   COMOSAR E-Field Probe Calibration Report




                                             Ref : ACR.88.1.13.SATU.A




     Shenzhen SEM.Test Technology Co., Ltd.
1/F, Building A, Hongwei Industrial Park, Liuxian 2nd Road,
                Bao'an District, Shenzhen,
                 P.R.C. (518101), CHINA
SATIMO COMOSAR DOSIMETRIC E-FIELD PROBE
                    SERIAL NO.: SN 09/13 EP168


                         Calibrated at SATIMO US
                 2105 Barrett Park Dr. - Kennesaw, GA 30144




                                    03/21/2014

    Summary:
    This document presents the method and results from an accredited COMOSAR Dosimetric E-Field
    Probe calibration performed in SATIMO USA using the CALISAR / CALIBAIR test bench, for use
    with a SATIMO COMOSAR system only. All calibration results are traceable to national
    metrology institutions.


                         COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    3/28/2014

Checked by :                 Jérôme LUC                         Product Manager                    3/28/2014

Approved by :          Kim RUTKOWSKI                            Quality Manager                    3/28/2014




                                                                       Customer Name
                                                                         SEM Test
                                       Distribution :                 Compliance Service
                                                                          Co., Ltd.




   Issue                   Date                                                           Modifications
     A                  3/28/2014                  Initial release




                                                               Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




                                                    TABLE OF CONTENTS



1    Device Under Test ................................................................................................. 4
2    Product Description................................................................................................ 4
    2.1     General Information _______________________________________________________ 4
3    Measurement Method............................................................................................. 4
    3.1     Linearity ________________________________________________________________ 4
    3.2     Sensitivity _______________________________________________________________ 5
    3.3     Lower Detection Limit _____________________________________________________ 5
    3.4     Isotropy _________________________________________________________________ 5
    3.5     Boundary Effect __________________________________________________________ 5
4    Measurement Uncertainty ...................................................................................... 5
5    Calibration Measurement Results ........................................................................... 6
    5.1     Sensitivity in air __________________________________________________________ 6
    5.2     Linearity ________________________________________________________________ 7
    5.3     Sensitivity in liquid ________________________________________________________ 7
    5.4     Isotropy _________________________________________________________________ 8
6    List of Equipment ................................................................................................ 10




                                                                  Page: 3/10

                   This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of SATIMO.


                          COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




1              DEVICE UNDER TEST

                                                      Device Under Test
      Device Type                                                    COMOSAR DOSIMETRIC E FIELD PROBE
      Manufacturer                                                   Satimo
      Model                                                          SSE5
      Serial Number                                                  SN 09/13 EP168
      Product Condition (new / used)                                 new
      Frequency Range of Probe                                       0.7 GHz-3GHz
      Resistance of Three Dipoles at Connector                       Dipole 1: R1=0.225 M
                                                                     Dipole 2: R2=0.235 M
                                                                     Dipole 3: R3=0.229 M

A yearly calibration interval is recommended.

2              PRODUCT DESCRIPTION

2.1          GENERAL INFORMATION

Satimo’s COMOSAR E field Probes are built in accordance to the IEEE 1528, OET 65 Bulletin C
and CEI/IEC 62209 standards.




                           Figure 1 – Satimo COMOSAR Dosimetric E field Dipole

      Probe Length                                                   330 mm
      Length of Individual Dipoles                                   4.5 mm
      Maximum external diameter                                      8 mm
      Probe Tip External Diameter                                    5 mm
      Distance between dipoles / probe extremity                     2.7 mm


         3     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC 62209 standards provide
recommended practices for the probe calibrations, including the performance characteristics of
interest and methods by which to assess their affect. All calibrations / measurements performed meet
the fore mentioned standards.

         3.1   LINEARITY
The evaluation of the linearity was done in free space using the waveguide, performing a power
sweep to cover the SAR range 0.01W/kg to 100W/kg.




                                                                Page: 4/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                       Ref: ACR.88.1.13.SATU.A




        3.2     SENSITIVITY
The sensitivity factors of the three dipoles were determined using a two step calibration method (air
and tissue simulating liquid) using waveguides as outlined in the standards.

        3.3     LOWER DETECTION LIMIT
The lower detection limit was assessed using the same measurement set up as used for the linearity
measurement. The required lower detection limit is 10 mW/kg.

        3.4     ISOTROPY
The axial isotropy was evaluated by exposing the probe to a reference wave from a standard dipole
with the dipole mounted under the flat phantom in the test configuration suggested for system
validations and checks. The probe was rotated along its main axis from 0 - 360 degrees in 15 degree
steps. The hemispherical isotropy is determined by inserting the probe in a thin plastic box filled
with tissue-equivalent liquid, with the plastic box illuminated with the fields from a half wave dipole.
The dipole is rotated about its axis (0°–180°) in 15° increments. At each step the probe is rotated
about its axis (0°–360°).

        3.5     BOUNDARY EFFECT
The boundary effect is defined as the deviation between the SAR measured data and the expected
exponential decay in the liquid when the probe is oriented normal to the interface. To evaluate this
effect, the liquid filled flat phantom is exposed to fields from either a reference dipole or waveguide.
With the probe normal to the phantom surface, the peak spatial average SAR is measured and
compared to the analytical value at the surface.

4             MEASUREMENT UNCERTAINTY

The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty associated with an E-field
probe calibration using the waveguide technique. All uncertainties listed below represent an
expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor
of k=2, traceable to the Internationally Accepted Guides to Measurement Uncertainty.
 Uncertainty analysis of the probe calibration in waveguide
                                                 Uncertainty             Probability                                                 Standard
         ERROR SOURCES                                                                            Divisor              ci
                                                  value (%)              Distribution                                             Uncertainty (%)
 Incident or forward power                            3.00%              Rectangular                    3              1               1.732%

 Reflected power                                      3.00%              Rectangular                    3              1               1.732%
                                                                                                   1.732051
 Liquid conductivity                                  5.00%              Rectangular                    3              1               2.887%
                                                                                                   1.732051
 Liquid permittivity                                  4.00%              Rectangular                    3              1               2.309%
                                                                                                   1.732051
 Field homogeneity                                    3.00%              Rectangular                    3              1               1.732%
                                                                                                   1.732051
 Field probe positioning                              5.00%              Rectangular                    3              1               2.887%
                                                                                                   1.732051
 Field probe linearity                                3.00%              Rectangular                    3              1               1.732%
                                                                                                   1.732051

                                                                  Page: 5/10

                   This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of SATIMO.


                        COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




  Combined standard uncertainty                                                                                                  5.831%
      Expanded uncertainty
                                                                                                                                 12.0%
    95 % confidence level k = 2



       5     CALIBRATION MEASUREMENT RESULTS


                                                Calibration Parameters
   Liquid Temperature                                      21 °C
   Lab Temperature                                         21 °C
   Lab Humidity                                            45 %


       5.1   SENSITIVITY IN AIR



                                  Normx dipole Normy dipole Normz dipole
                                  1 (μV/(V/m)2) 2 (μV/(V/m)2) 3 (μV/(V/m)2)
                                      4.96          6.25          4.19


                                  DCP dipole 1 DCP dipole 2 DCP dipole 3
                                    (mV)         (mV)         (mV)
                                      90           91           91


Calibration curves ei=f(V) (i=1,2,3) allow to obtain H-field value using the formula:
                                                    E  E1  E2  E3
                                                                 2        2        2




                                                              Page: 6/10

               This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
               The information contained herein is to be used only for the purpose for which it is submitted and is not to
               be released in whole or part without written approval of SATIMO.


                     COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




     5.2   LINEARITY




     5.3   SENSITIVITY IN LIQUID

  Liquid       Frequency             Permittivity         Epsilon (S/m)               ConvF
                (MHz +/-
               100MHz)
  HL850           835                    42.56                  0.88                   6.25
  BL850           835                    55.26                  0.96                   6.50
  HL900           900                    41.79                  0.96                   5.92
  BL900           900                    55.98                  1.04                   6.12
  HL1800          1750                   40.17                  1.38                   5.56
  BL1800          1750                   52.05                  1.48                   5.72
  HL1900          1880                   39.80                  1.43                   6.16
  BL1900          1880                   52.55                  1.50                   6.30
  HL2000          1950                   38.93                  1.44                   5.61
  BL2000          1950                   53.12                  1.51                   5.79
  HL2450          2450                   38.64                  1.82                   5.51
  BL2450          2450                   52.02                  1.94                   5.70



LOWER DETECTION LIMIT: 7mW/kg




                                                           Page: 7/10

            This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
            The information contained herein is to be used only for the purpose for which it is submitted and is not to
            be released in whole or part without written approval of SATIMO.


                       COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




       5.4   ISOTROPY
HL900 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.06 dB




HL1800 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.17 dB




                                                             Page: 8/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                       COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




HL2450 MHz
- Axial isotropy:                              0.05 dB
- Hemispherical isotropy:                      0.09 dB




                                                             Page: 9/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.88.1.13.SATU.A




        6    LIST OF EQUIPMENT


                                                  Equipment Summary Sheet

    Equipment                 Manufacturer /                        Current                                               Next Calibration
                                             Identification No.
    Description                  Model                          Calibration Date                                                Date
                                                                                        Validated. No cal               Validated.      No cal
    Flat Phantom                      Satimo                 SN-20/09-SAM71
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated.      No cal
COMOSAR Test Bench                  Version 3                          NA
                                                                                        required.                       required.
                              Rhode & Schwarz
  Network Analyzer                                               SN100132                         02/2013                        02/2016
                                   ZVA
                                                                                         Characterized prior to Characterized prior to
   Reference Probe                    Satimo                  EP 94 SN 37/08
                                                                                         test. No cal required. test. No cal required.

      Multimeter                 Keithley 2000                     1188656                        11/2013                        11/2016

   Signal Generator             Agilent E4438C                  MY49070581                        12/2013                        12/2016

                                                                                         Characterized prior to Characterized prior to
       Amplifier                  Aethercomm                        SN 046
                                                                                         test. No cal required. test. No cal required.

     Power Meter                   HP E4418A                    US38261498                        11/2013                        11/2016

    Power Sensor                 HP ECP-E26A                    US37181460                         11/2013                       11/2016

                                                                              Characterized prior to Characterized prior to
  Directional Coupler           Narda 4216-20                        01386
                                                                              test. No cal required. test. No cal required.
                                                                             Validated. No cal       Validated. No cal
     Waveguide                  Mega Industries             069Y7-158-13-712
                                                                             required.               required.
                                                                             Validated. No cal                          Validated. No cal
Waveguide Transition            Mega Industries             069Y7-158-13-701
                                                                             required.                                  required.
                                                                             Validated. No cal                          Validated. No cal
Waveguide Termination           Mega Industries             069Y7-158-13-701
                                                                             required.                                  required.
Temperature / Humidity
                       Control Company                             11-661-9                        3/2014                        3/2016
      Sensor




                                                                 Page: 10/10

                   This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of SATIMO.


      SAR Reference Dipole Calibration Report




                                       Ref : ACR.333.4.12.SATU.B




   Shenzhen SEM.Test Technology Co., Ltd.
1/F, Building A, Hongwei Industrial Park, Liuxian 2nd Road,
                Bao'an District, Shenzhen,
                 P.R.C. (518101), CHINA
    SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 900 MHZ
                      SERIAL NO.: SN 47/12 DIP 0G900-205
                          Calibrated at SATIMO US
                  2105 Barrett Park Dr. - Kennesaw, GA 30144




                                      11/26/2013

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    11/28/2013

Checked by :                 Jérôme LUC                         Product Manager                    11/28/2013

Approved by :          Kim RUTKOWSKI                            Quality Manager                    11/28/2013




                                                                       Customer Name
                                                                         Shenzhen
                                       Distribution :                 SEM.Test Technology
                                                                          Co., Ltd.




   Issue                  Date                                                            Modifications
     A                 11/28/2013                  Initial release
     B                 03/25/2014                  Add body measurement




                                                               Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                              SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




                                                     TABLE OF CONTENTS



1    Introduction ........................................................................................................... 4
2    Device Under Test ................................................................................................. 4
3    Product Description................................................................................................ 4
    3.1      General Information _______________________________________________________ 4
4    Measurement Method............................................................................................. 5
    4.1      Return Loss Requirements __________________________________________________ 5
    4.2      Mechanical Requirements ___________________________________________________ 5
5    Measurement Uncertainty ...................................................................................... 5
    5.1      Return Loss ______________________________________________________________ 5
    5.2      Dimension Measurement ___________________________________________________ 5
    5.3      Validation Measurement ____________________________________________________ 5
6    Calibration Measurement Results ........................................................................... 6
    6.1      Return Loss ______________________________________________________________ 6
    6.2      Mechanical Dimensions ____________________________________________________ 6
7    Validation measurement ......................................................................................... 7
    7.1      Measurement Condition ____________________________________________________ 7
    7.2      Head Liquid Measurement __________________________________________________ 7
    7.3      Measurement Result _______________________________________________________ 8
    7.4      Body Measurement Result __________________________________________________ 9
8    List of Equipment ................................................................................................ 10




                                                                   Page: 3/10

                    This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




1          INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2          DEVICE UNDER TEST




                                                   Device Under Test
        Device Type                                        COMOSAR 900 MHz REFERENCE DIPOLE
        Manufacturer                                       Satimo
        Model                                              SID900
        Serial Number                                      SN 47/12 DIP 0G900-205
        Product Condition (new / used)                     new

A yearly calibration interval is recommended.

3          PRODUCT DESCRIPTION

3.1     GENERAL INFORMATION

Satimo’s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                              Figure 1 – Satimo COMOSAR Validation Dipole



                                                             Page: 4/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




       4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of -20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5            MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURN LOSS
The following uncertainties apply to the return loss measurement:
                    Frequency band                                     Expanded Uncertainty on Return Loss
                      400-6000MHz                                                                0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                       Length (mm)                                         Expanded Uncertainty on Length
                            3 - 300                                                             0.05 mm


       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                     Scan Volume                  Expanded Uncertainty
                                1g                                                    16.19 %
                               10 g                                                   15.86 %

                                                             Page: 5/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




      6     CALIBRATION MEASUREMENT RESULTS

      6.1   RETURN LOSS




     Frequency (MHz)                                    Return Loss (dB)                                    Requirement (dB)
           900                                               -27.36                                               -20

      6.2   MECHANICAL DIMENSIONS
Frequency MHz                          L mm                                         h mm                                       d mm

                          required              measured                required             measured             required            measured

    300                 420.0 ±1 %.                                   250.0 ±1 %.                                 6.35 ±1 %.
    450                 290.0 ±1 %.                                   166.7 ±1 %.                                 6.35 ±1 %.
    750                 176.0 ±1 %.                                   100.0 ±1 %.                                 6.35 ±1 %.
    835                 161.0 ±1 %.                                    89.8 ±1 %.                                 3.6 ±1 %.
    900                 149.0 ±1 %.                PASS                83.3 ±1 %.               PASS              3.6 ±1 %.            PASS

    1450                 89.1 ±1 %.                                    51.7 ±1 %.                                 3.6 ±1 %.
    1500                 80.5 ±1 %.                                    50.0 ±1 %.                                 3.6 ±1 %.
    1640                 79.0 ±1 %.                                    45.7 ±1 %.                                 3.6 ±1 %.
    1750                 75.2 ±1 %.                                    42.9 ±1 %.                                 3.6 ±1 %.
    1800                 72.0 ±1 %.                                    41.7 ±1 %.                                 3.6 ±1 %.
    1900                 68.0 ±1 %.                                    39.5 ±1 %.                                 3.6 ±1 %.
    1950                 66.3 ±1 %.                                    38.5 ±1 %.                                 3.6 ±1 %.
    2000                 64.5 ±1 %.                                    37.5 ±1 %.                                 3.6 ±1 %.
    2100                 61.0 ±1 %.                                    35.7 ±1 %.                                 3.6 ±1 %.
    2300                 55.5 ±1 %.                                    32.6 ±1 %.                                 3.6 ±1 %.
    2450                 51.5 ±1 %.                                    30.4 ±1 %.                                 3.6 ±1 %.
    2600                 48.5 ±1 %.                                    28.8 ±1 %.                                 3.6 ±1 %.
    3000                 41.5 ±1 %.                                    25.0 ±1 %.                                 3.6 ±1 %.
    3500                 37.0±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.
    3700                 34.7±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.




                                                               Page: 6/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




        7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

        7.1    MEASUREMENT CONDITION
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Head Liquid Values: eps’ : 40.9 sigma : 0.96
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               900 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %

        7.2    HEAD LIQUID MEASUREMENT
                          Frequency
                                                 Relative permittivity (r’)              Conductivity (σ) S/m
                             MHz
                                                 required           measured            required          measured
                              300               45.3 ±5 %                               0.87 ±5 %
                              450               43.5 ±5 %                               0.87 ±5 %
                              750               41.9 ±5 %                               0.89 ±5 %
                              835               41.5 ±5 %                               0.90 ±5 %
                              900               41.5 ±5 %              PASS             0.97 ±5 %            PASS
                             1450               40.5 ±5 %                               1.20 ±5 %
                             1500               40.4 ±5 %                               1.23 ±5 %
                             1640               40.2 ±5 %                               1.31 ±5 %
                             1750               40.1 ±5 %                               1.37 ±5 %
                             1800               40.0 ±5 %                               1.40 ±5 %
                             1900               40.0 ±5 %                               1.40 ±5 %
                             1950               40.0 ±5 %                               1.40 ±5 %
                             2000               40.0 ±5 %                               1.40 ±5 %
                             2100               39.8 ±5 %                               1.49 ±5 %
                             2300               39.5 ±5 %                               1.67 ±5 %
                             2450               39.2 ±5 %                               1.80 ±5 %
                             2600               39.0 ±5 %                               1.96 ±5 %
                             3000               38.5 ±5 %                               2.40 ±5 %
                             3500               37.9 ±5 %                               2.91 ±5 %

                                                                Page: 7/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                       SAR REFERENCE DIPOLE CALIBRATION REPORT                                                      Ref: ACR.333.3.12.SATU.B




      7.3   MEASUREMENT RESULT
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

                      Frequency
                                                 1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                         MHz
                                             required           measured            required          measured
                          300                   2.85                                   1.94
                          450                   4.58                                   3.06
                          750                   8.49                                   5.55
                          835                   9.56                                   6.22
                          900                   10.9           11.12 (1.11)            6.99           6.99 (0.70)
                         1450                    29                                     16
                         1500                   30.5                                   16.8
                         1640                   34.2                                   18.4
                         1750                   36.4                                   19.3
                         1800                   38.4                                   20.1
                         1900                   39.7                                   20.5
                         1950                   40.5                                   20.9
                         2000                   41.1                                   21.1
                         2100                   43.6                                   21.9
                         2300                   48.7                                   23.3
                         2450                   52.4                                    24
                         2600                   55.3                                   24.6
                         3000                   63.8                                   25.7
                         3500                   67.1                                    25




                                                            Page: 8/10

             This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
             The information contained herein is to be used only for the purpose for which it is submitted and is not to
             be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




        7.4    BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body Liquid Values: eps’ : 54.1 sigma : 1.07
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               900 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                          measured                               measured
                              900                        11.30 (1.13)                           7.17 (0.72)




                                                                Page: 9/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.3.12.SATU.B




       8    LIST OF EQUIPMENT


                                                 Equipment Summary Sheet

   Equipment                 Manufacturer /                        Current                                               Next Calibration
                                            Identification No.
   Description                  Model                          Calibration Date                                                Date
                                                                                       Validated. No cal               Validated.      No cal
   Flat Phantom                      Satimo                 SN-20/09-SAM71
                                                                                       required.                       required.
                                                                                       Validated. No cal               Validated.      No cal
COMOSAR Test Bench                 Version 3                          NA
                                                                                       required.                       required.
                             Rhode & Schwarz
  Network Analyzer                                              SN100132                         02/2013                        02/2016
                                  ZVA

      Calipers                       Carrera                   CALIPER-01                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
  Reference Probe                    Satimo                EPG122 SN 18/11
                                                                                        test. No cal required. test. No cal required.

     Multimeter                 Keithley 2000                     1188656                        11/2013                        11/2016

  Signal Generator             Agilent E4438C                  MY49070581                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
      Amplifier                  Aethercomm                        SN 046
                                                                                        test. No cal required. test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        11/2013                        11/2016

   Power Sensor                 HP ECP-E26A                    US37181460                         11/2013                       11/2016

                                                                                        Characterized prior to Characterized prior to
 Directional Coupler           Narda 4216-20                        01386
                                                                                        test. No cal required. test. No cal required.
  Temperature and
                              Control Company                     11-661-9                        3/2014                        3/2016
  Humidity Sensor




                                                                Page: 10/10

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.


      SAR Reference Dipole Calibration Report




                                       Ref : ACR.333.4.12.SATU.B




   Shenzhen SEM.Test Technology Co., Ltd.
1/F, Building A, Hongwei Industrial Park, Liuxian 2nd Road,
                Bao'an District, Shenzhen,
                 P.R.C. (518101), CHINA
    SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 1800 MHZ
                      SERIAL NO.: SN 47/12 DIP 1G800-206
                          Calibrated at SATIMO US
                  2105 Barrett Park Dr. - Kennesaw, GA 30144




                                      11/26/2013

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    11/28/2013

Checked by :                 Jérôme LUC                         Product Manager                    11/28/2013

Approved by :          Kim RUTKOWSKI                            Quality Manager                    11/28/2013




                                                                       Customer Name
                                                                         Shenzhen
                                       Distribution :                 SEM.Test Technology
                                                                          Co., Ltd.




   Issue                  Date                                                            Modifications
     A                 11/28/2013                  Initial release
     B                 03/25/2014                  Add body measurement




                                                               Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                              SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




                                                     TABLE OF CONTENTS



1    Introduction ........................................................................................................... 4
2    Device Under Test ................................................................................................. 4
3    Product Description................................................................................................ 4
    3.1      General Information _______________________________________________________ 4
4    Measurement Method............................................................................................. 5
    4.1      Return Loss Requirements __________________________________________________ 5
    4.2      Mechanical Requirements ___________________________________________________ 5
5    Measurement Uncertainty ...................................................................................... 5
    5.1      Return Loss ______________________________________________________________ 5
    5.2      Dimension Measurement ___________________________________________________ 5
    5.3      Validation Measurement ____________________________________________________ 5
6    Calibration Measurement Results ........................................................................... 6
    6.1      Return Loss ______________________________________________________________ 6
    6.2      Mechanical Dimensions ____________________________________________________ 6
7    Validation measurement ......................................................................................... 7
    7.1      Measurement Condition ____________________________________________________ 7
    7.2      Head Liquid Measurement __________________________________________________ 7
    7.3      Measurement Result _______________________________________________________ 8
    7.4      Body Measurement Result __________________________________________________ 9
8    List of Equipment ................................................................................................ 10




                                                                   Page: 3/10

                    This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




1          INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2          DEVICE UNDER TEST




                                                   Device Under Test
        Device Type                                        COMOSAR 1800 MHz REFERENCE DIPOLE
        Manufacturer                                       Satimo
        Model                                              SID1800
        Serial Number                                      SN 47/12 DIP 1G800-206
        Product Condition (new / used)                     new

A yearly calibration interval is recommended.

3          PRODUCT DESCRIPTION

3.1     GENERAL INFORMATION

Satimo’s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                              Figure 1 – Satimo COMOSAR Validation Dipole



                                                             Page: 4/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




       4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of -20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5            MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURN LOSS
The following uncertainties apply to the return loss measurement:
                    Frequency band                                     Expanded Uncertainty on Return Loss
                      400-6000MHz                                                                0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                       Length (mm)                                         Expanded Uncertainty on Length
                            3 - 300                                                             0.05 mm


       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                     Scan Volume                  Expanded Uncertainty
                                1g                                                    16.19 %
                               10 g                                                   15.86 %

                                                             Page: 5/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




      6     CALIBRATION MEASUREMENT RESULTS

      6.1   RETURN LOSS




     Frequency (MHz)                                    Return Loss (dB)                                    Requirement (dB)
          1800                                               -26.65                                               -20

      6.2   MECHANICAL DIMENSIONS
Frequency MHz                          L mm                                         h mm                                       d mm

                          required              measured                required             measured             required            measured

    300                 420.0 ±1 %.                                   250.0 ±1 %.                                 6.35 ±1 %.
    450                 290.0 ±1 %.                                   166.7 ±1 %.                                 6.35 ±1 %.
    750                 176.0 ±1 %.                                   100.0 ±1 %.                                 6.35 ±1 %.
    835                 161.0 ±1 %.                                    89.8 ±1 %.                                 3.6 ±1 %.
    900                 149.0 ±1 %.                                    83.3 ±1 %.                                 3.6 ±1 %.
    1450                 89.1 ±1 %.                                    51.7 ±1 %.                                 3.6 ±1 %.
    1500                 80.5 ±1 %.                                    50.0 ±1 %.                                 3.6 ±1 %.
    1640                 79.0 ±1 %.                                    45.7 ±1 %.                                 3.6 ±1 %.
    1750                 75.2 ±1 %.                                    42.9 ±1 %.                                 3.6 ±1 %.
    1800                 72.0 ±1 %.                PASS                41.7 ±1 %.               PASS              3.6 ±1 %.            PASS

    1900                 68.0 ±1 %.                                    39.5 ±1 %.                                 3.6 ±1 %.
    1950                 66.3 ±1 %.                                    38.5 ±1 %.                                 3.6 ±1 %.
    2000                 64.5 ±1 %.                                    37.5 ±1 %.                                 3.6 ±1 %.
    2100                 61.0 ±1 %.                                    35.7 ±1 %.                                 3.6 ±1 %.
    2300                 55.5 ±1 %.                                    32.6 ±1 %.                                 3.6 ±1 %.
    2450                 51.5 ±1 %.                                    30.4 ±1 %.                                 3.6 ±1 %.
    2600                 48.5 ±1 %.                                    28.8 ±1 %.                                 3.6 ±1 %.
    3000                 41.5 ±1 %.                                    25.0 ±1 %.                                 3.6 ±1 %.
    3500                 37.0±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.
    3700                 34.7±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.




                                                               Page: 6/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




        7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

        7.1    MEASUREMENT CONDITION
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Head Liquid Values: eps’ : 39.9 sigma : 1.39
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               1800 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %

        7.2    HEAD LIQUID MEASUREMENT
                          Frequency
                                                 Relative permittivity (r’)              Conductivity (σ) S/m
                             MHz
                                                 required           measured            required          measured
                              300               45.3 ±5 %                               0.87 ±5 %
                              450               43.5 ±5 %                               0.87 ±5 %
                              750               41.9 ±5 %                               0.89 ±5 %
                              835               41.5 ±5 %                               0.90 ±5 %
                              900               41.5 ±5 %                               0.97 ±5 %
                             1450               40.5 ±5 %                               1.20 ±5 %
                             1500               40.4 ±5 %                               1.23 ±5 %
                             1640               40.2 ±5 %                               1.31 ±5 %
                             1750               40.1 ±5 %                               1.37 ±5 %
                             1800               40.0 ±5 %              PASS             1.40 ±5 %            PASS
                             1900               40.0 ±5 %                               1.40 ±5 %
                             1950               40.0 ±5 %                               1.40 ±5 %
                             2000               40.0 ±5 %                               1.40 ±5 %
                             2100               39.8 ±5 %                               1.49 ±5 %
                             2300               39.5 ±5 %                               1.67 ±5 %
                             2450               39.2 ±5 %                               1.80 ±5 %
                             2600               39.0 ±5 %                               1.96 ±5 %
                             3000               38.5 ±5 %                               2.40 ±5 %
                             3500               37.9 ±5 %                               2.91 ±5 %

                                                                Page: 7/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                       SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




      7.3   MEASUREMENT RESULT
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

                      Frequency
                                                 1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                         MHz
                                             required           measured            required          measured
                          300                   2.85                                   1.94
                          450                   4.58                                   3.06
                          750                   8.49                                   5.55
                          835                   9.56                                   6.22
                          900                   10.9                                   6.99
                         1450                    29                                     16
                         1500                   30.5                                   16.8
                         1640                   34.2                                   18.4
                         1750                   36.4                                   19.3
                         1800                   38.4           39.25 (3.92)            20.1          20.29 (2.03)
                         1900                   39.7                                   20.5
                         1950                   40.5                                   20.9
                         2000                   41.1                                   21.1
                         2100                   43.6                                   21.9
                         2300                   48.7                                   23.3
                         2450                   52.4                                    24
                         2600                   55.3                                   24.6
                         3000                   63.8                                   25.7
                         3500                   67.1                                    25




                                                            Page: 8/10

             This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
             The information contained herein is to be used only for the purpose for which it is submitted and is not to
             be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




        7.4    BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body Liquid Values: eps’ : 54.1 sigma : 1.50
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               1800 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                          measured                               measured
                             1800                        40.42 (4.04)                           21.42 (2.14)




                                                                Page: 9/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.333.4.12.SATU.B




       8    LIST OF EQUIPMENT


                                                 Equipment Summary Sheet

   Equipment                 Manufacturer /                        Current                                               Next Calibration
                                            Identification No.
   Description                  Model                          Calibration Date                                                Date
                                                                                       Validated. No cal               Validated.      No cal
   Flat Phantom                      Satimo                 SN-20/09-SAM71
                                                                                       required.                       required.
                                                                                       Validated. No cal               Validated.      No cal
COMOSAR Test Bench                 Version 3                          NA
                                                                                       required.                       required.
                             Rhode & Schwarz
  Network Analyzer                                              SN100132                         02/2013                        02/2016
                                  ZVA

      Calipers                       Carrera                   CALIPER-01                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
  Reference Probe                    Satimo                EPG122 SN 18/11
                                                                                        test. No cal required. test. No cal required.

     Multimeter                 Keithley 2000                     1188656                        11/2013                        11/2016

  Signal Generator             Agilent E4438C                  MY49070581                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
      Amplifier                  Aethercomm                        SN 046
                                                                                        test. No cal required. test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        11/2013                        11/2016

   Power Sensor                 HP ECP-E26A                    US37181460                         11/2013                       11/2016

                                                                                        Characterized prior to Characterized prior to
 Directional Coupler           Narda 4216-20                        01386
                                                                                        test. No cal required. test. No cal required.
  Temperature and
                              Control Company                     11-661-9                        3/2014                        3/2016
  Humidity Sensor




                                                                Page: 10/10

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.



Document Created: 2014-11-22 16:58:58
Document Modified: 2014-11-22 16:58:58

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC