SAR report part 3

FCC ID: 2AFZZ-RMSD1SG

RF Exposure Info

Download: PDF
FCCID_3856886

Calibration Laboratory of                                                      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                                   Service suisse d'etalonnage
                                                                               C   Servizio svizzero di taratura
 Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland                                   s   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.:   SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528-2013, "IEEE Recommended Practice for Determining the Peak Spatial-
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209-2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   c) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
  d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
  • Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
  • Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
  • Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required .
   • Electrical Delay: One-way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   • SAR measured: SAR measured at the stated antenna input power.
   • SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
   • SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D5GHzV2-1006_Sep16                              Page 2 of 12


Measurement Conditions
   DASY system confiQuration, as far as not c iven on page 1.
     DASY Version                                               DASY5                                    V52.8.8

     Extrapolation                                    Advanced Extrapolation

     Phantom                                        Modular Flat Phantom V5.0

     Distance Dipole Center - TSL                               10mm                                   with Spacer

     Zoom Scan Resolution                         dx, dy = 4.0 mm, dz = 1.4 mm             Graded Ratio = 1.4 (Z direction)
                                                        5250 MHz ± 1 MHz
     Frequency                                          5600 MHz± 1 MHz
                                                        5750 MHz + 1 MHz



Head TSL parameters at 5250 MHz
   Thtll   · parameters and ca11
     e o owing                     ·
                              cu at1ons were app l"d
                                                  1e
                                                                Temperature            Permittivity         Conductivity

     Nominal Head TSL parameters                                   22.0 °C                35.9               4.71 mho/m

     Measured Head TSL parameters                               (22.0 ± 0.2)   0
                                                                                   c   34.5 ±6 %          4.59 mho/m ± 6 %

     Head TSL temperature change during test                       < 0.5 °C                ----                    ----


SAR result with Head TSL at 5250 MHz
                              3
     SAR averaged over 1 cm (1 g) of Head TSL                    Condition

     SAR measured                                         100 mW input power                          8.13 W/kg

     SAR for nominal Head TSL parameters                      normalized to 1W             80.6 W/kg :t 19.9 % (k=2)


     SAR averaged over 10 cm3 (10 g) of Head TSL                  condition

     SAR measured                                         100 mW input power                          2.32 W/kg

     SAR for nominal Head TSL parameters                      normalized to 1W             22.9 W/kg ± 19.5 % (k=2)



Head TSL parameters at 5600 MHz
   The fo IIow1nq
              . parameters an d caIcu Iat1ons
                                         .    were appI"1ed
                                                                Temperature            Permittivity         Conductivity

     Nominal Head TSL parameters                                   22.0 °C                 35.5              5.07 mho/m

     Measured Head TSL parameters                               (22.0 ± 0.2) °C        34.0 ± 6 %         4.93 mho/m ± 6 %

     Head TSL temperature change during test                       < 0.5 °C                ----                    ----

SAR result with Head TSL at 5600 MHz

     SAR averaged over 1 cm3 (1 g) of Head TSL                   Condition

     SAR measured                                         100 mW input power                          8.47 W/kg

     SAR for nominal Head TSL parameters                      normalized to 1W             83.8 W I kg ± 19.9 % (k=2)


     SAR averaged over 1O cm3 (1 O g) of Head TSL                 condition

     SAR measured                                         100 mW input power                          2.40 W/kg

     SAR for nominal Head TSL parameters                      normalized to 1W              23.7 W/kg ± 19.5 % (k=2)



Certificate No: D5GHzV2-1006_Sep16                     Page 3 of 12


Head TSL parameters at 5750 MHz
   The following parameters and calculations were applied
                                                              Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                 22.0 °C           35.4               5.22 mho/m

     Measured Head TSL parameters                             (22.0 ± 0.2) °C   33.8 ±6 %          5.08 mho/m ± 6 %

     Head TSL temperature change during test                     < 0.5 °C           ----                   ----

SAR result with Head TSL at 5750 MHz

     SAR averaged over 1 cm3 (1 g) of Head TSL                 Condition
     SAR measured                                       100 mW input power                     8.14 W/kg
     SAR for nominal Head TSL parameters                    normalized to 1W        80.5 W/kg ± 19.9 % (k=2)

                                3
     SAR averaged over 10 cm (10 g) of Head TSL                condition
     SAR measured                                       100 mW input power                     2.30 W/kg

     SAR for nominal Head TSL parameters                    normalized to 1W        22.7 W/kg ± 19.5 % (k=2)




Body TSL parameters at 5250 MHz
   The following parameters and calculations were applied
                                                              Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                 22.0 °C           48.9               5.36 mho/m

     Measured Body TSL parameters                             (22.0 ± 0.2) °C   47.4±6%            5.52 mho/m ± 6 %

     Body TSL temperature change during test                     < 0.5 °C           ----                   ----


SAR result with Body TSL at 5250 MHz

     SAR averaged over 1 cm3 (1 g) of Body TSL                 Condition

     SAR measured                                       100 mW input power                     7.59 W/kg

     SAR for nominal Body TSL parameters                    normalized to 1W        75.5 W/kg ± 19.9 % (k=2)

                                3
     SAR averaged over 10 cm (10 g) of Body TSL                 condition

      SAR measured                                      100 mW input power                     2.13 W/kg

      SAR for nominal Body TSL parameters                   normalized to 1W         21.2 W/kg ± 19.5 % (k=2)




Certificate No: 0 5GHzV2-1006_Sep16                  Page 4 of 12


Body TSL parameters at 5600 MHz
   The foIIow1nq
             . parameters and caIculations were applied.
                                                              Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                 22.0 °c           48.5               5.77 mho/m

     Measured Body TSL parameters                             (22.0 ± 0.2) °C   46.8 ±6 %          6.00 mho/m ± 6 %

     Body TSL temperature change during test                     < 0.5 °C           ----                   ----

SAR result with Body TSL at 5600 MHz
                              3
     SAR averaged over 1 cm {1 g) of Body TSL                  Condition

     SAR measured                                       100 mW input power                     7.90 W/kg

     SAR for nominal Body TSL parameters                    normalized to 1W        78.6 W/kg ± 19.9 % (k=2)


     SAR averaged over 1O cm3 (1 O g) of Body TSL               condition

     SAR measured                                       100 mW input power                     2.22 W/kg

     SAR for nominal Body TSL parameters                    normalized to 1W         22.0 W/kg ± 19.5 % (k=2)


Body TSL parameters at 5750 MHz
   The following parameters and calculations were applied
                                                              Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                 22.0 °C            48.3              5.94 mho/m

     Measured Body TSL parameters                             (22.0 ± 0.2) °C    46.5 ±6 %         6.21 mho/m ± 6 %

     Body TSL temperature change during test                     < 0.5 °C           ----                   ----

SAR result with Body TSL at 5750 MHz
                              3
     SAR averaged over 1 cm (1 g) of Body TSL                   Condition

     SAR measured                                       100 mW input power                     7.50 W/kg

     SAR for nominal Body TSL parameters                    normalized to 1W         74.6 W/kg ± 19.9 % (k=2)

                                  3
     SAR averaged over 10 cm (10 g) of Body TSL                 condition

     SAR measured                                       100 mW input power                     2.10W/kg

     SAR for nominal Body TSL parameters                    normalized to 1W         20.8 W/kg ± 19.5 % (k=2)




Certificate No: D5GHzV2-1006_Sep16                   Page 5 of 12


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL at 5250 MHz

     Impedance, transformed to feed point                                           55.6 Q - 6.5 jQ

     Return Loss                                                                       - 21.8 dB


Antenna Parameters with Head TSL at 5600 MHz

     Impedance, transformed to feed point                                           57.5 Q - 4.7 jQ

     Return Loss                                                                       -21 .7 dB


Antenna Parameters with Head TSL at 5750 MHz

     Impedance, transformed to feed point                                           59.2 Q + 5.8 jQ

     Return Loss                                                                       - 20.1 dB


Antenna Parameters with Body TSL at 5250 MHz

     Impedance, transformed to feed point                                           55.5 Q - 3.4 jQ

      Return Loss                                                                      - 24.2 dB


Antenna Parameters with Body TSL at 5600 MHz

      Impedance, transformed to feed point                                           59.3 Q -1 .9 jQ

      Return Loss                                                                      - 21.2 dB


Antenna Parameters with Body TSL at 5750 MHz

      Impedance, transformed to feed point                                           58.8 Q + 8.9 jQ

      Return Loss                                                                      - 18.8 dB



General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.201 ns

                                                                                                     can be measured.
After long term use with 1OOW radiated power, only a slight warming of the dipole near the feedpoint

                                                                                                 is directly connected to the
The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line
                                                                                               the dipoles, small end caps
second arm of the dipole. The antenna is therefore short-circuited for DC-signals . On some of
                                                                                                  as explained in the
are added to the dipole arms in order to improve matching when loaded according to the position
                                                                                                   dipole length is still
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall
according to the Standard.
                                                                                                  connections near the
No excessive force must be applied to the dipole arms, because they might bend or the soldered
feedpoint may be damaged.

 Additional EUT Data

      Manufactured by                                                                    SPEAG

      Manufactured on                                                                August28 ,2003



 Certificate No: D5GHzV2-1006_Sep16                      Page 6 of 12


DASYS Validation Report for Head TSL
                                                                                       Date: 27.09.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole D5GHzV2; Type: D5GHzV2; Serial: D5GHzV2- SN:1006
Communication System: UID O - CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
                                                                               3
Medium parameters used: f = 5250 MHz; cr = 4.59 Sim; l':r = 34.5; p = 1000 kg/m
                                                                               3
Medium parameters used: f = 5600 MHz; cr = 4.93 Sim; Er = 34.0; p = 1000 kglm
                                                                               3
Medium parameters used: f = 5750 MHz; cr = 5.08 Sim; Er= 33.8; p = 1000 kglm
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEEIIECIANSI C63.19-201 l)
DASY52 Configuration:

   •   Probe: EX3DV4 - SN3503; ConvF(5.42, 5.42, 5.42); Calibrated: 30.06.2016, ConvF(4.89, 4 .89,
       4.89); Calibrated: 30.06.2016, ConvF(4.85, 4.85, 4.85); Calibrated: 30.06.2016;

   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

   •   Electronics: DAE4 Sn601; Calibrated: 30. 12.2015

   •   Phantom: Flat Phantom 5.0 (front); Type: QD000P50AA; Serial: 1001

   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Head Tissue/Pin=lOOmW, dist=lOmm, f=5250 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=l.4mrn
Reference Value = 72.67 Vim; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 29.4 W/kg
SAR(l g) =8.13 W/kg; SAR(lO g) =2.32 W/kg
Maximum value of SAR (measured) = 18.2 Wlkg

Dipole Calibration for Head Tissue/Pin=lOOmW, dist=lOmm, f=5600 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz= 1.4mm
Reference Value = 72.60 Vim; Power Drift = 0.04 dB
Peak SAR (extrapolated)= 33.1 W/kg
SAR(l g) =8.47 W/kg; SAR(lO g) =2.4 W/kg
Maximum value of SAR (measured)= 19.7 W/kg

Dipole Calibration for Head Tissue/Pin=lOOmW, dist=lOmm, f=5750 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=l .4mm
Reference Value = 70.56 Vim; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 33.0 W/kg
SAR(l g) =8.14 W/kg; SAR(lO g) =2.3 W/kg
Maximum value of SAR (measured) = 19.2 Wlkg




Certificate No: D5GHzV2-1006_Sep16             Page 7 of 12


      dB
      0

      -6.00


       -12.00

       -18.00

       -24.00


       -30.00
                 0 dB = 19.2 W/kg = 12.83 dBW/kg




Certificate No: D5GHzV2-1006_Sep16           Page 8 of 12


Impedance Measurement Plot for Head TSL




                                                                              27 Sep 2016  13:52:19
                 $11     1 U Fs                4: 59.648 a       —6.4844 n 4.6751 pF      5 250.000 a@o MHz
                                                               ,     T———
                                                        Fa _         (     >
                                                                          po       _              ,
          x                                                                    —                 «C
                                                                                                               &                                          CHL Markers
          Del                                                                            mz                    W\
                                                                                             4
                                                                                                                    *,
          Cor                                                         ~                                    ~oA

                                               o_ _—iz~—                                  1#+8+—}
                                               e
                                               \
                                                                                       ~900           t         "q
                                                                                                                            03
                                               \\
          fivg                                      \                                  L db           ue            5
          16                                                                                 =—            —

                                                                          7 —
          H1a                                             s
          CH2 Si1        _LOG______s dB/REF —20 8                 _                    _ 11—21.5909 dB                           5 250.000 900 MHz
                                                           |          1                                                                         1
                 obfnfi                                                                                                           cole          3177—— |

                                                                                                                                         |     $          CH2 Markers
                 |—                                                                —                                                           +}         2—21.707 0B
          tfor   |L                |                      |           |                 <t                                           —         |
                                                                                                                                               +—         S.60000 onz
                                   .   ‘j__'                                                                                                    |         3:—20.051 dB
                                       &                                           C                                             E_            154        5.75200 GHz
                 |—                                                                      5                                       &

          128                                  ‘v‘\\ ’/\//\/                                                                                 /}‘/

          HLd    |——                            f       \l                                                              ‘
                                                          |
                       SthRT 5 dod.a00 ado MHz




Certificate No: D5GHzV2—1006_Sep16                               Page 9 of 12


DASY5 Validation Report for Body TSL
                                                                                       Date: 26.09.2016

Test Laboratory: SPEAG, Zurich, Switzerland

OUT: Dipole D5GHzV2; Type: D5GHzV2; Serial: D5GHzV2 - SN:1006
Communication System: UID O - CW; Frequency: 5250 MHz, Frequency: 5600 MHz, Frequency: 5750 MHz
                                                                             3
Medium parameters used: f = 5250 MHz; cr = 5.52 Sim; Cr= 47.4; p = 1000 kg/m
                                                                             3
Medium parameters used: f = 5600 MHz; CJ = 6.00 Sim; gr = 46.8; p = 1000 kg/m
                                                                             3
Medium parameters used: f = 5750 MHz; cr = 6.21 Sim; gr = 46.5; p = 1000 kg/m
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY52 Configuration:
   •   Probe: EX3DV4 - SN3503; ConvF(4.85, 4.85, 4 .85); Calibrated: 30.06.2016, ConvF(4.35, 4.35,
       4.35); Calibrated: 30.06.2016, ConvF(4.3, 4.3, 4 .3); Calibrated: 30.06.2016;

   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

   •   Electronics: DAE4 Sn601; Calibrated: 30.12.2015

   •   Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Body Tissue/Pin=lOOmW, dist=lOmm, f=5250 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mrn, dz= l.4mm
Reference Value = 66.79 V/m; Power Drift = -0.07 dB
Peak SAR (extrapolated) = 28.8 W/kg
SAR(l g) =7.59 W/kg; SAR(lO g) =2.13 W/kg
Maximum value of SAR (measured)= 17.9 W/kg

Dipole Calibration for Body Tissue/Pin=lOOmW, dist=lOmm, f=5600 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz= l.4mm
Reference Value = 67.00 V/m; Power Drift = -0.07 dB
Peak SAR (extrapolated) = 32.5 W/kg
SAR(l g) =7.9 W/kg; SAR(lO g) =2.22 W/kg
Maximum value of SAR (measured) = 19 .0 W/kg

Dipole Calibration for Body Tissue/Pin=lOOmW, dist=lOmm, f=5750 MHz/Zoom Scan,
dist=l.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz= 1.4mm
Reference Value = 64.64 V/m; Power Drift= -0.03 dB
Peak SAR (extrapolated) = 32.2 W/kg
SAR(l g) = 7.5 W/kg; SAR(lO g) = 2.1 W/kg
Maximum value of SAR (measured) = 18.5 W/kg




Certificate No: D5GHzV2-1006_Sep16            Page 1O of 12


      dB
       0

      -6.00

      -12.00


      -18.00

      -24.00


       -30.00
                 0 dB = 18.5 W/kg = 12.67 dBW/kg




Certificate No: 05GHzV2-1006_Sep16          Page 11 of 12


Impedan ce Measure ment Plot for Body TSL




                                                                       25 Sep 2016       07:46:17
         [HI] S1.1     1 U FS           1: 55.521 ll     - 3.4180 r.: 8.8594 pF         5 250.000 000 MHz



         *                                                                                                   CH1 Markers
         Del                                                                                                  2: 59.314 s,
                                                                                                               -1.8555 !::
                                                                                                             5.50000 GHz
         Cor
                                                                                                              3: 58.801 !::
                                                                                                                 8 . 8906 •l
                                                                                                             5.75000 GHz


         Avg
         15


         Hld

          CH2   S11     LQ_G     _ 5 dB i filF - 20 dB                 1:-24.217 dB     5 ~ 50,000_000 MHz



                                                                                                             CH2 Markers

                                                                                                             2:- 2 1.2 21 dB
          Cor                                                                                                5 . 50000 GHz
                                                                                                             3:-1 8 . 8 15 dB
                                                                                                             5 . 750 00 GHz


          Avg
          15
                                 ....__- - -~ -
          Hld                                      +          +-       -          +

                      START 5 000. 000 000 MHz                                    STOP 6 000.000 000 MHz




 Certificate No: D5GHzV2-1006_Sep16                         Page 12 of 12


Phons sa1 1248 9700,Fan «d
inotspaay com, itspeon com



                                  IMPORTANT NOTICE


USAGE OF THE DAE4
The DAE unt is a delcate, high procision insrument and requires carefi treatment by the user. There are no
serviceable part inside the DAE. Specil attention shallbe given to he folowng ponts
Battery Exchange: The batley cover af the DAEA uniis lased using a serew, over lightening the serow may
cause the threads insde the DAE to wear out
Shipping of the DAE: Refore shipping the DAE to SPEAG for calbration, remove the batteris and pack the
DAE in an anitatie bag, This anistale bag shall en be packed nto larger box orcontainer which protects the
DAE from impacts during transporation. The package shallbe marked to indeate that a fragie instrument is
inside
EStop Fallures: Touch detecion may be malfinclioning due to broken magnels in e Estop. Rough handing
of the Etop may lead to damage of these magnels. Touch and calsionerars are often caused by dust and dit
accumulated in the Estop, To prevent Estop falure, the customer shall alvays maunt the probe to the DAE
earefily and keepthe DAE unt n a nor—dusty environimenti not used for measurements
Repalr: Minorrepats are performad at no extra cost duting the annval calbration, However, SPEAG reserves
the right to chargefor any repair especialy i rough unprofessional handing caused the defect
DASY Confiquration Files: Since the exact values of the DAE input resitances, as measured during the
eallration procedure ofa DAE un‘t, are not used by the DASY sofware, a norinal value o200 MOhm is ghen
in the corresponding conturaton fle:
important Note:
Warranty and calibration is void if the DAE unitis disassembled partly or fully by the
(Customer.

Important No
Never attompt to grease or ail the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by cortified SPEAG personnel only and is part of the annual
calibration procedure.


Important Not
[To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
\while turning the locking nut of the connector. The same care shall be used when
(disconnecting the probe from the DAE.


Scimid & Parner Engincemng
TNBRor0315AD DAES doe                                                                         11.12.2009


      r
Calibration Laboratory of                         «>                              SciweisrischrKaltrelonnt
Schmid & Partner                                  %                             z Servicesuisse tétstonnage
  Engineering AG                                  1oets                             Servzto avizero ttrntun
‘ghausstrase 4, 0004 zurich, Suttsatand           *fams                         . swis Caltration Serice

Accuiedby o Svis AccuctaionSevics (SAG)                                  reeredtston: SCS 0108
‘Tre Suiss AceredttionSrvieis oneofthesignstois ts the EA
HtatiterstAgreomentor threcogniton ofcalbratonconifetes
ce Sporton (Auden)                                                       Cetticas io: DAE4—1279_Jan18
CALIBRATION CERTIFICATE
 Oe                           DAE4 — SD 000 Do# BM — SN: 1279

 Cattaion pocedin(s           OA CAL—06.v20
                              Calibration procedure for the data acquisiion electronics (DAE)


 Catbaton ts                 January 03, 2018



 Thcalbaton corfeat decnats h ricnaviy o natena stardards,utich roatc h hyicaunte o messirenens S
 ‘The mossirements anh ureaininis wih conidanc obubityangiven on h otowingpages and e patofthe cotfeate
 Alcalbatons have boon condeto n thecosed aboratoylacieniorment temportiee(2 + 2and ramity «701
 CaltraionEnpmentused (UATE cntea for catbadon)
 Piman Stnduts                 o«                 calDate(contcwNo)                    scteded Catbraton
 Kathny MutinaarTypu 2001      it oiere           Sr¥eit Noz                           Aupis
 Sncondar Stamdarts            +                 tect Dare intouse)                    Scteded Creck
 Aco DNE Caitaten im           se is ons MA 1001.Osn17 in house crec                   in rousechesk sinto
 cittrBorvet                   se uis 0s M¥ 1002 Osvint7 n hoosectuct)                 Inhouse cteck ints




                               Name                         Functon                     Sqoauee
 Cattredby                     Atan Gaing                   LateratoyTechicin          A 77

 Appromby                      Snkim                        Dovuty mavage
                                                                                        i ,l/.(lgj UMMV'
                                                                                       issuet Jardarya, 2018
Ti cattatoncotfeatshal nebe mprdized ces n ul wthot wetenaeprovl t e latortary


Corticato No: DAE#—1270_Jant                        Page 1ot5


Calibration Laboratory of                                                 Schwoizerischor Kliriertanst
Schmid & Partner                                                        g Servien uisneialonnage
  Engineering AG                                                          Sorvito svizoro ditarurn
Hghousstraise i, 004 Zurich, Switzetand                                 S.—   swiss Calbration Sonvice

Accrudto by h SnisAccediaton Serice (GAS)                                Aestedtaton No: SCS 0108
The Suiss Aceredtation Service is oneofthesignntris to tEA
MulatoralAgroomentforthe recognitonofcalration corifentos

Glossary
Dae                     data acquisition electronics
Connector angle         information used in DASY system to align probe sensor X to the robot
                        coordinate system.
Methods Applied and Interpretation of Parameters
  * DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
        comparison with a calibrated instrument traceable to national standards. The figure given
        corresponds to the full scale range of the voltmeter in the respective range.
   * Gonnector angle: The angle of the connector is assessed measuring the angle
     mechanically by a tool inserted. Uncertainty is not required.
   *   The following parameters as documented in the Appendix contain technical information as a
       resultfrom the performance test and require no uncertainty.
       *    DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
            the nominal callbration voltage. Influence of offset voltageis included in this
            measurement.
       *    Common mode sensitivity: Influence of a positive or negative common modevoltage on
            the differential measurement.
       *    Channel separation:Influence of a voltage on the neighbor channels not subject to an
            input voltage.
        * AD Converter Values with inputs shorted: Values on the internal AD converter
          corresponding to zero input voltage
        * Input Offset Measurement: Output veltage and statistical results over a large number of
            zero voltage measurements.
        *   Input Offset Current: Typical value for information; Maximum channel input offset
            current, not considering the input resistance.
        *Input resistance: Typical value for information: DAE input resistance at the connector,
         during internal auto—zeroing and during measurement,
       * Low Battery Alarm Voltage: Typical valuefor information. Below this voltage, a battery
         alarm signal is generated.
       + Power consumption: Typical value for information. Supply currents in various operating
            modes.




Coriicate No: DAEA—1270_Janta                        Page 2015


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          wse»       ety ,         fullrange = ~100.,4300 mV
       Low Range:           ise        stov .        fullrange 1......c3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring tme: 3 seo
     Callbration Factors            x                       v                     z
     High Range            403.996 + 0.02% (k—2) 4nson+ 0.00% (k=2) 404.618 2 0.02% (ke2)
     Low Range             a.o4027 + 1.50% (ke2) a.000t0 + 1.500%(k=2)   a.ceoon + 1.50% (ce2)

Connector Angle
     Gonnector Angle to be used in DASY system                               ssroee1®




Cerifeate No: DAE4—1270.Jant                     Page3or$


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
     High Rango                               Reading (iV)       |_Ditterence (1¥)        Eiror (%)
     Ghannel x      + Input                     ressor64                10e                aco
     Channel X    + input                       2000480                280                 oo1
     Ghannel X____— input                       —rs09870               207                Eo
     Channel ¥    + input                       reosor3                 17e                oco
     Channel ¥     _ _+ input                   2000336                 rar               oo
     Ghannal Y     _ _— input                  —eoooe.7e               ce                  oo1
     Channel 2    + input                      19o007.co                186                oco
     Channel 2     _ _+ input                   20001.30               osr                —o00
     Channel 2     _ _— input                  2000042                 235                 oo
     Low Range                                Reaing (2V)        __|_Difference (iY)      Eror ()
     Ghannel X___+ input                        2ocnas                  ase                oos
     Ghannel X    + input                        z00e                    ts                asa
     Channel X____— input                       —r9082                   rar              se
     Channel Y    + input                       2oo1.79                 ogo                oce
     Channel ¥    + input                        zo18e                  oce                oo«
     Channel Y     _ — input                    19807                   ao1               00
     Ghannel 2    + input                       200213                  ase               oo3
     Channel 2    + input                        20117                  ws1
     Channel 2    > input                       1008                    omn               o«

2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 see; Measuring time: 3 seo
                    Common mode                     High Range                   Low Range
                    Input Vottage (m¥)        Average Reading (1¥)           Average Reading (1¥)
    Ghannel X                    200                   as51                       848
                                —200                   zes                         zore
    Channel ¥                    2o                     sao                         sar
                                —200                    se                         a15
    Channel2                     200                    osr                        c
                                200                     z90                        zse

3. Channel separation
   DASY measurement parameters: Auto Zere Time: 3 see; Messiring tme: see
                      Input Voltage (m¥)__|      Channe!X (@¥)     ChannelY (i¥)       ChannelZ (@¥)
    Channel xX                200                      —               se                  385
    Channel ¥                 200                    asr                                   4so
    Channelz                  200                    103                  sso                —

Cerifcate No: DAE#—1270_Jante                           Pagedot5


4. AD—ConverterValues with inputs shorted
   DASY mensurement parameters: Auto Zero Time: 3 see; Measuring tme: 3 sec
                                             High Range (L98)             Low Range (L98)
    Channel X                                       1so00                         18110
    Channel ¥                                       1sose                         17200
    Channel z.                                      1sros                         isriz

5. Input Offset Measurement
   DASY measurement parameters: Asto Zoro Time: 3 sec; Measuring time: 3 seo
   inour ton
                         Average (i¥)      min. Offset (i¥) max. offset yy)       * Z"“;;'“""
    Channol X                105                oz               2s                 oae
    Channal ¥                028                use              oss                o«2
    Channel 2               s08                 —1ss             205                asr

6. Input Offset Current
    Nominalinputcroutry ffset current on alchannels: «251A

7. Input Resistance (Tpicalvalues fo infomation)
                                            Zeroing (kOhm)                Measuring (Whm)
    Channel xX                                    2o                             2o
    Channol‘Y                                     200                            2o
    Channel2                                      200                            200

8. Low Battery Alarm Voltage(pica! values for information)
    Typical values                           Alarm Lovel (VDG)
    Supply (+ Vee)                                                   «9
    Supply (—Vee)                                                    2s

9. Power Consumption (Typicalvalues for nformaton)
    Typical values                      Switehed off (ma) Stand by (m)         |_Transmiting (ma)
    Supply (+ Veo)                           w00               «s                      w
    Supply (—Vee)                            —oo1                s                     s




Gorlicate No: DAEH—1270_Jantg                       PageSols


Calibration Laboratory of                                          S7                                           S   Schweizerischer Kalibrierdienst




                                                                                     8
                                                                 )§
Schmid & Partner                                                  M                                             C   Service suisse d‘étalonnage
  Engineering AG                                                                                                    Servizio svizzero di taratura
                                                                                                                S




                                                                 wl



                                                                                      II/’I
Zeughausstrasse 43, 8004 Zurich, Switzerland                        AZ       r~       NP                            Swiss Calibration Service
                                                                         //"Illlll\‘\\\


Accredited by the Swiss Accreditation Service (SAS)                                                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      Sporton (Auden)                                                                          Certificate No: EX3-3954_Jan 18



CALIBRATION CERTIFICATE
Object                                EX3DV4 — SN:3954


Calibration procedure(s)              QA CAL—01.v9, QA CAL—14.v4, QA CAL—23.v5, QA CAL—25.v6
                                      Calibration procedure for dosimetric E—field probes



Calibration date:                     January 31, 2018

This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70% .


Calibration Equipment used (M&TE critical for calibration)



 Primary Standards                   ID                                     Cal Date (Certificate No.)                Scheduled Calibration
 Power meter NRP                     SN: 104778                             04—Apr—17 (No. 217—02521/02522)           Apr—18
 Power sensor NRP—291                SN: 103244                             04—Apr—17 (No. 217—02521)                 Apr—18
 Power sensor NRP—291                SN: 103245                             04—Apr—17 (No. 217—02525)                 Apr—18
 Reference 20 dB Attenuator          SN: S5277 (20%)                        07—Apr—17 (No. 217—02528)                 Apr—18
 Reference Probe ES3DV2              SN: 3013                               30—Dec—17 (No. ES3—3013_Dec17)            Dec—18
 DAE4                                SN: 660                                21—Dec—17 (No. DAE4—660_Dec17)            Dec—18

 Secondary Standards                 ID                                     Check Date (in house)                     Scheduled Check
 Power meter E4419B                  SN: GB41293874                         06—Apr—16 (in house check Jun—16)         In house check: Jun—18
 Power sensor E4412A                 SN: MY41498087                         06—Apr—16 (in house check Jun—16)         In house check: Jun—18
 Power sensor E4412A                 SN: 000110210                          06—Apr—16 (in house check Jun—16)         In house check: Jun—18
 RF generator HP 8648C               SN: US3642U01700                       04—Aug—99 (in house check Jun—16)         In house check: Jun—18
 Network Analyzer HP 8753E           SN: US37390585                         18—Oct—01 (in house check Oct—17)         In house check: Oct—18

                                          Name                                        Function                          Signature
Calibrated by:                         Leif Klysner                                   Laboratory Technician
                                                                                                                     SfTp
Approved by:                           Katja Pokovic                                  Technical Manager
                                                                                                                    NedPX
                                                                                                                        Issued: February 1, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: EX3—3954_Jan18                                           Page 1 of 11


Cahbratlon Laboratory of                                     \\\“\‘y’fl’/@                §    Schweizerischer Kalibrierdienst
Schmid & Partner                                            SE\\\\:&EI
                                                                     IE//H":             _    Service suisse d‘étalonnage
  Engineering AG                                            Tz yy                        5    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 {’/"/,fi\\\‘\\}                   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                       Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
NORMx,y,z                    sensitivity in free space
ConvrF                       sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A, B, C, D                   modulation dependent linearization parameters
Polarization                  rotation around probe axis
Polarization 8               8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                             i.e., 8 = 0 is normal to probe axis
Connector Angle              information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
    a)   IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
         Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
         Techniques", June 2013
    b)   IEC 62209—1, ", "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from hand—
         held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)", July 2016
    c)   IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
         used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March 2010
    d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Methods Applied and Interpretation of Parameters:
         NORMx,y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; f > 1800 MHz: R22 waveguide).
         NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E—field
         uncertainty inside TSL (see below ConvF).
         NORM(F)x,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in the stated uncertainty of ConvF.
         DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
         PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
         characteristics
         Ax,y,z; Bx,y,z; Cx,y,z; Dx,y,z; VRx,y,z: A, B, C, D are numerical linearization parameters assessed based on
         the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
         media. VR is the maximum calibration range expressed in RMS voltage across the diode.
         ConvrF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
         Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
         to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
         MHz.
         Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
         Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.
         Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
         uncertainty required).


Certificate No: EX3—3954_Jan18                                  Page 2 of 11


EX3DV4 — SN:3954                                                                January 31, 2018




                      Probe EX3DV4

                                          SN:3954

                                  Manufactured:            August 6, 2013
                                  Calibrated:              January 31, 2018



                                 Calibrated for DASY/EASY Systems
                                    (Note: non—compatible with DASY2 system!)




Certificate No: EX3—3954_Jan18                    Page 3 of 11


EX3DV4— SN:3954                                                                                                              January 31, 2018




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954
Basic Calibration Parameters
                                                Sensor X                         Sensor Y                    Sensor Z              Unc (k=2)
 Norm (uV/(V/m))*                                  0.54                            0.44                         0.53               +10.1 %
 DCP (mVv)"                                        99.0                           101.1                         97.0

Modulation Calibration Parameters
 UID           Communication System Name                                    A             B          C          D          VR          Unc®
                                                                           dB        dBvuV                     dB         mV           (k=2)
 0             CW                                                 xX       0.0        0.0           1.0       0.00       149.4        +3.0 %
                                                                  Y        0.0        0.0           1.0                  139.2
                                                                  Z        0.0        0.0           1.0                  146.5


 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


* The uncertainties of Norm X,Y,Z do not affect the E*—field uncertainty inside TSL (see Pages 5 and 6).
® Numerical linearization parameter: uncertainty not required.
& Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: EX3—3954_Jan18                                     Page 4 of 11


EX3DV4— SN:3954                                                                                                             January 31, 2018




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954


Calibration Parameter Determined in Head Tissue Simulating Media
                        Relative         Conductivity                                                               Depth °          Unce
     f (MHz)°        |_Permittivity"        (S/m)"          ConvF X       ConvF Y        ConvFZ        Alpha®        _(mm)          (k=2)
        750               41.9                0.89            10.72         10.72         10.72         0.29         111          +12.0 %
        835               41.5                0.90            10.20         10.20         10.20         0.43         0.80         +12.0 %
        1750              40.1                1.37            8.65           8.65          8.65         0.29         0.85         +12.0 %
        1900              40.0                1.40            8.41           8.41          8.41         0.21         0.99         +12.0 %
       2000               40.0                1.40            8.33           8.33          8.33         0.30         0.84         +12.0 %
       2300               39.5                1.67            7.89           7.89          7.89         0.31         0.84         +12.0 %
       2450               39.2                1.80            7.49           7.49          7.49         0.35         0.84         +12.0 %
       2600               39.0                1.96            7.31           7.31          7.31         0.24         1.08         + 12.0 %

       5250               35.9               4.71             5.20           5.20          5.20         0.30         1.80         %13.1 %
       5600               35.6               5.07             4.59           4.59          4.59         0.40         1.80         +13.1 %
       5750               35.4               5.22             4.74           4.74          4.74         0.40         1.80         #13.1 %


€ Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and a) can be relaxed to + 10%if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3954_Jan18                                   Page 5 of 11


EX3DV4— SN:3954                                                                                                             January 31, 2018




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Calibration Parameter Determined in Body Tissue Simulating Media
                        Relative         Conductivity                                                              Depth °           Unc
     f (MHz)°        Permittivity"          (S/m)"          ConvF X       ConvF Y        ConvFZ        Alpha®        _(mm)          (k=2)
        750               55.5               0.96             10.21         10.21         10.21         0.41         0.84         + 12.0 %

        835               55.2               0.97             10.02         10.02         10.02         0.40         0.85         + 12.0 %

       1750               53.4                1.49            8.31           6.31          8.31         0.35         0.86         + 12.0 %

       1900               53.3                1.52            8.03           8.03          8.03         0.41         0.85         + 12.0 %

       2300               52.9                1.81            7.74           7.74          7.74         0.46         0.80         + 12.0 %

       2450               52.1                1.95            7.53           7.53          7.53         0.34         0.88         + 12.0 %

       2600               52.5               2.16             6.92           6.92          6.92         0.27         1.05         + 12.0 %

       5250              48.9                5.36             4.62          4.62           4.62         0.35         1.90         *13.1 %

       5600               48.5               5.77             4.05          4.05           4.05         0.40         1.90         * 13.1 %

       5750              48.3                5.94             4.18          4.18           4.18         0.40         1.90        + 13.1 %


© Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and &) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and a) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3954_Jan18                                    Page 6 of 11


EX3DV4— SN:3954                                                                                          January 31, 2018


                                             Frequency Response of E—Field
          Frequency response (normalized)         (TEM—Cell:ifi110 EXX, Waveguide: R22)




                                                     L*]
                                                     TEM
                                                                                            L*
                                                                                            R22


                                            Uncertainty of Frequency Response of E—field: £ 6.3% (k=2)




Certificate No: EX3—3954_Jan18                                     Page 7 of 11


EX3DV4— SN:3954                                                                                                                                                                                                                                                                                         January 31, 2018




                                                                                               Receiving Pattern (¢), 9 = 0°

                                                f=600 MHz,TEM                                                                                                                                                        f=1800 MHz,R22
                                                                                          80                                                                                                                                                               90
                                                                         +                *                ~+                                                                                                                                 *            #             »
                                                             x                                                             is                                                                                                     £                                                  .
                                                w                                                                                   "w                                                                               *                                                                        »
                                                                                          «.                                                                                                                                                               Heig
                                   135 "                         s                                     "rsy                                  * 45                                                       135 "                         *                                          &                     * 45
                                    *           a—~*——«                                                            2006             %                 ®                                                  ®                    +o.                                            5>*..                          ®
                               x            »           +                        *                 »                            .        «                \                                         x            #        &                           «          #                       .         «             s
                                        +                                                 a                                                   %                                                              +                                             x                              |             %
                           #                            &                                                                       *                             X                                 #                         *                                                              *                           #
                                    &                                                $o*                                                                                                                                                                  +0                                                >
                                                                                                               ao¥—                               I               4                        ®                          ax——*=s                                                s—*~—4                                          4
                       f                        e           *kz                      40        §
                                    o       >                        *    —                                            a                 +                                                  L            *       »                                    «1       o¢+                                 &0
                                                                      . e   _%                                                                                                             i                                                      s        t         +
                    180.                x                                —           A*rrfmciet                    _                         28                   i0                    180.                 w                                             9t                                           «                    a
                                                                                 )&                        "oz                 o4        0608                                              |                                                                t            ‘o2,            o4       o6        0s
                                    ®                                                L.       ¢%                                          *       %                                         |            £o*                          *               *\       \*                                 »o        0.
                       {                                *—*                                                        *+‘%——*                                        ¢                        t                                  y                                              *—s——*                                          *
                                    1                                                &0                                                           +                                                      &                                                604                                               +            }
                           +                        £                                     P                                     *                             4                                 %                         +                                                              &                           *
                                        &                                                                                                     +                                                              *                                             *
                               x            %       *                            ®                                              f        &                y                                         *            %        *                           x                                  J                       &
                                   l            *           .'       ®                                             *       %        *             y                                                      k           *        *           «                                  *       %        a



                                   225.                          *                                                     ®                     7315                                                       $35°.                                                                    *                     . 315
                                                                             *            &            *                                                                                                                                          *        +         *
                                                %                                                                                   »                                                                                «.                                                                       x
                                                            +         _                                        £           »                                                                                                  ¥                       M                              ¥
                                                                                          57               —                                                                                                                                              370            —

                                                                                                                                                              l                                                               €                                                                                      o_
                     Tot                                    x                                                              Y                                      Z                      Tot                                  x                                                      Y                                Z




                   &nmmdfrmemensennformmesnaefeorscnmscenmaJefrnnmstnencefevtminsmm
      Error [dB]




                   o
                   i S S ie
                   Brssssesfornsseeneccecrmsfereenseetetetiencdefemmmnsiisisemenihncnceesieesestisiiiieeerennccen
                                                                                                                                                                             Roll [*]
                                                10%10Hz                                                                                           600 MHz                                           1860 MHz .                                                                           2500 MHz

                                                                                 Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificate No: EX3—3954_Jan18                                                                                                                                         Page 8 of 11


EX3DV4— SN:3954                                                                                                                            January 31, 2018




                                                            Dynamic Range f(SARneaq)
                                                                      (TEM cell , feyi= 1900 MHz)




                                              105
                   Input Signal [uV]




                                              10%;



                                              10




                                                                      gds   |     rkas     |    sds     |            j    3.

                                                    108        102          101           10°          10           10°              10°
                                                                                  SAR [mW/icm3]
                                                               not compensated                          compensated
                                       Error [dB]

                                                    T




                                                                 I                                              T
                                                     103        102         101          100          101      102             103
                                                                                  SAR [mW/cm3]
                                                                      *
                                                               not compensated                        compensated


                                                           Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certificate No: EX3—3954_Jan18                                                    Page 9 of 11


EX3DV4— SN:3954                                                                                                                                                                                      January 31, 2018


                                                Conversion Factor Assessment
                        f= 835 MHz,WGLS R9 (H_convF)                                                                                f= 1900 MHz,WGLS R22 (H_convF)


                                                                                                 |
                                                                                                J              30

                                                                                                |
                                                                                                ||              5
                                                                                                               26—)   *n&
                                                                                                ‘                           *
                                                                                                                            &
                                                                                                                                *
                                                                                                                                 a
                                                                                                |          & wl                     *\
                                                                                                           >
      SAR [Wkgopw




                                                                                                           9
                                                                                                           #
                                                                                                           2
                    5




                                                                                                           $ o6—
                                                                                                           w


                                                        &*                                  |I                 10

                                                             K                              |

                                                                                            ||                  s                                                         —
                                                                                                                                                                                                * ces
                                                                                                                                                                                                           +
                                       1               _     A.,,.E..I.L,,.‘                                    gf    L4             .Lf .LLI                         boubec 1c3 cfo ced   ic    —uf 3 P 1O L4
                              10           15         20     25        30      36          40                                         5         10         15      20              25           30      36
                                                    2 [mm]                                                                                                       z [mm]
                             _o]                                      _¢]                                                                     e                                              Le]
                           analytcal                                measured                                                                analytical                                     measured



                                           Deviation from Isotropy in Liquid
                                                                        Error (6, 8), f = 900 MHz



                                       0.6 -—
                                   5 0.4 -=
                                   'fg 0.2 __—____
                                     0.0 |<
                                   8 .02 _
                                       0.4 _\
                                           —0.6 ==
                                           —0.8
                                           —1.0 ——
                                                0




                                   —1.0             —0.8     —0.6      —0.4         —0.2             0.0         0.2                  0.4            0.6        0.8             1.0
                                           Uncertainty of Spherical Isotropy Assessment: £ 2.6% (k=2)




Certificate No: EX3—3954_Jan18                                                       Page 10 of 11


EX3DV4— SN:3954                                              January 31, 2018




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Other Probe Parameters
 Sensor Arrangement                                              Triangular
 Connector Angle (°)                                                   70.4
 Mechanical Surface Detection Mode                                 enabled
 Optical Surface Detection Mode                                    disabled
 Probe Overall Length                                              337 mm
 Probe Body Diameter                                                10 mm
 Tip Length                                                           9 mm
 Tip Diameter                                                       2.5 mm
 Probe Tip to Sensor X Calibration Point                              1 mm
 Probe Tip to Sensor Y Calibration Point                              1 mm
 Probe Tip to Sensor Z Calibration Point                              1 mm
 Recommended Measurement Distance from Surface                      1.4 mm




Certificate No: EX3—3954_Jan18               Page 11 of 11



Document Created: 2019-07-10 12:25:03
Document Modified: 2019-07-10 12:25:03

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC