I19Z60716-SEM03_SAR_Rev1_part5

FCC ID: 2ACCJN031

RF Exposure Info

Download: PDF
FCCID_4343928

Calibration Laboratory of                                         $s/                                      S      Schweizerischer Kalibrierdienst
Schmid & Partner                                                  fnzx/k                                   C      Service suisse d‘étalonnage




                                                           |
  Engineering AG                                                 ;//a >                                           Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                     *,fi\\\\‘}                                 S      Swiss Calibration Service
                                                                      thilals

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CTTL (Auden)                                                                      Certificate No: D2450V2—853_Jul18

CALIBRATION CERTIFICATE
 Object                                D2450V2 — SN:853



 Calibration procedure(s)             QA CAL—05.v10
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    July 24, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                          Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                     Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                           Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—026783)                          Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                           Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06827        04—Apr—18 (No. 217—02683)                           Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                       Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                      Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                               Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
 Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                    In house check: Oct—18


                                        Name                                    Function
 Calibrated by:                        Claudio Leubler                          Laboratory Technician



 Approved by:                          Katia Pokovic                            Technical Manager



                                                                                                                      Issued: July 24, 2018
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D2450V2—853_Jul18                                     Page 1 of 8


Calibration Laboratory of                                                     S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                              C   Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                  S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—853_Jul18                               Page 2 of 8


)
(ugu
        S
       Measurement Conditions
           DASY system configuration, as far as not given on page 1.
             DASY Version                                              DASY5                              V52.10.1
             Extrapolation                                  Advanced Extrapolation
             Phantom                                         Modular Flat Phantom
             Distance Dipole Center— TSL                               10 mm                             with Spacer
             Zoom Scan Resolution                              dx, dy, dz =5 mm
             Frequency                                         2450 MHz + 1 MHz



       Head TSL parameters
          The following parameters and calculations were applied.
                                                                       Temperature       Permittivity         Conductivity
             Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
             Measured Head TSL parameters                              (22.0 + 0.2) °C   37.8 26 %          1.85 mho/m + 6 %
             Head TSL temperature change during test                      <0.5 °C            ~——                     _

       SAR result with Head TSL

            SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
             SAR measured                                       250 mW input power                      13.2 W/kg
             SAR for nominal Head TSL parameters                    normalized to 1W         51.7 Wikg + 17.0 % (k=2)


             SAR averaged over 10 cm‘ (10 g) of Head TSL                condition
             SAR measured                                       250 mW input power                  618 W/kg
             SAR for nominal Head TSL parameters                    normalized to 1W         24.2 Wikg £ 16.5 % (k=2)


       Body TSL parameters
          The following parameters and calculations were applied.
                                                                       Temperature       Permittivity         Conductivity
             Nominal Body TSL parameters                                  22.0 °C           §$2.7              1.95 mho/m
             Measured Body TSL parameters                              (22.0 + 0.2) °C   51.9 £6 %          2.02 mho/m x 6 %
             Body TSL temperature change during test                      <0.5 °C            ~———                    ~———


       SAR result with Body TSL

             SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
             SAR measured                                       250 mW input power                      18.1 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         51.3 W/kg + 17.0 % (k=2)

             SAR averaged over 10 cm* (10 g) of Body TSL                 condition
             SAR measured                                       250 mW input power                      6.10 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         24.1 Wikg # 16.5 % (k=2)



       Certificate No: D2450V2—853_Jul18                      Page 3 of 8


(EN,
       Appendix (Additional assessments outside the scope of SCS 0108)

       Antenna Parameters with Head TSL

             Impedance, transformed to feed point                                           54.1 Q + 8.9 jQ
             Return Loss                                                                       — 25.2 dB



       Antenna Parameters with Body TSL

             Impedance, transformed to feed point                                           48.5 Q + 6.0 jQ
             Return Loss                                                                      ~24.1 dB



       General Antenna Parameters and Design

           Iilectrical Delay (one direction)                                                   1.161 ns


       After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can be measured.

       The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
       second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
       are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
       "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
       according to the Standard.
       No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
       feedpoint may be damaged.


       Additional EUT Data

             Manufactured by                                                                    SPEAG
             Manufactured on                                                              November 10, 2009




       Certificate No: D2450V2—853_Jul18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                     Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.85 $/m; & = 37.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e    Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.3 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 26.1 W/kg
SAR(I g) = 13.2 W/kg; SAR(10 g) = 6.13 W/kg
Maximum value of SAR (measured) = 21.6 W/kg




                             1320


                             +17.60


                             —22.00

                                      0 dB = 21.6 W/kg = 13.34 dBW/kg




Certificate No: D2450V2—853_Jul18                    Page 5 of 8


Impedance Measurement Plot for Head TSL




     Eile ViewChannel Sweep Calbration Trace Scale Marker System Window .

                                                                        2450000 GHz      54.148 Q
                                                                           254.98 pH     3.9251 0
                                                                        2450000 GHz    54.794 mU
                                                                                          41.259 °




                   Ch1Aug= 20
          Ch1: Start 2.25000 GHz                                                       Stop 2.65000 GHz

        0.00
                                                                                       —2$.225 dB
       5.00
       0.00
        .00




          00

          00
          00
          00
          00       Ch 1     =
          Ch1: Start 2.25000 GHz   ——                                                  Stop 2.65000 GHz


       Status _   CH 1:                                                                     teos?




Certificate No: D2450V2—853_Jul18


               s                                                         I




)
@77‘                                                                             F
e




 DASY5 Validation Report for Body TSL

                                                                                      Date: 16.07.2018
 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:853

 Communication System: UID 0 — CW; Frequency: 2450 MHz
 Medium parameters used: f = 2450 MHz; 0 = 2.02 S/m; &, = 51.9; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

     *    Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

     e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

     *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

     «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 108.0 V/m; Power Drift =—0.03 dB
 Peak SAR (extrapolated) = 25.6 W/kg
 SAR(I g) = 13.1 W/kg; SAR(10 g) = 6.1 W/kg
 Maximum value of SAR (measured) = 21.0 W/kg


          dB
          0


          —4.00


          —8.00


          —12.00


          ~16.00


          —20.00
                    0 dB = 21.0 W/kg = 13.22 dBW/kg




 Certificate No: D2450V2—853_Jul18               Page 7 of 8


Impedance Measurement Plot for Body TSL




     Eile View Channel Sweep Calibration Trace Scale Marker System Window Help .
                                                                          2450000 GHz       48.464 O
                                                                             389.27 pH      5.9924 Q
                                                                          2450000 GHz     62.710 mU
                                                                                             100.89 °




                   Ch1Avg= 20
          Ch1: Start 2.25000 GHz   ——                                                     Stop 2.65000 GHz

         .00
        00
                                                                          2450000   GHz   —24.053 dB
        .00
        .00




         00

         00
         00
         00
         00       Ch 1     =
          Chi: Start 2.25000 GHz   ——

       Status    _ CH 1:




Certificate No: D2450V2—853_Jul18                 Page 8 of 8


                                         No. I19Z60716-SEM03
                                             Page 186 of 186



ANNEX I   Accreditation Certificate




                                      ©Copyright. All rights reserved by CTTL.



Document Created: 2019-07-04 10:53:03
Document Modified: 2019-07-04 10:53:03