RF Exposure Info

FCC ID: 2ACVD05601045

RF Exposure Info

Download: PDF
FCCID_2679701

         ORLAB!                                                                                         REPORT No. : $Z15040140S502




                                         RF EXPOSURE
                           EVALUATION REPORT


               APPLICANT                                    Testo Instruments (Shenzhen) Co., Ltd

                                                            Combined     infrared    and                                  penetration
               PRODUCT NAME
                                                            thermometer with Bluetooth®

               MODEL NAME                                   testo 104—IR BT


               TRADE NAME                                   Testo


                BRAND NAME                                  Testo


                FCC ID                                      2ACVDO05601045


               STANDARD(S)

                ISSUE DATE




             SHENZHEN MORLAB COMM

NOTE: This document is issued by MORLAB, the test report shall not be reproduced except in full without prior written permission of the
company. The test results apply only to the particular sample(s) tested and to the specific tests carried out which is available on request for
validation and information confirmed at our website.


                                         FL1—3, Building A, FeiYang Science Park, No.8 LongChang Road            Tel: 86—755—36698555     Fax: 86—755—30698525

M 0 RLAB             G Ro U P            Block67, BaoAn District, ShenZhen , GuangDong Province, P. R.   China   Hittp—//www.morlab.com   E—mail: service@morlab.cn


                                                                                                           REPORT No.:SZ15040140S02



                                                                DIRECTORY

 TEST REPORT DECLARATION ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙3


 1. TECHNICAL INFORMATION ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙4


 1.1. IDENTIFICATION OF APPLICANT ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙4
 1.2. IDENTIFICATION OF MANUFACTURER ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙4
 1.3. EQUIPMENT UNDER TEST (EUT) ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙4
 1.3.1. PHOTOGRAPHS OF THE EUT∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙5
 1.3.2. IDENTIFICATION OF ALL USED EUT∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙6
 1.4. APPLIED REFERENCE DOCUMENTS ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙6


 2.DEVICE CATEGORY AND RF EXPOSURE LIMIT ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙7


 3.MEASUREMENT OF CONDUCTED PEAK OUTPUT POWER ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙8


 4. RF EXPOSURE EVALUATION ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙8


 ANNEX A GENERAL INFORMATION ∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙9



                                                                   Change History
  Issue                Date                                                           Reason for change
    1.0           2015-07-08                                                                First edition




MORLAB GROUP
                                           FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                                           Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                                  Page 2 0f 9


                                                                                         REPORT No. : $Z15040140802



                        TEST REPORT DECLARATION

  Applicant                  Testo Instruments (Shenzhen) Co., Ltd
                             Block A, B4 Building, China Merchants Guangming
  Applicant Address          Sci&Tech Park, No.3009 Guan Guang Road, Guangming
                             New District, Shenzhen City
  Manufacturer               Testo Instruments (Shenzhen) Co., Ltd
                             Block A, B4 Building, China Merchants Guangming
  Manufacturer Address       Sci&Tech Park, No.3009 Guan Guang Road, Guangming
                             New District, Shenzhen City
                             Combined infrared and penetration thermometer with
  Product Name
                             Bluetooth®
  Model Name                 testo 104—IR BT

  Brand Name                 Testo

  HW Version                 V1.1

  SW Version                 V1.0
                             47CFR 2.1093;
  Test Standards
                             KDB 447498 D01 General RF Exposure Guidance v05r02

  Issue Date                 2015—07—09

  SAR Evaluation             Not Required




               Tested by                                                  Lin Jun
                                                                           Liu Jun


                 Reviewed by         _:                     é            ZAA/\ Z}\CW‘
                                                                    ~—   Zhu Zhan


               Approved by           _:                            ZfiM D{XQh
                                                                         Zené Dexin



                         FL1—3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86—755—36698555    Fax: 86—755—36608525
MORLAB GROUP             Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Hittp:/www.morlab.com   E—mail: service@morlab.cn

                                                                                                                                Page 3 Of 9


                                                                                            REPORT No.:SZ15040140S02



 1. TECHNICAL INFORMATION
 Note: the following data is based on the information by the applicant.

 1.1. Identification of Applicant
 Company Name:               Testo Instruments (Shenzhen) Co., Ltd
 Address:                    Block A, B4 Building, China Merchants Guangming Sci&Tech Park,
                             No.3009 Guan Guang Road, Guangming New District, Shenzhen City

 1.2. Identification of Manufacturer
 Company Name:               Testo Instruments (Shenzhen) Co., Ltd
 Address:                    Block A, B4 Building, China Merchants Guangming Sci&Tech Park,
                             No.3009 Guan Guang Road, Guangming New District, Shenzhen City

 1.3. Equipment Under Test (EUT)
 Model Name:                 testo 104-IR BT
 Trade Name:                 Testo
 Brand Name:                 Testo
 Hardware Version:           V1.1
 Software Version:           V1.0
 Frequency Bands:            Bluetooth4.0:2402-2480MHz;
 Modulation Mode:            Bluetooth4.0: GFSK
 Antenna type:               Fixed Internal Antenna
 Development Stage:          Identical prototype




MORLAB GROUP
                            FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                            Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                   Page 4 0f 9


                                                                                         REPORT No.:SZ15040140S02



 1.3.1. Photographs of the EUT


 1.   EUT front view




 2.   EUT rear view




MORLAB GROUP
                         FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                         Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                Page 5 0f 9


                                                                                            REPORT No.:SZ15040140S02



 1.3.2. Identification of all used EUT

 The EUT identity consists of numerical and letter characters, the letter character indicates the test
 sample, and the following two numerical characters indicate the software version of the test
 sample.


               EUT
                            Hardware Version                                         Software Version
             Identity
                   1#                   V1.1                                                 V1.0


 1.4. Applied Reference Documents
 Leading reference documents for testing:
 No.    Identity                            Document Title
 1      47 CFR§2.1093                       Radiofrequency Radiation Exposure Evaluation: portable
                                            devices
 2      KDB 447498 D01v05r02                General RF Exposure Guidance




MORLAB GROUP
                            FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                            Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                   Page 6 0f 9


                                                                                             REPORT No.:SZ15040140S02



 2. DEVICE CATEGORY AND RF EXPOSURE LIMIT

 Per user manual, this device is a thermometer with Bluetooth®. Based on 47CFR 2.1093, this
 device belongs to portable device category with General Population/Uncontrolled exposure.
 Portable Devices:
 47CFR 2.1093(b)
 For purposes of this section, a portable device is defined as a transmitting device designed to be
 used so that the radiating structure(s) of the device is/are within 20 centimeters of the body of the
 user.

 GENERAL POPULATION / UNCONTROLLED EXPOSURE
 47CFR 2.1093(d) (2)
 Limits for General Population/Uncontrolled exposure: 0.08 W/kg as averaged over the whole-body
 and spatial peak SAR not exceeding 1.6 W/kg as averaged over any 1 gram of tissue (defined as a
 tissue volume in the shape of a cube). Exceptions are the hands, wrists, feet and ankles where the
 spatial peak SAR shall not exceed 4 W/kg, as averaged over any 10 grams of tissue (defined as a
 tissue volume in the shape of a cube). General Population/Uncontrolled limits apply when the
 general public may be exposed, or when persons that are exposed as a consequence of their
 employment may not be fully aware of the potential for exposure or do not exercise control over
 their exposure. Warning labels placed on consumer devices such as cellular telephones will not be
 sufficient reason to allow these devices to be evaluated subject to limits for occupational/controlled
 exposure in paragraph (d)(1) of this section.




MORLAB GROUP
                             FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                             Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                    Page 7 0f 9


                                                                                              REPORT No.:SZ15040140S02



 3. MEASUREMENT OF CONDUCTED PEAK OUTPUT POWER


 1. BT 4.0 peak output power


                                                                                            Output
                                                             Frequency                    Power(dBm)
                       Band              Channel
                                                               (MHz)
                                                                                              GFSK
                                               0                  2402                        0.297
                         BT                   19                  2440                        -0.106
                                              39                  2480                        -0.627




 4. RF EXPOSURE EVALUATION
 The device only incorporates a Bluetooth transmitter, so standalone SAR evaluation is required for
 Bluetooth and simultaneous SAR is not required.

 Standalone transmission SAR evaluation

 According to KDB 447498 section 4.3.1, the 1-g SAR test exclusion thresholds at test separation
 Distances ≤ 50 mm are determined by:
 [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance,
 mm)]·[√f(GHz)] ≤ 3.0
 The maximum tune-up limit power is 1.07mW @ 2.402GHz
 When Bluetooth thermometer touch the body, BT antenna spacing 0mm from body, so use 5mm
 as the most conservative minimum test separation distance,
 [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance,
 mm)]·[√f(GHz)] =0.331 ≤ 3.0

 So SAR evaluation is not required for this device.




MORLAB GROUP
                              FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                              Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                     Page 8 0f 9


                                                                                           REPORT No.:SZ15040140S02




 ANNEX A GENERAL INFORMATION

 1. Identification of the Responsible Testing Laboratory
  Company Name:                              Shenzhen Morlab Communications Technology Co., Ltd.
  Department:                                Morlab Laboratory
  Address:                                   FL.3, Building A, FeiYang Science Park, No.8 LongChang
                                             Road, Block 67, BaoAn District, ShenZhen, GuangDong
                                             Province, P. R. China
  Responsible Test Lab Manager:              Mr. Su Feng
  Telephone:                                 +86 755 36698555
  Facsimile:                                 +86 755 36698525


 2. Identification of the Responsible Testing Location
  Name:                                      Shenzhen Morlab Communications Technology Co., Ltd.
                                             Morlab Laboratory
  Address:                                   FL.3, Building A, FeiYang Science Park, No.8 LongChang
                                             Road, Block 67, BaoAn District, ShenZhen, GuangDong
                                             Province, P. R. China


 3. Accreditation Certificate
   Accredited Testing Laboratory:            CNAS No. L3572
                                             (Shenzhen Morlab Communications Technology Co., Ltd.)


                                     ***** END OF REPORT *****




MORLAB GROUP
                           FL1-3, Building A, FeiYang Science Park, No.8 LongChang Road,         Tel: 86-755-36698555    Fax: 86-755-36698525
                           Block67, BaoAn District, ShenZhen , GuangDong Province, P. R. China   Http://www.morlab.com   E-mail: service@morlab.cn

                                                                                                                                  Page 9 0f 9



Document Created: 2015-07-09 10:24:17
Document Modified: 2015-07-09 10:24:17

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC