I19Z61162-SEM01_SAR_Rev1_3

FCC ID: 2ACCJH089

RF Exposure Info

Download: PDF
FCCID_4376071

    s      s                                                   wl
cahb"'ahon Laboratory of                                     s\‘@/”@           G   Schweizerischer Kalibrierdienst
Schmid & Partner                                            i‘m                C   Service suisse d‘étalonnage
  Engineering AG                                            $ ~—tma‘s              Servizio svizzero di taratura
                               .      —                     t ay               S      ;      Shwatt        —
Zeughausstrasse 43, 8004 Zurich, Switzerland                 4 lr\\ C              Swiss Calibration Service
                                                               uhilabs

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e)   DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
         Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
         Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
         Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
         Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
          SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D750V3—1017_Jul17                                Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.0
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom
     Distance Dipole Center — TSL                              15 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         750 MHz +1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           41.9               0.89 mho/m
     Measured Head TSL parameters                              (22.0 £0.2) °C    41.0 £ 6 %         0.89 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C                                    ~———

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     2.09 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         8.32 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                  1.36 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         5.42 Wikg + 16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                 22.0 °C            $5.5              0.96 mho/m
      Measured Body TSL parameters                             (22.0 + 0.2) °C   55.0 # 6 %         0.99 mho/m + 6 %
      Body TSL temperature change during test                     <0.5 °C            «+                      ween

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition
      SAR measured                                       250 mW input power                     2.22 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         8.66 W/kg a 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                     1.45 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W         5.68 Wikg £ 16.5 % (k=2)



Certificate No: D750V3—1017_Jul17                     Page 3 of 8


(lEM)
        Appendix (Additional assessments outside the scope of SCS 0108)

        Antenna Parameters with Head TSL

              Impedance, transformed to feed point                                           54.4 Q +0.5 jQ
              Return Loss                                                                       —27.5 dB



        Antenna Parameters with Body TSL

              Impedance, transformed to feed point                                           49.3 Q — 3.4 jQ
              Return Loss                                                                      —29.1 dB




        General Antenna Parameters and Design

              Electrical Delay (one direction)                                                  1.033 ns

        After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

        The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
        second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
        are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
        "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
        according to the Standard.
        No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
        feedpoint may be damaged.


        Additional EUT Data

              Manufactured by                                                                    SPEAG
              Manufactured on                                                                March 22, 2010




        Certificate No: D750V3—1017_Jul17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                     Date: 19.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: £ = 750 MHz; 0 = 0.89 $/m; & =41; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

       Probe: EX3DV4 — SN7349; ConvF(10.49, 10.49, 10.49); Calibrated: 31.05.2017;

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 $Sn601; Calibrated: 28.03.2017

       Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

       DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.85 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.22 W/kg
SAR(1 g) = 2.09 W/kg; SAR(10 g) = 1.36 W/kg
Maximum value of SAR (measured) = 2.83 W/kg


       dB
        0

       —2.20

       —4.40

        —6.60




        —11.00
                   0 dB = 2.83 W/kg =4.52 dBW/kg




Certificate No: D750V3—1017_Jul17              Page 5 of 8


(IIIII)
          Impedance Measurement Plot for Head TSL




                                                                             19 Jul    2017   ©8:47:38
                   (CHZ) sii     1 U FS         4: 54.355 a    0.5137 a   109.00 pH           750.000 000 MHz
                                                              sepes

                   Del

                   Ca




                   Av
                   163


                    H1d

                    CH2




                    Ca




                    Ay
                    16"

                    H1d

                               START 550.000 000 MHz                                  STOP 950.000 000 MHz




          Certificate No: D750V3—1017_Jul17                      Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                      Date: 19.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1017

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: £ = 750 MHz; 0 = 0.99 $/m; & = 55; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.35, 10.35, 10.35); Calibrated: 31.05.2017;

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.67 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 3.34 W/kg
SAR(1 g) = 2.22 W/kg; SAR(10 g) = 1.45 W/kg
Maximum value of SAR (measured) = 2.96 W/kg


        dB
        0

        —2.20

        —4.40

        —6.60

        —8.80

        —11.00
                   0 dB = 2.96 W/kg =4.71 dBW/kg




Certificate No: D750V3—1017_Jul17               Page 7 of 8


®
Impedance Measurement Plot for Body TSL




                                                              19 Jul 2017    14:53:13
         (CHI sii     1 U FS         1: 49.266 a —3.4219 n 62.015 pF        750.000 000 MHz
                                                lwe
                                                us       .




          Av
          16 $



          H1d

          CH2




          Ca




          Ay
          16°

          H1d

                    START 550.000 000 MHz                          STOP 950.000 000 MHz




Certificate No: D750V3—1017_Jul17                  Page 8 of 8


                                                       No. I19Z61162-SEM01
                                                           Page 150 of 251


835 MHz Dipole Calibration Certificate




                                         ©Copyright. All rights reserved by CTTL.


®    y      s
Calibration Laboratory of
Schmid & Partner
  Engineering AG
                             .      a
Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                 m
                                                                    w tm
                                                                  lt\3
                                                                 ze onl
                                                                  *« omay
                                                                   *4/7\\\ ®
                                                                      uhitabs
                                                                                                           g Schweizerischer Kalibrierdienst
                                                                                                           c Service suisse d‘étalonnage

                                                                                                           S
                                                                                                                   Servizio svizzero di taratura
                                                                                                                   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client      CTTL—BJ (Auden)                                                                    Certificate No: D835V2—40069_Jul17

CALIBRATION CERTIFICATE
 Object                               D835V2 — SN:4d069


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                    July 19, 2017


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical forcalibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
 Power sensor NRP—Z91                   SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
 Power sensor NRP—Z291                  SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
 Reference 20 dB Attenuator             SN: 5058 (20k)            07—Apr—17 (No. 217—02528)                             Apr—18
 Type—N mismatch combination            SN: 5047.2 / 06327        07—Apr—17 (No. 217—02529)                             Apr—18
 Reference Probe EX3DV4                 SN: 7349                  31—May—17 (No. EX3—7349_May17)                        May—18
 DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     in house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                        Name                                    Function                                 Signature
 Calibrated by:                         Johannes Kurikka                        Laboratory Technician                     / /



 Approved by:                           Katia Pokovic                           Technical Manager                     //%;_,



                                                                                                                        Issued: July 20, 2017
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d069_Jul17                                    Page 1 of 8




                                                                                     wrveuprg rrgr nc e wr ons


    f    .                                                     w91
Callbll'ahon Laboratory of                                 s\“\\\_//\_/"‘/,;                            G     Schweizerischer Kalibrierdienst
Schmid & Partner                                           iinax e                                      c     Service suisse d‘étalonnage
  Engineering AG                                           22— VE                                             Servizio svizzero di taratura
                              £       1                    z7z\wy
Zeughausstrasse 43, 8004 Zurich, Switzerland                */I//\\\\‘\\*‘                              S     Swiss Calibration Service
                                                               doatuas

Accredited by the Swiss Accreditation Service (SAS)                                                       Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
          Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
   e SAR measured: SAR measured at the stated antenna input power.
   e SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
      +   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D835V2—4d069_Jul17                              Page 2 of 8




                                                                               worvrvepay brg bee e mb ob ragr rvwr n neneves e nen meg se n n >


)
S
Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
    DASY Version
     Extrapolation
     Phantom
                                                            DASY5
                                                     Advanced Extrapolation
                                                      Modular Flat Phantom
                                                                                                     V52.10.0




     Distance Dipole Center — TSL                              15 mm                                with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                          835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature          Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C               41.5              0.90 mho/m
     Measured Head TSL parameters                             (22.0 + 0.2) °C        40.8 £ 6 %        0.91 mho/m + 6 %
     Head TSL temperature change during test                     <0.5 °C                                         se«


SAR result with Head TSL

     SAR averaged over 1 cm® (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                        2.37 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W            9.37 Wikg a 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                     1.53 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W            6.06 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature          Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C               55.2               0.97 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C        54.8 6 %           1.01 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                        2.43 Wi/kg
      SAR for nominal Body TSL parameters                    normalized to 1W            9.41 Wikg + 17.0 % (k=2)

      SAR averaged over 10 cm‘ (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                      1.57 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W             6.12 Wikg a 16.5 % (k=2)



Certificate No: D835V2—40069_Jul17                     Page 3 of 8




                                                                     wvripeg rrgr ie e m rrgr es e en n en e y    e n >


mm®
@TTL
uy
                                                                                                   1O l

 Appendix (Additional assessments outside the scope of SCS 0108)

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                            52.1 0 — 1.2 jQ
       Return Loss                                                                        — 32.4 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                            47.9 Q — $.9 jQ
       Return Loss                                                                       —26.9 dB



 General Antenna Parameters and Design

     LElectricaI Delay (one direction)                                                    1.392 ns

 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                     SPEAG
       Manufactured on                                                               November 09, 2007




 Certificate No: D835V2—4d069_Jul17                        Page 4 of 8




                                                                         wvriperg n ie n m n nge


(
C
DASY5 Validation Report for Head TSL



Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:40069
                                                                                                    Date: 19.07.2017




Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: £ = 835 MHz; 0 = 0.91 S/m; & = 40.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvF(10.07, 10.07, 10.07); Calibrated: 31.05.2017;

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

   «_   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.08 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 3.65 W/kg
SAR(1 g) = 2.37 W/kg; SAR(10 g) = 1.53 W/kg
Maximum value of SAR (measured) = 3.21 W/kg


        dB
        0


        —2.20

        ~4.40

        —6.60

        —8.80


        —11.00
                   0 dB = 3.21 W/kg = 5.07 dBW/kg




Certificate No: D835V2—4d069_Jul17              Page 5 of 8




                                                              N vvopeg n rgn ene e ee n rgrrene n ns se n e e e c


(III|I)°
           Impedance Measurement Plot for Head TSL




                                                                               19 Jul 2017       @8:54:25
                            §11     1 U FS         4: 52145 a   —1.1758 e   162.11 pF           835.000 000 MHz
                                                                ~ATO—




                     Av
                     16 *


                     H1d

                     CH2




                     Ca




                     Av
                     16 ®


                     H1d

                                  START 635.000 000 MHz                                STOP 1 035.000 000 MHz




           Certificate No: D835V2—40069_Jul17                      Page 6 of 8




                                                                                 worvr ty brag enc e mb n rgr rene e nenenec n mm n e c


(I1
      S
      DASY5 Validation Report for Body TSL

                                                                                                          Date: 19.07.2017
      Test Laboratory: SPEAG, Zurich, Switzerland

      DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d069

      Communication System: UID 0 — CW; Frequency: 835 MHz
      Medium parameters used: £ = 835 MHz; 0 = 1.01 S/m; & = 54.8; p = 1000 kg/m3
      Phantom section: Flat Section
      Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

      DASY52 Configuration:

         e   Probe: EX3DV4 — SN7349; ConvF(10.2, 10.2, 10.2); Calibrated: 31.05.2017;

         e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

         e   Electronics: DAE4 S$n601; Calibrated: 28.03.2017

         &   Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

         &   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


      Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
      Measurement grid: dx=5mm, dy=5mm, dz=5mm
      Reference Value = 59.35 V/m; Power Drift = —0.03 dB
      Peak SAR (extrapolated) = 3.67 W/kg
      SAR(1 g) = 2.43 W/kg; SAR(10 g) = 1.57 W/kg
      Maximum value of SAR (measured) = 3.22 W/kg


             dB
              0


             —2.20

             —4.40

              —6.60

              —8.80


              —11.00
                         0 dB = 3.22 W/kg = 5.08 dBW/kg




      Certificate No: D835V2—4d069_Jul17             Page 7 of 8




                                                                   N vvopeg n rgn ene e ee n rgrrene n ns se n e e e c


Impedance Measurement Plot for Body TSL




                                                                  19 Jul 2017       @9:16:58
         CHZ sii      1 U FS         4: 47.896 n   —3,0926 0   48.966 pF           835.000 000 MHz
                                                   +T_




         Av
         16 *


         H1d

          CH2




         Ca




         Aw
         :I.S9


          H1d

                    START 635.000 000 MHz                                 STOP 1 035.000 000 MHz




Certificate No: D835V2—40069_Jul17                    Page 8 of 8




                                                                    worvrvepabrgr nee n vr n ragirwwe n nene se n n e + n


                                                        No. I19Z61162-SEM01
                                                            Page 159 of 251


1750 MHz Dipole Calibration Certificate




                                          ©Copyright. All rights reserved by CTTL.


Calibration Laboratory of                                            ollUpy
      A                                                          s /n
                                                                faw‘/‘+                                   S       Schweizerischer Kalibrierdienst
Schm[d & I?artner                                               ilfifi                                      c Service suisse d‘étalonnage
   Englneerlng AG                                               ze                                                Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                    €4,/fi\\y}                                 S       Swiss Calibration Service
                                                                     "hilulst


Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      CTTL—BJ (Auden)                                                                   Certificate No: D1750V2—1003_Jul17

CALIBRATION CERTIFICATE
 Object                               D1750V2 — SN:1003


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                    July 21, 2017


 This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                     ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                       SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
 Power sensor NRP—Z91                  SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
 Power sensor NRP—Z91                  SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
 Reference 20 dB Attenuator            SN: 5058 (20k)            07—Apr—17 (No. 217—02528)                             Apr—18
 Type—N mismatch combination           SN: 5047.2 / 06327        07—Apr—17 (No. 217—02529)                             Apr—18
 Reference Probe EX3DV4                SN: 7349                  31—May—17 (No. EX3—7349_May17)                        May—18
 DAE4                                  SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


 Secondary Standards                    ID #                     Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                  SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783           07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317           07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
 Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—16)                    In house check: Oct—17


                                        Name                                    Function                                Signature
 Calibrated by:                        Michael Weber                            Laboratory Technician                  M—



 Approved by:                          Katia Pokovic                            Technical Manager                  M



                                                                                                                       Issued: July 24, 2017
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D1750V2—1003_Jul17                                    Page 1 of 8




                                                                                     Nn vperg nen n mn n en ce n n e en n n           ny   >      —



Document Created: 2019-07-26 15:43:35
Document Modified: 2019-07-26 15:43:35

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC