I19Z61162-SEM01_SAR_Rev1_8

FCC ID: 2ACCJH089

RF Exposure Info

Download: PDF
FCCID_4376076

                                                                              ow n   e se se k k Nes         N n en w




    P      s                                                    w4 uj
Callbl.'atlon Laboratory of                                  Q‘QM?Z           S   Schweizerischer Kalibrierdienst
Schmid & Partner                                            m                 c   Service suisse d‘étalonnage
  Engineering AG                                            s o amls              Servizio svizzero di taratura
                             ;      M
Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                            x4rosng
                                                               fan &
                                                                              S      i      iFSH          M
                                                                                  Swiss Calibration Service
                                                              Hlilals

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul18                              Page 2 of 8


3)
m
                                                                                                   ~<g > —        l >    —




    Measurement Conditions
       DASY system configuration, as far as not given on page 1.
         DASY Version                                              DASY5                             V52.10.1
         Extrapolation                                   Advanced Extrapolation
         Phantom                                         Modular Flat Phantom
         Distance Dipole Center — TSL                              10 mm                            with Spacer
         Zoom Scan Resolution                              dx, dy, dz =5 mm
         Frequency                                         2600 MHz + 1 MHz



    Head TSL parameters
       The following parameters and calculations were applied.
                                                                   Temperature      Permittivity          Conductivity
         Nominal Head TSL parameters                                  22.0 °C          39.0                1.96 mho/m
         Measured Head TSL parameters                              (22.0 £0.2) °C   37.2 26 %           2.02 mho/m +6 %
         Head TSL temperature change during test                      <0.5 °C           ——                      ——


    SAR result with Head TSL

         SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
         SAR measured                                        250 mW input power                    14.2 W/ikg
         SAR for nominal Head TSL parameters                     normalized to 1W       55.4 Wikg + 17.0 % (k=2)

         SAR averaged over 10 cm* (10 g) of Head TSL                condition
          SAR measured                                       250 mW input power                    6.33 W/kg
          SAR for nominal Head TSL parameters                    normalized to 1W       24.9 Wikg x 16.5 % (k=2)


    Body TSL parameters
       The following parameters and calculations were applied.                                                           .
                                                                   Temperature      Permittivity          Conductivity
          Nominal Body TSL parameters                                 22.0 °C          52.5                2.16 mho/m
          Measured Body TSL parameters                             (22.0 £0.2) °C   51.5 26 %           2.20 mho/m £ 6 %
          Body TSL temperature change during test                     <0.5 °C           =—

    SAR result with Body TSL

          SAR averaged over 1 cm‘ (1 g) of Body TSL                 Condition
          SAR measured                                       250 mW input power                    13.7 W/kg
          SAR for nominal Body TSL parameters                    normalized to 1W        54.1 Wikg #17.0 % (k=2)

          SAR averaged over 10 em* (10 g) of Body TSL                condition
          SAR measured                                       250 mW input power                 6.17 Wikg
          SAR for nominal Body TSL parameters                    normalized to 1W        24.5 Wikg # 16.5 % (k=2)



    Certificate No: D2600V2—1012_Jul18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 01 08)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           474 Q —7.4 jQ
      Return Loss                                                                       —21.9 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           44.1 Q — 4.9 jQ
      Return Loss                                                                      —21.8 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.154 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               October 30, 2007




Certificate No: D2600V2—1012_Jul18                        Page 4 of 8


mm>                                                                       De e e
/7/
—                                                                               t   e t i n n   n>




DASY5 Validation Report for Head TSL

                                                                                        Date: 26.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.02 S/m; & = 37.2; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

    e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    e    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    *    DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 118.3 V/m; Power Drift = —0.04 dB
Peak SAR (extrapolated) = 28.3 W/kg
SAR(1 g) = 14.2 W/kg; SAR(10 g) = 6.33 W/kg
Maximum value of SAR (measured) = 23.7 W/kg

        dB
        0

         —5.20

         —10.40

         —15.60

         —20.80

         —26.00
                    0 dB = 23.7 W/kg = 13.75 dBW/kg




 Certificate No: D2600V2—1012_Jul18              Page 5 of 8


Impedance Measurement Plot for Head TSL




File View      Channel Sweep         Calibration Trace Scale   Marker   System   Window   Help

                                                                                          2600000 GHz       47.371 Q
                                                                                              8.2358 pF    —7.4326 Q
                                                                                       \, 2600000 GHz     80.734 mU
                                                                                                            ~105.12 °




              Ch1Avg= 20
     Ch1: Start 2.40000 GHz                                                                               Stop 2.90000 GHz


                                                                                                          —2).859 dB




   35.00
      00       Ch       =
      Ch1: Start 2.40000 GHa   ———                                                                        Stop 2.20000 GHz

  Status      CH 1:    en            Te ira                             Avg=20 Delay                           LCL




Certificate No: D2600V2—1012_Jul18                               Page 6 of 8


8
DASY5 Validation Report for Body TSL



Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1012
                                                                                    Date: 26.07.2018




Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.2 S/m; & = 51.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   +    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.5 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 27.7 W/kg
SAR(I g) = 13.7 W/kg; SAR(10 g) = 6.17 W/kg
Maximum value of SAR (measured) = 22.6 W/kg


        dB
         0

        —5.20

        ~10.40

        —15.60

        —20.80

        —26.00
                   0 dB = 22.6 W/kg = 13.54 dBW/kg




Certificate No: D2600V2—1012_Jul18             Page 7 of 8


(IEI)
        Impedance Measurement Plot for Body TSL




         File         View    Channel    Sweep   Calibration    Trace    Scale   Marker   System   Window    Help

                                                                                                            2600000 GHz            44.132 O
                                                                                                                    12.374 pF     —4.9469 C
                                                                                                         y 2600000 GHz          81.418 mU
                                                                                                                                  —136.86 °




                          Ch1Avg= 20
                 Ch1: Start 2.40000 GHz       ——                                                                                Stop 2.80000 GHz

                 00
                                                                                                                                —21.786 dB
                00
           0.00




                 00       Ch1
                 Ch1: Start 2.                                                                                                  Stop 2.90000 GHz

           Status            CH 1:      E11                    | C" 1—Port                Avg=20 Delay                               LCL




        Certificate No: D2600V2—1012_Jul18                                         Page 8 of 8


We n ae   se k ns   wever im n we en uen n   ow w n n n n e in s —




                                                                                                 ap “
                                                                                                 "/A
                                                                                               i&\\   *E“T
                                                        In Collaboration with
                                                                                                                                     es
                            Afi

                            ©_     2j
                            ‘T];l; CALIBRATION LABORATORY                                                          CNAS
                                                                                                                      Bft




                                                                                                               h
                                                                                               \\\‘“




                                                                                                            l,/’
                                                                                                   /-\                               CALIBRATION
                         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China         l/"l'‘\‘\\
                                                                                                         1                           CNAS LO570
                         Tel: +86—10—62304633—2512            Fax: +86—10—62304633—2504                t3
                         E—mail: cttl@chinattl.com            Hitp:/www.chinattl.on
                Client :            CTTL                                                      Certificate No: 218—60359
          CALIBRATION CERTIFICATE

          Object                                           DAE4 — SN: 1525

          Calibration Procedure(s)                         FP—241—002—01

                                                           Calibration Procedure for the Data Acquisition Electronics
                                                           (DAEx)

          Calibration date:                                September 18, 2018

          This calibration Certificate documents the traceability to national standards, which realize the physical units of
          measurements(S!). The measurements and the uncertainties with confidence probability are given on the following
          pages and are part of the certificate.


          All calibrations have been conducted in the closed laboratory facility: environment temperature(2243)°c and
          humidity<70%.

          Calibration Equipment used (M&TE critical for calibration)


          Primary Standards                     ID #            Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration



          Process Calibrator 753             1971018                 20—Jun—18 (CTTL, No.J18X05034)                        June—19




                                               Name                       Function                                    Signature
          Calibrated by:                        Yu Zongying              SAR Test Engineer                            @.,db


          Reviewed by:                          Lin Hao                  SAR Test Engineer                          ( \f{f% $

          Approved by:                          Qi Dianyuan              SAR Project Leader                        y‘m

                                                                                                            Issued: September 20, 2018
          This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




               Certificate No: Z18—60359                                        Page 1 of 3


©
          e® |n Collsboration with

          iaaaaey»"                 CALIERATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hittp:/www.chinattl.cn

Glossary:
DAE                             data acquisition electronics
Connector angle                  information used in DASY system to align probe sensor X
                                to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e DC Voltage Measurement: Calibration Factor assessed for use in DASY
   system by comparison with a calibrated instrument traceable to national
   standards. The figure given corresponds to the full scale range of the
   voltmeter in the respective range.

e   Connector angle: The angle of the connector is assessed measuring the
    angle mechanically by a tool inserted. Uncertainty is not required.

e   The report provide only calibration results for DAE, it does not contain other
    performance test results.




Certificate No: Z18—60359                               Page 2 of 3


          WC |n, Collaboration with
         =‘["]‘], a
          uiaaazye"                 CALIBRATION LABORATORY
        Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.en


DC Voltage Measurement
    A/D — Converter Resolution nominal
        High Range:       1LSB =       6.1uV , full range =    ~100... +300 mV
        Low Range:        1LSB =        61nV , full range =    PAvcrssef +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


     Calibration Factors                     X                              Y               Z
     High Range                   405.933 + 0.15% (k=2) 405.969 + 0.15% (k=2)       405.417 + 0.15% (k=2)
     Low Range                    3.99161+0.7% (k=2)           4.01041+0.7% (k=2)   3.99418 +0.7% (k=2)



Connector Angle

     Connector Angle to be used in DASY system                                           59 @1°




Certificate No: Z18—60359                               Page 3 of 3


                                         No. I19Z61162-SEM01
                                             Page 251 of 251


ANNEX K   Accreditation Certificate




                                      ©Copyright. All rights reserved by CTTL.



Document Created: 2019-07-26 15:43:37
Document Modified: 2019-07-26 15:43:37

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC