I19Z60967-SEM03_SAR_Rev4_5

FCC ID: 2ACCJH089

RF Exposure Info

Download: PDF
FCCID_4321718

uy                                                                                           C




    .      .                                                   al19
Callbfat'on Laboratory of                                    Q\\\\\\_//’z;         g    Schweizerischer Kalibrierdienst
Schmid & Partner                                            m                      c    Service suisse d‘étalonnage
  Englneering AG                                            en itingl                   Servizio svizzero di taratura
                             i      P                       * ty
Zeughausstrasse 43, 8004 Zurich, Switzerland                7///\\\\\\*‘           S    gwiss Calibration Service
                                                                AARS

Accredited by the Swiss Accreditation Service (SAS)                                 Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Additional Documentation:
    e)   DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
         SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul17                               Page 2 of 8




                                                                               Nn en n n en n n n n e n n n         n n   ey   >


)
n
S
(II
      Measurement Conditions
         DASY system configuration, as far as not given onpage 1.
          DASY Version                                            DASYS5                                 V52.10.0
           Extrapolation                                   Advanced Extrapolation
           Phantom                                          Modular Flat Phantom
           Distance Dipole Center — TSL                              10 mm                              with Spacer
           Zoom Scan Resolution                              dx, dy, dz =5 mm
           Frequency                                         2600 MHz +# 1 MHz


      Head TSL parameters
         The following parameters and calculations were applied.
                                                                     Temperature       Permittivity          Conductivity
           Nominal Head TSL parameters                                  22.0 °C            39.0                 1.96 mho/m
           Measured Head TSL parameters                              (22.0 £0.2) °C     37.2 26 %          2.04 mho/m £ 6 %
           Head TSL temperature change during test                      <0.5 °C             ~—~                      =«»—

      SAR result with Head TSL

           SAR averaged over 1 cm° (1 g) of Head TSL                  Condition
           SAR measured                                        250 mW input power                      14.9 Wikg
           SAR for nominal Head TSL parameters                     normalized to 1W         57.9 Wikg # 17.0 % (k=2)

           SAR averaged over 10 cm* (10 g) of Head TSL                condition
           SAR measured                                        250 mW input power                  6.57 W/kg
           SAR for nominal Head TSL parameters                     normalized to 1W         25.8 Wikg a 16.5 % (k=2)


      Body TSL parameters
         The following parameters and calculations were applied.
                                                                     Temperature        Permittivity            Conductivity
           Nominal Body TSL parameters                                  22.0 °C            52.5                 216 mho/m
           Measured Body TSL parameters                              (22.0 £ 0.2) °C    51.6 46 %          2.22 mho/m * 6 %
            Body TSL temperature change during test                     <0.5 °C             m+>


      SAR result with Body TSL

            SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
            SAR measured                                       250 mW input power                      14.1 Wikg
            SAR for nominal Body TSL parameters                    normalized to 1W          55.5 Wikg + 17.0 % (k=2)

            SAR averaged over 10 cm* (10 g) of Body TSL                condition
            SAR measured                                       250 mW input power                      6.25 Wikg
            SAR for nominal Body TSL parameters                    normalized to 1W          24.8 Wikg 2 16.5 % (k=2)



      Certificate No: D2600V2—1012_Jul17                    Page 3 of 8




                                                                                       Nn opycg m n n       y    >


(llém)
         Appendix (Additional assessments outside the scope of SCS 0108)

         Antenna Parameters with Head TSL

               Impedance, transformed to feed point                                           46.8 Q — 5.0 jQ
               Return Loss                                                                       — 24.2 dB



         Antenna Parameters with Body TSL

               Impedance, transformed to feed point                                           43.5 Q — 5.3 jQ
               Return Loss                                                                      —21.0 dB




         General Antenna Parameters and Design

             fEIectricaI Delay (one direction)                                                   1.1§1 ns

         After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

         The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
         second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
         are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
         "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
         according to the Standard.
         No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
         feedpoint may be damaged.


         Additional EUT Data

               Manufactured by                                                                    SPEAG
               Manufactured on                                                               October 30, 2007




         Certificate No: D2600V2—1012_Jul17                        Page 4 of 8




                                                                                              ~8 y cans           o                      —


|N
GDASY5 Validation Report for Head TSL



 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012
                                                                                           Date: 20.07.2017




 Communication System: UID 0 — CW; Frequency: 2600 MHz
 Medium parameters used: f = 2600 MHz; 0 = 2.04 S/m; & = 37.2; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2011)

 DASY52 Configuration:

     *   Probe: EX3DV4 — SN7349; ConvF(7.96, 7.96, 7.96); Calibrated: 31.05.2017;

     *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

     *   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

     e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 113.6 V/m; Power Drift = 0.05 dB
 Peak SAR (extrapolated) = 32.3 W/kg
 SAR(1 g) = 14.9 W/kg; SAR(10 g) = 6.57 W/kg
 Maximum value of SAR (measured) = 25.0 W/kg


         dB
          0

         —5.00

         —10.00

         ~15.00

         —20.00

         —25.00
                    0 dB = 25.0 W/kg = 13.98 dBW/kg




 Certificate No: D2600V2—1012_Jul17              Page 5 of 8




                                                                       Nennma n n n en n m n nc   snn >   —   —


Impedance Measurement Plot for Head TSL




                                                                   20 Jul 2017 12:14:05
         CHI) sii     1 U FS          5: 46.936 a   ~5.0410 a   12.143 pF      2 600.000 000 MHz




         Av
         168


         H1d

          CH2




          Ca




          fivg
          16


          H1d

                      ART 2 400.000 000 MHz                              STOP 2 800.000 000 MHz




Certificate No: D2600V2—1012_Jul17                     Page 6 of 8




                                                                                 8ypyrcgene


DASY5 Validation Report for Body TSL

                                                                                          Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.22 $/m; & = 51.6; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(7.94, 7.94, 7.94); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   +   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.6 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 30.1 W/kg
SAR(I g) = 14.1 W/kg; SAR(10 g) = 6.25 W/kg
Maximum value of SAR (measured) = 23.4 W/kg


       dB
       0
       —5.20
        —10.40
        ~15.60
        —20.80
        —26.00
                   0 dB = 23.4 W/kg = 13.69 dBW/kg




Certificate No: D2600V2—1012_Jul17             Page 7 of 8




                                                                    Nnpeg n n en n en n en n n n ns n n n   ny   >


)
(I
     Impedance Measurement Plot for Body TSL




                                                                       24 Jul 2017 1i4:50:41
                     §11     i U Fs          1143.541 a   —5.3223 8 114.501 pF     2 600.000 000 MHz
                                                          _1


              Del

              Ca




               Av
               16a


               H1d

               CH2




               Ca




               Av
               16°

               H1d

                           START 2 400.000 000 MHz                           STOP 2 800.000 000 MHz




     Certificate No: D2600V2—1012_Jul17                      Page 8 of 8




                                                                                    Nnepg rrgn en n m n


                                           No. I19Z60967-SEM03
                                               Page 188 of 239


ANNEX I   DAE Calibration Certificate




                                        ©Copyright. All rights reserved by CTTL.


m*
/7L                                                                                             in Anioiniiiaeiiiontatie
uo




     &       &                                                  w‘ ty
 Callbl.'atlon Laboratory of                                  i\‘\\\\ij///'/';         S    Schweizerischer Kalibrierdienst
 Schmid & Partner                                            M                         c    Service suisse d‘étalonnage
    Engineering AG                                           o o mf                         Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                 * ///'\\\ y              S    Swiss Calibration Service
                                                                Carhitabe®®

 Accredited by the Swiss Accreditation Service (SAS)                                    Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates



 Glossary
 DAE                          data acquisition electronics
 Connector angle              information used in DASY system to align probe sensor X to the robot
                              coordinate system.

 Methods Applied and Interpretation of Parameters
       DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
      comparison with a calibrated instrument traceable to national standards. The figure given
      corresponds to the full scale range of the voltmeter in the respective range.
          Connector angle: The angle of the connector is assessed measuring the angle
          mechanically by a tool inserted. Uncertainty is not required.
          The following parameters as documented in the Appendix contain technical information as a
          result from the performance test and require no uncertainty.
                 DC Voltage Measurement Linearity: Verification of the Linearity at +10%and —10% of
                 the nominal calibration voltage. Influence of offset voltage is included in this
                 measurement.
                 Common mode sensitivity: Influence of a positive or negative common mode voltage on
                 the differential measurement.
                 Channel separation: Influence of a voltage on the neighbor channels not subject to an
                 input voltage.
                 AD Converter Values with inputs shorted: Values on the internal AD converter
                 corresponding to zero input voltage
                 Input Offset Measurement. Output voltage and statistical results over a large number of
                 zero voltage measurements.
                 Input Offset Current: Typical value for information; Maximum channel input offset
                 current, not considering the input resistance.
                 Input resistance: Typical value for information: DAE input resistance at the connector,
                 during internal auto—zeroing and during measurement.
                 Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
                 alarm signal is generated.
                 Power consumption: Typical value for information. Supply currents in various operating
                 modes.




 Certificate No: DAE4—1525_Oct17                                  Page 2 of 5




                                                                                 wervopg n ne a io rgn rewr a vewruer e cona wmz


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB =          6.1uV ,     full range =    —100...+300 mV
       Low Range:           1LSB =           61nV ,     full range =     —~1....... +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

     Calibration Factors                X                           ¥                           Z
     High Range             405.912 + 0.02% (k=2)     405.954 £0.02% (k=2)        405.400 £ 0.02% (k=2)
     Low Range              8.99166 £ 1.50% (k=2)     4.00980 + 1.50% (k=2)       3.99550 £1.50% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                             sG56°%1°          J




Certificate No: DAE4—1525_Oct17                       Page 3 of 5




                                                                                Ne n veperg n rag en e mb n ng eene n nene se n en meg n nc


()
 S
 Appendix (Additional assessments outside the scope of SCS0O108)

 1. DC Voltage Linearity
      High Range
      Channel X
      Channel X
                      + Input
                      + Input
                                                 Reading (uV)
                                                  200030.95
                                                   20004.11
                                                                      Difference (1V)
                                                                             242
                                                                              —0.05
                                                                                                  Error (%)
                                                                                                  —0.00
                                                                                                  —0.00
      Channel X       — Input                     —20003.75                   2.02                0.01
      Channel Y       + Input                     200031.20                  —2.23                 0.00
      Channel Y       + Input                      20001.46                   ~2.74               —0.01
      Channel Y       — Input                     ~20005.92                   —0.05                0.00
      Channel Z       + Input                     200032.03                   ~1.05                0.00
      Channel Z       + Input                      20001.94                   «2241                —0.01
      Channel Z       = Input                     —20006.15                   —0.20                0.00


      Low Range                                  Reading (V)          Difference (LV)             Error (%)
      Channel X       + Input                      2000.66                    0.19                 0.01
      Channel X       + Input                       200.40                    —0.18                —0.09
      Channel X       = Input                      ~198.67                    0.81                 —0.40
      Channel Y       + Input                      2000.90                    0.48                 0.02
      Channel Y       + Input                       199.98                    —0.58                —0.29
      Channel Y       — Input                       —200.18                   —0.62                0.31
      Channel Z       + Input                      2000.68                    0.32                 0.02
      Channel Z       + Input                       199.07                    ~1.45                —0.72

      Channel Z       — Input                       +201.14                   ~1.52                0.76



 2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuringtime: 3 sec
                         Common mode                         High Range                        Low Range
                         Input Voltage (mV)              Average Reading (1V)             Average Reading (1V)

     Channel X                            200                      18.92                        16.76
                                         — 200                     —15.78                       —17.08
     Channel Y                            200                      —20.47                       —20.86
                                         — 200                     20.66                        20.31
     Channel Z                            200                      18.43                        18.46
                                         ~ 200                     ~15.65                       ~15.97



 3. Channel separation
     DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                        Input Voltage (mV)           Channel X (uV)         Channel Y (uV)     Channel Z (uV)

     Channel X                     200                        —                  0.08               —3.66
     Channel Y                     200                       712                      —              1.80
     Channel Z                     200                    10.44                  4.52                    —


 Certificate No: DAE4—1525_Oct17                              Page 4 of 5




                                                                                          Nn epgrge en n c nnges


()
 S
 4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec



     Channel X

     Channel Y
                                                    High Range (LSB)

                                                            15817

                                                            16329
                                                                                      Low Range (LSB)

                                                                                             15005

                                                                                             14457
     Channel Z                                              15576                            15478



 5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                            Average (1V)        min. Offset (V)       max. Offset (uV)        Sid: [():\\’l;atlon

     Channel X                   0.63                 »0.54                 2.27                     0.51
     Channel Y                   ~2.07                ~3.42                ~1.02                     0.49
     Channel Z                   —0.89                —2.38                 0.83                     0.54



 6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA


 7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                      Measuring (MOhm)
     Channel X                                             200                                200
     Channel Y                                             200                                200
     Channel Z                                             200                                200



 8. Low Battery Alarm Voltage (Typical values for information)
     Typical values                                Alarm Level (VDC)
     Supply (+ Vee)                                                            +7.9
     Supply (— Vee)                                                            76



 9. Power Consumption (Typical values for information)
     Typical values                         Switched off (mA)        Stand by (mA)        Transmitting (mA)
     Supply (+ Vee)                                +0.01                  +6                        +14
     Supply (— Vec)                                —0.01                  —8                         —9




 Certificate No: DAE4—1525_Oct17                            Page 5 of 5




                                                                                       Ns eperg n ragi esn n en n nge ie >


                                                                  No. I19Z60967-SEM03
                                                                      Page 193 of 239


 ANNEX J         SPOT CHECK

 J.1 Dielectric Performance and System Validation


                        Table J.1-1: Targets for tissue simulating liquid
  Frequency(MHz)     Liquid Type Conductivity(σ) ± 5% Range Permittivity(ε)            ± 5% Range
       2450             Head            1.80         1.71~1.89          39.2           37.2~41.2
       1900             Body            1.52         1.44~1.60          53.3           50.6~56.0
       2600             Body            2.16         2.05~2.27          52.5           49.9~55.1

             Table J.1-2: Dielectric Performance of Head Tissue Simulating Liquid
  Measurement Date                              Permittivity    Drift       Conductivity       Drift
                         Type      Frequency
    (yyyy-mm-dd)                                    ε           (%)           σ (S/m)          (%)
      2019-6-17          Head      2450 MHz        39.55        0.89           1.771           -1.61
      2019-6-6           Body      1900 MHz        53.15        -0.28           1.54           1.32
      2019-6-6           Body      2600 MHz        51.66        -1.60          2.143           -0.79


                            Table J.1-3: System Validation of Head
Measurement                 Target value (W/kg)     Measured value(W/kg)               Deviation
    Date       Frequency     10 g         1g         10 g           1g            10 g          1g
(yyyy-mm-dd)                Average     Average     Average       Average        Average      Average
 2019-6-17       2450 MHz       24.2     51.7         24.5           52.1         1.24%         0.77%


                            Table J.1-4: System Validation of Body
Measurement                 Target value (W/kg)     Measured value(W/kg)               Deviation
    Date       Frequency     10 g         1g         10 g           1g            10 g          1g
(yyyy-mm-dd)                Average     Average     Average       Average        Average      Average
  2019-6-6       1900 MHz       21.4     40.4         21.35         40.17         -0.23%       -0.57%
  2019-6-6       2600 MHz       24.5     54.1         25.23         54.75         2.98%         1.20%




                                                               ©Copyright. All rights reserved by CTTL.


                                                                                      No. I19Z60967-SEM03
                                                                                          Page 194 of 239

         J.2 Conducted power of selected case
                                          Table J.2-1: The conducted Power for WIFI
                             Band             Channel       Frequency    Measured Power (dBm)
                        WLAN2450                 6           2437 MHz                       15.51
                      WCDMA1900                9662         1872.4 MHz                      23.01
                      WCDMA1900                9800          1880 MHz                       23.18
                      WCDMA1900                9938         1907.6MHz                       23.09
                           LTEBand2            18700         1860 MHz                       23.11
                           LTEBand7            21350         2560 MHz                       22.01


         J.3 SAR test result for spot check
                       Ambient Temperature: 22.5 oC                Liquid Temperature: 22.1 oC
  Frequency                    Figure       Conducted                   Measured       Reported     Measured        Reported   Power
                    Test                                Max. tune-up
                                No./         Power                      SAR(10g)      SAR(10g)      SAR(1g)         SAR(1g)    Drift
 Ch     MHz.      Position                              Power (dBm)
                                Note         (dBm)                        (W/kg)        (W/kg)       (W/kg)          (W/kg)    (dB)
  6     2437     Right check    Fig J.1      15.51         16.5           0.384          0.48        0.727           0.91      0.05
9400    1880       Bottom           /        23.18         23.5           0.575          0.62         1.06           1.14      -0.01
9262    1852.4     Bottom       Fig J.2      23.01         23.5           0.585          0.65         1.13           1.26      -0.16
9538    1907.6     Bottom           /        23.09         23.5           0.552          0.61         1.03           1.13      0.05
18700   1860       Bottom       Fig J.3       23.11        23.5           0.492          0.54        0.953           1.04      0.11
21350   2560       Bottom       Fig I.4      22.01         22.2           0.456          0.48         1.01           1.06      -0.09
9262    1852.4     Bottom      Headset1      23.01         23.5           0.561          0.63        1.082           1.21      0.08
9262    1852.4     Bottom      Headset2      23.01         23.5           0.561          0.63        1.090           1.22      0.07
9262    1852.4     Bottom      Headset3      23.01         23.5           0.559          0.63        1.113           1.25      0.06
9262    1852.4     Bottom      Headset4      23.01         23.5           0.545          0.63        1.095           1.23      -0.12
                   Note1: The distance between the EUT and the phantom bottom is 10mm


         J.4 Reported SAR Comparison

                                                                           Reported SAR              Reported SAR
          Exposure Configuration              Technology Band
                                                                       1g (W/kg): spot check        1g (W/kg): original

                  Head                                                         0.91                          0.99
                                                WLAN2450
        (Separation Distance 0mm)

                                               WCDMA1900                       1.26                          1.19

                  Hotspot
                                                 LTEBand2                      1.04                          1.19
        (Separation Distance 10MM)

                                                 LTEBand7                      1.06                          1.19

         Note: The spot check results marked blue are larger than the original result.




                                                                                   ©Copyright. All rights reserved by CTTL.


                                                                  No. I19Z60967-SEM03
                                                                      Page 195 of 239

                                Table J5.1: List of Main Instruments
No.          Name                 Type         Serial Number     Calibration Date       Valid Period
01     Network analyzer          E5071C         MY46110673       January 24, 2018         One year
02       Power meter              NRVD            102083
                                                                November 01,2017          One year
03       Power sensor            NRV-Z5           100542
04     Signal Generator          E4438C         MY49070393       January 02,2018          One Year
05         Amplifier             60S1G4           0331848              No Calibration Requested
06            BTS                CMW500           159889        December 20, 2017         One year
07       E-field Probe        SPEAG EX3DV4         7464         September 12,2017         One year
08            DAE              SPEAG DAE4          1525          October 02, 2017         One year
09    Dipole Validation Kit   SPEAG D1900V2        5d101           July 26,2017           One year
10    Dipole Validation Kit   SPEAG D2450V2         853            July 21,2017           One year
11    Dipole Validation Kit   SPEAG D2600V2        1012            July 21,2017           One year
                                   ***END OF REPORT BODY***




                                                               ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 196 of 239



J.5 GRAPH RESULTS
WLAN2450_CH6 Right Cheek
Date: 6/17/2019
Electronics: DAE4 Sn1525
Medium: head 2450 MHz
Medium parameters used: f = 2437 MHz; σ = 1.775 mho/m; εr = 38.71; ρ = 1000 kg/m3
Ambient Temperature: 22.5oC, Liquid Temperature: 22.3oC
Communication System: WLAN2450 2437 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.72,7.72,7.72)


Area Scan (91x151x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.15 W/kg


Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 13.74 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 1.46 W/kg
SAR(1 g) = 0.727 W/kg; SAR(10 g) = 0.384 W/kg
Maximum value of SAR (measured) = 1.17 W/kg




                                        Fig J.1




                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 197 of 239




WCDMA1900 _CH9262 Bottom
Date: 6/6/2019
Electronics: DAE4 Sn1525
Medium: body 1900 MHz
Medium parameters used: f = 1852.4 MHz; σ = 1.571 mho/m; εr = 53.24; ρ = 1000 kg/m3
Ambient Temperature: 22.1oC, Liquid Temperature: 22.2oC
Communication System: WCDMA1900-BII 1907.6 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.82,7.82,7.82)


Area Scan (161x91x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 1.72 W/kg


Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 20.79 V/m; Power Drift = -0.16 dB
Peak SAR (extrapolated) = 2.07 W/kg
SAR(1 g) = 1.13 W/kg; SAR(10 g) = 0.585 W/kg
Maximum value of SAR (measured) = 1.73 W/kg




                                         Fig J.2




                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 198 of 239


LTE1900-FDD2_CH18700 Bottom
Date: 6/6/2019
Electronics: DAE4 Sn1525
Medium: body 1900 MHz
Medium parameters used: f = 1860 MHz; σ = 1.525 mho/m; εr = 53.21; ρ = 1000 kg/m3
Ambient Temperature: 22.1oC, Liquid Temperature: 22.2oC
Communication System: LTE1900-FDD2 1900 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(8.32,8.32,8.32)


Area Scan (151x91x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.44 W/kg


Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 25.63 V/m; Power Drift = -0.11 dB
Peak SAR (extrapolated) = 1.73 W/kg
SAR(1 g) = 0.953 W/kg; SAR(10 g) = 0.492 W/kg
Maximum value of SAR (measured) = 1.45 W/kg




                                        Fig J.3




                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 199 of 239


LTE2500-FDD7_CH21350 Bottom
Date: 6/6/2019
Electronics: DAE4 Sn1525
Medium: body 2600 MHz
Medium parameters used: f = 2560 MHz; σ = 2.118 mho/m; εr = 53.21; ρ = 1000 kg/m3
Ambient Temperature: 22.1oC, Liquid Temperature: 22.2oC
Communication System: LTE2500-FDD7 2560 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.30,7.30,7.30)


Area Scan (141x91x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.62 W/kg


Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 11.50 V/m; Power Drift = -0.09 dB
Peak SAR (extrapolated) = 2.01 W/kg
SAR(1 g) = 1.01 W/kg; SAR(10 g) = 0.456 W/kg
Maximum value of SAR (measured) = 1.52 W/kg




                                        Fig J.4




                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 200 of 239


J.6 ANNEX System Verification Results

2450 MHz
Date: 6/17/2019
Electronics: DAE4 Sn1525
Medium: Head 2450 MHz
Medium parameters used: f = 2450 MHz; σ =1.771 mho/m; εr = 39.55; ρ = 1000 kg/m3
Ambient Temperature: 22.5oC Liquid Temperature: 22.3oC
Communication System: CW Frequency: 2450 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.72,7.72,7.72)

System Validation /Area Scan (81x191x1): Interpolated grid: dx=1.000 mm, dy=1.000
    mm
    Reference Value =113.68 V/m; Power Drift = 0.010 dB
    Fast SAR: SAR(1 g) = 12.89 W/kg; SAR(10 g) = 6.01 W/kg
Maximum value of SAR (interpolated) = 21.4 W/kg

    System Validation /Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
   dy=5mm, dz=5mm
   Reference Value =113.68 V/m; Power Drift = 0.010 dB
   Peak SAR (extrapolated) = 26.79 W/kg
   SAR(1 g) = 13.18 W/kg; SAR(10 g) = 6.18 W/kg
Maximum value of SAR (measured) = 21.83 W/kg




0 dB = 21.83 W/kg = 13.39 dB W/kg
                         Fig J.5 validation 2450 MHz 250mW

                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 201 of 239


1900 MHz
Date: 6/6/2019
Electronics: DAE4 Sn1525
Medium: Body 1900 MHz
Medium parameters used: f = 1900 MHz; σ =1.540 mho/m; εr = 53.15; ρ = 1000 kg/m3
Ambient Temperature: 22.5oC Liquid Temperature: 22.3oC
Communication System: CW Frequency: 1900 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.82,7.82,7.82)

System Validation /Area Scan (81x191x1): Interpolated grid: dx=1.000 mm, dy=1.000
    mm
    Reference Value =103.41 V/m; Power Drift = 0.07 dB
    Fast SAR: SAR(1 g) = 10.46 W/kg; SAR(10 g) = 5.48 W/kg
Maximum value of SAR (interpolated) = 17.78 W/kg

    System Validation /Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
   dy=5mm, dz=5mm
   Reference Value =103.41 V/m; Power Drift = 0.07 dB
   Peak SAR (extrapolated) = 17.91 W/kg
   SAR(1 g) = 10.51 W/kg; SAR(10 g) = 5.50 W/kg
Maximum value of SAR (measured) = 14.17 W/kg




0 dB = 14.17 W/kg = 11.51 dB W/kg
                           Fig J.6 validation 1900 MHz 250mW
                                                         ©Copyright. All rights reserved by CTTL.


                                                            No. I19Z60967-SEM03
                                                                Page 202 of 239


2600 MHz
Date: 6/6/2019
Electronics: DAE4 Sn1525
Medium: Body 2600 MHz
Medium parameters used: f = 2600 MHz; σ =2.143 mho/m; εr = 51.66; ρ = 1000 kg/m3
Ambient Temperature: 22.5oC Liquid Temperature: 22.3oC
Communication System: CW Frequency: 2600 MHz Duty Cycle: 1:1
Probe: EX3DV4 – SN7464 ConvF(7.30,7.30,7.30)

System Validation /Area Scan (81x191x1): Interpolated grid: dx=1.000 mm, dy=1.000
    mm
    Reference Value =108.81 V/m; Power Drift = 0.09 dB
    Fast SAR: SAR(1 g) = 13.90 W/kg; SAR(10 g) = 6.11 W/kg
Maximum value of SAR (interpolated) = 29.85 W/kg

    System Validation /Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
   dy=5mm, dz=5mm
   Reference Value =108.81 V/m; Power Drift = 0.09 dB
   Peak SAR (extrapolated) = 30.13 W/kg
   SAR(1 g) = 13.77 W/kg; SAR(10 g) = 6.13 W/kg
Maximum value of SAR (measured) = 23.93 W/kg




0 dB = 23.93 W/kg = 13.79 dB W/kg
                           Fig J.7 validation 2600 MHz 250mW
                                                         ©Copyright. All rights reserved by CTTL.


                                                       No. I19Z60967-SEM03
                                                           Page 203 of 239


Table J.6-1 Comparison between area scan and zoom scan for system verification
                                        Area scan    Zoom scan
   Date        Band         Position                                    Drift (%)
                                             (1g)        (1g)
2019-6-19      2450          Head          12.89         13.18             -2.20
 2019-6-6      1900          Body          10.46         10.51             -0.48
 2019-6-6      2600          Body           13.9         13.77              0.94




                                                    ©Copyright. All rights reserved by CTTL.


                                         No. I19Z60967-SEM03
                                             Page 204 of 239


ANNEX Probe Calibration Certificate




                                      ©Copyright. All rights reserved by CTTL.


         -IthL ;Co‘ld:orufionewith a


       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504
       E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


Glossary:
TSL              tissue simulating liquid
NORMx,y,z        sensitivity in free space
ConvF            sensitivity in TSL / NORMx,y,z
DCP              diode compression point
CF               crest factor (1/duty_cycle) of the RF signal
A,B,C,D          modulation dependent linearization parameters
Polarization ®   ® rotation around probe axis
Polarization 0   0 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                 8=0 is normal to probe axis
Con nector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
  M easurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
   July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KiDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
    NORMx,y,z: Assessed for E—field polarization 6=0 (fs900MHz in TEM—cell; {>1800MHz: waveguide).
    NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
    E —field uncertainty inside TSL (see below ConvF).
    NORM(P)x,y,z = NORMx,y,z* frequency_response (see Frequency Response Chart). This
    linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
    frequency response is included in the stated uncertainty of ConvF.
    DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
    (no uncertainty required). DCP does not depend on frequency nor media.
    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based onthe signal
    characteristics.
    Axy,2; Bx,y,z; Cx,y,2z; VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
    data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
    media. VR is the maximum calibration range expressed in RMS voltage across the diode.
    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
    Transfer Standard for fs800MHz) and inside waveguide using analytical field distributions based on
    power measurements for f >800MHz. The same setups are used for assessment of the parameters
    applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
    These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
    The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
    that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
    allows extending the validity from+50MHz to+100MHz.
    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
    phantom exposed by a patch antenna.
    Sensor Offset: The sensor offset corresponds to the offset ofvirtual measurement center from the
    probe tip (on probe axis). No tolerance required.
    Connector Angle: The angle is assessed using the information gained by determining the NORMx
    (no uncertainty required).


Certificate No: Z18—60357                             Page 2 of 11




                                                                           Nn eeg e c n regenen   ce ++   20


         fi'T\fl; 1;Collnbomdonewith 3


        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com     Hittp/www.chinattl.en




                       Probe EX3DV4


                                             SN: 7464


                                   Calibrated: September 30, 2018

                               Calibrated for DASY/EASY Systems
                                       (Note: non—compatible with DASY2 system!)




Certificate No: Z18—60357                              Page 3 of11




                                                                               Nn epeg m ige m c m mgree ce > se cy <> >


         Cie
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com     Hittp//www.chinattl.cn


     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7464

Basic Calibration Parameters
                                    Sensor X                Sensor Y              Sensor Z              Unc (k=2)
 Norm(pV/(V/im)?) A                 0.46                    0.44                  0.46                  £10.0%
 DCP(mV)®                           100.7                   101.4                 99.4


Modulation Calibration Parameters
 UID         Communication                        A            B            6            D           VR           Unc®
             System Name                          dB           dB/puV                    dB          mV           (k=2)
 0           Cw                          X        0.0          0.0          1.0          0.00         169.3       £2.2%
                                         Y        0.0          0.0          1.0                       163.1
                                         Z        0.0          0.0          1.0                       167.0



     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E—field uncertainty inside TSL (see Page 5 and Page6).
® Numerical linearization parameter: uncertainty not required.
© Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z18—60357                               Page 4 of 11




                                                                                         No epeg rrgr ie n m regeree n en en n en y en c   +



Document Created: 2019-06-18 18:53:28
Document Modified: 2019-06-18 18:53:28

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC