I19Z60710-SEM01_SAR_Rev2_part4

FCC ID: 2ACCJH103

RF Exposure Info

Download: PDF
FCCID_4317689

Impedance Measurement Plot for Body TSL




          File    View   Channel    Sweep   Calbration    Trace Scale Marker    System Window    Help

                                                                                                              46.118 O
                                                                                                             —5£1978 0
                                                                                                            67.394 mU
                                                                                                              ~123.56 °




                          Ch1Aug= 20
                 Ch1: Start §35.000 MHz   ——                                                                Stop 1.03500 GHz

                                                                                               835.000000   —2$.428 dB




                 o0       Ch 1     =
                 Ch1: Start €35.000 MHe   ———

            Status       CH 1:     511                   | C*1—Port             Avg=20 Delay                     LCL




Certificate No: D835V2—4d0069_Jul18                               Page 8 of 8


Calibration Laboratory of
Schmid & Partner                                                                                            S      Schweizerischer Kalibrierdienst
                                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                            C      Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                S      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CTTL (Auden)                                                                       Certificate No: D1750V2—1003_Jul18

CALIBRATION CERTIFICATE
 Object                                D1750V2 — SN:1003

 Calibration procedure(s)              QA CAL—05.v10
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     July 20, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical forcalibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                   Function                                 Signature
 Calibrated by:                         Manu Seitz                             Laboratory Technician                                +



 Approved by:                           Katia Pokovic                          Technical Manager

                                                                                                                   es
                                                                                                                       Issued: July 20, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1750V2—1003_Jul18                                    Page 1 of 8


uy                                                                                      C



   a    "                                                     so
ca"b"'atlon Laboratory of                                  s\‘\t\_.///”/,;   G   Schweizerischer Kalibrierdienst
Schmid & Partner                                           M                 c   Service suisse d‘étalonnage
  Engineering AG                                           zeol                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                */7/\\\\\\\*     S   Swiss Calibration Service
                                                               ulitas

Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) 1EC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
   e SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D1750V2—1003_Jul18                             Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     .Frequency                                        1750 MHz 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           40.1               1.37 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   39.0 26 %          1.34 mho/m #6 %
     Head TSL temperature change during test                      <0.5 °C                                    ——

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     8.91 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        35.9 W/kg x 17.0 %(k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                 4.71 Wikg
     .SAR for nominal Head TSL parameters                    normalized to 1W        18.9 Wikg 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C           53.4               1.49 mho/m
      Measured Body TSL parameters                             (22.0 + 0.2) °C   53.7 46 %          1.46 mho/m £ 6 %
      Body TSL temperature change during test                     <0.5 °C

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     8.97 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         36.4 Wikg a 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     4.79 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W         19.3 W/kg x 16.5 % (k=2)




Certificate No: D1750V2—1003_Jul18                    Page 3 of 8


4(S/
  Appendix (Additional assessments outside the scope of SCS 0108)

  Antenna Parameters with Head TSL
        Impedance, transformed to feed point                                           51.6 0 +1.3jQ
       . Return Loss                                                                      — 33.7 dB


  Antenna Parameters with Body TSL

        Impedance, transformed to feed point                                           47.3 Q + 1.2 jQ
        Return Loss                                                                      — 80.4 dB




  General Antenna Parameters and Design

       |Electrical Delay (one direction)                        |                         1.215 ns
  After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

  The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
  second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
  are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
  "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
  according to the Standard.
  No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
  feedpoint may be damaged.


  Additional EUT Data

        Manufactured by                                     °                              SPEAG
        Manufactured on                                                                 July 30, 2008




  Certificate No: D1750V2—1003_Jul18                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                     Date: 20.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.34 S/m; & = 39; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(8.5, 8.5, 8.5) @ 1750 MHz; Calibrated: 30.12.2017

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   +    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.5 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 16.3 W/kg
SAR(1 g) = 8.91 W/kg; SAR(10 g) = 4.71 W/kg
Maximum value of SAR (measured) = 13.8 W/kg

       dB
       0


        —4.00


        —8.00


        —12.00


        —16.00


        —20.00
                   0 dB = 13.8 W/kg = 11.40 dBW/kg




Certificate No: D1750V2—1003_Jul18              Page 5 of 8


©
Impedance Measurement Plot for Head TSL




          File    View Channel Sweep Calibration Trace Scale   Marker System Window   Help

                                                                                      1.750000 GHz      51.596 O
                                                                                          122.55 pH     13475 Q
                                                                                      1.750000 GHz    20.559 mU
                                                                                                         39.410 °




                          Ch1Aug= 20
                 Chi: Start 1.55000 GHz   ——                                                          Stop 1.95000 GHz




                 00

                 00
                 00
                 00
                 00       Ch 1
                 Ch1: Start 1.

            Status        CH 1:                                       Avg=20 Delay




Certificate No: D1750V2—1003_Jul18                     Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                    Date: 20.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.46 S/m; &, = 53.7; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(8.35, 8.35, 8.35) @ 1750 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 101.3 V/m; Power Drift =—0.01 dB
Peak SAR (extrapolated) = 15.7 W/kg
SAR(I g) = 8.97 W/kg; SAR(10 g) = 4.79 W/kg
Maximum value of SAR (measured) = 13.5 W/kg


       dB
        0


        —4.00


        —8.00


        —12.00


        —16.00


        —20.00
                   0 dB = 13.5 W/kg = 11.30 dBW/kg




Certificate No: D1750V2—1003_Jul18             Page 7 of 8


©
Impedance Measurement Plot for Body TSL




          File    View   Channel     Sweep   Calibration Trace Scale Marker   System Window    Help

                                                                                               1.750000 GHz      47.338 O
                                                                                                   112.37 pH     1.2356 O
                                                                                             y 11750000 GHz    30.149 mU
                                                                                                                  154.37 °




                          Ch1Avg= 20
                 Ch1: Start 1.55000 GHa   ——                                                                   Stop 1.95000 GHz




                 .00

                 00
                 .00
                 00
                 00                &
                 Ch1: Start 1.55000 GHa   ——                                                                   Stop 1.95000 GHz

            Status       CH 1:                                                Avg=20 Delay




Certificate No: D1750V2—1003_Jul18                            Page 8 of 8


     s       3                                                       «o
ca"bratlon Laboratory of                                           S\‘\\\\///’z}                            S      Schweizerischer Kalibrierdienst
Schmid & Partner                                                  M                                         c      Service suisse d‘étalonnage
  Englneerlng AG                                                  * sg                                             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       f'/,,Afi\“\?                              S      Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       CTTL (Auden)                                                                       Certificate No: D1900V2—5d101_Jul18

CALIBRATION CERTIFICATE
 Object                                D1900V2 — SN:5d101


 Calibration procedure(s)              QA CAL—05.v10
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     July 24, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                      Apr—19
 Power sensor NRP—Z291                  SN: 103244                04—Apr—18 (No. 217—02672)                            Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                            Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                       Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                   Function                                 Signature
 Calibrated by:                         Manu Seitz                             Laboratory Technician



 Approved by:                           Katja Pokovic                          Technical Manager                   /C,/@



                                                                                                                       Issued: July 24, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1900V2—5d101_Jul18                                   Page 1 of 8


uy




CallbI"atIOH Laboratory of                                   ¢\“\\L\I//I'7’7¢   g   Schweizerischer Kalibrierdienst
Schmid & Partner                                            ;ia\%\fl_%é          c   Service suisse d‘étalonnage
  Engineering AG                                            uzol                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 7'/,,//fi\“\?       S   swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
   * SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D1900V2—5d101_Jul18                             Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                              V52.10.1
      Extrapolation                                   Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                         1900 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
      Nominal Head TSL parameters                                  22.0 °C           40.0               1.40 mho/m
      Measured Head TSL parameters                              (22.0 + 0.2) °C   39.9 £ 6 %         1.34 mho/m +6 %
      Head TSL temperature change during test                      <0.5 °C                                    ————


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                       250 mW input power                      9.84 Wikg
      SAR for nominal Head TSL parameters                    normalized to 1W         40.4 Wikg # 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                       250 mW input power                      5.23 W/kg
      SAR for nominal Head TSL parameters                    normalized to 1W         21.3 Wikg #16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                  22.0 °C           53.3               1.52 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   54.3 £6 %          1.46 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C            ~———                    w


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                     9.83 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         40.4 Wikg x 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                       250 mW input power                      5.26 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         21.4 Wikg x 16.5 % (k=2)




Certificate No: D1900V2—5d101_Jul18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           50.7 Q + 5.3 jQ
      Return Loss                                                                       —25.6 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           44.9 Q + 6.5 jQ
      Return Loss                                                                      —21.2 dB




General Antenna Parameters and Design

     I Electrical Delay (one direction)                                                 1.205 ns                        J

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                March 28, 2008




Certificate No: D1900V2—5d101_Jul18                      Page 4 of 8


a
N
(I
     DASY5 Validation Report for Head TSL

                                                                                         Date: 24.07.2018

     Test Laboratory: SPEAG, Zurich, Switzerland

     DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

     Communication System: UID 0 — CW; Frequency: 1900 MHz
     Medium parameters used: f = 1900 MHz; 0 = 1.34 S/m; & = 39.9; p = 1000 kg/m*
     Phantom section: Flat Section
     Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

     DASY52 Configuration:

         *   Probe: EX3DV4 — SN7349; ConvF(8.18, 8.18, 8.18) @ 1900 MHz; Calibrated: 30.12.2017

         *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         *   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

         e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

         *   DASY32 52.10.1(1476); SEMCAD X 14.6.11(7439)


     Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
     Measurement grid: dx=5mm, dy=5mm, dz=5mm
     Reference Value = 110.6 V/m; Power Drift =—0.01 dB
     Peak SAR (extrapolated) = 17.9 W/kg
     SAR(1 g) = 9.84 W/kg; SAR(10 g) = 5.23 W/kg
     Maximum value of SAR (measured) = 15.0 W/kg




                                  12.00


                                  +16.00


                                  —20.00

                                           0 dB = 15.0 W/kg = 11.76 dBW/kg




     Certificate No: D1900V2—5d101_Jul18                  Page 5 of 8


®
Impedance Measurement Plot for Head TSL




          File    View    Channel    Sweep Calibration Trace Scale Marker System Window   Help

                                                                                        1.900000 GHz      50.660 O
                                                                                            441.05 pH     5.2653 O
                                                                                      , 1.900000 GHz    52.645 mU
                                                                                                           79.866 °




                          Ch1Aug= 20
                 Ch1: Start 1.70000 GHz   ——                                                            Stop 2.10000 GHz


                                                                                            00000 GHz   —2$.573 dB




                 00       Ch 1 Avg= |20
                 Ch1: Start 1.70000 GHa   ———                                                           Stop 2.10000 GHz




Certificate No: D1900V2—5d101_Jul18                        Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                    Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.46 S/m; & = 54.3; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.15, 8.15, 8.15) @ 1900 MHz; Calibrated: 30.12.2017

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   &    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.3 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.83 W/kg; SAR(10 g) = 5.26 W/kg
Maximum value of SAR (measured) = 14.8 W/kg




                            12.00


                            —16.00


                            —20.00
                                      0 dB = 14.8 W/kg = 11.70 dBW/kg




Certificate No: D1900V2—5d101_Jul18                  Page 7 of 8



Document Created: 2019-06-13 18:16:58