RF Exposure Info

FCC ID: 2AEHI-EL0215

RF Exposure Info

Download: PDF
FCCID_2694427

         RF Exposure Evaluation Report                                           Report No. : FA552245-01




                  RF Exposure Evaluation Report

                APPLICANT          : Chiaro Technology Ltd

                EQUIPMENT          : Elvie

                BRAND NAME         : Chiaro

                MODEL NAME         : EL02

                FCC ID             : 2AEHI-EL0215
                STANDARD           : FCC CFR 47 part 1, 1.1307(b) and 1.1310
                                     KDB 680106 D01v02

      We, SPORTON INTERNATIONAL INC., would like to declare that the device has been
      evaluated in accordance with FCC CFR 47 part 1, 1.1307(b) and 1.1310, and pass the limit.
      Without written approval of SPORTON INTERNATIONAL INC., the test report shall not be
      reproduced except in full.




      Reviewed by: Eric Huang / Deputy Manager




          Approved by: Jones Tsai / Manager



                             SPORTON INTERNATIONAL INC.
No.52, Hwa Ya 1st Rd., Hwa Ya Technology Park, Kwei-Shan District, Taoyuan City, Taiwan (R.O.C.)




SPORTON INTERNATIONAL INC.                                           Page Number        : 1 of 5
TEL : 886-3-327-3456                                                 Report Issued Date : Jul. 22, 2015
FAX : 886-3-328-4978                                                 Report Version     : Rev. 01
FCC ID : 2AEHI-EL0215


            RF Exposure Evaluation Report                                                                                     Report No. : FA552245-01

                                                         Table of Contents
1. ADMINISTRATION DATA ........................................................................................................................... 3
     1.1.      Testing Laboratory ............................................................................................................................. 3
2. DESCRIPTION OF EQUIPMENT UNDER TEST (EUT) ............................................................................. 3
3. RF EXPOSURE LIMIT INTRODUCTION .................................................................................................... 4
4. TEST MODE ................................................................................................................................................ 5
5. MEASUREMENT EQUIPMENT................................................................................................................... 5
6. RF EXPOSURE EVALUATION ................................................................................................................... 5
Appendix A. Test Setup Photo




                                                                Revision History
    REPORT NO.                 VERSION                                        DESCRIPTION                                              ISSUED DATE

    FA552245-01                 Rev. 01           Initial issue of report                                                               Jul. 22, 2015




SPORTON INTERNATIONAL INC.                                                                                     Page Number        : 2 of 5
TEL : 886-3-327-3456                                                                                           Report Issued Date : Jul. 22, 2015
FAX : 886-3-328-4978                                                                                           Report Version     : Rev. 01
FCC ID : 2AEHI-EL0215


seorron cas. _RF Exposure Evaluation Report                                                           Report No. : FA552245—01
  1. Administration Data

  1.1. Testing Laboratory

                                                          Testing Laboratory

  Test Site                               SPORTON INTERNATIONAL INC.
                                          No.52, Hwa Ya 1st Rd., Hwa Ya Technology Park, Kwei—Shan District, Taoyuan City,
                                          Taiwan (R.O.C.)
  Test Site Location
                                          TEL: +886—3—327—3456
                                          FAX: +886—3—328—4978

                                                               Applicant

 Company Name                          Chiaro Technology Ltd

 Address                               2nd Floor, 5—9 Hatton Wall, London, United Kingdom, EC1N 8HX


                                                             Manufacturer

 Company Name                          FU GANG ELECTRONIC(KUNSHAN})CO.,LTD
                                       NO.6 Zheng Wei West Road,Jin Xi Town, Kun Shan City, Jiang Su Province, 215324,
 Lebles                                China




  2. Description of Equipment Under Test (EUT)


                                   Elvie
                                   Chiaro
                                   ELO2
                                   2AEHI—ELO215
                                   110KHz ~ 205 KHz
                                   + ASK
                                   Wire
                                   Production Unit
  1                            Jun. 13, 2015
  Remark: The above EUT‘s information was declared by manufacturer. Please refer to the specifications or user‘s manual
  for more detailed description.




  SPORTON INTERNATIONAL INC.                                                              Page Number         : 3 of 5
  TEL : 886—3—327—3456                                                                    Report Issued Date : Jul. 22, 2015
  FAX : 886—3—328—4978                                                                    Report Version      : Rev. 01
  FCC ID : 2AEHI—ELO215


          RF Exposure Evaluation Report                                                             Report No. : FA552245-01
3. RF Exposure Limit Introduction
§ 1.1310 The criteria listed in table 1 shall be used to evaluate the environmental impact of human exposure to radio

frequency(RF) radiation as specified in § 1.1307(b), except in the case of portable devices which shall be evaluated

according to the provisions of § 2.1093 of this chapter.




(1) Occupational/controlled exposure limits apply in situations in which persons are exposed as a consequence of their

employment provided those persons are fully aware of the potential for exposure and can exercise control over their

exposure. Limits for occupational/controlled exposure also apply in situations when a person is transient through a

location where occupational/controlled limits apply provided he or she is made aware of the potential for exposure. The

phrase fully aware in the context of applying these exposure limits means that an exposed person has received written

and/or verbal information fully explaining the potential for RF exposure resulting from his or her employment. With the

exception of transient persons, this phrase also means that an exposed person has received appropriate training

regarding work practices relating to controlling or mitigating his or her exposure. Such training is not required for transient

persons, but they must receive written and/or verbal information and notification (for example, using signs) concerning

their exposure potential and appropriate means available to mitigate their exposure. The phrase exercise control means

that an exposed person is allowed to and knows how to reduce or avoid exposure by administrative or engineering

controls and work practices, such as use of personal protective equipment or time averaging of exposure.

(2) General population/uncontrolled exposure limits apply in situations in which the general public may be exposed, or in

which persons who are exposed as a consequence of their employment may not be fully aware of the potential for

exposure or cannot exercise control over their exposure.




SPORTON INTERNATIONAL INC.                                                              Page Number        : 4 of 5
TEL : 886-3-327-3456                                                                    Report Issued Date : Jul. 22, 2015
FAX : 886-3-328-4978                                                                    Report Version     : Rev. 01
FCC ID : 2AEHI-EL0215


                 RF Exposure Evaluation Report                                                            Report No. : FA552245-01
      4. Test Mode
      This device has been tested in the following charging conditions as below:
                      Test Mode                           Test Setup Configuration              Charging Current Condition
                         TM1                            Test w/ Client Device installed               < 1% Battery status
                         TM2                            Test w/ Client Device installed                50% Battery status
                         TM3                            Test w/ Client Device installed             Near 100% Battery status

      Support Equipment:
                   Name                       Manufacturer                     Model Name                       FCC ID
                   Elvie                        Chiaro                            EL01                       2AEHI-EL0115


      5. Measurement Equipment
       Instrument                 Manufacturer            Model No.        Serial No.      Freq Rang         Last Cal.         Due Date
Electric and Magnetic field
                               Narda S.T.S / PMM          EHP 200A        160WX41006      9KHz~30MHz       Nov. 03, 2014     Nov. 02, 2015
     Probe-Analyzey



      6. RF Exposure Evaluation
      1.    The equipment under test was placed on a wooden desk inside of shield room. The isotropic field probe was used to
            measure the field strength for 6 EUT surfaces, and during measurement a separation of 10cm is maintained between
            EUT surface and the center of the field probe. The detail setup photo please refer to Appendix A.
      2.    Per KDB 680106 D01v02, for devices designed for typical desktop applications, such a wireless charging pads, RF
            exposure evaluation should be conducted assuming a user separation distance of 10 cm. E and H field strength
            measurements or numerical modeling may be used to demonstrate compliance. Measurements should be made from
            all sides and the top of the primary/client pair, with the 10 cm measured from the center. of the probe(s) to the edge of
            the device. Emissions between 100 kHz to 300 kHz should be assessed versus the limits at 300 kHz in Table 1 of
            Section 1.1310: 614 V/m and 1.63 A/m and aggregate leakage fields at 10 cm surrounding the device from all
            simultaneous transmitting coils are demonstrated to be less than 30% of the MPE limit.

               Position                                             E-Field Measurement (V/m)
           (Distance 10cm)           A              B               C             D             E             F          30% of limit
                TM1                0.338          0.343           0.334         0.334        0.343          0.440
                TM2                0.316          0.338           0.326         0.311        0.342          0.413            184
                TM3                0.327          0.316           0.322         0.346        0.334          0.345


               Position                                             H-Field Measurement (A/m)
           (Distance 10cm)           A              B               C             D             E             F          30% of limit
                TM1                0.0528        0.0531          0.0515         0.0538       0.0510         0.0548
                TM2                0.0513        0.0527          0.0508         0.0532       0.0550         0.0560          0.489
                TM3                0.0522        0.0519          0.0522         0.0535       0.0510         0.0534


      Conclusion:
      The field strength limit refers to Part 1.1310 and the test result of exposure evaluation is compliant with 30%
      of the MPE limit. (E- Field: 184 V/m; H-field: 0.489A/m).




      SPORTON INTERNATIONAL INC.                                                              Page Number        : 5 of 5
      TEL : 886-3-327-3456                                                                    Report Issued Date : Jul. 22, 2015
      FAX : 886-3-328-4978                                                                    Report Version     : Rev. 01
      FCC ID : 2AEHI-EL0215



Document Created: 2019-09-23 20:23:13
Document Modified: 2019-09-23 20:23:13

© 2024 FCC.report