RF Exposure

FCC ID: 2ACN7-HRM826

RF Exposure Info

Download: PDF
FCCID_4186737

RF Exposure Requirements

Product Description: Heart Rate Monitor
Model No.: HRM826
FCC ID: 2ACN7-HRM826

According to the KDB 447498 D01 v06 section 4.3.1, for 100 MHz to 6 GHz and test separation
distances ≤ 50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the
following:
[(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance,
mm)] · [√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR, where
- f(GHz) is the RF channel transmit frequency in GHz
- Power and distance are rounded to the nearest mW and mm before calculation17
- The result is rounded to one decimal place for comparison

Calculation Result:

Tx frequency range: 2402-2480MHz
Min. test separation distance: 5mm
Maximum Conducted Output Power: 0.612dBm
Tune-Up output power: 1dBm
RF channel transmit frequency: 2480MHz
Result: 0.4
Limit: 3.0



The exclusion thresholds is 0.4≤3, so the transmitter complies with the RF exposure requirements
and the SAR is not required.



Document Created: 2019-03-01 10:42:34
Document Modified: 2019-03-01 10:42:34