SAR report part 2

FCC ID: 2AFZZ-RMSDG1

RF Exposure Info

Download: PDF
FCCID_3660750

C(
                                    &     in Collaboration with


                      TTL $B—.2—8_—BI—
             Add: No:51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2079     Fac +86—10—62304633—2504
             E—mail: cftl@chinatil.com       hitpy/wis.chinattl.en



     DASYS Validation Report for Body TSL                                        Date: 08.27.2017
     Test Laboratory: CTTL, Beijing, China
     DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40020
       Communication System: UID 0, CW; Frequency: $35 MHz; Duty Cycle: 1:1
        Medium parameters used: F= 835 MHz; 0 = 0.984 S/m; &; = 55.62; p =1000 kg/m’
        Phantom section: Right Section
        Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
     DASYS Configuration:

                 *    Probe: EX3DV4 — SN3617; ConvF(9.64,9.64, 9.64); Calibrated: 1/23/2017;
                 *«   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
                 *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
                 *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
                 *    Measurement SW; DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                      (7417)

             Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
             dy=Smm, dz=5mm
             Reference Value= 56.55 V/m; Power Drift = 0.02 dB
             Peak SAR (extrapolated)=3.71 W/kg
             SAR(I g) =2.46 W/kg:; SAR(10 g)=1.63 W/kg
             Maximum value of SAR (measured) = 3.29 W/kg


                dB
                0

                 ~2.05


                 410

                 5.14




                ~10.24
                              L.
                            0 dB =3.29 Wkg=5.17 dBW/kg




     Certificate No: Z17—97114                             Page 7 of 8


         Add: No,51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
         E—mail: ctil@chinattl.com      hitp://Anww.chinattl.en


         Impedance Measurement Plot for Body TSL

          Tri 512 Log nag 10. 00087   Ref 5.ooods [Fi]
            £0—00    71835100000 ana ~24.754 db
            40.00             ;                                              —
            30.00        —4                —        f=—       4x
            20.00    |————_.—           oceces




            50. 00
         WR si1 snith (rejx) Scate 1.000u [r2 pel]
          »2. ©25,00000 mz 46.762 n —4.5737 0. «1.67




Certificate No: Z17—97114                              Page 8 of 8


                                           in Collaboration with                   '\“:\%; A ::2;{
                                           CALIBR&ATION LABORATORY                      e CNAS Bik
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   -’«,        \.:                       CALIBRATION
             Tel: +86—10—62304633—2079         Fax: +86—10—02304633—2504            /"[.f,m\.“                         CNAS LO570
               E—mail: cttf@chinattl.com       hitp://www.chinattl.en
      Client                TA(Shanghai)                                      Certificate No:      217—97002

CALIBRATION CERTIFICATE

Object                                      D1750V2 — SN: 1033

Calibration Procedure(s)                    FD2z11—003—.01

                                            Calibration Procedures for dipole validation kits

Calibration date:                           January 10, 2017

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S!). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22:3)‘C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                   ID #                Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Power Meter         NRP2           101919             27—Jun—16 (CTTL, No.J16X04777)                           Jun—17
 Power sensor NRP—291               101547             27—Jun—16 (CTTL, No.J16X04777)                          Jun—17
 Reference Probe EX3DV4             SN 7307            19—Feb—16(SPEAG,No.EX3—7307_Feb16)                      Feb—17
 DAE4                               SN 771            02—Feb—16(CTTL—SPEAG,No.Z16—97011)                       Feb—17

 Secondary Standards                ID#               Cal Date(Calibrated by, Certificate No.)         Scheduled Calibration
 Signal Generator E443BC            MY49071430        O1—Feb—16 (CTTL, No.J16X00893)                           Jan—17
 Network Analyzer E5071C            MY46110673        26—Jan—16 (CTTL, No.J16X00894)                           Jan—17



                                     Name                          Function                                  Signature
Calibrated by:                     Zhao Jing                  SAR Test Engineer                               gz
Reviewed by:                       Oi Dianyuan                SAR Project Leader                   :;_fi,f\_/
Approved by:                       Lu Bingsong                 Deputy Director of the laboratory           % 1;%

                                                                                         Issued: January    12, 2017
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: 217—97002                                Page 1 of 8


C(
                                      in Collaboration with

                  TTLQ_MH_
                                      CALIBRATION LABORATORY

          Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
          Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
          E—mail: cttl@chinatl.com       hip/www.chinattkon


  Glossary:
  TSL                       tissue simulating liquid
  ConvF                     sensitivity in TSL / NORMx.y,z
  N/A                      not applicable or not measured


 Calibration is Performed According to the Following Standards:
 a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
    Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
 b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
    devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
      2005
 ) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
      30MHz to 6GHz)", March 2010
 d) KDBB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


 Additional Documentation:
 e) DASY4/5 System Handbook


 Methods Applied and Interpretation of Parameters:
 «_    Measurement Conditions: Further details are available from the Validation Report at the end
       of the certificate. All figures stated in the certificate are valid at the frequency indicated,
 ®     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
       point exactly below the center marking of the flat phantom section, with the arms oriented
       paralle! to the body axis.
 *     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
       positioned under the liquid filled phantom. The impedance stated is transformed from the
       measurement at the SMA connector to the feed point. The Return Loss ensures low
       reflected power. No uncertainty required.
 «_    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
       No uncertainty required.
 «_    SAR measured: SAR measured at the stated antenna input power.
 «     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
 &     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
       nominal SAR result.


      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement multiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probability of approximately 95%.


 Certificate No: Z17—97002                              Page2 of 8


                              *     In Collaboration with


                m‘ Immmmin
        Add: No.51 Xueyuan Road. Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl,com      hitp:/Awinw.chinattl.en


Measurement Conditions
   DASY systam configuration, as far as not given on page 1.
     DASY Version                                           DASY52                                   52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                           10 mm                                  with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       1750 MHz & 1 MHz


Head TSL parameters
   Thefollowing parameters and calculations were applied.
                                                            Temperature             Pormitivity            Conductivity

     Nominal Head TSL parameters                              22.0 °C                  40.1                1.37 mhoim

     Measured Head TSL parameters                        (22.0 £ 0.2) °C             39.4 +6 %           1.35 mho/m + 6 %

     Head TSL temperature change during tost                  €1.0 °C                   «——                      «—
SAR result with Head TSL
     SAR averaged over1_Ci" _(1 g) of Head TSL                          Condition
     SAR measured                                                   250 mW input power                 9.27 mW / q

     SAR for nominal Head TSL parameters                             normalized to 1W         37.2 mW /g + 20.8 % (k=2)

     SAR averaged over 10 cm" (10 g) of Head TSL                        Condition
     SAR measured                                                   250 mW input power                 4.90 mW /g

     SAR for nominal Head TSL parameters                             normalized to 1W         19.7 mW 7g £ 20.4 % (k=2)

Body TSL parameters
  The following parameters and calculations were applied.
                                                            Temperature             Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  §3.4                1.49 mha/m
     Measured Body TSL parameters                        (22.0£0.2) °C               §3.1 46 %           1,48 mho/m £ 6 %

     Body TSL temperature change during test                  «1.0 °C                   «~                       —

SAR result with Body TSL
     SAR averaged over 1 _Cm" _(1 g) of Body TSL                        Condition
     SAR measured                                                   250 mW input power                 9.40 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W         37.6 mW 1g # 20.8 % (k=2)

     SAR averaged over 10 Ci" (10 g) of Body TSL                        Condition
     SAR measured                                                   250 mW input power                 5.03 mW /g

     SAR for nominal Body TSL parameters                             normalized to 1W         20.1 mW /g + 20.4 % (k=2)

Certificate No: £17—97002                              Page 3 of 8


C                                                                                            .
                              *     In Collaboration with

                TTL a
                                    CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      hitps/wwi.chinattl.en


  Appendix

 Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                  49.80+ 0.93J0

     Return Loss                                                              ~40.30B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   44.70— 0. 1010

     Return Lass                                                              —25.008


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.327 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable, The center conductor of the feeding line is directly
 connected to the second arm ofthe dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in orderto improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still accarding to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend orthe soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufacturad by                                                             SPEAG




Certificate No: Z17—97002                             Page 4 of 8


C(
            fi 1; Cu{labomrinnewifll

                                     CALIBRATION LABORATORY

         Add: No.31 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fux: +86—10—62304633—2504
         E—mail: cttl@chinatt.com       hup?//www.chinatil.en


 DASYS Validation Report for Head TSL                                        Date: 01.10.2017
 Test Laboratory: CTTL, Beijing, China
 DUT: Dipole 1750 MHz; Type: D1750V2; Serial; D1750V2 — SN: 1033
     Communication System: UID 0, CW; Frequency: 1750 MHz; Duty Cycle: 1:1
     Medium parameters used: {= 1750 MHz; o = 1.352 $/m; er = 39.36; p = 1000 kg/m3
     Phantom section: Center Section
     Measurement Standard: DASYS (IEEE/IEC/ANST C63.19—2007)
 DASYS Configuration:

             +    Probe: EX3DV4 — SN7307; ConvF(8.37, 8.37, 8.37); Calibrated: 2/19/2016:
             +    Sensor—Surface; 1 Amm (Mechanical Surface Detection)
             «_   Electronics: DAFE4 $n771; Calibrated: 2/2/2016
             *«   Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
             *    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)


         System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0; Measurement grid:
         dx=5mm, dy=Smm, dz=5mm
         Reference Value = 98.21V/m; Power Drift= 0.03 dB
         Peak SAR (extrapolated) = 17.1 Wkg
         SAR(I g) =9.27 W/kg; SAR(10 g) =4.9 W/kg
         Maximum value of SAR (measured) = 14.4 W/kg


            dB
             0


             ~3.39


             —6.78


             1017


             13.56


             16.95
                       B
                       0 dB = 14.4 W/kg = 11.58 dBW/kg




 Certificate No: Z17—97002                             Page 5 of 8


C                                                                                   .
             t"                          In Collaboration with
             _‘/"7‘J, a
             sc                          CALIBRATION LABORATORY

         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079           Fimc: +86—10—62304633—2504
         E—mail: cttl@chinat],com            hitp://wwiy.chinattl.en




         Impedance Measurement Plot for Head TSL


      Tri i1 Log Mag 10.00087 Rer 0.00008 TFZ]
        5900 rex—repupoc ons —t0.33 oo
        «0. 00
        30.00
        20.00
        10. 00
        ©,000 y                                                                               C
       —10.00
       ~20.00

       ~30.00
       ~48. 00

      —50.00                                                                x
     MBR s11. smith (@+4>0 scale 1.000u {ri. on7]
       »1.   1.7500000 owz   49.755 o.     927.18 mo       84.322—pl

                                                       /




    it sunisseen                                                Fowuoo he       o   useaienes m




Certificate No: Z17—97002                                     Page 6 of 8


C(
              !-\*                      in Collaboration with
              @7/"/°J,
              ‘
                       a
                              g         CALIBRATION LABORATORY
         Add: No.31 Xueyuan Road, Haidian District, Beijing, 100191, China
          I        6—10—62304633—2079       Fax: +86—10—62304633—2504
         E—mail: cttl@chinatt.com           hitp//ww.chinattl.on




 DASYS Validation Report for Body TSL                                           Date: 01.10,2017
 Test Laboratory: CTTL, Beijing, China
 DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1033
     Communication System: UID 0, CW; Frequency: 1750 MHz; Duty Cycle: 1:1
     Medium parameters used: £= 1750 MHz; a= 1.484 S/m:; se = 53.05; p = 1000 kg/m‘
     Phantom section; Left Section
     Measurement Standard: DASYS (IEEE/IEC/ANST C63.19—2007)
 DASY3 Configuration:

              *      Probe: EX3DV4 — SN7307; ConvF(8.18, 8.18, 8.18); Calibrated: 2/19/2016:
              *«     Sensor—Surface: 1.4mm (Mechanical Surface Detection)
              *      Electronics: DAE4 Sn771: Calibrated: 2/2/2016
              *      Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
              *      Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

         System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0; Mcasurement grid:
         dx=5mm, dy=5mm, dz=Smm
         Reference Value = 86.52 V/m; Power Drift= 0.04 dB
         Peak SAR (extrapolated) = 16.6 W/kg
         SAR(I g) = 9.4 W/kg; SAR(1O g) = 5.03 W/kg
         Maximum value of SAR (measured) = 14.1 W/kg


              dB
              0


              —3.26


              5.52


              —9.78


              13.04


              "16.30

                          0 dB =14.1 W/kg =11.49 dBW/kg


 Certificate No: Z17—97002                               Page 7 of 8


         FCC
           C SAR Test Re
                       eport                                                              Report No: RXA1710-0339
                                                                                                                9SAR02R3




TA Technology (Shangh       hai) Co., Ltd.                     TA-MB-04-00   03S                                  Page 145 of 172
This report sh
             hall not be reprroduced exceptt in full, withoutt the written ap
                                                                            pproval of TA Teechnology (Shan
                                                                                                          nghai) Co., Ltd.


                                                                                                               gaf
            Add: No.51 Xueyuan                                                                                 CALIBRATION
            Tel: +86—1 0—62304633—2079 Heldian District, Beijing, 100191, China   %                            CNAS LO570
            E—mail: cttl@chinatil.com       Fax: +86—10—62304633—2504
                                            hups//www.chinattLen
                              Shan


Object                                   |Bigaove EK sgoeo mm .
                                                                                          ycelen      :
Calibration Procedure(s)                 PZTi
                                         (FF E—00
                                              e omm rma
                                                      3—01                            nmaagmemememe<
                                         \GallbrationProcedures for dipolevalidationkits. _
Calibration date:
                                         (August26,20170
This calibration Certificate documents the traceability to national standards, which
                                                                                     realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given
                                                                                               on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperatu
                                                                                              re(2#3)C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)

Primary Standards                 ID #                Cal Date(Calibrated by, Certificate No.)     Scheduled Calibration
 Power Meter    NRVD              102083             22—Sep—16 (CTTL, No.J16X05809)                      Sep—17
 Power sensor NRV—Z5              100595             22—Sep—16 (CTTL, No.J16X06809)                      Sep—17
 Reference Probe EX3DV4           SN 3617            23—Jan—17(SPEAG,No.EX3—3617_Jan1 7)                 Jan—18
 DAE4                             SN 1331            19—Jan—17(CTTL—SPEAG,No.217—97015)                  Jan—18

 Secondary Standards              ID#                Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Signal Generator E4438C          MY49071430         13—Jan—17 (CTTL, No.J17X00286)                      Jan—18
 Network Analyzer E5071G          MY46110673         13—Jan—17 (CTTL, No.J17X00285)                      Jan—18



                                                                 Fu_nt:}i9_n o aatminle rurtmnre      _Signature
Calibrated by                             ng      ||          BAR fost Engreer                           *
Reviewed by:                                     VC USAR TestEngineer
Approved by:                    (a Dianyuan               ~;AF§_Pm]ect Leader
                                Kss




     Certificate No: Z17—97115                               Page 1 of 8


                   %TTL CAliEr
                        a     anon LABORATORY
                Add: No.51 Xueyuan R         idian Distri
                Tel: +86—10—6
                      i
                              2304633.
                                a
                                       2079—onFax:Stict.
                                                    +86— Beijing, 100191,
                                                                     z
                                                                            io
                                                                          China
                E—mail: cttl@chinattl.com            hitp:”86 mc':fi(::l?: 2504


        Glossary:
        ‘cr:SL ¢                  tissug simulating liquid
        N?/I\W                    sensitivity in TSL / NORMx.y,z
                                  not applicable or not measured


         Calibration is Performed According to the Following
                                                             Standards:
         a) |EEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
           Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
           Communications Devices: Measurement Techniques", June 2013
         b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
            exposure to radio frequency fields from hand—held and body—mounted wireless
            communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
            6GHz)", July 2016
          ) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
             communication devices used in close proximity to the human body (frequency range of
             30MHz to 6GHz)", March 2010
          d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

          Additional Documentation:
          e) DASY4/5 System Handbook


          Methods Applied and Interpretation of Parameters:
          * Measurement Conditions: Further details are available from the Validation Report at the end
             of the certificate. All figures stated in the certificate are valid at the frequency indicated.
          & Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
             point exactly below the center marking of the flat phantom section, with the arms oriented
             paralle! to the body axis.
          & Feed Point Impedance and Return Loss: These parameters are measured with the dipole
             positioned under the liquid filled phantom. The impedance stated is transformed from the
             measurement at the SMA connector to the feed point. The Return Loss ensures low
             reflected power. No uncertainty required.
          * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
             No uncertainty required.
          e SAR measured: SAR measured at the stated antenna input power.
           e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
             connector.
           * SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
              nominal SAR result.


              The reported uncertainty of measurement is stated as the standard uncertainty of
              Measurement multiplied by the coverage factor k=2, which for a normal distribution
              Corresponds to a coverage probability of approximately 95%.



           Certificate No: Z17—97115                                 Page 2 of 8




                                                           mm nrcn>
nis report shuit HOL UE 1EprUUuteu EAUtpt n j um renerneme


                               ®                   &     :
           A                           In Collsboration with

           EI°TI, a
                  CALIBRATION LABORATORY
         Add:+86—
         Tel: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
              ' 10—62301 0163
                           6333—2079   Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com
                                        http://www.chinattLen
 MeDasurement Condition
                                   s
      ASY System configuration, as
                                   far as not given on page 1.
      DASY Version                                    DASY52                                            52.10.0.1446

      Extrapolation                                Advanced Extrapolation
       Phantom                                     Triple Flat Phantom 5.1C
       Distance Dipole Center — TSL                            10 mm                                  with Spacer
      Zoom Scan Resolution                             dx, dy, dz = 5 mm
       Frequency                                       1900 MHz +1 MHz


 Head TSL parameters
    The following parameters and calculations were applied.
                                                               Temperature             Permittivity           Conductivity
       Nominal Head TSL parameters                               22.0 °C                  40.0                1.40 mho/m
       Measured Head TSL parameters                          (22.0 £0.2) °C             39.9 £ 6 %          141 mho/m + 6 %
       Head TSL temperature change during test                   <1.0 °C                   —                        «=
 SAR result with Head TSL
      SAR averaged over 1_Cm°_(1 g) of Head TSL                            Condition
      SAR measured                                                     250 mW input power                 10.1 mW /g

      SAR for nominal Head TSL parameters                               normalized to 1W         40.1 mW /g £ 18.8 % (k=2)

      SAR averaged over 10 Cm° (10 g) of Head TSL                          Condition
      SAR measured                                                     250 mW input power                 5.19 mW /g

      SAR for nominal Head TSL parameters                               normalized to 1W         20.7 mW Ig + 18.7 % (k=2)

 Body TSL parameters
    The following parameters and calculations were applied.
                                                               Temperature             Permittivity           Conductivity

      Nominal Body TSL parameters                               22.0 °C                   53.3                1.52 mho/m
      Measured Body TSL parameters                           (22.0 £ 0.2) °C            53.6 46 %           1.53 mho/m £ 6 %

      Body TSL temperature change during test                   <1.0 °C                    —                        ——
SAR result with Body TSL
   SAR averaged over 1 cm* _(1 g) of Body TSL                              Condition
     SAR measured                                                   250 mW input power                    9.90 mW /g
     SAR for nominal Body TSL parameters                               normalized to 1W          39.5 mW /g + 18.8 % (k=2)

     SAR averaged over 10 cm° (10 g) of Body TSL                           Condition
     SAR measured                                                   250 mW input power                    5.21 mW /g

     SAR for nominal Body TSL parameters                               normalized to 1W          20.8 mW /g # 18.7 % (k=2)
Certificate No: Z17—97115                               Page 3 of 8


                                s                   nlue
                   Z"TL a
                        CausRaAnoNn LaBorATORY
          Add: No.ST Xueyuan Road. Haidian District, Beijing, 100191, China
          Tek:
          Ef' £$86—10—62301633—2070      Pax:+86—10—02300633—2508
            malls eftl@chinatt.com         hnpy/wwwchinatten
  Appendix (Additional assessments outside the scope of CNAS L0570)
  Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                   52.00+ 6.5910
       Return Loss                                                              —23.408

  Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                   §52.70+ 8.35]0
       Retumn Loss                                                              —21.40B

  General Antenna Parameters and Design

       Electrical Delay (one direction)                          l              1.302 ns             J



  After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can
  be measured.

  The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
  connected to the second arm of the dipole. The antennais therefore short—ciroulted for DC—signals. On some
  of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
  according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
  affected by this change. The overall dipole length is still according to the Standard,
  No excessive force must be applied to the dipole arms, because they might bend or the soldered
  connections near the feedpoint may be damaged.



  Additional EUT Data

      Manufactured by                                                              SPEAG




Certificate No: Z17—97115                              Page 4 of 8


C(
                ammime..® in Colleborationwith
                ©T7TL
                “(
                      a
                      CAuERATON Laporarory
              olo.51 Xueyuan Road. Haidion District, Beijing. 100191, China
                      10—62301633—2079      Fax: +86—10—62304633—2504
              E—mails Cl @chinatil.com      hitpo//wwew chinaiten
      PASYS Validation Report for Hea    d TSL
     Test Laboratory: CTTL, Bei
                                jing, China                                        Date: 08262017
           : Dipole 1900 MHz; Type: D19
                                        00V2; Serial; DL900V2 — SN: 540
         Com                                                           60
             munication System: UID 0, CW; Frequency
         Medium parameters used: f= 1900 MH         : 1900 MHz; Duty Cycle: 1:1
                                           z; a = 1.413 S/m; er= 39.85; p= 100
         Phantom section: Left Section                                         0 kg/m3
         Measurement Standard: DASY5 (IEE
                                          E/EC/ANSI C63.19—2007)
     DASY5 Configuration:

                  *    Probe: EX3DV4—SN3617; ConvF(8.26, 8.26, 8.26); Calibra
                                                                              ted: 1/23/2017;
                  *    Sensor—Surface: 1.4mm (Mechanical Surface Detecti
                                                                        on)
                  *    Electronics: DAE4 Sn1331; Calibrated: 1/19/20
                                                                     17
                  *   Phantom: Triple Flat Phantom 5. 1C; Type: QD 000 P51 CA; Serial:
                                                                                       1 161/1
                  *   Measurement SW: DASY52, Version 52,10 (0); SEMCAD X Version 14.6.10
                      (7417)

              System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measure
                                                                                 ment grid:
              dx=5mm, dy=Smm, dz=5mm
              Reference Value=94.94 V/m; Power Drift=0.01 dB
              Peak SAR (extrapolated) = 19.5 W/kg
              SAR(I g) = 10.1 W/kg: SAR(10 g)=5.19 Wikg
              Maximum value of SAR {measured)= 15.9 Wihkg




                 —3.64


                 .28

                 —10.93

                 1457


                 18.21
                            _4

                           0 dB =15.9 W/kg =12.01 dBW/kg




     Certificate No: Z17—97115                           Page 5 of8


         Add: No,51 Xueyuan Road, Haldian Distri Beijing.    100191, China
        Tel: +86—10—62304633—2079         Faxs +a:'1.|o.eg‘3n&533-zsnl
         E—mail: ent@chinatil.com         hitpi/wanschinattlen




         Impedance Measurement Plot for Head TSL


          Tei. S11 Log nag 10. 00087 ReF o.oo600 [FI]
            so.
                * mi—covovoren cBuar as
            «0.00    —     <=—L_.
            30.00         ———        —
            20.00      —_——         ——
            20.00       ——

            0. 000 B
           —20.00
           ~20,00

           —30.00
           ~40.00

           ~50.00
         IMBRM 511 Sntch (ref) scale 1.000u (F1 be1]
           31. 2.0000000 gkz 31.991 a 6. 5866 o. $51.7,




Certificate No: Z17—97115                               Page 6 of 8


C(
                mm8                       in Coliborationvith
               %‘TTL a
                     CAUBRATION L&gorarory
             Add: No.51 Xucyuan Rond, Haidian District,
             Tel *8(:-|u-623s.|g'3'3_;;7§”wm“    District, Befjing.
                                                              Belji 100191, China
                                                                               $
              Eotany        *                Fax: +86—10—62304633—2504
             E—mail: cftl@chinatl.com        hipifwinkchinati.en

     DASYS Validation Report for Body TSL
     Test Laboratory: CTTL, Beijing,
                                     China
                                                                                    Date: 08.26.2017
     DUT: Dipole 1900 MHz; Type: D1900V2; Serial; D190
                                                       0V2 — SN: 50060
        Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
        Medium parameters used: f= 1900 MHz; a = 1.528 S/m: &, = 53.55; p = 1000
                                                                                 kg/m*
        Phantom section: Center Section
        Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
     DASYS Configuration;

                       Probe: EX3DV4 — SN3617; ConvF(7.95, 7.95, 7.95); Calibrated: 1/23/2017;
                  *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
                  *«   Electronies: DAE4 Sn1331; Calibrated: 1/19/2017
                  *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
                  *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                       (7417)

             System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
             dx=Smm, dy=Smm, dz=5mm
             Reference Value =91.19 V/m; Power Drift= 0.01 dB
             Peak SAR (extrapolated)= 18.1 W/kg
             SAR(I g) = 9.9 W/kg; SAR(1O g) =5.21 W/kg
             Maximum value of SAR (measured)= 15.3 W/kg


               dB
              fl 0

                1 +3.48

                 —6.96

                 —10.44

                 "13.92

                 17.40       A

                            0 dB=15.3 W/kg= 11.85 dBW/kg




     Certificate No: Z17—971 15                               Page 7 of 8


                 TTL e
                     $b —a           CALIERATION LABORATORY
         Add: No.51 Xueyuan Road
                                    ,
         Tel: +86—10—62304633 207 Haidian District, Bcijing, 100191, Chin
                                    8  Fax: +86—10—62304633—2504
         E—mail; ctil@chinattl.co
                                  m    huip:/lwwwehinattl.cn

         Impedance Measurement Plot for
                                        Bod                     y TSL

           Tri i2 Log mag 10, 00de7 nor o.o00
                                              d8 TFZ)
             s0.00 >
                     1.5000000 ane <21.405 de
            40.00     ~——
            30.00        ———
            20.00
            20.00
            ©.coop)
           ~20.00
           ~20.00
           ~30.00
           ~40.00
           ~50,00
          MR si2 snith (rsfx) scale 1.000u [F1 oe1]
            ». 2.9000000 anz s2.660 o. 8.3483 o. 699.39—p




Certificate No: Z17—97115                             Page 8 of 8


                     TTL s_—p_e_a S48 _ sgey,
                                  ®     in Collaborstion with                                 olWn                            fBQA
                                                                                          a           2




             Add: No.5          Road, Haidian     District, Beijing, 100191, China   C%                   §  $   v           gl.litamou
             Teb +be—i0—523dieas—207)       Pax:r86—10—8B01633—.2004 _                4.,/,,/?,,\—"\.\.\\*                   CNAS LOS70
             E—mail: cf@chinatl.com             bups/fwwwchinat.en
      Client             [TA(Shanghai)                                          [Certificate No: Z17—97118" V


Object

Calibration Procedure(s)


Calibration date:

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S!). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed faboratory facility: environment temperature(22s3)C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)

Primary Standards                ID #                    Cal Date(Calibrated by, Certificate No.)                Scheduled Calibration
 Power Meter   NRVD              102083                 22—Sep—16 (CTTL, No.J16X06809)                                 Sep—17
 Power sensor   NRV—Z5           100595                 22—Sep—16 (CTTL, No.J16X06809)                                  Sep—17
 Reference Probe EX3DV4         SN 3617                 23—Jan—17(SPEAG,No.EX3—3617_Jan17)                             Jan—18
 DAE4                           SN 1331                 19—Jan—17(CTTL—SPEAG.No.217—97015)                             Jan—18

 Secondary Standards            ID#                     Cal Date(Calibrated by, Certificate No.)                 Scheduled Calibration
 Signal Generator E4438C        MY49071430              13—Jan—17 (CTTL, No.J17X00286)                                 Jan—18
 Network Analyzer E5071C        MY46110673              13—Jan—17 (CTTL, No.J17X00285)                                 Jan—18



                                                                    Function _ «/.

arated 59.                                                     SAR Test Engineer
                                                               as Teét Eng];Ip e
Reviewed by:



tCt34                           i Dianyuan                     S_Aii—fir;]e-s;t_-uader

                                                                                 Issued: September 1, 2017
This calibration certificate shall not be reproduced except in full without written approval of the laboratary.



    Certificate No: Z17—97116                                  Page 1 of8


o
                      7TL a
                          § °p_e           CALIBRATION LABORATORYy

             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
             E—mail: cttl@chinattl.com       http://wwwchinattl.en


    Glossary:
    TSL                        tissue simulating liquid
    ConvF                       sensitivity in TSL / NORMx,y,z
    N/A                        not applicable or not measured

    Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
      Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
       exposure to radio frequency fields from hand—held and body—mounted wireless
       communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
       6GHz)", July 2016
    c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
       communication devices used in close proximity to the human body (frequency range of
       30MHz to 6GHz)", March 2010
    d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


    Additional Documentation:
    e) DASY4/5 System Handbook

    Methods Applied and Interpretation of Parameters:
    * Measurement Conditions: Further details are available from the Validation Report at the end
       of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    ® Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
       point exactly below the center marking of the flat phantom section, with the arms oriented
       parallel to the body axis.
    * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
       positioned under the liquid filled phantom. The impedance stated is transformed from the
       measurement at the SMA connector to the feed point. The Return Loss ensures low
       reflected power. No uncertainty required.
    e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
       No uncertainty required.
       SAR measured: SAR measured at the stated antenna input power.
       SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
    * SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.


       The reported uncertainty of measurement is stated as the standard uncertainty of
       Measurement multiplied by the coverage factor k=2, which for a normal distribution
       Corresponds to a coverage probability of approximately 95%.



    Certificate No: Z17—97116                             Page 2 of 8


                                        *       in Collaboration with
                          TTL a
                              CAUBRATION LABORATORYy
                  Add: No.S1 X         Rond, Haidian District, Beljing, 100191, Ching
                  Te+86—10—2504003—2070            Faxt+86—10—62304633—.2504
                  E—mail: ettl@chinattl.com        hitp:/vwwchinattlen

          Measurement Conditions
             DASY system configuration, as far as not given on page 1.
               DASY Version                                          DASY52                                     52.10,0,1448

               Extrapolation                                Advanced Extrapolation

               Phantom                                      Triple Flat Phantom 5.1C

               Distance Dipole Center — TSL                          10. mm                                   with Spacer
               Zoom Scan Resolution                             dx, dy, dz =5 mm
               Frequency                                        2450 MHz & 1 MHz


          Head TSL parameters
             The following parameters and calculations were applied.
                                                                      Temperature              Permittivity             Conductivity
               Nominal Head TSL parameters                              22.0 °C                   39.2                  1.80 mho/m

               Measured Head TSL parameters                          (22.0 £0.2) °C            39.7 +6 %            1,82 mho/m £ 6 %

               Head TSL temperature change during test                  <1.0 °C                    —                        =»
          SAR result with Head TSL
             SAR averaged over 1_cm°_(1 g) of Hoad TSL                             Condition
               SAR measured                                                   250 mW input power                  13.2 mW /g
               SAR for nominal Head TSL parameters                             normalized to 1W          §2.6 mW /g x 18.8 % (k=2)

               SAR avoraged over 10 on‘ (10 g) of Head TSL                         Condition
               SAR measured                                                    250 mW input power                 6.16 mW /g

               SAR for nominal Head TSL parameters                             normalized to 1W          24.6 mW /g £ 18.7 % (k=2)
          Body TSL parameters
             Thefollowing parameters and calculations were applied.
                                                                      Temperature              Permittivity             Conductivity

               Nominal Body TSL parameters                               22.0°C                   §2.7                  1.95 mho/m

               Measured Body TSL parameters                          (22.0£0.2) °C              52546 %             1.94 mhoim + 6 %

               Body TSL temperature change during test                   <1.0 °C                   —                        ——
          SAR result with Body TSL
               SAR averaged over 1 _Chr"_(1 g) of Body TSL                         Condition
               SAR measured                                                    250 mW input power                 12.7 mW /g
               SAR for nominal Body TSL parameters                              normalized to 1W         50.8 mW ig & 18.8 % (k=2)

               SAR averaged over 10 cm (10 g) of Body TSL                          Condition
               SAR measured                                                    250 mW input power                 5.87 mW 1 g

               SAR for nominal Body TSL parameters                              normalized to 1W         23.5 mW ig £ 18.7 % (ke2)


          Certificate No: Z17—97116                              Page 3 of 8




rind cepore ortan ze in ropreraacice cneapee ie pany emnimews noe mm oppecin n c n yc es                          u2l       2


           mm*                          in Collaboration with
          &
          T 7/"7‘J,                   s
                                      a e_ a
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86-10-6?304633-2504
        E—mail: cttl@chinattl.com         hup://www.chinatt.en

 Appendix (Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL

     Impedance, transformed to feed point                                   5340+ 4.2910
     Return Loss                                                              —25.5dB


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   51.00+ 6.6110

     Return Loss                                                             ~23.6dB


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                        1.265 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

    [ Manufactured by                                                           SPEAG




Certificate No: Z17—97116                             Page 4 of 8


C(
               !\'                         in Collaboration with

               %TTL _3
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
             E—mail: eti@chinattl.com        hitpo/www.chinatt.en

     DASYS Validation Report for Head TSL                                        Date: 08.29.2017
     Test Laboratory: CTTL, Beijing, China
     DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 786
        Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
        Medium parameters used: {= 2450 MHz; 0 = 1.822 $/m; &r =39.65; p = 1000 kg/m3
        Phantom section: Left Section
       Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
     DASY5 Configuration:

                 *    Probe: EX3DV4 — SN3617; ConvF(7.74, 7.74, 7.74); Calibrated: 1/23/2017;
                 *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
                 *    Electronics: DAE4 Sn1331; Calibrated: 1/19/2017
                 *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
                 *    Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                      (7417)
             Dipole Calibration/Zcom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
             dy=5mm, dz=5mm
             Reference Value = 105.1 V/m; Power Drift=—0.03 dB
             Peak SAR (extrapolated) =27.5 W/kg
             SAR(I g)=13.2 W/kg; SAR(10 g)=6.16 W/kg
             Maximum value of SAR (measured) =22.2 W/kg


                dB
                 0

                 ~4.33

                 —8.66

                 ~12.98

                 17.31


                 21.64
                              £—
                           0 dB =22.2 W/kg=13.46 dBW/kg




     Certificate No: Z17—97116                            Page 5 of8


Cz
                                     *         In Collaboration with


                                               CALIBRATION LABORArory
             Add: No.S1 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2079           Fax: +86—10—62304633—2504
             E—mail; ett!@chinatil.com           hiip//wwwchinatt.en




             Impedance Measurement Plot for Head TSL


              Tri 311 Log wag 10.00007 we 0.000d8 [s1]
                50. 00 on
                       TRA B9

                40.00
                30.00
                20.00
                10.00
                6.co0                                                            a
               ~20.00
               ~20.00
               =30. 60
               ~40.00
               50. on
             PIIB si1 saith (@+]x) scale 1.000u fFi. oel]
               31.   2,4500000 anz       $3.436 o.   4.2806 o.   270. $§%.9r




     Certificate No: Z17—97116                                   Page 6 of 8


C(

               &TTL £‘6eca
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2079     Fax: +xn—|u—fi_zm 633—2504
             E—mail: etti@chinattl.com       hitp:/fwwchinatthen

     DASY5 Validation Report for Body TSL                                        Date: 08.29.2017
     Test Laboratory: CTTL, Beijing, China
     DUT: Dipole 2450 MHz; Type: D2450V2; Serial; D2450V2 — SN: 786
       Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
       Medium parameters used: {= 2450 MHz; a= 1.943 S/m; s, = 52.45; p = 1000 kg/m‘
       Phantom section: Right Section
       Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
     DASY5 Configuration:

                 *    Probe: EX3DV4 — SN3617; ConvF(7.8, 7.8, 7.8); Calibrated: 1/23/2017;
                 +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
                 *    Electronics; DAE4 Sn1331; Calibrated: 1/19/2017
                 *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
                 *«   Measurement SW: DASY52, Version 52.10 (0); SEMCAD X Version 14.6.10
                      (7417)

             Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
             dy=5mm, dz=5mm
             Reference Value = 92.28 V/m; Power Drift=—0.01 dB
             Peak SAR (extrapolated) = 27.0 W/kg
             SAR(I g) = 12.7 W/kg; SAR(10 g) = 5.87 W/kg
             Maximum value of SAR (measured) =21.5 W/kg




                 ~4.38

                 —8.75


                 13.13

                 ~17.50


                 ~21.88

                            0 dB =21,5 W/kg =13.32 dBW/kg




     Certificate No: Z17—97116                            Page 7 of 8



Document Created: 2017-12-01 13:09:46
Document Modified: 2017-12-01 13:09:46

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC