3.97366 + 0.7% (k=2)   |




Connector Angle

   (finnector Angle to be used in DASY system                                            82554 1°




Certificate No: Z18—60487                               Page 3 of 3


Calibration Laboratory of                                                                       . Seiwelzerischer Kalbrertfonst
Schmid & Partner                                                                                q Service suisse d‘étaonnage
  Engineering AG                                                                                g Servitlo sviezero d taratura
Zoughausstrasse 43, 8004 zurich, Switzedand                                                       Swiss Caliration Service
Accredted by he Swise Accredtation Senvce (SAS)                                                 Accreditation No.: SCS 0108
The Swiss Acsreditation Service is one ofthe signatories to the EA
MuTtLateral Agrooment fothe recognition of calibrtion cortficatos
ciew       (Sporton                                                                   (CarificateNo: ES3—3203_Oct18

|CALIBRATION CERTIFICATE
Object                            EsspV3—sN:3203                                  2

Catbraton pacedure(s)



Caltration tate                   October 25, 2018
This calbraion cerficate documentsthe raceably tnatonastandards, which reatze the physical untsof measurements (S)
The measurements and the unceriainies wih confdence probatily are guen on thefolowing pages and are part ofthe certfcate
Allcalbeations have been conducted in the closed laborelary facity; enviranmenttemperatre 22 + 3°C and bumidty« 70%
Catbration Ecusment used (MBTE criicalfo calbraton)

 Priary Standarts                o                              GalDate Coriicate No)                 Scneduled Caltvation
 Power meterNR®                  si torrra                      otAp18 (No.217—orerznzere)            Apcie
 Power senso: NRP:231            sn rosaid                      ot—Anct® (No 217—000r2)               Apeto
 Power senser NRP—231            sNt rosms                      ot—Ap—18(No217—0z0r0)                 Apeto
 Reference 20 dB Atenater        SN 5277 (20x)                  Ot—An—18 No 217—0r000)                Apers
 Reference Prbe ESTOV2           ENe                            30.Dee—17 (No: ES3—3013Dect)          Decis
 onga                            sit seo                        2i—Dee17 (No: DaE—a00Decin)           beci#
 Secondary Sandarts              io                             Check bate i house)                   Scheduled Check
 Power meterE«e198               su corzosers                   06—A—16 in house chack Jun—16)        in house chack Jn20
 Power sensor E44128             stt mversssosr                 06—Ape16 in house check Jun—18)       in house chacks Jun20
 Power sensor E442A              shcoomezto                     06—Aor16 in house chedk Jun—18)       in house chacks Jin20
 R generstor e soi8G             sn usssszuorr00                0t—Aup49 in house check Jun—18)       in nouse cneet Jun—20
 Netwo Anazer EesseA             sh: userosour?                 31—Mar—18 n house check Oct—18)       in house theck Oct—to
                                   Name                               Functon                         Sgnature
Calbrated ty                       \oton Kasiad                       UibomtoyTechnician             q 5 E%_,

Approved by                        KaisPobovie                        Techrical
                                                                                               erve    Issued: October 25, 2018
 This caltraien corficale shalnot be rapreduced sxcent in fut wihout witen approval fthe aboratary


Certficate No: E83—3203_Octt8                                 Page 1of 11


Calibration Laboratory of                                                                   Setwelzorichor Kallbrarcionst
Schmid & Partner                                                                       g Service suisso d‘ittonnage
  Engineering AG                                                                         Servizio svizzero d taratura
Zeughausstrasse 43, 004 Zurich, Svitzeriand                                            $ suise Caltration Service

Accinted by ns Sss Accredtaton Serice(SA)                                                Accreditation No:: SCS 0108
The Swiss Accreditation Service is one ofthe signatoriesto the EA
MultlateralAgreementfor the recognition ofcalbration cortiicates

Glossary:
TSt                        tissue simulating liquid
NORMxyz —                  sensitvily in free space
Gony®                      sensitivity in TSL / NORMcy.z:
DCP                        diode compression point
or                         rest factor (1iduty..cycle) of the RF signal
A.8,.C.0                   modulation dependent linearization parameters
Polarization o             9 rotation around probe axis
Polarization 8             8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                           i. 8 = 0 is normal to probe axis
Connector Angle            information used in DASY system to align probe sensor X to the robat coordinate system
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
        Absorption Rate (SAR)in the Human Head from Wireless Communications Devices: Measurement
        Techniques®, June 2013
    b) 1E $2209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from hand—
         held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz), July 2016
    0 EC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
         used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)®, March 2010
    d KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
        NORMy,z: Assessed for E—field polarization 8 = 0 (f < 900 Mz in TEM—celt 1 > 1800 Mriz: R22 wavegulde)
        NORMxy,z are only intermediate values,i. the uncertainties of NORMy,z does not affect the E—feld
        uncertainty inside TSL (see below Conv),
         NORM(Jxy.z = NORMy.z * frequency..response (see Frequency Response Chart).This Inearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in the stated uncertainty of ConvF.
         DCPxy,z: DCP are numerical Inearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
        PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
        characteristis
        Axy.e: Buy.z: Coy.a: Day.z; VRey.a: A, B, C, D are numerical inearization parameters assessed based on
        the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
        media. VR is the maximum callbration range expressed in RMS voltage across the diode
         ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—leld (or Temperature Transfer
         Standard for f 800 MHz) and inside waveguide using analytical field distrbutions based on power
         measurements for 1 > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
         to NORM,y.z " ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validityfrom + 50 MHz to x 100
         MHz.
         Sphericalisotropy (3D deviation from isotropy}; in a fild of tow gradients realized using a fiat phantom
         exposed by a patch antenna
         Sensor Offset: The sensor offset corresponds to the offset of virtval measurement center from the probe tip
         (on probe axis), No tolerance required
         Connector Angle: The angle is assessed using the information gained by determining the NORM (no
         uncertainty required).

Certficale No: ES3—3203_Oct18                              Page 20(11


essove — sNazse                                                                October 25, 2018




                      Probe ES3DV3

                                         SN:3293

                                 Manufactured:            July 6, 2010
                                 Repaired:                October 19,2018
                                 Calibrated:              October 25, 2018



                                Calibrated for DASY/EASY Systems
                                   (Note: non—compatible with DASY2 system!)




Certficate No: E53—3203_Oct18                    Page 3 of 11


Esspvs— sN—3z0                                                                                                     October 25, 2018


DASY/EASY — Parameters of Probe: ES3DV3 — SN:3293
Basic Calibration Parameters
                                   |        SensorX                       ‘Sensor Y                 SensorZ             Un (ke2)
 Norm (@Vi(VimYY                               109                           080                      o73               £10.1%
[DGP (my)"                                    1043                           1055                     1078

Modulation Calibration Parameters
 uo           Communication System Name                              A           s          c          o         va         ind
                :                                                    a8       dBviV                   c          mV         (ke2)
 o            ow                                             x       oo         so          10       "|_ou0     tss        #0%
                                                             y       oo         00          t                   mer
                                                             2|      oo          e      |   10                  1era

 The reported uncertainty of measurement is stated as thestandard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


1 The unceriaties of Norm XY2 do not affetthe Eid uncertainty nside TSL (see Pages 5 and 6)
" Numereaincarzaton parameter uncertny no eauied
* Uncertantis deweninad using the max. devition fom Inear response applying rectanguladstibution and is expressed fo the square of the
feld value




Cortficate No: ES3—3203_Oct18                                 Page 4 of 11


Essove— sN:3z0s                                                                                                    October 25, 2018



DASY/EASY — Parameters of Probe: ES3DV3 — SN:3293

Calibration Parameter Determined in Head Tissue Simulating Media

     1 (Mite)©     wm:;y al           cn&n‘&‘"m         ConvEX       ConyEY       ConvEZ        Aipha®     D:'n?'n??       r:t':zc)
       750              419               o8              678          se           s.78         ose         120        £120%
        sas _           415               o.%0            sa7          47           ear          o4          180        £120%
       EM               a1s               oo7             634          a34          634          os0         120        +120%
       1750             401               137             540          540          s4o          ore         120        +120%
       1900             400               140             5.19         s10           s1e         ose         iss        £120%
       2000             400               140             sa7          547          517          os0         141        +120%
       2300             sas               167             483          4ss          ass          ore         is1        +120%
       2450             sa2               1.80            ass          453          ass          oso         i23        £120%
       2600             ES                196             a44          444          as4          os0         131        £120%

© Frequeney valdty above 910 Mz ot 2 100 kz only applis for DASY v4.t and hgher (se0 Page 2), se t s restictd t 2 50 Nite. The
uncertinty s the RSS ofthe Gorv® unceriity t calbraton feauency and the uncertantforthe inicated roquency band. Frequency valdty
below 300 itz is2 10,25,49, 50 and 70 Ntz for Conveassessments t 30, 64, 128, 150 and 220 Ntz respectvay. Above 5 Gite hrequency
yaldty can be extended to 2 110 Mitz
" Attrequencis below 3 GHz, the valdty of tesue parameters(cand a) can be relased to : 10% i lavcompensation formula is agpled to
measured SA values. At hequencies above 3 GHz,the vality f lisue parameters (and o) s restieled to 2 5% The uncerlanty is the RSS of
the Corv uncertainyfoindcated trgettssue paramoters
* AphoDepth are daternined durng caltration. SPEAG warrants tht the remaining deviaton du to the boundary effctater compensaton is
always less than + 1% fofeguenciasbelow 3 Gitz and below + 23 forfreguencies between 3—6 GHteat any clstance largerthan hat the probe t
clameterrom the boundary.




Certficate No: ES3—3209OcitB                                 Page Sof 11


esspvs— sh—azes                                                                                                  October 25, 2018



DASY/EASY — Parameters of Probe: ES3DV3 — SN:3293

Calibration Parameter Determined in Body Tissue Simulating Media

     +eg®          h?r:m:i.ty’       Corzg’:nc’fi,my ConvEX           ConvEY       ConvEZ Aipha® "('..‘.’,";.‘,“          (I::zn)
      750              ss5               096         634              634          634    oso      112                  £120%
        sss _           se2               o9r             a24         624           624         seo         i18_|      +120%
       1750             sea               149             sor         sor           sor         oss         18         +120%
       1900             sas               152             474         474           474         os0_|       144        £120%
       2300             529               181             451         451           a5s1        oo          128        £120%
       2450             s2r               195             430         430           439         os0         122|       £120%
       2500             s25               216             428          428          428         o%0         120        +120%

© Frequeney valdy above 300 NKz o s 100 Mz ony apples for DASY i.4 and nigher (seo Page 2is s resticted to + 50Mz The
uncetainy is he RSS of the Gorv® uncertint atcalivatonfreauency and the uneatainy for the indlatod requency band. Frequency valdty
below 300 Nike is 10,25,40, 50 and 70 Kz forGonv® assessments at 30. 64, 128, 180 and 220 NHerespectvey. Above 5 Git frequency
yaldity cabe entendadto s 110 itz
* M frequencies below 3 GHz,the valdtyoftssue parameter (and ) can be relaxedto : 10% i lquid compansation fomuta s anpled to
measured SA values.Atrequencies above 2 GHz, the valdtyof esue parametars(sand )is restcted o 2 5%. The uncertiny s the RSS of
the GomvE uncetainy forincicatedtarget tssue paramaters
" AhalDap‘h are deternined durng calbrationSPEAG warrants tht theremaining devaton dueto the boundary effetatercompensation is
alvays less than 2 13 for requencies below 3 Gite and balow 2 2% for requencies batween 3.6 GHz atany dtance argerthan half e probe tp
lamaterfrom the boundary




Cortficate No: E83—3203_Oct18                                Page 6 of 11


Esspvs— sviares                                                                                         October 25, 2018

                                            Frequency Response of E—Field
         Frequency response {normalized)         (TEM—Cell:if110 EXX, Waveguide: R22)




                                           Uncertainty of Frequency Response of E—field: 2 6.3% (k=2)




Gertficate No: E3—3203_Oct1                                       Page 7 of 11


Essovs— shiazes                                                                             October 25, 2018



                                 Receiving Pattern (¢), 9 = 0°

                    1600 MHz,TEM                                        11800 MHz.R22




                                                                                                 BC
     Einor {4B)




                   To            To       04s                  t       &           io        o
                                                            RoWl]
                     whide                   shiitle                  w                 witthe
                               Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Cortficate No: E3—4203_Octta                           Page 8 of 11


Essovs— N3203                                                                                                          October 25, 2018




                                                    Dynamic Range f(SARneaq)
                                                           (TEM cell , fa,= 1900 MHz)


                                     10°



                                     10%
                 Input Signal [uV)
                                     a




                                     103




                                     10                           debtdiiihndeds
                                                                         t
                                           103                           107            19
                                                                         SAR {mWiem3]




                                                                                             4
                                                                                             NN
                                           i 109          103            101           100                 1ot   10t
                                                                         SAR [nWiem?]                  .
                                                            [+]
                                                       not compensated                           compansated

                                                   Uncertainty of Linearity Assessment: 0.6% (k=2)




Certficate No: ES3—3208_Oct18                                            Page Sof 11


Essovs— sazes                                                                                    October 25, 2018

                                Conversion Factor Assessment
                1= 200 MHz.WGLS RO (4_cons£)                          1= 1750 MiizWNGLS R22 (H_con?)




                                «h    *       5                                     «ires
                    walite                  wotk

                             Deviation from Isotropy in Liquid
                                              Error (6, 9), £= 900 MHz




                         —10    08    —hs     cos   cor   o0     o2    04    o8   o8        10
                             Uncertainty of Spherical Isotropy Assessment: # 2.6% (K=2)


Cortficate No: ES3—3203_Oot18                        Page 10 of 11


Esapvs— sNigags                                           October 25, 2018



DASY/EASY — Parameters of Probe: ES3DV3 — SN:3293

Other Probe Parameters
 Sensor Arangement                                            Triangular
 Connector Angle (°)                                                ‘on
 Mechanical Surface Detection Mode                              enabled
 Optical Surlace Detection Mode                                 disabled
 Probe Overall Length                                           337 mm
 Probe Body Diamater                                             10 mm
Tip Lengin                                                       10 mm
 Tip Diameter                                                      4 mm
 Probe Tip to Sensor X Callbration Point                           2 mm
 Probe Tip to Sensor Y Calibration Point                           2 mm
 Probe Tip to Sensor 2 Callbration Point                           2 mm
 Recommended Measurement Distance from Surface                     3 mm




Certficate No: ES3—4203_Oct18                Page 11011



Document Created: 2019-03-22 00:49:00
Document Modified: 2019-03-22 00:49:00

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC