I18Z61172-SEM01_SAR_Rev0_part4

FCC ID: 2ACCJH094

RF Exposure Info

Download: PDF
FCCID_3946044

)
n
S
                                                                                                    C


   a    s                                                     you 1
Callbratlon Laboratory of                                   s\‘\t\_////‘/,;             S     Schweizerischer Kalibrierdienst
Schm[d & Partner                                           m                            c     Service suisse d‘étalonnage
  Engmeenng AG                                             T ong                              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                {f/,,/I//_\\\\‘\\‘}         S     Swiss Calibration Service
                                                                rhiitabe

Accredited by the Swiss Accreditation Service (SAS)                                      Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
         300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
         SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1750V2—1003_Jul17                               Page 2 of 8




                                                                                  Ns veperg n ragn en n en n en en n e n n n   y >


)
@777                                                                                          1 l
uo




 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                               V52.10.0
      Extrapolation                                   Advanced Extrapolation
      Phantom                                         Modular Flat Phantom

      Distance Dipole Center — TSL                              10 mm                              with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                         1750 MHz & 1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature        Permittivity         Conductivity
      Nominal Head TSL parameters                                  22.0 °C             40.1              1.37 mho/m

      Measured Head TSL parameters                              (22.0 + 0.2) °C     39.0 + 6 %        1.35 mho/m £ 6 %

      Head TSL temperature change during test                      <0.5 °C                                     ———

 SAR result with Head TSL

      SAR averaged over 1 cm‘ (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                      9.15 W/kg
      SAR for nominal Head TSL parameters                     normalized to 1W          36.7 Wikg * 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                      4.84 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W          19.4 Wikg £ 16.5 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature        Permittivity         Conductivity

       Nominal Body TSL parameters                                 22.0 °C             58.4              1.49 mho/m
       Measured Body TSL parameters                             (22.0 £ 0.2) °C     53.3 26 %         1.49 mho/m £ 6 %
       Body TSL temperature change during test                     <0.5 °C                                     ———

 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
       SAR measured                                       250 mW input power                      9.29 W/kg
       SAR for nominal Body TSL parameters                    normalized to 1W          37.1 Wikg + 17.0 % (k=2)

       SAR averaged over 10 cm* (10 g) of Body TSL                condition
       SAR measured                                       250 mW input power                   4.94 Wikg
       SAR for nominal Body TSL parameters                    normalized to 1W          19.8 Wikg * 16.5 % (k=2)



 Certificate No: D1750V2—1003_Juf17                    Page 3 of 8




                                                                                  Nn epeg n n en n n n en c ++ >


(fléN)
        Appendix (Additional assessments outside the scope of SCS 0108)

        Antenna Parameters with Head TSL

              Impedance, transformed to feed point                                          50.9 Q + 1.1 jQ
              Return Loss                                                                      —37.1 dB


        Antenna Parameters with Body TSL

              Impedance, transformed to feed point                                          47.0 Q + 0.1 jQ
              Return Loss                                                                      — 30.2 dB




        General Antenna Parameters and Design

            | Electrical Delay (one direction)                                                  1.213 ns
        After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

        The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
        second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
        are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
        "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
        according to the Standard.
        No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
        feedpoint may be damaged.



        Additional EUT Data

              Manufactured by                                                                    SPEAG
              Manufactured on                                                                July 30, 2008




        Certificate No: D1750V2—1003_Jul17                       Page 4 of 8




                                                                                            ~nnmuan n n en c n c ogn on >


DASY5 Validation Report for Head TSL

                                                                                        Date: 21.07.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f= 1750 MHz; 0 = 1.35 S/m; & = 39; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(8.73, 8.73, 8.73); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.4 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 17.0 W/kg
SAR(1 g) = 9.15 W/kg; SAR(10 g) = 4.84 W/kg
Maximum value of SAR (measured) = 14.0 W/kg


        dB


        —3.60
        —£.20
        —10.80
        —14.40
        —18.00
                   0 dB = 14.0 W/kg = 11.46 dBW/kg




Certificate No: D1750V2—1003_Jul17              Page 5 of 8




                                                                     Nspeg n snn n n n en n n n e n en   ny   >


Impedance Measurement Plot for Head TSL




                                                                    21 Jul 2017 11:01:47?
          CHI) sii     i u Fs          1: 50.939 a    1.0566 a   96.097 pH      1 750.000 0oo MHz
                                               ,"/‘              ‘\‘-




          Av
          16 *


          H1d

          CH2




          Ca




          Av
          16°

          H1d

                     START 1 550.000 @00 MHz                              STOP 1 950.000 000 MHz




Certificate No: D1750V2—1003_Jul17                      Page 6 of 8




                                                                                 Ns epgrge es


DASY5 Validation Report for Body TSL

                                                                                           Date: 20.07.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: £= 1750 MHz; o = 1.49 S/m; & = 53.3; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(8.46, 8.46, 8.46); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 99.34 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 16.4 W/kg
SAR(1 g) = 9.29 W/kg; SAR(10 g) = 4.94 W/kg
Maximum value of SAR (measured) = 13.4 W/kg

        dB
        0

        —3.40


        —6.80

        —10.20

        —13.60


        —17.00
                   0 dB = 13.4 W/kg = 11.27 dBW/kg




Certificate No: D1750V2—1003_Jul17             Page 7 of 8




                                                                     Nnpeg n n en n en n en n n n ns n n n   ny   >


)
(IIII
        Impedance Measurement Plot for Body TSL




                                                                          20 Jul 2017 14137137
                 CHI sii      1 U FS          1: 46.994 a   0,0703 2   6.3944 pH      1 750.000 @00 MHz




                 Av
                 159


                 H1d

                  CH2




                  Ca




                  Av
                  16°

                  H1d

                            START 1 550.000 000 MHz                             STOP 1 950.000 000 MHz




        Certificate No: D1750V2—1003_Jul17                    Page 8 of 8




                                                                                        Nn y rg nn m nge>


—( )
m
~Calibration Laboratory of                                                                                           Schweizerischer Kalibrierdienst




                                                                                                            n 0
 Schmid & Partner                                                                                                    Service suisse d‘étalonnage
   Engineering AG                                                                                                    Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                        Swiss Calibration Service



  Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multilateral Agreementfor the recognition of calibration certificates

  Client      CTTL—BJ (Auden)                                                                    Certificate No: D1900V2—5d101_Jul17

  CALIBRATION CERTIFICATE
  Object                                D1900V2 — SN:5d101


   Calibration procedure(s)             QA CAL—05.v9
                                        Calibration procedure for dipole validation kits above 700 MHz



   Calibration date:                    July 26, 2017


   This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
   The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


   All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


   Calibration Equipment used (M&TE critical for calibration)


   Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
   Power meter NRP                        SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
   Power sensor NRP—Z91                   SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
   Power sensor NRP—Z91                   SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
   Reference 20 dB Attenuator             SN: 5058 (20k)            07—Apr—17 (No. 217—02528)                             Apr—18
   Type—N mismatch combination            SN: 5047.2 / 06327        O7—Apr—17 (No. 217—02529)                             Apr—18
   Reference Probe EX3DV4                 SN: 7349                  31—May—17 (No. EX8—7349_May17)                        May—18
   DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


   Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
   Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
   Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                          Name                                   Function
   Calibrated by:                         Johannes Kurikka                       Laboratory Technician



   Approved by:                           Katia Pokovic                          Technical Manager

                                                                                                                      ePX Issued: July 26, 2017
   This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


  Certificate No: D1900V2—5d101_Jul17                                   Page 1 of 8


m©
777
T.                                                                                             ty —            200




         s      x                                                   al m
     Callbll'atlon Laboratory of                                  Q\\‘L\\:////’/,;   S   Schweizerischer Kalibrierdienst
     Schmid & Partner                                            ;Ia\&_/fl            c   Service suisse d‘étalonnage
       Engineering AG                                            ;//—/%\3                Servizio svizzero di taratura
     Zeughausstrasse 43, 8004 Zurich, Switzerland                 *g/fi‘\\\\*         S   Swiss Calibration Service
                                                                     "Whilal?



     Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
     The Swiss Accreditation Service is one of the signatories to the EA
     Multilateral Agreement for the recognition of calibration certificates

     Glossary:
     TSL                          tissue simulating liquid
     ConvrF                       sensitivity in TSL / NORM x,y,z
     N/A                          not applicable or not measured

     Calibration is Performed According to the Following Standards:
        a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
           Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
           Communications Devices: Measurement Techniques", June 2013
        b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
            (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
            300 MHz to 6 GHz)", July 2016
        c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
            communication devices used in close proximity to the human body (frequency range of 30
            MHz to 6 GHz)", March 2010
        d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
     Additional Documentation:
         e) DASY4/5 System Handbook

     Methods Applied and Interpretation of Parameters:
          Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
          Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
          Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
          Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
               connector.
               SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
               nominal SAR result.
       The reported uncertainty of measurement is stated as the standard uncertainty of measurement
       multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
       probability of approximately 95%.




     Certificate No: D1900V2—5d101_Jul17                              Page 2 of 8


(EM)
                                                                                                          w




       Measurement Conditions
          DASY system configuration, as far as not given on page 1.
            DASY Version                                              DASY5                              V52.10.0
            Extrapolation                                   Advanced Extrapolation
            Phantom                                         Modular Flat Phantom
            Distance Dipole Center — TSL                              10 mm                             with Spacer
            Zoom Scan Resolution                              dx, dy, dz =5 mm
            Frequency                                         1900 MHz £ 1 MHz



       Head TSL parameters
          The following parameters and calculations were applied.
                                                                      Temperature       Permittivity           Conductivity
            Nominal Head TSL parameters                                  22.0 °C           40.0                 1.40 mho/m
            Measured Head TSL parameters                              (22.0 + 0.2) °C   40.7 #6 %             1.39 mho/m + 6 %
            Head TSL temperature change during test                      <0.5 °C            ~——                     ———

       SAR result with Head TSL

            SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
            SAR measured                                        250 mW input power                     9.93 W/kg
            SAR for nominal Head TSL parameters                     normalized to 1W         40.0 W/kg + 17.0 % (k=2)


            SAR averaged over 10 cm* (10 g) of Head TSL                 condition
            SAR measured                                        250 mW input power                     5.23 W/kg
            SAR for nominal Head TSL parameters                     normalized to 1W         21.0 W/kg + 16.5 % (k=2)


       Body TSL parameters
          The following parameters and calculations were applied.
                                                                      Temperature       Permittivity            Conductivity
            Nominal Body TSL parameters                                   22.0 °C           53.3                 1.52 mho/m
            Measured Body TSL parameters                              (22.0 £0.2) °C    54.1 #6 %             1.50 mho/m « 6 %
            Body TSL temperature change during test                      <0.5 °C            ——


       SAR result with Body TSL

            SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
             SAR measured                                       250 mW input power                     10.0 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         40.5 Wikg + 17.0 % (k=2)

             SAR averaged over 10 cm‘ (10 g) of Body TSL                condition
             SAR measured                                       250 mW input power                     5.33 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         21.5 Wikg * 16.5 % (k=2)



       Certificate No: D1900V2—5d101_Jul17                   Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          51.7 Q + 5.8 jQ
      Return Loss                                                                      — 24.5 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          46.2 Q + 6.6 jQ
      Return Loss                                                                      —22.0 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1,203 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   SPEAG
      Manufactured on                                                               March 28, 2008




Certificate No: D1900V2—5d101_Jul17                      Page 4 of 8


(Iulm?
   N
         S
     DASY5 Validation Report for Head TSL



     Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                           Date: 26.07.2017



     DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

     Communication System: UID 0 — CW; Frequency: 1900 MHz
     Medium parameters used: £ = 1900 MHz; 0 = 1.39 S/m; & = 40.7; p = 1000 kg/m3
     Phantom section: Flat Section
     Measurement Standard: DASYS5 (IEEE/IEC/ANSI C€63.19—2011)

     DASY52 Configuration:

            e   Probe: EX3DV4 — SN7349; ConvF(8.43, 8.43, 8.43); Calibrated: 31.05.2017;

            e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

            e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

            e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

            &   DASY32 52.10.0(1446); SEMCAD X 14.6.10(7417)


     Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
         Measurement grid: dx=5mm, dy=5mm, dz=5mm
         Reference Value = 106.3 V/m; Power Drift = —0.04 dB
         Peak SAR (extrapolated) = 18.4 W/kg
         SAR(1 g) = 9.93 W/kg; SAR(10 g) = 5.23 W/kg
         Maximum value of SAR (measured) = 14.9 W/kg




                —3.60

                —7.20

                 —10.80

                 ~14.40

                 —18.00
                            0 dB = 14.9 W/kg = 11.73 dBW/kg




         Certificate No: D1900V2—5d101_Jul17            Page 5 of 8


)
~
~
(I
     Impedance Measurement Plot for Head TSL




                                                                     26 Jul 2eir os:s2:00
              CH] sin      i uors          4t 51.727 a   5.8418 0 489.34 pH      1 900.000 oog MHz
                                                         ~~T




              Av
              16 *


              H1d

               CH2




              Ca




              Av
              16°

               H1d

                         START 1 700.000 @00 MHz                           STOP 2 100.000 000 MHz




     Certificate No: D1900V2—5d101_Jul17                   Page 6 of 8


(Ilflll,
           S
          DASY5 Validation Report for Body TSL



      Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                         Date: 26.07.2017



      DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

          Communication System: UID 0 — CW; Frequency: 1900 MHz
          Medium parameters used: f = 1900 MHz; 0 = 1.5 S/m; & = 54.1; p = 1000 kg/m*
          Phantom section: Flat Section
          Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

          DASY52 Configuration:

             e   Probe: EX3DV4 — SN7349; ConvF(8.2, 8.2, 8.2); Calibrated: 31.05.2017;

             e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

             e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

             e   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

             e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


          Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
          Measurement grid: dx=5mm, dy=5mm, dz=5mm
          Reference Value = 101.8 V/m; Power Drift = —0.09 dB
          Peak SAR (extrapolated) = 17.6 W/kg
          SAR(1 g) = 10 W/kg; SAR(10 g) = 5.33 W/kg
          Maximum value of SAR (measured) = 14.4 W/kg


                 dB
                 0

                 —3.40

                 —6.80

                  —10.20

                  —13.60

                  —17.00
                             0 dB = 14.4 W/kg = 11.58 dBW/kg




          Certificate No: D1900V2—5d101_Jul17            Page 7 of 8


@TTL


)
ue




 Impedance Measurement Plot for Body TSL




                                                               26 Jul 2017    11:15:02
          CHI sii      i uo ors        1: 46.164 o 6.5918 0 55247 pH         1 900.000 000 MHz
                                                 mpe




                     START 1 700.000 000 MHz                         STOP 2 100.000 000 MHz




 Certificate No: D1900V2—5d101_Jul17                 Page 8 of 8


«
n es   e en n n nge n         e ns




         Calibration Laboratory of                                             wl n
                                                                           Sasy"/»                                  S       Schweizerischer Kalibrierdienst




                                                                       §               8
         Schmid & Partner




                                                                                       S
                                                                          4
                                                                                                                    C       Service suisse d‘étalonnage
           Engineering AG                                                                                                   Servizio svizzero di taratura




                                                                                     \>>
                                                                                     4”:,/
                                                                         i
         Zeughausstrasse 43, 8004 Zurich, Switzerland                                                               S       swiss Calibration Service
                                                                              Aulutuls®

         Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
         The Swiss Accreditation Service is one of the signatories to the EA
         Multilateral Agreement for the recognition of calibration certificates

         Client      CTTL—BJ (Auden)                                                                    Certificate No: D2450V2—853_Jul17

         CALIBRATION CERTIFICATE
          Object                               D2450V2 — SN:853


          Calibration procedure(s)             QA CAL—05.v9
                                               Calibration procedure for dipole validation kits above 700 MHz



          Calibration date:                    July 21, 2017

          This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
          The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


          All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity <70%.


          Calibration Equipment used (M&TE critical for calibration)


          Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
          Power meter NRP                        SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
          Power sensor NRP—Z91                   SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
          Power sensor NRP—Z91                   SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
          Reference 20 dB Attenuator             SN: 5058 (20k)            O7—Apr—17 (No. 217—02528)                             Apr—18
          Type—N mismatch combination            SN: 5047.2 / 06327        O7—Apr—17 (No. 217—02529)                             Apr—18
          Reference Probe EX3DV4                 SN: 7349                  31—May—17 (No. EX3—7349_May17)                        May—18
          DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


          Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
          Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
          RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
          Network Analyzer HP 8758E              SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                                 Name                                     Function                                Signature
          Calibrated by:                         Michael Weber                            Laboratory Technician
                                                                                                                                    e


          Approved by:                           KatJa Pokovic                            Technical Manager

                                                                                                                            DERE
                                                                                                                                 Issued: July 24, 2017
          This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


         Certificate No: D2450V2—853_Jul17                                      Page 1 of 8


    +      #                                                  uon
Callbratlon Laboratory of                                   i\‘\\w_&/’n}      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           M                  c   Service suisse d‘étalonnage
  Engineering AG                                           22—R                   Servizio svizzero di taratura
                              £       *                    y n
Zeughausstrasse 43, 8004 Zurich, Switzerland               *g/7/\\\\‘\\*      S   Swiss Calibration Service
                                                              thilals

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
      e   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
      e   SAR measured: SAR measured at the stated antenna input power.
      e   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
      e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—853_Jul17                               Page 2 of 8



Document Created: 2018-07-27 14:33:15
Document Modified: 2018-07-27 14:33:15

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC