I19Z61572-SEM01_SAR_Rev0_4

FCC ID: 2ACCJH108

RF Exposure Info

Download: PDF
FCCID_4471791

Calibration Laboratory of                                   s\“\‘&%’z                    Schweizerischer Kalibrierdienst




                                                                                  in o
Schmid & Partner                                           in *                          Service suisse d‘étalonnage
  Engineering AG                                           i;‘aezmfis                     Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                7'/,1/fi\\xb                  Swiss Calibration Service
                                                               "ehluls



Accredited by the Swiss Accreditation Service (SAS)                                 Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
         paralle! to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1750V2—1003_Jul19                             Page 2 of 8




                                                                             w n veperg n ragen e n n nagr nwne n nene en nenn meg n n n


m..
‘/"7‘/,                                                                                              ttg s    n n >         > ~




  Measurement Conditions
      DASY system configuration, as far as not given on page 1.
        DASY Version                                              DASYS                                V52.10.2
        Extrapolation                                  Advanced Extrapolation
        Phantom                                         Modular Flat Phantom
        Distance Dipole Center — TSL                              10 mm                               with Spacer
       Zoom Scan Resolution                               dx, dy, dz =5 mm
        Frequency                                         1750 MHz # 1 MHz



  Head TSL parameters
     The following parameters and calculations were applied.
                                                                  Temperature         Permittivity           Conductivity
        Nominal Head TSL parameters                                  22.0 °C              40.1               1.37 mho/m
        Measured Head TSL parameters                              (22.0 £ 0.2) °C     40.2 + 6 %          1.34 mho/m + 6 %
        Head TSL temperature change during test                      <0.5 °C               ———                    =———

  SAR result with Head TSL

        SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
        SAR measured                                        250 mW input power                       9.01 Wikg
        SAR for nominal Head TSL parameters                    normalized to 1W            36.6 Wikg * 17.0 % (k=2)


        SAR averaged over 10 cm* (10 g) of Head TSL                 condition
        SAR measured                                        250 mW input power                       4.77 Wikg
        SAR for nominal Head TSL parameters                    normalized to 1W            19.3 Wikg * 16.5 % (k=2)



  Body TSL parameters
     The following parameters and calculations were applied.
                                                                  Temperature         Permittivity           Conductivity

        Nominal Body TSL parameters                                  22.0 °C              53.4               1.49 mho/m
        Measured Body TSL parameters                              (22.0 £ 0.2) °C      53.5 + 6 %         1.46 mho/m * 6 %
        Body TSL temperature change during test                      <0.6 °C               ~——                    ~———

  SAR result with Body TSL

        SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
        SAR measured                                        250 mW input power                       9.07 W/kg
        SAR for nominal Body TSL parameters                    normalized to 1W            36.8 Wikg * 17.0 % (k=2)

        SAR averaged over 10 cm* (10 g) of Body TSL                 condition
        SAR measured                                        250 mW input power                       4.83 Wikg
        SAR for nominal Body TSL parameters                    normalized to 1W            19.5 Wikg £ 16.5 % (k=2)



  Certificate No: D1750V2—1003_Jul19                     Page 3 of 8




                                                                                    wervopg n nee a in on rew a vewruer e cona wmz d d ies


mm©                                                                                       22
@777                                                                                                 ty s      n n >   9     n —
uo




 Appendix (Additional assessments outside the scope of SCS 0108)

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           51.0 Q + 0.1 jQ
       Return Loss                                                                       — 39.9 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           45.6 Q + 0.2 jQ
       Return Loss                                                                       —26.7 dB




 General Antenna Parameters and Design

       Electrical Delay (one direction)                                                   1.213 ns


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                    SPEAG




 Certificate No: D1750V2—1003_Jul19                       Page 4 of 8




                                                                                     wervopg n nee a in on rew a vewruer e cona wmz d d ies


DASY5 Validation Report for Head TSL

                                                                                              Date: 16.07.2019

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: £ = 1750 MHz; 0 = 1.34 S/m; s = 40.2; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.67, 8.67, 8.67) @ 1750 MHz; Calibrated: 29.05.2019

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

   e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   +    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.1 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 16.6 W/kg
SAR(I g) =9.01 W/kg; SAR(10 g) = 4.77 W/kg
Maximum value of SAR (measured) = 13.9 W/kg


       dB
        0


        —3.40


        —6.80


        —10.20


        —13.60


        —17.00
                   0 dB = 13.9 W/ikg=11.43 dBW/kg




Certificate No: D1750V2—1003_Jul19              Page 5 of 8




                                                                     Nes veperg n ragnean n en en n e n e n n n   y > 0


mm
=‘7/"/‘],                                                                                                     o



    Impedance Measurement Plot for Head TSL




        LEile    View    Channel    Sweep   Calibration Trace   Scale   Marker   System   Window   Help

                                                                                                  1.750000 GHz       51.016 0
                                                                                                      12.826 pH    141.03 m
                                                                                                , 1.750000 GHz     10.156 mU
                                                                                                                       78224 °




                         Ch1Aug= 20
                Chi: Start 1.55000 GHz   ——                                                                        Stop 1.95000 GHz


                                                                                                     50000   QHz   —




                00

                00
                00
                00
                00       Ch 1
                Ch1: Start 1                                                                                       Stop 1.95000 GHz

           Status        CH 1:     511                ] C" 1—Port                Avg=20 Delay                           LCL




    Certificate No: D1750V2—1003_Jul19                              Page 6 of 8




                                                                                                   Nn yc n >


®
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                          Date: 16.07.2019



DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.46 S/m; & = 53.5; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.45, 8.45, 8.45) @ 1750 MHz; Calibrated: 29.05.2019

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

   e    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   &    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 101.3 V/m; Power Drift =—0.06 dB
Peak SAR (extrapolated) = 16.0 W/kg
SAR(1 g) = 9.07 W/kg; SAR(10 g) = 4.83 W/kg
Maximum value of SAR (measured) = 13.6 W/kg




        —3.20


        —6.40


        —9.60


        ~12.80


        —16.00
                   0 dB = 13.6 W/kg = 11.34 dBW/kg




Certificate No: D1750V2—1003_Jul19             Page 7 of 8




                                                                    Nes perg n rag esn n mn en n e n e en n en y e s s#


Impedance Measurement Plot for Body TSL




     File    View   Channel     Sweep     Calibration   Trace   Scale Marker System   Window    Help

                                                                                               1.750000 GHz          45.607 C
                                                                                                   17.866 pH       196. 45 mQ
                                                                                         %1750000 GHz              45.992 mU
                                                                                                                       177.32°




                     Ch1Aug= 20
            Chi: Start 1.55000 GHz   ——                                                                            Stop 1.95000 GHz

                                                                                               1,j50000     QGHz   —2§. 746 dB




            00

            00
            00
            00
            00                Z
            Ch1: Start 1.55000 GHa   ——                                                                            Stop 1.95000 GHz

       Staus        CH i: _XXXE t:Por




Certificate No: D1750V2—1003_Jul19                                 Page 8 of 8




                                                                                               Nepeg n en en n n n en s n n n >


                                                                                                           es e sn
TTL



     s      .                                                        w\ We
Calibration Laboratory of                                         Q\\“\\\\_/;/"/3                         S       Schweizerischer Kalibrierdienst
Schmid & Partner                                                 ilaé//ufi}:                               c Service suisse d‘étalonnage
  Engineering AG                                                 ze                                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      * /\ y                                  S       Swiss Calibration Service
                                                                    Talutabe®

Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client      CTTL (Auden)                                                                       Certificate No: D1900V2—5d101_Jul19

CALIBRATION CERTIFICATE
 Object                               D1900V2 — SN:5d101


 Calibration procedure(s)             QA CAL—O5.v11
                                      Calibration Procedure for SAR Validation Sources between 0.7—3 GHz



 Calibration date:                    July 17, 2019


 This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                     Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                       SN: 104778                03—Apr—19 (No. 217—02892/02893)                       Apr—20
 Power sensor NRP—Z91                  SN: 103244                 03—Apr—19 (No. 217—02892)                            Apr—20
 Power sensor NRP—Z91                  SN: 103245                 03—Apr—19 (No. 217—02893)                            Apr—20
 Reference 20 dB Attenuator            SN: 5058 (20k)             04—Apr—19 (No. 217—02894)                            Apr—20
 Type—N mismatch combination           SN: 5047.2 / 06327         04—Apr—19 (No. 217—02895)                            Apr—20
 Reference Probe EX3DV4                 SN: 7349                  29—May—19 (No. EX3—7349_May19)                       May—20
 DAE4                                   SN: 601                   30—Apr—19 (No. DAE4—601_Apr19)                       Apr—20

 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Power meter E4419B                    SN: GB39512475             30—Oct—14 (in house check Feb—19)                    In house check: Oct—20
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—18)                    In house check: Oct—20
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—18)                    In house check: Oct—20
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—18)                    In house check: Oct—20
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—18)                    In house check: Oct—19


                                        Name                                    Function                                Signature
 Calibrated by:                        Michael Weber                            Laboratory Technician



 Approved by:                          KatJa Pokovic                            Technical Manager                  W



                                                                                                                       Issued: July 19, 2019
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D1900V2—5d101_Jul19                                   Page 1 of 8


3
Calibration Laboratory of
Schmid:
        & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                              )9
                                                            SA/+
                                                            il!,\_/fi:.
                                                            Neeg
                                                             z7 //;\\\ w
                                                               Trlitube®®
                                                                                     S
                                                                                     C
                                                                                     S
                                                                                            Schweizerischer Kalibrierdienst
                                                                                            Service suisse d‘étalonnage
                                                                                            Servizio svizzero di taratura
                                                                                            Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                      Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1900V2—5d101_Jul19                             Page 2 of 8




                                                                              non n ue    n nnn n n nngnn e n n en mm meg n n e e


                                                                                                Tg—    6 > _—>       .—




Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.2
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         1900 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           40.0               1.40 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   41.5 16 %          1.37 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            ~—~                     «—+—


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     9.71 W/kg
      SAR for nominal Head TSL parameters                    normalized to 1W         39.7 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                       250 mW input power                  5.12 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W         20.8 Wikg * 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C           53.3               1.52 mho/m
      Measured Body TSL parameters                             (22.0 £ 0.2) °C   54.1 +6 %          1.48 mho/m + 6 %
      Body TSL temperature change during test                     <0.5 °C

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     9.74 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         39.7 W/kg + 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     5.17 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         20.9 W/kg x 16.5 % (k=2)




Certificate No: D1900V2—5d101_Jul19                   Page 3 of 8


(Héll)°
          Appendix (Additional assessments outside the scope of SCS 0108)

          Antenna Parameters with Head TSL

                Impedance, transformed to feed point                                           50.7 Q + 4.6 jQ
                Return Loss                                                                       — 26.6 dB



          Antenna Parameters with Body TSL

                Impedance, transformed to feed point                                           46.5 Q + 6.4 jQ
                Return Loss                                                                       — 22.4 dB




          General Antenna Parameters and Design

                Electrical Delay (one direction)                                                   1.203 ns


          After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

          The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
          second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
          are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
          "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
          according to the Standard.
          No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
          feedpoint may be damaged.


          Additional EUT Data

                Manufactured by                                                                    SPEAG




          Certificate No: D1900V2—5d101_Jul19                      Page 4 of 8


                                                                                       8
                                                                                       C
                                                                                        (
DASY5 Validation Report for Head TSL
                                                                                             Date: 15.07.2019

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.37 S/m; & =41.5; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.44, 8.44, 8.44) @ 1900 MHz; Calibrated: 29.05.2019

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

   e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   +    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 108.4 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 17.6 W/kg
SAR(I g) = 9.71 W/kg; SAR(10 g) = 5.12 W/kg
Maximum value of SAR (measured) = 14.9 W/kg




        —3.40


        —6.80


        —10.20


        —13.60


        —17.00
                   0 dB = 14.9 W/kg = 11.73 dBW/kg




Certificate No: D1900V2—5d101_Jul19             Page 5 of 8




                                                                        L2     s   >       n n ne nn e y e >


TTL
‘/




Impedance Measurement Plot for Head TSL




 File    Yiew Channel Swesp         Calibration Trace   Scale Marker   System   Window Help

                                                                                       1.900000 GHz        50.681 0
                                                                                           389.25 pH       4.6469 0
                                                                                       1.900000 GHz      46.599 mU
                                                                                                            79.018 °




                 Chi1Avg= 20
        Chi: Start 1.70000 GHz    ——                                                                     Stop 2.10000 GHz



                                                                                              0000   2   —2§$.633 dB




        00

        00
        00
        00
        00        Ch 1     =
        Chi1: Start 1.70000 GHz   ———                                                                    Stop 2.10000 GHz

   Staus         CH1. B1________JCiFor                                 Avg=20 Delay                           LCL




Certificate No: D1900V2—5d101_Jul19                             Page 6 of 8




                                                                                                                            Roiny 2s


ETTL                                                                     es ce seimes




)
                                                                          ge c2 c el es



 DASY5 Validation Report for Body TSL
                                                                                     Date: 17.07.2019

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

 Communication System: UID 0 — CW; Frequency: 1900 MHz
 Medium parameters used: f = 1900 MHz; 0 = 1.48 S/m; & = 54.1; p = 1000 kg/m‘
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

 DASY52 Configuration:

    e    Probe: EX3DV4 — SN7349; ConvF(8.42, 8.42, 8.42) @ 1900 MHz; Calibrated: 29.05.2019

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 30.04.2019

    *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

    e    DASY52 52.10.2(1504); SEMCAD X 14.6.12(7470)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 104.3 V/m; Power Drift = —0.02 dB
 Peak SAR (extrapolated) = 17.3 W/kg
 SAR(1 g) = 9.74 W/kg; SAR(10 g) = 5.17 W/kg
 Maximum value of SAR (measured) = 14.7 W/kg


        dB
        0


         —4.00


         —8.00


         —12.00


         —16.00


         —20.00
                    0 dB = 14.7 W/kg = 11.67 dBW/kg




 Certificate No: D1900V2—5d101_Jul19            Page 7 of 8


                                                                                                                             exzxe on   d w id




Impedance Measurement Plot for Body TSL




 File View     Channel     Sweep   Calibration    Trace      Scale Marker   System   Window    Help

                                                                                              1.900000 GHz           46.481 O
                                                                                                  537.32 pH          6.4145 O
                                                                                              1.900000 GHz         75.664 mU
                                                                                                                      114.94 °




             Ch1Avg= 20
     Chi: Start 1.70000 GHz                                                                                       Stop 2.10000 GHz


                                                                                                      000   GHz    —2?2.422 dB




     00

     00
     00
     00
     .00
     Chi: Start 1.       He                                                                                       Stop 2.10000 GHz

  Status      CH 1:      §11                     C" 1—Port                  Avg=20 Delay                               LCL




Certificate No: D1900V2—5d101_Jul19                                  Page 8 of 8




                                                                                                                  snn ngeee ce > mssn m se es


     i      4                                                      svlgn
Calibration
      .     Laboratory of                                        a° J /y/s
                                                                 fam                                        S Schweizerischer Kalibrierdienst




                                                                              3
Schmid & Partner                                                 ila%                                       c Service suisse d‘étalonnage
  Engineering AG                                                  uge S                                            Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       &4/ mnN<
                                                                        *\ \&                               S      Swiss Calibration Service
                                                                     YAnlotabs®

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CTTL (Auden)                                                                       Certificate No: D2450V2—853_Jul19

CALIBRATION CERTIFICATE
 Object                                D2450V2 — SN:853


 Calibration procedure(s)              QA CAL—O5.v11
                                       Calibration Procedure for SAR Validation Sources between 0.7—3 GHz



 Calibration date:                     July 17, 2019

 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                03—Apr—19 (No. 217—02892/02893)                       Apr—20
 Power sensor NRP—Z91                   SN: 103244                03—Apr—19 (No. 217—02892)                             Apr—20
 Power sensor NRP—Z91                   SN: 108245                03—Apr—19 (No. 217—02893)                             Apr—20
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—19 (No. 217—02894)                             Apr—20
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—19 (No. 217—02895)                             Apr—20
 Reference Probe EX3DV4                 SN: 7349                  29—May—19 (No. EX3—7349_May19)                        May—20
 DAE4                                   SN: 601                   30—Apr—19 (No. DAE4—601_Apr19)                        Apr—20


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter E4419B                     SN: GB39512475            30—Oct—14 (in house check Feb—19)                     In house check: Oct—20
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—18)                     In house check: Oct—20
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—18)                     In house check: Oct—20
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—18)                     In house check: Oct—20
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—18)                     In house check: Oct—19


                                        Name                                      Function                              Signature
 Calibrated by:                         Michael Weber                             Laboratory Technician



 Approved by:                           Katia Pokovic                             Technical Manager



                                                                                                                       Issued: July 17, 2019
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D2450V2—853_Jul19                                     Page 1 of 8



Document Created: 2019-09-29 09:39:46
Document Modified: 2019-09-29 09:39:46

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC